Performance of a dispersion-compensating scanning X-ray spectrometer for Compton profile measurements
aSynchrotron Radiation Facility (ESRF), BP 220, 38043 Grenoble CEDEX 9, France, and bDepartment of Physics, University of Helsinki, PO Box 64, FIN-0014 Helsinki, Finland
*Correspondence e-mail: hiraoka@spring8.or.jp
A new X-ray spectrometer has been constructed for Compton profile measurements at beamline ID15B of the ESRF. The spectrometer is based on a novel idea, dispersion compensation, which was proposed earlier. A cylindrically bent Laue monochromator focuses ∼90 keV synchrotron radiation at about 0.7 m before the sample, and produces a well defined energy or wavelength gradient on the sample. A cylindrically bent Laue analyser almost perfectly compensates this wavelength gradient. Using an Al sample, it has been confirmed that the new spectrometer improves the
by a factor of two compared with the previously constructed 30 keV and 60 keV spectrometers, with a comparable momentum resolution. Because of reduced absorption owing to use of high-energy X-rays, the enhancement of the is spectacular for heavy-element materials.Keywords: X-ray spectrometer; Compton scattering; dispersion compensation.
1. Introduction
Compton-scattered X-rays from a material provide us with information about the electron momentum distribution. Within the impulse approximation (Eisenberger & Platzman, 1970), the energy of scattered X-rays () is simply determined by the energy of the incident X-rays (), the scattering angle () and the primary electron momentum (p1),
where m and c are the electron mass and the velocity of light, respectively. The p1·K term gives the projection of p1 onto the scattering vector, K, which is defined by , and . Therefore, the energy spectrum provides the momentum density [(p1)] projected onto K, i.e. the Compton profile,
The z-axis is typically taken parallel to K, and p1z is the z-component of p1, i.e. the projection of p1 onto K ( p1z = p1·K/|K|). When p, and the wavelengths of the incident X-rays () and scattered X-rays () are used, (1) yields a simple result for the Compton peak,
where h is the Planck constant. This relation is important for the operation of the dispersion-compensating spectrometer, as discussed later. In order to investigate momentum densities in materials, various kinds of X-ray spectrometers have been developed for Compton scattering experiments. Currently two methods are used for energy analysis of scattered radiation. One uses a Ge solid-state detector (SSD) and the other uses a crystal analyser. The latter makes possible experiments with three to five times higher resolution. This improvement is very significant for examining solids, particularly in investigating Fermi surfaces in momentum space. However, the efficiency of a crystal analyser is rather low. Therefore, there was limited application of this kind of spectrometer to heavy-element materials, which strongly absorb the X-rays, before the recent advent of intense high-energy synchrotron radiation sources.
Compton spectrometers using a crystal analyser can be classified into either the Cauchois type or the scanning type. In the Cauchois type, all the components are basically stationary. A spectrum is recorded as a function of position on a position-sensitive detector (Loupias & Petiau, 1980; Shiotani et al., 1989; Berthold et al., 1992; Sakurai et al., 1992; Hiraoka et al., 2001; Itou & Sakurai, 2004). In the scanning type, a spectrum is recorded as a function of the of the crystal analyser (), where the analyser and the detector move along several axes, synchronized with (Suortti et al., 1999). If each component performs adequately, one can have similar resolutions and count rates using both types of spectrometers. However, problems arise in the high-energy region. High-energy X-rays of ∼100 keV are not focused or reflected well by bent crystals, and are detected with less efficiency and poor spatial resolution by position-sensitive detectors compared with X-rays of ≤30 keV.
As for the monochromator, the Rowland circle geometry with a bent Bragg crystal has been used widely so far for both types of spectrometers. Here, a light source and a sample are placed on the Rowland circle defined by the bending radius of the monochromator. The finite thickness of the crystal, however, generally causes a rather large aberration, i.e. a large focal size for high-energy X-rays. For scanning spectrometers, the focusing properties are particularly important because the resolution is mainly determined by the focal size or the width of the slits placed at the focal point. Schulze et al. (1998) demonstrated the ability of a cylindrically bent Laue crystal to eliminate this aberration. Nevertheless, it was not straightforward to apply a bent Laue monochromator to spectrometers for Compton scattering. This is because the light source can no longer be placed on the Rowland circle in the Laue geometry, and thus the reflected beam has a rather wide range of energies, or a large wavelength gradient.
A few years ago we proposed a new combination of a monochromator, an analyser and a detector, so as to compensate the wavelength gradient and to optimize a scanning spectrometer in the ∼100 keV region (Suortti et al., 2001). In the report, we showed the potential of the dispersion-compensating spectrometer with results of a test experiment. After having implemented several improvements, the spectrometer is now used for actual experiments. In the present report we show the current performance of the dispersion-compensating spectrometer. Some improvements are still necessary, but the spectrometer is already close to its ultimate performance.
2. Dispersion-compensating spectrometer
The idea of the dispersion-compensating spectrometer has already been described in our earlier work (Suortti et al., 2001). We repeat some essential parts here to help understand the principle. Fig. 1 illustrates the spectrometer. The monochromator is a cylindrically bent Laue crystal. The bent Laue monochromator leads to high reflectivity, an optimized bandpass and aberration-free focusing. As already mentioned, the Laue monochromator also has a wide range of ( in Fig. 1), i.e. a large energy or wavelength gradient along the horizontal axis (d/ds), where is the wavelength and s is a position along the tangential axis to the curvature of the crystal (see Fig. 1). The wavelength gradient on the monochromator is given by
where dM is the d-spacing of the monochromator (Si 511), P is the distance between the source and the monochromator, and P0 = R RM is the bending radius and is the asymmetric cut of the monochromator. Equation (4) is equivalent to equation (4) of Suortti et al. (2001). The beam converges to the focus at a distance Q from the monochromator, and at the sample the gradient is reversed and magnified by a factor Q/L, where L is the distance between the focal point and the sample. The wavelength gradient on the sample, (, is
where x is a position along the horizontal axis on the sample. Similarly, the scattered beam reflected by the analyser has a wavelength gradient (d/dx)A owing to the width of the beam at the sample. The width of the analyser reflectivity curve can be adjusted by the asymmetric cut (Suortti et al., 1997). For the energy resolution required in the present case, is small, only about 1.5°. The , is also small at high photon energies, so that the dispersion relation simplifies to
where dA and are the d-spacing of the analyser (Si 400) and the respectively, and RA is the bending radius. This equation is equivalent to equation (5) of Suortti et al. (2001) at the limit of small and . When the dispersion-compensation condition is fulfilled, the two gradients coincide, i.e. (d/dx)M = (d/dx)A. If perfect gradients and their compensation are achieved, the spectrum measured as a function of shows a single peak, like the experimental result obtained for (see Fig. 3), but the analyser actually reflects X-rays of different energies.
The condition of the dispersion compensation is precisely valid only for one gradient. Once the dispersion compensation is optimized for the that the gradient is the same at p1z = 0, so that the dispersion compensation is exact. Away from the Compton peak, there is an extra broadening of the resolution function owing to the uncompensated dispersion, but the effect is very small, only a few percent of the resolution function, which can be neglected (Suortti et al., 2001). It is noted that (5) assumes the x-axis on the sample to be parallel to the s-axis on the monochromator. If the x-axis or the sample surface is rotated, the dispersion-compensation condition should be re-adjusted by changing RA.
of X-rays, it is seen from (3)3. Performance tests
3.1. Wavelength gradient
The wavelength gradient and the intensity distribution of the incident beam were measured by the following experiment. A 200 µm-thick gold plate, placed at the sample position, was rotated so that the surface normal was at an angle of 60° from the incident beam, and the edge of the plate was irradiated by X-rays (see the inset in Fig. 2). Using a Ge SSD, placed at a scattering angle of 150°, the spectrum was collected as a function of the position of the gold plate. In this geometry, scattering from the surfaces was strongly absorbed, and the signal was dominated by the scattering from the edge. Hence, it was possible to obtain the wavelength gradient and the intensity distribution with a spatial resolution of 200 µm. Fig. 2 plots the peaks and integrated intensities of spectra as a function of the position x. Least-squares fitting shows that = 1.65 Å m−1 = 1.65 × 10-10. This agrees well with the gradient of 1.70 Å m−1 expected from the geometry of the monochromator. Using the observed we obtain RA = 1.60 m from (6).
3.2. Resolution function
As already mentioned, the line shape of the elastic line provides the exact resolution function of a Compton profile at p1z = 0. Fig. 3 shows the elastic lines from the gold plate, which was mounted so that the surface normal was parallel to the scattering vector (see the inset). For a scanning spectrometer, the resolution is a function of the sample thickness (t) or the penetration length of X-rays (lp) in the sample. In the geometry of our spectrometer, the deterioration of the resolution is seen when both t and lp are 1 mm. The gold plate, having a much smaller lp, ∼40 µm, simulates an ideal sample. Using a 100 µm-wide slit, we have obtained a resolution of 120 eV (FWHM). This corresponds to a momentum resolution of 0.08 a.u. at the Compton peak (here a.u. means and 1 a.u. = 1.99 × 10-24 kg m s-1). Similarly, we have obtained a resolution of 90 eV (0.06 a.u.) with a 60 µm-wide slit. The 60 µm-wide slit simply causes an intensity loss of 40% compared with the 100 µm-wide slit. Table 1 shows several main contributions to the resolutions. The observed resolutions agree well with our expectations.
|
3.3. Compton profiles
Fig. 4(a) shows the Compton profile of Al. It was measured using a 100 µm-wide slit on a 0.5 mm-thick Al sample, using the 90 keV spectrometer. The scattering angle was 172°. The resolution of 0.08 a.u. is confirmed from the FWHM of the elastic line. The observed spectrum is almost perfectly reproduced with the valence-electron Compton profile based on band theory (Papanicolaou et al., 1991), the core-electron Compton profile from the Hartree–Fock calculation (Biggs et al., 1975), a multiple scattering spectrum obtained by Monte Carlo simulation (Fajardo et al., 1998), and a parabolic background. The at the Compton peak was 1200 counts s−1 at ∼60 mA storage-ring current.
One of most important comparisons is between the performance of the present 90 keV spectrometer and that of the previously constructed 60 keV and 30 keV ones having a standard Rowland circle geometry. The 60 keV spectrometer consists of a Si 311 monochromator and a Ge 440 analyser, while that of 30 keV consists of a Si 111 monochromator and a Si 400 analyser (see Suortti et al., 1999). The slit widths of the 60 keV and 30 keV spectrometers are typically 200 µm, while that of the 90 keV spectrometers is 100 µm. The scattering angles are common, 172°, for the three spectrometers. The comparisons are summarized in Table 2. The is about twice the rate achieved with 60 keV and 30 keV conventional spectrometers. The resolution is better by a factor of ∼2 than that of the 60 keV spectrometer, and the same as that of the 30 keV spectrometer. The background is 3% of the peak intensity, and most of the background is from air scattering around the sample. If we had used a vacuum chamber, already available at the beamline, the background would have been significantly reduced.
‡Hiraoka, Buslaps, Honkimäki, Minami & Uwe (2005). §Hiraoka, Buslaps, Honkimäki, Itou et al. (2005). |
Compton profile measurements on heavy-element materials, utilizing high-energy synchrotron radiation are one of the biggest motivations for the construction of this spectrometer. Fig. 4(b) shows the experimental Compton profile of an unconventional superconductor, Sr2RuO4, along with the theoretical Compton profile calculated based on band theory (Hiraoka, Buslaps, Honkimäki, Itou et al., 2005). Reduced absorption in the = 90 keV experiment increases the Compton scattering intensity by more than three and 15 times, compared with in the = 60 keV and 30 keV experiments on this sample, respectively. Namely, the total improvement in the count rate is a factor of 6 and 30. In this experiment, almost perfect collimation was achieved for scattered X-rays from the sample mounted on the 20 K cryostat, and the background-to-signal ratio was ∼0.01%. The overall shape is almost perfectly reproduced without assuming any background. The resolution is deteriorated by a factor of about two compared with the experiment on Al. The extra peak broadening was due to a mismatch of the energy gradients of the monochromator and analyzer because of the following two reasons: firstly, the surface was not flat and, secondly, the Compton profile was measured for the [100] direction, which made an angle of 22° with the surface normal. This could be re-adjusted as mentioned at the end of the last section.
4. Summary and future improvements
The new Compton spectrometer, based on a novel idea, i.e. dispersion compensation, was constructed at beamline ID15 of the ESRF. This spectrometer has been confirmed with an Al sample to improve the by a factor of two with a resolution comparable with the conventional 60 keV and 30 keV spectrometers. The enhancement of the is spectacular in measurements for heavy-element materials owing to the reduced absorption of high-energy X-rays. The peak is ∼1000 counts s−1 for samples including 4d or 5p elements. With this typically it is possible to reach a Compton profile with 100000 counts at the peak within a bin of 0.02 a.u. in half a day, and thus eight to ten Compton profiles in a beam time of one week. In fact, for all the samples shown in Table 2 (except for Al), the momentum densities were reconstructed from Compton profiles acquired in one week. They all showed significant results (Hiraoka, Buslaps, Honkimäki, Guyot & Schlenker, 2005; Hiraoka, Buslaps, Honkimäki, Minami & Uwe, 2005; Hiraoka, Buslaps, Honkimäki, Itou et al., 2005).
Although the new spectrometer already shows high performance, there is still room for improvements. The first point is the cooling system of the monochromator. Most of the performance tests were made in the 16 bunch mode, where the average ring current was ∼60 mA. We also tried some tests in the normal mode, where the average ring current was 190–180 mA, but the . An effective improvement for the cooling system does not only lead to an improved but may also provide a smaller focal size (theoretically, 30 µm is possible). This would make higher resolution possible without loss of beam intensity.
did not increase linearly with the ring current, and the gain was 2.5 at most. Furthermore, the intensity distribution of the beam on a sample separates into several peaks, unlike the broad single peak in Fig. 2The second point is the scattering geometry. At present, the spectrometer has the so-called horizontal scattering geometry, where the analyser is placed at a scattering angle of 172° in the horizontal plane. In the vertical scattering geometry, where the analyser is below the incident beam, the resolution is independent of the 2RuO4, should be eliminated. The project to modify the scattering geometry has already started.
in the sample and of the orientation and/or the roughness of the irradiated surface. Therefore, the deterioration of the resolution, seen in the data of SrFootnotes
‡Present address: Taiwan Beamline, SPring-8, 1-1-1 Kouto, Mikazuki, Sayo, Hyogo 679-5198, Japan.
Acknowledgements
We would like to thank T. Martin (ESRF) for helping us with monitoring the deformation of the analyser crystal with a two-dimensional scintillator. The data of Sr2RuO4 were taken in collaboration with Y. Sakurai (SPring-8) and Y. Maeno (Kyoto University).
References
Berthold, A., Mourikis, S., Schmitz, J. R., Schülke, W. & Schulte-Schrepping, H. (1992). Nucl. Instrum. Methods, A317, 373–382. CrossRef CAS Google Scholar
Biggs, F., Mendelsohn, L. B. & Mann, J. B. (1975). Atom. Data Nucl. Data Tables, 16, 201–309. CrossRef CAS Google Scholar
Eisenberger, P. & Platzman, P. M. (1970). Phy. Rev. A, 2, 415–423. CrossRef Web of Science Google Scholar
Fajardo, P., Honkimäki, V., Buslaps, T. & Suortti, P. (1998). Nucl. Instrum. Methods, B134, 337–345. CrossRef Google Scholar
Hiraoka, N., Buslaps, T., Honkimäki, V., Guyot, H. & Schlenker, C. (2005). Phys. Rev. B, 71, 125417-1–125417-5. Google Scholar
Hiraoka, N., Buslaps, T., Honkimäki, V., Itou, M., Sakurai, Y., Mao, Z. Q. & Maeno, Y. (2005). In preparation. Google Scholar
Hiraoka, N., Buslaps, T., Honkimäki, V., Minami, H. & Uwe, H. (2005). Phys. Rev. B, 71, 205106-1–205106-6. Google Scholar
Hiraoka, N., Itou, M., Ohata, T., Mizumaki, M., Sakurai, Y. & Sakai, N. (2001). J. Synchrotron Rad. 8, 26–32. Web of Science CrossRef CAS IUCr Journals Google Scholar
Itou, M. & Sakurai, Y. (2004). AIP Conf. Proc. 705, 901–904. CrossRef Google Scholar
Loupias, G. & Petiau, J. (1980). J. Phys. 41, 265–271. CrossRef CAS Google Scholar
Papanicolaou, N. I., Bacalis, N. C. & Papaconstantopoulos, D. A. (1991). Handbook of Calculated Electron Momentum Distributions, Compton Profiles, and X-ray Form Factors of Elemental Solids, pp. 104–105. Boca Raton, FL: CRC Press. Google Scholar
Sakurai, Y., Ito, M., Urai, T., Tanaka, Y., Sakai, N., Iwazumi, T., Kawata, H., Ando, M. & Shiotani, N. (1992). Rev. Sci. Instrum. 63, 1190–1193. CrossRef CAS Web of Science Google Scholar
Schulze, C., Lienert, U., Hanfland, M., Lorenzen, M. & Zontone, F. (1998). J. Synchrotron Rad. 5, 77–81. Web of Science CrossRef CAS IUCr Journals Google Scholar
Shiotani, N., Sakai, N., Itoh, F., Sakurai, M., Kawata, H., Amemiya, Y. & Ando, M. (1989). Nucl. Instrum. Methods, A275, 447–452. CrossRef CAS Google Scholar
Suortti, P., Buslaps, T., DiMichiel, M., Honkimäki, V., Lienert, U., McCarthy, J. E., Merino, J. M. & Shukla, A. (2001). Nucl. Instrum. Methods, A467/468, 1541–1544. CrossRef Google Scholar
Suortti, P., Buslaps, T., Fajardo, P., Honkimäki, V., Kretzschmer, M., Lienert, U., McCarthy, J. E., Renier, M., Shukla, A., Tschentscher, Th. & Meinander, T. (1999). J. Synchrotron Rad. 6, 69–80. Web of Science CrossRef CAS IUCr Journals Google Scholar
Suortti, P., Lienert, U. & Schulze, C. (1997). AIP Conf. Proc. 389, 175–192. CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.