Figure 1
Schematic representation of the basic beamline concept of the DELTA beamline BL8. The first mirror M1 collimates the wiggler beam vertically to a double-crystal monochromator with three different pairs of monochromator crystals that can be exchanged by remote control. The second mirror M2 refocuses the beam onto the sample on the diffractometer endstation. The third mirror system M3 provides harmonic rejection for low-energy photon beams (E < 4 keV). The transfer line between the second optics hutch and the experimental hutch is not shown in this figure. |