short communications
Non-uniform nanosecond gate-delay of hybrid pixel detectors
aDeutsches Elektronen-Synchrotron (DESY), D-22603 Hamburg, Germany, bPhysics Department, University of Hamburg, Germany, cEuropean Synchrotron Radiation Facility (ESRF), Grenoble, France, and dHelmholtz-Zentrum Berlin für Materialien und Energie GmbH, Wilhelm-Conrad-Röntgen Campus, BESSY II, Albert-Einstein-Strasse 15, 12489 Berlin, Germany
*Correspondence e-mail: roman.shayduk@desy.de
A simple experiment to characterize the gating properties of X-ray area detectors using pulsed X-ray sources is presented. For a number of time-resolved experiments the gating uniformity of area detectors is important. Relative gating delays between individual modules and readout chips of PILATUS2 series area X-ray detectors have been observed. For three modules of a PILATUS 300K-W unit the maximum gating offset between the modules is found to be as large as 30 ns. On average, the first photosensor module is found to be triggered 15 ns and 30 ns later than the second and the third modules, respectively.
1. Introduction
Synchrotron storage rings generate X-ray radiation that consists of a series of short X-ray photon bursts separated by, essentially, photon-free gaps. Many storage rings are operated in the so-called hybrid and/or single-bunch mode providing an X-ray pulse obtained from an isolated electron bucket that is separated from the rest of the bunches by two electron-free gaps hundreds of nanoseconds long. It has been shown that these isolated pulses can be gated by modern area X-ray detectors (Ejdrup et al., 2009; Reinhardt et al., 2016). Here all the unwanted signal is masked and it makes it possible to carry out synchrotron-based pump–probe time-resolved X-ray scattering experiments with time resolution down to hundreds of femtoseconds (Schoenlein et al., 2000) without the need of a mechanical chopper (Wulff et al., 2003).
Although hybrid pixel area X-ray detectors use a large-area single-crystal photosensor, the spatial resolution is achieved via hybrid pixel architecture (Broennimann et al., 2006; Dinapoli et al., 2011; Ponchut et al., 2002). The readout electronics constituting a photosensor are usually modular. When such a detector receives a gate signal it is important that the signal is passed to and processed by all the readout modules simultaneously.
In this short communication we describe a simple experiment that is aimed at characterizing the area X-ray detector triggering uniformity. In particular we report on the gating non-uniformity of two area X-ray detectors PILATUS 300K-W and PILATUS 100K employed in real synchrotron-based pump–probe X-ray diffraction experiments. The carried-out `pulsed flat intensity field' experiments could serve as a standard tool to characterize the gating uniformity of a whole class of hybrid pixel area X-ray detectors.
2. Experimental setups
2.1. ESRF setup
The experiment with the PILATUS 300K-W unit was carried out at the ID03 surface X-ray diffraction beamline at the European Synchrotron Radiation Facility (ESRF). The storage ring was operated in hybrid 7/8 bunch-filling mode with 868 electron bunches filled with 0.23 mA per bunch and one isolated single 1 mA bunch placed in the center of a 356 ns-wide electron bunch-free gap as schematically shown in Fig. 1(b). We used monochromatic X-rays with an energy of 15 keV. The direct beam intensity was of the order of 1012 photons s−1.
Fig. 1(a) shows a sketch of the experiment. The X-ray detector was placed in the region of strong diffuse scattering in the vicinity of the non-attenuated direct X-ray beam. The detector was operated in the so-called `external enable' triggering mode1 and received a TTL gate signal from a DG645 delay generator (Stanford Inc.). The signal had sharp rising and falling edges of around 200 ps and a width of 50 ns. In the experiment, the DG645 signals were phase-locked to the storage-ring revolution frequency and the repetition rate of the signals were frequency divided to reach a rate of ∼1 kHz. The reduction of the repetition rate to 1 kHz was required by the laser used in the ESRF experiment.
2.2. BESSY setup
The experiment with the PILATUS 100K single-module unit was carried out at the KMC3 beamline of the BESSY II storage ring in Berlin, Germany. During the experiment, BESSY II was operated in the so-called `low-α multi-bunch hybrid mode'. In this operation mode an isolated electron bunch is placed near the center of a 184 ns-wide bunch-free gap as schematically shown in Fig. 1(b). As in the ESRF experiment the detector was operated in the `external enable' triggering mode and received a phase-locked 20 ns-long TTL gate signal. The repetition rate of the signal in this experiment was set to 1.25 MHz, equal to the BESSY revolution frequency. The X-ray energy used was 11.9 keV.
In practice, one readout frame of a PILATUS detector in the `external enable' gated mode synthesizes a digital image that stores photon hits in each pixel due to multiple ultrashort gated exposures. For example, for a 1 kHz trigger rate a 1 s `exposure' produces one digital image that is the sum of 1000 gated exposures. In the `external enable' triggered mode it was possible to trigger the detector exposure at a repetition rate that exceeds by far the chip readout-limited rate of around 300 Hz. The internal mechanism of a single ultrashort gate exposure controlled by an external gate is a part of the PILATUS detector architecture (Kraft et al., 2009) and will not be discussed in this article.
3. Results
3.1. Test of PILATUS 300K-W
During the experiment the detector was kept at a fixed position. The readout period in the experiments was 1 s, equivalent to 1000 exposures at 1 kHz. The starting point of the gate signal with respect to the single bunch was changed in steps of 1 ns over the revolution period of 2.8 µs. The integrated counts of different regions of interest (ROIs) of the photosensor were recorded as a function of the gate delay. Fig. 2(b) shows two overlayed histograms of integrated counts as a function of gate delay taken from two different ROIs. To compare how different regions of a detector respond to the gate signal, ROI(1,1) is placed near the upper left corner of the detector and ROI(14,2) is the region in the center of the detector as indicated in Fig. 2(b).
From the data we unambiguously determine at which gate delay only those X-ray photons are counted that correspond to the isolated electron bunch i.e. `single bunch' (SB) X-ray photons. The `zero point' of the gate delay in the plot is chosen at the single-bunch center-of-mass (COM) of ROI(1,1). These photons produce count events that form the trapezoidal distribution around zero. One can also see the ∼90 ns-long counts-free regions corresponding to the two 178 ns-long bunch-free time gaps of the 7/8+1 ESRF filling mode (one bunch placed in the center of the 356 ns-long bunch-free gap). The higher counting statistics on the left and on the right sides of the histograms are due to the capture of X-ray photons generated by 2.8 ns-separated electron bunches of ESRF. Due to the detector response time limitations, the multi-bunch time structure is not resolved in this experiment.
One striking feature of this particular PILATUS 300K-W module is the 30 ns offset of the COM of ROI(1,1) counts compared with the ROI(14,2) counts COM. In addition, the SB counts falling edge of ROI(14,2) is observed about 20 ns before the SB counts falling edge of ROI(1,1). The pulse length of a single X-ray pulse is of the order of 100 ps and the spatial distribution of the synchrotron radiation
over the detector photosensor is rather homogeneous in this experiment. Therefore, these systematic time offsets in the histograms which depend on the position of the photosensor area reflect the gating property of the detector readout electronics. It is unlikely that the observed time offsets of 20–30 ns stem from signal propagation in the detector. This would require a cable length of 4–6 m in the detector. We exclude any environmental effects as the X-ray hutch is equipped with air-conditioning that supports constant temperature and humidity. Furthermore, the environment or the detector setting fluctuations during the experiment would most likely affect the absolute delays. The observed statistically averaged relative delays as a function of photosensor region would stay intact.Figs. 2(b) and 2(c) show color-coded maps of the timings of a SB's rising and falling edges for the whole area of the PILATUS 300K detector for all 48 different readout chips of the detector. The rising and the falling edges were determined by fitting Gaussians to the differentiated data of Fig. 2(a) type. The standard deviations of the fits were of the order of 3.8 ns. These data clearly show differences in the measured SB position depending on the module. The maximum gating difference for this PILATUS 300K-W unit was of the order of 30 ns as determined in this experiment and was between the first and the third modules. It is not yet clear whether this offset is peculiar to this particular detector or whether it is a typical property of the PILATUS series of multi-module detectors. Within each module, the variation in timing behavior is significantly smaller (less than 12 ns) and forms a fairly random pattern. On average, the first photosensor module is triggered 15 ns and 30 ns later than the second and the third modules, respectively.
In Fig. 2(d) one can see an example of a gated PILATUS 300K-W image taken in a real pump–probe surface X-ray diffraction experiment. The image is taken at zero relative delay [see Fig. 2(a)]. The complex intensity structure in the PILATUS image is essentially coming from the X-ray powder diffraction rings of the Be chamber used in the experiment and from the diffuse scattering that is shadowed by the detector flight tube. However, on top of the complex intensity structure one clearly distinguishes rectangular regions of `dimmed' count rates. These regions reveal the 2 × 8 readout chip structure of a single module. The `dimmed chip' effect nearly disappears when the relative gate delay is set to −10 ns and the gate width is extended to 100 ns. For the 50 ns-long gate the chip contrast was clearly present in detector images for all gate delays. Hence, using the 50 ns-long gate the simultaneous gating of all the three modules was not possible.
3.2. Test of PILATUS 100K
Fig. 3 represents the data similar to those shown in Figs. 2(a)–2(c) but acquired for a different single-module PILATUS 100K area X-ray detector at the BESSY synchrotron. In this experiment the data were taken with much higher statistics. Fig. 3(a) shows the number of counted photons as a function of gate delay near the SB for two different ROIs. Here we used ROIs of 6 × 6 PILATUS pixels. Such ROIs were distributed on a grid to cover the entire PILATUS 100K photosensor. In Fig. 3(b) we plot a color map of SB COM position as a function of ROI on the grid. By doing so, the 2 × 8 readout chip structure of the PILATUS 100K detector module becomes visible. Although the apparent delays are quite small now, some chips still represent a systematic gate offset of up to 5 ns on average. There is also an apparent COM position gradient within each chip observed. The standard deviation of the COM determination in this experiment is limited only by the counting statistics and is of the order of 200 ps.
4. Conclusion
We describe a simple experiment to characterize the gating properties of X-ray area detectors using pulsed X-ray sources.
The gating performance of two PILATUS 300K-W and PILATUS 100K detectors employed in real synchrotron-based pump–probe X-ray diffraction experiments is discussed. Systematic gating offsets of up to 30 ns between three modules constituting the PILATUS 300K-W photosensitive area were observed. On average, the second and the third photosensor modules were found to be triggered 15 ns and 30 ns later than the first module, respectively. Based on the results one can speculate that the gate delay would grow progressively with the number of modules used, i.e. for a PILATUS 6M it could be as large as 300 ns. The results of our experiments on a PILATUS 100K detector are in agreement with earlier work (Ejdrup et al., 2009), demonstrating that the detector can be used to achieve X-ray pulse duration-limited time resolution in synchrotron-based pump–probe experiments. However, the multi-modular unit PILATUS 300K-W requires the analysis of gating performance in multiple regions of the photosensitive area to determine the optimal gate timing in a pump–probe experiment.
The experiment described in this article is proposed as the standard evaluation procedure for hybrid pixel area X-ray detectors gating capabilities. The results of these experiments deliver important information required for the proper implementation of area X-ray detectors in synchrotron-based pump–probe experiments.
Footnotes
1As described in the PILATUS manual.
References
Broennimann, Ch., Eikenberry, E. F., Henrich, B., Horisberger, R., Huelsen, G., Pohl, E., Schmitt, B., Schulze-Briese, C., Suzuki, M., Tomizaki, T., Toyokawa, H. & Wagner, A. (2006). J. Synchrotron Rad. 13, 120–130. Web of Science CrossRef CAS IUCr Journals Google Scholar
Dinapoli, R., Bergamaschi, A., Henrich, B., Horisberger, R., Johnson, I., Mozzanica, A., Schmid, E., Schmitt, B., Schreiber, A., Shi, X. T. & Theidel, G. (2011). Nucl. Instrum. Methods Phys. Res. A, 650, 79–83. Web of Science CrossRef CAS Google Scholar
Ejdrup, T., Lemke, H. T., Haldrup, K., Nielsen, T. N., Arms, D. A., Walko, D. A., Miceli, A., Landahl, E. C., Dufresne, E. M. & Nielsen, M. M. (2009). J. Synchrotron Rad. 16, 387–390. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kraft, P., Bergamaschi, A., Brönnimann, C., Dinapoli, R., Eikenberry, E. F., Graafsma, H., Henrich, B., Johnson, I., Kobas, M., Mozzanica, A., Schlepütz, C. M. & Schmitt, B. (2009). IEEE Trans. Nucl. Sci. 56, 758–764. Web of Science CrossRef CAS Google Scholar
Ponchut, C., Visschers, J. L., Fornaini, A., Graafsma, H., Maiorino, M., Mettivier, G. & Calvet, D. (2002). Nucl. Instrum. Methods Phys. Res. A, 484, 396–406. Web of Science CrossRef CAS Google Scholar
Reinhardt, M., Koc, A., Leitenberger, W., Gaal, P. & Bargheer, M. (2016). J. Synchrotron Rad. 23, 474–479. Web of Science CrossRef CAS IUCr Journals Google Scholar