For X-ray absorption spectroscopy, either in transmission mode with concentrated samples or for dilute samples in fluorescence mode, it is advantageous to improve the signal-to-noise ratio by implementing a slit apparatus. Several investigations into the improvement of measurements when slits and filters are employed have been reported; however, these have always been for a particular design and are not transferable between dissimilar systems. A generalized approach to Soller slit design will be presented which enables a target level of noise rejection to be achieved by varying the number, size and placement of the filter and Soller slit assembly. A procedure for determining the reduction in efficiency of the Soller slits with respect to misalignment with the sample will also be discussed.