X-ray powder diffraction at the XRD1 beamline at LNLS

CROSSMARK_Color_square_no_text.svg

aLaboratório Nacional de Luz Síncrotron, CNPEM, Campinas, SP 13083-970, Brazil, bICT, UNIFESP, São José dos Campos, SP 12231-280, Brazil, and cUnidade Acadêmica de Física, UFCG, Campina Grande, PB 58429-900, Brazil
*Correspondence e-mail: alexandre.carvalho@lnls.br

Edited by I. Schlichting, Max Planck Institute for Medical Research, Germany (Received 31 March 2016; accepted 5 August 2016; online 6 October 2016)

Various upgrades have been completed at the XRD1 beamline at the Brazilian synchrotron light source (LNLS). The upgrades are comprehensive, with changes to both hardware and software, now allowing users of the beamline to conduct X-ray powder diffraction experiments with faster data acquisition times and improved quality. The main beamline parameters and the results obtained for different standards are presented, showing the beamline ability of performing high-quality experiments in transmission geometry. XRD1 operates in the 5.5–14 keV range and has a photon flux of 7.8 × 109 photons s−1 (with 100 mA) at 12 keV, which is one of the typical working energies. At 8 keV (the other typical working energy) the photon flux at the sample position is 3.4 × 1010 photons s−1 and the energy resolution ΔE/E = 3 × 10−4.

1. Introduction

The XRD1 beamline at the Brazilian Synchrotron Light Laboratory (LNLS) is a dedicated X-ray diffraction beamline (Cusatis et al., 1998[Cusatis, C., Kobayashi Franco, M., Kakuno, E., Giles, C., Morelhão, S., Mello, V. & Mazzaro, I. (1998). J. Synchrotron Rad. 5, 491-493.]) in operation since 1998. The beamline is installed at the D12B bending magnet (1.67 T) and operates in the 5.5–14 keV range.

This beamline was upgraded recently. The mechanical stress on the optical elements and floor instabilities began causing beam stability problems as well as affecting the repeatability of the monochromator positions. Furthermore, user requirements for in situ powder diffraction experiments and faster detection speeds meant that the X-ray Powder Diffraction beamline (XPD) was becoming increasingly congested. The XRD1 diffractometer did not allow for easy installation of cell reactors and fast detection systems based on a set of linear detectors. Furthermore, the heavy detection system, which included a metallic casing, also gave problems. These limitations were the driving force for the beamline upgrade that began in 2011. Increasing beam instabilities led the LNLS to undertake a comprehensive restructuring of the optical hutch, including the construction of an independent concrete slab below the mirror and monochromator. Furthermore, the mechanics inside the monochromator were completely changed as well as the replacement of the two monochromator crystals (Canova et al., 2014[Canova, H., Fontoura, A., Neuenschwander, R. T., Diaz, B. & Rodella, C. B. (2014). J. Phys. Conf. Ser. 493, 012004.]).

Commissioning tests of the optical elements showed that XRD1 gave a similar performance of photon flux at the sample position, energy resolution, motor repeatability and beam stability as the XPD beamline but with the advantage of a more spacious experimental hutch. Thus, the renovation continued to the experimental hutch with the installation of a three-circle heavy-duty Newport diffractometer in 2013. The new diffractometer allows for faster measurements when combined with the MYTHEN 24K system from Dectris® (MYTHEN, 2014[MYTHEN (2014). MYTHEN Microstrip X-ray Detectors, https://www.dectris.com/tl_files/root/products/mythen_family/MYTHEN%20Brochures/MYTHEN_8Seiter_Jun2014_final_web.pdf.]), and a robotic arm for rapid automated sample changing. Moreover, in 2014, a second experimental hutch was opened after the first hutch of XRD1. Both use the same X-ray beam and operate alternately. The second end-station (named XTMS) has a heavy-duty Huber diffractometer installed inside a thermo-mechanical simulator (Gleeble® Synchrotron system). The main goal of this facility is to perform structural studies on bulk metallic and non-metallic samples during thermic and/or mechanical tests (XTMS, 2015[XTMS (2015). XTMS Experimental Station, https://lnnano.cnpem.br/laboratories/cpm/facilities/xtms/.]). The parameters of the XTMS end-station, which is not used for X-ray powder diffraction, will not be discussed in the present work.

The commissioning of the first end-station of the XRD1 beamline is now complete and optimized for structural characterization by X-ray diffraction experiments in Debye–Scherrer geometry, i.e. transmission mode of low-absorption materials. Samples are placed in glass or quartz capillaries for X-ray diffraction analyzes. Due to the relatively low energy of the beamline, sample thickness, particle size and sample packing have to be determined to optimize X-ray transmission. In general, low-absorption materials such as pharmaceuticals, ceramics, biomass, clays, catalysts and soils can be studied at XRD1. Carvalho et al. (2016[Carvalho, A. M. G., Nunes, R. S. & Coelho, A. A. (2016). arXiv: 1608.00565 [physics. ins-det].]) showed that it is also possible to obtain high-quality data for high-absorption materials using capillaries in the same geometry, despite the relatively low energies of the beamline. Besides the fast-automated experiments at room temperature, it is also possible to perform measurements at temperatures ranging from 100 K and 1070 K using two different devices. In addition, a capillary cell reactor has been developed to allow controlled flow of gases and/or vapor through the samples and its commissioning is in progress.

2. Beamline overview

2.1. Beamline optics

The optical layout of the XRD1 beamline is shown schematically in Fig. 1[link]. The source is a 1.67 T bending magnet of the LNLS ring operating at 1.37 GeV (Craievich & Rodrigues, 1997[Craievich, A. F. & Rodrigues, A. R. D. (1997). Braz. J. Phys. 27, 417-424.]; Rodrigues et al., 1998[Rodrigues, A. R. D., Craievich, A. F. & Gonçalves da Silva, C. E. T. (1998). J. Synchrotron Rad. 5, 1157-1161.]), with a typical initial average current around 250 mA and an 11 h lifetime (February 2016).

[Figure 1]
Figure 1
Schematic diagram of XRD1 beamline (excluding the second end-station). The diagram shows the bending magnet; white-beam slits (WBS); the vertical focusing mirror (VFM); the double-crystal monochromator (DCM); slits (S1 and S2); and the sample position on the heavy-duty diffractometer.

From the source, the beam passes through a set of home-made automated slits (identified as WBS in Fig. 1[link]). The slits are independent and water-cooled. This set of slits has the main function of limiting the horizontal and vertical dimensions and divergence of the incident beam. The horizontal divergence of the beam is regulated by the horizontal slits, which are typically separated by 27.4 mm, resulting in a divergence of 6 mrad.

After passing through the white-beam slits, the beam reaches a ULE (ultra-low expansion) Rh-coated glass mirror, located at about 5.4 m from the source, inside the optical hutch. The mirror is used to focus/collimate the white beam vertically, as well as filter off the high-energy photons. The mirror was fabricated with a curvature radius of 2.8 km. The cut-off energy of the photons is ∼15 keV and it is defined by the rhodium coating as well as the incident angle of the white beam on the mirror, which is about 3.7 mrad. The mirror dimensions are 750 mm × 80 mm. The mirror is accommodated in a vacuum chamber produced in-house, under a pressure of ∼10−7 Pa. Three independent motors (with a Heidenhain encoder) allow the mirror position to be adjusted (height and vertical and horizontal angles). Another motor (with a potentiometer-like encoder) is used to bend the mirror. The position of this motor is read by a 12-bit LOCON board. Home-made software (SPEGULO) controls the mirror movements and positions.

Monochromatization of the X-ray beam is performed using a double-crystal Si (111) monochromator developed in-house. The first crystal is flat (80 mm × 80 mm × 3 mm) and is water-cooled. The second crystal is bent for sagittal focusing. The double-crystal monochromator is mounted over a disc-shaped goniometer under high vacuum (∼10−5 Pa), located inside the optical hutch 6.7 m from the source. With this system, the incident angle of the white beam over the first crystal (θm) is determined with precision and provides good stability.

There are two other sets of slits (identified as S1 and S2 in Fig. 1[link]) inside the first-end station that are used to define the beam size at the sample position. The first set, located 12.5 m from the source, are similar to the WBS, but not water-cooled. The second set, closer to the diffractometer (15.5 m from the source), are a compact set of automated slits (from ADC company).

2.2. Diffractometer, detection system and accessories

The XRD1 beamline is equipped with a three-circle N3050-P1 diffractometer from Newport® (Fig. 2[link]), which is installed 18.2 m from the source. Its goniometer is based on a one-axis object-positioning θ-circle and in line, centered on the same horizontal axis as the second- and third-circle detector holder. The 2θ- and δ-circles are designed to support a variety of heavy accessories such as detectors. Loadings of more than 100 kg for the 2θ-circle and 60 kg for the δ-circle are possible. The θ-circle is capable of supporting a 35 kg load, up to 20 kg with the open frame XYZ sample positioner. According to the manufacturer, the three circles have the same angular resolution (0.0001°), accuracy (0.001°) and zero position error (0.001°).

[Figure 2]
Figure 2
Photograph of the heavy-duty diffractometer at XRD1 beamline. (1) δ-circle; (2) 2θ-circle; (3) θ-circle; (4) XYZ sample positioner; (5) housing of the MYTHEN 24K system; (6) Gas Blower system.

Although it is possible to use the θ- and 2θ-circles to perform X-ray powder diffraction in Bragg–Brentano geometry, in practice we have only been using the δ-circle, which supports the detectors for fast measurements in Debye–Scherrer geometry. Rigidly fixed onto the δ-circle is a housing for the MYTHEN 24K system (Dectris), from Dectris®. Inside the housing, there is a continuous flow of helium in order to reduce the absorption and scattering of the diffracted beam. The MYTHEN 24K system consists of 24 linear MYTHEN detectors installed 760 mm from the sample position, leading to a fixed angular resolution of 0.0037°. These detectors are 320 µm thick and recommended to operate between 5 and 12.5 keV, according to the manufacturer. A complete description of the MYTHEN detectors was reported by Bergamaschi et al. (2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]) and the instrumental profile of a MYTHEN detector in Debye–Scherrer geometry is discussed by Gozzo et al. (2010[Gozzo, F., Cervellino, A., Leoni, M., Scardi, P., Bergamaschi, A. & Schmitt, B. (2010). Z. Kristallogr. 225, 616-624.]). These kinds of position-sensitive detectors have been used in different synchrotron facilities around the world, such as the Swiss Light Source (Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]), Diamond Light Source (Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]) and Beijing Synchrotron Radiation Facility (Du et al., 2016[Du, R., Cai, Q., Chen, Z., Gong, Y., Liu, H. & Wu, Z. (2016). Instrum. Sci. Technol. 44, 1-11.]).

The angular calibration of the MYTHEN 24K system was performed using a well determined energy at 12 keV. In order to convert from channel number to 2θ angle, a series of diffraction patterns of the silicon standard were acquired while shifting the detector by 0.1°. Different reflections of the silicon were fitted with Gaussians to determine their positions in terms of channel numbers. With a linear regression of the data, we obtain the number of channels per degree (ci) and the central channel (Ccenter i) for each module i. The correct distance Ri, in millimeters, of the module i from the diffractometer center was calculated using the relation

[{R^{\,i}} = {p \over {\tan\left({{1/{{c^{\,i}}}}}\right)}}, \eqno(1)]

where p is the pixel (channel) size, 0.05 mm, and ci is the number of channels per degree in each module i. The corresponding angular value for each channel j inside each module i was obtained using the relation

[\theta_j^{\,\prime\,i} = \arctan\left[ {{ p\left(C_j^{\,i}-C_{\rm{center}}^{\,i}\right) }\over{ R^{\,i} }} \right], \eqno(2)]

where Cj i is the corresponding number of the channel j of the module i; Ccenter i is the central channel of each module. Finally, the angular offset of each module, [\theta_0^{\,i}], was obtained through a diffraction pattern of the LaB6 NIST standard and the angular position refinement of the observed reflections in comparison with the calculated values. Therefore, the refined angular value for each channel j inside each module i is given by

[\theta_j^{\,i} = \theta_j^{\,\prime\,i} + \theta_0^{\,i}. \eqno(3)]

The flat-field correction for the detectors was performed by the manufacturer.

At the XRD1 beamline, it is possible to perform X-ray powder diffraction experiments as a function of temperature. We have been using two different devices for variable temperature measurements: from 100 K up to 460 K, using the Cryojet5® (OXFORD Instruments); from room temperature to 1070 K, using the Gas Blower GSB 1300 (FMB Oxford) (see Fig. 2[link]). According to the manufacturers, the working temperature ranges are larger. Nevertheless, we have limited these ranges after appropriate calibrations were conducted.

In order to increase the efficiency of the beamline during measurements, we have installed a robotic arm (Yaskawa-Motoman) to change samples quickly and automatically. This robotic arm is also used to hold both Cryojet5® and Gas Blower systems when performing measurements as a function of temperature.

2.3. Beamline parameters and performance

The photon flux at the sample position was measured at different X-ray wavelengths (Fig. 3[link]) using a silicon photodiode (model AXUV-20HE1). The current from the photodiode was measured using a picoammeter (Keithley 6485). The maximum photon flux is about 3.8 × 1010 photons s−1 at 1.771 Å (7 keV), at 100 mA. Comparing with the results obtained in 2013, reported by Canova et al. (2014[Canova, H., Fontoura, A., Neuenschwander, R. T., Diaz, B. & Rodella, C. B. (2014). J. Phys. Conf. Ser. 493, 012004.]), we observe that the photon flux has been increased, mainly at higher wavelengths. Generally, the wavelength is kept at 1.033 Å (12 keV), unless it is necessary to perform resonant X-ray diffraction experiments or to avoid fluorescence from specific atoms. At this wavelength, the photon flux is about 7.8 × 109 photons s−1, at 100 mA. The beam is focused on the sample position with a size of 2.0 mm (horizontal) and between 0.6 and 0.7 (vertical), depending on the energy. These and other information are listed in Table 1[link].

Table 1
XRD1 information

Beamline name XRD1
Source type Bending magnet
Monochromator Si(111) double-crystal
Mirror Rh-coated
Energy range 5.5–14 keV
Photon flux (photons s−1 @ 100 mA) 7.8 × 109 photons s−1 at 12 keV
3.4 × 1010 photons s−1 at 8 keV
Beam size (focused) 2.0 mm (horizontal)
  0.6–0.7 (vertical)
Beam divergence (at the sample position) 116 µrad (vertical) (unbent mirror)
Resolution (ΔE/E) 3 × 10−4 at 8 keV
Diffractometer Three-circle N3050-P1 diffractometer from Newport®
Detector MYTHEN 24K from Dectris®
Temperature range (at the sample) 100–1070 K
[Figure 3]
Figure 3
Photon flux at the sample position as a function of the X-ray wavelength. The data obtained in 2013 were reported by Canova et al. (2014[Canova, H., Fontoura, A., Neuenschwander, R. T., Diaz, B. & Rodella, C. B. (2014). J. Phys. Conf. Ser. 493, 012004.]).

The vertical size (FWHM) of the X-ray beam was determined by translating a silicon single-crystal across the beam at the sample position. The section of the beam not blocked by the crystal was measured by an ionization chamber. We observe that the vertical size of the beam has a minimum value at 15% curvature of the mirror (Fig. 4[link]).

[Figure 4]
Figure 4
Wavelength resolution and vertical size of the beam at the sample position as functions of the curvature of the mirror [0% and 100% correspond to minimum (3.5 km) and maximum (1.4 km) allowed curvature, respectively].

In order to determine the wavelength resolution of the X-ray beam with λ = 1.54877 (4) Å, rocking curves of the (111) and (333) reflections of a silicon single-crystal were recorded for different curvatures of the mirror. The wavelength resolution is evaluated using the following relation (Ferreira et al., 2006[Ferreira, F. F., Granado, E., Carvalho Jr, W., Kycia, S. W., Bruno, D. & Droppa Jr, R. (2006). J. Synchrotron Rad. 13, 46-53.]),

[{{\Delta\lambda}\over{\lambda}} = {{\Delta{E}}\over{E}} = {\left\{{{\left({w}_{333}^{2}-{w}_{111}^{2}\right)-\left({w}_{{\rm{D}}333}^{2}-{w}_{{\rm{D}}111}^{2}\right)}\over{{\left[\tan\left({\theta}_{\rm{m}}\right)-\tan\left({\theta}_{2}\right)\right]}^{2}-{\left[\tan\left({\theta}_{\rm{m}}\right)-\tan\left({\theta}_{1}\right)\right]}^{2}}}\right\}}^{1/2}\!\!, \eqno(4)]

where w111 and w333 are the widths of the rocking curves; wD111 and wD333 are the theoretical Darwin widths of the (111) and (333) reflections, respectively, of the silicon crystal used; θ1 and θ2 are the angles of the (111) and (333) reflections, respectively; θm is the monochromator angle. At 8 keV, the theoretical Darwin widths for silicon are wD111 = 6.79 arcsec and wD333 = 2.03 arcsec. In Fig. 4[link], we observe that the variation of the wavelength resolution with the curvature of the mirror is almost linear. Since the best resolution is achieved with the smallest curvature of the mirror, we keep this condition during experiments.

X-ray powder diffraction measurements were conducted on NIST standard reference materials (SRM): Si (SRM 640d), Al­2O3 (SRM 676a) and LaB6 (SRM 660b). The LaB6 standard was diluted in corn starch (LaB6 34% wt and corn starch 66% wt) in order to reduce the X-ray absorption. The SRMs were placed inside 0.3 mm-diameter borosilicate capillaries, which are the thinnest capillaries used at XRD1 beamline. The capillaries are fixed in ferromagnetic stainless steel holders. Each capillary holder is fitted with a magnetic tip attached to the three-circle diffractometer. This magnetic tip is able to rotate, spinning the sample (∼300 r.p.m.) during measurements. The measurements were conducted using the standard operating conditions (parameters) of the beamline: minimum curvature of the mirror; beam size of ∼2.0 mm × 0.7 mm at the sample position; MYTHEN 24K. Two different wavelengths were used, 1.54877 (4) Å (8 keV) and 1.03324 (3) Å (12 keV), keeping the temperature at 296 (1) K. Representative Bragg peaks of each diffraction pattern were fitted with a pseudo-Voigt function, using the software PyMca (Version 5.0.3). The linewidths (FWHM) of the peaks of each standard sample as a function of 2θ (Fig. 5[link]) were fitted with the Caglioti function (Caglioti et al., 1958[Caglioti, G., Paoletti, A. & Ricci, F. P. (1958). Nucl. Instrum. 3, 223-228.]),

[{\rm{FWHM}} = \left(U\tan^2\theta+V\tan\theta+W\right)^{1/2}. \eqno(5)]

The parameters U, V and W used to obtain the fits shown in Fig. 5[link] are listed in Table 2[link]. The largest contribution for the linewidths of the standard Si comes from the Gaussian part of the pseudo-Voigt function. Since this standard is claimed to have no microstrain (NIST, 2010[NIST (2010). SRM 640d; Silicon Powder Line Position and Line Shape Standard for Powder Diffraction. National Institute of Standards and Technology, Gaithersburg, MD, USA.]), its linewidths may be taken as good estimates of the instrumental resolution (including the contribution of the 0.3 mm-diameter capillary).

Table 2
Agreement R-factors and wavelength values from the structural refinements of NIST standards Si and Al2O3 measured at 12.0 keV

U, V and W parameters used to fit the experimental linewidths of these standards (Fig. 5[link]) are also listed.

        λ = 1.54877 Å λ = 1.03324 Å
Standards Rexp Rwp λ (Å) U V W U V W
Si (SRM 640d) 4.82 10.32 1.033303 (1) 0.001983 0.000182 0.000759 0.002774 0.000002 0.000415
Al­2O3 (SRM 676a) 3.08 6.75 1.033167 (2) 0.002906 0.001209 0.002886 0.004289 0.000548 0.001102
[Figure 5]
Figure 5
Bragg peak widths (full width at half-maximum in 2θ) as a function of 2θ for the standards Si and Al2O3. The lines are fits for the experimental FWHM values.

The X-ray powder diffraction patterns of the standards obtained at 12.0 keV were analyzed using the Rietveld method (Rietveld, 1969[Rietveld, H. M. (1969). J. Appl. Cryst. 2, 65-71.]) and TOPAS (Version 4.1) software. The agreement R-factors for Si and Al2O3 obtained from the Rietveld refinements are listed in Table 2[link]. The measured and calculated X-ray diffraction profiles for the NIST standard Al2O3 are shown in Fig. 6[link]. The inset shows five consecutive reflections with more detail: [(\bar3\,4\,2)], [(\bar1\,3\,8)], (0 2 10), [(0\,0\,\bar12)] and [(\bar1\,4\,4)] at 2θ ≃ 54.62°, 54.70°, 56.07°, 56.99° and 57.24°, respectively. The differences between the calculated and observed profiles for the three standard samples are relatively small (the fittings are very satisfactory), as exemplified for Al2O3 (Fig. 6[link]). The refinement of the LaB6 standard diluted in corn starch resulted in the agreement R-factors Rexp = 4.47 and Rwp = 9.09.

[Figure 6]
Figure 6
Observed and calculated X-ray powder diffraction intensities for Al2O3 reference powder (NIST SRM 676a). Acquisition time: 140 s.

The influence of the capillary diameter on the linewidths can be observed in Fig. 7(a)[link]. The NIST standard Si (SRM 640d) was measured inside capillaries with three different diameters: 0.3, 0.5 and 0.7 mm. These measurements were executed off the optimal conditions of resolution. It is easy to notice that the linewidth becomes larger with the size of the capillaries in a non-linear way. The widths of the observed peaks are dominated by the contribution of large capillaries. In Fig. 7(b)[link] we show the experimental data and the refinement results for two silicon reflections at high angles. The data were obtained using a 0.3 mm-diameter capillary. Comparing with the results obtained on I11 beamline at Diamond Light Source (Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]), we noticed that, for a similar angle range, the linewidths from our results are about 54% larger. Nevertheless, at low angles, the linewidths are small enough to solve the great majority of the scientific cases related to X-ray powder diffraction submitted to LNLS nowadays.

[Figure 7]
Figure 7
(a) Reflection (311) of NIST standard Si (SRM 640d) analyzed inside capillaries with three different diameters (0.3, 0.5 and 0.7 mm). (b) Reflections (642) and (553) of Si obtained inside a 0.3 mm-diameter capillary.

3. Facility access

Access to the Brazilian Synchrotron Light Laboratory is merit-based. Researchers submit scientific proposals to XRD1 beamline once a semester. All proposals are independently peer-reviewed and scored by an external scientific committee. The execution of the accepted proposals is always (and only) in the following semester. In the near future, it will be possible to conduct experiments at XRD1 remotely. Further information can be requested from the beamline coordinator by email (alexandre.carvalho@lnls.br).

4. Conclusions

The commissioning of the first end-station of XRD1 beamline is now complete and we are able to conduct high-quality X-ray powder diffraction experiments. This is optimized for experiments with low-absorption materials (such as medicines) in powder form. The instrumental resolution is sufficient to solve the great majority of the scientific cases related to X-ray powder diffraction submitted to LNLS nowadays. The devices for variable-temperature measurements are operating well. For low-absorption materials, we obtain good quality data using either dynamic or isothermal processes. For superior results with high-absorption materials, we suggest diluting them before the measurements. A capillary reactor cell has been developed and its commissioning is in progress. This and other developments and improvements will be transferred to the new X-ray powder diffraction beamline (Paineira) at the new fourth-generation Brazilian synchrotron (Sirius).

Acknowledgements

We thank all the technicians and scientists that worked to build and improve the XRD1 beamline. We also acknowledge LNLS and the financial support from FAPESP and CNPq.

References

First citationBergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653–668.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationCaglioti, G., Paoletti, A. & Ricci, F. P. (1958). Nucl. Instrum. 3, 223–228.  CrossRef CAS Web of Science Google Scholar
First citationCanova, H., Fontoura, A., Neuenschwander, R. T., Diaz, B. & Rodella, C. B. (2014). J. Phys. Conf. Ser. 493, 012004.  CrossRef Google Scholar
First citationCarvalho, A. M. G., Nunes, R. S. & Coelho, A. A. (2016). arXiv: 1608.00565 [physics. ins-det].  Google Scholar
First citationCraievich, A. F. & Rodrigues, A. R. D. (1997). Braz. J. Phys. 27, 417–424.  CrossRef CAS Google Scholar
First citationCusatis, C., Kobayashi Franco, M., Kakuno, E., Giles, C., Morelhão, S., Mello, V. & Mazzaro, I. (1998). J. Synchrotron Rad. 5, 491–493.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDu, R., Cai, Q., Chen, Z., Gong, Y., Liu, H. & Wu, Z. (2016). Instrum. Sci. Technol. 44, 1–11.  Web of Science CrossRef CAS Google Scholar
First citationFerreira, F. F., Granado, E., Carvalho Jr, W., Kycia, S. W., Bruno, D. & Droppa Jr, R. (2006). J. Synchrotron Rad. 13, 46–53.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGozzo, F., Cervellino, A., Leoni, M., Scardi, P., Bergamaschi, A. & Schmitt, B. (2010). Z. Kristallogr. 225, 616–624.  Web of Science CrossRef CAS Google Scholar
First citationMYTHEN (2014). MYTHEN Microstrip X-ray Detectors, https://www.dectris.com/tl_files/root/products/mythen_family/MYTHEN%20Brochures/MYTHEN_8Seiter_Jun2014_final_web.pdf.  Google Scholar
First citationNIST (2010). SRM 640d; Silicon Powder Line Position and Line Shape Standard for Powder Diffraction. National Institute of Standards and Technology, Gaithersburg, MD, USA.  Google Scholar
First citationRietveld, H. M. (1969). J. Appl. Cryst. 2, 65–71.  CrossRef CAS IUCr Journals Web of Science Google Scholar
First citationRodrigues, A. R. D., Craievich, A. F. & Gonçalves da Silva, C. E. T. (1998). J. Synchrotron Rad. 5, 1157–1161.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationThompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637–648.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationXTMS (2015). XTMS Experimental Station, https://lnnano.cnpem.br/laboratories/cpm/facilities/xtms/Google Scholar