Optical performance and radiation stability of polymer X-ray refractive nano-lenses

CROSSMARK_Color_square_no_text.svg

aImmanuel Kant Baltic Federal University, Nevskogo 14, Kaliningrad 236041, Russia, bEuropean Molecular Biology Laboratory, Hamburg Unit, Notkestraße 85, Hamburg, Germany, cFaculty of Physics, Lomonosov Moscow State University, Moscow 119991, Russia, dFrumkin Institute of Physical Chemistry and Electrochemistry, Russian Academy of Sciences, Moscow 119071, Russia, eEuropean Synchrotron Radiation Facility (ESRF), 71 avenue des Martyrs, 38043 Grenoble, France, and fInstitute of Microelectronics Technology RAS, Chernogolovka, Russia
*Correspondence e-mail: mpolikarpov@embl-hamburg.de

Edited by A. Momose, Tohoku University, Japan (Received 26 October 2018; accepted 29 January 2019; online 1 April 2019)

Full-field X-ray imaging and microscopy with polymer compound refractive nano-lenses is demonstrated. Experiments were carried out at beamline ID13 at the European Synchrotron and yielded a resolution of 100 nm. The lenses were demonstrated to be functioning even after an absorbed dose of ∼107 Gy. This article also discusses issues related to lens aberrations, astigmatism and radiation stability, and thus ways of improving the lens further are considered. Polymer nano-lenses are versatile and are promissing for nano-focusing and compact X-ray microscopy.

1. Introduction

In recent years, a new (fourth) generation of light sources has been emerging worldwide. Diffraction-limited machines with increased brilliance, coherence and flux of the X-ray beam are opening up ways to various promising experimental techniques. However, it also presents novel challenges for X-ray optics (Dimper et al., 2014[Dimper, R., Reichert, H., Raimondi, P., Ortiz, L. S., Sette, F. & Susini, J. (2014). ESRF Upgrade Programme Phase II (2015-2022) Technical Design Study, https://www.esrf.fr/Apache_files/Upgrade/ESRF-orange-book.pdf.]) which should preserve unique radiation properties and be capable of beam transport, nano-focusing, phase-contrast imaging and full-field microscopy. The last issue becomes a complex task as highly coherent X-ray radiation easily interacts on its path with various imperfections of optical elements, spoiling the outgoing wavefront with aberrations.

X-ray refractive lenses (Snigirev et al., 1996[Snigirev, A., Kohn, V., Snigireva, I. & Lengeler, B. (1996). Nature, 384, 49-51.]) are used as standard beamline optical elements (Lengeler et al., 1999[Lengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153-1167.], 2005[Lengeler, B., Schroer, C. G., Kuhlmann, M., Benner, B., Günzler, F. T., Kurapova, O., Zontone, F., Snigirev, A. & Snigireva, I. (2005). J. Phys. D Appl. Phys. 38, A218-A222.]; Aristov et al., 2000[Aristov, V., Grigoriev, M., Kuznetsov, S., Shabelnikov, L., Yunkin, V., Weitkamp, T., Rau, C., Snigireva, I., Snigirev, A., Hoffmann, M. & Voges, E. (2000). Appl. Phys. Lett. 77, 4058-4060.]; Snigirev et al., 2007[Snigirev, A., Hustache, R., Duboc, P., Massonnat, J.-Y., Claustre, L., Van Vaerenbergh, P., Snigireva, I., Grigoriev, M. & Yunkin, V. (2007). Proc. SPIE, 6705, 670511.]; Polikarpov et al., 2016[Polikarpov, M., Polikarpov, V., Snigireva, I. & Snigirev, A. (2016). Phys. Procedia, 84, 213-220.]) at many third-generation synchrotrons. Nowadays, beryllium and silicon lenses are the most commonly used owing to their well developed manufacturing process. Nevertheless, both types of lenses have their drawbacks.

Rotationally parabolic beryllium lenses are made of a polycrystalline metal with grain sizes of 10–100 µm. Hence, the grain structure introduces parasitic phase contrast to the propagating wavefront, reducing the overall image quality in the full-field microscopic regime. In the context of nano-focusing applications, the diffraction-limited lens resolution is inversely proportional to the numerical aperture, which is the ratio of the lens aperture divided by its focal distance. The focal distance decreases with a decrease in the lens radius which allows one to achieve a higher resolution when approaching the diffraction limit. However, currently it is technologically difficult to produce beryllium lenses with a radius smaller than 50 µm, which in its turn limits the potentially achievable lens resolution to 50 nm, while only 100 nm has been yet experimentally demonstrated (Snigireva et al., 2011[Snigireva, I., Vaughan, G. B. M., Snigirev, A., McNulty, I., Eyberger, C. & Lai, B. (2011). AIP Conf. Proc. 1365, 188-191.]; Lengeler, 2004[Lengeler, B. (2004). AIP Conf. Proc. 705, 748-751.]; Snigirev & Snigireva, 2008[Snigirev, A. & Snigireva, I. (2008). C. R. Phys. 9, 507-516.]).

In contrast, parabolic silicon lenses are manufactured using the Bosch process (Aristov et al., 2001[Aristov, V. V., Grigoriev, M., Kuznetsov, S., Shabelnikov, L., Yunkin, V., Rau, C., Snigirev, A., Snigireva, I., Weitkamp, T., Hoffmann, M. & Voges, E. I. (2001). Proc. SPIE, 4145, 285-293.]). They have ultra-small radii of about 5 µm and are capable of focusing down to 50 nm (Schroer et al., 2005[Schroer, C. G., Kurapova, O., Patommel, J., Boye, P., Feldkamp, J., Lengeler, B., Burghammer, M., Riekel, C., Vincze, L., Van Der Hart, A. & Küchler, M. (2005). Appl. Phys. Lett. 87, 1-3.]). Made of monocrystalline material, silicon lenses exhibit high structural perfection and are transparent to X-rays, thereby not introducing any parasitic phase contrast. However, as a consequence of the manufacturing by lithographic and etching processes, silicon lenses can only be structured along one dimension. This strongly limits their versatility and thus potential applicability for microscopy. When one needs two-dimensional focusing [e.g. for ptychographic or coherent diffraction imaging applications (Marchesini et al., 2003[Marchesini, S., He, H., Chapman, N., Hau-Riege, P., Noy, A., Howells, R., Weierstall, U. & Spence, H. (2003). Phys. Rev. B, 68, 140101.])], silicon lenses require complex cross-alignment.

For the aforementioned applications, polymer refractive lenses were recently demonstrated to be efficient (Petrov et al., 2017[Petrov, A. K., Bessonov, V. O., Abrashitova, K. A., Kokareva, N. G., Safronov, K. R., Barannikov, A. A., Ershov, P. A., Klimova, N. B., Lyatun, I. I., Yunkin, V. A., Polikarpov, M., Snigireva, I., Fedyanin, A. A. & Snigirev, A. (2017). Opt. Express, 25, 14173-14181.]; Sanli et al., 2018[Sanli, U. T., Ceylan, H., Bykova, I., Weigand, M., Sitti, M., Schütz, G. & Keskinbora, K. (2018). Adv. Mater. 30, 1802503.]; Mirzaeimohri et al., 2018[Mirzaeimoghri, M., Morales Martinez, A., Panna, A., Bennett, E. E., Lucotte, B. M., DeVoe, D. L. & Wen, H. (2018). PLoS One, 13, e0203319.]). Produced from a homogeneous polymer material by additive manufacturing, polymer lenses have compact dimensions, small radii and two-dimensional parabolic shapes. Thus, they can potentially be used for both uniform nano-focusing and short-focus high-resolution full-field microscopy. In the present article, we investigate their focusing and imaging performance on third-generation synchrotrons. We also evaluate their radiation stability.

2. Materials and methods

The experiments presented in this paper were carried out at the ID13 beamline of The European Synchrotron, where an in-vacuum undulator produces X-ray radiation 30 µm × 150 µm in the vertical and horizontal directions at the source point, respectively. Then, X-rays pass through the cryogenically cooled Si(111) channel-cut monochromator to the experimental hutch, located 90 m from the source. We used 12.7 keV photons in all experimental modes of this article.

In some parts of the experimental procedure, we used pre-collimation with a white-beam compound refractive lens (CRL), consisting of three beryllium double-concave parabolic refractive lenses with radius R of 200 µm. The CRL was located 28.5 m from the undulator source and upstream of the monochromator. The pre-focusing CRL focused 12.7 keV X-rays to a distance of 35.5 m downstream, thus increasing the photon flux at the sample position.

The sample environment was represented with several micro-manipulators that could carry pinholes or test objects for imaging and microscopy. The manipulators had three orthogonal translations for the perfect sample alignment and were located upstream of the objective nano-goniometer.

As an objective lens, we used a polymer compound refractive lens (PCRL30, Fig. 1[link]), comprised of 30 individual double-concave parabolic polymer lenses, described by Petrov et al. (2017[Petrov, A. K., Bessonov, V. O., Abrashitova, K. A., Kokareva, N. G., Safronov, K. R., Barannikov, A. A., Ershov, P. A., Klimova, N. B., Lyatun, I. I., Yunkin, V. A., Polikarpov, M., Snigireva, I., Fedyanin, A. A. & Snigirev, A. (2017). Opt. Express, 25, 14173-14181.]). The printing parameters were 13 mW and 400 µm s−1. The polymer material had the formula C21H36O8SiS and a density of 1.2 g cm−3. Each single lens was expected to have a radius of curvature, R, of 5 µm and a physical aperture, A, of 24 µm. According to the calculations, the focal distance and the diffraction-limited resolution (Schroer et al., 2013[Schroer, C. G., Brack, F.-E., Brendler, R., Hönig, S., Hoppe, R., Patommel, J., Ritter, S., Scholz, M., Schropp, A., Seiboth, F., Nilsson, D., Rahomäki, J., Uhlén, F., Vogt, U., Reinhardt, J. & Falkenberg, G. (2013). Proc. SPIE, 8848, 884807.]) of the PCRL30 at 12.7 keV X-ray energy were expected to be F = 5 cm and S = 70 nm, respectively. PCRL30 was installed on the six-axis nano-goniometer allowing for all necessary translations and rotations with nanometre accuracy.

[Figure 1]
Figure 1
Scanning electron micrograph of (a) the PCRL30 and (b) one individual lens.

For image acquisition, we used the PCO-2000 CCD camera coupled with a monochromatic beam optical system (Optique Peter) and a 20-fold objective. It gave an effective pixel size of 0.37 µm within the 2048 pixels × 2048 pixels field of view. The detector was mounted downstream of the PCRL30 objective at a separate bench with all necessary translations, allowing us to vary the PCRL30-to-detector distance from several millimetres to 3 m.

3. X-ray focusing

First of all, we tested PCRL30 in focusing mode (Fig. 2[link]) to check its optical properties such as focal length, spot size and absence (or presence) of aberrations. Knowing that the ideal PCRL30 would have a focal distance of 5 cm, we scanned the space near this position along the lens optical axis, recording images of the focal spot with the detector. A strong astigmatism was discovered, which manifested itself in the rays propagating in two perpendicular planes and having different foci (Hecht, 2002[Hecht, E. (2002). Optics, 4th ed. New York: Addison-Wesley.]). Thus, vertical and horizontal lines of the object were in sharp focus at two different distances: Fv = 37 mm and Fh = 45 mm, respectively [Figs. 3[link](a) and 3(b)]. At these distances, the spots had an equal minimum size of 1.3 µm (full width at half-maximum; FWHM), which is significantly larger than the diffraction limit S of 70 nm even if one takes into account the finite camera pixel of 0.37 µm.

[Figure 2]
Figure 2
Layout of the focusing experiment.
[Figure 3]
Figure 3
X-ray images of lens sharp foci in the (a) vertical (Fv = 37 mm) and (b) horizontal (Fh = 45 mm) directions. (c) The circle of least confusion is located in between at F1 = 41 mm.

Between Fv and Fh, we found the circle of least confusion [Fig. 3[link](c)]. It was located at a distance of F1 = 41 mm, where PCRL30 focused the radiation symmetrically into a 1.6 µm × 1.6 µm (FWHM) spot. Yet the experimentally determined value F1 differs from the expected value of F = 50 mm, indicating that each single biconcave lens has an average curvature radius R = 4 µm that is smaller than designed value of 5 µm.

The astigmatism can be described with the relative parameter Δ, which shows an offset of both foci and lens curvature radii from their average values,

[\Delta = \left|{{F_{\rm v}-F_{\rm h}}\over {F_{\rm v}+F_{\rm h}}}\right| \times 100\% = \left|{{R_{\rm v}-R_{\rm h}}\over {R_{\rm v}+R_{\rm h}}}\right| \times 100\%. \eqno (1)]

In the present case, Δ1 = 10%, meaning that lens radii in the vertical and horizontal directions are different and equal to Rv = 3.6 µm and Rh = 4.4 µm, respectively.

The presence of astigmatism can account for the manufacturing procedure. During the two-photon polymerization, a minimal polymerizable volume (voxel) is an ellipsoid of revolution (spheroid), vertically elongated along the optical axis of the laser beam. The elongation depends on the laser irradiation power and becomes more pronounced with its increase (Sakellari et al., 2012[Sakellari, I., Kabouraki, E., Gray, D., Purlys, V., Fotakis, C., Pikulin, A., Bityurin, N., Vamvakaki, M. & Farsari, M. (2012). ACS Nano, 6, 2302-2311.]). The further lens shape is an envelope of elementary spheroids with the centers lying on a given parabolic trajectory [Fig. 4[link](a)]. Estimating the voxel shape to be a circle (with radius a = b = 0.3 µm) in the horizontal plane [Fig. 4[link](b)] and a vertically elongated ellipse (a = 0.5 µm and b = 0.3 µm) in the vertical plane, the shape of the envelope y may be calculated using the following system of equations for a one-parameter family of plane curves,

[\eqalign {f(x,y,t) = 0, \cr f'(x,y,t) = 0.} \eqno(2)]

For our specific cases of an ellipse and a circle, this system transforms to

[\eqalign {{{{{{\left({x - t} \right)}^2}} \over {{a^2}}} + {{{{\left({y - {{{t^2}} / {2{R_0}}}} \right)}^2}} \over {b^2}} = 1}, \cr {{{\left({x - t} \right)} \over {{a^2}}} + {t \over {\;{R_0}}}{{\left({y - {{t^2} / {2{R_0}}}} \right)} \over {{b^2}}} = 0,}} \eqno(3)]

where R0 = 5 µm is the designed lens parabolic parameter (radius); (t,   t2 / 2R0) and (x, y) are the plane coordinates of the voxel center and envelope, respectively [Fig. 4[link](a)]. After solving equation (3[link]), we found envelope shapes for both the vertical [yv(x)] and horizontal [yh(x)] planes. If every small portion of each envelope is then approximated by its own tangential parabola with curvature radii of Rh,v(x) = 1/yh,v′′(x), the radii will take the values that are shown in Fig. 4[link](c). These numbers mean that the lens shape strongly deviates from the ideal one in the center, while becoming closer to the designed one towards the edges.

[Figure 4]
Figure 4
Schematic representation of (a) how the resulting lens envelope deviates from the initial laser trajectory due to (b) voxel ellipticity in the vertical plane. (c) Deviation of radii in the vertical and horizontal lens planes from their designed value of 5 µm.

During the subsequent developing step (dissolving of unpolymerized resin), the volume of the polymer material further shrinks due to solvent evaporation. Moreover, the shrinkage of the lens is not isotropic – its bottom is attached to the substrate and cannot change dimensions similarly to the upper part. Altogether, the above leads to the reduction of the lens radii, spherical aberrations and to astigmatism which we have experimentally observed.

4. Full-field X-ray microscopy

PCRL30 was also tested in full-field X-ray microscopy mode (Fig. 5[link]) to refine the maximum resolution of the lens and the image quality. In this mode, we used the pre-focusing CRL located at the optics hutch. A Siemens star and a 25 µm pinhole were also mounted upstream of the PCRL30 on the micro-manipulators. A Siemens star is a test object [Fig. 6[link](a)] that consists of a pattern of `spokes' etched in SiN and covered by a 500 nm layer of Ta (XRESO-50HC, NTT-AT, Japan). The Siemens star was located at a distance L1 of 43 mm from the PCRL30. According to the thin lens formula

[{1 \over F} = {1 \over {L_1}} + {1 \over {L_2}} \eqno(4)]

and assuming that the focal distance F is equal to F1 = 41 mm (as from the previous paragraph), we located the X-ray detector at L2 = 885 mm from the PCRL30. Thus, we achieved a lens magnification of L2/L1 = 20.6, giving an effective detector pixel size of 18 nm.

[Figure 5]
Figure 5
Layout of the experiment in X-ray microscopy mode. Note that parts of the drawing before and after the gap have different scales.
[Figure 6]
Figure 6
(a) Scanning electron and (b) flat-field-corrected X-ray images of the Siemens star. (c) The resolution remains the same after 5.5 h X-ray exposure.

To ensure the best image quality, we used a speckle suppressor (Goikhman et al., 2015[Goikhman, A., Lyatun, I., Ershov, P., Snigireva, I., Wojda, P., Gorlevsky, V., Semenov, A., Sheverdyaev, M., Koletskiy, V. & Snigirev, A. (2015). J. Synchrotron Rad. 22, 796-800.]) – a rotating plate of porous nanoberyllium that homogeneously scatters X-rays upstream of the sample. To improve the contrast and signal-to-noise ratio, we also applied a flat-field correction [Fig. 6[link](b)], dividing the image with the mounted Siemens star by the image with the unmounted one. It allowed us to resolve the 100 nm and 200 nm bars in the vertical and horizontal directions, respectively. This size was much closer to the 70 nm theoretical diffraction limit of the PCRL30. However, different image resolution in the vertical and horizontal directions indicated an asymmetry in the contrast transfer function due to the presence of astigmatism in the optical system.

5. Radiation stability

To estimate the value of possible radiation damage to the lens, we simply exposed PCRL30 to the X-ray beam with a flux density of 7 × 1011 photons s−1 mm−2. The procedure lasted for 5.5 h. During that time, individual lenses of the PCRL30 absorbed an average dose of ∼15 × 106 Gy each. X-ray radiograms of the PCRL30 before and after exposure (Fig. 7[link]) demonstrate non-uniform polymer lens shrinkage; lenses were almost unchanged around the substrate–lens conjunction because of the mechanical stress and therefore higher stability. However, they shrunk significantly around the freestanding top. After the exposure, we roughly estimated the decrease in size to be 17 ± 2% on average, irrespective of the complexity of the lens shape.

[Figure 7]
Figure 7
X-ray radiograms of the PCRL30 were taken under the same experimental conditions (a) before and (b) after 5.5 h X-ray exposure.

We also repeated the focusing experiment to see the change in the focal distance. The PCRL30 focused the radiation in the circle of least confusion with 1.8 µm × 1.8 µm size and at a focal distance F2 = 35 mm. The distance F2 is 15% smaller than F1, which corresponds to our observations of geometrical size reduction. The PCRL30 still had astigmatism, with sharp vertical and horizontal foci at Fv2 = 32 mm and Fh2 = 38 mm, respectively. Estimating the astigmatism again with the parameter Δ, we obtain a value Δ2 = 8.5%.

In subsequent experiments, we studied whether the degraded PCRL30 preserved the imaging qualities. We repeated the X-ray microscopy experiment with the corrected distances L1 = 36.5 mm and L2 = 851 mm [as in equation (4[link])]. As a result, we obtained a 23-fold magnified X-ray image [Fig. 6[link](c)]. The image was almost identical to the one obtained before the radiation damage – 200 nm bars were the last ones clearly visible and the lenses were still successfully working as microscopic objective lenses, even though their radii were non-uniformly reduced.

The experiment has revealed that X-ray irradiation causes a continuous degradation of polymer material, namely polymer chain scission or crosslinking. A detailed description of the lens degradation mechanism requires further research and is beyond the scope of this paper. From a practical point of view, one could only state that, although lenses suffer from continuous degradation-induced shrinkage, there are no volume cracks, voids or pores, because those would be immediately seen (by having a strong effect) on the X-ray micrographs.

6. Conclusions

Our experiments showed that the polymer lenses worked successfully, achieving a resolution of 100–200 nm in full-field imaging mode, which was relatively close to their diffraction limit of 70 nm using the current setup. Nevertheless, lenses had strong astigmatism arising from the manufacturing process, which led to image distortions in microscopy mode. It also reduced their focusing performance, not allowing one to achieve the expected nanometre size of the focal spot.

One could make several changes to improve the lens geometry, radiation stability and overall quality. First of all, the voxel ellipticity could be reduced by decreasing the average dose delivered by the laser in two-photon polymerization. Second, knowing the voxel size and shape, we could introduce the corrected laser trajectory in order to achieve a parabolic envelope as a result. Then, there are several ways to deal with polymerization shrinkage as it is a well known problem in two-photon lithography. One could use special polymer materials with significantly reduced shrinkage (Ovsianikov et al., 2009[Ovsianikov, A., Shizhou, X., Farsari, M., Vamvakaki, M., Fotakis, C. & Chichkov, B. N. (2009). Opt. Express, 17, 2143-2148.]). However, we should pay special attention to the chemical composition of a polymer, because the presence of heavy elements in the CRL will lead to increased X-ray absorption. Another approach could be used to compensate for the shrinkage on the stage of lens design (Sun et al., 2004[Sun, H. B., Suwa, T., Takada, K., Zaccaria, R. P., Kim, M. S., Lee, K. S. & Kawata, S. (2004). Appl. Phys. Lett. 85, 3708-3710.]) – to reduce the mechanical stress in the substrate–lens conjunction with the use of sponge-like or anchor-like supporting structures (Maruo et al., 2009[Maruo, S., Hasegawa, T. & Yoshimura, N. (2009). Opt. Express, 17, 20945-20951.]). This will lead to a more uniformed shrinkage (including that induced by X-ray exposure), thus reducing the astigmatism. Finally, we may improve the mechanical and X-ray radiation resistance of the lens, applying post UV-curing methods with an increased degree of crosslinking while preserving the same resolution of 3D printing (Oakdale et al., 2016[Oakdale, J. S., Ye, J., Smith, W. L. & Biener, J. (2016). Opt. Express, 24, 27077-27086.]).

The improvements proposed above could help to enhance the lens quality, allowing sufficient resolution for nano-focusing applications to be achieved. To summarize, even though the polymer refractive lenses still require some technological improvement, they are already promising as versatile compact devices for X-ray microscopy. Due to their short focal distance and in contrast to beryllium CRLs, they can be easily integrated into most optical schemes. Besides synchrotrons, polymer lenses could also be extremely useful at laboratory X-ray sources where compactness affects the recorded X-ray intensity.

Acknowledgements

We thank A. Petrov (MSU) and ESRF ID13 staff for their help in this research.

Funding information

The research was supported by The Russian Ministry of Education and Science (grant No. 14.Y26.31.0002, 14.W03.31.0008 – lens printing and tests), the Russian Science Foundation (15-12-00065 – scanning electron microscopy), the Russian Academic Excellence Project at IKBFU and by the Quantum technology centre at Moscow State University. VY was supported by the State Task No. 075-00475-19-00.

References

First citationAristov, V., Grigoriev, M., Kuznetsov, S., Shabelnikov, L., Yunkin, V., Weitkamp, T., Rau, C., Snigireva, I., Snigirev, A., Hoffmann, M. & Voges, E. (2000). Appl. Phys. Lett. 77, 4058–4060.  Web of Science CrossRef CAS Google Scholar
First citationAristov, V. V., Grigoriev, M., Kuznetsov, S., Shabelnikov, L., Yunkin, V., Rau, C., Snigirev, A., Snigireva, I., Weitkamp, T., Hoffmann, M. & Voges, E. I. (2001). Proc. SPIE, 4145, 285–293.  CrossRef Google Scholar
First citationDimper, R., Reichert, H., Raimondi, P., Ortiz, L. S., Sette, F. & Susini, J. (2014). ESRF Upgrade Programme Phase II (2015–2022) Technical Design Study, https://www.esrf.fr/Apache_files/Upgrade/ESRF-orange-book.pdfGoogle Scholar
First citationGoikhman, A., Lyatun, I., Ershov, P., Snigireva, I., Wojda, P., Gorlevsky, V., Semenov, A., Sheverdyaev, M., Koletskiy, V. & Snigirev, A. (2015). J. Synchrotron Rad. 22, 796–800.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationHecht, E. (2002). Optics, 4th ed. New York: Addison–Wesley.  Google Scholar
First citationLengeler, B. (2004). AIP Conf. Proc. 705, 748–751.  CrossRef CAS Google Scholar
First citationLengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153–1167.  Web of Science CrossRef IUCr Journals Google Scholar
First citationLengeler, B., Schroer, C. G., Kuhlmann, M., Benner, B., Günzler, F. T., Kurapova, O., Zontone, F., Snigirev, A. & Snigireva, I. (2005). J. Phys. D Appl. Phys. 38, A218–A222.  Web of Science CrossRef CAS Google Scholar
First citationMarchesini, S., He, H., Chapman, N., Hau-Riege, P., Noy, A., Howells, R., Weierstall, U. & Spence, H. (2003). Phys. Rev. B, 68, 140101.  Web of Science CrossRef Google Scholar
First citationMaruo, S., Hasegawa, T. & Yoshimura, N. (2009). Opt. Express, 17, 20945–20951.  Web of Science CrossRef PubMed CAS Google Scholar
First citationMirzaeimoghri, M., Morales Martinez, A., Panna, A., Bennett, E. E., Lucotte, B. M., DeVoe, D. L. & Wen, H. (2018). PLoS One, 13, e0203319.  Web of Science CrossRef PubMed Google Scholar
First citationOakdale, J. S., Ye, J., Smith, W. L. & Biener, J. (2016). Opt. Express, 24, 27077–27086.  Web of Science CrossRef CAS PubMed Google Scholar
First citationOvsianikov, A., Shizhou, X., Farsari, M., Vamvakaki, M., Fotakis, C. & Chichkov, B. N. (2009). Opt. Express, 17, 2143–2148.  Web of Science CrossRef PubMed CAS Google Scholar
First citationPetrov, A. K., Bessonov, V. O., Abrashitova, K. A., Kokareva, N. G., Safronov, K. R., Barannikov, A. A., Ershov, P. A., Klimova, N. B., Lyatun, I. I., Yunkin, V. A., Polikarpov, M., Snigireva, I., Fedyanin, A. A. & Snigirev, A. (2017). Opt. Express, 25, 14173–14181.  Web of Science CrossRef CAS PubMed Google Scholar
First citationPolikarpov, M., Polikarpov, V., Snigireva, I. & Snigirev, A. (2016). Phys. Procedia, 84, 213–220.  CrossRef CAS Google Scholar
First citationSakellari, I., Kabouraki, E., Gray, D., Purlys, V., Fotakis, C., Pikulin, A., Bityurin, N., Vamvakaki, M. & Farsari, M. (2012). ACS Nano, 6, 2302–2311.  Web of Science CrossRef CAS PubMed Google Scholar
First citationSanli, U. T., Ceylan, H., Bykova, I., Weigand, M., Sitti, M., Schütz, G. & Keskinbora, K. (2018). Adv. Mater. 30, 1802503.  Web of Science CrossRef Google Scholar
First citationSchroer, C. G., Brack, F.-E., Brendler, R., Hönig, S., Hoppe, R., Patommel, J., Ritter, S., Scholz, M., Schropp, A., Seiboth, F., Nilsson, D., Rahomäki, J., Uhlén, F., Vogt, U., Reinhardt, J. & Falkenberg, G. (2013). Proc. SPIE, 8848, 884807.  CrossRef Google Scholar
First citationSchroer, C. G., Kurapova, O., Patommel, J., Boye, P., Feldkamp, J., Lengeler, B., Burghammer, M., Riekel, C., Vincze, L., Van Der Hart, A. & Küchler, M. (2005). Appl. Phys. Lett. 87, 1–3.  Web of Science CrossRef Google Scholar
First citationSnigirev, A., Hustache, R., Duboc, P., Massonnat, J.-Y., Claustre, L., Van Vaerenbergh, P., Snigireva, I., Grigoriev, M. & Yunkin, V. (2007). Proc. SPIE, 6705, 670511.  CrossRef Google Scholar
First citationSnigirev, A., Kohn, V., Snigireva, I. & Lengeler, B. (1996). Nature, 384, 49–51.  CrossRef CAS Web of Science Google Scholar
First citationSnigirev, A. & Snigireva, I. (2008). C. R. Phys. 9, 507–516.  Web of Science CrossRef CAS Google Scholar
First citationSnigireva, I., Vaughan, G. B. M., Snigirev, A., McNulty, I., Eyberger, C. & Lai, B. (2011). AIP Conf. Proc. 1365, 188–191.  CrossRef Google Scholar
First citationSun, H. B., Suwa, T., Takada, K., Zaccaria, R. P., Kim, M. S., Lee, K. S. & Kawata, S. (2004). Appl. Phys. Lett. 85, 3708–3710.  Web of Science CrossRef CAS Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds