short communications\(\def\hfill{\hskip 5em}\def\hfil{\hskip 3em}\def\eqno#1{\hfil {#1}}\)

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775

A setup for synchrotron-radiation-induced total reflection X-ray fluorescence and X-ray absorption near-edge structure recently commissioned at BESSY II BAMline

CROSSMARK_Color_square_no_text.svg

aChemistry Department, Washington State University, Pullman, WA 99164, USA, bBAM, Bundesanstalt für Materialforschung und -prüfung, 12489 Berlin, Germany, cChemistry and Chemistry Education Department, Europa-Universität Flensburg, 24943 Flensburg, Germany, dChemistry Department, University of Hamburg, 20146 Hamburg, Germany, and eAtominstitut, Vienna University of Technology, Vienna 1020, Austria
*Correspondence e-mail: ursula.fittschen@wsu.edu

Edited by P. A. Pianetta, SLAC National Accelerator Laboratory, USA (Received 1 October 2015; accepted 1 February 2016; online 23 March 2016)

An automatic sample changer chamber for total reflection X-ray fluorescence (TXRF) and X-ray absorption near-edge structure (XANES) analysis in TXRF geometry was successfully set up at the BAMline at BESSY II. TXRF and TXRF-XANES are valuable tools for elemental determination and speciation, especially where sample amounts are limited (<1 mg) and concentrations are low (ng ml−1 to µg ml−1). TXRF requires a well defined geometry regarding the reflecting surface of a sample carrier and the synchrotron beam. The newly installed chamber allows for reliable sample positioning, remote sample changing and evacuation of the fluorescence beam path. The chamber was successfully used showing accurate determination of elemental amounts in the certified reference material NIST water 1640. Low limits of detection of less than 100 fg absolute (10 pg ml−1) for Ni were found. TXRF-XANES on different Re species was applied. An unknown species of Re was found to be Re in the +7 oxidation state.

1. Introduction

1.1. TXRF and TXRF-XANES

Synchrotron-radiation-based total reflection X-ray fluorescence analysis (SR-TXRF) is an analytical technique allowing for quasi non-destructive, simultaneous multi-element analysis of minute amounts of a sample (<1 mg). Coupled with X-ray absorption near-edge structure spectroscopy (XANES), this technique now provides information about the electronic structure of the elements of interest. Early work on XAS in total reflection geometry was performed in the late 1980s and early 1990s (Heald et al., 1986[Heald, S. M., Tranquada, J. M. & Chen, H. J. (1986). J. Phys. 47(C-8), 293-298.]; Dartyge et al., 1986[Dartyge, E., Fontaine, A., Tourillon, G., Cortes, R. & Jucha, A. (1986). Phys. Lett. A, 113, 384-388.]; Asakura et al., 1993[Asakura, K., Shirai, M. & Iwasawa, Y. (1993). Catal. Lett. 20, 117-124.]; Kawai et al., 1993[Kawai, J., Hayakawa, S., Kitajima, Y., Suzuki, S., Maeda, K., Urai, T., Adachi, H., Takami, M. & Gohshi, Y. (1993). Proc. Jpn Acad. Ser. B, 69, 179-184.]). The advantages of using SR-TXRF and SR-TXRF/XANES lie in the fact that limits of absolute detection are in the low femtogram (fg µL−1) and picogram (pg µL−1) regions, respectively, whereas, using laboratory TXRF setups, typical lower limits of detection fall in the picogram range (Klockenkämper & von Bohlen, 2015[Klockenkamper, R. & von Bohlen, A. (2015). Total-Reflection X-ray Fluorescence Analysis and Related Methods, 2nd ed. New York: Wiley.]). TXRF-XANES requires tunable radiation by maintaining relatively high primary beam flux densities and is, therefore, usually solely pursued at synchrotron facilities.

Synchrotron-based TXRF has been proven very useful for different applications. One is the determination of elements in atmospheric aerosols to monitor atmospheric events (short time scale, 1 h sampling time), an application that requires the analysis of minute amounts of sample (sub-µg). Several groups have used this technique to study elemental concentration in aerosol particulate matter (Fittschen et al., 2008[Fittschen, U. E. A., Meirer, F., Streli, C., Wobrauschek, P., Thiele, J., Falkenberg, G. & Pepponi, G. (2008). Spectrochim Acta B, 63, 1489-1495.]; Osán et al., 2010[Osán, J., Meirer, F., Groma, V., Török, S., Ingerle, D., Streli, C. & Pepponi, G. (2010). Spectrochim Acta B, 65, 1008-1013.]).

Speciation of elements in p.p.m. concentration in these minute amounts of sample is challenging for conventional setups but achievable in TXRF-XANES geometry (Fittschen et al., 2008[Fittschen, U. E. A., Meirer, F., Streli, C., Wobrauschek, P., Thiele, J., Falkenberg, G. & Pepponi, G. (2008). Spectrochim Acta B, 63, 1489-1495.]; Meirer et al., 2007[Meirer, F., Pepponi, G., Streli, C., Wobrauschek, P., Mihucz, V. G., Záray, G., Czech, V., Broekaert, J. A. C., Fittschen, U. E. A. & Falkenberg, G. (2007). X-ray Spectrom. 36, 408-412.]). Meirer et al. report a limit of detection of 32 p.p.b. for speciation of As in the xylem sap of plants. The results of the study indicate that cucumber plants reduce AsV to AsIII (Meirer et al., 2007[Meirer, F., Pepponi, G., Streli, C., Wobrauschek, P., Mihucz, V. G., Záray, G., Czech, V., Broekaert, J. A. C., Fittschen, U. E. A. & Falkenberg, G. (2007). X-ray Spectrom. 36, 408-412.]).

Other potential applications are the speciation of trace concentrations of hazardous elements in pore water (water held within a soil or rock, in gaps between particles) or valuable elements from sediments. The oxidation state is critical for the mobility of elements such as Tc and actinides found in sediments of sites contaminated by the nuclear industry, or valuable elements such as Re which have a complex redox chemistry. Identifying their oxidation state under a certain environmental condition allows one to understand and predict the dynamic behaviour of the respective element. This will eventually advance remediation of contaminated sites and exploitation of valuable elements such as noble and rare-earth metals.

1.2. Advantage of a sample changer

TXRF spectrometry is sensitive to the deviation of the angle of incidence and the solid angle captured by the detector. However, fine tuning of the angle of incidence and assuring low angle deviation over the irradiated surface area is time-consuming and limits the output achievable within the time frame of normal user beam time. Accordingly, equipment that ensures optimal and constant measurement conditions increases output and accuracy of measured data. A few examples for beamlines more frequently used for TXRF analysis are: PTB at BESSY II, BAMline at BESSY II and beamline L at Doris III. Only beamline L was equipped with an automated sample chamber for SR-TXRF until the Doris synchrotron ring was shut down in 2012. This chamber was modified to fit the BAMline and figures-of-merit were determined.

2. Sample chamber and periphery

The general design of the sample chamber is described by Streli et al. (2006[Streli, C., Pepponi, G., Wobrauschek, P., Jokubonis, C., Falkenberg, G., Záray, G., Broekaert, J., Fittschen, U. & Peschel, B. (2006). Spectrochim Acta B, 61, 1129-1134.]). The setup allows users to record SR-TXRF and SR-TXRF-XANES spectra with high reproducibility on 30 mm-diameter quartz sample holders. The sample holder has been designed to allow synchrotron light to interact with the sample at a grazing angle of below 0.1° with the detector setup in the polarization plane of the incoming radiation. The setup consists of an Al chamber with a fitting for a vacuum line, two Kapton windows (8 µm) and a vacuum feedthrough fitting for an energy-dispersive XRF detector. The first Kapton window is the entrance for the primary beam into the chamber and the second Kapton window is the exit for both the reflected beam and the primary beam passing over the reflector without reflection. The core of the chamber is the sample positioning unit in front of the detector, consisting of three metal balls and a pressure spring sample holder that keeps the sample carriers in a defined plane relative to the primary beam and the detector. A Ta wedge on the side of the incoming beam ensures a minimum of scattering from the reflector rim. The chamber is designed for vacuum operation, an ideal feature for reducing remaining scatter from air. It also removes the Ar K line arising from the Ar content of ambient air and reduces absorption allowing for low detection limits in light-element detection.

The chamber was installed at the BAMline (Bundesanstalt für Materialforschung und -prüfung) of the Berlin electron storage ring (BESSY II). Details of the BAMline design are given by Riesemeier et al. (2005[Riesemeier, H., Ecker, K., Görner, W., Müller, B. R., Radtke, M. & Krumrey, M. (2005). X-ray Spectrom. 34, 160-163.]). This experimental station is a high-energy beamline for XRF measurements at BESSY II at the Helmholtz-Zentrum Berlin. At this beamline the primary radiation source is a 7 T wavelength shifter (Wiedemann, 2010[Wiedemann, H. (2010). Synchrotron Radiation. Berlin/Heidelberg: Springer.]; Borovikov et al., 1998[Borovikov, V. M., Craft, B., Fedurin, M. G., Jurba, V., Khlestov, V., Kulipanov, G. N., Li, O., Mezentsev, N. A., Saile, V. & Shkaruba, V. A. (1998). J. Synchrotron Rad. 5, 440-442.]; Görner et al., 2001[Görner, W., Hentschel, M. P., Müller, B. R., Riesemeier, H., Krumrey, M., Ulm, G., Diete, W., Klein, U. & Frahm, R. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 703-706.]).

The chamber was mounted on an XYZ stage and goniometer table (Huber, Rimsting, Germany). The angle between the reflector plane and the incoming beam was adjusted by the goniometer. A silicon drift detector (SDD) (XFlash 5010; Bruker AXS, Berlin, Germany) with a 30 mm2 active area was used to collect XRF spectra. Fig. 1[link] shows a sketch of the general setup with the chamber; Fig. 2[link] shows an image of the sample positioning unit, two of three steel balls forming the reference plane and the Ta wedge; Fig. 3[link] shows the setup at the BAMline.

[Figure 1]
Figure 1
Sketch of the sample chamber showing all major components and its positioning at the BAMline.
[Figure 2]
Figure 2
Inside the chamber showing the sample holder, the steel balls forming the reference plane as well as the Ta wedge.
[Figure 3]
Figure 3
View into the open chamber showing the sample changer.

3. Data acquisition and application

A double multilayer monochromator was used for the analysis of Ni samples and analysis of NIST water 1640. A beam size of 1.5 mm × 1.5 mm (width × height) was used and an acquisition time of 1000 s was chosen for the purpose.

For the Re XANES acquisition, the energy of the X-ray beam was tuned by a double-crystal Si (111) monochromator. A beam size of 4 mm × 4 mm was used and the acquisition time was 100 s per energy.

The software Athena from the IFEFFIT suite was used to process the XANES data (Ravel & Newville, 2005[Ravel, B. & Newville, M. (2005). J. Synchrotron Rad. 12, 537-541.]; Newville, 2001[Newville, M. (2001). J. Synchrotron Rad. 8, 322-324.]). All measurements were performed in air.

3.1. Limits of detection (LOD) for Ni

TXRF is a very common analytical tool in the Si-wafer industry. An established procedure in Si-wafer TXRF is the determination of the LOD for low concentrations of Ni contamination. In this respect, LOD for Ni were determined in this work. Specimens containing Ni in concentrations of 100 pg and 10 pg were prepared from pipetting 10 µL of diluted Ni single-element standard solutions (Merck Millipore, Darmstadt, Germany) and ultra-pure water (Omnitrace®, Merck Millipore, Darmstadt, Germany) on a 30 mm-diameter quartz reflector. The LOD were determined according to the following equation,

[{\rm{LOD}}= {{ 3\,Ci\,(N_{\rm{B}})^{1/2} }\over{ N_i }} ,\eqno(1)]

where Ci is the concentration of Ni, NB is the background count and Ni is the net count of Ni Kα.

For the 10 pg Ni specimens and 1000 s measurement time, a LOD value of 60 fg was found. This current value is around ten times more than our earlier findings (Streli et al., 2006[Streli, C., Pepponi, G., Wobrauschek, P., Jokubonis, C., Falkenberg, G., Záray, G., Broekaert, J., Fittschen, U. & Peschel, B. (2006). Spectrochim Acta B, 61, 1129-1134.]), but is explained by the fact that at the time of the experiments presented here the detector had a smaller active area and had not yet been incorporated into the beamline electronics, and the measurements were performed under air atmosphere. The LODs are considerably lower compared with a conventional laboratory setup. A comparison is given in Table 1[link]. As the detection limit is an extrapolated value based on statistical considerations, it is to be expected that the 10 pg sample is not detectable using the TXRF laboratory setup.

Table 1
LODs for Ni determined using the newly installed setup at the BAMline and a laboratory TXRF setup (calculated for 1000 s lifetime)

Amount of Ni (pg) LODs Sy-TXRF (pg) LOD conventional TXRF (pg) low-power TXRF
10 0.06 <detection limit
100 0.07 8
A Low Power Mo Tube TXRF Spectrometer with Electrically Cooled Silicon Drift Detector (Wobrauschek et al., 2011[Wobrauschek, P., Felling, C., Prost, J. & Streli, C. (2011). XRF Newsl. 22(2), 16.]).

3.2. Accuracy of water analysis

Trace elemental determination in minute amounts of aqueous samples (<1 ml) is a general application area for TXRF, e.g. analysis of pore water, Xylem sap from plants, etc.

In order to determine the accuracy of the analysis of small amounts of an aqueous sample, 10 µL samples of NIST water 1640 were prepared on a quartz reflector and analyzed for 100 s. The spectra were fitted using PyMCA (Solé et al., 2007[Solé, V., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). Spectrochim Acta B, 62, 63-68.]); Cu was used as internal standard and a fundamental parameter approach was used to determine the concentration of all detectable elements. Fig. 4[link] shows the spectrum acquired at the BAMline and fitted using PyMCA. Table 2[link] shows the detected and certified elements and their concentrations as certified and determined at the BAMline. All elements above the detection limits were found to deviate by at most <10% from the certified concentrations. However, the Ni, Cr and Ba concentrations deviated by 20–30%.

Table 2
Elemental concentrations certified for NIST water 1640 and determined at the BAMline with SR-TXRF

Element Line Certified amount (µg L−1) Determined amount (µg L−1) Deviation (%)
K Kα 994.00 912.44 8.20
Ca Kα 7045.00 6872.37 2.45
V Kα 12.99 12.77 1.66
Cr Kα 38.60 26.84 30.47
Mn Kα 121.50 118.91 2.13
Fe Kα 34.30 34.05 0.73
Co Kα 20.28 18.13 10.62
Ni Kα 27.40 32.69 19.30
Cu(IS) Kα 85.20 85.20 0.00
Zn Kα 53.20 56.56 6.31
Ba Lα 148.00 118.18 20.15
[Figure 4]
Figure 4
SR-TXRF spectrum of the NIST water 1640 specimen fitted using PyMCA, acquired at the BAMline.

3.3. Re L3-edge XANES

Elemental speciation of metal ions in low concentrations is maybe the most valuable and also the most challenging application of SR-TXRF; valuable because of its unique capabilities for trace element speciation offering p.p.b. LOD, while ensuring minimum interference with the sample and, therefore, reducing the risk of alteration, and challenging because self-absorption effects well known in fluorescence yield XANES increase for small angles as applied in TXRF. However, a number of publications addressed this problem in the past and made modelling and experimental solutions available to solve this problem (Lühl et al., 2012[Lühl, L., Mantouvalou, I., Malzer, W., Schaumann, I., Vogt, C., Hahn, O. & Kanngiesser, B. (2012). Anal. Chem. 84, 1907-1914.]; Meirer et al., 2008[Meirer, F., Pepponi, G., Streli, C., Wobrauschek, P., Kregsamer, P., Zoeger, N. & Falkenberg, G. (2008). Spectrochim Acta B, 63, 1496-1502.], 2009[Meirer, F., Pepponi, G., Streli, C., Wobrauschek, P. & Zoeger, N. (2009). J. Appl. Phys. 105, 074906.]; Menzel et al., 2013[Menzel, M., Schlifke, A., Falk, M., Janek, J., Fröba, M. & Fittschen, U. E. A. (2013). Spectrochim Acta B, 85, 62-70.]).

Here we focused on the speciation of Re. Re has a complex redox chemistry, is of interest as a catalyst and, therefore, stands as an example for a valuable element. Re is also a non-radioactive substitute for Tc, which is a critically hazardous element regarding nuclear contaminated sites. Fluorescence yield XANES usually requires spectra at the L3-edge using the Lα line. The allowed transition from the 2p state to the 5d state is dominant and leads to the distinct white line. For the 5d states located at the central atom, the energy of the ground and the excited states changes during oxidation, though the shift at the L3-edge is quite small. The height of the white line is a measure of the density of excited states and accordingly of the oxidation state as well. Tougerti et al. (2012[Tougerti, A., Cristol, S., Berrier, E., Briois, V., La Fontaine, C., Villain, F. & Joly, Y. (2012). Phys. Rev. B, 85, 125136.]) found a shift of ∼1 eV per formal oxidation state for Re oxides at the L3-edge.

Here, an unknown aqueous Re solution and two known Re compounds with defined oxidation states NaReO4 (ReVII) and ReI3 (ReIII) were prepared from suspensions to ensure a minimum of alteration of the specimen. The procedure was described by Fittschen et al. (2008[Fittschen, U. E. A., Meirer, F., Streli, C., Wobrauschek, P., Thiele, J., Falkenberg, G. & Pepponi, G. (2008). Spectrochim Acta B, 63, 1489-1495.]). The concentration of Re in the compounds was 5 ng and 6 ng, respectively, determined using a laboratory TXRF instrument (Picofox S2, Bruker, Berlin, Germany).

The TXRF-XANES study on the reference compounds showed an edge shift of 3 eV between NaReO4 and ReI3, which is shown in Fig. 5[link]. Compared with the results of Tougerti et al. (2012[Tougerti, A., Cristol, S., Berrier, E., Briois, V., La Fontaine, C., Villain, F. & Joly, Y. (2012). Phys. Rev. B, 85, 125136.]), this corresponds to a change in oxidation state from +7 to +4; however, ReI3 is formally in a +3 oxidation state. It is likely that a part of the ReI3 was oxidized by oxygen in ambient air to Re +7, which is a quite stable state. The difference in the heights of the white line of both compounds is also apparent. Following normalization to the post edge, the white line for ReI3 rises to 1.50 units above the post-edge line and to 2.20 units for NaReO4, which is explained by the higher density of excited states in Re +7 (see paragraph above).

[Figure 5]
Figure 5
XANES spectra of NaReO4 (ReVII), ReI3 (ReIII) and an unknown compound.

The speciation of 1 ng Re in an unknown Re complex revealed a good correlation to the XANES spectrum of NaReO4. The amount was determined prior to the TXRF-XANES measurements using a laboratory TXRF instrument (Picofox S2, Bruker, Berlin, Germany).

Fig. 5[link] shows the normalized XANES spectra for the three compounds. Fig. 6[link] shows the derivatives of the XANES spectra. The edge shifts of the unknown and NaReO4 were not significantly different. It can be assumed that both compounds have a formal oxidation state of +7. However, the small shift indicates a difference in the chemical species of the unknown. The height of the white line of the unknown was 2.50 units. This is different from the height found for NaReO4 and is caused by self-absorption effects due to differences in the total amount of Re in the respective specimen. The unknown with the lowest absolute mass of 1 ng Re shows a minimum in self-absorption effects.

[Figure 6]
Figure 6
Derivative of the edge regions of the XANES spectra shown in Fig. 5.[link]

4. Conclusion

The TXRF setup at the BAMline at BESSY II is now available to all users and is currently the only available synchrotron-based total reflection X-ray spectrometer of its kind. It is part of a growing TXRF community that has recently been reviewed by Klockenkämper & von Bohlen (2014[Klockenkämper, R. & von Bohlen, A. (2014). Spectrochim Acta B, 99, 133-137.]).

Acknowledgements

We acknowledge the Helmholtz-Zentrum Berlin for provision of synchrotron radiation beam time at the BAMline of BESSY II. CS and PW thankfully acknowledge the financial support from HZB. The research leading to these results has received funding from the European Community's Seventh Framework Programme (FP7/2007–2013) under grant agreement No. 312284. The authors thank Katharina Peikert and Dr Paul Benny for providing Re materials. The presented work has received funding from the Austrian Science Fund (FWF), project No. P23832-N20.

References

First citationAsakura, K., Shirai, M. & Iwasawa, Y. (1993). Catal. Lett. 20, 117–124.  CrossRef CAS Web of Science Google Scholar
First citationBorovikov, V. M., Craft, B., Fedurin, M. G., Jurba, V., Khlestov, V., Kulipanov, G. N., Li, O., Mezentsev, N. A., Saile, V. & Shkaruba, V. A. (1998). J. Synchrotron Rad. 5, 440–442.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDartyge, E., Fontaine, A., Tourillon, G., Cortes, R. & Jucha, A. (1986). Phys. Lett. A, 113, 384–388.  CrossRef Web of Science Google Scholar
First citationFittschen, U. E. A., Meirer, F., Streli, C., Wobrauschek, P., Thiele, J., Falkenberg, G. & Pepponi, G. (2008). Spectrochim Acta B, 63, 1489–1495.  Web of Science CrossRef Google Scholar
First citationGörner, W., Hentschel, M. P., Müller, B. R., Riesemeier, H., Krumrey, M., Ulm, G., Diete, W., Klein, U. & Frahm, R. (2001). Nucl. Instrum. Methods Phys. Res. A, 467468, 703–706.  Google Scholar
First citationHeald, S. M., Tranquada, J. M. & Chen, H. J. (1986). J. Phys. 47(C-8), 293–298.  Google Scholar
First citationKawai, J., Hayakawa, S., Kitajima, Y., Suzuki, S., Maeda, K., Urai, T., Adachi, H., Takami, M. & Gohshi, Y. (1993). Proc. Jpn Acad. Ser. B, 69, 179–184.  CrossRef CAS Web of Science Google Scholar
First citationKlockenkämper, R. & von Bohlen, A. (2014). Spectrochim Acta B, 99, 133–137.  Google Scholar
First citationKlockenkamper, R. & von Bohlen, A. (2015). Total-Reflection X-ray Fluorescence Analysis and Related Methods, 2nd ed. New York: Wiley.  Google Scholar
First citationLühl, L., Mantouvalou, I., Malzer, W., Schaumann, I., Vogt, C., Hahn, O. & Kanngiesser, B. (2012). Anal. Chem. 84, 1907–1914.  Web of Science PubMed Google Scholar
First citationMeirer, F., Pepponi, G., Streli, C., Wobrauschek, P., Kregsamer, P., Zoeger, N. & Falkenberg, G. (2008). Spectrochim Acta B, 63, 1496–1502.  Web of Science CrossRef Google Scholar
First citationMeirer, F., Pepponi, G., Streli, C., Wobrauschek, P., Mihucz, V. G., Záray, G., Czech, V., Broekaert, J. A. C., Fittschen, U. E. A. & Falkenberg, G. (2007). X-ray Spectrom. 36, 408–412.  Web of Science CrossRef CAS Google Scholar
First citationMeirer, F., Pepponi, G., Streli, C., Wobrauschek, P. & Zoeger, N. (2009). J. Appl. Phys. 105, 074906.  Web of Science CrossRef Google Scholar
First citationMenzel, M., Schlifke, A., Falk, M., Janek, J., Fröba, M. & Fittschen, U. E. A. (2013). Spectrochim Acta B, 85, 62–70.  Web of Science CrossRef CAS Google Scholar
First citationNewville, M. (2001). J. Synchrotron Rad. 8, 322–324.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationOsán, J., Meirer, F., Groma, V., Török, S., Ingerle, D., Streli, C. & Pepponi, G. (2010). Spectrochim Acta B, 65, 1008–1013.  Google Scholar
First citationRavel, B. & Newville, M. (2005). J. Synchrotron Rad. 12, 537–541.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationRiesemeier, H., Ecker, K., Görner, W., Müller, B. R., Radtke, M. & Krumrey, M. (2005). X-ray Spectrom. 34, 160–163.  Web of Science CrossRef CAS Google Scholar
First citationSolé, V., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). Spectrochim Acta B, 62, 63–68.  Google Scholar
First citationStreli, C., Pepponi, G., Wobrauschek, P., Jokubonis, C., Falkenberg, G., Záray, G., Broekaert, J., Fittschen, U. & Peschel, B. (2006). Spectrochim Acta B, 61, 1129–1134.  Web of Science CrossRef Google Scholar
First citationTougerti, A., Cristol, S., Berrier, E., Briois, V., La Fontaine, C., Villain, F. & Joly, Y. (2012). Phys. Rev. B, 85, 125136.  Web of Science CrossRef Google Scholar
First citationWiedemann, H. (2010). Synchrotron Radiation. Berlin/Heidelberg: Springer.  Google Scholar
First citationWobrauschek, P., Felling, C., Prost, J. & Streli, C. (2011). XRF Newsl. 22(2), 16.  Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds