research papers
Coherent Small-Angle Scattering on a Bending-Magnet Beamline at the ESRF
A coherent X-ray beam is obtained from the D2AM bending-magnet beamline (BM2) of the European Synchrotron Radiation Facility. As this line has permanent convergent optics, monochromated by an Si(111) double monochromator, coherence conditions are satisfied by selecting a part of the beam close to the focal point. Low intensity (106 X-rays s−1) is partially compensated by a high degree of coherence and by the use of a high-resolution direct-illumination CCD area detector. The stability of the small-angle set-up makes quantitative analysis possible. The calculated and measured degrees of coherence are compared. The distributions of speckle intensities are explained by a beam composed of two-thirds coherent and one-third incoherent parts. This incoherent component is caused by Kapton windows, which will be removed in future experiments.
Keywords: coherent X-rays; small-angle scattering; speckle.
1. Introduction
High-brilliance synchrotron radiation sources provide an excellent tool for obtaining coherent X-ray beams (Sutton et al., 1991). This makes it possible to carry out experiments where short-wavelength X-ray beams are used like laser beams, mainly for speckle and X-ray photon correlation spectroscopy measurements in the Bragg peak region (Brauer et al., 1995; Dufresne, 1995; Bley et al., 1995) and in the small-angle domain (Dierker et al., 1995; Mochrie et al., 1997; Mainville et al., 1997).
The basis of the improvements of third-generation synchrotron radiation machines is the high quality of electron optics at a few GeV. For instance, the initial ESRF goals were to obtain emittances of 7 nm (H) × 0.7 nm (V) for an intensity of 100 mA. Now (in 1997), one routinely obtains 4 nm (H) × 0.04 nm (V), with a 200 mA (nominal) intensity. This is an increase of a factor of 35 in −4) is limited by the energy of the electrons (6 GeV), and only a decrease of the source size (from 0.187 × 0.128 mm to 0.142 × 0.037 mm) improves the The total increase for a bending-magnet beamline is close to 12. At the same time, the beam stability has been improved in order that variations do not exceed 10% of the beam size and divergence. With such source performances, a demand exists for better X-ray optics and detectors.
and this can induce a corresponding increase in available coherent beam intensities for undulator sources. In the case of bending-magnet sources, the beam divergence (∼10For a monochromatic beam of wavelength λ, the basic technique for obtaining coherence is to fulfil the condition
where ∊ is the beam divergence and φ is the beam size. In synchrotron radiation, the source size is fixed, and the simplest method can be to limit ∊ using slits. The synchrotron radiation source is smaller in the vertical direction (φv ≃ 40 µm), and the corresponding transverse coherence length, Λt, for λ ≃ 1.5 Å, at L ≃ 40 m, is Lλ/φv ≃ 150 µm. This coherence length is about four times smaller in the horizontal direction. If the whole coherent surface is selected, detection of the speckle structure needs an experimental resolution of the order of λ/Λt, here within the microradian range. In the available experiment set-up, the resolution is ten times larger. In this case, the size of the coherent beam must be less than 10 µm. If no focusing optics are used, only a small part of the coherent available intensity is used, and the high quality of the source is useless. For this reason, focusing optics are an efficient improvement, provided that source degradation due to optical elements is not too strong.
In this paper we report tests with the D2AM beamline at the ESRF for coherent small-angle X-ray scattering (SAXS) experiments. This beamline is located on a bending-magnet source and has high-quality focusing optics. In this case, it is possible to control ∊ and φ simultaneously close to the focused image of the source. This makes possible a quantitative comparison between calculated and experimental intensities and coherence. The effect of inserting various devices into the beam is discussed (windows, monochromators, mirrors). This study provides new ways for improving X-ray optics and X-ray detection, in order to take better advantage of the advances in the synchrotron radiation sources. Results are compared with the interpretation of experimental results (Abernathy et al., 1998) from an undulator source beamline (Troika, ID10).
2. Characteristics of the beamline and the experimental set-up
2.1. Source
of the bending-magnet source can be estimated as (Elleaume, 1986in . This formula holds roughly for 0.8 < λ/λc < 3.0, where λc is the wavelength cut-off: λc = 0.64 Å for E = 6 GeV and for a bending-magnet field of 0.85 T. I is the electron beam intensity (I = 0.02 A). In our experiment, λ = 1.56 Å, and we find
The b, of the source can be deduced from the source dimensions given by ESRF data sheets (Roth et al., 1997). In the case of σh = 142 µm and σv = 37 µm (r.m.s.), we find
From this result, one can roughly estimate the maximum coherent intensity with an Si(111) double monochromator (δλ/λ = 1.4 × 10−4), assuming ∊φ = λ, as
2.2. Optics
The D2 beamline has pseudo-symmetrical double-mirror–double-monochromator optics (Roth et al., 1997). The incoming vertically diverging beam is made parallel by the first curved mirror in order to obtain an excellent monochromatization by the Si crystals. The second monochromator crystal is sagittally curved in order to focus the beam horizontally. As the dimensions of the beam can be 4 mm (V) × 80 mm (H), the second crystal is reinforced with ribs (period 1.2 mm) in order to limit anticlastic curvature. This results in a non-constant and periodic curvature. This is observed for underfocused configurations of the bent monochromator, where the image of the beam close to the sample consists of a horizontal alignment of spots, corresponding to the 1.2 mm period. One of these spots is selected by closing the primary slits S1 (23 m from the source) to 1 × 1 mm in order to optimize the horizontal focusing. The heat load becomes negligible and this improves the experimental stability. The second mirror is a vertically focusing mirror and, as only a small region of the optics is used in this experiment, optical defects are less probable.
The quality of the focusing optics was tested for small beams. In this case, the final limitation of the beam is obtained by the S2 secondary slits, with an opening dh × dv. The source-to-optics distance was 27 m and the 10 µm pinhole was used for scans after optimization of focusing. The pinhole was close to the sample position, 11 m after the centre of the optics and 9.5 m after S2. The image dimensions sh and sv (FWHM) were measured in two different configurations: (i) 1/3 filling, 124 mA, dv × dh = 0.75 × 1.0 mm, sh = 112 µm, sv = 59 µm; (ii) 16 bunches, 90 mA, dv × dh = 0.2 × 0.2 mm, sh = 90 µm, sv = 55 µm.
Focusing is better in the second configuration. This is mainly the result of closing the secondary slits S2, which minimize optical defects. The (FWHM) area of the beam can be a-priori estimated from the (r.m.s.) dimensions of the source and from the ratio between the source-to-optics distance and the optics-to-pinhole distance (2.5 here). This area is σh × σv × (8ln2)/(2.5)2 = 4.7 × 10−3 mm2, which compares well with the best value given here, namely sh × sv = 5 × 10−3 mm2. The focusing optics can be considered as excellent.
2.3. Coherence
In this type of experiment, the number of optical elements is usually minimized in order to limit distortions of the beam. Each of these optical elements acts as a secondary source, so that stable and reliable speckle structures are difficult to obtain. In our case, as permanent focusing optics are present, we take advantage of the large distance (∼10 m) between slits S2 at the end of the optics and the SAXS camera. Coherence conditions will be discussed by considering the square (dh = dv = d) beam across slits S2 as a secondary source. The experimental set-up is sketched in Fig. 1. The pinhole limiting the beam size, of diameter φ, is placed at the entrance of the SAXS camera, at a distance D = 9.5 m from S2. Transverse coherence will be obtained by choosing suitable values of the ratio
This ratio will be essential for the quantitative discussion.
Due to excellent monochromaticity, the longitudinal coherence length, Λl = λ/(δλ/λ), is of the order of 1 µm. For small-angle experiments (here 2θ is less than 0.01 rad), with pinhole diameters φ ≤ 10 µm, and sample thicknesses e ≤ 1 mm, it can be assumed that the conditions of coherence are met (roughly φθ << Λl and eθ2 << Λl).
2.4. Windows
Transmission through windows has strong effects on the beam coherence. There are two 500 µm-thick Be windows on the beam; the first is before the optics and the second is between the first mirror and the monochromators. There is a 50 µm-thick Kapton window just before slits S2. For vacuum security, two thin Kapton windows (thicknesses 20 µm and 50 µm), separated by 1.5 cm of air, are placed 41 cm before the pinhole.
2.5. SAXS camera
The SAXS camera is housed in an evacuated cell. The pinhole holder is followed, after 21 cm, by guard slits (S4). Samples can be placed just after these slits. The beam stop is at a distance of 2.1 m from the sample, just before the end of the evacuated tube. A filter holder is inserted before the detector.
2.6. Detection
A `deep depletion' CCD area detector from Princeton Instruments was used in direct X-ray illumination. The pixels are squares of size r0 = 22.5 µm. At 2.2 m from the sample, the pixel resolution of the CCD (10 µrad) is sufficient to resolve the speckle structure (λ/φ = 15.6 µrad for φ = 10 µm). For a 576 × 384 pixel area, the detector surface is of the order of 1 cm2. The absorption of an X-ray on the surface of the CCD creates very locally about 2000 electronic charges which, as the detector is cooled at 213 K, is more than 100 times larger than the noise. It is easy to calculate the position (Livet et al., 1998) of isolated X-rays with an accuracy of 1 pixel. A custom-designed `droplet' program has been developed that extracts the position of individual photons from images obtained by frequent readings of the detector. With this program, the analog CCD device is transformed into an area photon-counting detector. The overall detection (DQE) of the detector is estimated to be 55% at the 7.95 keV photon energy used here.
3. Results
3.1. Diffraction of the pinholes
One of the standard methods for checking coherence properties of the experiment is to observe Fraunhofer fringes from the pinholes used for the selection of the coherent beam.
In these tests, 0.95Ir0.05 80 µm-thick sheet. Simple calculations show that at 7.95 keV the fraction of X-rays crossing the edges of the aperture is less than 3% for the 10 µm pinhole diameter, and we neglect this effect. This contribution will, however, increase rapidly with photon energy.
pinholes were used. The advantage of these is the high quality of their edges, as can be verified by For their use in X-ray beam selection, these pinholes have relatively thin edges (here, close to 10 µm). The hole is at the centre of a conical hollow in a PtThe pinhole–detector distance is L = 2.41 m, and the measured intensity is strongly localized at the centre of the detector. For that reason, an attenuation by a factor of 103 to 104 is necessary, and the recorded region of the detector is limited to an area of 100 × 100 pixels. This reduces the total time for ADC conversion and for disk-saving to about 0.1 s with a 430 kHz ADC unit and a standard Pentium-based PC computer.
In practice, 2000 frames of 0.03 s each were recorded, each one corresponding to a few hundred detected X-rays. Our `droplet' program was then used. For each frame, the position of each X-ray was registered. Frames were then deleted and a new measurement started.
For studies where time resolution is not required, the number of detected X-rays for each pixel is eventually accumulated. The centre of the pattern (x0, y0) is calculated from the first moment of the distribution. The distance R of each pixel to the centre of the pattern can be deduced.
A series of results obtained with the 10 µm pinhole are plotted in Fig. 2 with various values of d, the aperture of the secondary slits S2: for (a), (b), (c), (d) and (e), d = 50, 100, 150, 200 and 300 µm, respectively. Averages have been taken along right and left half-circles, and R was incremented by 0.25 of a pixel. In Figs. 2(a) and 2(b), Fraunhofer fringes are clearly visible, and there is no significant difference between the two images. Only the different number of accumulated X-rays [3.4 × 105 and 7.5 × 105 for Figs. 2(a) and 2(b), respectively] accounts for the greater noise in the case of Fig. 2(a). The dynamics of the observed intensities is close to five orders of magnitude. Plots of IR3 [Figs. 2(f)–2(j) corresponding to Figs. 2(a) to 2(e)] allow a detailed discussion of the results. The asymptotic behaviour of our measurements can be compared with the asymptotic shape of the diffraction pattern of a pinhole of diameter φ at a distance L from the detector plane,
In Fig. 2, the distances R are expressed in pixel units: n = R/r0. From the period of the oscillations, a precise estimate of φ can be given: φ = 9.35 (5) µm, which is close to the nominal 10 µm diameter. Many oscillations can be observed in Figs. 2(f) and 2(g). This means that the diameter is well characterized and that the edges of the hole are of good quality. Moreover, the curves are very symmetrical, showing that, in the region of the CCD shown in Fig. 2 (60 pixels diameter, i.e. 1.3 mm on the CCD), scattering from the surface of the hole is negligible. Irregular intensity tails are observed for larger R values, and closing of the guard slits (S4) to 70 µm suppresses this parasitic scattering.
Although the oscillations in Figs. 2(f) and 2(g) are not damped, their amplitudes are about 40% of the predicted value. One of the reasons for this is the lack of resolution due to the pixel size. The period of the oscillations is 1.79 pixels. If one approximately calculates the amplitude of the oscillations from the pixel resolution, one obtains
This partly explains the small amplitude of the oscillations. In Figs. 2(c), 2(d) and 2(e) the Fraunhofer oscillations become indistinguishable and the FWHM of the beam increases. A simple formula for that width is
This equation is a first approximation for the beam coherence. Pinhole z in equation (6) is small: z = 0.5 and 1.0 for Figs. 2(a) and 2(b), respectively.
are dominant when the dimensionless ratioA similar discussion may be applied to the 5 µm pinhole. In Fig. 3(a), d = 100 µm and z = 0.5, and the oscillations are visible with a larger period as expected. In Fig. 3(b), d = 300 µm, and z = 1.5, and the oscillations are difficult to observe.
Table 1 shows the observed beam FWHM and the results of equation (9), as a function of φ (and z). Agreement is satisfactory, if account is taken of the 22.5 µm pixel resolution.
|
From filter transmission, the beam intensity for the 10 µm pinhole and d = 200 µm is about 106 X-rays s−1, for a 200 mA electron beam in the storage ring. As the product ∊φ is here close to λ (1.5λ, in fact), this value is to be compared with that calculated from the source characteristics, 1.1 × 107.
The coherence of the beam is difficult to estimate quantitatively from these patterns. Better estimates are obtained from test samples.
3.2. Quantitative study of coherence
3.2.1. Sample
In order to discuss the beamline capacity for obtaining SAXS speckle patterns, scattering from a 0.5 mm-thick Teflon test sample was measured. The sample-to-CCD distance was L = 2.2 m. The guard slits were closed to 70 µm so that, with a 2 mm diameter beamstop located 10 cm before the CCD, the X-ray background was negligible.
The whole detector area was used, and the `droplet' algorithm was applied to 1000 or 2000 frames of 1 s or 2 s each. For such times, the maximum intensity was of the order of 10−2 X-rays pixel−1 frame−1.
In a standard (incoherent) experiment, Teflon has an isotropic scattering pattern. This intensity, 〈I(q)〉, can be obtained from a circular average of the intensity across the area detector.
For example, in Fig. 4 〈I(q)〉 is plotted for the q-region discussed here. The units are counts pixel−1, and the results correspond to an accumulation of 1000 frames of 2 s each, with d = 100 µm and a 10 µm pinhole. The electron beam intensity was 80 mA (16-bunch mode). The averages were carried out over circular bands 3 pixels wide. This provides a reasonable number of pixels (from 1000 to 2000), and the incoherent intensity varies negligibly across these regions. The average number of photons per pixel is low (from 2 to 20 for a total counting time of 30 min), and Poisson statistics have to be taken into account for a quantitative discussion.
3.2.2. Coherence and statistics
Among regions of constant 〈I〉, coherence leads to strong variations of measured intensity. For full coherence, the intensity probability distribution is
In the case of partial coherence, it can be assumed that the scattering observed corresponds to a set of α (α need not be an integer) identical `coherence volumes' scattering independently (ICV assumption). The usual methods of statistical convolution (Brown, 1993) lead to the distribution
where I0 = 〈I〉/α and the global degree of coherence of the experiment is β = 1/α < 1.
3.2.3. Calculated coherence from experimental geometry
In order to test the ICV hypothesis, β has to be calculated. The result will be compared with experimental estimations. We have to consider two averaging processes leading to β < 1. First, we use Young's technique, where a small angular region of an incoherent source (here the beam across S2) is selected. This always leads to partial coherence. Second, the speckle structure is somewhat averaged by the finite size of the CCD pixel. As φ < (r0, d), we will consider these two averages as statistically independent.
The beam coherence, βb, can be obtained from the complex mutual coherence, μ(r), between two points at distance r in the pinhole plane. μ(r) corresponds to the beam radiated by slits S2, acting as a secondary source. The average over the pinhole surface has to be calculated: βb = 〈|μ(r)2|〉. The intensity across S2 must be assumed to be uniform, but the shape of the wavefront is not planar: the beam is convergent. On the other hand, as the beam is focused, it can be considered to be planar close to the pinhole. For this reason, βb is calculated in the reverse manner: the pinhole is considered as a planar incoherent secondary source, and mutual coherence is obtained between two points in the plane of S2, separated by a distance x (Mandel & Wolf, 1995),
and βb is estimated by averaging |μ(x)2| across slits S2,
where z is defined in equation (6).
The intensity radiated by the sample is averaged by the detector. This contribution, βd, can also be obtained by averaging |μ(x)2| over the pixel size. The values of μ(x) are also calculated by considering the beam across the pinhole as a planar incoherent secondary source. The pinhole–sample distance can be neglected. From detector resolution (r0 = 22.5 µm), z = πφr0/2λL ≃ 1 and, from (13), βd = 0.85.
The global degree of coherence of the experiment can be expressed as the product of the two averages, i.e. β0 = βbβd. This assumes that no correlation occurs, on average, between the two calculated values of μ(r). The result of the calculation is given in Table 2 for the set of values of d where scattering from Teflon has been measured. As d varies from 100 to 250 µm, and as only the 10 µm pinhole was used, z lies between 1 and 2.5.
|
In the case of low z, the result of the calculation depends only weakly on the exact shape of the beam. It is not necessary to know precisely the intensity and phase across the incoherent source S2. For instance, βb was also estimated for a planar source across slit S2,
3.2.4. Simple estimation of coherence from the experiment
The coherence can be estimated from the mean square deviation (MSD) of the intensity around circles,
As the Poisson counting statistics are independent of electromagnetic field fluctuations, its corresponding contribution is added to the MSD estimate. In this equation, βa is an apparent coherence.
βa is usually obtained from a plot of {〈I2(q)〉 − [〈I(q)〉]2}/〈I(q)〉 versus 〈I(q)〉. The constant slope of the plot gives an estimate of βa, the error of which is close to 0.01 here, and which provides a test of the experimental stability. Such a plot is shown in Fig. 5 for the same results as in Fig. 4. Here, βa = 0.31 (1) and the calculated value of β is 0.72. Such a discrepancy is observed for all the results: all βa are about 2.5 times lower than β0. Obviously, βa is large enough to observe speckle patterns. This nevertheless leads to a strong decrease in the signal-to-noise ratio and, for the same intensity, it takes 2.5 times longer for the same speckle contrast measurement. Moreover, if the shape of the incoming electromagnetic waves is not fully understood, improvements in the beam and sample stability can be difficult, as well as the interpretation of static measurements.
3.2.5. A detailed statistical study of intensity distribution
One of the origins of the low value of βa can be mechanical instabilities. In fact, after a few hours a non-linear shape is observed on the curves as in Fig. 6. For low intensities, corresponding to large angles (q ≃ 8 × 10−3 Å−1), after some hours the slope of the plot decreases indicating long-term drift of the experiment. This is the only evidence of instability observed, and an intensity decrease is simultaneously observed. This intensity decrease can be compensated by a ±30 µm vertical movement of the second mirror, indicating that instabilities are related mainly to fluctuations in the optics alignment. An approximative daily cycle was observed. The SAXS experiment (including pinholes and detector) is itself mounted on a monolithic granite optical bench and is very stable. This type of support can, however, experience vibrations with a period shorter than the time resolution of 1 s. The main effect of vibrations is to increase the number of coherence volumes (McKechnie, 1975). As in this case, equation (11) should hold with α = 1/βa; this hypothesis can be tested by studying the intensity distribution.
In order to compare equation (11) with the observed intensity probability distribution, it is necessary to take account of Poisson counting statistics, and the probability of obtaining a discrete value of k for I is calculated by a Poisson transform (Mandel formula),
Fig. 6 compares the observed and calculated distributions for various angles for the same results as Fig. 4. The six curves correspond to the six upper points of the figure. The dotted curves are obtained from (16), with α = 1/β = 3.2, as discussed earlier with reference to Fig. 5. As has been already observed (Abernathy et al., 1998), the calculated distribution does not fit the results: the number of low counts is too small, and the observed distribution is more peaked than that calculated. Only the tail of the curve fits well.
The shape of the distribution can be better discussed in terms of moments. Fig. 7 compares the calculated second-order moment of the distribution [equation (15)] (same results as Fig. 4) with the right-hand side of equation (15) (dotted line, with βa from Table 2). Obviously, the fit with βa = 0.31 is excellent. From the third and fourth moments, the skewness is calculated (Mandel & Wolf, 1995) from the integer values of I(q),
and from the curtosis, from which we subtract the Gaussian constant 3,
where Poisson counting statistics introduce correcting second-order terms {O[1/〈I(q)〉2]}, which are calculated numerically. Results of these two calculations are given in Figs. 8 and 9. Open circles correspond to the results of Fig. 4 and closed circles correspond to a second measurement under identical conditions. The comparison between the results of the two measurements (dotted line) and the results of equations (17) and (18), with α = 1/βa (Table 2), clearly shows a discrepancy. In both cases, the experimental results are larger than calculated. Moreover, extrapolation towards low angles (large intensities) should provide an estimate of 2/α1/2 ≃ 1.6 [equation (17)] and 6/α ≃ 3.8 [equation (18)]. This is consistent with β = 0.64, much larger than βa = 0.31, and closer to β0 = 0.72 of Table 2. Skewness and curtosis are independent of the average value of a distribution. This means that a constant shift of the origin of I gives a distribution that is in better agreement with the predicted distribution.
3.2.6. Two beam components
A more realistic statistical model can be deduced from this discussion. Following Abernathy et al. (1998), we assume that the incoming beam can be decomposed into two different contributions. The first, of relative intensity δ, is completely incoherent and gives incoherent small-angle scattering with only Poisson statistics. The second contribution, of relative intensity 1 − δ, is partially coherent with a degree of coherence β. Qualitatively, this model explains why low counts are rarely observed (see Fig. 6). The corresponding distribution has been calculated taking account of Poisson counting statistics contributions. Typical results are shown in Fig. 6 (continuous lines) for the distributions and in Figs. 8 and 9 for the moments, with δ = 0.37 and β = 0.72. The hybrid model explains the observed distribution. In Fig. 7, the continuous line corresponds to the two-beam model, and it is observed that the second-order moment cannot distinguish between the two models discussed here: they both fit the experimental results.
This model was applied to all the measurements of Table 2. All the results agree well with a constant incoherent component, δ ≃ 0.37, of the beam. In this case, β is in agreement with our estimates (β0) from the geometry of the experiment. Although standard coherence theory cannot explain the presence of the incoherent beam, this latter has an intensity that is proportional to the incoming beam. It seems obvious that there is a secondary source that re-emits X-rays under irradiation. This explains the reduction of β: βa ≃ β0(1 − δ)2 ≃ β0/2.5.
3.2.7. Windows
The only possible secondary source is the two thin Kapton windows (20 and 50 µm). These are close to the focus, and one can consider that the irradiated area is 100 µm (H) × 50 µm (V) (FWHM). These windows are about 41 cm before the pinhole. Normally, such windows have only a low absorption, and Kapton is not usually considered as a strong scatterer. In fact, the SAXS intensity is relative to the q-region considered. In the case of these windows, if they scatter strongly in a cone of angle 10−4 rad (in 2θ units), they give a strong contribution to the intensity crossing the pinhole. This intensity has very poor coherence (the corresponding value of z is close to 10). Conversely, a significant part of the beam travelling towards the pinhole is slightly deviated and lost from the experiment. Obviously, this picture is simplified: the whole beam wavefront is distorted on passing through the windows.
This incoherent component of the beam crossing the pinhole can also contribute to the low amplitude of the intensity oscillations in Figs. 2(f) and 2(g).
3.3. Intensity/coherence optimization
If no β can be obtained by closing slits S2. This is obviously at the expense of intensity, and practical criteria for the experiment can be briefly discussed.
is observed, a high value ofThe experimental speckle contrast is of the order of 〈I(q)〉β(z)1/2. Because it depends on the number of coherence areas, β(z) decreases as z−2 for large z, and 〈I(q)〉 increases as z2. We are interested in the square of the signal-to-noise ratio (s/n), where
where the square of the Poisson counting noise is 〈I(q)〉. Starting from the calculated coherence [equation (13) gives the first terms of the series expansion], the product z2βb(z) is π for z = ∞. Results for the z values of interest are given in Table 2. The values of the product still show a significant increase for z > 2. For instance, the condition z = 2 means βb = 0.53, and the product is 2.15, which is significantly smaller than π. For z = 3, βb = 0.28, and the product is 2.6. For these values, pinhole Fraunhofer oscillations are difficult to observe (see Fig. 2). This means that the observation of these oscillations is not necessary for the experimental study of speckle contrast. In practice, z cannot be very large. In our case, z is varied by opening slits S2, and inhomogeneities or instabilities of the beam may also have a strong influence. As a 25% speckle contrast is often considered to be excellent, z may range from 2 to 3 in practical experiments, if is eliminated.
3.4. Beam intensity
The quality of the source is in agreement with the ESRF data sheets, and no significant degradation is observed after the focusing optics. This latter was originally designed to obtain a stable focus size of 300 µm × 300 µm (1/10 mm2) and, provided that only a small fraction is used, we obtain a stable FWHM area of 1/200 mm2.
The beam intensity was observed to be about 106 photons s−1 for ∊φ ≃ 1.3λ (z ≃ 2). The intensity estimated from the source characteristics is 1.1 × 107 photons s−1. Some losses have been neglected: for beryllium windows (∼0.1 mm), for which we calculate a transmission of 0.75, the reflectivity of each mirror is roughly 70%, yielding a total loss factor of 3. This leaves a residual factor of about 4. As the focus is excellent, the mirror and the monochromator optical elements are clear of suspicion.
The main explanation for this intensity loss can be found in the above discussion. There are two unpolished beryllium windows prior to the monochromator and a Kapton window after the optics, before slits S2. The intensity diffracted at very small angles by these windows is completely eliminated in the case of the very high experimental resolution needed here. Qualitatively, this means that a large fraction of the improvements in synchrotron radiation source performance are lost because of these windows.
4. Discussion
We can now briefly compare these results with those obtained on an ESRF undulator beamline (Troïka), as discussed by Mainville et al. (1997) and Abernathy et al. (1998). In the latter paper, the degree of coherence is estimated for a very different set-up: (i) by means of plane mirrors, a wide energy bandwidth (δλ/λ = 1.3%), corresponding to the third harmonic of the undulator, is selected, and the beam intensity is three orders of magnitude larger (109 photons s−1); (ii) the main causes of loss of coherence are the windows which decrease β by a factor of 10 (Abernathy et al., 1998), and the energy bandwidth which decreases β by a factor of 3 for q ≃ 0.01 Å−1. This also explains the elongated shape of the speckles observed.
The 3 times greater than that of a bending-magnet beamline, and this is sufficient to explain the observed intensity difference.
of an undulator is 10Apart from this obvious difference, we have here chosen to improve the degree of coherence and to have a better control of its value and stability. This is partly at the expense of intensity, but a smaller intensity can be compensated by greater coherence in a speckle experiment. Low intensity and high coherence are also more compatible with the use of a CCD area detector, at least with the droplet algorithm. Currently, CCDs are read at a low rate (less than 1 MHz conversion frequency for Princeton Instruments), and our algorithm works if individual photons are easily identified (in practice, the intensity must be less than one-tenth of an X-ray per pixel in the main part of the detector).
Windows introduce strong degradation of the synchrotron radiation source properties. Here, the
is decreased because the windows act as secondary sources in the beam trajectory.5. Conclusions
In this paper, we have shown that excellent results can be obtained in static coherent experiments with the D2AM beamline. The discussion clearly shows that the optical performance is at a high level. Moreover, the set-up is stable for several hours and for q smaller than 0.01 Å−1. Removal of the Kapton windows will increase the degree of coherence and make its measurement more precise. The arrangement provides a tool for the study of slow processes, of small spatial extension (from 100 to 2000 Å).
For the future of such measurements, focusing and monochromated beamlines from window-free undulators will be very efficient.
Acknowledgements
The authors wish to thank Mark Sutton for suggesting realistic calculations of the beam coherence, Erik Geissler, Jacques Mainville, René Caudron, Doug Abernathy and G. Grübel for fruitful collaboration, and René Caillot and Stephan Arnaud for technical help.
References
Abernathy, D. L., Grübel, G., Brauer, S., McNulty, I., Stephenson, G. B., Mochrie, S. G., Sandy, A. R., Mulders, N. & Sutton, M. (1998). J. Synchrotron Rad. 5, 37–47. Web of Science CrossRef CAS IUCr Journals
Bley, F., Livet, F., Leroux, J. C., Simon, J. P., Abernathy, D., Als-Nielsen, D., Grübel, G., Vignaud, G., Dolino, G., Legrand, J. F., Camel, D., Menguy, N. & Papoular, M. (1995). Acta Cryst. A51, 746–750. CrossRef CAS Web of Science IUCr Journals
Brauer, S., Stephenson, G. B., Sutton, M., Brüning, R., Dufresne, E., Mochrie, S. G. J., Grübel, G., Als-Nielsen, J. & Abernathy, D. L. (1995). Phys. Rev. Lett. 74, 2010–2013. CrossRef PubMed CAS Web of Science
Brown, W. (1993). Editor. Dynamic Light Scattering, p. 90. Oxford: Clarendon Press.
Dierker, S. B., Pindak, R., Fleming, R. M., Robinson, I. K. & Berman, L. (1995). Phys. Rev. Lett. 75, 449–452. CrossRef PubMed CAS Web of Science
Dufresne, E. (1995). PhD thesis, McGill University, Canada.
Elleaume, P. (1986). Summer School on Synchrotron Radiation, Aussois. Grenoble: CNRS.
Livet, F., Bley, F., Sutton, M., Mainville, J., Geissler, E., Dolino, G. & Caudron, R. (1998). To be published.
McKechnie, T. S. (1975). Opt. Commun. 13, 35–39. CrossRef Web of Science
Mainville, J., Bley, F., Livet, F., Geissler, E., Legrand, J. F., Abernathy, D. L., Grübel, G., Mochrie, S. G. J. & Sutton, M. (1997). J. Appl. Cryst. 30, 828–832. Web of Science CrossRef CAS IUCr Journals
Mandel, L. & Wolf, E. (1995). Optical Coherence and Quantum Optics. Cambridge University Press.
Mochrie, S. G. J., Mayes, A. M., Sandy, A. R., Sutton, M., Brauer, S., Stephenson, G. B., Abernathy, D. L. & Grubel, G. (1997). Phys. Rev. Lett. 78, 1275–1278. CrossRef CAS Web of Science
Roth, M., Ferrer, J. L., Simon, J. P. & Berar, J. F. (1997). ESRF Beamline Handbook, p. 43. ESRF, Grenoble, France.
Sutton, M., Mochrie, S. G. J., Greytak, T., Nagler, S. E., Berman, L. E., Held, G. A. & Stephenson, G. B. (1991). Nature (London), 352, 608–610. CrossRef Web of Science
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.