q2xafs2017 workshop
Improving the quality of
dataaHigh Energy Accelerator Research Organization (KEK), Tsukuba, Japan, bElettra-Sincrotrone Trieste, Basovizza, Italy, cDiamond Light Source, Oxfordshire, UK, dUniversity of Notre Dame, Notre Dame, IN, USA, eEuropean Synchrotron Radiation Facility, Grenoble, France, and fPaul Scherrer Institute, Villigen, Switzerland
*Correspondence e-mail: sakura@esrf.fr
Following the Q2XAFS Workshop and Satellite to IUCr Congress 2017 on `Data Acquisition, Treatment, Storage – quality assurance in in operando. Advice on how to avoid or deal with these problems and a few good practice recommendations are reported, including how to correctly report results.
spectroscopy', a summary is given of the discussion on different aspects of a experiment that affect data quality. Some pertinent problems ranging from sources and minimization of noise to harmonic contamination and uncompensated monochromator glitches were addressed. Also, an overview is given of the major limitations and pitfalls of a selection of related methods, such as photon-out spectroscopies and energy-dispersive and of increasingly common applications, namely studies at high pressure, and time-resolved investigations of catalystsKeywords: XAFS; quality of XAFS data; sources of noise in XAFS; harmonic rejection in XAFS; good practice in XAFS experiments.
1. Introduction
The et al., 1971; Lee et al., 1981), who developed the method, and evolved to include thousands of experts in a plethora of different scientific areas, including biology, geology, chemistry, cultural heritage and many more, often with little or no academic background in physics. In parallel, methods have also evolved, triggered by facility scientists developing their in-house research programmes, while accelerator-based sources developed into higher-brilliance third-generation and diffraction-limited storage rings and free-electron lasers. Today we see more complex experimental setups aimed at a more exhaustive investigation of systems, including more relevant composition, space and time resolution, under a variety of external stimuli and adopting new XAFS-based methods based on photon-out spectroscopies. Many of these developments are accompanied by larger data sets and more complex data reduction protocols. In parallel, is becoming an increasingly popular characterization technique, used more and more by non-experts. is also becoming available on the laboratory scale, providing easier access to this method by non-expert users (Seidler et al., 2014; Padamati et al., 2017). To maintain modern applications accessible to the larger user community without compromising the accuracy of the information, it is important to make all aware of the outstanding issues and to continue efforts toward experimental standardization. For a review on the history of the efforts devoted to establishing standards and criteria, see, for example, Oyanagi & Stern (2017).
(X-ray absorption fine-structure) community has been constantly growing since the first quantitative work exploiting synchrotron radiation, almost 50 years ago. We have seen a gradual transition in the nature of this community, that started with a few scientists, mainly physicists (SayersIt is in this spirit that a session of the Q2XAFS1 Workshop Satellite to IUCr Congress 2017 was devoted to a discussion on the outstanding and unsolved problems affecting data quality. Issues closely related to beamline optics and detection were discussed at large, with focus on sources and minimization of noise, harmonic contamination and uncompensated monochromator glitches. This discussion is summarized in §2. In §3 we report some issues that were addressed on the limitations and difficulties affecting a selection of XAFS-related methods, and in particular energy-dispersive and photon-out spectroscopies. Good practice recommendations are reported here, including how to correctly report results. Finally, in §4 we focus on two increasingly common applications at high-brilliance sources, i.e. high-pressure studies and time-resolved investigations of catalysts. We report on recent developments and offer some advice on how to avoid or deal with inherent difficulties. There are of course other possible sources of error that were not discussed at the workshop. For example:
(i) Sample thickness and et al., 1982).
and effects on the amplitude of (Goulon(ii) Non-linearity of detectors when working at the high-flux condition: ionization chambers (Pettifer et al., 1999), fluorescence detectors (Woicik et al., 2010; Walko et al., 2011).
(iii) Radiation damage and associated spectral changes (Holton, 2007; Bertrand et al., 2015).
(iv) Experimental challenges with dilute systems and complex molecules (Chantler et al., 2012a).
Many of these issues and how to deal with them are addressed in the comprehensive paper by Chantler et al. (2012b), an outcome of the 2011 Q2XAFS Workshop. Finally, the book by G. Bunker (Bunker, 2010) also provides, in ch. 3, practical issues of sample preparation, experimental methodology, choice of detectors, sources of measurement errors and measures of how to avoid them.
2. Sources of errors related to beamline optics and detection
2.1. Sources and minimization of noise
`Noise' in e.g. counting statistics, (ii) electronic noise, both analog and digital, (iii) X-ray beam instability from the X-ray source, and (iv) mechanical motion of optical elements or sample. In terms of data quality, it is useful to generalize this to include other artefacts such as slowly varying thermal fluctuations in monochromators, mirrors, etc., sample issues, e.g. inhomogeneity, crystallinity, etc., and other monochromator or mirror issues.
includes contributions from (i) stochastic noise,Stochastic noise, e.g. counting statistics, is generally only a problem in low-count-rate and/or high-background situations, e.g. photon-out spectroscopies such as dilute fluorescence, etc. Note that even with high absolute count rates, scattered or fluorescence background may dominate and the actual `signal' count rate might be relatively low. In these common cases, the statistical fluctuation in background counts generally dominates over fluctuations in signal photon count rate. This has recently been covered in detail in explorations of the limits of fluorescence detection at high-flux beamlines (Heald, 2015).
To optimize the signal-to-noise ratio [S/N = Ns/(Ns+Nb)1/2 where Ns and Nb are the counts of the signal and of the background, respectively], one wants to reduce the background signal, but generally the signal will be reduced by these measures as well. This can be accomplished by the use of X-ray filters and Soller slits, photon-counting detectors (e.g. Si or Ge solid state detectors) allowing energy discrimination, or the use of crystal or multilayer X-ray analyzers. Each of these approaches has its own strengths and limitations (see, for example, Bunker, 2010; Heald, 2015).
Electronic noise or artefacts can also contaminate the signal, but can generally be minimized by improving shielding of detectors and pre-amplifiers and eliminating ground loops due to shared signal paths. Another source of error includes pulse-counting issues such as pulse pile-up, dead-time corrections, or analyser window shifts for high count rates.
Transmission data can also be susceptible to noise if the `dark current' offset voltage is improperly calibrated. This offset is necessary to avoid low-current nonlinearity in voltage-to-frequency converters, but, if not removed after voltage-to-frequency conversion, will introduce nonlinearities of its own as intensity fluctuations will not be correctly normalized.
X-ray beam instability from the X-ray source is an additional source of error, affecting data quality. Insertion devices might cause issues more often than bending-magnet sources. For instance, the high heat load from wigglers can cause thermal stability issues with beamline components, while the peaked spectrum from undulators either require scanning the undulator gap (possibly increasing noise) or, if the gap is tapered to broaden the spectrum, spatial structure can be introduced in the beam that can also introduce artefacts as a function of energy.
Mechanical stability is very important to minimize point-to-point intensity fluctuations, and thermal stability essential to avoid long-term data distortion. Additionally, monochromator `glitches', i.e. spurious reflections from monochromator crystals, can cause unwanted changes in intensity, beam direction, harmonic content or polarization. This can depend strongly on sample details with inhomogeneous samples much more sensitive to these artefacts. While some care in sample preparation can help, sometimes the only real solution is to use a different monochromator crystal set with different φ orientation. Many beamlines have glitch spectra mapped out for users (Stanford Synchrotron Radiation Lightsource, 1999).
A very important issue which is often underestimated is detector linearity. Intensity fluctuations due to source position/angle or X-ray optics would ideally cancel out if both Is and incident I0 are measured and then the ratio taken. If the response is nonlinear due to detectors, electronics, harmonic contamination or sample issues, fluctuations will not cancel resulting in increased noise or other artefacts. This is important to minimize in beamline design, beamline configuration for a particular experiment, and experimental setup by users. This is important in transmission as well as fluorescence measurements, and for ion chambers can be exacerbated by ion-chamber recombination for very intense X-ray beams.
2.2. Contamination from harmonics
One of the important factors to determine the quality of et al., 1976; Sainctavit et al., 1988). The primary X-ray beam used for measurements is usually monochromated by double-crystal monochromators (DCMs) of Si(111) or Si(311), and the higher harmonics should be rejected by mirrors. Harmonic contamination in the X-ray beam can produce distortions in the spectrum that can be misinterpreted, yielding incorrect results.
spectra is how to handle and reject harmonics of the primary X-ray beam (BonseIt is not easy to become aware of harmonic contaminations in I0 normalization (i.e. monochromator glitches or other features present in the energy dependence of the I0 are not fully normalized out) and non-efficient harmonics rejection should be suspected. Fig. 1 illustrates the XANES of the spectra recorded on a reference Ti foil under different experimental conditions. The dashed line is measured using a detuning of 50% of the DCM, while the solid line on the other hand is measured using a fully tuned DCM. The comparison shows that in the case of a fully tuned DCM a spike is visible at ∼5060 eV, caused by the lack of normalization of a monochromator glitch. In the regime, oscillations are small and harmonic contamination, particularly in the high-k region, can seriously damage data analysis.
of samples that are being studied, while it is relatively easy to notice those of standard materials such as commercially available metal foils and oxide powders. One simple way is to examine the spectra carefully. If some reproducible non-statistical noise, or spike feature, is found in the spectra, this could be due to a lack of properThe Fourier transforms (FTs) of Ti K-edge on Ti foil are shown in Fig. 2. In the 50% detuned case (dashed line), the peak at R ≃ 2.5 Å in the FT can be well reproduced using a single Ti–Ti path, with an R-factor of 0.009. However, in the fully tuned case (solid line), besides the presence of an unphysical signal between 1 and 1.8 Å, a ghost shoulder structure is present around 2.2 Å and the peak can no longer be correctly fitted by a single Ti–Ti path.
This example illustrates the importance of carefully planning the best optical configuration to achieve efficient harmonic rejection and obtain correct and reliable −5 or less in the of photons at the energy of the third harmonic with respect to the of photons at the fundamental energy. Detuning to this extent by the monochromator is not feasible. Therefore a pair of harmonic rejection mirrors, with suitable choice of coatings and operational grazing angles, is mandatory. Information on the amount of harmonic rejection by different mirror coatings/substrates are given by Henke et al. (1993). For guidelines to reduce harmonic content, see Tran et al. (2003) and Glover & Chantler (2009).
spectra. Here we have illustrated the effect of insufficient harmonic rejection on a thin homogeneous metallic foil, often used as reference sample for However, as already mentioned above, the effects on `real' samples can be much more difficult to detect. Therefore, it is good practice to assure a degree of harmonic rejection that will allow to work safely on any kind of sample. The generally accepted level of rejection is to achieve a ratio of 102.3. Non-compensated monochromator glitches
The occurrence of multiple reflections within a monochromator crystal gives rise to sharp dips or spikes in the intensity of the diffracted X-ray beam at specific energies. When collecting monochromator glitches are usually removed from the spectra by normalizing the signal transmitted by the sample to the incoming beam intensity. Unfortunately, this approach does not always work. Reasons for this include the presence of harmonics in the beam, inhomogeneity and non-uniformity of the sample thickness, and the non-linearity of the two detectors used in the measurements (Stern & Lu, 1982; Bridges et al., 1992; Li et al., 1994).
data, these so-calledAdditionally, it has been recently found that, when collecting the fluorescence signal for very dilute samples using a large-area multi-element detector, these glitches do not compensate (Sutter et al., 2016). This occurs even after carefully tackling all the experimental issues mentioned above. A detailed investigation of the scattering contributions revealed that this effect is due to changes in the spatial distribution of the quasi-elastically scattered photons over the detector when passing through a glitch. As shown in Fig. 3(a), the intensity drop of the integrated scattering strongly depends on the vertical position of the element in the detector. In very dilute samples, this contribution is significant and cannot be completely separated from the fluorescence signal, thus preventing proper normalization of the glitch features in the spectral data (see Fig. 3b).
A fitting procedure treating coherent and Compton scattering developed by Sutter et al. (2016) has shown that the spatial distribution of the quasi-elastically scattered intensity induces changes in the polarization of the incident beam on the sample as a monochromator glitch is traversed. In particular, when the vectors of the extra reflections do not lie in the of the main diffracted beam, the coupling between the incident and the diffracted beams within the crystal induces a change in the polarization (mainly a rotation of the polarization). This result has been corroborated by multiple-beam theory calculations. Post-processing routines to compensate the glitches in data could be developed that include corrections due to these effects.
Although the glitches are an intrinsic phenomenon of the monochromatization process and cannot be avoided, the azimuthal angle of the monochromator crystals can be carefully optimized to minimize the number of glitches appearing within the working energy range. In that respect, some efforts have been made to predict the best configuration for each type of crystal (van der Laan & Thole, 1988; Tang et al., 2015).
3. Related methods: limitations and recommendations
3.1. Photon-in/photon-out spectroscopy
Photon-in/photon-out spectroscopy may be defined for an experimental setup that enables measurement of incoming and emitted (scattered) photons with an energy bandwidth of the order of the core-hole lifetime broadening. This distinguishes it from fluorescence-detected absorption spectroscopy using conventional solid state detectors with energy bandwidth approximately two orders of magnitude larger than the core-hole lifetime broadening. High-energy-resolution fluorescence-detected (HERFD) X-ray absorption spectroscopy provides sharper spectral features but a priori does not record a spectrum that is proportional to the photoelectric absorption (Carra et al., 1995). Excitations into strongly localized orbitals (pre-edges) may show strong deviations from the in HERFD-XAS while excitations into delocalized orbitals (bands that give rise to the main edge) may appear in HERFD-XAS as an with increased spectral resolution (Glatzel et al., 2013).
HERFD-XAS is always distorted by over-absorption (or incident beam self-absorption) just as standard fluorescence-detected absorption spectroscopy (Bunker, 2010). In many cases, this distortion cannot be corrected [using, for example, the FLUO code (Haskel, 1999)] because the sample composition and other experimental parameters are not sufficiently well defined. In order to understand whether a spectral change arises from this experimental artefact or a real change of the we propose to fit the spectra using the general formula for fluorescence-detected absorption spectroscopy (Fig. 4) (Bianchini & Glatzel, 2012). In this case, instead of correcting the distortion it is reproduced using fitting parameters. While this procedure does not provide the correct spectrum, it may help to identify the cause of a spectral change. We note that the correction and fitting shown in Fig. 4 can likely be improved by using a formalism that treats the angles of the incoming and outgoing beam in a more sophisticated fashion.
Some more remarks:
(i) A HERFD-XAS spectrum always requires the chosen emission energy to be reported. As the emission energy may show a chemical dependence, it must be mentioned whether the emission energy was kept constant or changed to stay on the maximum intensity.
(ii) The full captured scattering angle must always be reported as it is an important parameter for photon-in/photon-out experiments.
(iii) The scattering
for the different fluorescence lines varies by more than three orders of magnitude. The required must be considered for detector linearity. Higher harmonics in the incident beam may give background arising from strong fluorescence lines in the energy range of a weak line of interest.(iv) X-ray emission spectroscopy always depends on the e.g. KL-edge) (Glatzel et al., 2003) and background from Compton scattering.
even when chosen well above the because of the onset of multiple electron excitations ((v) The instrumental energy bandwidth should be reported. This requires some care as the
has a different angular dependence than the inelastic process that is recorded in the actual experiment. In the case of a multi-analyzer crystal instrument, each analyzer is likely to have a different bandwidth because of the limited reproducibility of the manufacturing process.3.2. Energy-dispersive XAS
The energy-dispersive spectrometer employs a curved crystal to disperse and focus a polychromatic fan of X-rays onto the sample (Matsushita & Phizackerley, 1981). The transmitted beam is detected by a position-sensitive detector where energy is correlated to position. Besides following time-dependent phenomena (Mathon et al., 2016), these spectrometers can be used for very specific applications (Pascarelli & Mathon, 2010, and references therein; Torchio et al., 2016; Ihli et al., 2017).
The energy bandwidth ΔE diffracted by the crystal is proportional to the variation along the beam footprint, multiplied by the cotangent of the This leads to the first limitation due to a reduced ΔE at low energies: often only XANES is acquired at E < 7 keV.
However, the most important limitation of EDXAS is linked to having a polychromatic beam on the sample, so photon-out spectroscopies are not applicable.
Another intrinsic limitation stems from the fact that the i.e. SAXS), which destroys energy resolution and may introduce artefacts in the spectra. This scattering can be eliminated by using a filter (Hagelstein et al., 1998), or parallel detection is given up and a monochromatic beam, selected by a slit on the polychromatic fan, is used to measure the spectrum in a step-by-step fashion (Pascarelli et al., 1999).
spectrum is acquired as a one-dimensional image on the position-sensitive detector. In particular, the energy-direction correlation established at the polychromator must be preserved all the way to the detector, which poses restrictions on sample microstructure to avoid scattering (The typical illustrates an example of a pellet containing Ge micrometer-sized powder in a boron nitride (BN) matrix in a 1:10 ratio. Data have been acquired in transmission mode, using both the step-by-step variant of EDXAS (Pascarelli et al., 1999) and ion chambers to measure I0 and I1, and in the conventional EDXAS mode, with parallel acquisition of the full spectrum on the position-sensitive detector. The figure illustrates that the latter data are affected by strong broadening of the XANES features, due to small-angle scattering from the sample which perturbs the energy–direction correlation established at the polychromator. This scattering is often due to the matrix (i.e. BN) of the pellet. Heterogeneous catalysts are another class of very challenging samples, often because of diffusion from the porous microstructure of the support (i.e. zeolite).
pellet, for example, is very difficult to handle. Fig. 5Another related problem is the high sensitivity to defects in optical elements and samples, enhanced by the increasing coherence length of X-rays from lower-emittance storage rings. In particular, the surface quality of the polychromator crystal plays a major role. Finally, the method is very sensitive to beam instabilities, especially if I0 and I1 are not measured simultaneously, but the great advances in source stability nowadays make this a minor problem.
The advent of lower-emittance storage rings will bring additional challenges to EDXAS. Since the spectrum is acquired as a one-dimensional image, the larger horizontal coherence length and its negative effects on the spatial et al., 2016)]. On the other hand, the sharper energy spectrum of undulator emission is not expected to limit the effective energy bandwidth of the data, which will likely remain limited by the quality of the polychromator crystals. Diffraction-limited storage rings are expected to boost applications that exploit the unique capacity of EDXAS to acquire a full spectrum in a single shot, where the total number of photons per X-ray bunch, more than is the relevant parameter.
of the beam will affect EDXAS to a greater extent than other XAS-based methods. Also, the reduction in horizontal source size will not translate to smaller horizontal spot sizes at the facilities which are already today at the diffraction limit [such as ID24 at the ESRF (Pascarelli4. Increasingly common applications
4.1. High-pressure XAS
Matter undergoes changes in its physical, chemical and structural characteristics when subject to high pressures. Under pressure, atoms are forced closer in a smaller volume and the energy of the atomic bonds is changed. In addition, pressure serves as a tool for synthesizing new materials and is especially important in the study of the rocks and minerals constituting the interior of the Earth and other planets. High-pressure research has been progressing rapidly in the last decades thanks to the concomitant development of high-pressure technology and to the increased accuracy of probing methods (Mao et al., 2016). Synchrotron sources offer now a large portfolio of techniques that can be applied to understand a material's behaviour under extreme environment from all aspects. X-ray absorption spectroscopy is one of these techniques. Diamond anvil cells (DACs) are the most common device for pressure generation (Eremets, 1996) and are widely used for at high pressure. They can generate conditions above 750 GPa (Dubrovinsky et al., 2015). The concept of a DAC consists of forcing together two flawless diamonds against a microgram-sized sample (Eremets, 1996).
An issue related to the use of DAC for . The main solution to this issue is to reduce the thickness of the diamonds using a combination of fully perforated diamond as diamond backing plate with a miniature anvil and a partially perforated diamond. In this way the diamond thickness along the X-ray path can be reduced down to less than 1 mm maintaining at the same time a good mechanical strength to reach pressures in excess of 100 GPa (Dadashev et al., 2001). As an alternative, the DAC can be oriented so that X-rays travel through a low-Z gasket such as Be (Itié et al., 2007). The fluorescence signal is collected at 90° with respect to the incident beam. Besides safety issues related to machining of Be, the main drawback of this geometry is that the beam path through the gasket changes with pressure.
regards the high absorption of the diamond anvils at energies below 7 keV. The transmission of 4 mm of diamonds, corresponding to the total thickness of a couple of standard anvils, is shown in Fig. 6The other and principal technical issue related to the use of DACs for et al., 2003; Ishimatsu et al., 2012). These consist of randomly oriented single-phase diamond grains of several tens of nanometers. Since the grains are randomly oriented, Bragg's law is always satisfied independently of the energy, therefore the intensity of the diffracted X-ray changes moderately with energy and consequently the NPD imparts a smooth background to the absorption profile.
is the of the diamond anvils. In fact, in the large energy range in which the is measured, Bragg conditions are satisfied for the of the diamond causing the X-rays to be diffracted by the anvils. This removes photons at precise energy values leading to large dips in the transmitted appearing as large peaks in the spectra. The ultimate solution to this is to use nano-polycrystalline diamond (NPD) anvils instead of single-crystal diamond anvils (Irifune4.2. Transient studies of catalysts
Using hard X-ray (>4.5 keV) based techniques, such as i.e. in a catalytic reactor. Such studies have been made since the early 1980s at synchrotron light sources and have led to an understanding of structure–performance relationships of many heterogeneous catalytic systems (Bare & Ressler, 2009). Many such studies (at least on the higher weight loaded samples) can nowadays potentially be carried out using a commercial laboratory-based X-ray spectrometer.
to study catalytic reaction mechanisms has a major advantage that catalytic materials can be studied under operating conditions,To rationally design and improve catalytic processes, an understanding of the et al. (Bravo-Suárez et al., 2008) showed that the observation of a specific structure by under operating conditions does not necessarily mean that this structure is involved in the catalytic This structure, resembling a particular structural site, might well be a spectator species to the catalytic reaction. Transient which can only be performed at a synchrotron, allows the rate of structural change to be measured. When the rate of structural change matches the rate of the overall reaction as measured for example using a then this structure, or alternatively called active site, is involved in the rate-limiting step.
including reaction intermediates is essential. In a seminal paper, OyamaThe question then arises, what time resolution is needed to decipher the rate-limiting step in a heterogeneous catalytic reaction? When one considers the rates of important industrial catalytic processes, one will observe that these rates rarely exceed 10 ms. For example, the rate of ammonia synthesis is of the order of a second (Nørskov et al., 2009). As a consequence, the spectroscopic technique employed needs to have an ultimate time resolution of a few milliseconds to catch the species involved in the rate-limiting step. Time-resolved with this time resolution can for example be obtained with energy-dispersive (§3.2) or with the quick-EXAFS technique (Müller et al., 2016). Both techniques can be applied for in situ studies in transmission mode, whereas only quick-EXAFS can be combined with fluorescence/emission detection, by using for example a PIPS diode. In the case of quick-EXAFS, the time resolution that can ultimately be achieved is not limited by the mechanics/driving mechanism of the monochromator system but rather by the response time of the detectors used. When one considers that the collection of an spectrum takes about 10 ms, and that such an spectrum consists of 1000 eV where individual data points need to be 1 eV apart, one sees that the response time of the detectors needs to be at least 10 ms per 1000 energy points = 10 µs per energy point. PIPS detectors, used for the detection of the fluorescence, typically have a sub-microsecond rise-time, whereas ionization chambers, used for the detection of the incoming and transmitted X-rays, are limited by the drift time of the ions and typically have a response time of hundreds of microseconds. Gridded ionization chambers, where a grid is placed between the high-voltage plate and the ground, have a rise time of a few microseconds (Müller et al., 2013) and overcome this limitation. The next step to be taken is to improve the response time of the current amplifier, which is on the order of 5–10 µs for gains of 6 and 7.
When one uses a dispersive crystal spectrometer to detect X-ray emission spectra with a resolution of around 1 eV in a single shot (Kopelent et al., 2016), at (fixed incident) energies just before the X-ray then time-resolved resonant X-ray emission spectroscopy can be performed with sub-second time resolution and has a high sensitivity to small changes. This technique was recently used to determine small changes in the redox state of Ce during CO oxidation (Kopelent et al., 2015). Active Ce3+ species in a ceria-supported platinum catalyst during CO oxidation are short-lived and therefore cannot be observed under steady-state conditions. Using resonant X-ray emission spectroscopy it could be shown that the initial rate of Ce3+ formation under transient conditions is quantitatively correlated to the overall rate of CO oxidation under steady-state conditions and thus that ceria reduction is a kinetically relevant step in CO oxidation, whereas a fraction of Ce3+ was present as spectators.
5. Conclusions and outlook
We report here a summary of the problems discussed at the Q2XAFS Workshop and Satellite to IUCr Congress 2017 on different aspects of a
experiment that affect data quality. The aim is to maintain the larger community informed on the outcome of the meeting, and not to give a comprehensive overview of the field. From the selection of topics reported here, we see that a few outstanding problems persist and ways to overcome some of them are addressed. Novel developments in methods or in the field of two increasingly common applications are presented, coupled to good practice recommendations.In the coming years, with the arrival on the synchrotron scene of diffraction-limited storage rings, and with the high-energy free-electron lasers coming into full operation, our
community will be confronted with important challenges. The continuing growth of a strong community will rely more and more on our ability to maintain close communication between `method developers' and inexperienced users, which, given the multidisciplinary nature of the technique, are expected to span a wider and wider horizon of different scientific domains. In this context, implementing a more robust strategy for the application of standards and criteria will be increasingly important and desirable.Footnotes
1The Q2XAFS acronym was coined at the First International Workshop on `Improving data Quality and Quantity for experiments', that took place on 8–9 December 2011 at Photon Factory, Tsukuba, Japan.
Acknowledgements
The authors acknowledge the following synchrotron radiation facilities where data were collected: Diamond Light Source (I20), European Synchrotron Radiation Facility (ID24, ID26), Photon Factory KEK (BL-9C, PF-PAC No. 2015G551) and Swiss Light Source (SuperXAS). PG acknowledges Sara Lafuerza for providing the experimental data in Fig. 4.
References
Bare, S. & Ressler, T. (2009). Adv. Catal. 52, 339–465. Google Scholar
Bertrand, L., Schöeder, S., Anglos, D., Breese, M. B. H., Janssens, K., Moini, M. & Simon, A. (2015). Trends Anal. Chem. 66, 128–145. Web of Science CrossRef Google Scholar
Bianchini, M. & Glatzel, P. (2012). J. Synchrotron Rad. 19, 911–919. Web of Science CrossRef IUCr Journals Google Scholar
Bonse, U., Materlik, G. & Schröder, W. (1976). J. Appl. Cryst. 9, 223–230. CrossRef IUCr Journals Web of Science Google Scholar
Bravo-Suárez, J. J., Bando, K. K., Lu, J., Haruta, M., Fujitani, T. & Oyama, T. (2008). J. Phys. Chem. C, 112, 1115–1123. Google Scholar
Bridges, F., Li, G. G. & Wang, X. (1992). Nucl. Instrum. Methods Phys. Res. A, 320, 548–555. CrossRef Web of Science Google Scholar
Bunker, G. (2010). Introduction to XAFS: A Practical Guide to X-ray Absorption Fine Structure Spectroscopy. Cambridge University Press. Google Scholar
Carra, P., Fabrizio, M. & Thole, B. T. (1995). Phys. Rev. Lett. 74, 3700–3703. CrossRef PubMed CAS Web of Science Google Scholar
Chantler, C. T., Barnea, Z., Tran, C. Q., Rae, N. A. & de Jonge, M. D. (2012b). J. Synchrotron Rad. 19, 851–862. Web of Science CrossRef IUCr Journals Google Scholar
Chantler, C. T., Rae, N. A., Islam, M. T., Best, S. P., Yeo, J., Smale, L. F., Hester, J., Mohammadi, N. & Wang, F. (2012a). J. Synchrotron Rad. 19, 145–158. Web of Science CrossRef IUCr Journals Google Scholar
Dadashev, A., Pasternak, M. P., Rozenberg, G. K. & Taylor, R. D. (2001). Rev. Sci. Instrum. 72, 2633–2637. Web of Science CrossRef CAS Google Scholar
Dubrovinsky, L., Dubrovinskaia, N., Bykova, E., Bykov, M., Prakapenka, V., Prescher, C., Glazyrin, K., Liermann, H.-P., Hanfland, M., Ekholm, M., Feng, Q., Pourovskii, L. V., Katsnelson, M. I., Wills, J. M. & Abrikosov, I. A. (2015). Nature (London), 525, 226–229. Web of Science CrossRef Google Scholar
Eremets, M. I. (1996). High-Pressure Experimental Methods. Oxford University Press. Google Scholar
Glatzel, P., Bergmann, U., de Groot, F. M. F. & Cramer, S. P. (2003). AIP Conf. Proc. 652, 250–255. CrossRef Google Scholar
Glatzel, P., Weng, T. C., Kvashnina, K., Swarbrick, J., Sikora, M., Gallo, E., Smolentsev, N. & Mori, R. A. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 17–25. Web of Science CrossRef Google Scholar
Glover, J. L. & Chantler, C. T. (2009). X-ray Spectrom. 38, 510–512. Web of Science CrossRef CAS Google Scholar
Goulon, J., Goulon-Ginet, C., Cortes, R. & Dubois, J. M. (1982). J. Phys. Fr. 43, 539–548. CrossRef CAS Web of Science Google Scholar
Hagelstein, M., Lienert, U., Ressler, T., San Miguel, A., Freund, A., Cunis, S., Schulze, C., Fontaine, A. & Hodeau, J.-L. (1998). J. Synchrotron Rad. 5, 753–755. Web of Science CrossRef CAS IUCr Journals Google Scholar
Haskel, D. (1999). FLUO, https://www.aps.anl.gov/~haskel/fluo.html. Google Scholar
Heald, S. M. (2015). J. Synchrotron Rad. 22, 436–445. Web of Science CrossRef IUCr Journals Google Scholar
Henke, B. L., Gullikson, E. M. & Davis, J. C. (1993). At. Data Nucl. Data Tables, 54, 181–342. CrossRef CAS Web of Science Google Scholar
Holton, J. M. (2007). J. Synchrotron Rad. 14, 51–72. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ihli, J., Ferreira Sanchez, D., Jacob, R. R., Cuartero, V., Mathon, O., Krumeich, F., Borca, C., Huthwelker, T., Cheng, W., Shu, Y., Pascarelli, S., Grolimund, D., Menzel, A. & van Bokhoven, J. A. (2017). Angew. Chem. Int. Ed. 56, 14031–14035. Web of Science CrossRef Google Scholar
Irifune, T., Kurio, A., Sakamoto, S., Inoue, T. & Sumiya, H. (2003). Nature (London), 421, 599–600. Web of Science CrossRef PubMed CAS Google Scholar
Ishimatsu, N., Matsumoto, K., Maruyama, H., Kawamura, N., Mizumaki, M., Sumiya, H. & Irifune, T. (2012). J. Synchrotron Rad. 19, 768–772. Web of Science CrossRef CAS IUCr Journals Google Scholar
Itié, J. P., Flank, A.-M., Lagarde, P., Polian, A., Couzinet, B. & Idir, M. (2007). AIP Conf. Proc. 879, 1329–1332. Google Scholar
Kopelent, R., van Bokhoven, J. A., Nachtegaal, M., Szlachetko, J. & Safonova, O. V. (2016). Phys. Chem. Chem. Phys. 18, 32486–32493. Web of Science CrossRef Google Scholar
Kopelent, R., van Bokhoven, J. A., Szlachetko, J., Edebeli, J., Paun, C., Nachtegaal, M. & Safonova, O. V. (2015). Angew. Chem. Int. Ed. 54, 8728–8731. Web of Science CrossRef Google Scholar
Lee, P. A., Citrin, P. H., Eisenberger, P. & Kincaid, B. M. (1981). Rev. Mod. Phys. 53, 769–806. CrossRef CAS Web of Science Google Scholar
Li, G. G., Bridges, F. & Wang, X. (1994). Nucl. Instrum. Methods Phys. Res. A, 340, 420–426. CrossRef Web of Science Google Scholar
Mao, H. K., Chen, B., Chen, Y., Li, K., Lin, J.-F., Yang, W. & Zheng, H. (2016). Matter Radiat. Extremes, 1, 59–75. CrossRef Google Scholar
Mathon, O., Kantor, I. & Pascarelli, S. (2016). X-ray Absorption and X-ray Emission Spectroscopy: Theory and Applications, edited by J. A. van Bokhoven and C. Lamberti, pp. 185–212. Chichester: John Wiley and Sons.. Google Scholar
Matsushita, T. & Phizackerley, R. P. (1981). Jpn. J. Appl. Phys. 20, 2223–2228. CrossRef CAS Web of Science Google Scholar
Müller, O., Nachtegaal, M., Just, J., Lützenkirchen-Hecht, D. & Frahm, R. (2016). J. Synchrotron Rad. 23, 260–266. Web of Science CrossRef IUCr Journals Google Scholar
Müller, O., Stötzel, J., Lützenkirchen-Hecht, D. & Frahm, R. (2013). J. Phys. Conf. Ser. 425, 092010. Google Scholar
Nørskov, J. K., Bligaard, T., Rossmeisl, J. & Christensen, C. H. (2009). Nat. Chem. 1, 37–46. Google Scholar
Oyanagi, H. & Stern, E. A. (2017). Q2XAFS Meetings as a means of experimental XAFS standardization, https://www.ixasportal.net/ixas/index.php?option=com_content&view=article&id=43&Itemid=198. Google Scholar
Padamati, S. K., Angelone, D., Draksharapu, A., Primi, G., Martin, D. J., Tromp, M., Swart, M. & Browne, W. R. (2017). J. Am. Chem. Soc. 139, 8718–8724. Web of Science CrossRef Google Scholar
Pascarelli, S. & Mathon, O. (2010). Phys. Chem. Chem. Phys. 12, 5535–5546. Web of Science CrossRef CAS PubMed Google Scholar
Pascarelli, S., Mathon, O., Mairs, T., Kantor, I., Agostini, G., Strohm, C., Pasternak, S., Perrin, F., Berruyer, G., Chappelet, P., Clavel, C. & Dominguez, M. C. (2016). J. Synchrotron Rad. 23, 353–368. Web of Science CrossRef IUCr Journals Google Scholar
Pascarelli, S., Neisius, T. & De Panfilis, S. (1999). J. Synchrotron Rad. 6, 1044–1050. Web of Science CrossRef CAS IUCr Journals Google Scholar
Pettifer, R. F., Borowski, M. & Loeffen, P. W. (1999). J. Synchrotron Rad. 6, 217–219. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sainctavit, Ph., Petiau, J., Manceau, A., Rivallant, R., Belakhovsky, M. & Renaud, G. (1988). Nucl. Instrum. Methods Phys. Res. A, 273, 423–428. CrossRef Web of Science Google Scholar
Sayers, D. E., Stern, E. A. & Lytle, F. W. (1971). Phys. Rev. Lett. 27, 1204–1207. CrossRef CAS Web of Science Google Scholar
Seidler, G. T., Mortensen, D. R., Remesnik, A. J., Pacold, J. I., Ball, N. A., Barry, N., Styczinski, M. & Hoidn, O. R. (2014). Rev. Sci. Instrum. 85, 113906. Web of Science CrossRef Google Scholar
Stanford Synchrotron Radiation Lightsource (1999). Monochromator Crystal Glitch Library, https://www-ssrl.slac.stanford.edu/~xas/glitch/glitch.html. Google Scholar
Stern, E. A. & Lu, K.-Q. (1982). Nucl. Instrum. Methods Phys. Res. 195, 415–417. CrossRef CAS Web of Science Google Scholar
Sutter, J. P., Boada, R., Bowron, D. T., Stepanov, S. A. & Díaz-Moreno, S. (2016). J. Appl. Cryst. 49, 1209–1222. Web of Science CrossRef IUCr Journals Google Scholar
Tang, Z., Zheng, L., Chu, S., Wu, M., An, P., Zhang, L. & Hu, T. (2015). J. Synchrotron Rad. 22, 1147–1150. Web of Science CrossRef CAS IUCr Journals Google Scholar
Torchio, R., Occelli, F., Mathon, O., Sollier, A., Lescoute, E., Videau, L., Vinci, T., Benuzzi-Mounaix, A., Headspith, J., Helsby, W., Bland, S., Eakins, D., Chapman, D., Pascarelli, S. & Loubeyre, P. (2016). Sci. Rep. 6, 26402. Web of Science CrossRef Google Scholar
Tran, C. Q., Barnea, Z., de Jonge, M. D., Dhal, B. B., Paterson, D., Cookson, D. J. & Chantler, C. T. (2003). X-ray Spectrom. 32, 69–74. Web of Science CrossRef CAS Google Scholar
Van Der Laan, G. & Thole, B. T. (1988). Nucl. Instrum. Methods Phys. Res. A, 263, 515–521. CrossRef Web of Science Google Scholar
Walko, D. A., Arms, D. A., Miceli, A. & Kastengren, A. L. (2011). Nucl. Instrum. Methods Phys. Res. A, 649, 81–83. Web of Science CrossRef Google Scholar
Woicik, J. C., Ravel, B., Fischer, D. A. & Newburgh, W. J. (2010). J. Synchrotron Rad. 17, 409–413. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.