Figure 9
Top: motor position versus reconstructed position giving an indication of the reconstruction error. A 20 µm segment spanning two strips, with 0 µm and 20 µm corresponding to adjacent strip centres, is shown. Blurring of the line is a maximum at the strip centres. Bottom: quantifying the reconstruction error. This is approximately equivalent to the detector PSF for which the σx of the experimental profile is measured to be 1.8 µm. The simulated PSF for a 2 µm-wide square beam with ENC = 340 e− and a coupling of K = 0.06 is also shown. |