In-house texture measurement using a compact neutron source

CROSSMARK_Color_square_no_text.svg

aMaterials Sciences Research Center, Japan Atomic Energy Agency, 2-4 Shirakata, Tokai, Ibaraki 319-1195, Japan, and bCenter for Advanced Photonics, RIKEN, 2-1 Hirosawa, Wako, Saitama 351-0198, Japan
*Correspondence e-mail: xu.pingguang@jaea.go.jp

Edited by G. Kostorz, ETH Zurich, Switzerland (Received 30 June 2019; accepted 24 February 2020; online 25 March 2020)

In order to improve the instrumental accessibility of neutron diffraction techniques, many emerging compact neutron sources and in-house neutron diffractometers are being developed, even though the precision level of neutron diffraction experiments performed on such instruments was thought to be incomparable with that of large-scale neutron facilities. As a challenging project, the RIKEN accelerator-driven compact neutron source (RANS) was employed here to establish the technical environment for texture measurements, and the recalculated pole figures and orientation distribution functions of an interstitial-free steel sheet obtained from RANS were compared with the results from another two neutron diffractometers well established for texture measurement. These quantitative comparisons revealed that the precise neutron diffraction texture measurement at RANS has been realized successfully, and the fine region division of the neutron detector panel is invaluable for improving the stereographic resolution of texture measurements. Moreover, through selectively using the parts of the obtained neutron diffraction patterns that exhibit good statistics, the Rietveld texture analysis improves the reliability of the texture measurement to a certain extent. These technical research results may accelerate the development of other easily accessible techniques for evaluation of engineering materials using compact neutron sources, and also help to improve the data-collection efficiency for various time-resolved scattering experiments at large-scale neutron facilities.

1. Introduction

Neutron diffraction is widely thought of as a powerful probe for texture evaluation of advanced materials and even coarse-grained geological samples (Wenk, 1991[Wenk, H.-R. (1991). J. Appl. Cryst. 24, 920-927.]; Wenk et al., 1991[Wenk, H. R., Larson, A. C., Vergamini, P. J. & Schultz, A. J. (1991). J. Appl. Phys. 70, 2035-2040.]; Brokmeier, 1999[Brokmeier, H. G. (1999). Textures Microst. 33, 13-33.]; Jansen et al., 2000[Jansen, E., Schäfer, W. & Kirfel, A. (2000). J. Struct. Geol. 22, 1559-1564.]), because its large spot size and high penetration enable one to acquire bulk-averaged orientation information from a polycrystalline sample with a volume of the order of 1 cm3 (Vogel, 2013[Vogel, S. C. (2013). ISRN Mater. Sci. 302408.]; Malamud et al., 2014[Malamud, F., Santisteban, J. R., Vicente Alvarez, M. A., Bolmaro, R., Kelleher, J., Kabra, S. & Kockelmann, W. (2014). J. Appl. Cryst. 47, 1337-1354.]; Yusuf & Kumar, 2017[Yusuf, S. M. & Kumar, A. (2017). Appl. Phys. Rev. 4, 031303.]; Hayashi et al., 2018[Hayashi, M., Root, J. H., Rogge, R. B. & Xu, P. G. (2018). QuBS, 2, 21.]). Other crystalline information, e.g. lattice parameters, phase fractions, strains and stresses, may be also extracted through a combined Rietveld analysis (Lutterotti et al., 1997[Lutterotti, L., Matthies, S., Wenk, H. R., Schultz, A. J. & Richardson, J. (1997). J. Appl. Phys. 81, 594-600.], 2004[Lutterotti, L., Chateigner, D., Ferrari, S. & Ricote, J. (2004). Thin Solid Films, 450, 34-41.]; Wenk et al., 2003[Wenk, H. R., Lutterotti, L. & Vogel, S. C. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]; Xu et al., 2018[Xu, P. G., Harjo, S., Ojima, M., Suzuki, H., Ito, T., Gong, W., Vogel, S. C., Inoue, J., Tomota, Y., Aizawa, K. & Akita, K. (2018). J. Appl. Cryst. 51, 746-760.]). These bulk-averaged textures and related crystalline information are extremely valuable for developing various advanced structural and functional materials (Xu et al., 2015[Xu, P. G., Yin, J. & Zhang, S. Y. (2015). Acta Metall. Sin. 51, 1297-1305.]; Liss et al., 2016[Liss, K. D., Funakoshi, K., Dippenaar, R. J., Higo, Y., Shiro, A., Reid, M., Suzuki, H., Shobu, T. & Akita, K. (2016). Metals, 6, 165.]; Yusuf & Kumar, 2017[Yusuf, S. M. & Kumar, A. (2017). Appl. Phys. Rev. 4, 031303.]; Mo et al., 2018[Mo, F. J., Sun, G. A., Li, J., Zhang, C. S., Wang, H., Chen, Y., Liu, Z., Yang, Z. K., Li, H. J., Yang, Z. L., Pang, B. B., Huang, Y. L., Tian, Y., Gong, J., Chen, B. & Peng, S. M. (2018). QuBS, 2, 15.]; Takajo et al., 2018[Takajo, S., Hollis, K. J., Cummins, D. R., Tegtmeier, E. L., Dombrowski, D. E. & Vogel, S. C. (2018). QuBS, 2, 12.]). For example, with the aim of further improving the strength–ductility balance of transformation-induced-plasticity low-alloy steels, the lattice parameters and volume fraction of retained austenite have been precisely measured through a neutron diffraction texture measurement (Xu et al., 2017[Xu, P. G., Tomota, Y., Arakaki, Y., Harjo, S. & Sueyoshi, H. (2017). Mater. Charact. 127, 104-110.]; Tomota et al., 2017[Tomota, Y., Sekido, N., Xu, P. G., Kawasaki, T., Harjo, S., Tanaka, M., Shinohara, T., Su, Y. H. & Taniyama, A. (2017). Tetsu Hagane, 103, 570-578.]). If the effect of residual stresses can be omitted or reasonably considered, the carbon concentration of retained austenite may be evaluated indirectly from the lattice parameters of austenite and ferrite through experiential relationships (Onink et al., 1993[Onink, M., Brakman, C. M., Tichelaar, F. D., Mittemeijer, E. J., van der Zwaag, S., Root, J. H. & Konyer, N. B. (1993). Scr. Metall. Mater. 29, 1011-1016.]; Sugimoto et al., 2003[Sugimoto, K., Hayakawa, A., Hojo, T., Hashimoto, S. & Ikeda, S. (2003). Tetsu Hagane, 89, 1233-1239.]; Chen et al., 2006[Chen, S. C., Tomota, Y., Shiota, Y., Toomine, Y. & Kamiyama, T. (2006). Tetsu Hagane, 92, 557-561.]). With the aim of understanding the structural and magnetic properties and further optimizing the magnetic phase microstructure at the nanoscale, neutron diffraction texture measurements have been employed to investigate permanent magnets (Wroblewski et al., 1999[Wroblewski, T., Jansen, E., Schäfer, W. & Skowronek, R. (1999). Nucl. Instrum. Methods Phys. Res. A, 423, 428-434.]), morphotropic piezoceramics (Hinterstein et al., 2015[Hinterstein, M., Hoelzel, M., Rouquette, J., Haines, J., Glaum, J., Kungl, H. & Hoffman, M. (2015). Acta Mater. 94, 319-327.]), and other advanced magnetic materials for modern computers, medical instruments, ultrasonic motors, electric generators, telecommunications and transportation (Yusuf & Kumar, 2017[Yusuf, S. M. & Kumar, A. (2017). Appl. Phys. Rev. 4, 031303.]).

In order to obtain good straight and strong neutron beams for precise neutron diffraction measurements, nuclear reactors, proton accelerators or other large-scale neutron facilities are usually thought essential (Liss, 2017a[Liss, K. D. (2017a). Metals, 7, 266.],b[Liss, K. D. (2017b). QuBS, 1, 1.]; Argyriou & Allen, 2018[Argyriou, D. N. & Allen, A. J. (2018). J. Appl. Cryst. 51, 567-569.]). A huge budget is required to cover the establishment of the related hardware and software, the general maintenance, and the ordinary running costs. As a result, the number of available neutron beam instruments across the world is quite limited (Jansen et al., 2004[Jansen, E., Schäfer, W. & Kirfel, A. (2004). Physica B, 350, E569-E572.]). Though some rapid measurement platforms based on robotic sample-exchange systems (Hoshikawa et al., 2009[Hoshikawa, A., Ishigaki, T., Nagai, M., Kobayashi, K., Sagehashi, H., Kamiyama, T., Yonemura, M., Aizawa, K., Sakuma, T., Tomota, Y., Arai, M., Hayashi, M., Ebata, K., Takano, Y. & Kasao, T. (2009). Nucl. Instrum. Methods Phys. Res. A, 600, 203-206.]; Reiche & Vogel, 2010[Reiche, H. M. & Vogel, S. C. (2010). Rev. Sci. Instrum. 81, 093302.]; Brokmeier et al., 2011[Brokmeier, H. G., Gan, W., Randau, C., Völler, M., Rebelo-Kornmeier, J. & Hofmann, M. (2011). Nucl. Instrum. Methods Phys. Res. A, 642, 87-92.]) have been attempted, acquiring the necessary neutron beam time is still highly competitive. If no progress is made in increasing the availability of neutron diffractometers, such high competition will continue for several decades because of the gradually increasing application needs. These applications may include averaged texture optimization for developing low-carbon strip-cast steels (Xu et al., 2006[Xu, P. G., Yin, F. X. & Nagai, K. (2006). Mater. Sci. Eng. A, 441, 157-166.]), heavy-gauge shipbuilding steel plates (Hase et al., 2016[Hase, K., Handa, T., Eto, T., Yamamura, N. & Aoki, M. (2016). Mater. Jpn, 56, 82-84.]; Nishimura et al., 2007[Nishimura, K., Handa, T. & Hashimoto, M. (2007). JFE GIHO, 18, 18-22.]; Nishimura & Takeuchi, 2014[Nishimura, K. & Takeuchi, Y. (2014). Tetsu Hagane, 100, 1097-1103.]), delayed fracture-resistant ultrahigh streel plates (Xu et al., 2019[Xu, P. G., Ishijima, Y., Qiu, H., Morooka, S., Gong, W. & Harjo, S. (2019). J-PARC Symposium 2019, 23-26 September 2019. Poster.]), lean duplex stainless steels with low Ni content (Takahashi et al., 2020[Takahashi, O., Shibui, Y., Xu, P. G., Harjo, S., Suzuki, T. & Tomota, Y. (2020). QuBS, 4, 16.]) and other formable high-strength lightweight metallic materials. Moreover, high statistically averaged textures are also expected during the numerical simulation and process optimization for press-forming behavior of high-strength metallic materials (Delannay et al., 2006[Delannay, L., Jacques, P. J. & Kalidindi, S. R. (2006). Int. J. Plast. 22, 1879-1898.]; Takamura et al., 2013[Takamura, M., Nishimura, S. & Sunaga, H. (2013). Key Eng. Mater. 554-557, 1331-1337.]; Choi et al., 2013[Choi, S. H., Kim, E. Y., Woo, W., Han, S. H. & Kwak, J. H. (2013). Int. J. Plast. 45, 85-102.]; Hama et al., 2015[Hama, T., Matsudai, R., Kuchinomachi, Y., Fujimoto, H. & Takuda, H. (2015). ISIJ Int. 55, 1067-1075.]). The research and development significance of alternative compact neutron sources has already been emphasized by the JFE Techno-Research Corporation (Sato et al., 2017[Sato, K., Nakamichi, N. & Sueyoshi, H. (2017). Phys. World, October 2017, https://live.iop-pp01.agh.sleek.net/physicsworld/reader/#!edition/editions_neutron_2017/article/page-22576.]), because, as a common measurement tool within manufacturing industries, the neutron diffraction technique will accelerate the innovation of steel production technology through various strong industry–academic community links.

Fortunately, many compact neutron source facilities are being developed around the world (Anderson et al., 2016[Anderson, I. S., Andreani, C., Carpenter, J. M., Festa, G., Gorini, G., Loong, C. K. & Senesi, R. (2016). Phys. Rep. 654, 1-58.]), and the Union for Compact Accelerator-Driven Neutron Sources has also been organized to promote technical information exchange about small accelerator-based neutron sources and related neutron scattering techniques. Compared with the large-scale neutron source facilities, these compact neutron source facilities (Anderson et al., 2016[Anderson, I. S., Andreani, C., Carpenter, J. M., Festa, G., Gorini, G., Loong, C. K. & Senesi, R. (2016). Phys. Rep. 654, 1-58.]), such as the RIKEN accelerator-driven compact neutron source (RANS) (Yamagata et al., 2015[Yamagata, Y., Hirota, K., Ju, J., Wang, S., Morita, S., Kato, J., Otake, Y., Yaketani, A., Seki, Y., Yamada, M., Ota, H., Bautista, U. & Jia, Q. G. (2015). J. Radioanal. Nucl. Chem. 305, 785-794.]; Otake et al., 2017[Otake, Y., Seki, Y., Wakabayashi, Y., Ikeda, Y., Hashiguchi, T., Yoshimura, Y., Sunaga, H., Taketani, A., Mizuta, M., Oshima, Y. & Ishida, M. (2017). J. Disaster Res. 12, 585-592.]; Otake, 2018a[Otake, Y. (2018a). Applications of Laser-Driven Particle Acceleration, edited by P. Bolton, K. Parodi & J. Schreiber, ch. 19, pp. 291-314. Boca Raton: CRC Press.],b[Otake, Y. (2018b). Encyclopedia of Analytical Chemistry, edited by R. A. Meyers, Neutron Diffraction and Neutron Imaging with Compact Neutron Source. Chichester: John Wiley & Sons.]) and the Jülich High-Brilliance Neutron Source (Rücker et al., 2016[Rücker, U., Cronert, T., Voigt, J., Dabruck, J. P., Doege, P. E., Ulrich, J., Nabbi, R., Beßler, Y., Butzek, M., Büscher, M., Lange, C., Klaus, M., Gutberlet, T. & Brückel, T. (2016). Eur. Phys. J. Plus, 131, 19.]), usually have a shorter flight beam path, a larger beam divergence angle, a lower beam power output and higher background noise. Consequently, the most urgent technical problem is how to realize precise neutron diffraction measurements using a weak neutron beam facility.

Recently, RANS has been developed as an in-house multipurpose neutron facility for nondestructive inspection (Otake et al., 2017[Otake, Y., Seki, Y., Wakabayashi, Y., Ikeda, Y., Hashiguchi, T., Yoshimura, Y., Sunaga, H., Taketani, A., Mizuta, M., Oshima, Y. & Ishida, M. (2017). J. Disaster Res. 12, 585-592.]; Ikeda et al., 2017[Ikeda, Y., Otake, Y. & Mizuta, M. (2017). J. Adv. Concr. Technol. 15, 603-609.]), steel corrosion imaging (Taketani et al., 2017[Taketani, A., Yamada, M., Ikeda, Y., Hashiguchi, T., Sunaga, H., Wakabayashi, Y., Ashigai, S., Takamura, M., Mihara, S., Yanagimachi, S., Otake, Y., Wakabayashi, T., Kono, K. & Nakayama, T. (2017). ISIJ Int. 57, 155-161.]) and the volume fraction analysis of retained austenite (Ikeda et al., 2016[Ikeda, Y., Taketani, A., Takamura, M., Sunaga, H., Kumagai, M., Oba, Y., Otake, Y. & Suzuki, H. (2016). Nucl. Instrum. Methods Phys. Res. A, 833, 61-67.], 2018[Ikeda, Y., Takamura, M., Hakoyama, T., Otake, Y., Kumagai, M. & Suzuki, H. (2018). Tetsu Hagane, 104, 138-144.]). Because the texture optimization of advanced materials has attracted wider attention for better strength–ductility balance and functional performance, we have attempted to establish a neutron texture measurement environment at RANS through various technical optimizations. A well evaluated interstitial-free (IF) steel sheet sample (Xu et al., 2008[Xu, P. G., Tomota, Y., Suzuki, H., Suzuki, T., Machiya, S. & Yin, F. X. (2008). Mater. Trans. 49, 2033-2039.]) was employed here as a typical texture material for reference, and the reliability of the texture measurement was evaluated through carefully comparing the bulk textures obtained at RANS and two other well established time-of-flight neutron diffractometers. Such technical research was carried out to accelerate the establishment of other easily accessible techniques for evaluation of engineering materials using neutron diffraction and to promote the wider application of compact neutron diffraction techniques.

2. Experimental procedures

2.1. Low-energy nuclear reaction for producing neutrons

In contrast to the conventional neutron production method using the thermal chain fission reaction in a 235U or 239Pu nuclear reactor or using a spallation reaction driven by a high-energy proton accelerator (Anderson et al., 2016[Anderson, I. S., Andreani, C., Carpenter, J. M., Festa, G., Gorini, G., Loong, C. K. & Senesi, R. (2016). Phys. Rep. 654, 1-58.]), RANS has been developed to produce neutrons through the nuclear reaction between the beryllium metallic target 49Be(p,n) and low-energy protons (Hawkesworth, 1977[Hawkesworth, M. R. (1977). At. Energy Rev. 15, 169-220.]), as follows:

[\rm {}_{1}^{1}H+{}_{4}^{9}Be(p,n)\to {}_{5}^{9}B(p,n)+{}_{0}^{1}n-1.85 \,\,MeV. \eqno(1)]

Here, the negative reaction energy means that this nuclear reaction is endothermic, and it requires a net energy input with a critical energy of 1.85 MeV. However, considering the initial kinetic energy of the protons, the practical threshold energy is Eth = 2.06 MeV (Hawkesworth, 1977[Hawkesworth, M. R. (1977). At. Energy Rev. 15, 169-220.]; Yamagata et al., 2015[Yamagata, Y., Hirota, K., Ju, J., Wang, S., Morita, S., Kato, J., Otake, Y., Yaketani, A., Seki, Y., Yamada, M., Ota, H., Bautista, U. & Jia, Q. G. (2015). J. Radioanal. Nucl. Chem. 305, 785-794.]; Anderson et al., 2016[Anderson, I. S., Andreani, C., Carpenter, J. M., Festa, G., Gorini, G., Loong, C. K. & Senesi, R. (2016). Phys. Rep. 654, 1-58.]; Hirota, 2018[Hirota, K. (2018). Hamon, 28, 29-32.]).

Generally, a higher net release of nuclear energy corresponds to stronger radiation, especially for the thermal fission reaction at a nuclear reactor. So, the low-energy nuclear reaction for neutron production at RANS simplifies the radiation shielding of the target station and other equipment (Yamagata et al., 2015[Yamagata, Y., Hirota, K., Ju, J., Wang, S., Morita, S., Kato, J., Otake, Y., Yaketani, A., Seki, Y., Yamada, M., Ota, H., Bautista, U. & Jia, Q. G. (2015). J. Radioanal. Nucl. Chem. 305, 785-794.]; Ma et al., 2018[Ma, B. L., Otake, Y., Wang, S., Sunaga, H., Yamagata, Y., Taketani, A., Hu, H. S., Jia, Q. G., Hu, G. & Bautista, U. (2018). Appl. Radiat. Isot. 137, 129-138. ]). The RANS incident proton energy is 7 MeV, so the total neutron yield is ∼1012 n s−1 at a full-power averaged proton beam current of 100 µA (Otake, 2018a[Otake, Y. (2018a). Applications of Laser-Driven Particle Acceleration, edited by P. Bolton, K. Parodi & J. Schreiber, ch. 19, pp. 291-314. Boca Raton: CRC Press.]).

2.2. RIKEN accelerator-driven compact neutron source

Fig. 1[link](a) shows the RANS system, about 15 m total length and 25 ton total weight, consisting of an ion source (30 keV), linear proton accelerators [0.03 → 3.5 MeV radio-frequency quadrupole (RFQ) and 3.5 → 7.0 MeV drift-tube linac (DTL))], a target station for neutron beam generation, movable neutron guide tubes, a sample stage and a neutron detector. Such compact equipment helps the rapid exchange of experimental setup between neutron diffraction, neutron scattering and neutron imaging.

[Figure 1]
Figure 1
Neutron diffraction texture experiment at RANS. (a) Overview. (b) View of the sample setup. The rolling direction of the reference sample is parallel to the mouth-opening direction of the Eulerian cradle in the horizontal plane. (c) Geometric parameters for the neutron flight paths (L1, L2) and the incident beam slit size (30 × 30 mm).

In the target station, 50 mm-diameter 0.3 mm-thick 49Be(p,n) metal is employed as the target film, whose thickness is sufficient to slow down the proton beam. A 4.5 mm-thick vanadium plate is used as the backing material of the target beryllium film, because vanadium has the highest hydrogen diffusional coefficient and the highest hydrogen embrittlement resistance. This vanadium plate is further cooled by water flowing in a 5 mm-thick titanium cavity (Yamagata et al., 2015[Yamagata, Y., Hirota, K., Ju, J., Wang, S., Morita, S., Kato, J., Otake, Y., Yaketani, A., Seki, Y., Yamada, M., Ota, H., Bautista, U. & Jia, Q. G. (2015). J. Radioanal. Nucl. Chem. 305, 785-794.]). Reaction (1)[link] generates fast neutrons with a maximum energy of about 5 MeV and with a flux peak at around 1.5 MeV (Otake et al., 2017[Otake, Y., Seki, Y., Wakabayashi, Y., Ikeda, Y., Hashiguchi, T., Yoshimura, Y., Sunaga, H., Taketani, A., Mizuta, M., Oshima, Y. & Ishida, M. (2017). J. Disaster Res. 12, 585-592.]). These fast neutrons are further slowed down by using a 40 mm-thick polyethyl­ene moderator (the moderator may be replaced with 20 or 60 mm-thick moderators, according to experimental needs) to obtain the thermal neutrons. Cubic graphite reflector blocks (of side length 400 mm) surrounding the polyethyl­ene moderator are employed to increase the low-energy neutron flux.

Fig. 1[link](b) gives an overview of the experimental layout for the neutron diffraction texture measurement. The movable sample stage is surrounded by movable 5 mm-thick B4C rubber shielding plates with reduced background radiation disturbance. The incident neutron flight path (L1) from the sample center to the surface of the polyethyl­ene moderator and the diffracted neutron flight path (L2) from the sample center to the flattened neutron detector array may be freely adjusted together with the multiple-purpose sample stages for a specific neutron scattering experiment. Fig. 1[link](c) illustrates the geometric parameters L1 = 5250 mm and L2 = 315 mm employed in this texture measurement. According to our practical experience, these allow us to realize a good balance between high intensity and high time-of-flight resolution (Johnson & Daymond, 2002[Johnson, M. W. & Daymond, M. R. (2002). J. Appl. Cryst. 35, 49-57.]). Optimization of the compact neutron source and the moderator system for stronger neutron beam flux and higher time-of-flight resolution is still ongoing.

Fig. 2[link] shows the neutron energy spectra of fast and thermal neutrons measured at two `feature' positions of the RANS, as marked by red points in Fig. 1[link](c): (a) the exit of the target station and (b) the nominal center of the sample stage, 1.4 and 5.0 m from the moderator surface, respectively. The peak energy of the thermal neutrons within an energy range of E = 10–360 meV is around 50 meV, suitable for general diffraction experiments (Otake et al., 2017[Otake, Y., Seki, Y., Wakabayashi, Y., Ikeda, Y., Hashiguchi, T., Yoshimura, Y., Sunaga, H., Taketani, A., Mizuta, M., Oshima, Y. & Ishida, M. (2017). J. Disaster Res. 12, 585-592.]), corresponding to a maximum beam flux at 1.28 Å within a wavelength range of 2.86–0.48 Å according to the equation λ (Å) = [81.787/E (meV)]1/2 (Windsor, 1981[Windsor, C. G. (1981). Pulsed Neutron Scattering pp. 34-35, 88, 232, 268 and 288. London: Taylors & Francis Ltd.]). The peak energy of the fast neutrons within an energy range of E = 0.1–5.0 MeV is around 2.0 MeV, suitable for fast neutron imaging (Taketani et al., 2017[Taketani, A., Yamada, M., Ikeda, Y., Hashiguchi, T., Sunaga, H., Wakabayashi, Y., Ashigai, S., Takamura, M., Mihara, S., Yanagimachi, S., Otake, Y., Wakabayashi, T., Kono, K. & Nakayama, T. (2017). ISIJ Int. 57, 155-161.]). The full-power averaged current of the proton beam is 100 µA, where the beam pulse width and the pulse frequency can be adjusted in the ranges 8–180 µs and 10–180 Hz, respectively. During this texture measurement, the averaged current of the proton beam flux was about 32 µA, the beam pulse width was 30 µs, the pulse frequency was 115 Hz and the duty cycle was about 1.3%. Moreover, because of the clear difference in the neutron flight distances of the two feature positions, the beam flux at the nominal center of the sample stage (5.0 m) is lower than that at the exit of the target station (1.4 m), about 4.2 × 104 versus 7.7 × 105 n [s cm2 (100 µA) lethargy]−1, respectively, for the thermal neutrons and about 1.2 × 105 versus 2.0 × 106 n [s cm2 (100 µA) lethargy]−1 for the fast neutrons. The RANS thermal neutron flux at the nominal center of the sample stage is about 0.43%, 0.17% of the corresponding values from the TAKUMI neutron diffractometer {about 9.6 × 106 n [s cm2 (66 µA)]−1 in this paper; the full-power design value will be about 4.8 × 107 n [s cm2 (330 µA)]−1} and the HIPPO neutron diffractometer {about 2.4 × 107 n [s cm2 (120 µA)]−1; Wenk et al., 2003[Wenk, H. R., Lutterotti, L. & Vogel, S. C. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]}.

[Figure 2]
Figure 2
The neutron energy spectra of the RANS pulsed neutron beam at 1.4 and 5.0 m away from moderator surface, simulated by the PHITS code (Sato et al., 2013[Sato, T., Niita, K., Matsuda, N., Hashimoto, S., Iwamoto, Y., Noda, S., Ogawa, T., Iwase, H., Nakashima, H., Fukahori, T., Okumura, K., Kai, T., Chiba, S., Furuta, T. & Sihver, L. (2013). J. Nucl. Sci. Technol. 50, 913-923.]), expressed in neutrons per second per cm2 per lethargy at the RANS full-powder averaged current of 100 µA. The moderator thickness is 4.0 cm, and the neutron lethargy (i.e. logarithmic energy decrement) u = ln(E0/E) is a dimensionless logarithm of the ratio of the energy of source neutrons to the energy of product neutrons after a collision.

2.3. The RANS neutron diffractometer and its 2D neutron detector array

Fig. 3[link](a) shows the 2D detector array, consisting of eight Ø12.7 mm × 600 mm position-sensitive detector tubes filled with 3He gas at 10 standard atmospheres (1 atm = 101.325 kPa), having a spatial resolution of 10 mm along the tube axial direction and a neutron time-of-flight resolution of several microseconds. The bottom boundary of the detector array is aligned with an initial azimuthal angle η = 0°, where the maximum spanning of azimuthal angle (Δη) is about 17.8°. This detector array is vertically set up along the horizontal direction from Z = −300 mm to Z = 300 mm, obtaining a scattering angle spanning Δ2θ = 87.2° (or 2θ = 46.4–133.6°). Considering the balance between the diffraction intensity and the time-of-flight resolution of the expected neutron diffraction patterns (Johnson & Daymond, 2002[Johnson, M. W. & Daymond, M. R. (2002). J. Appl. Cryst. 35, 49-57.]), the detector array panel was divided into 16 panel regions [2 (vertical) × 8 (horizontal)], and their azimuthal angles for the geometrical centers of the bottom and top regions were 13.4 and 4.5° (or, the azimuthal angle span is Δη = 17.8°), correspondingly.

[Figure 3]
Figure 3
(a) Illustration for 16 divided panel regions of the 2D detector array containing eight position-sensitive detector tubes. (b) Orientation coverage achieved with the nominal central locations of divided detector panel regions through all 48 sample rotations in the equal-area pole figure projection; the diffraction patterns are summarized into eight bank groups, and the orientation coverages for two sample rotations (χ, ϕ) = (90°, 0°) and (90°, 45°) are marked in light blue.

For the neutron diffraction pattern from each divided region, the following equations were employed to prepare the necessary instrumental parameters according to the de Broglie wavelength equation and the Bragg diffraction law (Jorgensen et al., 1989[Jorgensen, J. D., Faber, J., Carpenter, J. M., Crawford, R. K., Haumann, J. R., Hitterman, R. L., Kleb, R., Ostrowski, G. E., Rotella, F. J. & Worlton, T. G. (1989). J. Appl. Cryst. 22, 321-333.]):

[{\lambda }_{i} = {{h}\over{m {v}_{i}}} = {{h {t}_{i}}\over{m {l}_{i}}} = {{h {t}_{i}}\over{\left\{m \big[{L}_{1}+({{{x}_{i}^{2}+y}_{i}^{2}+{z}_{i}^{2}} )^{1/2}\big]\right\}}} = 2{d}_{i}\sin{\theta }_{i}, \eqno(2)]

[{Q}_{i} = {{2\pi }\over{{d}_{i}}} = {{4\pi m}\over{h}} {{ \sin{\theta }_{i}}\over{{t}_{i}}} \left[{L}_{1}+({{x}_{i}^{2}+{y}_{i}^{2}+{z}_{i}^{2}} )^{1/2}\right], \eqno(3)]

where v, t and λ are the flight speed, the flight time and the wavelength of the incident neutrons, respectively, h is the Planck constant, m is the neutron mass, l is the total flight path of the neutrons, Q is the momentum transfer, d is the lattice plane spacing, usually abbreviated as the d spacing, and (xi, yi, zi) and 2θi are the coordinates and the scattering angle of the geometrical center of the panel region No. i. Here i = 1, 2,…, 16 and xi = L2 = 315 mm. The resolution of the time-of-flight neutron diffraction can be calculated as follows (Windsor, 1981[Windsor, C. G. (1981). Pulsed Neutron Scattering pp. 34-35, 88, 232, 268 and 288. London: Taylors & Francis Ltd.]; Jorgensen et al., 1989[Jorgensen, J. D., Faber, J., Carpenter, J. M., Crawford, R. K., Haumann, J. R., Hitterman, R. L., Kleb, R., Ostrowski, G. E., Rotella, F. J. & Worlton, T. G. (1989). J. Appl. Cryst. 22, 321-333.]):

[{{\Delta d}\over{d}} = -{{\Delta Q}\over{Q}} = {\left[{\left({{\Delta t}\over{t}}\right)}^{2}+{\left({{\Delta L}\over{L}}\right)}^{2}+{\left(\cot\theta \Delta \theta \right)}^{2}\right]}^{{{1}/{2}}}. \eqno(4)]

Accordingly, a small wavelength interval (Δλ = 0.005 Å) corresponding to a binning time width (Δt ≃ 10.63 µs) was employed here for each divided panel region to avoid any additional increase of the instrumental error.

A pure body-centered cubic (b.c.c.) Fe powder sample filled in a Ø8 mm × 60 mm vanadium can (S), a similar empty vanadium can (C), a Ø15 mm × 40 mm vanadium–nickel alloy sample (V) and the no-sample background (B) were each measured for 60 min, and the following ratio involving the sample cross section (Windsor, 1981[Windsor, C. G. (1981). Pulsed Neutron Scattering pp. 34-35, 88, 232, 268 and 288. London: Taylors & Francis Ltd.]) was employed to correct the intensity distribution I(λ) or I(t) of the neutron diffraction pattern from the divided panel region No. i (i = 1, 2,…, 16):

[{\left({{{\rm d}\sigma }\over{{\rm d}\Omega }}\right)}_{\rm S} = {\left({{{\rm d}\sigma }\over{{\rm d}\Omega }}\right)}_{\rm V}\, \left({\rm {S-C}\over{\rm V-B}}\right). \eqno(5)]

2.4. Texture measurement and Rietveld texture analysis

70% cold-rolled and annealed IF steel with composition 0.0018 C–0.01 Si–0.17 Mn–0.013 P–0.006 S–0.01 Cu–0.01 Ni–0.02 Cr–0.003 V–0.03 Ti–0.026 Nb–0.033 Alsolute–0.0014 Ntotal (mass%) (Xu et al., 2008[Xu, P. G., Tomota, Y., Suzuki, H., Suzuki, T., Machiya, S. & Yin, F. X. (2008). Mater. Trans. 49, 2033-2039.]) was employed to prepare a rounded-edged 15 × 15 × 15 mm reference sample and examine the reliability of texture measurement at RANS. The sample was chosen to be larger than the usual 10 × 10 × 10 mm reference sample for large-scale neutron facilities, because the long measurement time due to the weak beam flux at RANS may be shortened through using a larger gauge volume.

After section fine-cutting of the IF steel sheet and an electrochemical polishing treatment of the new surfaces, the grain orientation characteristics were observed using electron backscatter diffraction (EBSD) with a Hitachi S-4300SE field-emission scanning electron microscope. Fig. 4[link] shows the 3D grain orientation distribution characteristics [previously represented in Figs. 5 and 6 of Xu et al. (2008[Xu, P. G., Tomota, Y., Suzuki, H., Suzuki, T., Machiya, S. & Yin, F. X. (2008). Mater. Trans. 49, 2033-2039.])]. Through the inverse pole figure maps referring to the normal direction (ND), it is found that the IF steel sheet has a strong {111}〈hkl〉 fiber recrystallization texture together with a weak {001}〈hkl〉 texture component, and the ferrite grains in the surface layer are a little finer than those in the center layer.

[Figure 4]
Figure 4
Crystallographic orientation gradient distribution from surface to center of the interstitial-free steel. The inverse pole figure maps for different sample sections are presented here in a three-dimensional way through referring to the normal direction of the steel sheet. ND: normal direction; RD: rolling direction; TD: transverse direction.

During the neutron diffraction measurements of the IF steel sample, the incident beam slit size was set to 30 × 30 mm to ensure that the sample was completely bathed in the neutron beam. The IF steel sample was measured with 48 (χ, ϕ) rotations by using the Eulerian cradle, where χ and ϕ are, respectively, the angles between the ND and RD (rolling direction) of the IF steel sample and the cradle-dependent nominal scattering vector Q in the horizontal plane. When the ND is set up along the vertical direction then χ = 90°; when the ND is along the horizontal direction then χ = 0° [Fig. 3[link](a)]. Fig. 3[link](b) shows that the orientation coverage for all the diffraction patterns from 48 rotations is more than 90% of the stereographic angle of a complete pole figure; the orientation coverages for two (χ, ϕ) rotations are marked with light blue (Takajo & Vogel, 2018[Takajo, S. & Vogel, S. C. (2018). J. Appl. Cryst. 51, 895-900.]). The collection time for neutron diffraction patterns of the IF steel sample at each step was 5 min, and the total time including the sample rotation was about 300 min.

In future, another two neutron detector arrays will be added to RANS for a larger azimuthal angle span (Δη ≃ 43.6°) and the proton beam current will be increased. Through the reduced sample rotations and the shortened neutron collection time for each sample orientation, the RANS neutron texture measurement may be finished within 60 min for the large-sized IF steel reference sample.

All of the divided panel and intensity-corrected neutron diffraction patterns (Nos. 1–16) of the IF steel sample obtained from 48 rotations were simultaneously refined using the Materials Analysis Using Diffraction (MAUD) software (Lutterotti et al., 1997[Lutterotti, L., Matthies, S., Wenk, H. R., Schultz, A. J. & Richardson, J. (1997). J. Appl. Phys. 81, 594-600.]), and Rietveld texture analysis was carried out using the extended Williams–Imhof–Matthies–Vinel (E-WIMV) texture algorithm implemented in MAUD (Matthies & Vinel, 1982[Matthies, S. & Vinel, G. W. (1982). Phys. Status Solidi B, 112, K111-K114.]; Lutterotti et al., 2004[Lutterotti, L., Chateigner, D., Ferrari, S. & Ricote, J. (2004). Thin Solid Films, 450, 34-41.]), as shown in the pole figure [Fig. 3[link](b)], through eight different bank groups according to their scattering angles 2θ: (a) 51.9–52.6° for Group A (No. 1 and No. 9); (b) 62.8–63.4° for Group B (No. 2 and No. 10); (c) 73.7–74.1° for Group C (No. 3 and No. 11); (d) 84.6–84.7° for Group D (No. 4 and No. 12); (e) 95.4–95.3° for Group E (No. 5 and No. 13); (f) 106.3–105.9° for Group F (No. 6 and No. 14); (g) 117.2–116.6° for Group G (No. 7 & No. 15); (h) 128.1–127.4° for Group H (No. 8 and No. 16). During the Rietveld texture analysis, the crystallographic orientations were fitted at an orientation distribution function (ODF) resolution of 5°, sample symmetry was not presumed, and the pre-setup orientation angle (χ0, ϕ0) and the geometrical center (x0, y0, z0) of the sample were alignment adjusted and refined automatically.

In order to evaluate the reliability of the RANS texture measurement technique, the Rietveld texture analysis results of the same IF steel sample based on two well established time-of-flight neutron diffractometers were employed here for comparison: one was the HIPPO neutron diffractometer at the Los Alamos Neutron Source Center (LANSCE), Los Alamos National Laboratory (Wenk et al., 2003[Wenk, H. R., Lutterotti, L. & Vogel, S. C. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]), and the other was the TAKUMI engineering diffractometer at J-PARC (Xu et al., 2018[Xu, P. G., Harjo, S., Ojima, M., Suzuki, H., Ito, T., Gong, W., Vogel, S. C., Inoue, J., Tomota, Y., Aizawa, K. & Akita, K. (2018). J. Appl. Cryst. 51, 746-760.]). During the HIPPO neutron experiment in December 2009, the sample was measured through four rotations, ω = 0, 45, 67.5 and 90°, and the neutron diffraction patterns from the scattering angles 2θ = 144, 90 and 39° were collected for 10 min per rotation. During the TAKUMI neutron experiment in February 2013, the same sample was measured through 120 rotations to obtain the complete pole figures in high stereographic resolution. In order to compare the texture quantitatively, the corresponding ODFs were calculated using the spherical-harmonic function series expansion method (Bunge, 1982[Bunge, H. J. (1982). Texture Analysis in Materials Science. London: Butterworths.]).

It should be mentioned that HIPPO has since been upgraded through adding the new 2θ = 120 and 60° detector bank groups to acquire in total 53 diffraction patterns by each sample rotation (Reiche et al., 2012[Reiche, H. M., Vogel, S. C., Mosbrucker, P., Larson, E. J. & Daymond, M. R. (2012). Rev. Sci. Instrum. 83, 053901.]), so that three rotations may realize about 90% pole figure coverage of sample orientations (Takajo & Vogel, 2018[Takajo, S. & Vogel, S. C. (2018). J. Appl. Cryst. 51, 895-900.]). Recently, comparable measurements at TAKUMI using other texture samples confirmed that 19 rotations using a pseudo-equal-area scanning routine (Gnäupel-Herold & Creuziger, 2011[Gnäupel-Herold, T. & Creuziger, A. (2011). Mater. Sci. Eng. A, 528, 3594-3600.]) are practicable to obtain a good stereographic coverage of the 1/4 pole figure for reliable texture analysis of multiphase steels and other high-crystal-symmetry materials.

3. Results and discussion

3.1. Stereographic region division and instrumental scattering characteristics

Fig. 5[link] gives the incoherent vanadium and background scattering characteristics collected from all the divided panel regions (Nos. 1–16). The vanadium scattering patterns No. 1 and No. 9 of Bank Group A at 2θ = 51.9–52.6° show a typical intensity distribution combining thermal neutrons at a lattice plane spacing range of d = 1.0–2.0 Å with epithermal neutrons at d < 0.5 Å (Windsor, 1981[Windsor, C. G. (1981). Pulsed Neutron Scattering pp. 34-35, 88, 232, 268 and 288. London: Taylors & Francis Ltd.]). The scattering patterns No. 8 and No. 16 of Bank Group H at 2θ = 128.1–127.4° show a strong neutron intensity distribution around d = 0.6 Å, and the local vanadium scattering intensity at d > 1.5 Å is almost comparable to the background intensity. The other diffraction patterns of the Bank Groups B–H show a gradient transition between these extreme intensity distributions. This clear difference in scattering characteristics between various bank groups reveals that an appropriate region division of the detector panel(s) is essential to utilize the neutron diffraction patterns effectively for a compact neutron source.

[Figure 5]
Figure 5
Incoherent vanadium and background scattering characteristics of all 16 divided panel regions, measured with a Ø15 mm × 40 mm vanadium–nickel alloy and without any sample, respectively.

Fig. 6[link] shows that the neutron diffraction patterns of the b.c.c.-Fe powder sample (in discrete points) are appropriately corrected and Rietveld refined. The 110 diffraction peak of the measured diffraction patterns No. 1 and No. 9 has a relatively larger FWHM in comparison with that of the measured diffraction patterns No. 8 and No. 16, revealing that the high scattering angle leads to a better instrumental resolution Δd/d = −ΔQ/Q. Meanwhile, it is found that for the diffraction patterns No. 1 and No. 9 the diffraction intensity at a lattice plane spacing range of d < 0.8 Å is at a lower statistical precision because of the corresponding local strong background intensity; for the diffraction patterns No. 8 and No. 16, the diffraction intensity at d > 1.5 Å is at a lower statistical precision owing to the weak incident long-wavelength neutrons. Moreover, through the good Rietveld refinement of the neutron diffraction patterns, shown here as solid lines, the necessary parameters for the time-of-flight/spacing (t/d) conversion, the peak-shape refinement etc. were extracted to prepare the instrumental file for the MAUD texture analysis together with the geometrical parameters of all the panel regions including the scattering angles (2θ), the azimuthal angles (η) and the flight paths (L2).

[Figure 6]
Figure 6
Neutron diffraction patterns of the pure b.c.c.-Fe powder sample filled in a Ø10 mm × 60 mm vanadium can. The measured patterns after various corrections are shown as discrete points, and their fitted curves obtained using the MAUD software are shown as solid lines. (a) Diffraction patterns No. 1 and No. 9 of Group A at 2θ = 51.9–52.6°. (b) Diffraction patterns No. 8 and No. 16 of Group H at 2θ = 128.1–127.4°.

3.2. Texture analysis of cold-rolled and annealed IF steel

Fig. 7[link] shows typical neutron diffraction patterns of cold-rolled and annealed IF steel and their Rietveld-fitted results. In general, the Rietveld texture refinement is highly satisfactory considering that the peak intensities in each diffraction pattern are not strong. Even for a same-sample rotation, the diffraction patterns from the neighboring panel regions Nos. 3 and 11 and Nos. 5 and 13 show a clear difference in the diffraction intensities of the 211 and 220 peaks (marked by red frames), suggesting that the panel region division is very valuable for a reliable texture measurement. However, for Group A at a lattice plane spacing range of d < 0.8 Å, the background noise is stronger and the deviation of the intensity of the measured diffraction patterns from the Rietveld-refined curves is relatively larger; for Group E at a lattice plane spacing range of d > 1.4 Å, the weak incident beam flux results in a relatively large deviation of the intensity of the measured diffraction patterns from the Rietveld-refined curves. Therefore, 11 peak reflections were employed from the diffraction patterns of Bank Groups A–E, and only nine peak reflections were employed from the diffraction patterns of Bank Groups F–H, hereafter referred to as `unequal d ranges' (i.e. for Groups A–E: d = 0.6–2.4 Å; for Groups F–H: d = 0.6–1.4 Å). The slightly different case of using 11 peak reflections from Bank Groups A–H is referred to as `equal d ranges', or d = 0.6–2.4 Å.

[Figure 7]
Figure 7
Partial neutron diffraction patterns of the IF steel sample and their partially Rietveld refined patterns: (a) Bank Group A (No. 1 and No. 9, at 2θ = 51.9–52.6°); (b) Bank Group C (No. 3 and No. 11, at 2θ = 73.7–74.1°); (c) Bank Group E (No. 5 and No. 13, at 2θ = 95.4–95.3°); (d) Bank Group H (No. 8 and No. 16, at 2θ = 128.1–127.4°).

Fig. 8[link] shows the recalculated pole figures in equal-area projection obtained by using the diffraction patterns measured from RANS, TAKUMI and HIPPO, together with the φ2 = 45° ODF sections calculated by using the series expansion texture calculation method at an expansion series of Lmax = 32. Moreover, the corresponding values of the texture index F2, i.e. the integral of the square of the texture function f(g) (Bunge, 1982[Bunge, H. J. (1982). Texture Analysis in Materials Science. London: Butterworths.]), are given here for reference. The preferred orientations in the pole figures in Fig. 8[link](a) from RANS using `unequal d ranges' are similar to those in Fig. 8[link](c) from TAKUMI and Fig. 8[link](d) from HIPPO using `equal d ranges', including good orientation symmetry, no ghost orientations and almost consistent pole density distribution for typical texture orientations, suggesting the RANS technical environment for texture measurement has been established reliably. In contrast, Fig. 8[link](b) shows an unexpected stronger density distribution in the 110 pole figure, mostly due to the uncertainty in the intensities of the diffraction patterns of Banks F–H at d > 1.4 Å. Moreover, from the φ2 = 45° ODF sections in Figs. 8[link](a), 8[link](c) and 8[link](d), the strong {111}〈uvw〉 γ fiber at around φ1 = 0–90°, ϕ = 55°, φ2 = 45° together with a weak {001}〈110〉 rotated-cube component at around φ1 = 0°, ϕ = 0 and 90°, φ2 = 45° confirms the crystallographic orientation characteristics observed from the EBSD mapping results (Fig. 4[link]) at a higher statistical level. Fig. 8[link](b) shows overestimated γ-fiber and underestimated rotated-cube components, which is almost unacceptable as a high-statistical-precision measurement result.

[Figure 8]
Figure 8
Recalculated pole figures (in equal-area projection, left) and orientation distribution functions (φ2 = 45° section, right) of the IF reference steel obtained from the measured neutron diffraction patterns through the Rietveld texture analysis: (a) from RANS using unequal d ranges (for Groups A–E: d = 0.6–2.4 Å; for Groups F–H: d = 0.6–1.4 Å), Imax = 4.85 m.r.d. (multiples of a random distribution) in pole figures, fmax = 10.89 m.r.d. in ODF, texture index F2 = 6.73; (b) from RANS using equal d ranges (d = 0.6–2.4 Å), Imax = 5.23 m.r.d. in pole figures, fmax = 13.91 m.r.d. in ODF, texture index F2 = 7.51; (c) from TAKUMI using equal d ranges (d = 0.6–2.4 Å), Imax = 5.19 m.r.d. in pole figures, fmax = 11.76 m.r.d. in ODF, texture index F2 = 7.02: (d) from HIPPO using equal d ranges (d = 0.6–2.4 Å), Imax = 5.24 m.r.d. in pole figures, fmax = 9.30 m.r.d. in ODF, texture index F2 = 6.61.

Since HIPPO is a well established neutron diffractometer for texture measurement (Wenk et al., 2003[Wenk, H. R., Lutterotti, L. & Vogel, S. C. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575-588.]), which has been widely employed in many academic studies, and TAKUMI has been established for high-stereographic-resolution texture measurement (Xu et al., 2018[Xu, P. G., Harjo, S., Ojima, M., Suzuki, H., Ito, T., Gong, W., Vogel, S. C., Inoue, J., Tomota, Y., Aizawa, K. & Akita, K. (2018). J. Appl. Cryst. 51, 746-760.]), the above results reveal that, at the least, RANS and the related texture measurement environment are satisfactory for the quantitative texture measurement of steels and other high-symmetry materials. For RANS texture analysis in equal d ranges, the reliability of the texture measurement may be improved through longer neutron collection times, and the texture measurements of some complicated materials including titanium alloys and magnesium alloys may be realized. Moreover, because the panel region division at RANS enables us to acquire many concurrent neutron diffraction patterns with good statistics, Rietveld texture analysis in unequal d ranges is extremely valuable for a compact neutron source such as RANS.

On the other side, since the same IF steel was measured at HIPPO for 10 min per rotation and at TAKUMI for 0.5 min per rotation, the reliability of the RANS texture measurement from the weak diffraction patterns reveals that the neutron collection time per rotation at HIPPO, TAKUMI and other large-scale neutron facilities could be further shortened to provide more possibility for time-resolved neutron diffraction studies (Onuki et al., 2016[Onuki, Y., Hoshikawa, A., Sato, S., Xu, P. G., Ishigaki, T., Saito, Y., Todoroki, H. & Hayashi, M. (2016). J. Appl. Cryst. 49, 1579-1584.]).

4. Conclusions and prospect

In order to improve the instrumental accessibility of the neutron diffraction technique and provide a rapid response to the on-site research needs for advanced metallic materials, the quantitative texture measurement technique was established at RANS. The fine division of the diffraction detector panel ensured the high stereographic resolution of time-of-light neutron diffractograms, and the concurrent Rietveld full-profile refinement of all obtained neutron diffractograms decreased the necessary sample rotations to a certain extent.

All of the 768 neutron diffraction patterns of a cold-rolled and annealed IF steel sample collected through 48 rotations were Rietveld texture analyzed through selectively using parts of the obtained neutron diffraction patterns that exhibited good statistics, and the obtained pole figures were almost consistent with the corresponding results from two well equipped neutron diffraction texture measurement instruments, TAKUMI at J-PARC, Japan, and HIPPO at LANSCE, USA, revealing that the precise neutron diffraction texture measurement at RANS has been realized successfully.

Such technical progress using a compact neutron source to realize precise neutron diffraction texture measurement suggests that we are entering a new era of neutron scattering technology and its engineering applications, and that RANS-like compact neutron sources will be widely used together with in-house neutron scattering facilities for higher instrument accessibility.

Acknowledgements

The authors thank Dr Sven C. Vogel at Los Alamos National Laboratory, USA, for his HIPPO experimental support (proposal No. 20092165) at the Los Alamos Neutron Science Center (LANSCE). The authors also thank Dr Stefanus Harjo, Dr Takayoshi Ito and Dr Jun Abe at J-PARC Center, Japan Atomic Energy Agency, Japan, for their TAKUMI experimental support (proposal No. 2012I0019) at the Materials and Life Science Experimental Facility (MLF). The authors appreciate the highly valuable advice of Professor Luca Lutterotti (University of Trento, Italy) on the use of the Rietveld texture analysis software MAUD. The cubic sample of IF steel was supplied by Professor Yo Tomota at Ibaraki University (now at National Institute of Advanced Industrial Science and Technology), Japan.

Funding information

This work has been supported by the Japan Society for the Promotion of Science through the Grant-in-Aid for Scientific Research B: JSPS KAKENHI grant No. 17H03161.

References

First citationAnderson, I. S., Andreani, C., Carpenter, J. M., Festa, G., Gorini, G., Loong, C. K. & Senesi, R. (2016). Phys. Rep. 654, 1–58.  Google Scholar
First citationArgyriou, D. N. & Allen, A. J. (2018). J. Appl. Cryst. 51, 567–569.  Google Scholar
First citationBrokmeier, H. G. (1999). Textures Microst. 33, 13–33.  Google Scholar
First citationBrokmeier, H. G., Gan, W., Randau, C., Völler, M., Rebelo-Kornmeier, J. & Hofmann, M. (2011). Nucl. Instrum. Methods Phys. Res. A, 642, 87–92.  Google Scholar
First citationBunge, H. J. (1982). Texture Analysis in Materials Science. London: Butterworths.  Google Scholar
First citationChen, S. C., Tomota, Y., Shiota, Y., Toomine, Y. & Kamiyama, T. (2006). Tetsu Hagane, 92, 557–561.  Google Scholar
First citationChoi, S. H., Kim, E. Y., Woo, W., Han, S. H. & Kwak, J. H. (2013). Int. J. Plast. 45, 85–102.  Google Scholar
First citationDelannay, L., Jacques, P. J. & Kalidindi, S. R. (2006). Int. J. Plast. 22, 1879–1898.  Google Scholar
First citationGnäupel-Herold, T. & Creuziger, A. (2011). Mater. Sci. Eng. A, 528, 3594–3600.  Google Scholar
First citationHama, T., Matsudai, R., Kuchinomachi, Y., Fujimoto, H. & Takuda, H. (2015). ISIJ Int. 55, 1067–1075.  Google Scholar
First citationHase, K., Handa, T., Eto, T., Yamamura, N. & Aoki, M. (2016). Mater. Jpn, 56, 82–84.  Google Scholar
First citationHawkesworth, M. R. (1977). At. Energy Rev. 15, 169–220.  Google Scholar
First citationHayashi, M., Root, J. H., Rogge, R. B. & Xu, P. G. (2018). QuBS, 2, 21.  Google Scholar
First citationHinterstein, M., Hoelzel, M., Rouquette, J., Haines, J., Glaum, J., Kungl, H. & Hoffman, M. (2015). Acta Mater. 94, 319–327.  Web of Science CrossRef CAS Google Scholar
First citationHirota, K. (2018). Hamon, 28, 29–32.  Google Scholar
First citationHoshikawa, A., Ishigaki, T., Nagai, M., Kobayashi, K., Sagehashi, H., Kamiyama, T., Yonemura, M., Aizawa, K., Sakuma, T., Tomota, Y., Arai, M., Hayashi, M., Ebata, K., Takano, Y. & Kasao, T. (2009). Nucl. Instrum. Methods Phys. Res. A, 600, 203–206.  Google Scholar
First citationIkeda, Y., Otake, Y. & Mizuta, M. (2017). J. Adv. Concr. Technol. 15, 603–609.  Google Scholar
First citationIkeda, Y., Takamura, M., Hakoyama, T., Otake, Y., Kumagai, M. & Suzuki, H. (2018). Tetsu Hagane, 104, 138–144.  Google Scholar
First citationIkeda, Y., Taketani, A., Takamura, M., Sunaga, H., Kumagai, M., Oba, Y., Otake, Y. & Suzuki, H. (2016). Nucl. Instrum. Methods Phys. Res. A, 833, 61–67.  Google Scholar
First citationJansen, E., Schäfer, W. & Kirfel, A. (2000). J. Struct. Geol. 22, 1559–1564.  Web of Science CrossRef Google Scholar
First citationJansen, E., Schäfer, W. & Kirfel, A. (2004). Physica B, 350, E569–E572.  Google Scholar
First citationJohnson, M. W. & Daymond, M. R. (2002). J. Appl. Cryst. 35, 49–57.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationJorgensen, J. D., Faber, J., Carpenter, J. M., Crawford, R. K., Haumann, J. R., Hitterman, R. L., Kleb, R., Ostrowski, G. E., Rotella, F. J. & Worlton, T. G. (1989). J. Appl. Cryst. 22, 321–333.  Google Scholar
First citationLiss, K. D. (2017a). Metals, 7, 266.  Google Scholar
First citationLiss, K. D. (2017b). QuBS, 1, 1.  Google Scholar
First citationLiss, K. D., Funakoshi, K., Dippenaar, R. J., Higo, Y., Shiro, A., Reid, M., Suzuki, H., Shobu, T. & Akita, K. (2016). Metals, 6, 165.  Google Scholar
First citationLutterotti, L., Chateigner, D., Ferrari, S. & Ricote, J. (2004). Thin Solid Films, 450, 34–41.  Web of Science CrossRef CAS Google Scholar
First citationLutterotti, L., Matthies, S., Wenk, H. R., Schultz, A. J. & Richardson, J. (1997). J. Appl. Phys. 81, 594–600.  CrossRef CAS Google Scholar
First citationMa, B. L., Otake, Y., Wang, S., Sunaga, H., Yamagata, Y., Taketani, A., Hu, H. S., Jia, Q. G., Hu, G. & Bautista, U. (2018). Appl. Radiat. Isot. 137, 129–138.   Google Scholar
First citationMalamud, F., Santisteban, J. R., Vicente Alvarez, M. A., Bolmaro, R., Kelleher, J., Kabra, S. & Kockelmann, W. (2014). J. Appl. Cryst. 47, 1337–1354.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMatthies, S. & Vinel, G. W. (1982). Phys. Status Solidi B, 112, K111–K114.  CrossRef Web of Science Google Scholar
First citationMo, F. J., Sun, G. A., Li, J., Zhang, C. S., Wang, H., Chen, Y., Liu, Z., Yang, Z. K., Li, H. J., Yang, Z. L., Pang, B. B., Huang, Y. L., Tian, Y., Gong, J., Chen, B. & Peng, S. M. (2018). QuBS, 2, 15.  Google Scholar
First citationNishimura, K., Handa, T. & Hashimoto, M. (2007). JFE GIHO, 18, 18–22.  Google Scholar
First citationNishimura, K. & Takeuchi, Y. (2014). Tetsu Hagane, 100, 1097–1103.  Google Scholar
First citationOnink, M., Brakman, C. M., Tichelaar, F. D., Mittemeijer, E. J., van der Zwaag, S., Root, J. H. & Konyer, N. B. (1993). Scr. Metall. Mater. 29, 1011–1016.  Google Scholar
First citationOnuki, Y., Hoshikawa, A., Sato, S., Xu, P. G., Ishigaki, T., Saito, Y., Todoroki, H. & Hayashi, M. (2016). J. Appl. Cryst. 49, 1579–1584.  CrossRef IUCr Journals Google Scholar
First citationOtake, Y. (2018a). Applications of Laser-Driven Particle Acceleration, edited by P. Bolton, K. Parodi & J. Schreiber, ch. 19, pp. 291–314. Boca Raton: CRC Press.  Google Scholar
First citationOtake, Y. (2018b). Encyclopedia of Analytical Chemistry, edited by R. A. Meyers, Neutron Diffraction and Neutron Imaging with Compact Neutron Source. Chichester: John Wiley & Sons.  Google Scholar
First citationOtake, Y., Seki, Y., Wakabayashi, Y., Ikeda, Y., Hashiguchi, T., Yoshimura, Y., Sunaga, H., Taketani, A., Mizuta, M., Oshima, Y. & Ishida, M. (2017). J. Disaster Res. 12, 585–592.  Google Scholar
First citationReiche, H. M. & Vogel, S. C. (2010). Rev. Sci. Instrum. 81, 093302.  Web of Science CrossRef PubMed Google Scholar
First citationReiche, H. M., Vogel, S. C., Mosbrucker, P., Larson, E. J. & Daymond, M. R. (2012). Rev. Sci. Instrum. 83, 053901.  Google Scholar
First citationRücker, U., Cronert, T., Voigt, J., Dabruck, J. P., Doege, P. E., Ulrich, J., Nabbi, R., Beßler, Y., Butzek, M., Büscher, M., Lange, C., Klaus, M., Gutberlet, T. & Brückel, T. (2016). Eur. Phys. J. Plus, 131, 19.  Google Scholar
First citationSato, K., Nakamichi, N. & Sueyoshi, H. (2017). Phys. World, October 2017, https://live.iop-pp01.agh.sleek.net/physicsworld/reader/#!edition/editions_neutron_2017/article/page-22576Google Scholar
First citationSato, T., Niita, K., Matsuda, N., Hashimoto, S., Iwamoto, Y., Noda, S., Ogawa, T., Iwase, H., Nakashima, H., Fukahori, T., Okumura, K., Kai, T., Chiba, S., Furuta, T. & Sihver, L. (2013). J. Nucl. Sci. Technol. 50, 913–923.  Google Scholar
First citationSugimoto, K., Hayakawa, A., Hojo, T., Hashimoto, S. & Ikeda, S. (2003). Tetsu Hagane, 89, 1233–1239.  Google Scholar
First citationTakahashi, O., Shibui, Y., Xu, P. G., Harjo, S., Suzuki, T. & Tomota, Y. (2020). QuBS, 4, 16.  Google Scholar
First citationTakajo, S., Hollis, K. J., Cummins, D. R., Tegtmeier, E. L., Dombrowski, D. E. & Vogel, S. C. (2018). QuBS, 2, 12.  Google Scholar
First citationTakajo, S. & Vogel, S. C. (2018). J. Appl. Cryst. 51, 895–900.  Google Scholar
First citationTakamura, M., Nishimura, S. & Sunaga, H. (2013). Key Eng. Mater. 554–557, 1331–1337.  Google Scholar
First citationTaketani, A., Yamada, M., Ikeda, Y., Hashiguchi, T., Sunaga, H., Wakabayashi, Y., Ashigai, S., Takamura, M., Mihara, S., Yanagimachi, S., Otake, Y., Wakabayashi, T., Kono, K. & Nakayama, T. (2017). ISIJ Int. 57, 155–161.  Google Scholar
First citationTomota, Y., Sekido, N., Xu, P. G., Kawasaki, T., Harjo, S., Tanaka, M., Shinohara, T., Su, Y. H. & Taniyama, A. (2017). Tetsu Hagane, 103, 570–578.  Google Scholar
First citationVogel, S. C. (2013). ISRN Mater. Sci. 302408.  Google Scholar
First citationWenk, H.-R. (1991). J. Appl. Cryst. 24, 920–927.  CrossRef Web of Science IUCr Journals Google Scholar
First citationWenk, H. R., Larson, A. C., Vergamini, P. J. & Schultz, A. J. (1991). J. Appl. Phys. 70, 2035–2040.  Google Scholar
First citationWenk, H. R., Lutterotti, L. & Vogel, S. C. (2003). Nucl. Instrum. Methods Phys. Res. A, 515, 575–588.  CrossRef CAS Google Scholar
First citationWindsor, C. G. (1981). Pulsed Neutron Scattering pp. 34–35, 88, 232, 268 and 288. London: Taylors & Francis Ltd.  Google Scholar
First citationWroblewski, T., Jansen, E., Schäfer, W. & Skowronek, R. (1999). Nucl. Instrum. Methods Phys. Res. A, 423, 428–434.  Google Scholar
First citationXu, P. G., Harjo, S., Ojima, M., Suzuki, H., Ito, T., Gong, W., Vogel, S. C., Inoue, J., Tomota, Y., Aizawa, K. & Akita, K. (2018). J. Appl. Cryst. 51, 746–760.  Google Scholar
First citationXu, P. G., Ishijima, Y., Qiu, H., Morooka, S., Gong, W. & Harjo, S. (2019). J-PARC Symposium 2019, 23–26 September 2019. Poster.  Google Scholar
First citationXu, P. G., Tomota, Y., Arakaki, Y., Harjo, S. & Sueyoshi, H. (2017). Mater. Charact. 127, 104–110.  CrossRef CAS Google Scholar
First citationXu, P. G., Tomota, Y., Suzuki, H., Suzuki, T., Machiya, S. & Yin, F. X. (2008). Mater. Trans. 49, 2033–2039.  CrossRef CAS Google Scholar
First citationXu, P. G., Yin, F. X. & Nagai, K. (2006). Mater. Sci. Eng. A, 441, 157–166.  Google Scholar
First citationXu, P. G., Yin, J. & Zhang, S. Y. (2015). Acta Metall. Sin. 51, 1297–1305.  CAS Google Scholar
First citationYamagata, Y., Hirota, K., Ju, J., Wang, S., Morita, S., Kato, J., Otake, Y., Yaketani, A., Seki, Y., Yamada, M., Ota, H., Bautista, U. & Jia, Q. G. (2015). J. Radioanal. Nucl. Chem. 305, 785–794.  Google Scholar
First citationYusuf, S. M. & Kumar, A. (2017). Appl. Phys. Rev. 4, 031303.  Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds