beamlines
beamline B23 at the Diamond Light Source
aDiamond Light Source, Chilton, Didcot, Oxfordshire OX11 0DE, UK
*Correspondence e-mail: rohanah.hussain@diamond.ac.uk, giuliano.siligardi@diamond.ac.uk
Synchrotron radiation
(SRCD) is a well established technique in structural biology. The first UV-VIS beamline, dedicated to at Diamond Light Source Ltd, a third-generation synchrotron facility in south Oxfordshire, UK, has recently become operational and it is now available for the user community. Herein the main characteristics of the B23 SRCD beamline, the ancillary facilities available for users, and some of the recent advances achieved are summarized.Keywords: synchrotron radiation circular dichroism (SRCD); vacuum-ultraviolet (VUV); secondary structure; protein; denaturation; capillary.
1. Introduction
; Berova et al., 2000). Unfortunately the usefulness of conventional CD instruments is limited by several factors: (i) the low especially in the far-UV region which gives limited use in sample stability and binding differentiation studies (Hussain et al., 2012); (ii) a larger beam spot size at the sample position, making it impossible to perform measurements in small-pathlength microcapillary cells unless a focusing lens is applied after the beam polarizer that can potentially introduce artifacts; and (iii) a divergent light beam that does not allow measurement in long-pathlength and, at the same time, low-volume (<800 µl) cells. The development of the B23 SRCD beamline at Diamond, making use of the very intense light of the 3 GeV synchrotron source, can help us overcome these limitations. The beamline has been designed to produce a highly collimated beam which not only extends the utility of the technique down to the vacuum-UV (VUV; <190 nm) region (Sutherland et al., 1980) but also allows measurements in very low volume capillary cells without the need of additional optical elements. This feature can also be exploited by measuring very dilute samples in long-pathlength cells which require less than a millilitre of sample. Contrary to B23, most of the other SRCD beamlines around the world, in order to reduce stray light, employ solar-blind (PMT) detectors, which reduce the available wavelength range to about 140–320 nm, excluding a good part of the near-UV (and the visible) regions. At B23, because of the double-grating subtractive monochromators, stray light is greatly reduced making it possible to reach a much wider spectral range (see Table 1 for details). In most SRCD beamlines the beam spot size at the sample position is also considerably larger than that of B23, making the density significantly poorer (see Table 2).
(CD) is a versatile spectroscopic technique which is often used to obtain low-resolution structural information about a wide variety of chiral materials in solution such as small molecules, proteins and (bio)polymers (Fasman, 1996
|
|
2. Beamline overview
The B23 SRCD beamline is based on one of the 1.4 T dipole bending magnets (BMs) of the synchrotron storage ring (Hussain et al., 2008). The light emerging from the BM has an aperture of 9 mrad (vertical) × 20 mrad (horizontal). The beamline can be divided into two parts: the front-end, inside the storage ring tunnel, and the experimental hutch, outside the shielding wall (Fig. 1). The front-end contains the beamline optics: two silicon mirrors which reflect and also focus the light onto the slit of a monochromator in the experimental hutch. The first mirror (M1), a water-cooled plane mirror, oriented at a 15° grazing angle, reflects the light up onto a toroidal focusing mirror (M2), oriented at a −15° grazing angle. M1 is further protected by a water-cooled copper finger (8 mm diameter × 23 cm long), which absorbs the hard X-ray component of the synchrotron radiation. The distance between the source and M2 as well as that of M2 and the entrance slit of module A (or that of module B) is 11.6 m.
In the experimental area there are two CD spectrographs (module A and module B) and a small silicon plane mirror (M3), the position of which determines the amount of light entering into each of the modules. The components of both CD modules are the same: a MgCl2 Rochon polarizing prism followed by a photo-elastic modulator (PEM) made of CaF2 for module A and fused silica for module B, sample chamber and two PMT detectors (Electrontubes, UK, model 9407B with MgF2 window for module A and model 9924QB for module B). Both modules can operate either in a single or double beam mode. The Rochon prism splits the incident light into two linearly polarized beams that are both modulated by the PEM to generate the left and right circularly polarized light at 50 kHz frequency. With both modules the CD is measured by the direct subtracting method, which was developed by On-Line Instruments Systems (Olis, USA). This method is based on the measurement of the intensities for each circular polarization (left and right) component by using fast analogue-to-digital converters. The data are then processed by a computer and the CD is calculated as the difference between the measurement for the left and right circularly polarized light. Contrary to the conventional approach, which is based on a lock-in amplifier system and still applied in many bench-top instruments and other synchrotron beamlines, the direct subtraction method means that the result is absolute so there is no need for calibrating a proportionality constant, and even large CD signals are measured correctly (Johnson, 1996). The two CD modules target slightly different applications: module A, having an ultra-high-vacuum (UHV) monochromator (RMP srl, Italy) with the first grating made of silicon and the second made of MgF2-coated aluminium (1200 lines mm−1, JY Horiba) and a CaF2 window behind the exit slit, is designed to measure solid samples (films) down to about 140 nm, while module B, with a monochromator employing MgF2-coated aluminium gratings, under nitrogen gas atmosphere and a sapphire vacuum window in front of its entrance slit, is suitable for liquid samples down to 165 nm. The choice for the different vacuum window materials was based on their UV transparency and resistance to UV damage. For module A, the vacuum window made of CaF2 is transparent down to 120 nm but is susceptible to colour centre formation. As the window is after the exit slit of the UHV monochromator, the monochromator filters out the wavelengths below 140 nm protecting the CaF2 window from UV damage (colour centre formation). For module B, the vacuum window made of sapphire with crystal orientation along the c-axis (0001) to substantially minimize the crystal birefringence, and UV transparent down to 160 nm, is much less susceptible to colour centres formation induced by the synchrotron light beam than any other material tested (LiF, MgF2, CaF2, SrF2, BaF2 and Suprasil fused silica) and can withstand the synchrotron `white' beam for longer. To increase the window life time the sapphire window is attached to an UHV XY stage coupled with a shutter that is opened only during measurements. When the window irradiated area becomes too dark, and hence less transparent (see Fig. 2), a clear area can be selected. In this manner the window is expected to last up to 2.5–3 years of beamline operation.
3. Ancillary facilities
The control and data acquisition of both modules is achieved by using the Olis SpectralWorks program suite. The advantage of this program is that it also contains modules for various data processing tasks such as secondary or analysis of kinetic traces based on different reaction schemes. It also allows the integration of a script file for sample auto-measurements such as variable temperature and kinetics assay measurements. The program also controls the external shutters in the vacuum pipe: the shutters are opened during data collection and closed when the measurement is finished, hence protecting the gratings and vacuum windows from UV damage.
A simple data converter with graphical user interface, developed in-house, is also available for the users to bring the collected data files into the format suitable for off-site data processing. The sample preparation area is mainly in the beamline experimental cabin and also in a peripheral laboratory which are equipped with a UV-VIS nanophotometer, pH meters, microbalances, Zeiss microscope, sonicator bath, concentrator and 15000 r.p.m. centrifuge for Eppendorf-type vials. Other requirements may be accommodated using other peripheral laboratories.
4. Facility access
The User Office is the first point of contact for all users. There are two calls for proposals per year, one with a deadline of 1 April and the second of 1 October. Access to B23 is available via direct access or programme access, for long-term research project proposals. Users can contact the beamline staff for initial discussion [https://www.diamond.ac.uk/Home/ForUsers.html] on the suitability of projects for submission. Proposals will be peer-reviewed by an external panel and all applicants notified accordingly of the outcome of the proposal review.
5. Highlights
Previously we have reported the ability of exploiting the microbeam by measuring CD spectra of protein in a capillary cell of 0.018 mm pathlength and 2 mm width (Jávorfi et al., 2010; Siligardi et al., 2010). To extend further the capability and functionality of the B23 collimated microbeam, attempts were made to measure low-volume diluted samples of nanomolar concentration. This was achieved by designing a 10 cm capillary cell with volume requirement of just under 800 µl. Fig. 3 shows a spectrum of 10 nM essentially fatty acid free human serum albumin (HSAff) using a 10 cm pathlength cell compared with that obtained with a 10 µM sample in a 0.02 cm-pathlength cylindrical cell. This cannot be achieved with existing bench-top instruments owing to the larger (in diameter) and more divergent beam. This could be especially important as in many cases obtaining suitable samples in sufficiently large quantities is either very labour intensive or dreadfully expensive.
6. Discussion and conclusions
In summary, the B23 end-stations have the capability to perform measurements from 165 to 650 nm for module B with variable beam spot size (0.25 mm2 to 4 mm2), depending on user requirements, or 155 nm to 500 nm for module A, which is more suitable for films and solid samples. A temperature-controlled sample environment is available on module B with the capability to run scripts for automated measurements. In the upgrade pipeline, we have a stopped-flow attachment for kinetic analysis and a high-throughput screening system based on a vertical sample chamber for 96 well plates that are in the commissioning phase.
References
Berova, N., Nakanishi, K. & Woody, R. W. (2000). Circular Dichroism Principles and Applications, 2nd ed. New York: Wiley. Google Scholar
Fasman, G. D. (1996). Circular Dichroism and Conformational Analysis of Biomolecules. New York: Plenum. Google Scholar
Hussain, R., Jávorfi, T. & Siligardi, G. (2012). Comprehensive Chirality, edited by G. Tranter. The Netherlands: Elsevier. In the press. Google Scholar
Hussain, R., Krumpa, N., Strachan, J., Clarke, D., Wagner, U., Macdonald, B., Reading, D., Cobb, T., Gillingham, I., Price, A., Smith, M. & Siligardi, G. (2008). Adv. Synchrotron Rad. 1, 265–270. CrossRef Google Scholar
Jávorfi, T., Hussain, R., Myatt, D. & Siligardi, G. (2010). Chirality, 22(Suppl. 1), E149–E153. Google Scholar
Johnson, W. C. Jr (1996). In Circular Dichroism and Conformational Analysis of Biomolecules, edited by G. D. Fasman. New York: Plenum. Google Scholar
Miles, A. J., Janes, R. W., Brown, A., Clarke, D. T., Sutherland, J. C., Tao, Y., Wallace, B. A. & Hoffmann, S. V. (2008). J. Synchrotron Rad. 15, 420–422. Web of Science CrossRef CAS IUCr Journals Google Scholar
Siligardi, G., Hussain, R., Myatt, D. & Jávorfi, T. (2010). Diamond Light Source Proc. 1(Issue SRMS-7), e104. Google Scholar
Sutherland, J. C., Desmond, E. J. & Takacs, P. Z. (1980). Nucl. Instrum. Methods, 172, 195–199. CrossRef Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.