Characteristics of a tapered undulator for the X-ray absorption fine-structure technique at PLS-II
aPohang Accelerator Laboratory, Pohang University of Science and Technology, Pohang 790-834, Republic of Korea
*Correspondence e-mail: ijlee@postech.ac.kr
An in-vacuum undulator (IVU) with a tapered configuration was installed in the 8C nanoprobe/XAFS beamlime (BL8C) of the Pohang Light Source in Korea for hard X-ray nanoprobe and X-ray absorption fine-structure (XAFS) experiments. It has been operated in planar mode for the nanoprobe experiments, while gap-scan and tapered modes have been used alternatively for K-edge spectra were obtained in each mode. Owing to the large variation around this energy, spectra of the planar and gap-scan modes show considerable spectral distortions in comparison with the tapered mode. This indicates that the tapered mode, owing to the smooth profile at the expense of the highest and most stable intensity, can be an alternative for experiments where the gap-scan mode gives a considerable intensity variation; it is also suitable for quick-XAFS scanning.
experiments. To examine the features of the BL8C IVU for experiments, spectral distributions were obtained theoretically and experimentally as functions of the gap and gap taper. Beam profiles at a of the X-ray beam were acquired using a slit to visualize the intensity distributions which depend on the gap, degree of tapering and harmonic energies. To demonstrate the effect of tapering around the lower limit of the third-harmonic energy, VKeywords: undulator; high-order harmonic; X-ray absorption fine structure; gap-scan mode; tapered mode.
1. Introduction
Synchrotron radiation from an undulator source has high et al., 1989; Kim, 1986; Attwood et al., 1985; Boyanov et al., 1994), and is therefore suitable for applications such as X-ray diffraction or nanoprobe/microprobe experiments. The characteristic feature of an undulator is that the is concentrated in a fundamental and its harmonics by interference (Kim, 1986; Alferov et al., 1974; Kincaid, 1977). The 8C nanoprobe/X-ray absorption fine-structure (XAFS) beamline (BL8C) of the Pohang Light Source (PLS-II) in Korea was built for nanoprobe and experiments. The technique requires high and a spectral range as wide as 1 keV to obtain valuable spectra in regions of interest. In order to obtain a high insertion devices, such as wigglers or undulators, have generally been used as the photon source. A wiggler has been used for the beamline because of its wide and intense owing to its large deflection parameter, originating from the strong magnetic field (Welter, 2012). However, because a wiggler has a stronger magnetic field in comparison with an undulator, the emittance of the storage-ring electron beam is increased such that it is higher than that of an undulator (Sasaki et al., 1990). Therefore, an undulator was chosen as the photon source for BL8C in order to minimize the emittance of the PLS-II storage-ring electron beam and to have the required for nanoprobe experiments.
and narrow spectral range (ChapmanAlthough a planar undulator (planar mode) emits brilliant X-rays, the spectral width of each harmonic is too narrow to be used in et al., 1994; Walker, 1988; Mori et al., 2004; Barrea et al., 2005; Curtin et al., 1990; Hagelstein et al., 1997; Fischetti et al., 2004; Nonaka et al., 2012; Sekizawa et al., 2013; Inada et al., 2007; Maréchal, 1998), and (ii) scanning the gap of a planar undulator (gap-scan mode) producing a smooth profile of the X-rays with highest intensity (Boyanov et al., 1994; Mori et al., 2004; Barrea et al., 2005; Curtin et al., 1990; Hagelstein et al., 1997; Fischetti et al., 2004; Nonaka et al., 2012; Sekizawa et al., 2013; Inada et al., 2007; Maréchal, 1998; Shih & Caponi, 1982; Saitoh et al., 2012; Tanaka et al., 2001; Gauthier et al., 1999; Rogalev et al., 1998; Igarashi et al., 2013; Takahashi et al., 2013; Somogyi et al., 2013). The gap-scan mode can be applicable to the continuous-scan (Saitoh et al., 2012; Tanaka et al., 2001) or high-precision step-scan (Tanida & Ishii, 2001; Oyanagi et al., 2000) by changing the undulator gap synchronized with a double-crystal monochromator (DCM) angle. On the other hand, the tapered mode is used for time-resolved experiments (Nonaka et al., 2012; Tanaka et al., 2001). Because the tapered mode widens the spectrum at the expense of reducing the of the undulator, a reproducible gap switch to planar mode has to be realised for nanoprobe experiments. These modes can be applied alternatively if an undulator has a tapered configuration. To realise this, a tapered in-vacuum undulator (IVU) has been installed on BL8C and operated alternatively in tapered mode and gap-scan mode for experiments.
experiments. To alleviate this problem, two methods can be applied: (i) using an undulator with a gap tapering configuration whose magnetic field gradient is along the electron beam trajectory and so has a wide spectral range (tapered mode) (BoyanovIn this paper we present the theoretical and experimental
profiles of the incoming X-rays as a function of the gap and tapering. In addition, we discuss the effect of the IVU parameters (Table 1) on beam intensity distribution profiles and spectra in terms of gap tapering.2. Experimental procedure
The radiation source of BL8C (Fig. 1) of PLS-II is a tapered IVU. PLS-II has been operating at 3.0 GeV and ≥200 mA in top-up injection mode. Synchrotron radiation from the IVU is collimated by slits and focused by a Rh-coated horizontal focusing mirror (HFM). The X-ray beam is monochromated by a Si(111) DCM, and finally delivered to the sample through a secondary source aperture. At the sample position the beam diameter is ∼0.3 mm. A water-cooled HFM and liquid-nitrogen-cooled DCM are located at positions 20 m and 24 m from the IVU, respectively. The DCM was designed to be operated in the energy range 2.5–35.0 keV but the available energy range of the BL8C is limited to 4.0–22.0 keV due to insufficient The undulator, DCM and all devices are controlled using Experimental Physics and Industrial Control System (EPICS)/LabVIEW software.
When the system is operated in gap-scan mode, the gaps of the IVU are adjusted following formulae which were obtained by tracing the intensity maximums as a function of the gap and required energy. In contrast, in tapered mode, the downstream gap of the undulator was controlled in the range 0–1.2 mm with respect to the upstream gap which depends on the experimental energy range. To compare with experimental results, theoretical calculations for the tapered undulator were carried out using the XYaup code (Boyanov et al., 1994; Sanchez del Rio & Dejus, 1998, 2004).
Beam profiles with and without tapering at energies of interest were acquired utilizing a slit located between the radiation source and HFM (Fig. 1). The slit aperture was opened to 0.1 mm × 0.1 mm for high spatial resolution, and was raster-scanned at 0.1 mm intervals from −1.0 to 1.0 mm in the horizontal direction (x) and from −0.5 to 0.5 mm in the vertical direction (y), both with respect to the beam center. Other apertures downstream of the slit were opened enough so as not to screen the X-rays. Beam profiles and distribution profiles were obtained with a nitrogen-filled detector. spectra were recorded at the V K-edge in planar and tapered modes to demonstrate the effect of tapering on the spectrum quality in comparison with the results of the gap-scan mode. The sample was prepared with VO2 powder (Aldrich, 99.9%) spread thinly on a polyimide film. The incident and transmitted X-rays were monitored using two detectors. Because the Rh-coated HFM cannot remove contamination by high-order harmonics below 7.5 keV, was detuned by more than 30% to eliminate harmonic contamination. In order to reduce spectral noise, spectra were obtained using the step-scan method.
3. Results and discussion
The IVU (Table 1) is a hybrid and asymmetric type, of length 1.4 m with 20 mm period and 140 poles. The gap range of the IVU is 5.1–20.0 mm and the available X-ray energy range is 4.0–22.0 keV. The photon beam size at the source point is 12 µm × 187 µm (V × H) and the beam divergence is 16 µrad × 49 µrad (V × H). The IVU gaps and tapering can be controlled interactively by beamline users through EPICS-based software to adjust the magnetic field using two stepping motors. In the case of the gap-scan mode, because the intensity of harmonics decreases significantly with increasing gap size, several formulae of the IVU gap were made using high-order harmonics with a small gap to deliver intense X-rays for all available energy ranges. On the other hand, the tapered mode needs a proper gap and tapering, depending on the energy region, to cover an energy range of the spectrum. Beamline software recommends a proper gap formula and tapering for the gap-scan and tapered modes, respectively, depending on the operation mode.
|
We compared theoretical (Fig. 2a) and experimental (Fig. 2b) profiles of the planar mode in the energy range 4.0–20.0 keV. To show the dependency of spectral variations on undulator parameters, the gap was changed from 5.1 mm to 6.3 mm in 0.4 mm increments without tapering. Spectral features of the experimental spectra matched well with the theoretical calculations except at high energy. The first peaks correspond to the minimum energies of third-order harmonics with different gaps. Other peaks are high-order harmonics. Due to interference by the IVU radiation, peaks moved towards the high-energy side as the gap increased but the maximum intensities decreased gradually. Because the lower limit of the IVU gap is 5.1 mm (Table 1), the peak near 5.45 keV corresponds to the minimum energy of the third-harmonic X-rays. To obtain precise harmonic energies, the secondary source aperture (SSA) (Fig. 1) was located at the central spot of the radiation cone by finding the intensity maximum. Otherwise, the intensity maximum of a harmonic moved towards the lower energy side as a function of the vertical distance between the SSA and central spot of the radiation cone, which is contrary to the report of Guo et al. (1999). Theoretically, even-order harmonic radiation is not included in undulator spectra for an ideal undulation. However, this was observed in Fig. 2(a) with relatively low intensity, and is ascribed to the non-zero emittance of electrons in the storage ring (Walker, 1988).
We also compared theoretical (Fig. 3a) and experimental (Fig. 3b) profiles of the tapered mode in the energy range 4.0–20.0 keV. To show that spectral variations depend on the undulator parameters, the gap tapering was changed from 0 to 1.2 mm in 0.4 mm increments with IVU tapering. As the degree of tapering increased, intensities decreased significantly and the spectral widths of peaks increased to form plateau shapes. Moreover, intensity plateaus moved towards the higher energy side proportional to the energy, resulting in a considerable decrease of the intensities at the peak position of zero-tapering. For the same tapering, the spectral broadness was larger at high energy than at low energy. These results indicate that a larger tapering is needed at low energy for low-order harmonics in order to have a wide spectral range but lower X-ray intensity.
To observe the intensity distribution of the IVU radiation as functions of X-ray energy and IVU parameters, X-ray beam profiles of cross sections of X-rays were obtained utilizing a slit located between the IVU and HFM (Fig. 1). Profiles of the third harmonics (Fig. 4), normalized to their maximum intensity, were obtained while keeping the gap of the IVU at 5.1 mm in planar and tapered modes at 5.45 keV [Figs. 4(a) and 4(c)] and 5.60 keV [Fig. 4(b) and 4(d)], which correspond to energies of the first maximum and minimum intensity of the third harmonics, respectively. In the planar mode the profile at 5.45 keV (Fig. 4a) shows a peak at the center with an elliptical peaked top, whereas the profile at 5.60 keV (Fig. 4c) has a valley form. In the tapered mode, the profile at 5.45 keV (Fig. 4b) has a minimum at the center contrary to Fig. 4(a), whereas the profile at 5.60 keV (Fig. 4d) has a peak at the center with a gentle slope and high intensity. These features show that intensity variation is smaller in the tapered mode than in the planar mode.
Profiles of the fourth [Figs. 5(a) and 5(b)] and fifth [Figs. 5(c) and 5(d)] harmonics with 5.1 mm IVU gap were obtained in planar and tapered modes at the first peak positions of both harmonics. In the planar mode, peaks at the center of the profiles and absolute intensities of the odd-order harmonics [Figs. 4(a) and 5(c)] were sharper and higher than that of the even-order harmonics (Fig. 5a). In contrast, profiles of tapered modes [Figs. 4(b) and 5(b)] show bowl shapes due to the blue-shift of the intensity plateaus. The profile of the fifth harmonics in tapered mode (Fig. 5d) showed higher intensity at the center than other tapered mode profiles. These results are ascribed to the fact that the intensity plateau of the tapered fourth harmonics overlap the zero-tapered fifth-order harmonic peak (Fig. 3).
The best experimental condition for
experiments can be achieved using the gap-scan mode. However, because the minimum energy of the third harmonics is 5.45 keV, the spectrum in the range below 5.45 keV could be obtained in the tapered or gap-scan mode using the fundamental or second-harmonic X-rays, but the intensities of those harmonics are very low.To demonstrate the tapering effect clearly, V K-edge spectra of VO2 were obtained in transmission mode in the range 5.2–6.4 keV which includes the minimum energy of the third harmonics. Although the fundamental or second harmonics give higher intensities than the third harmonics below 5.45 keV, third-harmonic X-rays were used for the gap-scan and planar modes because they give more intense X-rays than the fundamental or second harmonics above this energy. Second-harmonic X-rays were used for the tapered mode. The X-ray intensities and spectra were measured at the V K-edge in three operation modes with a gap of 5.1 mm. Above the minimum energy of the third harmonics (5.45 keV), the (Fig. 6a) of the tapered mode was less than 10% of the gap-scan mode, and had intensity variations in the plateau range which were not observed clearly in the theoretical calculations. Below 5.45 keV, the of the gap-scan mode was the same as that of the planar mode.
To observe variations of the . Below 5.45 keV, the beam profile shows two peaks split vertically. These peaks were merged into a single peak at 5.45 keV with increased intensity. In contrast, as the X-ray energy increased, the peak was divided into two peaks horizontally with decreased intensity. The intensity profile was changed to a valley shape at the energy of the first intensity-minimum of the third-order harmonics. Beam profiles of below and above 5.45 keV indicate that lower- or higher-energy harmonic X-rays, in comparison with the minimum energy of the on-axis harmonics, can be obtained by selecting an off-axis peak. The energy differences between the on-axis and off-axis X-rays are given as functions of the distance between the peak and central spot of the radiation cone. Although off-axis odd-order harmonics can be applicable instead of low-intensity even-order harmonics where the energy region is below the minimum energy of the harmonics, its intensity is decreased as a function of the distance between the peak and central spot of the radiation cone.
distribution around 5.45 keV, beam profiles were obtained. The coordinates of the beam intensity profiles in this figure are the same as used in Fig. 5XAFS spectra taken in planar and gap-scan modes (Fig. 6b) show distorted spectral features around the beginning and (Fig. 6b, inset) in comparison with the spectrum of the tapered mode. These distortions are ascribed to significant intensity variations around the minimum of the third-harmonic energy accompanied with the non-linear responses of the detectors (Tanaka et al., 2001). The distortions were reduced when the linearity was improved by reducing the variation, but this reduction contradicts the purpose of securing the maximum intensity. Actually, when the BL8C IVU was used in continuous-scan or step-scan mode, the gap-scan mode gave better spectra than the tapered mode for available energy regions due to the intense X-rays. However, the tapered mode can be an alternative where the gap-scan has a considerable intensity variation and is suitable for quick-XAFS scanning.
4. Conclusion
We present the characteristic features of the BL8C IVU, operated in planar, tapered and gap-scanning modes, as a function of gap and gap tapering. The experimental spectral distributions of the IVU agree well with theoretical calculations in the range 4–20 keV. Beam intensity profiles around the energy of the intensity maximum of the third harmonic were obtained in different IVU operation modes to examine how the spectral variations depend on the IVU parameters and X-ray energies. Furthermore, K-edge were obtained to show the intensity variation and the effect of the IVU parameters on spectral quality. The spectra indicate that gap-scan mode, owing to the intense X-rays, is suitable for experiments for available energy regions when the IVU was used in continuous-scan or step-scan mode. The tapered mode, as well as being suitable for quick-XAFS scanning, can be an alternative where the gap-scan gives a considerable intensity variation.
spectra at the VAcknowledgements
The
experiments were performed with approval of authorities from the Pohang Light Source (PLS). This research was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education, Science and Technology (2011-0022268).References
Alferov, D. F., Bashmakov, Yu. A. & Bessonov, E. G. (1974). Sov. Phys. Tech. Phys. 18, 1336–1339. Google Scholar
Attwood, D., Halbach, K. & Kim, K. J. (1985). Science, 228, 1265–1272. CrossRef PubMed CAS Web of Science Google Scholar
Barrea, R. A., Fischetti, R., Stepanov, S., Rosenbaum, G., Kondrashkina, E., Bunker, G. B., Black, E., Zhang, K., Gore, D., Heurich, R., Vukonich, M., Karanfil, C., Kropf, A. J., Wang, S. & Irving, T. C. (2005). Phys. Scr. 115, 867–869. CrossRef Google Scholar
Boyanov, B. I., Bunker, G., Lee, J. M. & Morrison, T. I. (1994). Nucl. Instrum. Methods Phys. Res. A, 339, 596–603. CrossRef CAS Web of Science Google Scholar
Chapman, K., Lai, B., Cerrina, F. & Vicarro, J. (1989). Nucl. Instrum. Methods Phys. Res. A, 283, 88–99. CrossRef Web of Science Google Scholar
Curtin, M., Bhowmik, A., Brown, J., McMullin, W., Metty, P., Benson, S. V. & Madey, J. M. J. (1990). Nucl. Instrum. Methods Phys. Res. A, 296, 69–74. CrossRef Web of Science Google Scholar
Fischetti, R., Stepanov, S., Rosenbaum, G., Barrea, R., Black, E., Gore, D., Heurich, R., Kondrashkina, E., Kropf, A. J., Wang, S., Zhang, K., Irving, T. C. & Bunker, G. B. (2004). J. Synchrotron Rad. 11, 399–405. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gauthier, C., Solé, V. A., Signorato, R., Goulon, J. & Moguiline, E. (1999). J. Synchrotron Rad. 6, 164–166. Web of Science CrossRef CAS IUCr Journals Google Scholar
Guo, J.-H., Butorin, S. M., Sathe, C., Nordgren, J., Bassler, M., Feifel, R., Werin, S., Georgsson, M., Andersson, A., Jurvansuu, M., Nyholm, R. & Eriksson, M. (1999). Nucl. Instrum. Methods Phys. Res. A, 431, 285–293. Web of Science CrossRef CAS Google Scholar
Hagelstein, M., San Miguel, A., Fontaine, A. & Goulon, J. (1997). J. Phys. IV, 7(C2), 303–308. Google Scholar
Igarashi, N., Shimizu, N., Koyama, A., Mori, T., Ohta, H., Niwa, Y., Nitani, H., Abe, H., Nomura, M., Shioya, T., Tsuchiya, K. & Ito, K. (2013). J. Phys. Conf. Ser. 425, 072016. CrossRef Google Scholar
Inada, Y., Suzuki, A., Niwa, Y. & Nomura, M. (2007). AIP Conf. Proc. 879, 1230–1233. CrossRef CAS Google Scholar
Kim, K.-J. (1986). Nucl. Instrum. Methods Phys. Res. A, 246, 67–70. CrossRef Web of Science Google Scholar
Kincaid, B. M. (1977). J. Appl. Phys. 48, 2684. CrossRef Web of Science Google Scholar
Maréchal, X.-M. (1998). J. Synchrotron Rad. 5, 401–402. Web of Science CrossRef IUCr Journals Google Scholar
Mori, T., Nomura, M., Sato, M., Adachi, H., Uchida, Y., Toyoshima, A., Yamamoto, S., Tsuchiya, K., Shioya, T. & Kawata, H. (2004). AIP Conf. Proc. 705, 255–258. CrossRef CAS Google Scholar
Nonaka, T., Dohmae, K., Araki, T., Hayashi, Y., Hirose, Y., Uruga, T., Yamazaki, H., Mochizuki, T., Tanida, H. & Goto, S. (2012). Rev. Sci. Instrum. 83, 083112. CrossRef PubMed Google Scholar
Oyanagi, H., Ishii, M., Lee, C.-H., Saini, N. L., Kuwahara, Y., Saito, A., Izumi, Y. & Hashimoto, H. (2000). J. Synchrotron Rad. 7, 89–94. Web of Science CrossRef CAS IUCr Journals Google Scholar
Rogalev, A., Gotte, V., Goulon, J., Gauthier, C., Chavanne, J. & Elleaume, P. (1998). J. Synchrotron Rad. 5, 989–991. Web of Science CrossRef CAS IUCr Journals Google Scholar
Saitoh, Y., Fukuda, Y., Takeda, Y., Yamagami, H., Takahashi, S., Asano, Y., Hara, T., Shirasawa, K., Takeuchi, M., Tanaka, T. & Kitamura, H. (2012). J. Synchrotron Rad. 19, 388–393. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sanchez del Rio, M. & Dejus, R. J. (1998). Proc. SPIE, 3448, 340–345. CrossRef Google Scholar
Sanchez del Rio, M. & Dejus, R. J. (2004). Proc. SPIE, 5536, 171–174. CrossRef Google Scholar
Sasaki, S., Youngman, B. & Winick, H. (1990). Nucl. Instrum. Methods Phys. Res. A, 291, 401–407. CrossRef Web of Science Google Scholar
Sekizawa, O. et al. (2013). J. Phys. Conf. Ser. 430, 012020. CrossRef Google Scholar
Shih, C. C. & Caponi, M. Z. (1982). Phys. Rev. A, 26, 438–450. CrossRef CAS Web of Science Google Scholar
Somogyi, A., Kewish, C. M., Ribbens, M., Moreno, T., Polack, F., Baranton, G., Desjardins, K. & Samama, J. P. (2013). J. Phys. Conf. Ser. 463, 012027. CrossRef Google Scholar
Takahashi, K., Imamura, M., Yamamoto, I., Azuma, J., Ogawa, K., Kamada, M., Ohkuma, H. & Yamamoto, S. (2013). J. Phys. Conf. Ser. 425, 072007. CrossRef Google Scholar
Tanaka, T., Matsubayashi, N., Imamura, M. & Shimada, H. (2001). J. Synchrotron Rad. 8, 345–347. Web of Science CrossRef CAS IUCr Journals Google Scholar
Tanida, H. & Ishii, M. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 1564–1567. Web of Science CrossRef CAS Google Scholar
Walker, R. P. (1988). Nucl. Instrum. Methods Phys. Res. A, 267, 537–546. CrossRef Web of Science Google Scholar
Welter, E. (2012). J. Synchrotron Rad. 19, 905–910. Web of Science CrossRef CAS IUCr Journals Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.