research papers
Development of an X-ray imaging system to prevent scintillator degradation for white synchrotron radiation
aDiamond Light Source Ltd, Harwell Science and Innovation Campus, Didcot OX11 0DE, UK
*Correspondence e-mail: hongchang.wang@diamond.ac.uk
The high
of the white X-ray beams from third-generation synchrotron light sources can significantly benefit the development of high-speed X-ray imaging, but can also bring technical challenges to existing X-ray imaging systems. One prevalent problem is that the image quality deteriorates because of dust particles accumulating on the scintillator screen during exposure to intense X-ray radiation. Here, this problem has been solved by embedding the scintillator in a flowing inert-gas environment. It is also shown that the detector maintains the quality of the captured images even after days of X-ray exposure. This modification is cost-efficient and easy to implement. Representative examples of applications using the X-ray imaging system are also provided, including fast tomography and multimodal phase-contrast imaging for biomedical and geological samples.Keywords: white-beam detector; fast imaging; phase contrast; scintillators; X-ray exposure; inert gas.
1. Introduction
Third-generation synchrotron sources generate hard X-rays with significantly higher et al., 2010), and the stripes are difficult to remove by flat-field correction in post-processing. To maximize the usage of the high of the synchrotron radiation and further decrease the data acquisition time, a pink (with filters) or white beam can be used by removing the monochromator and other X-ray optics. In addition, image artefacts caused by imperfections in the beamline optics can also be avoided by using a pink or white beam. Taking advantage of such a high-flux beam, time-resolved X-ray imaging opened many new avenues of scientific endeavor. Examples include studies of fluid dynamics by X-ray imaging, whereby the inaccuracy from the refraction and reflection in visible light photography can be avoided (Lee et al., 2013); imaging the dynamic behavior under extreme conditions of materials that are nontransparent in visible light (Eakins & Chapman, 2014; Luo et al., 2012; Workman et al., 2010); and four-dimensional imaging with high-speed X-ray tomography revealing the dynamics of the internal three-dimensional structure of the samples (Walker et al., 2014; Baker et al., 2012).
than laboratory X-ray sources. Nevertheless, with monochromatic X-rays selected by crystal monochromators, only a small portion of the can be used, and the lowered inevitably limits the data acquisition speed. For many X-ray imaging applications, the requirement for temporal coherence is moderate, in which case the can be significantly increased by using multilayer monochromators with a larger spectral bandpass. However, the image quality often suffers from the irregular stripe modulations caused by imperfections in the multilayers (RackTo further enhance the image contrast and retrieve complementary information to absorption-contrast imaging, diverse X-ray phase-contrast imaging techniques have been developed over the last two decades. Importantly, several of the phase-contrast imaging techniques, such as the propagation-based, the grating-based and the speckle-based methods, are compatible with a polychromatic beam and can therefore benefit from the high et al., 1996; Momose et al., 2009; Wang, Kashyap, Cai et al., 2016).
of white beam radiation (WilkinsWhen using a white beam, current X-ray imaging systems face several challenges. The high
and high energy of the beam can easily damage the scintillator, objective and other instruments in the beam path through excessive heating or radiation damage. The scintillator screen can become electrostatically charged because of the intense As a result, dust particles are attracted to the screen, degrading the image in a non-systematic manner and in some cases appearing as over-saturated spots which cannot be flat-field corrected. Scattered high-energy photons directly striking the camera sensor can also lead to saturated white spots (commonly known as `zingers'). An X-ray imaging system designed in order to overcome some of these problems is presented in this paper. We hope it will benefit other research groups in the community facing similar challenges. Experimental results from a tomographic scan and from a scan using the speckle-based phase-contrast technique are presented as demonstrations of potential applications of the X-ray imaging system.2. Experimental setup and detector design
Tests of the detector were conducted at the bending magnet (BM) beamline B16 (Sawhney et al., 2010) at Diamond Light Source (Diamond). The BM source produces a continuous spectrum of X-rays with a of 8.4 keV, with the spectrum extending to beyond 100 keV on the high-energy side. The experimental layout is illustrated in Fig. 1. Raw emission, with no narrowband spectral filtering from monochromators or focusing optics, was used. A 0.5 mm-thick aluminium attenuator was used to protect the downstream components in the beam path, such as beryllium and Kapton windows, from excessive radiation. Several beam slits were placed at various positions along the beam path to reduce the stray light and heat load on components placed upstream of the camera. Filters with different thicknesses and materials could be moved into the beam, with motorized stages prior to the sample, to tune the peak energy of the broadband beam. Spectra simulated with the X-ray Oriented Program (XOP) (Sanchez del Rio & Dejus, 2004), using aluminium and copper filters, are given as examples of this approach in Fig. 1. The simulation shows that the would be dramatically reduced by increasing the thickness of the copper filter, and the peak energy of the spectra would shift towards a higher energy region. Pre-hardened X-rays may be preferred, especially for samples that are thick or contain high-atomic-number materials, as the low-energy photons only increase the absorbed dose in the samples yet contribute little to achieving image contrast. The BM source was 41 m from the beam slit, and 47.5 m from the detector, between which the sample to be imaged could be placed flexibly. The camera and optics housing were lined with a lead shield protecting them from scattered X-rays and to reduce the bright pixels caused by scattered high-energy photons.
The design of the detector (Fig. 1) follows the concept of the indirect detector approach (Bonse et al., 1991; Stampanoni et al., 2002; Koch et al., 1998), with the scintillator imaged by a CMOS camera via a lens. The camera and the lens are each independently mounted on a linear stage. By adjusting the positions of the lens and the camera, the geometric magnification can be varied continuously without the need to change to objectives with different magnifications. An additional goniometric stage is mounted in order to align the orientation of the camera sensor to the rotation axis of the sample stage for tomographic scans. With the current lens (50 mm f/2.8 Schneider macro lens) and motion-stages (150 mm and 25 mm travel range), the magnification can be varied between 1.7 and 5.1, shown as positions A and B in Fig. 2. Two radiographic images of a resolution sample (Micro-CT Bar Pattern NANO) at these two different geometrical magnifications are also given in Fig. 2. The patterns with linewidths of 4 and 2 µm are shown in the images, where we can see that the finer line pattern with a linewidth of 2 µm is at the limit of the resolution with the higher magnification setting. If a higher magnification is desired with the same overall camera dimensions, then a macro lens with shorter focal length can be used, as shown by the dashed lines in the plot in Fig. 2. Compared with microscope objectives with relatively small magnifications, a macro lens has a larger numerical aperture, and hence higher Long-working-distance microscope objectives with higher magnifications are good alternatives for applications that require even higher spatial resolution. The pco.edge camera has a pixel size of 6.5 µm and 2560 × 2160 pixels. The field of view (FOV) and effective pixel size changes along with the varying of magnification. This tradeoff between a larger FOV and higher spatial resolution is illustrated in the plot in Fig. 2. The maximum frame rate of the camera is 100 frames s−1 at full resolution. With a reduction of the total pixel number by using only a region of the sensor, the frame rate can be increased. Depending on the application, existing cameras on the market can reach frame rates as high as MHz and capture even single-bunch X-ray images (Etoh et al., 2003; Olbinado et al., 2017), with a sacrifice of either FOV or spatial resolution.
The scintillator used was LuAG:Ce from Crytur with a thickness of 80 µm and a diameter of 25 mm. A glassy carbon window was mounted in front of the scintillator to block visible light and minimize contamination of the et al. have shown that LuAG:Ce is one of the scintillator materials whose emission spectra is best matched to CMOS sensor (Olbinado et al., 2017). The thickness of the scintillator affects the spatial resolution and the conversion efficiency, which can be altered depending on the requirement of the experimental application. It should be noted that the scintillator can be replaced with others that have shorter decay times, such as LYSO:Ce or LuI3:Ce (Marton et al., 2014; Pidol et al., 2004), which are more suitable for ultrafast imaging.
by dust. OlbinadoThe scintillator was mounted on a 45° prism mirror cage, so that the lens and the camera could be mounted transverse to the X-ray beam direction allowing it to be protected from radiation damage caused by the X-rays that passed through the scintillator. Dry nitrogen gas with a positive gas pressure constantly flowed into the mirror cage with a slow speed so that the dust from the surrounding environment could not enter the mirror cage and accumulate on the charged scintillator screen during X-ray exposures. Gas flow is easier to handle and more cost-efficient than a vacuum environment for the scintillator. As can be seen from the example given in Fig. 3(a), without any gas flow, a significant amount of dust accumulated after using the camera with a pink beam for 38 h, which manifested as clouds of white spots in the X-ray image. The image quality was significantly degraded compared with that at the beginning of the test. In this case, the scintillator normally needed to be cleaned, or sometimes replaced. In Fig. 3(b), when the gas flow was on, the FOV was still clean after being used for 43 h. This gave a much more stable image background for long duration or high-throughput experiments, thereby enabling a better comparison against time for applications that involve dynamically changing samples. The different efficiencies of individual scintillators and the slight variation in camera focus when manually switching the scintillators can also bring inconsistency into the collected images, which is best avoided for most applications. Eliminating the need to change and clean scintillators during experiments also offers benefits of reduced costs and downtime.
3. Applications
3.1. Fast imaging
With the improved X-ray imaging system installed, we will be able to take full advantage of the much higher total et al., 1990; Lak et al., 2008), or time-resolved experiments.
of a white or pink beam compared with a monochromatic beam, and the required acquisition time can be significantly reduced. A more efficient use of beam time is therefore possible, as well as lessening the risk of artefacts from instabilities in fixation of sample or mechanical drifts during long exposures. This is especially valuable for experiments including tomographic scans and multi-dimensional raster scans. The high of the white beam and a camera that is compatible with such high radiation also makes it possible for fly scans (continuous rotation mode) (KalenderA tomographic scan was conducted at B16 during camera tests, and is shown in Fig. 4. Half of a human primary tooth was imaged with an exposure time of 50 ms for each projection with 601 projections taken over 180°. The effective pixel size was 3.8 µm. The beam was filtered by 0.5 mm Al and 0.5 mm Cu, leading to a peak X-ray energy of 43 keV as shown in Fig. 1. Fig. 4(a) shows a projection image from the scan. The image is flat-field and dark image corrected. The standard deviation of the background is 0.025. The inset shows part of the dentine in the tooth, where giant tubules or microcanals can be observed (Agematsu et al., 1990, 2005; Sumikawa et al., 1999). The enamel, dentine and pulp of the tooth are clearly distinguishable, thanks to edge-enhancements from phase contrast. Fig. 4(b) shows the three-dimensional rendering of the tomographic scan. Fig. 4(c) shows a transverse slice at the position indicated by the dashed line in Fig. 4(a), showing that the microcanals are aligned along the mesiodistal direction, in agreement with previous studies (Sumikawa et al., 1999; Agematsu et al., 2005). Fig. 4(d) is a vertical slice taken from the dotted line in Fig. 4(c), showing the pulp chamber and the propagating direction of the microcanals. From the inset with enhanced contrast, we can also see the structure of tubules in the dentine.
Previous studies on the microstructure of teeth have been mostly carried out using light microscopy or ). X-rays, on the other hand, can image the internal three-dimensional structure non-destructively. The sample can be preserved in a more natural state to avoid artefacts and damage that can be introduced by the sample preparation process; therefore, more realistic information about the sample can be extracted. Fast and large-FOV tomography can also be applied as a pre-scan before the destructive sample preparation process for other imaging tools with higher resolution, such as SEM, or more recently, ptychography (Zanette, Enders et al., 2015), so that sample cutting can be controlled and targeted at the region of interest.
(SEM). Both methods require the sample to be cut into thin slices, in order to observe the surface structure. Prior knowledge is required to find the region of interest for sample preparation. Drying and gold sputtering are also normally needed for SEM observations (Janda, 19953.2. High-energy phase-contrast imaging
When the white-beam from the BM is filtered, low-energy photons are absorbed and a spectrum with a higher average energy is achieved. Materials that have high atomic numbers can be imaged and characterized, for example, with phase-contrast imaging methods (Spanne et al., 1999; Wang, Kashyap, Cai et al., 2016; Diemoz et al., 2016; Willner et al., 2013). With multimodal phase-contrast imaging, results about phase, absorption and scattering can be obtained simultaneously, which offers complementary information to the material properties compared with commonly used attenuation-contrast imaging (Pfeiffer et al., 2008; Berujon, Wang et al., 2012; Modregger et al., 2016; Pagot et al., 2003). In the example given here, the speckle-based phase-contrast technique with one-dimensional scan mode has been applied (Wang et al., 2015, 2016), where a stack of sandpaper (P800) is used as the static diffuser. An image of the sandpaper is given in Fig. 5(a), with the random structures resembling `speckles'. The effective pixel size was 3.8 µm. In this experiment, 0.5 mm-thick Al and 1.5 mm-thick Cu filters were used, which gave a hardened spectrum with peak energy at 57.5 keV, as shown in Fig. 1. In this case, the visibility of the speckle pattern, defined as the normalized standard deviation here (Zanette, Zdora et al., 2015), is about 7%. The speckle size calculated as the FWHM of the autocorrelation is 12 µm. Other materials, such as steel wool (Wang, Kashyap, Cai et al., 2016), can also be used as diffuser, which can give a different visibility and speckle size.
The principle of the speckle-based phase-contrast imaging is to measure the refraction of the X-ray beam induced by the sample, by tracking the shift of the speckle pattern. With the proportional relation = , where and are the phase gradients in the horizontal and vertical, respectively, and are the refraction angles and λ is the wavelength of the X-rays, the phase shift Φ can be calculated by two-dimensional integration (Morgan et al., 2012; Berujon, Ziegler et al., 2012). Simultaneously, transmission and dark-field images can also be obtained, from the changes in the intensity and the visibility of the speckle pattern induced by the sample, revealing information about absorption and scattering (Berujon, Wang et al., 2012).
As a demonstration of the potential applications in geological science, we selected one picrite (an important chemical end-member magma) sample from Iceland. It is composed of olivine macrocrystals set in a microcrystalline matrix that also contains some olivine. The thickness of the sample is 530 µm. Three-dimensional textural information is key to understanding cooling histories, and also to contextualize the chemical information of the materials. Attenuation contrast works well for distinguishing between crystals and air for porous samples. However, because of the chemical similarity between the macrocrystal and the matrix there is very little difference in attenuation, which means that phase contrast is highly valuable (Polacci et al., 2010). The development of this imaging system allows for new dynamic experiments to be designed that test long-held assumptions regarding the origin and evolution of these and other magmas. As shown in Fig. 5, the multimodal images of the picrite sample are retrieved using the speckle-based technique. The transmission image is shown in Fig. 5(e), and it exhibits poor absorption contrast between the olivine and the host groundmass. While this image can be interpreted by eye, it is difficult to segment digitally. The phase contrast image in Fig. 5(d), integrated from the horizontal and vertical refraction angles in Figs. 5(b) and 5(c), produces better contrast which can be segmented more easily. Further work is under way to apply this technique to tomographic imaging, where virtual slices can be reconstructed without the need for preparing thin sections of the sample. Combined with tomography, multimodal images will allow quantitative retrieval of the complex refractive indices of the materials in the sample, and can be applied for material characterization and quantitative analysis (Zanette, Zdora et al., 2015). Multimodal imaging also offers the extra scattering information that traditional attenuation imaging cannot. For example, it can be used to study water transportation across construction materials (Yang et al., 2015). With the development of the improved white-beam detecting system, such scientific applications are also possible at beamline B16. By enhancing the applicability and utility of BM beamlines for imaging, synchrotron X-ray users have expanded the opportunity to conduct a greater number of experiments, by providing an alternative to heavily over-subscribed insertion-device-based imaging beamlines.
4. Conclusion
A white-beam imaging system has been designed, tested and established at the B16 beamline at Diamond. Continuous nitrogen flow in the space between the scintillator and the reflecting mirror has proved to be an effective way of preventing accumulation of dust particles on the scintillator, which would otherwise significantly degrade image quality over time. Similar concepts can be applied to other existing X-ray imaging systems. The spatial resolution and FOV can be adjusted continuously by varying the positions of the lens and the camera with motorized stages, without the need to change to objectives with different magnifications. Demonstrations for fast tomography and multimodal phase-contrast imaging show that the X-ray imaging system developed will find wide applications, such as in the biomedical and geological sciences. With the further development of the detector, the white-beam imaging system will be routinely used, for example to study dynamic behavior of samples in fly-scan mode. The design of the imaging system is cost-efficient and flexible, and can be implemented easily on other synchrotron beamlines that provide intense X-ray flux.
Acknowledgements
This work was carried out with the support of the Diamond Light Source Ltd, UK. The authors wish to thank Matthew Pankhurst and Thor Thordarson for access to the geological samples and for analyzing the results. We would like to acknowledge Andrew Malandain and Hiten Patel for their technical assistance, and Ian Pape and Oliver Fox for their assistance during beam time at B16.
References
Agematsu, H., Abe, S., Shiozaki, K., Usami, A., Ogata, S., Suzuki, K., Soejima, M., Ohnishi, M., Nonami, K. & Ide, Y. (2005). Bull. Tokyo Dent. Collect. 46, 7–15. CrossRef Google Scholar
Agematsu, H., Watanabe, H., Yamamoto, H., Fukayama, M., Kanazawa, T. & Miake, K. (1990). Bull. Tokyo Dent. Collect. 31, 163–173. CAS Google Scholar
Baker, D. R., Brun, F., O'Shaughnessy, C., Mancini, L., Fife, J. L. & Rivers, M. (2012). Nat. Commun. 3, 1135. Web of Science CrossRef PubMed Google Scholar
Berujon, S., Wang, H. & Sawhney, K. (2012). Phys. Rev. A, 86, 063813. Web of Science CrossRef Google Scholar
Bérujon, S., Ziegler, E., Cerbino, R. & Peverini, L. (2012). Phys. Rev. Lett. 108, 158102. Web of Science PubMed Google Scholar
Bonse, U., Nusshardt, R., Busch, F., Pahl, R., Kinney, J. H., Johnson, Q. C., Saroyan, R. A. & Nichols, M. C. (1991). J. Mater. Sci. 26, 4076–4085. CrossRef CAS Google Scholar
Diemoz, P. C., Bravin, A., Sztrókay-Gaul, A., Ruat, M., Grandl, S., Mayr, D., Auweter, S., Mittone, A., Brun, E., Ponchut, C., Reiser, M. F., Coan, P. & Olivo, A. (2016). Phys. Med. Biol. 61, 8750–8761. CrossRef Google Scholar
Eakins, D. E. & Chapman, D. J. (2014). Rev. Sci. Instrum. 85, 123708. Web of Science CrossRef PubMed Google Scholar
Etoh, T. G., Poggemann, D., Kreider, G., Mutoh, H., Theuwissen, A. J. P., Ruckelshausen, A., Kondo, Y., Maruno, H., Takubo, K., Soya, H., Takehara, K., Okinaka, T. & Takano, Y. (2003). IEEE Trans. Electron Devices, 50, 144–151. CrossRef Google Scholar
Janda, R. (1995). Biomaterials, 16, 209–217. CrossRef CAS Google Scholar
Kalender, W. A., Seissler, W., Klotz, E. & Vock, P. (1990). Radiology, 176, 181–183. CrossRef CAS Google Scholar
Koch, A., Raven, C., Spanne, P. & Snigirev, A. (1998). J. Opt. Soc. Am. A, 15, 1940–1951. Web of Science CrossRef CAS Google Scholar
Lak, M., Néraudeau, D., Nel, A., Cloetens, P., Perrichot, V. & Tafforeau, P. (2008). Microsc. Microanal. 14, 251–259. CrossRef CAS Google Scholar
Lee, J. S., Weon, B. M. & Je, J. H. (2013). J. Phys. D, 46, 494006. CrossRef Google Scholar
Luo, S. N., Jensen, B. J., Hooks, D. E., Fezzaa, K., Ramos, K. J., Yeager, J. D., Kwiatkowski, K. & Shimada, T. (2012). Rev. Sci. Instrum. 83, 073903. Web of Science CrossRef PubMed Google Scholar
Marton, Z., Nagarkar, V. V., Miller, S. R., Brecher, C., Bhandari, H. B., Kenesei, P., Ross, S. K., Almer, J. D. & Singh, B. (2014). J. Phys. Conf. Ser. 493, 012017. CrossRef Google Scholar
Modregger, P., Cremona, T. P., Benarafa, C., Schittny, J. C., Olivo, A. & Endrizzi, M. (2016). Sci. Rep. 6, 30940. CrossRef Google Scholar
Momose, A., Yashiro, W., Maikusa, H. & Takeda, Y. (2009). Opt. Express, 17, 12540–12545. Web of Science CrossRef PubMed CAS Google Scholar
Morgan, K. S., Paganin, D. M. & Siu, K. K. W. (2012). Appl. Phys. Lett. 100, 124102–124104. CrossRef Google Scholar
Olbinado, M. P., Just, X., Gelet, J.-L., Lhuissier, P., Scheel, M., Vagovic, P., Sato, T., Graceffa, R., Schulz, J., Mancuso, A., Morse, J. & Rack, A. (2017). Opt. Express, 25, 13857–13871. CrossRef CAS Google Scholar
Pagot, E., Cloetens, P., Fiedler, S., Bravin, A., Coan, P., Baruchel, J., Härtwig, J. & Thomlinson, W. (2003). Appl. Phys. Lett. 82, 3421–3423. Web of Science CrossRef CAS Google Scholar
Pfeiffer, F., Bech, M., Bunk, O., Kraft, P., Eikenberry, E. F., Brönnimann, C., Grünzweig, C. & David, C. (2008). Nat. Mater. 7, 134–137. Web of Science CrossRef PubMed CAS Google Scholar
Pidol, L., Kahn-Harari, A., Viana, B., Virey, E., Ferrand, B., Dorenbos, P., de Haas, J. T. M. & van Eijk, C. W. E. (2004). IEEE Trans. Nucl. Sci. 51, 1084–1087. CrossRef CAS Google Scholar
Polacci, M., Mancini, L. & Baker, D. R. (2010). J. Synchrotron Rad. 17, 215–221. CrossRef CAS IUCr Journals Google Scholar
Rack, A., Weitkamp, T., Riotte, M., Grigoriev, D., Rack, T., Helfen, L., Baumbach, T., Dietsch, R., Holz, T., Krämer, M., Siewert, F., Meduňa, M., Cloetens, P. & Ziegler, E. (2010). J. Synchrotron Rad. 17, 496–510. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sanchez del Rio, M. & Dejus, R. J. (2004). Proc. SPIE, 5536, 171–174. CrossRef Google Scholar
Sawhney, K. J. S., Dolbnya, I. P., Tiwari, M. K., Alianelli, L., Scott, S. M., Preece, G. M., Pedersen, U. K., Walton, R. D., Garrett, R., Gentle, I., Nugent, K. & Wilkins, S. (2010). AIP Conf. Proc. 1234, 387–390. CrossRef CAS Google Scholar
Spanne, P., Raven, C., Snigireva, I. & Snigirev, A. (1999). Phys. Med. Biol. 44, 741–749. Web of Science CrossRef PubMed CAS Google Scholar
Stampanoni, M., Borchert, G., Wyss, P., Abela, R., Patterson, B., Hunt, S., Vermeulen, D. & Rüegsegger, P. (2002). Nucl. Instrum. Methods Phys. Res. A, 491, 291–301. Web of Science CrossRef CAS Google Scholar
Sumikawa, D. A., Marshall, G. W., Gee, L. & Marshall, S. J. (1999). Pediatr. Dent. 21, 439–444. CAS Google Scholar
Walker, S. M., Schwyn, D. A., Mokso, R., Wicklein, M., Müller, T., Doube, M., Stampanoni, M., Krapp, H. G. & Taylor, G. K. (2014). PLoS Biol. 12, e1001823. Web of Science CrossRef PubMed Google Scholar
Wang, H., Kashyap, Y., Cai, B. & Sawhney, K. (2016). Sci. Rep. 6, 30581. CrossRef Google Scholar
Wang, H., Kashyap, Y. & Sawhney, K. (2015). Opt. Express, 23, 23310–23317. CrossRef CAS Google Scholar
Wang, H., Kashyap, Y. & Sawhney, K. (2016). Sci. Rep. 6, 20476. Web of Science CrossRef PubMed Google Scholar
Wilkins, S. W., Gureyev, T. E., Gao, D., Pogany, A. & Stevenson, A. W. (1996). Nature (London), 384, 335–338. CrossRef CAS Web of Science Google Scholar
Willner, M., Bech, M., Herzen, J., Zanette, I., Hahn, D., Kenntner, J., Mohr, J., Rack, A., Weitkamp, T. & Pfeiffer, F. (2013). Opt. Express, 21, 4155–4166. Web of Science CrossRef PubMed Google Scholar
Workman, J., Cobble, J., Flippo, K., Gautier, D. C., Montgomery, D. S. & Offermann, D. T. (2010). Rev. Sci. Instrum. 81, 10E520. CrossRef Google Scholar
Yang, F., Griffa, M., Bonnin, A., Mokso, R. C. D. I. B., DI Bella, C., Münch, B., Kaufmann, R. & Lura, P. (2015). J. Microsc. 261, 88–104. CrossRef CAS Google Scholar
Zanette, I., Enders, B., Dierolf, M., Thibault, P., Gradl, R., Diaz, A., Guizar-Sicairos, M., Menzel, A., Pfeiffer, F. & Zaslansky, P. (2015). Sci. Rep. 5, 9210. Web of Science CrossRef PubMed Google Scholar
Zanette, I., Zdora, M.-C., Zhou, T., Burvall, A., Larsson, D. H., Thibault, P., Hertz, H. M. & Pfeiffer, F. (2015). Proc. Natl Acad. Sci. USA, 112, 12569–12573. CrossRef CAS Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.