research papers
Development of a single-shot CCD-based data acquisition system for time-resolved
at an X-ray free-electron laser facilityaResearch Infrastructure Group, RIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148, Japan, bExcitation Order Research Team, RIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148, Japan, cHarima Branch, Institute for Solid State Physics, The University of Tokyo, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5198, Japan, dInstitute for Solid State Physics, The University of Tokyo, Kashiwanoha 5-1-5, Kashiwa, Chiba 277-8581, Japan, eUVSOR, Institute for Molecular Science, Myodaiji, Okazaki, Aichi 444-8585, Japan, fResearch Laboratory for Surface Science, Okayama University, Okayama 700-8530, Japan, gGraduate School of Engineering Science, Osaka University, Toyonaka, Osaka 560-8531, Japan, hXFEL, JASRI, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5198, Japan, iXFEL Research and Development Division, RIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148, Japan, and jRIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148, Japan
*Correspondence e-mail: oura@spring8.or.jp
In order to utilize high-brilliance photon sources, such as X-ray free-electron lasers (XFELs), for advanced time-resolved f photoelectrons with intentionally controlled The system has also been applied to hard X-ray PES (HAXPES) in `ordinary sweep' mode as well as shot-to-shot image mode at the 27 m-long undulator beamline BL19LXU of SPring-8 and also at the SACLA XFEL facility. The XFEL-induced Ti 1s core-level spectrum of La-doped SrTiO3 is reported as a function of incident power density. The Ti 1s core-level spectrum obtained at low power density is consistent with the spectrum obtained using the synchrotron source. At high power densities the Ti 1s core-level spectra show space-charge effects which are analysed using a known mean-field model for ultrafast electron packet propagation. The results successfully confirm the capability of the present data acquisition system for carrying out the core-level HAXPES studies of condensed matter induced by the XFEL.
(TR-PES), a single-shot CCD-based data acquisition system combined with a high-resolution hemispherical electron energy analyzer has been developed. The system's design enables it to be controlled by an external trigger signal for single-shot pump–probe-type TR-PES. The basic performance of the system is demonstrated with an offline test, followed by online core-level photoelectron and in `single-shot image', `shot-to-shot image (image-to-image storage or block storage)' and `shot-to-shot sweep' modes at soft X-ray undulator beamline BL17SU of SPring-8. In the offline test the typical repetition rate for image-to-image storage mode has been confirmed to be about 15 Hz using a conventional pulse-generator. The function for correcting the shot-to-shot intensity fluctuations of the exciting photon beam, an important requirement for the TR-PES experiments at FEL sources, has been successfully tested at BL17SU by measuring Au 4Keywords: data acquisition system; shot-to-shot measurement; soft X-ray photoelectron spectroscopy; hard X-ray photoelectron spectroscopy; X-ray free-electron laser.
1. Introduction
With the advent of recent high-brilliance photon sources such as third-generation synchrotron radiation, high-harmonic generation (HHG) and free-electron laser (FEL) sources, time-resolved
(TR-PES) has become one of the fascinating spectroscopic techniques providing `real time' direct information on the ultrafast (picosecond and femtosecond) dynamics in electronic structure of condensed matter. Most of the TR-PES experiments performed up to the present rely upon a pump–probe measurement scheme which combines femtosecond optical and HHG (or FEL) laser pulses, where the temporal evolution of selected electronic states can be determined as a function of the delay between optical pump and HHG (or FEL) probe pulses.Early studies on TR-PES were mainly based on the utilization of a β-BaB2O2 crystal to produce a frequency-doubled UV pulse as HHG probe from a Ti:sapphire laser (Aeschlimann et al., 2000; Bauer & Aeschlimann, 2002; Perfetti et al., 2006, 2007, 2008; Schmitt et al., 2008; Azuma et al., 2009). The recent trend for sources in TR-PES has been to adopt the harmonic generation process using a rare gas (Lewenstein et al., 1994; Haight & Peale, 1994), which can give significantly high intensities even for much higher harmonics (Dakovski et al., 2010; Ishizaka et al., 2011; Dachraoui et al., 2011; Rohwer et al., 2011; Petersen et al., 2011). Hence, it provides us with an opportunity to study the photo-induced transient electronic structure of condensed matter even by means of core-level TR-PES, thereby allowing element-specific studies (Ishizaka et al., 2011; Dachraoui et al., 2011; Siffalovic et al., 2002). Nowadays, the HHG source can cover the energy range of the water window (Chen et al., 2010), but the resultant light is not suitable for spectroscopy due to limitations of tunability, number of photons per pulse and its broad bandwidth, although its time structure can be as narrow as 11 attoseconds full width at half-maximum (FWHM).
The emergence of FEL sources may lead to a breakthrough in these limitations for applying such femtosecond tunable light pulses to core-level TR-PES in materials science. Pietzsch et al. (2008) and Hellmann et al. (2010, 2012) have carried out pioneering work on TR-PES with monochromated FEL probe pulses at FLASH (Ackermann et al., 2007), but to date there has been no other report on photoemission spectroscopy for condensed matter using a FEL source. This is perhaps due to some difficulties concerning pulse-to-pulse intensity fluctuations and wavelength change inherent to the self-amplified (SASE) process. Furthermore, FEL-induced vacuum space-charge effects (Hellmann et al., 2012) cause an additional difficulty in performing TR-PES with a FEL probe pulse. Although the difficulty concerning wavelength change is reduced by adopting the monochromator, it is better that a data acquisition (DAQ) system with shot-to-shot measurement scheme is synchronized to the recording system for the corresponding FEL pulse properties. In previous studies (Pietzsch et al., 2008; Hellmann et al., 2010, 2012) photoemission spectra were acquired using a shot-to-shot measurement system with a gated CCD camera synchronized to the FEL macropulse with repetition rate of 5 Hz. In order to realise the shot-to-shot measurement system synchronized with a FEL single pulse of 10 Hz repetition rate without counting-loss we have developed a single-shot DAQ system for TR-PES by using a CCD-based high-resolution hemispherical electron energy analyzer to advance spectroscopic studies on transient phenomena in condensed matter. The system is aimed at performing TR-PES with a high-brilliance FEL source operating at a low repetition rate and designed to be controlled by an external trigger signal, e.g. a master trigger signal from the accelerator.
The purpose of this paper is to describe the single-shot CCD-based DAQ system and demonstrate its feasibilities for the shot-to-shot measurement scheme with the aid of the characterizing experiments using soft and hard X-ray synchrotron radiation beams. The system has also been applied to hard et al., 2012).
(HAXPES) at an X-ray free-electron laser (XFEL) facility, SACLA (Ishikawa2. apparatus combined with a single-shot DAQ system
2.1. Hardware and software for the single-shot DAQ system
The single-shot DAQ system, which can be triggered by an external signal and is developed for the CCD-based hemispherical electron energy analyzer, consists of a multifunction DAQ device, e.g. National Instruments (NI) PCIe-6363 (National Instruments, 2012), and a conventional CCD camera (480 × 433 pixel arrays). The maximum readout speed of the present CCD camera is 70 frames s−1 with a FireWire (IEEE 1394) connection. The system can be used for acquiring an energy-analyzed photoelectron image in pump–probe-type TR-PES for materials science using the FEL source.
In the present system the DAQ device can work as an analog-to-digital converter (ADC), a digital-to-analog converter (DAC) and/or a counter. The sampling rate of ADC channels is 2 MS s−1 for one analog input. Output from the DAC channel is configured to be a transistor-transistor-logic (TTL) pulse for multipurpose use, and the counter channel can be used to monitor the voltage output from any electronics by combining with a voltage-to-frequency converter. The system can be used for single-shot measurement and is designed to proceed iteratively at 10 Hz without counting-loss in the shot-to-shot measurement scheme.
The software to control the NI PCIe-6363 is entirely coded using LabVIEW and is also configured in order to facilitate the external control of the electron analyzer, e.g. VG SCIENTA R4000-10kV, with the aid of the SESWrapper library (VG Scienta, 2008). In the software, `single-shot image', `shot-to-shot image' and `shot-to-shot sweep' modes are packaged. Actually, shot-to-shot image mode corresponds to an iteration of the single-shot image mode. The shot-to-shot sweep mode, on the other hand, additionally includes a sweep of the hemispherical analyzer. Two-dimensional image data of the energy-analyzed photoelectrons are sequentially acquired via the CCD camera, and the data are immediately transferred to the stacked memory and finally stored in mass storage. As for the data storage, there are two storage modes for shot-to-shot image mode. One is the image-to-image storage mode, and the other is the block storage mode. In the block storage mode the DAQ system accumulates consecutive images in the stacked memory and, when the number of images reaches a certain number, e.g. 1000, all the images in the stacked memory are stored in mass storage. During the data storage of the block storage mode, e.g. typically a few seconds, a new image accumulation is inhibited until the end of the storage. Besides the image data, various information, such as the intensity and wavelength of each exciting photon pulse, can be acquired simultaneously by using the NI PCIe-6363. Therefore pulse-to-pulse intensity fluctuations and wavelength changes inherent to the SASE process can be corrected. Offline analysis software is also prepared to analyze the enormous amount of image data in order to convert them to spectra.
2.2. Achieved performance of the single-shot DAQ system
We have measured the typical processing time of the single-shot DAQ system. The external trigger signal is generated by using a conventional pulse-generator. When the NI PCIe-6363 is triggered, the hardware starts the A/D conversions and counter, where the accumulation time for the counter has been set to be 20 ms. Simultaneously, a command to control the electronic shuttering of the CCD camera is issued. In the performance test the CCD camera has been set to acquire a still image. After a certain exposure time, typically 14 ms, image data from pixel arrays are read out and immediately transferred to the stacked memory on the PC, and the data are finally stored in mass storage. When the data transfer to the stacked memory is finished, the CCD and DAQ device can be cleared and the DAC channel outputs the TTL pulse instantly for multipurpose use. When the repetition rate of the external trigger is higher than the processing speed of the system, a new trigger signal is inhibited during the processes mentioned above and is treated as counting-loss.
In order to estimate the optimum frequency enabling the data acquisition without counting-loss, we have measured the relation between , we can recognize that the is gradually decreasing from 100% at 10 Hz to ∼94% at 15 Hz, and there is a sudden drop in the when the trigger rate exceeds 15 Hz. In the trigger rate region that gives rise to the sudden drop in the it is found that the block storage mode has an efficiency about 1.6 times larger than that of the image-to-image storage mode. In the lower half of Fig. 1, the measured processing time of the image-to-image storage mode is summarized. In this summary the processing times are shown in milliseconds relative to the reference time of the trigger input. From these measurements the total processing time to complete one cycle of the procedure for the image-to-image storage mode, i.e. the time difference between the reference trigger input and re-initialization of the DAQ device, is measured to be 55 ± 10 ms and an acceptable repetition rate of 15 Hz is confirmed to be realised, although the counting-loss is about 6%. At the moment the present system can be operated with a trigger rate of up to 15 Hz and data acquisition without counting-loss can be carried out at a 10 Hz repetition rate, but we are making efforts to further improve the performance of the system.
and trigger rate for the image-to-image storage and block storage modes by changing the trigger rate of the pulse-generator. As seen in Fig. 12.3. Photoelectron spectrometer
We have integrated the present DAQ system with the existing et al., 2005, 2007). The apparatus mainly consists of the hemispherical electron energy analyzer R4000-10kV (VG SCIENTA) mounted on an ultra-high-vacuum system. In addition to the analyzer, a sample manipulator with a motorized XYZΘ stage, a flow-type He cryostat for sample cooling, and two turbomolecular pumps are equipped on the main chamber. This analysis system together with load-lock and preparation chambers are mounted on position-adjustable stages. The vacuum of the main chamber is of the order of 10−8 Pa, and the lowest sample temperature that can be achieved is about 20 K.
apparatus (TakataThe single-shot DAQ system combined with the e.g. from UV to hard X-rays. Fig. 2 depicts a typical layout of the measurement system for TR-PES. As shown in Fig. 2, an external trigger signal from the photon source, e.g. the master trigger signal from the accelerator, is fed into the NI PCIe-6363 via SCB-68 and issues a command to start a procedure in single-shot image, shot-to-shot image or shot-to-shot sweep mode. An image of energy-analyzed photoelectrons is acquired through the CCD camera and immediately transferred to the stacked memory before being stored in mass storage. At the same time the acquired photoelectron image is displayed on the monitor. Some useful information, such as intensity of the photon beam and information about the wavelength of the exciting photon beam (if possible), can be acquired via the ADC channels or counter channel of the NI PCIe-6363.
apparatus, with the aid of an appropriate lens parameter of the analyzer, can be utilized for TR-PES with a variety of wavelengths of the exciting photons,3. Feasibility experiments using synchrotron radiation
In order to check the feasibility of the present single-shot DAQ system combined with the et al., 2007; Senba et al., 2007; Yabashi et al., 2001).
apparatus, we have carried out online test experiments at the soft/hard X-ray undulator beamlines of BL17SU and BL19LXU of SPring-8 (OhashiIn the case of test experiments at soft X-ray beamline BL17SU, the apparatus was installed downstream of the exit slit of the beamline optics. Target materials were chosen to be a thin Au layer vacuum-evaporated onto a copper plate and a La-doped SrTiO3 (STO) chip. The beam size on the target was roughly estimated to be 230 µm (H) × 200 µm (V). In between the apparatus and the exit-slit there is a beam diagnostic section, where the incoming intensity of the soft X-ray beam can be monitored via the ADC or counter channel of the NI PCIe-6363. The of the photon beam was fixed to be 600.2 or 798.5 eV for the Au sample and 598.7 eV for the STO sample, and the bandwidth of the photon beam was about 200 meV. The electron analyzer was set to the condition with a pass energy (Ep) of 200 eV and a slit of 0.5 mm. The emitted photoelectrons or Auger electrons were measured in single-shot image, shot-to-shot image and shot-to-shot sweep modes. The `X channel' of the CCD camera was suitably determined by using a standard sample and a well calibrated incident soft X-ray beam. This is a very important process for calibrating the scale in the shot-to-shot sweep mode. We have also checked the capability of the software for correcting the intensity fluctuations of the exciting photon beam by adjusting the width of the exit slit.
In the case of test experiments at hard X-ray beamline BL19LXU, the apparatus was installed in experimental hutch EH3. The experimental set-up was almost the same as that previously reported (Takata et al., 2007). A high-resolution, ΔE ≃ 60 meV, monochromatic X-ray beam of 7.941 keV was focused onto the STO target. The spot size at the sample position was about 35 µm (H) × 50 µm (V). The set-up of the electron analyzer was the same as for the measurement at BL17SU. The Ti 1s HAXPES and KLL spectra were measured in ordinary sweep mode as reference spectra.
Fig. 3 shows a typical example of image and projected spectra of Au 4f7/2,5/2 photoelectrons emitted by 798.5 eV photon irradiation. This image is one of the 201 consecutive images recorded in shot-to-shot sweep mode, in which the is scanned from 705 to 725 eV with 0.1 eV step size. The two-dimensional intensity map (upper-right panel) represents the energy-analyzed photoelectron image accumulated at 14 ms exposure time. The horizontal scale corresponds to the of the photoelectrons, and the vertical scale provides information about the spatial origin of the photoelectrons. Thus the `X channel' projected spectrum (lower panel) gives us the spectrum and the `Y channel' projection gives us the intensity profile as a function of position along the footprint of the synchrotron radiation beam on the sample. Although the image and spectra were recorded as a single-shot image for 14 ms exposure time, the storage ring was operated with the multi-bunch operational condition and thus the image was actually the sum of a large but fixed number of cumulative shots.
In Fig. 4 the upper half of the figure shows the source image data measured in the shot-to-shot sweep mode and the lower half represents the resultant spectrum of Au 4f7/2,5/2 photoelectrons. As mentioned above, 201 spectra were consecutively measured by sweeping the Each photoelectron image was projected onto the axis, and all the spectra were summed by adjusting the shift, which is inherent to the sweep mode, of energy with suitable calibration. The red solid curve represents the final spectrum summed with appropriate calibration, while the blue solid curve, on the other hand, depicts a typical example with inappropriate calibration.
We have measured the soft X-ray induced electron spectra from an STO target with different acquisition modes for comparison. Fig. 5 shows a comparison of the wide-range spectra of STO measured by means of ordinary sweep and shot-to-shot sweep modes. The wide-range spectrum measured in shot-to-shot sweep mode is composed of 2151 consecutive images. In the spectra, pronounced peaks of Sr 3d photoelectrons and also the broad structure of Ti LMM as well as O KVV Auger electrons are observed. The large difference in the statistics comes from the difference in the total exposure time of the CCD. It is shown, however, that the acquired spectra are essentially the same. This result confirms that shot-to-shot sweep mode works properly.
Fig. 6 presents results of a demonstration exhibiting the correction of intensity fluctuations of an exciting photon beam during a measurement for hours in shot-to-shot image mode. The intensity of the photon beam was intentionally changed by adjusting the width of the exit slit and monitored by measuring the photoion intensity at the beam diagnostic section. In Fig. 6, red open triangles show the integrated Au 4f photoelectrons, blue open circles represent the intensity of photoions, and black closed circles show the resultant Au 4f photoelectron intensity corrected by using the measured photoion intensity. A small hump seen in the corrected intensity in the region between 580 and 680 shot number is considered to be due to the saturation of the photoion intensity monitor. Figs. 7(a) and 7(b) show Au 4f intensity maps, before and after correction, as a function of shot number. As can be seen from these figures, the intensity correcting function works correctly. This indicates that shot-to-shot image or shot-to-shot sweep mode can be suitably applied to the measurement of photoelectrons emitted by an intensity fluctuating photon source, such as a SASE-FEL source.
4. Application of the system to HAXPES at an X-ray free-electron laser facility
As mentioned earlier, our motivation to develop the present system is to advance pump–probe-type TR-PES using low-repetition-rate and high-brilliance photon sources. Time-resolved hard e.g. surface insensitivity and large probing depth, compared with PES with UV or soft X-ray photon beam (Takata et al., 2005, 2007). Thus we have applied the present single-shot DAQ system to HAXPES at the SACLA XFEL facility (Ishikawa et al., 2012).
(TR-HAXPES) will be one of the targets in possible applications of the present system, since HAXPES has some advantages,The apparatus was installed in experimental hutch EH2 of BL3. The electron analyzer was set to the condition with Ep = 200 eV and a slit of 4 mm so as to increase the efficiency. The target material was La-doped STO and we measured the Ti 1s HAXPES spectra in shot-to-shot image and ordinary sweep modes using 8 keV monochromatic XFEL beam with ΔE ≃ 1 eV. Under these conditions the FWHM of the Ti 1s HAXPES peak can be theoretically estimated to be about 2.7 eV if we employ a natural width value of 0.8 eV (Krause, 1979) for the Ti 1s hole state. The typical intensity of the XFEL beam obtained from the double-crystal monochromator was about 3.85 µJ per pulse, i.e. 3 × 109 photons per pulse, with the spot size at the sample position of about 0.6 mm in diameter.
As shown in previous studies (Pietzsch et al., 2008; Hellmann et al., 2010, 2012), the photoelectron spectrum will be broadened and energy-shifted due to the FEL-induced space-charge effects. Although the statistics are rather poor because of the small at 8 keV, we have confirmed that the Ti 1s photoelectron spectra show significant tailing toward the lower side as the XFEL power increases, as shown in Fig. 8. This tailing is supposed to be due to the space-charge effects caused by preceding electron packets of Sr 2p photoelectrons as well as Ti KLL Auger electrons, which have higher kinetic energies than the Ti 1s photoelectron. The Ti 1s photoelectron packet is therefore pushed back due to the repulsive force created by these preceding electron packets. The broadening, on the other hand, is considered to be due to the sum of intra-packet and inter-packet Coulomb repulsive forces.
In order to determine the appropriate XFEL power for the HAXPES measurements with acceptable space-charge effects, we have carried out a simple analysis of all the spectra shown in Fig. 8 after subtracting Shirley's type background (Shirley, 1972) from each spectrum. The peak position of the Ti 1s core-level for each condition of XFEL power was determined by the least-squares fitting procedure by employing the asymmetric double sigmoidal function. The results are plotted together with an exponential fit (dashed curve) in the upper half of Fig. 9. On increasing the XFEL power, the peak centers shift to the higher side and the widths of the lower side become larger giving rise to significant tailing. As mentioned above, this tailing tendency may be attributed to space-charge effects in which the preceding electron packets repel back the Ti 1s photoelectron on the way to the analyzer. The peak width of the Ti 1s core-level, on the other hand, for each XFEL power was also determined by the fitting procedure as shown in the lower half of Fig. 9. As described above, the bare peak width of the Ti 1s core-level for the present experimental conditions can be theoretically calculated to be about 2.7 eV. Thus the measured width of 2.6 eV for the lowest XFEL power is consistent with the theoretical peak width. We have then applied the `mean-field model' (Siwick et al., 2002) of femtosecond electron packet propagation to estimate the energy broadening of the Ti 1s peak width. According to the criterion discussed in the previous work (Hellmann et al., 2009), the Ti 1s photoelectron packet can be considered as a quasi-two-dimensional disk in front of the sample surface. The mean-field model predicts a square-root dependence of the space-charge broadening on the number of electrons per pulse. Thus we estimate the number of emitted photoelectrons per pulse by using the following simple expressions,
where ρ is the density of STO (g cm−3), t is the thickness of the target (cm), φ is the number of photons per pulse, NA is Avogadro's number (mol−1), M is the molecular weight (g mol−1), ΔΩ is the solid angle subtended by the electron analyzer (sr), w is the % of Ti in STO, σnl is the Ti 1s (cm2) (Scofield, 1973), βnl is the energy-dependent asymmetry parameter, P2(cos θ) = (1/2)(3 cos2θ − 1), and θ is the angle between the polarization vector and photoelectron direction. Substituting the values for our experimental conditions as well as relevant information for the target materials in the above expression, we have estimated the space-charge broadening. For example, when the XFEL power is 1.0 µJ per pulse the resultant number of emitted photoelectrons per pulse can be calculated to be about 165, which gives rise to a space-charge-induced bandwidth of about 4 eV. Similarly, the peak width has been calculated with the corresponding XFEL power for each case shown in Fig. 8, and the results are shown as the shaded area in the lower half of Fig. 9. As can be seen in Fig. 9, the for lower XFEL power are reasonably explained by the mean-field model. Only for the highest XFEL power, on the other hand, does the measured peak width show a deviation from the model. This deviation might be due to the emergence of the influence of the inter-packet Coulomb repulsive force. In order to fully understand such femtosecond electron packet propagation dynamics, we probably need to consider a more rigorous simulation taking into account inter-packet Coulomb repulsion.
In order to avoid these space-charge effects, the intensity of the XFEL beam had to be reduced to 7.6% (∼0.29 µJ per pulse) of the monochromator output by inserting the Al filter. Under this condition we have tried to acquire the core-level photoelectron image in shot-to-shot image mode. Fig. 10 shows the integrated Ti 1s photoelectron image of 9000 shots and its projections onto the `X channel (or energy channel)' and `Y channel (or position channel)'. In the upper-right of the figure the histogram showing the number of detector hits per shot is also indicated. According to this histogram we found that about half of the total shots did not contain any measured photoelectrons. The number of photoelectrons for each shot is quite a lot less, i.e. the averaged detector hits for 9000 shots is only about 1.2, so that we have measured all the spectra induced by the XFEL beam by means of ordinary sweep mode with long exposure time so as to smooth out intensity fluctuations. Under the same conditions, the Ti 1s HAXPES spectrum has been measured in ordinary sweep mode at SACLA and is compared with that measured in the same mode at BL19LXU of SPring-8. In Fig. 11 the spectra in the vicinity of the Ti 1s core-level are represented for comparison. The top panel indicates the spectrum measured at BL19LXU and the middle panel depicts the resultant spectrum measured at SACLA. Although the statistics of the HAXPES spectrum of SACLA are smaller, it definitely shows a nice spectrum which was measured in ∼1 h. In the bottom panel the Ti 1s HAXPES spectrum measured in shot-to-shot image mode is also shown for comparison. As can be seen from Fig. 11, the Ti 1s HAXPES spectra measured in ordinary sweep mode, i.e. both of the spectra shown in the top and middle panels, reflect the total instrumental energy resolution including the X-ray bandwidth. In the case of the Ti 1s HAXPES spectrum measured in shot-to-shot image mode, the analyzer was operated in fixed mode so that the resultant spectral width becomes larger compared with the case of the energy-sweep mode. These results confirm the capability for core-level HAXPES spectra of condensed matter induced by the XFEL beam.
Finally, with the aid of the fact that the KLL spectrum of STO by using the non-monochromatic XFEL beam, the so-called pink-beam, centered at 8 keV with an intensity of 250 µJ per pulse. Fig. 12 represents the Ti KLL spectra measured at SACLA (red curve), with that obtained at BL19LXU (blue curve) for comparison. The accumulation time was 500 ms per step and the spectrum at SACLA is just the result of one sweep. The power of the XFEL pink-beam was reduced to be 2%, i.e. 5 µJ per pulse. We have found that the spectrum is significantly distorted due to the space-charge effects so that we need to reduce the intensity of the XFEL beam by about two orders of magnitude. We come to the fact that spectroscopic studies of condensed matter with core-level using XFELs is achievable. However, if we try to perform the time-resolved measurement, especially for the temporal evolution measurement of the transient electronic state by means of pump–probe-type it is rather time-consuming.
spectrum can be acquired using white beam, we have tried to measure the Ti5. Perspective
For the TR-HAXPES using the shot-to-shot measurement scheme, the repetition rate of SACLA is suitable for the present DAQ system. At the moment SACLA is operated with a 20 Hz repetition rate, so that the shot-to-shot image or shot-to-shot sweep mode of the system is applicable by combining with the pulse selector (Kudo et al., 2009). The minimum exposure time of the present CCD camera is 14 ms, long enough to acquire all the electrons in the single shot. The of the XFEL pulse from SACLA is measured to range between 4.5 and 31 fs (Inubushi et al., 2012). The ultimate resolution of the time-resolved pump–probe measurement will be around several tens of femtoseconds, if we can tune the optical pump laser to have a comparable with the probe pulse. By tuning the delay between the optical pump and XFEL probe pulses, the temporal evolution of the intrinsic electronic structure of various materials induced by the pump laser will be probed by the XFEL pulse. We have recently carried out a TR-HAXPES experiment at SACLA using the pump–probe technique. The theoretical analyses of the experimental results are in progress and will be published elsewhere.
Here we have opened up the opportunity of studying the transient electronic structure of condensed matter on a sub-picosecond time scale by TR-HAXPES at the SACLA XFEL facility. Although the measurement of the
seems to be difficult because of the small the ultrafast core-level dynamics can be extensively studied using the present single-shot DAQ system combined with the high-resolution hemispherical electron energy analyzer. In the near future, when the SACLA is available for providing a soft X-ray FEL beam, the measurement of the transient electronic structure in the will be challenged.Furthermore, we are also planning to carry out TR-HAXPES at BL19LXU of SPring-8. Although the ultimate time resolution is restricted to 40 ps, which is the
of the synchrotron radiation beam from SPring-8, a higher repetition rate of 1 kHz with negligible space-charge effects would be applicable to the TR-HAXPES experiment to study the fast dynamics on a sub-nanosecond time scale.6. Summary
In conclusion, we have developed a single-shot CCD-based data acquisition system combined with a high-resolution hemispherical electron energy analyzer. It can be triggered by an external signal to carry out the shot-to-shot measurement for hard
(HAXPES) using a low-repetition-rate and high-brilliance photon source such as a free-electron laser source. The system has been successfully characterized with offline and online tests of feasibility experiments using soft/hard at high-brilliance photon source. The results of these characterizations satisfactorily demonstrate its performance of the shot-to-shot measurement scheme for near-future time-resolved HAXPES using XFEL sources.Acknowledgements
The synchrotron radiation experiments were performed at BL17SU and BL19LXU of SPring-8 with the approval of RIKEN (proposal Nos. 20120001, 20120038 and 20130036). The experiment at SACLA was performed with the approval of JASRI (proposal No. 2013A8050). This work is partially supported by MEXT program `X-ray Free-Electron Laser Priority Strategy Program'.
References
Ackermann, W. et al. (2007). Nat. Photon. 1, 336–342. Web of Science CrossRef Google Scholar
Aeschlimann, M., Bauer, M., Pawlik, S., Knorren, R., Bouzerar, G. & Bennemann, K. H. (2000). Appl. Phys. A, 71, 485–491. CrossRef CAS Google Scholar
Azuma, J., Tokudomi, S., Takahashi, K. & Kamada, M. (2009). Phys. Status Solidi C, 6, 307–310. CrossRef CAS Google Scholar
Bauer, M. & Aeschlimann, M. (2002). J. Electron Spectrosc. Relat. Phenom. 124, 225–243. Web of Science CrossRef CAS Google Scholar
Chen, M. C., Arpin, P., Popmintchev, T., Gerrity, M., Zhang, B., Seaberg, M., Popmintchev, D., Murnane, M. M. & Kapteyn, H. C. (2010). Phys. Rev. Lett. 105, 173901. Web of Science CrossRef PubMed Google Scholar
Dachraoui, H., Michelswirth, M., Siffalovic, P., Bartz, P., Schäfer, C., Schnatwinkel, B., Mattay, J., Pfeiffer, W., Drescher, M. & Heinzmann, U. (2011). Phys. Rev. Lett. 106, 107401. Web of Science CrossRef PubMed Google Scholar
Dakovski, G. L., Li, Y., Durakiewicz, T. & Rodriguez, G. (2010). Rev. Sci. Instrum. 81, 073108. CrossRef PubMed Google Scholar
Haight, R. & Peale, D. R. (1994). Rev. Sci. Instrum. 65, 1853–1857. CrossRef CAS Google Scholar
Hellmann, S., Beye, M., Sohrt, C., Rohwer, T., Sorgenfrei, F., Redlin, H., Kalläne, M., Marczynski-Bühlow, M., Hennies, F., Bauer, M., Föhlisch, A., Kipp, L., Wurth, W. & Rossnagel, K. (2010). Phys. Rev. Lett. 105, 187401. Web of Science CrossRef PubMed Google Scholar
Hellmann, S., Rossnagel, K., Marczynski-Bühlow, M. & Kipp, L. (2009). Phys. Rev. B, 79, 035402. Web of Science CrossRef Google Scholar
Hellmann, S., Sohrt, C., Beye, M., Rohwer, T., Sorgenfrei, F., Marczynski-Bühlow, M., Kalläne, M., Redlin, H., Hennies, F., Bauer, M., Föhlisch, A., Kipp, L., Wurth, W. & Rossnagel, K. (2012). New J. Phys. 14, 013062. Web of Science CrossRef Google Scholar
Inubushi, Y., Tono, K., Togashi, T., Sato, T., Hatsui, T., Kameshima, T., Togawa, K., Hara, T., Tanaka, T., Tanaka, H., Ishikawa, T. & Yabashi, M. (2012). Phys. Rev. Lett. 109, 144801. Web of Science CrossRef PubMed Google Scholar
Ishikawa, T. et al. (2012). Nat. Photon. 6, 540–544. Web of Science CrossRef CAS Google Scholar
Ishizaka, K., Kiss, T., Yamamoto, T., Ishida, Y., Saitoh, T., Matsunami, M., Eguchi, R., Ohtsuki, T., Kosuge, A., Kanai, T., Nohara, M., Takagi, H., Watanabe, S. & Shin, S. (2011). Phys. Rev. B, 83, 081104(R). CrossRef Google Scholar
Krause, M. O. (1979). J. Phys. Chem. Ref. Data, 8, 307. CrossRef Google Scholar
Kudo, T., Hirono, T., Nagasono, M. & Yabashi, M. (2009). Rev. Sci. Instrum. 80, 093301. Web of Science CrossRef PubMed Google Scholar
Lewenstein, M., Balcou, Ph., Ivanov, M. Yu., L'Huillier, A. & Corkum, P. B. (1994). Phys. Rev. A, 49, 2117–2132. CrossRef PubMed Google Scholar
National Instruments (2012). The National Instruments X Series multifunction data acquisition (DAQ) devices with a multithreaded NI-DAQmx driver software. Google Scholar
Ohashi, H. et al. (2007). AIP Conf. Proc. 879, 523–526. CrossRef CAS Google Scholar
Perfetti, L., Loukakos, P. A., Lisowski, M., Bovensiepen, U., Berger, H., Biermann, S., Cornaglia, P. S., Georges, A. & Wolf, M. (2006). Phys. Rev. Lett. 97, 067402. Web of Science CrossRef PubMed Google Scholar
Perfetti, L., Loukakos, P. A., Lisowski, M., Bovensiepen, U., Eisaki, H. & Wolf, M. (2007). Phys. Rev. Lett. 99, 197001. Web of Science CrossRef PubMed Google Scholar
Perfetti, L., Loukakos, P. A., Lisowski, M., Bovensiepen, U., Wolf, M., Berger, H., Biermann, S. & Georges, A. (2008). New J. Phys. 10, 053019. CrossRef Google Scholar
Petersen, J. C., Kaiser, S., Dean, N., Simoncig, A., Liu, H. Y., Cavalieri, A. L., Cacho, C., Turcu, I. C., Springate, E., Frassetto, F., Poletto, L., Dhesi, S. S., Berger, H. & Cavalleri, A. (2011). Phys. Rev. Lett. 107, 177402. Web of Science CrossRef PubMed Google Scholar
Pietzsch, A., Föhlisch, A., Beye, M., Deppe, M., Hennies, F., Nagasono, M., Suljoti, E., Wurth, W., Gahl, C., Döbrich, K. & Melnikov, A. (2008). New J. Phys. 10, 033004. CrossRef Google Scholar
Rohwer, T., Hellmann, S., Wiesenmayer, M., Sohrt, C., Stange, A., Slomski, B., Carr, A., Liu, Y., Avila, L. M., Kalläne, M., Mathias, S., Kipp, L., Rossnagel, K. & Bauer, M. (2011). Nature (London), 471, 490–493. Web of Science CrossRef CAS PubMed Google Scholar
Schmitt, F., Kirchmann, P. S., Bovensiepen, U., Moore, R. G., Rettig, L., Krenz, M., Chu, J. H., Ru, N., Perfetti, L., Lu, D. H., Wolf, M., Fisher, I. R. & Shen, Z. X. (2008). Science, 321, 1649–1652. Web of Science CrossRef PubMed CAS Google Scholar
Scofield, J. H. (1973). Report UCRL-51326. Lawrence Livermore Laboratory, CA, USA. Google Scholar
Senba, Y., Ohashi, H., Kishimoto, H., Miura, T., Goto, S., Shin, S., Shintake, T. & Ishikawa, T. (2007). AIP Conf. Proc. 879, 718–721. CrossRef CAS Google Scholar
Shirley, D. A. (1972). Phys. Rev. B, 5, 4709–4714. CrossRef Google Scholar
Siffalovic, P., Drescher, M. & Heinzmann, U. (2002). Europhys. Lett. 60, 924–930. Web of Science CrossRef CAS Google Scholar
Siwick, B. J., Dwyer, J. R., Jordan, R. E. & Miller, R. J. D. (2002). J. Appl. Phys. 92, 1643–1648. Web of Science CrossRef CAS Google Scholar
Takata, Y., Horiba, K., Matsunami, M., Shin, S., Yabashi, M., Tamasaku, K., Nishino, Y., Miwa, D., Ishikawa, T., Ikenaga, E., Kobayashi, K., Arita, M., Shimada, K., Namatame, H., Taniguchi, M., Nohira, H., Hattori, T., Södergren, S. & Wannberg, B. (2007). AIP Conf. Proc. 879, 1597–1602. CrossRef CAS Google Scholar
Takata, Y., Yabashi, M., Tamasaku, K., Nishino, Y., Miwa, D., Ishikawa, T., Ikenaga, E., Horiba, K., Shin, S., Arita, M., Shimada, K., Namatame, H., Taniguchi, M., Nohira, H., Hattori, T., Södergren, S., Wannberg, B. & Kobayashi, K. (2005). Nucl. Instrum. Methods Phys. Res. A, 547, 50–55. CrossRef CAS Google Scholar
VG Scienta (2008). The SESWrapper library for remote control of the Scienta spectrometer, https://www.vgscienta.com/ . Google Scholar
Yabashi, M., Mochizuki, T., Yamazaki, H., Goto, S., Ohashi, H., Takeshita, K., Matsushita, T., Tamasaku, K., Tanaka, Y. & Ishikawa, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 678–681. CrossRef CAS Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.