research papers
A device for selecting and rejecting X-ray harmonics in synchrotron radiation beams
aPhysics Division of BCPS Department, Illinois Institute of Technology, Chicago, IL 60616, USA, and bDepartment of Anatomy and Cell Biology, College of Medicine, University of Saskatchewan, Saskatoon, Canada SK S7N 5E5
*Correspondence e-mail: karacah@iit.edu
A practical device (`beam cleaner') is described that selects a specific harmonic (e.g. third order) from other nearby harmonics (e.g. fourth or fifth orders) that are transmitted by the beamline monochromator. The ability to isolate and use a clean harmonic beam effectively extends the energy range of existing beamlines. This device is a tunable medium-resolution X-ray which consists of a double bent Laue crystal post-monochromator. It is a simple and inexpensive device that does not require precision goniometry and can be operated in air. The device can also be used to passively compensate for vertical beam motion as the primary beamline monochromator is scanned. Experimental results are presented that demonstrate the improvement in data quality under high harmonic conditions.
Keywords: X-ray harmonics; XAFS; X-ray monochromators; synchrotron instrumentation; beamlines.
1. Introduction
Double-crystal monochromators commonly use low orders of diffraction [e.g. Si(111), Si(220)] to prepare X-ray beams for research at synchrotron sources. These reflections provide high intensity (compared with higher-order reflections) and are adequate for many experiments. However, in some cases it becomes desirable to use the higher-order harmonics from the monochromator to access higher X-ray energies, or alternatively to improve the energy resolution. The use of a higher-order harmonic presents a new problem for the researcher: the selection of that specific harmonic while rejecting the fundamental and other harmonics.
What is needed is a medium-bandwidth optic that would allow the experimenter to select particular harmonics. We have previously shown (Karanfil et al., 2000) that a single bent crystal in Laue geometry can serve this purpose. Here we describe our approach, which uses diffraction from a bent Laue double-crystal system as a moderate bandwidth post-monochromator to select the desired harmonic energy. This can easily be scanned in energy to track the primary monochromator. Perfect crystals are not ideal in this situation because of their very narrow reflectivity widths. However, the use of highly bent crystals makes the mechanism simpler to construct and align since the crystals provide a large reflectivity width as well as surprisingly high reflectivity (Erola et al., 1990).
2. Bent-crystal optics
Over the past several years bent Laue crystals have been used for a number of applications, including medical imaging (Karanfil et al., 2000; Zhong, 1996; Zhong et al., 1997) and fluorescence detection (Karanfil et al., 2000). A surprising aspect of diffraction from an optimized Laue geometry is the high efficiency that such a bent Laue crystal may have, sometimes as high as 80%. A major advantage of a bent-crystal geometry is the enhanced reflectivity width due to the bending of the lattice planes in the transmission geometry. The reflectivity width can be calculated from geometrical and lattice compression considerations to be (Erola et al., 1990)
where T is the crystal thickness, is the bending radius, is the asymmetry angle of the lattice planes with respect to the crystal surface, is the and is the Poisson ratio for the crystal. Equation (1) gives the expected reflection width of the diffraction that occurs in the crystal. The intensity that is diffracted is more complicated to calculate. For most purposes a kinematic approximation can be made for the diffracted intensity. We will assume that the crystal is bent sufficiently for the crystal to behave as a fully In this case the integrated intensity is that given by the kinematic theory and is assumed to be uniform in intensity over the reflectivity width (Zachariasen, 1967).
3. Gap and offset determination of the monochromator and a double bent crystal system
The `beam cleaner' is a double-crystal device that has two bent crystals in Laue geometry with a gap between them set appropriately (Fig. 1). A single crystal is inconvenient to use as a cleaner of the incident beam because the experimental apparatus must follow the diffracted beam if the energy is changed. It can be shown that the variation of the double-crystal beam offset of the Laue-case system is in the opposite sense to the double-crystal Bragg system. For example, in terms of the perpendicular crystal separation GM and the offset HM of a double-crystal symmetric monochromator in the Bragg case,
When the monochromator is rotated about an axis passing through the diffraction surface of the first crystal, the variation in the offset is given by
Similarly, for the symmetric Laue (SL) case the offset HSL and variation are
Again, for a rotation axis passing through the diffracting region of the first Laue crystal, the variation in the offset is
Since the offset change for the Bragg case (monochromator) is opposite to that for the Laue cleaner the variation in offset can be removed at an energy. For simplicity, it is assumed that the same reflection is used ( = = ) with a given gap for the cleaner. The combined variation in offset of a tandem monochromator and cleaner system is then
Since the angle change of each
is the same for a matched thenTherefore the overall offset variation can be removed around the design energy if
This condition gives
This relationship can be used to define the Laue crystal separation GSL for a given monochromator gap GM. Similar equations for the offset and variation can be derived for the asymmetric Laue (AL) geometry. The asymmetry angle of the lattice planes is and is measured from the normal direction,
and
Here the upper and lower signs refer to the side of the lattice planes used in relation to the bending radius of the crystal. The upper sign refers to the geometry when the source and bending radius are on the same side of the diffraction planes.
The ratio of crystal separations of the monochromator and cleaner can be found as
Thus the possibility of having the beamline monochromator operate in `channel cut' mode in tandem with the cleaner can result in a cleaned beam with no or very little offset variation. Note that these equations assume that the crystals are parallel. This is the only practical crystal arrangement that allows the crystals to be placed in close proximity. These arrangements are on the first crystal and on the second crystal or vice versa. Arrangements of (or ) incident on both crystals result in the crystals intercepting each other for reasonable values of and crystal size, and these configurations are not considered in this work.
4. Experimental set-up and results
Experiments were performed at the MR-CAT (sector 10) beamline at the Advanced Photon Source at Argonne National Laboratory. The beamline monochromator was a Si(111) double-crystal mechanism. No mirror or post-monochromator optical elements were in place to affect the harmonic spectra.
4.1. Reflectivity measurements
Reflectivity measurements were performed at a number of energies. Results at 18 keV will be shown and are representative of those obtained at other energies. At 18 keV the beamline monochromator was set at 6 keV from the Si(111) and thus Si(333) passed 18 keV. The undulator was set with the fundamental also at 6 keV and the 18 keV harmonic was used as well.
Both crystals of the device were mounted on a common plate. This plate can be rotated to change the energy using a rotary stage (Huber 410). The first crystal was mounted on the rotation axis of the stage. The second crystal was mounted on a translation stage to adjust the perpendicular distance between the crystals (gap). The second crystal also has a rotation adjustment using a second translation stage to move a tangent bar. Both crystals have adjustable bending radii using a screw assembly with spring.
Two argon-filled ion chambers were used to measure the diffracted and transmitted beams from the first and second crystals. The direct beam was blocked downstream of the cleaner assembly by a lead stop. One of the ion chambers was positioned to intercept the diffracted beam from the first crystal passed through the second crystal. The second ion chamber intercepted the beam diffracted by the second crystal. This chamber is used to set the first cleaner crystal to the shows a comparison between the calculated and the measured rocking curve of a pair of 200 µm-thick Si crystals at 18 keV. The measured data are shown as the solid curve. A calculated rocking curve for the two-crystal cleaner is shown as the dashed curve for comparison. This calculation is based on the convolution of the calculated plane-wave rocking curve for each crystal using the REFLECT computer code (Eteläniemi et al., 1989). The measured diffraction efficiency of the double bent crystal arrangement is about 45%. The measurements at the other energies (15, 20, 25, 30 keV) with different bending radii are also consistent with those obtained by the REFLECT computer program (Figs. 3, 4 and 5).
This beam is the one to be used in the beam cleaner experiment. The plot in Fig. 24.2. Offset measurements
To operate the cleaner in tandem with a beamline monochromator, beam motion effects can be eliminated in both devices by using the correct gap between the cleaner crystals [see equations (10) and (12)]. The beamline monochromator offset is 35 mm. The calculated gap required for the beam cleaner to correct for the vertical beam motion in the monochromator fixed-gap mode is 4.5 mm at 19.5 keV. In this case the (333) reflection of the monochromator is selected by the cleaner (111). To confirm the dependence of the offset as a function of gap at a given monochromator energy, the crystal gap was chosen to be large enough for the beam displacements to be easily measurable. Fig. 6 shows the measurements of gap and offset for an energy of 18 keV. There is good agreement between measured and calculated values within the estimated error of the gap–offset measurements.
4.3. measurements with the beam cleaner
Spectral purity of the X-ray beam is essential for accurate X-ray absorption spectra. If harmonics are present, noise is increased because variations in intensity of the X-ray beam do not divide out properly in the ratio of detector signals, and spectra are distorted. This effect is especially noticeable in transmission mode because of `thickness effects' (Koningsberger & Prins, 1988; Parrat et al., 1957; Stern & Kim, 1981).
We have previously established the feasibility of selecting harmonics with a single-crystal beam cleaner (Karanfil et al., 2000). In that paper the efficiency of a single-crystal beam cleaner was approximately 70%. In Fig. 3 of that paper the energy spectrum clearly showed that only the desired energy harmonic was transmitted.
In order to demonstrate the ability of a double beam cleaner to reject X-rays at undesired energies during an energy scan, we carried out transmission K edge, 17.998 keV) using the Si(333) reflection of the beamline monochromator, the undulator third harmonic and Si(111) reflection of the beam cleaner. The beam cleaner angle was scanned in tandem with the primary monochromator in much the same way that undulator gaps are scanned. Tracking is straightforward because of the relatively large energy bandwidth of the beam cleaner (∼50 eV). The monochromator was set to 6 keV (fundamental) for the first configuration and the (333) reflection of 18 keV was used (Fig. 7a). The spectrum for this case is shown in Fig. 8.
measurements on a zirconium foil (It is useful to consider the effect of harmonics on the signal-to-noise ratio in transmission ln(I0/I1), where I0 and I1 are the intensities measured in detectors placed before and after the sample. In the presence of undesired harmonics, h0 and h1 are measured in the two detectors, and the measured is given by = ln[(I0+h0)/(I1+h1)]. However, if the harmonic contamination is small compared with the signal, a first-order expansion gives the error in the as = (h0/I0)-(h1/I1). Each of the signals I0, I1, h0 and h1 fluctuate in time, but I0 and I1 are statistically highly correlated with each other, as are h0 and h1. Note that, even if the harmonic contamination were constant, the fluctuations in I0 and I1 generally do not cancel out because the fraction of harmonic contamination will generally be different in the front and rear detectors. The implication is that the error in the is of the same order of magnitude as the ratio of the harmonic content to the total signal. The variation in owing to is typically of the order of several percent. So harmonic contamination of the order of a few percent in the beam can completely obscure the This effect is seen in Fig. 8. In addition, the absorption-edge jumps are reduced. Consider a simple calculation of the edge jump under conditions in which the harmonic content (h0 and h1) are both 10% of I0. If the product of the sample thickness and the increases from 1.0 to 2.0 over the (an edge jump of 1.0), the apparent edge jump in the presence of harmonics would be reduced to 0.69, a 31% reduction.
If no undesired harmonics were present, the would be given by =In a real experiment the fundamental would be filtered out in order to reduce contamination from 6 keV. Therefore an aluminium attenuator of thickness 0.3 mm is placed in front of the slits (Fig. 7b) to block almost all of the fundamental energy (6 keV) but to transmit 18 keV (70%). The spectrum for this case is shown in Fig. 9(a). As seen in Fig. 9(a), the spectrum of Zr foil without the beam cleaner is severely distorted and noisy with a greatly reduced measured edge step. The edge step is substantially reduced from its true value of 17.998 keV, and it is very noisy because fluctuations in measured incident-beam intensity do not divide out correctly. With the beam cleaner (Fig. 7d), the spectrum is excellent (Fig. 9b).
In order to simulate a thick sample (to further exacerbate thickness effects) we placed a 1.22 mm-thick aluminium attenuator over the sample surface (Figs. 7c and 7e). This attenuator has the primary effect of `hardening' the beam, i.e. increasing the proportion of high-energy harmonics in the transmitted beam, thereby distorting the spectra further. Fig. 10 shows the benefit of using the beam cleaner. Without the beam cleaner (Fig. 10a) the spectrum of the Zr foil plus 1.22 mm aluminium is even worse, as expected because of the beam hardening effect of the aluminium, i.e. the proportion of harmonics in the transmitted beam is increased by the absorption of the aluminium. Even in this extreme case, the beam cleaner produces an excellent spectrum, which is the same as the original spectrum but shifted upward by the aluminium absorption (Fig. 10b). They differ only by the additive absorption of the aluminium.
The edge steps and [background-subtracted Fig. 11(a), and Fourier transform (FT), Fig. 11(b)]. These clearly show no differences between the spectra of the sample with the aluminium and without it. This demonstrates that there is no amplitude reduction due to beam hardening; the beam cleaner effectively removes the harmonic contamination.
amplitudes are not affected by adding the aluminium, as can be seen in Fig. 115. Conclusion
The beam cleaner has been shown to be an effective medium-resolution bandpass monochromator that can be used to select a harmonic from a beamline monochromator. It can improve data quality by eliminating undesired energies in the beam. A useful feature is that it can be operated in a mode that compensates for beam height changes from the primary monochromator when operated in channel-cut mode. In many applications a very simple water or liquid-nitrogen-cooled channel-cut monochromator could be used on a beamline provided it was followed by an appropriate beam cleaner.
Beam cleaners can extend the effective energy range of monochromators by making high-order reflections of the primary monochromator more useful, eliminating the need to change crystals to reach high energies. Bent Laue crystals have wide reflectivity curves (and high reflectivity), which allows easy alignment and scanning.
Acknowledgements
The authors wish to acknowledge useful discussions regarding the device with Zhong Zhong. In addition, the authors would like to thank Thomas Gog and Jeremy Kropf for the loan of some of the equipment used in this experiment. The assistance of Nadia Leyarovska is also gratefully acknowledged. Work performed at MRCAT was supported in part by funding from the Department of Energy under grant number DEFG0200ER45811. We wish to acknowledge the support in part by US Army grant DAMD17-99-1-9217, US DOE contract DE-AC02-CH10886, State of Illinois Higher Education Cooperative Agreement, National Institutes of Health RR08630 and R43 RR018635-01.
References
Erola, E., Eteläniemi, V., Suortti, P., Pattison, P. & Thomlinson, W. (1990). J. Appl. Cryst. 23, 35–45. CrossRef Web of Science IUCr Journals
Eteläniemi, V., Suortti, P. & Thomlinson, W. (1989). Report BNL-43247. National Synchrotron Light Source, Brookhaven National Laboratory, Upton, NY, USA.
Karanfil, C., Chapman, L. D., Bunker, G. B., Segre, C. U. & Leyarovska, N. E. (2000). AIP Conf. Proc. 521, 276–282. CrossRef
Karanfil, C., Zhong, Z., Chapman, L. D., Fischetti, R., Bunker, G. B., Segre, C. U. & Bunker, B. A. (2000). AIP Conf. Proc. 521, 178–182. CrossRef CAS
Koningsberger, D. C. & Prins, R. (1988). X-ray Absorption: Principles, Applications, Techniques of EXAFS, SEXAFS and XAFS. New York: Wiley.
Parrat, L. G., Hempsted, C. F. & Jossem, E. L. (1957). Phys. Rev. 105, 1228–1232. CrossRef Web of Science
Stern, E. A. & Kim, K. (1981). Phys. Rev. B, 23, 3781–3787. CrossRef CAS Web of Science
Zachariasen, W. H. (1967). Theory of X-ray Diffraction in Crystals. New York: Dover.
Zhong, Z. (1996). PhD thesis, SUNY at Stony Brook, USA.
Zhong, Z., Chapman, D., Menk, R., Richardson, J., Theophanis, S. & Thomlinson, W. (1997). Phys. Med. Biol. 42, 1751–1762. CrossRef CAS PubMed Web of Science