In situ monitoring of hydrothermal reactions by X-ray diffraction with Bragg–Brentano geometry

CROSSMARK_Color_square_no_text.svg

aHochschule Wismar, Philipp-Müller-Strasse 14, 23966 Wismar, Germany, and bUniversity of Greifswald, Friedrich-Ludwig-Jahn-Strasse 17A, 17489 Greifswald, Germany
*Correspondence e-mail: karsten.mesecke@stud.uni-greifswald.de

Edited by J. M. García-Ruiz, Instituto Andaluz de Ciencias de la Tierra, Granada, Spain (Received 24 February 2020; accepted 1 May 2020; online 18 June 2020)

This note describes an autoclave chamber developed and constructed by Anton Paar and its application for in situ experiments under hydrothermal conditions. Reactions of crystalline phases can be studied by successive in situ measurements on a conventional laboratory X-ray diffractometer with Bragg–Brentano geometry at temperatures <483 K and saturated vapour pressure <2 MPa. Variations in the intensity of X-ray diffraction reflections of both reactants and products provide quantitative information for studying the reaction kinetics of both dissolution and crystal growth. Feasibility is demonstrated by studying a cementitious mixture used for autoclaved aerated concrete production. During a period of 5.7 h at 466 K and 1.35 MPa, the crystallization of torbermorite and the partial consumption of quartz were monitored.

1. Introduction

A wide range of commercial and non-commercial reaction cells have been developed for studying chemical reactions in situ by X-ray or neutron diffraction (Van Beek & Pattison, 2019[Van Beek, W. & Pattison, P. (2019). International Tables for Crystallography, Vol. H, Powder Diffraction, pp. 189-197. Chester: International Union of Crystallography.]), but so far hydrothermal reactions have been studied using such techniques exclusively in transmission geometry (Norby, 2006[Norby, P. (2006). Curr. Opin. Colloid Interface Sci. 11, 118-125.]). Such transmission cells can be obtained by modifications of laboratory autoclaves (Van Beek & Pattison, 2019[Van Beek, W. & Pattison, P. (2019). International Tables for Crystallography, Vol. H, Powder Diffraction, pp. 189-197. Chester: International Union of Crystallography.]) and are suited to studying the formation of tobermorite and other calcium silicate hydrates at saturated vapour temperatures of 463 K (Bernstein, 2011[Bernstein, S. (2011). PhD thesis, Ludwig Maximilian University of Munich, Germany.]), 463–583 K (Shaw et al., 2000[Shaw, S., Clark, S. & Henderson, C. (2000). Chem. Geol. 167, 129-140.]) and 463 K (Kikuma et al., 2010[Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A., Matsui, K. & Sato, M. (2010). J. Am. Ceram. Soc. 93, 2667-2674.]; Matsui et al., 2011[Matsui, K., Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A. & Sato, M. (2011). Cem. Concr. Res. 41, 510-519.]). However, beamline time is limited and the lengths of these experiments vary between 5 h (Shaw et al., 2000[Shaw, S., Clark, S. & Henderson, C. (2000). Chem. Geol. 167, 129-140.]) and 13.5 h (Matsui et al., 2011[Matsui, K., Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A. & Sato, M. (2011). Cem. Concr. Res. 41, 510-519.]). Cells used to study these hydrothermal reactions with conventional laboratory X-ray diffractometers and reflection geometry have not yet been documented. This laboratory note details such a cell, which was developed by Anton Paar (Anton Paar GmbH, Graz, Austria), and the accompanying analytical setup used for in situ monitoring of tobermorite formation in association with the production of autoclaved aerated concrete (AAC).

2. Experimentation

2.1. The device

The autoclave chamber (Fig. 1[link]) is a device constructed by Anton Paar for temperatures <483 K and saturated vapour pressure <2 MPa. It is a flat-plate chamber comprising a stainless steel body, thermal insulation and a front face screw-on lid. X-rays are able to pass through beryllium windows protected by Kapton polyimide foil; the associated intensity loss is ca 45% under ambient conditions. The formation of condensate droplets on the inside is minimized by applying a surfactant solution prior to each experiment. The chamber is mounted in the middle of the goniometer and can be adjusted in height by ±3 mm to compensate for different specimen thicknesses [Fig. S1(e) in the supporting information]. Because of the design the temperature has no significant influence on the Bragg–Brentano geometry. A slight tilt can be corrected by adjusting the θ offset.

[Figure 1]
Figure 1
A schematic cross section through the autoclave chamber, a flat-plate chamber for temperatures <483 K and saturated vapour pressure <2 MPa. The front face screw-on lid is not shown in this figure.

Saturated steam is supplied from an external boiler equipped with a tray of calcium hydroxide for scrubbing carbon dioxide. As the steam enters the chamber through the stainless steel body, a homogeneous temperature is maintained, which is measured below the specimen by a PT100 sensor. Constant conditions can be achieved with a precision of ±0.5 K. Pressure is measured at the condensate outlet by an analogue gauge with an accuracy of ±0.025 MPa. Condensate can drain through a vent or reflux to the steam supply boiler supported by an oscillation pump. Separate tubing within the stainless steel body is connected to an external circulation thermostat. The respective pipes are fitted to the diffractometer (in this case a Bruker D8 Advance) without modification of the safety enclosure (here a D8 Discover enclosure).

In the Bragg–Brentano geometry an even specimen surface is crucial for obtaining valid X-ray diffraction (XRD) data. Cracking, buckling or tilting of the surface may obstruct ideal focusing conditions. Therefore cementitious mixtures are favourable to achieve these conditions. A stable surface is generally obtained after 12 h of hydration prior to measurement. The sample trays constructed for the device are 40 × 40 × 14 mm, made out of nickel, and available with specimen depths of either 12 or 2 mm. The 12 mm deep trays can be partially filled with hollow stainless steel cylinders [Fig. S1(a)] to reduce the amount of sample needed and to improve surface stability. Condensate leaching is prevented by a curved sample cover made out of 0.5 mm poly(tetrafluorethylene) foil featuring a 40 × 13 mm slot for X-rays.

2.2. Sample preparation

A test sample was prepared analogously to industrial AAC production with a water-to-solid ratio of 0.73 and 48.3% quartz (99.1% SiO2; D50 23 µm), 38.1% ordinary Portland cement, 8.7% lime (93% CaO), 5.0% calcium fluorite (>99%; sintered at 973 K for 3 h) and 0.1% aluminium powder. This slurry was poured directly into the sample trays and an even surface obtained by covering the sample trays with thread-seal tape, polyethylene (PE) foil and a planar weight. Analogously to AAC production, the added aluminium powder induces hydrogen gas release, and subsequent foaming yields a thickness of 3–5 mm. The viscosity increases rapidly and the sample is expected to set within 1 h. For further cement hydration, it was kept at 343 K for 7 h and afterwards at room temperature for a total of two days prior to in situ measurement. Moisture loss was prevented by the PE foil [Fig. S1(b)].

2.3. In situ measurement

The Bruker D8 Advance used in this study was equipped with a conventional X-ray source (Cu 40 kV, 40 mA), a position-sensitive detector (LYNXEYE, opening degree 2.56°) and an Ni Kβ filter. A fixed divergence slit of 2 mm on the primary beam yields an irradiated area of ca 13 × 11 mm. The sample and its cover were slid into the autoclave chamber [Fig. S1(d)], with two cones under the tray matching cavities in the steel body. After sealing of the front face screw-on lid, the chamber was pre-heated (<363 K) using the circulation thermo­stat. With the vent open, steam was flushed through the chamber to ensure saturated water vapour inside the system. Sealing the vent allowed the pressure to build up (0 min). Hydrothermal conditions at 466 K and 1.35 MPa were obtained by steadily increasing the boiler and thermostat temperatures (230 min). After a reaction period of 5.7 h, the steam supply boiler was switched off (570 min) and the system cooled down slowly. In situ monitoring was carried out by subsequent θ–2θ measurements from 14.2 to 50.8° 2θ, each taking 13.22 min. A total of 44 out of 51 diffractograms were used for evaluation. Reflection intensities were obtained by exporting areas (c.p.s. × °) for respective 2θ ranges (DIFFRAC.EVA; Bruker AXS, Karlsruhe, Germany) and subtracting the background. Normalization was done according to the first or last evaluated measurement. Afterwards, control measurements of the original and then homogenized sample were carried out under ambient conditions.

3. Results and discussion

Fig. 2[link] shows four diffractograms ranging from 26 to 29.3° 2θ, close to and at 466 K (>230 min). Changes in the reflection intensities of quartz 101 [SiO2, Powder Diffraction File (PDF; International Centre for Diffraction Data, Newtown Square, Pennsylvania, USA; https://www.icdd.com) card 46-1045; Kern & Eysel, 1993[Kern, A. & Eysel, W. (1993). ICDD Grant-in-Aid. Mineralogisch-Petrograph Institut, Universität Heidelberg, Germany.]] and tobermorite 110 [Ca4.5Si6O14(OH)3 2H2O, PDF card 83-1520; Hamid, 1981[Hamid, S. A. (1981). Z. Kristallogr. 154, 189-198.]] indicate that a hydrothermal reaction takes place. Fluorite 111 (CaF2, PDF card 35-0816; Swanson et al., 1985[Swanson, H. E., McMurdie, H. F., Morris, M. C. & Evans, E. H. (1985). Standard X-ray Diffraction Powder Patterns, National Bureau of Standards Monograph 25 - Section 5, pp. 21 and 52. Gaithersburg: National Bureau of Standards.]) is seen to be essentially inert and serves as an internal standard.

[Figure 2]
Figure 2
X-ray diffraction results close to and after reaching constant hydrothermal conditions at 466 K and 1.35 MPa (230 min). The tobermorite 110 reflection intensity increases as the quartz 101 reflection intensity decreases. Fluorite serves as internal standard. (See Fig. S2 in the supporting information for the full range.)

In Fig. 3[link] normalized reflection intensities are shown. Below 443 K the fluorite and quartz intensities decrease proportionally in response to increasing temperature and pressure. The steam density inside the autoclave chamber changes from 0.59 kg m−3 at 373 K to 6.86 kg m−3 at 466 K (Harvey, 1998[Harvey, A. (1998). NIST Interagency/Internal Report No. 5078, pp. 5-7. National Institute of Standards and Technology, Gaithersburg, MD, USA.]), which is accompanied by an increase in X-ray absorption. Up to 466 K (230 min) the fluorite reflection intensity decreases by ca 50%. It is fitted by an empirical exponential function (dashed line) and used to make a temperature correction. Above 443 K, the quartz intensity decreases significantly faster than the fluorite intensity, which marks the beginning of hydrothermal dissolution. The quartz dissolution can be approximated as a first-order reaction behaviour after temperature correction (k = 7.8 × 10−5 s−1, solid line). The intensities in the 2θ range of tobermorite reflections start to increase at >443 K (>130 min), with a distinct increase of the 110 reflection starting at 210 min, and then a continuous increase according to a first-order reaction behaviour (k = 6.1 × 10−5 s−1, solid line). In compliance with previous studies, tobermorite forms via poorly crystalline calcium silicate hydrates, initially limited by the dissolution of quartz (Matsui et al., 2011[Matsui, K., Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A. & Sato, M. (2011). Cem. Concr. Res. 41, 510-519.]).

[Figure 3]
Figure 3
Time-resolved normalized reflection intensities from 373 K (0 min) to constant conditions at 466 K and 1.35 MPa (>230 min). Below 443 K, the fluorite and quartz intensities decrease proportionally in response to increasing temperature and pressure, while above 443 K further decrease of the quartz intensities is related to hydrothermal dissolution. The approximation (solid line) is a combination of temperature correction (dashed line) and a first-order reaction behaviour. Distinct tobermorite crystallization starts at 210 min and a first-order reaction behaviour can be assumed (solid line).

To interpret these reaction rates, surface sensitivity has to be considered. At an incident angle of 20° the penetration depth is expected to be less than 120 µm. Despite a constant steam density at 466 K (>230 min), Fig. 3[link] shows a slight decrease in the fluorite 111 reflection intensity. This may reflect the preferred growth of tobermorite crystals on the sample surface, which could mask fluorite and quartz. Accordingly, in Fig. 4[link] the intensity of the fluorite 111 reflection after hydrothermal reaction is lower than before. Upon homogenization, fluorite 111 regains its original reflection intensity while tobermorite 110 remains lower in intensity. Homogenization is accompanied by a notable increase in the intensity of the quartz 101 reflection, which could be related to the initial gravitational settling of larger less-reactive quartz grains. Therefore, the reaction rates derived from the decrease of quartz 101 reflection intensities could be overestimated.

[Figure 4]
Figure 4
Plots of θ-corrected X-ray diffraction results before (11 min) and after hydrothermal reaction. Homogenization significantly increases the intensity of the quartz 101 reflection, indicating some surface sensitivity of the in situ measurements. (See Fig. S3 for the full range)

4. Conclusions

This reflection setup is suitable for in situ monitoring of hydrothermal reactions and can provide routine information for time-resolved phase quantification. However, in contrast with transmission setups, there are limitations in time resolution, lower signal-to-noise ratios, additional temperature influences and limitations in penetration depth. An internal standard can be used to reveal migration within the sample, correct the influence of temperature on reflection intensities or quantify amorphous content by Rietveld analysis. Due to differences in experimental setup, reaction rates might differ from previous studies of cementitious mixtures in association with the production of autoclaved aerated concrete (Kikuma et al., 2010[Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A., Matsui, K. & Sato, M. (2010). J. Am. Ceram. Soc. 93, 2667-2674.]; Matsui et al., 2011[Matsui, K., Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A. & Sato, M. (2011). Cem. Concr. Res. 41, 510-519.]). However, the surface sensitivity of this reflection setup could provide additional spatial information on the nature of hydrothermal reactions at interfaces. In situ monitoring of other systems is possible if, in accordance with Bragg–Brentano geometry, a stable and even surface can be obtained.

5. Supporting data

Measurement data are available at https://drive.google.com/open?id=1vKxa3fHu92Cc03_U3eGD43c5m2R5EMpK.

Supporting information


Acknowledgements

Josef Gautsch and his colleagues from Anton Paar are thanked for their support during installation and trial of the autoclave chamber. Arkadiusz Kopaczynski from Bruker AXS is thanked for his support during implementation of motorized drives to the diffractometer control software.

Funding information

Funding for this research was provided by Arbeitsgemeinschaft industrieller Forschungsvereinigungen `Otto von Guericke' e.V. (grant No. 19464 BR/1).

References

First citationBernstein, S. (2011). PhD thesis, Ludwig Maximilian University of Munich, Germany.  Google Scholar
First citationHamid, S. A. (1981). Z. Kristallogr. 154, 189–198.  CrossRef ICSD CAS Web of Science Google Scholar
First citationHarvey, A. (1998). NIST Interagency/Internal Report No. 5078, pp. 5–7. National Institute of Standards and Technology, Gaithersburg, MD, USA.  Google Scholar
First citationKern, A. & Eysel, W. (1993). ICDD Grant-in-Aid. Mineralogisch-Petrograph Institut, Universität Heidelberg, Germany.  Google Scholar
First citationKikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A., Matsui, K. & Sato, M. (2010). J. Am. Ceram. Soc. 93, 2667–2674.  Web of Science CrossRef CAS Google Scholar
First citationMatsui, K., Kikuma, J., Tsunashima, M., Ishikawa, T., Matsuno, S., Ogawa, A. & Sato, M. (2011). Cem. Concr. Res. 41, 510–519.  Web of Science CrossRef CAS Google Scholar
First citationNorby, P. (2006). Curr. Opin. Colloid Interface Sci. 11, 118–125.  Web of Science CrossRef CAS Google Scholar
First citationShaw, S., Clark, S. & Henderson, C. (2000). Chem. Geol. 167, 129–140.  Web of Science CrossRef CAS Google Scholar
First citationSwanson, H. E., McMurdie, H. F., Morris, M. C. & Evans, E. H. (1985). Standard X-ray Diffraction Powder Patterns, National Bureau of Standards Monograph 25 – Section 5, pp. 21 and 52. Gaithersburg: National Bureau of Standards.  Google Scholar
First citationVan Beek, W. & Pattison, P. (2019). International Tables for Crystallography, Vol. H, Powder Diffraction, pp. 189–197. Chester: International Union of Crystallography.  Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds