X-ray analyzer-based phase-contrast imaging is combined with computed lamino­graphy for imaging regions of interest in laterally extended flat specimens of weak absorption contrast. The optics discussed here consist of an asymmetrically cut collimator crystal and a symmetrically cut analyzer crystal arranged in a nondispersive (+, −) diffraction geometry. A generalized algorithm is given for calculating multi-contrast (absorption, refraction and phase contrast) images of a sample. Basic formulae are also presented for lamino­graphic reconstruction. The feasibility of the method discussed was verified at the vertical wiggler beamline BL-14B of the Photon Factory. At a wavelength of 0.0733 nm, phase-contrast sectional images of plastic beads were successfully obtained. Owing to strong circular artifacts caused by a sample holder, the field of view was limited to about 6 mm in diameter.

1. Introduction

When an X-ray beam passes through an object, its propagation direction is slightly refracted owing to the inhomogeneous electron density in the object. In X-ray analyzer-based imaging, this small refraction is detected by an analyzer crystal for nondestructive observation of inner structures of the object (Davis et al., 1995[Davis, T. J., Gao, D., Gureyev, T. E., Stevenson, A. W. & Wilkins, S. W. (1995). Nature (London), 373, 595-598.]; Chapman et al., 1997[Chapman, D., Thomlinson, W., Johnston, R. E., Washburn, D., Pisano, E., Gmür, N., Zhong, Z., Menk, R., Arfelli, F. & Sayers, D. (1997). Phys. Med. Biol. 42, 2015-2025.]). As the refraction contrast is mathematically equivalent to the differential phase contrast, X-ray analyzer-based imaging has much higher sensitivity to low-Z elements compared with conventional imaging techniques based on absorption contrast. On account of this advantage, X-ray analyzer-based phase-contrast imaging is widely used for observing various specimens, such as biological soft tissues and soft materials, at many synchrotron radiation facilities.

In X-ray imaging, volume reconstruction is usually carried out using computed tomography (CT) (Hounsfield, 1973[Hounsfield, G. N. (1973). Br. J. Radiol. 46, 1016-1022.]). There are, however, several problems with the standard CT methods. The first problem is that the sample size must be smaller than the field of view (FOV) of the detector and beam size, which is generally a few centimeters when using synchrotron radiation (SR). This limitation has hindered the application of SR-CT to large samples. The second problem is that the thickness of the sample must be small enough for measuring the intensity of transmitted X-rays with sufficient accuracy. To mitigate these problems, synchrotron radiation computed lamino­graphy (SR-CL) was developed (Helfen et al., 2005[Helfen, L., Baumbach, T., Mikulík, P., Kiel, D., Pernot, P., Cloetens, P. & Baruchel, J. (2005). Appl. Phys. Lett. 86, 071915.], 2011[Helfen, L., Myagotin, A., Mikulík, P., Pernot, P., Voropaev, A., Elyyan, M., Di Michiel, M., Baruchel, J. & Baumbach, T. (2011). Rev. Sci. Instrum. 82, 063702.]) by adopting the principles of tomosynthesis (Dobbins & Godfrey, 2003[Dobbins, J. T. III & Godfrey, D. J. (2003). Phys. Med. Biol. 48, R65-R106.]). Whereas the standard CT methods are characterized by tomographic rotation around an axis perpendicular to the beam path, SR-CL is characterized by the more general geometry of an inclined rotation axis. On account of this generalized geometry, SR-CL is applicable to a wider variety of samples compared with SR-CT. For example, SR-CL is especially effective for observing regions of interest (ROIs) in laterally extended samples such as flip-chip bonded devices (Helfen et al., 2011[Helfen, L., Myagotin, A., Mikulík, P., Pernot, P., Voropaev, A., Elyyan, M., Di Michiel, M., Baruchel, J. & Baumbach, T. (2011). Rev. Sci. Instrum. 82, 063702.]), leaves (Verboven et al., 2015[Verboven, P., Herremans, E., Helfen, L., Ho, Q. T., Abera, M., Baumbach, T., Wevers, M. & Nicolaï, B. M. (2015). Plant J. 81, 169-182.]) and thin organic objects (Krug et al., 2008[Krug, K., Porra, L., Coan, P., Wallert, A., Dik, J., Coerdt, A., Bravin, A., Elyyan, M., Reischig, P., Helfen, L. & Baumbach, T. (2008). J. Synchrotron Rad. 15, 55-61.]).

Since its development, SR-CL has been applied not only to microscopy (Watanabe et al., 2009[Watanabe, N., Hoshino, M., Yamamoto, K., Aoki, S., Takeuchi, A. & Suzuki, Y. (2009). AIP Conf. Ser. 186, 012022.]; Hoshino et al., 2011[Hoshino, M., Uesugi, K., Takeuchi, A., Suzuki, Y. & Yagi, N. (2011). Rev. Sci. Instrum. 82, 073706.]; Xu et al., 2012[Xu, F., Helfen, L., Suhonen, H., Elgrabli, D., Bayat, S., Reischig, P., Baumbach, T. & Cloetens, P. (2012). PLoS One, 7, e50124.]) but also to phase-contrast imaging based on propagation (Helfen et al., 2009[Helfen, L., Baumbach, T., Cloetens, P. & Baruchel, J. (2009). Appl. Phys. Lett. 94, 104103.]) and grating interferometry (Harasse et al., 2010[Harasse, S., Hirayama, N., Yashiro, W. & Momose, A. (2010). Proc. SPIE, 7804, 780411.]; Altapova et al., 2012[Altapova, V., Helfen, L., Myagotin, A., Hänschke, D., Moosmann, J., Gunneweg, J. & Baumbach, T. (2012). Opt. Express, 20, 6496-6508.]). Recently it was shown that analyzer-based phase-contrast imaging provides a better sensitivity with respect to propagation-based and grating-based phase-contrast imaging under certain experimental conditions (Diemoz et al., 2012a[Diemoz, P. C., Bravin, A. & Coan, P. (2012a). Opt. Express, 20, 2789-2805.],b[Diemoz, P. C., Bravin, A., Langer, M. & Coan, P. (2012b). Opt. Express, 20, 27670-27690.]). Furthermore, analyzer-based phase-contrast imaging is free from the phase unwrapping problem (Judge & Bryanston-Cross, 1994[Judge, T. R. & Bryanston-Cross, P. J. (1994). Opt. Lasers Eng. 21, 199-239.]) associated with the fringe-scanning method (Bruning et al., 1974[Bruning, J. H., Herriott, D. R., Gallagher, J. E., Rosenfeld, D. P., White, A. D. & Brangaccio, D. J. (1974). Appl. Opt. 13, 2693-2703.]) used in grating-based phase-contrast imaging. Because of these advantages, the combination of X-ray analyzer-based phase-contrast imaging and SR-CL is expected to be very useful.

Currently, two kinds of X-ray optics are mainly used for analyzer-based phase-contrast imaging: optics that make use of a pencil-beam and optics that make use of an expanded beam. Whereas a symmetrically cut collimator crystal is used in the former, an asymmetrically cut collimator crystal is used in the latter for expanding the beam. Because a large field of view is desirable for observing a large sample, we have used the latter optics so far (Hirano, 2003[Hirano, K. (2003). J. Phys. D, 36, 1469-1472.], 2011[Hirano, K. (2011). Jpn. J. Appl. Phys. 50, 026402.]). Another advantage of the asymmetrically cut collimator crystal is that it has a larger acceptance angle than the symmetrically cut collimator crystal. We have also developed a generalized algorithm for obtaining multi-contrast (absorption, refraction and phase contrast) images.

In this paper, we propose the introduction of SR-CL to analyzer-based phase-contrast imaging for observing ROIs in laterally extended flat samples of weak absorption contrast. We also describe our generalized algorithm for obtaining multi-contrast images of a sample, and present the basic formulae for lamino­graphic reconstruction.

2. Theoretical background

2.1. Optical setup

An optical setup of X-ray analyzer-based phase-contrast computed lamino­graphy is schematically shown in Fig. 1[link]. The optics consist of an asymmetrically cut first crystal (collimator) and a symmetrically cut second crystal (analyzer) arranged in a nondispersive (+, −) diffraction geometry. In this optics, the diffraction plane is defined as the plane that includes the optical path. The incident monochromatic X-ray beam is collimated and expanded in the diffraction plane by the first crystal and propagates through the sample. The refraction caused by the sample is analyzed by the second crystal. The beam diffracted by the second crystal is observed by an X-ray area detector.

[Figure 1]
Figure 1
X-ray optics for analyzer-based phase-contrast computed laminography

Here, we introduce the laboratory coordinate systems (x1,y1,z1) for the sample and (x2,y2,z2) for the detector as shown in Fig. 1[link]. Here the zj axis (j = 1, 2) is antiparallel to the beam direction and the yj axis (j = 1, 2) is perpendicular to the diffraction plane. Using these coordinate systems, we can define an object plane (x1y1 plane) right after the sample and a detector plane (x2y2 plane) right before the detector. Notice that a point p,q) in the object plane is projected by the analyzer to a point (-p,q) in the detector plane.

2.2. Principles of X-ray analyzer-based phase-contrast imaging

The optical properties of the sample are described by the complex refraction index [n = 1 - \delta + i\beta]. The phase shift produced by the sample is given by

[\Phi ({x_1},{y_1}) = {{2\pi } \over \lambda }\int {\delta \left({{x_1},{y_1},{z_1}} \right)}\, {\rm d}{z_1}, \eqno(1)]

where λ is the wavelength. The beam refraction due to the phase shift is given by (λ/2π)∇Φ. The analyzer is only sensitive to the refraction component in the diffraction plane, which is given by

[{\psi _{{x_1}}}\left({{x_1},{y_1}} \right) = {\lambda \over {2\pi }}{{\partial \Phi } \over {\partial {x_1}}} = {\partial \over {\partial {x_1}}}\int {\delta \left({{x_1},{y_1},{z_1}} \right)}\, {\rm d}{z_1}. \eqno(2)]

Because this equation is similar to a Radon transformation, we can estimate the δ function from a set of refraction maps using the volume reconstruction algorithm.

Before estimating the refraction map of the sample, [{\psi _{{x_1}}}({{x_1},{y_1}})], it is necessary to first measure the rocking curve of the analyzer without the sample. Fig. 2[link] shows an example of the rocking curve, [{R_{\rm ref}}(\theta)], measured at a reference point near the optical axis on the detector plane. (The profile of this rocking curve was obtained by the experiments described in §3[link].) Here, Rref is the intensity of the diffracted beam and θ is the offset angle from the diffraction condition. According to the dynamical theory of X-ray diffraction (Zachariasen, 1945[Zachariasen, W. H. (1945). Theory of X-ray Diffraction in Crystals. New York: Wiley.]; Batterman & Cole, 1964[Batterman, B. W. & Cole, H. (1964). Rev. Mod. Phys. 36, 681-717.]; Ishikawa & Kohra, 1991[Ishikawa, T. & Kohra, K. (1991). Handbook on Synchrotron Radiation, Vol. 3, edited by G. S. Brown and D. E. Moncton, pp. 63-104. Amsterdam: North-Holland.]; Authier, 2001[Authier, A. (2001). Dynamical Theory of X-ray Diffraction. New York: Oxford University Press.]), the full width at half-maximum (FWHM) of [{R_{\rm ref}}(\theta)] is usually in the order of 0.1–10 arcsec. In the case of X-ray optics that make use of a symmetrically cut collimator crystal, the shape of [{R_{\rm ref}}(\theta)] is close to triangular. However, in the case of our optics, which make use of an asymmetrically cut collimator crystal, [{R_{\rm ref}}(\theta)] is no longer triangular but close to the silk-hat shape of the intrinsic rocking curve. This is why it is necessary to use the generalized algorithm shown below for image analyses.

[Figure 2]
Figure 2
An example of the rocking curve, [{R_{\rm ref}}(\theta)], measured at the reference point without the sample.

In general, the profile of the rocking curve gradually varies with the location of the measuring point. Here, we assume that the rocking curve measured at (x2,y2) without the sample is expressed as

[{R_{\rm o}}({\theta, {x_2},{y_2}}) = {a_{\rm o}}({{x_2},{y_2}}) {R_{\rm ref}}[{\theta - {b_{\rm o}}({{x_2},{y_2}} )} ], \eqno(3)]

where ao(x2,y2 ) is proportional to the incident beam intensity and bo(x2,y2 ) is the shift amount. At the reference point, ao = 1 and bo = 0. Both ao(x2,y2) and bo(x2,y2) depend on the optical setting and experimental conditions. For the estimation of ao(x2,y2) and bo(x2,y2), it is necessary to acquire two images on each side of the rocking curve, one at the low-angle side, [{\theta _{\rm LO}}], and one at the high-angle side, [{\theta _{\rm HO}}] (Fig. 2[link]). The acquired images are represented by [{R_{\rm o}}({{\theta _{\rm LO}},{x_2},y_2^{}} )] and [{R_{\rm o}}({{\theta _{\rm HO}},{x_2},y_2^{}})], respectively. Dividing the former by the latter, we can extract the effect of bo(x2,y2):

[\eqalignno{{{{R_{\rm o}}({{\theta _{\rm HO}},{x_2},{y_2}})} \over {{R_{\rm o}}({{\theta _{\rm LO}},{x_2},{y_2}})}} &= {{{a_{\rm o}}({{x_2},{y_2}}){R_{\rm ref}}[{{\theta _{\rm HO}} - {b_{\rm o}}({{x_2},{y_2}})}]} \over {{a_{\rm o}}({{x_2},{y_2}} ){R_{\rm ref}}[{{\theta _{\rm LO}} - {b_{\rm o}}({{x_2},{y_2}} )}]}}\cr &= {{{R_{\rm ref}}[{{\theta _{\rm HO}} - {b_{\rm o}}({{x_2},{y_2}} )} ]} \over {{R_{\rm ref}}[{{\theta _{\rm LO}} - {b_{\rm o}}({{x_2},{y_2}})} ]}}. &(4)} ]

For the estimation of bo(x2,y2), we introduce the following function:

[{V_{\rm o}}(\varepsilon) \equiv { {{R_{\rm ref}}({\varepsilon + {{\Delta {\theta _{\rm o}}}/2)}}} \over {{R_{\rm ref}}({\varepsilon - {{\Delta {\theta _{\rm o}}}/2)}}} } , \eqno(5)]

where [\Delta {\theta _{\rm o}}] is the difference between [{\theta _{\rm LO}}] and [{\theta _{\rm HO}}] ([\Delta {\theta _{\rm o}} \equiv {\theta _{\rm HO}} - {\theta _{\rm LO}}]). [\Delta {\theta _{\rm o}}] is usually set around the FWHM of the rocking curve, [{R_{\rm ref}}(\theta)]. For example, Fig. 3[link] shows the calculated [{V_{\rm o}}(\varepsilon)] curve from the rocking curve, [{R_{\rm ref}}(\theta)], shown in Fig. 2[link]. Using this [{V_{\rm o}}(\varepsilon)] curve as a lookup table, we can estimate the value, [{\varepsilon _{\rm o}}], which corresponds to the measured value of [{{{R_{\rm o}}({{\theta _{\rm HO}},{x_2},{y_2}})}/{{R_{\rm o}}({{\theta _{\rm LO}},{x_2},{y_2}})}}]. Then we can obtain bo(x2,y2) by

[{b_{\rm o}}({{x_2},{y_2}}) = - {\varepsilon _{\rm o}}\left({{x_2},{y_2}} \right) + {{{\theta _{\rm LO}} + {\theta _{\rm HO}}} \over 2}. \eqno(6)]

As shown in Fig. 3[link], the estimation of [{\varepsilon _{\rm o}}] is only possible in the limited region between [{\varepsilon _{\rm L}}] and [{\varepsilon _{\rm H}}]. The width of this region is approximately the FWHM of the rocking curve, [{R_{\rm ref}}(\theta)], and is usually sufficient for observing most samples. Once the bo(x2,y2) is obtained, ao(x2,y2) is given by

[\eqalignno{{a_{\rm o}}({{x_2},{y_2}}) &= {{{R_{\rm o}}({{\theta _{\rm HO}},{x_2},{y_2}})} \over {{R_{\rm ref}}[{{\theta _{\rm HO}} - {b_{\rm o}}({{x_2},{y_2}})}]}}\cr &= {{{R_{\rm o}}({{\theta _{\rm LO}},{x_2},{y_2}})} \over {{R_{\rm ref}}[{{\theta _{\rm LO}} - {b_{\rm o}}({{x_2},{y_2}})}]}}. &(7)}]

[Figure 3]
Figure 3
Example of the calculated [{V_{\rm o}}(\varepsilon)] curve from the rocking curve, [{R_{\rm ref}}(\theta )], shown in Fig. 2[link].

When the sample is inserted into the optical path, the X-ray beam is absorbed and refracted by the sample. Here, we disregard the effect of X-ray small-angle scattering. In this case, the rocking curve of the analyzer is expressed as

[\eqalignno{ {{R_{\rm s}}({\theta, {x_2},{y_2}})} = &\,\,{a_{\rm o}}({{x_2},{y_2}})\,{t_{\rm s}}({{x_2},{y_2}} )\,{R_{\rm ref}}[{\theta} - {b_{\rm o}}({{x_2},{y_2}}) \cr &- {r_{\rm s}}({{x_2},{y_2}}) ], &(8)}]

where ts(x2,y2) is the transmissivity and rs(x2,y2) is the refraction angle. The former corresponds to the absorption-contrast image (absorption map) and the latter to the refraction-contrast image (refraction map). We can estimate the refraction map, rs(x2,y2), from two images recorded at [{\theta _{\rm LS}}] and [{\theta _{\rm HS}}]. Dividing [{R_{\rm s}}({{\theta _{\rm HS}},{x_2},y_2^{}})] by [{R_{\rm s}}({{\theta _{\rm LS}},{x_2},y_2^{}})], we can extract the effect of rs(x2,y2),

[{{{R_{\rm s}}({{\theta _{\rm HS}},{x_2},{y_2}})} \over {{R_{\rm s}}({{\theta _{\rm LS}},{x_2},{y_2}})}} = {{{R_{\rm ref}}[{{\theta _{\rm HS}} - {b_{\rm o}}({{x_2},{y_2}}) - {r_{\rm s}}({{x_2},{y_2}})}]} \over {{R_{\rm ref}}[{{\theta _{\rm LS}} - {b_{\rm o}}({{x_2},{y_2}}) - {r_{\rm s}}({{x_2},{y_2}})}]}}. \eqno(9)]

For the estimation of rs(x2,y2), we reintroduce the following function:

[{V_{\rm s}}(\varepsilon) \equiv {{R_{\rm ref}}({\varepsilon + {\Delta {\theta _{\rm s}}}/2}) \over {{R_{\rm ref}}({\varepsilon - {\Delta {\theta _{\rm s}}}/2} ) }}, \eqno(10)]

where [\Delta {\theta _{\rm s}}] is the difference between [{\theta _{\rm LS}}] and [{\theta _{\rm HS}}] [(\Delta {\theta _{\rm s}} \equiv {\theta _{\rm HS}} - {\theta _{\rm LS}})]. Using this function as the lookup table, we can estimate the value, [{\varepsilon _{\rm s}}], which corresponds to the measured value of [{{{R_{\rm s}}({{\theta _{\rm HS}},{x_2},{y_2}})} / {{R_{\rm s}}({{\theta _{\rm LS}},{x_2},{y_2}})}}]. Then we can obtain rs(x2,y2) from

[\eqalignno{{r_{\rm s}}({{x_2},{y_2}}) = &\,\,{\varepsilon _{\rm o}}({{x_2},{y_2}}) - {\varepsilon _{\rm s}}({{x_2},{y_2}})\cr &+ ({{{\theta _{\rm LS}} + {\theta _{\rm HS}} - {\theta _{\rm LO}} - {\theta _{\rm HO}}}) / 2}. & (11)}]

If [{\theta _{\rm LS}} = {\theta _{\rm LO}}] and [{\theta _{\rm HS}} = {\theta _{\rm HO}}], this equation is simplified as [{r_{\rm s}}({{x_2},{y_2}}) = {\varepsilon _{\rm o}}({{x_2},{y_2}}) - {\varepsilon _{\rm s}}({{x_2},{y_2}})]. Once rs(x2,y2) is obtained, ts(x2,y2) is given by

[\eqalignno{ {a_{\rm o}}({{x_2},{y_2}}){t_{\rm s}}({{x_2},{y_2}}) &= {{{R_{\rm s}}({{\theta _{\rm HS}},{x_2},{y_2}})} \over {{R_{\rm ref}}[{{\theta _{\rm HS}} - {b_{\rm o}}({{x_2},{y_2}}) - {r_{\rm s}}({{x_2},{y_2}})}]}}\cr &= {{{R_{\rm s}}({{\theta _{\rm LS}},{x_2},{y_2}})} \over {{R_{\rm ref}}[{{\theta _{\rm LS}} - {b_{\rm o}}({{x_2},{y_2}}) - {r_{\rm s}}({{x_2},{y_2}})} ]}}. &(12)}]

Note that both ts(x2,y2) and rs(x2,y2) are the images on the detector plane. To obtain the corresponding images on the object plane (x1y1 plane), it is necessary to invert the images along the x2 axis. Without this procedure, lamino­graphic reconstruction will result in failure.

2.3. Lamino­graphic reconstruction

For lamino­graphic reconstruction, it is necessary to obtain a set of refraction maps by rotating the sample through 360°. Here we introduce a coordinate system (u,v,w) attached to the sample as shown in Fig. 1[link]. The rotation axis ([\hat \varphi] axis) coincides with the v axis, and is tilted at angle α from the y1 axis. When φ = 0° and α = 0°, the object coordinate system (u,v,w) coincides with the laboratory coordinate system (x1,y1,z1). The relationship between the coordinates of a given point in the object coordinate system (u,v,w) and in the laboratory coordinate system (x1,y1,z1) is given by

[{\pmatrix{ {{x_1}} \cr {{y_1}} \cr {{z_1}} \cr } } = {\pmatrix{ {\cos \varphi } & 0 &{ - \sin \varphi } \cr {\sin \alpha \sin \varphi } & {\cos \alpha } & {\sin \alpha \cos \varphi } \cr {\cos \alpha \sin \varphi } & { - \sin \alpha } & {\cos \alpha \cos \varphi } \cr } }{\pmatrix{ u \cr v \cr w \cr } } . \eqno(13)]

X-ray lamino­graphic reconstruction is based on the filtered backprojection method and is represented by the following equation,

[\eqalignno{{\delta _{\rm fbp}}({u,v,w}) &= {1 \over {2\pi_{\vphantom{\big|}} }}\int_0^{2\pi } {p({\varphi, {x_1},{y_1}})}\, {\rm d}\varphi, \cr &\left\{ {\matrix{ {{x_1} = u\cos \varphi - w\sin \varphi, }\hfill\cr {{y_1} = u\sin \alpha \sin \varphi + v\cos \alpha + w\sin \alpha \cos \varphi,\hfill } \cr } } \right. &(14)}]

where [p({\varphi, {x_1},{y_1}})] is a filtered projection image for each φ angle. A reconstructed sectional image of [{\delta _{\rm fbp}}] at a given v value is parallel to the uw plane. The three-dimensional volume data can be obtained by stacking the reconstructed images along the v axis. The filtered projection image, [p({\varphi, {x_1},{y_1}})], is given by

[\eqalignno{p({\varphi, {x_1},{y_1}}) = & \int {\left[\,\,{\int {{\psi_{{x_1}}}({\varphi, {x_1},{y_1}})\exp ({ - 2\pi i{x_1}\omega } )\, {\rm d}{x_1}} }\right]}\cr & \times H(\omega)\exp ({2\pi i{x_1}\omega })\, {\rm d}\omega, &(15)}]

where [{\psi _{{x_1}}}({\varphi, {x_1},{y_1}})] is the refraction map on the object plane and [H(\omega)] is the one-dimensional filter function. This equation can be simply expressed as

[p({\varphi, {x_1},{y_1}}) = {F^{\, - 1}}[{F({{\psi _{{x_1}}}}) \times H}], \eqno(16)]

where F denotes the one-dimensional Fourier transform and F - 1 the one-dimensional inverse Fourier transform. The filter function is given by

[H(\omega) = - i{{\cos \alpha } \over 2}{\rm sgn}(\omega) \eqno(17)]

with

[{\rm sgn} (\omega) = \left\{ {\matrix{ { - 1\quad {\rm if}\quad \omega \,\lt\, 0} \cr {\,\,\,\,\,\,0\quad {\rm if}\quad \omega = 0} \cr { + 1\quad {\rm if}\quad \omega \,\gt\, 0} \cr } } \right.]

(Harasse et al., 2010[Harasse, S., Hirayama, N., Yashiro, W. & Momose, A. (2010). Proc. SPIE, 7804, 780411.]).

Note that this lamino­graphic reconstruction is the generalization of the tomosynthetic approach (Maksimenko et al., 2007[Maksimenko, A., Yuasa, T., Ando, M. & Hashimoto, E. (2007). Appl. Phys. Lett. 91, 234108.]). It is also worth noting that the [{\delta _{\rm fbp}}] given by equation (14[link]) is only an approximation of δ due to (i) incomplete sampling in Fourier space and (ii) the filtering process in equation (15[link]).

3. Experiments and results

To verify the feasibility of X-ray analyzer-based phase-contrast computed lamino­graphy, we performed experiments at the vertical wiggler beamline BL-14B of the Photon Factory. At first, the white beam from the light source was monochromated at 0.0733 nm by a Si(111) double-crystal monochromator. Then, the monochromatic beam linearly polarized in the vertical direction was incident to the X-ray optics shown in Fig. 1[link]. For minimizing the intensity loss caused by the polarization effect, the diffraction plane of the optics was set in the horizontal plane. This means that the polarization of the incident beam corresponded to σ-polarization for the optics. For the collimator, we used an asymmetric Si(220) crystal. The angle between the crystal surface and the diffracting lattice planes was 10° and the Bragg angle was 11°. As a result, the incident beam was expanded 10.9 times in the horizontal plane. Then the central part of the expanded beam was selected by a slit. For the analyzer, we used a symmetric Si(220) crystal. Both the collimator and the analyzer were made of non-doped float-zone silicon crystal (ρ ≥ 2000 Ω cm), and their surfaces were mechanochemically polished in order to remove defects and strain fields. The beam diffracted by the analyzer was recorded on a fiber-coupled X-ray CCD camera (Photonic Science Limited, X-ray FDI 1:1) consisting of a GdO2S:Tb scintillator (7 mg cm−2), a 1:1 glass fiber plate and a CCD. At the scintillator, X-rays were converted to visible light, which was transmitted through the glass fiber plate and detected by the CCD. The pixel size was 6.45 µm (H) × 6.45 µm (V) and the number of pixels was 1392 (H) × 1040 (V). The distance between the sample and the analyzer was about 25 cm, and the distance between the analyzer and the X-ray CCD camera was about 25 cm.

First, we measured the rocking curve of the analyzer without the sample. The aperture of the slit was set at 1 mm (H) × 1 mm (V) in size. The shape of the obtained rocking curve was close to the silk-hat shape of the intrinsic rocking curve. The FWHM was estimated to be 2.5 arcsec. This value agrees well with the theoretical value of 2.3 arcsec. The slight discrepancy is considered to be caused by deformation of the collimator and/or analyzer. Next, we set the aperture of the slit at 10 mm (H) × 10 mm (V) in size, and recorded the beam diffracted by the analyzer at θLO = −1.25 arcsec and θHO = +1.25 arcsec. The exposure time was 1 s per image. Using the calculated [{V_{\rm o}}\left(\varepsilon \right)] curve as the lookup table, we estimated bo(x2,y2 ).

Then, we inserted a sample into the optical path. We used plastic beads made from acrylic resin as the sample. The diameter of each bead was about 3 mm. Fig. 4[link] is a photograph showing the area around the sample holder. The [\hat \alpha] axis is on the horizontal plane and is perpendicular to the optical path. The φ stage was mounted on the α stage so that the [\hat \varphi] axis would always become perpendicular to the [\hat \alpha] axis and perpendicular to the optical path at α = 0°. In the experiments, we fixed the α angle at 30°. A photograph of the sample holder is also shown in the inset of Fig. 4[link]. The sample holder was made by attaching an acrylic sheet to a polypropylene tube of 6 mm diameter with an adhesive made of dichloro­methylene.

[Figure 4]
Figure 4
Photograph showing the area around the sample holder. The [\hat \alpha] axis is in the horizontal plane and perpendicular to the optical path. The φ stage is mounted on the α stage so that the [\hat \varphi] axis would always become perpendicular to the [\hat \alpha] axis and perpendicular to the optical path at α = 0°. The lower-left inset is a photograph of the sample holder. The sample holder is made by attaching an acrylic sheet to a polypropyl­ene tube of 6 mm in diameter with an adhesive made of dichloromethylene.

To obtain the refraction maps of the sample, we set the analyzer angle at θLS = θLO = −1.25 arcsec and θHS = θHO = +1.25 arcsec. It took less than a few seconds to change the analyzer angle and stabilize the analyzer crystal. At each analyzer angle we recorded a set of images by rotating the sample around the [\hat \varphi] axis through 360° in 1° steps. The exposure time was 1 s per image and the total measurement time was ∼30 min. About 60% of the measurement time was spent on rotation of the sample and data transfer from the X-ray CCD camera to the workstation. We will be able to reduce this dead time by using an X-ray detector with short readout time such as a scientific CMOS camera and by implementing an on-the-fly scan.

From the set of obtained refraction maps, we calculated the phase-contrast sectional images of the sample. Fig. 5[link] shows the results obtained at different depth positions: (a) v = 0 mm and (b) v = 0.97 mm. Depth-dependent features can be clearly seen, which shows that the lamino­graphic reconstruction was successful. In each image, a strong circular artifact is also observed. These artifacts originate from the joint of the sample holder. Due to these artifacts, the FOV was limited to about 6 mm in diameter. Clear phase-contrast sectional images can be obtained as long as the ROI is located inside this FOV. We will be able to expand the FOV by replacing the polypropylene tube of the sample holder with a larger one. For comparison, corresponding absorption-contrast images are also shown in Figs. 5[link](c) and 5[link](d). The phase-contrast images are much clearer than the absorption-contrast images. It is also seen that the edge of each bead is slightly enhanced due to Fresnel diffraction in Figs. 5[link](c) and 5[link](d).

[Figure 5]
Figure 5
Reconstructed phase-contrast sectional images of the plastic beads at different depth positions: (a) v = 0 mm and (b) v = 0.97 mm. For comparison, corresponding absorption-contrast images are also shown in (c) and (d). The strong circular artifact in each image originates from the joint of the sample holder. The size of each scale bar is 1 mm.

In this study, we assumed that the effect of ultra-small-angle X-ray scattering (USAX) caused by a sample is negligible. Regarding the plastic beads used in our experiment, this assumption is valid. However, for a sample where this assumption is no longer valid, our approach will not provide a correct image. For such a sample, multiple-image radiography (Pagot et al., 2003[Pagot, E., Cloetens, P., Fiedler, S., Bravin, A., Coan, P., Baruchel, J., Härtwig, J. & Thomlinson, W. (2003). Appl. Phys. Lett. 82, 3421-3423.]) will be useful for obtaining correct multi-contrast (absorption, refraction, USAX and phase contrast) sectional images.

4. Conclusion

We have developed X-ray analyzer-based phase-contrast computed lamino­graphy for imaging ROIs in laterally expanded flat specimens of weak absorption contrast. Our optics consist of an asymmetrically cut collimator crystal and a symmetrically cut analyzer crystal arranged in a nondispersive (+, −) diffraction geometry. We showed the generalized algorithm for calculating multi-contrast (absorption, refraction and phase contrast) images of a sample. Furthermore, we presented the basic formulae for lamino­graphic reconstruction. In the experiments performed at the vertical wiggler beamline BL-14B of the Photon Factory, we successfully obtained phase-contrast sectional images of the plastic bead sample at a wavelength of 0.0733 nm. Strong circular artifacts originating from the joint of the sample holder were also observed. Due to these artifacts, the FOV was limited to about 6 mm in diameter. It will be possible to enlarge the FOV by replacing the polypropylene tube of the sample holder with a larger one.

Acknowledgements

This work was supported by the Cross-Ministerial Strategic Innovation Promotion Program – Unit D66 – Innovative measurement and analysis for structural materials (SIP-IMASM) led by the Cabinet Office, Government of Japan. The experiments were performed with the approval of the Program Advisory Committee of the Photon Factory (2015S2-002 and 2016S2-001).

References

First citationAltapova, V., Helfen, L., Myagotin, A., Hänschke, D., Moosmann, J., Gunneweg, J. & Baumbach, T. (2012). Opt. Express, 20, 6496–6508.  Web of Science CrossRef PubMed Google Scholar
First citationAuthier, A. (2001). Dynamical Theory of X-ray Diffraction. New York: Oxford University Press.  Google Scholar
First citationBatterman, B. W. & Cole, H. (1964). Rev. Mod. Phys. 36, 681–717.  CrossRef CAS Web of Science Google Scholar
First citationBruning, J. H., Herriott, D. R., Gallagher, J. E., Rosenfeld, D. P., White, A. D. & Brangaccio, D. J. (1974). Appl. Opt. 13, 2693–2703.  CrossRef CAS PubMed Web of Science Google Scholar
First citationChapman, D., Thomlinson, W., Johnston, R. E., Washburn, D., Pisano, E., Gmür, N., Zhong, Z., Menk, R., Arfelli, F. & Sayers, D. (1997). Phys. Med. Biol. 42, 2015–2025.  CrossRef CAS PubMed Web of Science Google Scholar
First citationDavis, T. J., Gao, D., Gureyev, T. E., Stevenson, A. W. & Wilkins, S. W. (1995). Nature (London), 373, 595–598.  CrossRef CAS Web of Science Google Scholar
First citationDiemoz, P. C., Bravin, A. & Coan, P. (2012a). Opt. Express, 20, 2789–2805.  CrossRef CAS PubMed Google Scholar
First citationDiemoz, P. C., Bravin, A., Langer, M. & Coan, P. (2012b). Opt. Express, 20, 27670–27690.  CrossRef CAS PubMed Google Scholar
First citationDobbins, J. T. III & Godfrey, D. J. (2003). Phys. Med. Biol. 48, R65–R106.  Web of Science CrossRef PubMed Google Scholar
First citationHarasse, S., Hirayama, N., Yashiro, W. & Momose, A. (2010). Proc. SPIE, 7804, 780411.  CrossRef Google Scholar
First citationHelfen, L., Baumbach, T., Cloetens, P. & Baruchel, J. (2009). Appl. Phys. Lett. 94, 104103.  Web of Science CrossRef Google Scholar
First citationHelfen, L., Baumbach, T., Mikulík, P., Kiel, D., Pernot, P., Cloetens, P. & Baruchel, J. (2005). Appl. Phys. Lett. 86, 071915.  Web of Science CrossRef Google Scholar
First citationHelfen, L., Myagotin, A., Mikulík, P., Pernot, P., Voropaev, A., Elyyan, M., Di Michiel, M., Baruchel, J. & Baumbach, T. (2011). Rev. Sci. Instrum. 82, 063702.  Web of Science CrossRef PubMed Google Scholar
First citationHirano, K. (2003). J. Phys. D, 36, 1469–1472.  CrossRef CAS Google Scholar
First citationHirano, K. (2011). Jpn. J. Appl. Phys. 50, 026402.  CrossRef Google Scholar
First citationHoshino, M., Uesugi, K., Takeuchi, A., Suzuki, Y. & Yagi, N. (2011). Rev. Sci. Instrum. 82, 073706.  Web of Science CrossRef PubMed Google Scholar
First citationHounsfield, G. N. (1973). Br. J. Radiol. 46, 1016–1022.  CrossRef CAS PubMed Google Scholar
First citationIshikawa, T. & Kohra, K. (1991). Handbook on Synchrotron Radiation, Vol. 3, edited by G. S. Brown and D. E. Moncton, pp. 63–104. Amsterdam: North-Holland.  Google Scholar
First citationJudge, T. R. & Bryanston-Cross, P. J. (1994). Opt. Lasers Eng. 21, 199–239.  CrossRef Google Scholar
First citationKrug, K., Porra, L., Coan, P., Wallert, A., Dik, J., Coerdt, A., Bravin, A., Elyyan, M., Reischig, P., Helfen, L. & Baumbach, T. (2008). J. Synchrotron Rad. 15, 55–61.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMaksimenko, A., Yuasa, T., Ando, M. & Hashimoto, E. (2007). Appl. Phys. Lett. 91, 234108.  CrossRef Google Scholar
First citationPagot, E., Cloetens, P., Fiedler, S., Bravin, A., Coan, P., Baruchel, J., Härtwig, J. & Thomlinson, W. (2003). Appl. Phys. Lett. 82, 3421–3423.  Web of Science CrossRef CAS Google Scholar
First citationVerboven, P., Herremans, E., Helfen, L., Ho, Q. T., Abera, M., Baumbach, T., Wevers, M. & Nicolaï, B. M. (2015). Plant J. 81, 169–182.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWatanabe, N., Hoshino, M., Yamamoto, K., Aoki, S., Takeuchi, A. & Suzuki, Y. (2009). AIP Conf. Ser. 186, 012022.  Google Scholar
First citationXu, F., Helfen, L., Suhonen, H., Elgrabli, D., Bayat, S., Reischig, P., Baumbach, T. & Cloetens, P. (2012). PLoS One, 7, e50124.  CrossRef PubMed Google Scholar
First citationZachariasen, W. H. (1945). Theory of X-ray Diffraction in Crystals. New York: Wiley.  Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds