beamlines
The GALAXIES inelastic hard X-ray scattering end-station at Synchrotron SOLEIL
aSynchrotron SOLEIL, L'Orme des Merisiers, BP 48, Saint Aubin, 91192 Gif-sur-Yvette, France, bInstitut de Minéralogie, de Physique des Matériaux et de Cosmochimie, Sorbonne Université, UMR CNRS 7590, MNHN, IRD UMR 206, 4 Place Jussieu, F-75005 Paris, France, cLaboratoire de Physique des Solides, 91400 Orsay, France, dÉcole Nationale Supérieure d'Ingénieurs de Limoges, France, and eSorbonne Université, CNRS, Laboratoire de Chimie Physique – Matiére et Rayonnement, LCPMR, F-75005 Paris, France
*Correspondence e-mail: james.ablett@synchrotron-soleil.fr
GALAXIES is an in-vacuum undulator hard X-ray micro-focused beamline dedicated to the study of the electronic structure of materials with high energy resolution using both
and inelastic X-ray scattering and under both non-resonant (NR-IXS) and resonant (RIXS) conditions. Due to the penetrating power of hard X-rays and the `photon-in/photon-out' technique, the sample environment is not a limitation. Materials under extreme conditions, for example in diamond anvil cells or catalysis chambers, thus constitute a major research direction. Here, the design and performance of the inelastic X-ray scattering end-station that operates in the energy range from ∼4 keV up to 12 keV is reported, and its capabilities are highlighted using a selection of data taken from recently performed experiments. The ability to scan `on the fly' the incident and scattered/emitted X-ray energies, and the sample position enables fast data collection and high experimental throughput. A diamond X-ray transmission phase retarder, which can be used to generate circularly polarized light, will also be discussed in the light of the recent RIXS–MCD approach.1. Introduction
et al., 2009; Llorens et al., 2012; Verbeni et al., 2009; Huotari et al., 2017; Moretti Sala et al., 2018; Glatzel et al., 2009a), the Advanced Photon Source (Fister et al., 2006; Xiao et al., 2016), the Stanford Synchrotron Radiation Light Source (Sokaras et al., 2012, 2013), SPring-8 (Ishii et al., 2013), the Diamond Light Source (Diaz-Moreno et al., 2018), the Swiss Light Source (Kleymenov et al., 2011), PETRA-III (Ketenoglu et al., 2018; Sundermann et al., 2017) and the Shanghai Synchrotron Radiation Facility (Ni et al., 2018; Duan et al., 2016; Gao et al., 2013). Spectrometers based on cylindrically bent crystal analyzers using the Von Hamos geometry, and which are particularly suited to at X-ray free-electron lasers, are also becoming more prevelent (Hoszowska & Dousse, 2004; Alonso-Mori et al., 2012; Szlachetko et al., 2017).
using hard X-rays (≥3 keV) is continuing to develop into a well established spectroscopic tool. Insertion device development, advancements in single-crystal analysers and improved detector technology have been the impetus in developing facilities at third-generation synchrotron radiation sources over recent years. In the hard X-ray range, instruments come in two flavours: very high resolution instruments with ∼1 meV resolution for phonon excitations, and medium-resolution instruments (∼100 meV to 1 eV) mostly employed for resonant X-ray Raman scattering or X-ray emission. Examples of medium-resolution state-of-the-art hard X-ray instruments can be found at the European Synchrotron Radiation Facility (HazemannThe various ; Rueff & Shukla, 2010, 2013; Rovezzi & Glatzel, 2014; Glatzel & Bergmann, 2005; Hayashi, 2013; Glatzel et al., 2009b) and theoretical (Schülke, 1986, 2007; van Veenendaal, 2014; Ament et al., 2011; De Groot & Kotani, 2008) points of view. We have recently reported on the overall characteristics of the GALAXIES beamline (Rueff et al., 2015a) and also on the hard (HAXPES) end-station (Céolin et al., 2013). This article completes the description of our beamline facilities.
processes (resonant and non-resonant) can be divided into several distinct techniques such as high-energy-resolved fluorescence detection (HERFD) and X-ray Raman spectroscopy (XRS). These often require very different experimental demands in terms of resolution and apparatus and there are many excellent reviews of these from both experimental (Rueff, 20091.1. Major components of the GALAXIES beamline
In this section, we describe the important aspects of the GALAXIES beamline that are relevant to the operation of the 2Fe14B undulator (98 periods, 1.96 m long) with a peak magnetic field of 1.04 T at a minimum gap of 5.5 mm. A schematic of the optical layout of the beamline is shown in Fig. 1, where all X-ray mirrors operate at a grazing incidence angle of 5 mrad.
end-station. The source consists of a 20 mm-period in-vacuum NdThe first major optical component of the beamline is the silicon (111) liquid-nitrogen-cooled fixed-exit double-crystal monochromator (DCM) that provides an energy resolution of ≃ 1.4 × 10−4 FWHM. Higher energy resolution can be obtained with the DCM using the third- and fourth-order harmonics starting from 6.9 keV and 9.2 keV, respectively. Circular polarization can be generated with the diamond X-ray phase retarder (XPR), which is currently installed immediately downstream of the DCM (see Section 7). X-rays are then reflected by the M1 spherical collimating mirror, that provides harmonic rejection using selectable carbon or palladium coatings. For high energy resolution, a four-bounce high-resolution monochromator (HRM) can be inserted into the beam delivering δE ≤ 100 meV FWHM bandwidth over the entire operating range for measurements (Ablett et al., 2013). The HRM is currently undergoing commissioning, and high-resolution measurements performed so far have relied on the higher-order harmonics of the DCM. Two focusing modes are available at the experimental end-station (Fig. 1): with the M2A toroidal mirror, the X-rays are focused onto the HAXPES experimental end-station that acts as a virtual X-ray source for the KB focusing mirrors (equipped with rhodium and boron carbide coatings) to provide a micro-focused beam of ∼5 µm × 5 µm FWHM. With M2A withdrawn from the beam, the beam is focused by the M2B toroidal mirror that gives higher compared with the micro-focused mode (about a factor of three more due to the acceptance of the KB mirrors and reflectivity) but with a larger beam size of 30 µm (vertical) × 90 µm (horizontal) FWHM. The end-station is separated from the beamline vacuum by a 50 µm thin beryllium window, thus ensuring minimum absorption loss at low X-ray energies. In order to provide an accurate estimation of the photon intensity delivered for planning a particular experiment, a measurement of the X-ray delivered to the end-station at the sample position, with the DCM and M2B toroidal mirror, is shown in Fig. 2. These measurements were made using a calibrated 104 µm-thick silicon photodiode and corrected for absorption effects including air paths, detector efficiency and at a storage ring current of 430 mA.
2. The GALAXIES spectrometers
The GALAXIES . The spectrometer sits on a high-precision five-axis motorized base using stepper motors that allows the spectrometer to be aligned accurately with respect to the incident beam position. The base rests on anti-skid anti-vibrational pads which ensure excellent vibration damping. The six-axis sample stage, which can accommodate loads of up to 150 kg, allows for in-plane x–s translations (±25 mm) and height adjustments (±100 mm). A horizontal rotation stage (±180°) is mounted beneath these translation stages, underneath which are located two concentric goniometer cradles (±10°). For certain experiments, a sample support comprising a compact horizontal rotation stage, small manual goniometer and a helium/rough-vacuum environment is available for users. For all experiments and in order to minimize absorption losses, helium flight paths are used to eliminate air paths in the set-up.
spectrometers are two integrated analyser systems that share the same support base and sample stage positioners. Both operate in the vertical use 10 cm-diameter spherical analyser crystals with Bragg angles from 70° to 89° in the Rowland circle geometry and use continuous `on-the-fly' scanning. The single-axis spectrometer, support base, multi-analyser arm and sample positioning stages were built by the Newport company. All stages have around 1 µm or 1 milli-degree of translation or rotational resolutions, respectively, and a schematic of the spectrometer is shown in Fig. 32.1. Single-analyser spectrometer
The single-analyser spectrometer uses analysers with between 1 m and 2 m radii of curvature. It is optimized for high-resolution studies in both energy and momentum space and allows for a wide range of scattering geometries. The scattering angle can be varied in both the horizontal and vertical planes by the use of the δ1 and γ1 rotations, respectively (Fig. 3). The analyser and detector are positioned on the Rowland circle by the use of five DC servo-motor-driven high-precision translation and rotation stages. Fig. 4 shows a photograph of this set-up superimposed with the Rowland circle geometry.
2.2. Multi-analyser spectrometer
The multi-analyser spectrometer uses four 10 cm-diameter crystal analysers with radii of curvature of 0.5 m or 1 m and was designed in collaboration with and fabricated at IMPMC-SU (Institut de Minéralogie, Physique des Matériaux et Cosmochimie, Sorbonne Université) (Gélebart et al., 2007). The multi-analyser system is fixed onto the short horizontal arm of the spectrometer (Fig. 3), and can be easily removed when required. For RIXS experiments, where the emitted energy is scanned on-the-fly, the configuration is that shown on the right-hand side of Fig. 5. In this configuration, the scan of the emitted energy is realised by changing the height of the analysers while adjusting the analyser–sample and detector–sample distances to maintain the Rowland circle condition. For non-resonant or measurements, where only the incident X-ray energy is scanned and the of the analyser crystals remains fixed, we typically use a 2 × 2 analyser configuration in order to increase the momentum resolution and provide a more compact arrangement (left-hand side of Fig. 5), and has been used for XRS measurements on CeCuGe2 (Rueff et al., 2015b), lithium borate glasses (Lelong et al., 2014, 2017) and ancient materials (Gueriau et al., 2017). Each analyser crystal is contained in a module that comprises highly rigid spring-loaded gimble mounts that are mounted onto a 50 mm-long linear high-precision translation. Each module is then mounted on a 30 cm-long vertical translation stage that allows the to be selected from 75.6° to 90° for 1 m-radius analysers and covers 70° to 90° for 0.5 m-radius analysers. Currently, Si(110) and Si(111) four-crystal 1 m-radius sets are available for users.
3. Analyser crystals
We are constantly adding to our portfolio of analyser crystals, which are shown in Table 1 and are updated under the user guide section of the GALAXIES website (https://www.synchrotron-soleil.fr/fr/lignes-de-lumiere/galaxies). As well as acquiring commercially available analysers, we are able to use anodic bonding for in-house fabrication (Collart et al., 2005). An extensive list of analyser crystals for many absorption and emission energies, together with their energy resolution and reflectivities, has also been published which is useful for planning experiments (Gog et al., 2013). In order to determine the overall energy resolution for an measurement, it is necessary to combine the incident beam resolution [given by the bandwidth of the DCM and/or HRM monochromators (Ablett et al., 2013)], the intrinsic resolution of the analyser crystal as well as other geometrical effects such as the vertical source size contribution and Johann error (Bergmann & Cramer, 1998). A signifcant improvement in predicting the reflectivity curves for spherical analyser crystals without strain relief, that includes the dominant effect of angular compression due to the analyser bending, has recently been discussed (Honkanen et al., 2014, 2016, 2017).
|
Fig. 6 shows the estimated total resolution (FWHM) of the GALAXIES spherically bent analyers using their full available area. All geometrical contributions (Bergmann & Cramer, 1998) and the effect of strain (Honkanen et al., 2014, 2016, 2017) (see Table 1) have been included in the calculation. The highest resolution is obtained by reducing the incident X-ray beam bandwidth [currently using the Si(333) reflection of the DCM] and using 2 m-radius analyser crystals or analyzer crystals that have been strain relieved (Table 1). The Si(110) and Si(111) strain-relieved analyzers are currently used for X-ray Raman spectroscopy at Bragg angles close to 89° and achieve <300 meV FWHM total resolution for these type of measurements.
4. Sample environments
We have a wide range of sample environments available for experiments and users can install their own specialized systems. For high-pressure measurements (Ablett et al., 2015; Dorfman et al., 2015, 2016; Baledent et al., 2015; Martin et al., 2016; Baledent & Rue, 2016), we have several types of gas-driven membrane diamond anvil cells, and a high-pressure laboratory is available for preparing and loading them. A ruby fluorescence laser measuring system is typically mounted within the experimental set-up in order to quickly determine the applied pressure. High-pressure measurements can also be made down to 4 K with a specially designed helium cryostat (Sundermann et al., 2015; Rueff et al., 2015b) and up to several hundreds of degrees Celsius with a resistive heater. A helium compressor cryostat, for measuring samples down to 15 K, a special oven allowing temperature exceeding 900°C and cryo-streams are also available. To apply magnetic fields to the sample (Zimmermann et al., 2018), a compact electromagnet with drilled hole pieces that allow the incident X-ray beam to pass can deliver ±0.7 T. A catalytic cell developed in collaboration with the ROCK beamline at SOLEIL and adapted to the RIXS geometry is also available. Finally, an experimental setup for the study of gas phase samples has also been developed (Ismail et al., 2018).
5. Detectors
On the X-ray spectrometer detector front, we use either a silicon drift detector (SDD), an 2, a thick 25 µm-thick beryllium window and can be used within the entire operating energy range of the end-station. The detector is connected to an XSPRESS mini (Quantum Detectors, UK) fast digital pulse processing unit which can be used up to and above a million counts per second without saturation and incorporates accurate dead-time correction. The high APD (FMB Oxford) has a 10 mm × 10 mm area, a 200 µm-thick silicon sensor and a 40 µm opaque Kapton window. The Merlin pixel (Quantum Detectors, UK) single (256 × 256 pixels) and quad chip (512 × 512 pixels) detectors are also available, for energies above 7 keV. These hybrid detectors have 500 µm-thick sensors with 55 µm × 55 µm size pixels and exhibit a very high One of the important benefits of these detectors is that they can increase the energy resolution when used in conjunction with dispersive analyser systems that are essential when measuring low-energy excitations (Shvyd'ko, 2004; Huotari et al., 2005, 2006; Ketenoglu et al., 2015). In addition, the ability to define specific `regions of interest' can help reduce background contributions and can also be used for imaging applications (Huotari et al., 2011).
(APD) or hybrid pixel detectors, depending upon experimental conditions. The fast SDD (SiriusSD, RaySpec, UK) has a 450 µm sensor, a collimated area of 80 mm6. Continuous on-the-fly scanning
The incident X-ray energy (combined scanning of the DCM and undulator), the energies of the single- and multi-analyser spectrometers and the sample position stages all use on-the-fly continuous scans. On-the-fly scanning is an essential prerequisite for the operation of the GALAXIES K-edge on a 6 µm-thick Fe foil, using a silicon (531) crystal analyser tuned to the Fe Kβ emission line, with a dwell time of 0.25 s over a 0.3 eV energy interval, can be performed within 90 s with adequate statistics, which is ∼15 times faster than an equivalent scan performed in step-by-step mode. For these type of scans, the undulator speed is continuously adjusted in order to keep it synchronized with the DCM, and is typically within 1 eV of the undulator maximum intensity throughout the scan. This excellent synchronization is highlighted in Fig. 7 (left) over a 950 eV scan range with a 0.5 s dwell time over a 0.2 eV energy interval, with a total aquisition time of ∼45 min. The diagonal line is the superposition of both the DCM and undulator energies (left axis) and the blue curve is the energy difference between the two (right axis). This energy difference is more than adequate in order to stay on the maximum of the undulator emission as highlighted in Fig. 7 (right).
end-station as it enables data collection with no dead-time thus allowing for high data collection throughput. This is particularly advantageous for radiation-sensitive materials. As an example, a scan around the ironAll on-the-fly scans are defined via commands explicit to the beamline which are based on the Python language. These commands are frequently used in macros in order to automatize the experiment once it has been optimized. A schematic of the control architecture is shown in Fig. 8. The configuration of the hardware is automatically applied once the scan information has been entered by the user. The master clock and controller(s) are simultaneously initiated at the start of the scan. During a scan, the master clock delivers individual TTL pulses of duration t (the integration time specified by the scan) and separated by a short millisecond delay (which is mostly dictated by the read-out of the detector modules). These pulses ensure synchronization between the various hardware components during a scan. A counting card, the detector electronics and certain Newport XPS controllers are triggered by these pulses. The counting card, which is gated by the TTL pulses, is used to record the position of the the DCM and single-analyser Bragg angles (and thus their energies). The multi-analyser energy is determined by the position of the height translations of the four analyser motors that are read out synchronously from the XPS controller, in exactly the same manner in which the positions of the six-axis sample stage are determined from a different XPS controller. For the energy scans of the spectrometers and the sample position, the motor trajectories are controlled by the XPS controllers. All data are stored on a server that is labelled `File Server' in Fig. 8 in both NEXUS and ASCII formats.
An example of on-the-fly scanning is shown in Fig. 9 where X-ray Raman spectroscopy has been used to elucidate the carbon speciation in ancient and historic materials such as fossils and pigments (Gueriau et al., 2017). Fig. 9 shows an XRS spectrum on a few mm-thick graphite sample around the carbon K-edge using the third order of four Si(111) analyser crystals at a of 86° (incident X-ray energy = 5.9 keV) in a back-scattering geometry (momentum transfer ≃ 5.8 Å−1). The total resolution for this experiment was 0.8 eV FWHM and the plot represents the background subtracted average of 43 scans with an overall acquisition time of 2.2 h.
7. Diamond X-ray phase retarder
Polarization can be controlled in a relatively straightforward manner in the hard X-ray range using XPRs. These devices, which are fabricated from low-absorbing perfect-crystal materials such as diamond or silicon, allow the horizontal polarization of the synchrotron X-ray beam to be transformed into circular, elliptical or vertical polarizations and have long been in operation at numerous hard X-ray synchrotron facilities worldwide (Logan et al., 2016; Hirano & Maruyama, 1997; Hirano et al., 1993; Giles et al., 1994a,b, 1995; Freeland et al., 2002; Lang & Srajer, 1995; Strempfer et al., 2013; Paolasini et al., 2007). The GALAXIES XPR mechanical assembly is currently fitted with a 500 µm-thick single-crystal synthetic diamond (111). The XPR operates in the Laue asymmetric transmission geometry where the reflecting planes subtend an angle of 19.47° to the XPR [111] surface and is very efficient in terms of transmission and polarization in the energy range from 6 to 9 keV. Table 2 shows the important parameters for the XPR.
|
One area of research where the XPR is currently being exploited is in hard X-ray resonant inelastic X-ray scattering combined with s2p RIXS in transition metal compounds, where the Kα(2p → 1s) transition is recorded after resonant excitation of a 1s core electron (1s → 3d), is of particular interest because the final state in the RIXS process 2p53dn+1 is similar to that of the corresponding soft XMCD (Caliebe et al., 1998). The attractiveness of RIXS–MCD is not only in its potential high contrast and high penetrating power of the incident and emitted X-rays but also in the ability to tune the incident X-ray energy to a specific chemical species in mixed valent systems and perform site-selective magnetic hysteresis measurements (Sikora et al., 2012).
(RIXS–MCD). This technique allows bulk-sensitive, element- and site-selective magnetic information on complex materials to be obtained. In particular, 1The first RIXS–MCD measurements at GALAXIES were performed on Fe3O4 as a previously measured test example (Sikora et al., 2010). For this experiment, a 1 m-radius Ge(440) single analyser crystal was used to record the Kα1 emission line as a function of incident energy, circular polarization (left and right) and magnetic field direction (along the direction of X-ray beam propagation). An electromagnet, with 1.5 mm-diameter holes drilled through the pole pieces, allowed a magnetic field to be applied to the sample along the direction of the incident X-rays, with a maximum field of ±0.7 T. Fig. 10 shows these results where the upper left-hand plot shows the Fe Kα1 intensity as a function of incident X-ray energy for both left and right circularly polarized light with an applied field of +0.7 T along the incident X-ray beam direction, in the pre-edge region of the K-edge of Fe3O4. The dramatic strong magnetic contrast of ∼16% in the pre-edge region of the K-edge was found, compared with conventional K-edge absorption measurements, due to reduced lifetime broadening and increased splitting of the final-state spectral features (Sikora et al., 2010). The difference between these signals for the two magnetic field directions is shown in the bottom left of Fig. 10 and a RIXS–MCD map is displayed on the right-hand side of Fig. 10, showing the dramatic magnetic contrast in the pre-edge region as previously reported (Sikora et al., 2010). We have also recently used 1s2p RIXS–MCD in order to determine the magnetic ground state of CrO2 where there is a similar but smaller magnetic contrast compared with magnetite (Zimmermann et al., 2018).
8. Conclusion
In this article, we have described and demonstrated the capabilities of the
end-station at the GALAXIES beamline. The facility enables a wealth of different experimental techniques to be performed such as X-ray emission, X-ray Raman, resonant- and non-resonant scattering and spectroscopies. Hard X-ray is developing into a readily accessible tool for studying the electronic properties of matter at synchrotron radiation facilities. The performance of the facility will be enhanced once the high-resolution monochromator has been commissioned and an upgrade for X-ray Raman spectroscopy comprising a 40-crystal analyser spectrometer is currently underway.Acknowledgements
We would like to thank Dr Palyanov and co-workers at the Institute of Geology and Mineralogy of the Russian Academy of Science (Novossibirsk, Russia) for the growth of the synthetic diamond used in the XPR. Much appreciation is given to Professor T. A. Tyson for supplying us with silicon crystals at the start of this project and Dr P. Glatzel from the ESRF for lending us analyser crystals from time to time. We are extremely grateful to Professor A. Juhin and Professor Ph. Sainctavit for help and supplying the sample for the RIXS–MCD measurements on magnetite. Finally, we express our gratitude to the various SOLEIL teams including the design, engineering, computing, detector and accelerator groups for their essential support throughout this project.
References
Ablett, J. M., Dubuisson, J. M., Moreno, T., Céolin, D., Prieur, D., Raimon, E., Corruble, D., Coquet, J., Lestrade, A., Bourgoin, C. & Rueff, J. P. (2013). J. Phys. Conf. Ser. 425, 052007. CrossRef Google Scholar
Ablett, J., Rueff, J.-P., Shieh, S., Kao, C. & Wang, S. (2015). Phys. Rev. B, 92, 014113. CrossRef Google Scholar
Alonso-Mori, R., Kern, J., Sokaras, D., Weng, T.-C., Nordlund, D., Tran, R., Montanez, P., Delor, J., Yachandra, V. K., Yano, J. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 073114. Web of Science PubMed Google Scholar
Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705–767. Web of Science CrossRef CAS Google Scholar
Balédent, V. & Rueff, J. (2016). High. Press. Res. 36, 371–380. Google Scholar
Balédent, V., Rullier-Albenque, F., Colson, D., Ablett, J. M. & Rueff, J.-P. (2015). Phys. Rev. Lett. 114, 177001. Google Scholar
Bergmann, U. & Cramer, S. P. (1998). Proc. SPIE, 3448, 198–209. CrossRef CAS Google Scholar
Caliebe, W. A., Kao, C. C., Hastings, J. B., Taguchi, M., Kotani, A., Uozumi, T. & de Groot, F. M. F. (1998). Phys. Rev. B, 58, 13452–13458. Web of Science CrossRef CAS Google Scholar
Céolin, D., Ablett, J. M., Prieur, D., Moreno, T., Rueff, J. P., Marchenko, T., Journel, L., Guillemin, R., Pilette, B., Marin, T. & Simon, M. (2013). J. Electron Spectrosc. Relat. Phenom. 190, 188–192. Google Scholar
Collart, E., Shukla, A., Gélébart, F., Morand, M., Malgrange, C., Bardou, N., Madouri, A. & Pelouard, J.-L. (2005). J. Synchrotron Rad. 12, 473–478. Web of Science CrossRef CAS IUCr Journals Google Scholar
De Groot, F. M. F. & Kotani, A. (2008). Core Level Spectroscopy of Solids. CRC Press/Taylor & Francis Group. Google Scholar
Diaz-Moreno, S., Amboage, M., Basham, M., Boada, R., Bricknell, N. E., Cibin, G., Cobb, T. M., Filik, J., Freeman, A., Geraki, K., Gianolio, D., Hayama, S., Ignatyev, K., Keenan, L., Mikulska, I., Mosselmans, J. F. W., Mudd, J. J. & Parry, S. A. (2018). J. Synchrotron Rad. 25, 998–1009. Web of Science CrossRef IUCr Journals Google Scholar
Dorfman, S. M., Badro, J., Rueff, J., Chow, P., Xiao, Y. & Gillet, P. (2015). Am. Mineral. 100, 2246–2253. CrossRef Google Scholar
Dorfman, S., Dutton, S., Potapkin, V., Chumakov, A., Rueff, J., Chow, P., Xiao, Y., Cava, R., Duffy, T. S., McCammon, C. & Gillet, P. (2016). Am. Mineral. 101, 1659–1667. CrossRef Google Scholar
Duan, P., Gu, S., Cao, H., Li, J. & Huang, Y. (2016). X-ray Spectrom. 46, 12–18. CrossRef Google Scholar
Fister, T., Seidler, G., Wharton, L., Battle, A., Ellis, T., Cross, J., Macrander, A., Elam, W., Tyson, T. & Qian, Q. (2006). Rev. Sci. Instrum. 77, 063901. CrossRef Google Scholar
Freeland, J. W., Lang, J. C., Srajer, G., Winarski, R., Shu, D. & Mills, D. M. (2002). Rev. Sci. Instrum. 73, 1408–1410. Web of Science CrossRef CAS Google Scholar
Gao, X., Gu, S., Gao, Q., Zou, Y., Jiang, Z., Zhang, S., Wei, X., Yu, H., Sheng, G., Duan, P. & Huang, Y. (2013). X-ray Spectrom. 42, 502–507. CrossRef Google Scholar
Gélebart, F., Morand, M., Dermigny, Q., Giura, P., Rueff, J.-P. & Shukla, A. (2007). AIP Conf. Proc. 879, 1837–1840. Google Scholar
Giles, C., Malgrange, C., Goulon, J., de Bergevin, F., Vettier, C., Dartyge, E., Fontaine, A., Giorgetti, C. & Pizzini, S. (1994b). J. Appl. Cryst. 27, 232–240. CrossRef IUCr Journals Google Scholar
Giles, C., Malgrange, C., Goulon, J., de Bergevin, F., Vettier, C., Fontaine, A., Dartyge, E. & Pizzini, S. (1994a). Nucl. Instrum. Methods Phys. Res. A, 349, 622–625. CrossRef Google Scholar
Giles, C., Malgrange, C., Goulon, J., de Bergevin, F., Vettier, C., Fontaine, A., Dartyge, E., Pizzini, S., Baudelet, F. & Freund, A. (1995). Rev. Sci. Instrum. 66, 1549–1553. CrossRef CAS Web of Science Google Scholar
Glatzel, P. & Bergmann, U. (2005). Coord. Chem. Rev. 249, 65–95. Web of Science CrossRef CAS Google Scholar
Glatzel, P., de Groot, F. M. F. & Bergmann, U. (2009b). Synchrotron Radiat. News, 22(2), 12–16. CrossRef Google Scholar
Glatzel, P., Sikora, M., Smolentsev, G. & Fernández-García, M. (2009a). Catal. Today, 145, 294–299. CrossRef Google Scholar
Gog, T., Casa, D. M., Said, A. H., Upton, M. H., Kim, J., Kuzmenko, I., Huang, X. & Khachatryan, R. (2013). J. Synchrotron Rad. 20, 74–79. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gueriau, P., Rueff, J. P., Bernard, S., Kaddissy, J., Goler, Z., Sahle, C. J., Sokaras, D., Wogelius, R. A., Manning, P. L., Bergmann, U. & Bertrand, L. (2017). Anal. Chem. 89, 10819–10826. CrossRef Google Scholar
Hayashi, H. (2013). Encyclopedia of Analytical Chemistry, edited by R. A. Meyers, pp. 1–31. Chichester: John Wiley. Google Scholar
Hazemann, J.-L., Proux, O., Nassif, V., Palancher, H., Lahera, E., Da Silva, C., Braillard, A., Testemale, D., Diot, M.-A., Alliot, I., Del Net, W., Manceau, A., Gélébart, F., Morand, M., Dermigny, Q. & Shukla, A. (2009). J. Synchrotron Rad. 16, 283–292. Web of Science CrossRef CAS IUCr Journals Google Scholar
Hirano, K., Ishikawa, T., Nakamura, I., Mizutani, M. & Kikuta, S. (1993). Jpn. J. Appl. Phys. 33, L689–L692. CrossRef Google Scholar
Hirano, K. & Maruyama, H. (1997). Jpn. J. Appl. Phys. 36, L1272–L1274. CrossRef Google Scholar
Honkanen, A.-P., Monaco, G. & Huotari, S. (2016). J. Appl. Cryst. 49, 1284–1289. CrossRef CAS IUCr Journals Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Monaco, G. & Huotari, S. (2014). J. Synchrotron Rad. 21, 104–110. Web of Science CrossRef CAS IUCr Journals Google Scholar
Honkanen, A.-P., Verbeni, R., Simonelli, L., Moretti Sala, M., Monaco, G. & Huotari, S. (2017). J. Synchrotron Rad. 24, 545–546. CrossRef IUCr Journals Google Scholar
Hoszowska, J. & Dousse, J.-C. (2004). J. Electron Spectrosc. Relat. Phenom. 137–140, 687–690. Web of Science CrossRef CAS Google Scholar
Huotari, S., Albergamo, F., Vankó, G., Verbeni, R. & Monaco, G. (2006). Rev. Sci. Instrum. 77, 053102. Web of Science CrossRef Google Scholar
Huotari, S., Pylkkänen, T., Verbeni, R., Monaco, G. & Hämäläinen, K. (2011). Nat. Mater. 10, 489–493. Web of Science CrossRef CAS PubMed Google Scholar
Huotari, S., Sahle, C. J., Henriquet, C., Al-Zein, A., Martel, K., Simonelli, L., Verbeni, R., Gonzalez, H., Lagier, M.-C., Ponchut, C., Moretti Sala, M., Krisch, M. & Monaco, G. (2017). J. Synchrotron Rad. 24, 521–530. CrossRef CAS IUCr Journals Google Scholar
Huotari, S., Vankó, Gy., Albergamo, F., Ponchut, C., Graafsma, H., Henriquet, C., Verbeni, R. & Monaco, G. (2005). J. Synchrotron Rad. 12, 467–472. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ishii, K., Jarrige, J., Yoshida, M., Ikeuchi, K., Inami, T., Murakami, Y. & Mizuki, J. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 127–132. CrossRef Google Scholar
Ismail, I., Guillemin, R., Marchenko, T., Travnikova, O., Ablett, J., Rueff, J.-P., Piancastelli, M.-N., Simon, M. & Journel, L. (2018). Rev. Sci. Instrum. 89, 063107. CrossRef Google Scholar
Ketenoglu, D., Harder, M., Klementiev, K., Upton, M., Taherkhani, M., Spiwek, M., Dill, F.-U., Wille, H.-C. & Yavaş, H. (2015). J. Synchrotron Rad. 22, 961–967. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ketenoglu, D., Spiekermann, G., Harder, M., Oz, E., Koz, C., Yagci, M. C., Yilmaz, E., Yin, Z., Sahle, C. J., Detlefs, B. & Yavaş, H. (2018). J. Synchrotron Rad. 25, 537–542. CrossRef IUCr Journals Google Scholar
Kleymenov, E., van Bokhoven, J., David, C., Glatzel, P., Janousch, M., Alonso-Mori, R., Studer, M., Willimann, M., Bergamaschi, A., Henrich, B. & Nachtegaal, M. (2011). Rev. Sci. Instrum. 82, 065107. CrossRef Google Scholar
Lang, J. C. & Srajer, G. (1995). Rev. Sci. Instrum. 66, 1540–1542. CrossRef CAS Web of Science Google Scholar
Lelong, G., Cormier, L., Hennet, L., Michel, F., Rueff, J. P., Ablett, J. M. & Monaco, G. (2017). J. Non-Cryst. Solids, 472, 1–8. CrossRef Google Scholar
Lelong, G., Radtke, G., Cormier, L., Bricha, H., Rueff, J. P., Ablett, J. M., Cabaret, D., Gélébart, F. & Shukla, A. (2014). Inorg. Chem. 53, 10903–10908. CrossRef Google Scholar
Llorens, I., Lahera, E., Delnet, W., Proux, O., Braillard, A., Hazemann, J.-L., Prat, A., Testemale, D., Dermigny, Q., Gelebart, F., Morand, M., Shukla, A., Bardou, N., Ulrich, O., Arnaud, S., Berar, J.-F., Boudet, N., Caillot, B., Chaurand, P., Rose, J., Doelsch, E., Martin, P. & Solari, P. (2012). Rev. Sci. Instrum. 83, 063104. CrossRef Google Scholar
Logan, J., Harder, R., Li, L., Haskel, D., Chen, P., Winarski, R., Fuesz, P., Schlagel, D., Vine, D., Benson, C. & McNulty, I. (2016). J. Synchrotron Rad. 23, 1210–1215. CrossRef IUCr Journals Google Scholar
Martin, N., Deutsch, M., Itié, J., Rueff, J.-P., Rössler, U., Koepernik, K., Fomicheva, L., Tsvyashchenko, A. & Mirebeau, I. (2016). Phys. Rev. B, 93, 214404. CrossRef Google Scholar
Moretti Sala, M., Martel, K., Henriquet, C., Al Zein, A., Simonelli, L., Sahle, C., Gonzalez, H., Lagier, M.-C., Ponchut, C., Huotari, S., Verbeni, R., Krisch, M. & Monaco, G. (2018). J. Synchrotron Rad. 25, 580–591. Web of Science CrossRef CAS IUCr Journals Google Scholar
Ni, D., Kang, X., Yan, S., Huang, X. C., Xiong, T., Liang, D. X., Yang, K. & Zhu, L. F. (2018). Rev. Sci. Instrum. 89, 085108. CrossRef Google Scholar
Paolasini, L., Detlefs, C., Mazzoli, C., Wilkins, S., Deen, P. P., Bombardi, A., Kernavanois, N., de Bergevin, F., Yakhou, F., Valade, J. P., Breslavetz, I., Fondacaro, A., Pepellin, G. & Bernard, P. (2007). J. Synchrotron Rad. 14, 301–312. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rovezzi, M. & Glatzel, P. (2014). Semicond. Sci. Technol. 29, 023002. CrossRef Google Scholar
Rueff, J.-P. (2009). Magnetism and Synchrotron Radiation, edited by E. Beaurepaire, H. Bulou, F. Scheurer & J. P. Kappler, Vol. 133 in Springer Proceedings in Physics, pp. 263–277. Springer. Google Scholar
Rueff, J.-P., Ablett, J. M., Céolin, D., Prieur, D., Moreno, T., Balédent, V., Lassalle-Kaiser, B., Rault, J. E., Simon, M. & Shukla, A. (2015a). J. Synchrotron Rad. 22, 175–179. CrossRef IUCr Journals Google Scholar
Rueff, J.-P., Ablett, J. M., Strigari, F., Deppe, M., Haverkort, M. W., Tjeng, L. H. & Severing, A. (2015b). Phys. Rev. B, 91, 201108. CrossRef Google Scholar
Rueff, J.-P. & Shukla, A. (2010). Rev. Mod. Phys. 82, 847–896. Web of Science CrossRef CAS Google Scholar
Rueff, J.-P. & Shukla, A. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 10–16. CrossRef CAS Google Scholar
Schülke, W. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 491–499. Google Scholar
Schülke, W. (2007). Electron Dynamics by Inelastic X-ray Scattering. Oxford University Press. Google Scholar
Shvyd'ko, Y. (2004). X-ray Optics – High-Energy-Resolution Applications. Springer Optical Sciences. Google Scholar
Sikora, M., Juhin, A., Simon, G., Zając, M., Biernacka, K., Kapusta, C., Morellon, L., Ibarra, M. R. & Glatzel, P. (2012). J. Appl. Phys. 111, 07E301. CrossRef Google Scholar
Sikora, M., Juhin, A., Weng, T. C., Sainctavit, P., Detlefs, C., de Groot, F. M. F. & Glatzel, P. (2010). Phys. Rev. Lett. 105, 037202. CrossRef Google Scholar
Sokaras, D., Nordlund, D., Weng, T.-C., Mori, R. A., Velikov, P., Wenger, D., Garachtchenko, A., George, M., Borzenets, V., Johnson, B., Qian, Q., Rabedeau, T. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 043112. CrossRef PubMed Google Scholar
Sokaras, D., Weng, T.-C., Nordlund, D., Alonso-Mori, R., Velikov, P., Wenger, D., Garachtchenko, A., George, M., Borzenets, V., Johnson, B., Rabedeau, T. & Bergmann, U. (2013). Rev. Sci. Instrum. 84, 053102. Web of Science CrossRef PubMed Google Scholar
Strempfer, J., Francoual, S., Reuther, D., Shukla, D. K., Skaugen, A., Schulte-Schrepping, H., Kracht, T. & Franz, H. (2013). J. Synchrotron Rad. 20, 541–549. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sundermann, M., Chen, K., Yavaş, H., Lee, H., Fisk, Z., Haverkort, M., Tjeng, L. & Severing, A. (2017). Europhys. Lett. 117, 17003. CrossRef Google Scholar
Sundermann, M., Strigari, F., Willers, T., Winkler, H., Prokofiev, A., Ablett, J. M., Rueff, J.-P., Schmitz, D., Weschke, E., Sala, M. M., Al-Zein, A., Tanaka, A., Haverkort, M. W., Kasinathan, D., Tjeng, L. H., Paschen, S. & Severing, A. (2015). Sci. Rep. 5, 17937. CrossRef Google Scholar
Szlachetko, J., Nachtegaal, M., Grolimund, D., Knopp, G., Peredkov, P., Czapla–Masztafiak, J. & Milne, C. (2017). Appl. Sci. 7, 899. CrossRef Google Scholar
Veenendaal, M. van (2014). Theory of Inelastic Scattering of X-rays. Cambridge University Press. Google Scholar
Verbeni, R., Pylkkänen, T., Huotari, S., Simonelli, L., Vankó, G., Martel, K., Henriquet, C. & Monaco, G. (2009). J. Synchrotron Rad. 16, 469–476. Web of Science CrossRef CAS IUCr Journals Google Scholar
Xiao, Y., Chow, P. & Shen, G. (2016). High. Press. Res. 36, 315–331. CrossRef Google Scholar
Zimmermann, P., Bouldi, N., Hunault, M. O. J. Y., Sikora, M., Ablett, J. M., Rueff, J.-P., Lebert, B., Sainctavit, P., de Groot, F. M. F. & Juhin, A. (2018). J. Electron Spectrosc. Relat. Phenom. 222, 74–87. CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.