|
Figure 2
(a) Linear and circular polarization fraction as a function of X-ray phase retarder angular offset, where Δθ is the deviation from the ideal Bragg condition. The linear (red curve) and circular (blue curve) polarization fraction are calculated for a monochromator bandwidth of ΔE/E = 1.07 × 10−4. Black circles show experimental values for the linear polarization measured with a Ge(333) analyzer crystal at 8 keV. (b) Air-scattered intensity measured before and after a 12 h time interval. Changes correspond to a slow drift in the XPR alignment of 0.001° in 12 h. |
Open
access
