radiation damage
Dose dependence of radiation damage for protein crystals studied at various X-ray energies
aStructural Biology Group, Research and Utilization Division, Japan Synchrotron Radiation Research Institute, 1-1-1 Kouto, Sayo, Sayo-gun, Hyogo 679-5198, Japan, and bDivision of Synchrotron Radiation Instrumentation, RIKEN SPring-8 Center, Hyogo, Japan
*Correspondence e-mail: yamamoto@postman.riken.go.jp
Radiation damage to protein crystals is the most serious problem in obtaining accurate structures from protein crystallography. In order to examine the photon energy dependence of radiation damage, 12 to 15 data sets from each of nine tetragonal lysozyme crystals were collected at nine different X-ray energies (6.5, 7.1, 8.3, 9.9, 12.4, 16.5, 20.0, 24.8 and 33.0 keV) using beamline BL41XU at SPring-8. All results were compared on the basis of absorbed dose, expressed in Gray (Gy). Crystallographic statistics, such as the values of lattice constants, Rmerge and I/σ(I), for each data set degraded at all nine energies as the exposure time for each crystal increased. In all data sets, radiation damage was observed after the absorbed dose exceeded 106 Gy. However, from the point of view of crystallographic statistics normalized to the absorbed dose, no clear dependence on photon energy was observed in these results. Structural showed that the average B-factor for the last data set was larger than that for the first data set at all energies tested. However, no energy dependence of radiation damage on B-factor was found. Furthermore, disruption of disulfide bonds due to radiation damage was observed in electron density maps even at the highest photon energy (33 keV) used in this study. Therefore, these results suggest that radiation damage in the energy range investigated could be evaluated based on absorbed dose without energy dependence, and that it is important to minimize the absorbed dose in a crystal sample for obtaining an accurate protein structure.
Keywords: radiation damage; X-ray energy; dose dependence.
1. Introduction
High-brilliance X-ray beams from synchrotron radiation are one of the most powerful tools for modern protein crystallography. However, this technology often results in serious problems due to radiation damage. Although the mechanism of radiation damage in cryo-cooled protein crystals has not been characterized in detail, its effect is considered to be due to several interactions. When a protein crystal absorbs X-ray photons, direct interactions occur first, causing temperature increases and dissociation of the covalent bonds. As a result of these interactions, reactive electrons, in other words free radicals, are produced (O'Neill et al., 2002). These radicals excite other covalent bonds and water molecules. This excitation propagates in the crystal, finally disrupting not only covalent bonds but also intermolecular networks forming crystal lattices. As these events are caused by absorbed photons in the crystal, it is important to consider the number of photons and their energy when discussing radiation damage.
The cryo-crystallographic method is almost always utilized for diffraction experiments at the synchrotron beamline in order to reduce radiation damage (Garman & Schneider, 1997). This is effective in significantly reducing radiation damage at beamlines where the density is around 1012 photons s−1 mm−2, such as bending-magnet beamlines at SPring-8, and the damage occurs gradually during one data collection. However, at beamlines equipped with an insertion device as the light source, where the density reaches ∼1014 photons s−1 mm−2, a protein crystal would receive marked radiation damage before the collection of one data set was complete, even for measurements at 100 K (Garman & Owen, 2006). The density clearly differs between these two types of beamline. However, although the probability of an interaction between an atom and an absorbed photon in a crystal can essentially be defined by the this coefficient changes with photon energy. For these reasons the absorbed dose is widely used as the most basic factor for evaluating radiation damage (e.g. Sliz et al., 2003; Leiros et al., 2006). The absorbed dose is calculated with knowledge of a number of parameters derived from properties not only of the incident X-ray beam but also of the crystal sample. For a cryo-crystallographic measurement, the maximum absorbed dose per unit mass for the diffraction intensity to fall by half through radiation damage has been calculated by analogy with observations to be 2 × 107 Gy [1 Gray (Gy) = 1 J kg−1] (Henderson, 1990). The Henderson limit is a predicted value; some experimentally obtained limit values have recently been reported by Owen et al. (2006), which indicated that a value of 3 × 107 Gy would cause an overall intensity decrease of 30% and was appropriate; beyond this, the quality of biological information obtained from the structure was likely to be compromised.
As the X-ray exposure of a protein crystal increases, broadening of diffraction spots and decreases in diffracted intensity are often observed (Blake & Philips, 1962). Crystallographic statistics, for example Rmerge, and scaling B-factor also increase with radiation damage progression. Various techniques to try to avoid this phenomenon in protein crystallography have been designed and reported. For example, use of a helium gas cryostream lowers the sample temperature to around 35 K (Nakasako et al., 2001; Hanson et al., 2002; Teng & Moffat, 2002).
Utilization of a high X-ray energy as an alternative possible method for reducing radiation damage has been suggested, since damage is generally thought to decrease at higher X-ray energy as the absorption is lower. Over the past few decades a considerable number of studies have been conducted on the effect of photon energy on protein crystallography (Arndt, 1984; Gonzalez et al., 1994; Polikarpov, 1997; Polikarpov et al., 1997; Teplyakov et al., 1998; Mueller-Dieckmann et al., 2004, 2005; Djinović Carugo et al., 2005; Weiss et al., 2005). Some theoretical considerations on the effects of X-ray energy on diffraction intensity have been reported (Arndt, 1984; Polikarpov, 1997). Polikarpov et al. (1997) also postulated that an X-ray wavelength of 0.9 Å (∼13.8 keV) is short enough to suppress radiation damage in a protein crystal. However, following an experimental investigation, Weiss et al. (2005) found that there was no dependence on X-ray wavelength for radiation damage between 1.0 Å (∼12.4 keV) and 2.0 Å (∼6.2 keV). For the elucidation of dependence of radiation damage on photon energy it is important to experimentally examine the effects over a wide energy range. In the case of organic light-atom crystals, Müller et al. (2002) reported that radiation damage is reduced at around 0.6 Å (∼20.7 keV) to 1.0 Å (∼12.4 keV) on the basis of investigation using several energies between 0.45 Å (∼27.6 keV) and 1.5 Å (∼8.3 keV). In the study reported here, 12 to 15 data sets from the same lysozyme crystal per energy were collected sequentially, utilizing nine different X-ray energies (6.5, 7.1, 8.3, 9.9, 12.4, 16.5, 20.0, 24.8 and 33.0 keV). The effects of radiation damage were observed at all nine energies when the absorbed dose reached more than 106 Gy. However, no clear energy dependence was observed in the statistical parameters. Therefore, the effect of the damage appears to depend on the absorbed dose at all X-ray energies. Consistently, disruption of disulfide bonds was observed even at the highest energy (33.0 keV) used in this study. These findings are discussed in terms of the dependence of radiation damage on photon energy and absorbed dose.
2. Materials and methods
2.1. Sample preparation
Hen egg-white lysozyme crystals in the tetragonal form (P43212) were used as the sample for this study. Crystallization was performed by hanging-drop vapor diffusion at 293 K, pH 4.2 with 100 mM acetate buffer including 1.25 M NaCl. Crystals of approximate size 100 µm × 150 µm × 150 µm were selected for diffraction measurement. Paratone-N (Hampton Research) was selected as the cryoprotectant for measurements made in a 100 K nitrogen gas stream.
2.2. Beamline conditions
Diffraction experiments were performed at beamline BL41XU of SPring-8. At BL41XU an in-vacuum undulator (Kitamura, 2000) is installed as the light source, with a magnetic periodicity of 3.2 cm and a total length of 4.5 m. The magnetic gap can be adjusted from 9.6 to 50 mm. Energy ranges emitted by this undulator are from 6.3 to 18 keV for the first harmonic and from 19 to 54 keV for the third harmonic. The beam generated is monochromated using a rotated-inclined double-crystal monochromator, with a pair of Si(111) crystals. The first crystal has a pin-post structure for direct water cooling (Yamazaki et al., 1999) and the second is indirectly cooled by water. The monochromator provides X-rays of high over a wide energy range from 6.5 to 17.5 keV for the first harmonic and from 19 to 37.5 keV for the third harmonic. As the beamline 33 cm-long rhodium-coated bent mirror for vertical focusing is currently not in use, the beam is focused only in the horizontal direction by the 70 cm-long rhodium-coated bent mirror. When the energy range of the first harmonic is used, the glancing angle of this mirror is set to 3.5 mrad to eliminate higher harmonics. For utilization of X-rays from the third harmonic, the angle is altered to 1.5 mrad. The focused beam size with no cut-off at the sample position is ∼130 µm (horizontal) × ∼280 µm (vertical) (FWHM). The focused beam is shaped by two sets of quadrant slits installed on the diffractometer (see below). Energy calibration of the beamline has been performed by measuring the absorption edges of the metals, Ni K (8.3328 keV), Au LIII (11.9787 keV), Br K (13.4737 keV), Mo K (19.9995 keV) and Ag K (25.5140 keV). The at the sample position on BL41XU is highly stable. Changes in intensity for a beam size of 100 µm × 100 µm are less than 2% over 8 h (data not shown). Since all measurements at one energy were completed within 5 h, the effect of the intensity change is negligible.
2.3. and dose calculation
In order to compare the effect of radiation damage on the basis of absorbed dose, the dose value for each data set was estimated using the program RADDOSE (Murray et al., 2004). The incident necessary for estimating the absorbed dose was calculated from the photocurrent value measured directly with a PIN photodiode (S3590-09, HAMAMATSU Photonics) installed at the sample position (Fig. 1). The beam size at the sample position was measured by scanning with a 30 µm pinhole across the beam. Two sets of beam shaper slits were adjusted so that the beam size at the sample position would be 100 µm × 100 µm (FWHM). Fig. 2 shows the incident at the sample position as a function of X-ray energy for this beam size.
2.4. Experimental and analysis conditions
Table 1 shows the experimental conditions for the nine different photon energies. One data set comprises 180 images each with an oscillation angle of 1.0°. The detector was a Quantum 315 (Area Detector Systems Corporation) which is restricted to a minimum distance of 155 mm from the sample on beamline BL41XU. The exposure time was modified in the range from 0.3 to 5 s depending on the observed diffraction intensity. All 130 data sets were processed and scaled using HKL2000 (Otwinowski & Minor, 1997); the processing results for the first and last data sets at each energy are listed in Table 2. Average B-factors and scale factors were derived from the Wilson plot estimated using Truncate (French & Wilson, 1978). Using the first data set collected at 12.4 keV, which had the best quality of all data sets, a standard structure model was first developed. After a procedure using coordinates of the protein atoms included in Protein Data Bank deposition 2BLX (Nanao et al., 2005) up to 4 Å resolution, atomic positional and B-factor refinements were performed up to the maximum resolution (1.33 Å). and structural were carried out using AMoRe (Navaza, 1994) and REFMAC5 (Murshudov et al., 1997), respectively. Subsequently, 214 water molecules were automatically added to the model structure with the arp_waters feature (Lamzin & Wilson, 1993) from REFMAC5 in CCP4i. After additional positional/B-factor refinements using REFMAC5, the standard model structure was obtained.
|
|
This model was used to evaluate other data sets with different energies or exposure times. Utilizing the protein atoms of the standard structure model, rigid-body B-factor refinements to 2.33 Å, these water molecules were simply merged into the protein coordinates. The final structure for each energy was acquired after additional positional/B-factor refinements using REFMAC5. The maximum resolution of all data sets was cut off at 2.33 Å at the stage, which is the resolution limit of the worst diffracting data set collected at 6.5 keV. The for the first and last data sets are listed in Table 3 for four of the photon energies measured.
was performed to 4 Å resolution. Water molecules were excluded from the standard model and, after positional/
|
3. Results and discussion
In order to compare the dependence of radiation damage on X-ray energy in a protein crystal, it was necessary to conduct diffraction data collection over a wide range of X-ray energies. High-flux X-rays that can cause radiation damage in a relatively short time are suitable for carrying out repetitive measurement. As the light source of BL41XU at SPring-8 is an in-vacuum undulator, high
and a wide X-ray energy range (6.5–37.5 keV) are both available. Therefore, BL41XU is considered to be a suitable beamline for performing such comparisons of energy dependence.3.1. Processing results
Table 1 shows the experimental conditions at each X-ray energy. The detector distances were chosen based on the maximum resolution of diffraction patterns. However, the diffraction intensities at wide angles cannot be recorded at energies lower than 9.9 keV owing to the limitation on the minimum detector distance (155 mm). Exposure times at 33.0 and 24.8 keV were longer than at the other energies so as to obtain data with enough statistical accuracy, necessary owing to the phosphor efficiency of the CCD detector being optimized for 12.4 keV (∼1 Å) photons. Using different exposure times for the various energies caused no difficulties with the methodology of the experiment, since these studies were carried out on the basis of absorbed dose calculated using RADDOSE, which takes into account the exposure time.
Table 2 shows the processing results of the first and the last data sets at the nine energies investigated. Lattice constants of the first data sets are similar at all energies. For data sets measured at 12.4 keV and higher photon energies, resolution limits were chosen so that the values of Rmerge in the highest-resolution shell would not exceed 30%. Resolution limits for the 9.9 keV data sets and lower photon energies were determined by the detector edge because of the limitation on the closest detector distance. In both cases the resolution limit for each data set at the same X-ray energy was fixed on that for the first set to enable quantitative comparison of statistics. Almost all of these parameters deteriorated at all energies when the first and the last data set were compared. Considerable changes were observed for Rmerge and 〈I/σI〉 in the highest-resolution shell, indicating, as expected, that crystalline deterioration owing to radiation damage influences crystallographic statistics, especially at higher resolutions. Values of completeness at 24.8 and 12.8 keV, for which the change in Rmerge and 〈I/σI〉 is relatively large, were smaller than at other energies. Although re-processing of these outlier data sets was conducted, the reason for the smaller changes could not be determined. This behavior could be due to differences in the experimental conditions; for example, the volume of cryoprotectant around the crystal sample. Average B-factors in a Wilson plot increase during the progress of the measurement, obviously showing the effect of radiation damage. Increases of Wilson scale factors were also observed at all energies except at 33.0 keV; this outlier result observed at 33.0 keV is not understood.
3.2. Relationship between radiation damage and dose
The effect of radiation damage on each statistical parameter obtained from the diffraction data was examined. In Fig. 3, changes in lattice constants are compared over the nine energies. The parameter `Irradiated time' means the total exposure time per single data set, and one data point in Figs. 3(a1) and 3(a2) corresponds to one data set. Lattice constants increased along with irradiated time for all nine X-ray energies tested. Changes in lattice constants are plotted against the absorbed dose (Figs. 3b1 and 3b2). For all the data sets, radiation damage was observed after the absorbed dose exceeded 106 Gy. Curves plotted for all X-ray energies could almost be superimposed on one another (Figs. 3b1 and 3b2). Further comparisons were performed for changes in mosaicity values and are shown in Fig. 3(c). When plotted against absorbed dose, as for Figs. 3(b1) and 3(b2), all curves approximately overlapped. These findings indicate that the effect of radiation damage depends on absorbed dose in this X-ray energy range. However, for mosaicity values, large amounts of change were observed only at lower energies. This effect was not clearly observed in Figs. 3(b). In order to confirm whether this phenomenon was truly caused by energy dependence, values of relative mosaicity for the last data set were normalized using total absorbed dose. These normalized mosaicity values plotted against photon energy are shown in Fig. 3(d). Dependence of X-ray energy was not observed, whereas values varied from 1.9 to 0.6 between 33.0 and 6.5 keV when the value at 12.4 keV was considered to be 1.0. Thus, this result indicates that the large change in mosaicity at low energies could be explained based on increase in absorbed dose, not on energy dependence.
3.3. Structural effects of radiation damage
The interaction between absorbed photons and atoms in a crystal cause structural changes of amino acids in a protein (Burmeister, 2000; Ravelli & McSweeney, 2000; Weik et al., 2000). Disulfide bonds formed between sulfur atoms of two cysteine residues are easily influenced and disrupted by photon absorption. As the lysozyme molecule has four disulfide bonds, it is suitable for evaluating structural change caused by absorbed photons. In order to examine the effect due to radiation damage in an structural refinements were performed against all data sets. Typical results for the structures refined against the first and last data sets are summarized in Table 3. Average B-factors for all atoms also increase as well as Wilson B-factors. Relative changes in average B-factors for all atoms at each energy are shown in Fig. 4(a). The amount of change at low X-ray energy was larger than that at high energy. This result was also observed in Fig. 3(c). However, when the values for the last data set were normalized to total absorbed dose, the change of B-factor also had no X-ray energy dependence (Fig. 4b). These findings indicate that for structural it is important to evaluate radiation damage based on absorbed dose. The effect of radiation damage was not clearly seen in the values for crystallographic R-factor and Rfree. Although the difference between R-factor and Rfree became large, this might be reduced if further model buildings were performed.
Fig. 5 shows the electron density maps for the last data set at 8.3, 12.4 and 33.0 keV, obtained at 2.33 Å resolution as representative data. These maps are superimposed on two of four disulfide bonds in lysozyme, Cys76/Cys94 or Cys6/Cys127, respectively. Although the level of radiation damage is different for each bond and between energies, breakage of the disulfide bonds due to radiation damage was apparent in the even at the highest energy tested, 33.0 keV. Identical breakages were also observed not only for the other two disulfide bonds but also at all energies tested (data not shown). These results indicate that utilization of high-energy X-rays could not produce data without radiation damage. Therefore, for obtaining an accurate structure, it is essential to find the best measurement conditions to minimize the absorbed dose.
4. Conclusions and future investigations
A systematic study of radiation damage to lysozyme crystals has been conducted for various X-ray energies at beamline BL41XU at SPring-8. In this study an X-ray energy dependence was clearly not observed and changes in crystallographic statistics, such as mosaicity and B-factor, were dependent on the absorbed dose. The possible X-ray energy dependence of radiation damage has already been investigated based on theoretical and experimental approaches (Arndt, 1984; Gonzalez et al., 1994; Polikarpov, 1997; Polikarpov et al., 1997; Teplyakov et al., 1998; Mueller-Dieckmann et al., 2005; Weiss et al., 2005). In these reports, several parameters (including absorbed dose, scattering coefficient and crystal size) were considered for evaluation of energy dependence. As the absorbed dose is a physical quantity fundamentally including those parameters, the validity of using it to estimate the radiation damage has been demonstrated in the present study. However, it is controversial to evaluate the energy dependence of the detector efficiency. Although CCD detectors are generally used at beamlines in synchrotron facilities, their efficiency at high X-ray energy, such as 33.0 keV, is not well defined. In this study the exposure time for detecting an accurate diffraction signal at higher X-ray energy was necessarily long. From the point of view of minimizing the absorbed dose, it might be difficult to directly compare two sets of high-energy data (24.8 and 33.0 keV) due to insufficient and detector sensitivity. If both the beamline optics and the detector could be optimized for high energy, energy dependence of radiation damage could be investigated in more detail.
Utilization of high energy is not an effective approach to reducing radiation damage under similar experimental environments (same kind of detector and optics) as that used at lower energy. Breakage of disulfide bonds owing to radiation damage was observed even at the highest energy tested in this study, 33.0 keV. Therefore, minimizing the absorbed dose of crystals for diffraction data collection is extremely important using any X-ray energy suitable for obtaining an accurate structure.
Acknowledgements
We would like to thank Dr E. F. Garman (Oxford University) for valuable advice and discussions. We are also grateful to all members of both the Structural Biology Group at JASRI/SPring-8 and the Synchrotron Radiation Instrumentation Group at RIKEN SPring-8 Center.
References
Arndt, U. W. (1984). J. Appl. Cryst. 17, 118–119. CrossRef CAS Web of Science IUCr Journals Google Scholar
Blake, C. C. F. & Philips, D. C. (1962). Biological Effects of Ionizing Radiation at the Molecular Level, pp. 183–191. Vienna: International Atomic Energy Agency. Google Scholar
Burmeister, W. P. (2000). Acta Cryst. D56, 328–341. Web of Science CrossRef CAS IUCr Journals Google Scholar
Djinović Carugo, K., Helliwell, J. R., Stuhrmann, H. & Weiss, M. S. (2005). J. Synchrotron Rad. 12, 410–419. Web of Science CrossRef IUCr Journals Google Scholar
French, G. S. & Wilson, K. S. (1978). Acta Cryst. A34, 517. CrossRef IUCr Journals Web of Science Google Scholar
Garman, E. F. & Owen, R. L. (2006). Acta Cryst. D62, 32–47. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. & Schneider, T. R. (1997). J. Appl. Cryst. 30, 211–237. CrossRef Web of Science IUCr Journals Google Scholar
Gonzalez, A., Denny, R. & Nave, C. (1994). Acta Cryst. D50, 276–282. CrossRef CAS Web of Science IUCr Journals Google Scholar
Hanson, B. L., Harp, J. M., Kirschbaum, K., Schall, C. A., DeWitt, K., Howard, A., Pinkerton, A. A. & Bunick, G. J. (2002). J. Synchrotron Rad. 9, 375–381. CrossRef CAS IUCr Journals Google Scholar
Henderson, R. (1990). Proc. R. Soc. London Ser. B, 241, 6–8. CrossRef CAS Web of Science Google Scholar
Kitamura, H. (2000). J. Synchrotron Rad. 7, 121–130. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lamzin, V. S. & Wilson, K. S. (1993). Acta Cryst. D49, 129–147. CrossRef CAS Web of Science IUCr Journals Google Scholar
Leiros, H.-K. S., Timmins, J., Ravelli, R. B. G. & McSweeney, S. M. (2006). Acta Cryst. D62, 125–132. Web of Science CrossRef CAS IUCr Journals Google Scholar
Mueller-Dieckmann, C., Panjikar, S., Tucker, P. A. & Weiss, M. S. (2005). Acta Cryst. D61, 1263–1272. Web of Science CrossRef CAS IUCr Journals Google Scholar
Mueller-Dieckmann, C., Polentarutti, M., Djinović Carugo, K., Panjikar, S., Tucker, P. A. & Weiss, M. S. (2004). Acta Cryst. D60, 28–38. Web of Science CrossRef CAS IUCr Journals Google Scholar
Müller, R., Weckert, E., Zellner, J. & Drakopoulos, M. (2002). J. Synchrotron Rad. 9, 368–374. Web of Science CrossRef IUCr Journals Google Scholar
Murray, J. W., Garman, E. F. & Ravelli, R. B. G. (2004). J. Appl. Cryst. 37, 513–522. Web of Science CrossRef CAS IUCr Journals Google Scholar
Murshudov, G. N., Vagin, A. A. & Dodson, E. J. (1997). Acta Cryst. D53, 240–255. CrossRef CAS Web of Science IUCr Journals Google Scholar
Nakasako, M., Sawano, M. & Kawamoto, M. (2001). Rigaku J. pp. 47–53. Google Scholar
Nanao, M. H., Sheldrick, G. M. & Ravelli, R. B. G. (2005). Acta Cryst. D61, 1227–1237. Web of Science CrossRef CAS IUCr Journals Google Scholar
Navaza, J. (1994). Acta Cryst. A50, 157–163. CrossRef CAS Web of Science IUCr Journals Google Scholar
O'Neill, P., Stevens, D. L. & Garman, E. F. (2002). J. Synchrotron Rad. 9, 329–332. Web of Science CrossRef CAS IUCr Journals Google Scholar
Otwinowski, Z. & Minor, W. (1997). Methods Enzymol. 276, 307–326. CrossRef CAS Web of Science Google Scholar
Owen, R. L., Rudiño-Piñera, E. & Garman, E. F. (2006). Proc. Natl. Acad. Sci. USA, 103, 4912–4917. Web of Science CrossRef PubMed CAS Google Scholar
Polikarpov, I. (1997). J. Synchrotron Rad. 4, 17–20. CrossRef CAS Web of Science IUCr Journals Google Scholar
Polikarpov, I., Teplyakov, A. & Oliva, G. (1997). Acta Cryst. D53, 734–737. CrossRef CAS Web of Science IUCr Journals Google Scholar
Ravelli, R. B. G. & McSweeney, S. M. (2000), Structure, 8, 315–328. Web of Science CrossRef PubMed CAS Google Scholar
Sliz, P., Harrison, S. C. & Rosenbaum, G. (2003). Structure, 11, 13–19. Web of Science CrossRef PubMed CAS Google Scholar
Teng, T.-Y. & Moffat, K. (2002). J. Synchrotron Rad. 9, 198–201. Web of Science CrossRef CAS IUCr Journals Google Scholar
Teplyakov, A., Oliva, G. & Polikarpov, I. (1998). Acta Cryst. D54, 610–614. Web of Science CrossRef CAS IUCr Journals Google Scholar
Weik, M., Ravelli, R. B. G., Kryger, G., McSweeney, S. M., Raves, M. L., Harel, M., Gros, P., Silman, I., Kroon, J. & Sussman, J. L. (2000). Proc. Natl. Acad. Sci. USA, 97, 623–628. Web of Science CrossRef PubMed CAS Google Scholar
Weiss, M. S., Panjikar, S., Mueller-Dieckmann, C. & Tucker, P. A. (2005). J. Synchrotron Rad. 12, 304–309. Web of Science CrossRef CAS IUCr Journals Google Scholar
Yamazaki, H., Kimura, H., Kagaya, I., Yamashita, C. & Ishikawa, T. (1999). Proc. SPIE, 3773, 21–29. CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.