research papers
Correcting for surface topography in
imagingaLaboratory of Atomic and Solid State Physics, Cornell University, Ithaca, NY 14853, USA
*Correspondence e-mail: ret6@cornell.edu
Samples with non-planar surfaces present challenges for
imaging analysis. Here, approximations are derived to describe the modulation of fluorescence signals by surface angles and topography, and suggestions are made for reducing this effect. A correction procedure is developed that is effective for analysis of samples having a uniform matrix, and requires only a fluorescence map from a single detector. This procedure is applied to fluorescence maps from an incised gypsum tablet.Keywords: X-ray fluorescence; imaging; surface angle; surface orientation; topography.
1. Introduction
In synchrotron-based et al., 1992; Mantler & Schreiner, 2000; Solé et al., 2007). This technique can furnish important information about a wide range of objects, including biological specimens (Limburg et al., 2007), mineral specimens (Lintern et al., 2013), paintings (Dik et al., 2008) and archaeological artifacts, including ancient documents on parchment and stone (Bergmann & Knox, 2009; Powers et al., 2005, 2009).
imaging (XRFI), a small-diameter beam of synchrotron-generated X-ray radiation is scanned across the surface of an object. The fluorescence signal from each point is measured using an energy-dispersive detector, and then analyzed to yield a two-dimensional map of elemental concentrations across the object's surface (RiversAnalysis of XRFI data is most straightforward for objects with flat surfaces. For objects with surface curvature or relief, such as vases or tablets with incised letters (Powers et al., 2005, 2009), variations in local surface orientation relative to the incident X-ray beam modulate the intensity of received at the detector (Bonizzoni et al., 2006). These geometry-related intensity modulations cannot easily be distinguished from those due to variations in sample composition. The problem of quantitative (XRF) analysis of objects with arbitrary geometries is difficult and often ill-posed (Urbach & de Bokx, 2001). Computationally intensive Monte Carlo methods (Scot et al., 2007; Trojek et al., 2010; Trojek & Wegrzynek, 2010; Trojek, 2011, 2012) have been successfully applied in some cases.
Here, we derive simple expressions for the variation of fluorescence intensity with surface orientation, and discuss a procedure to correct fluorescence maps for the effects of surface relief. Although the necessary assumptions about sample surface relief and composition and about measurement geometry are fairly restrictive, the analysis is still useful for many real-world samples, as we will illustrate.
2. Variation of detected fluorescence intensity with surface and detector angles
To analyze the effect of surface relief, we make the following simplifying assumptions:
(i) Variations in surface orientation and sample composition are small on the scales of the incident X-ray beam width (∼1 µm–1 mm) and
(∼500 µm), respectively.(ii) The detector is far from the surface, so that differences in path length and angle from various points in the X-ray illuminated and fluorescing sample volume to the detector can be neglected, and integration over the detector's solid angle is unnecessary.
(iii) The sample's surface relief and the sample and detector orientations are such that there is an unobstructed path from the incident beam's interaction point to the detector for all points on the sample surface, i.e. the path to the detector from one point on the sample surface is not blocked by another part of the surface.
(iv) There is no secondary fluorescence.
As illustrated in Fig. 1 (left), we choose coordinates so that the beam intersects the surface at the origin, and define unit vectors b along the incident X-ray beam direction, n along the surface normal at the incident beam's interaction point, and d from the interaction point to the detector.
The intensity of the fluorescence signal at the detector is
where λ is a constant involving the and detector solid angle; = [ = ] is the sample at the incident [fluorescent] energy, and si [sf] is the length of the line from the sample surface to the interaction point along the beam direction [detector direction], as shown in Fig. 1 (right).
To determine sf, note that the interaction point is located at , and that the vector must lie in the surface plane. The equation of the surface is = 0, so that
Thus,
where k is a geometrical factor. (Note that sf is negative or fails to exist when the surface is angled away from the detector.) Then,
In our coordinate system the beam points along the +y direction ( = ), and the detector lies in the xy (horizontal) plane, making an angle α with −y. Then, if = (nx,ny,nz) and = ,
where θ is the rotation of the surface normal around the z axis; = 0 for a surface parallel to the xz plane. Finally, we have
Fig. 2 shows versus surface orientation angle θ at several detector angles α calculated using equation (5) and = 14 (appropriate for Ca fluorescence in a calcite matrix, with an incident energy of 17 keV).
The above calculation leads to several observations. One, changes in the surface orientation θ result in changes in the detected fluorescence intensity and thus, if no correction is performed, in apparent changes in elemental composition (Fig. 3). The magnitude of this intensity variation with θ increases with increasing , and so is often most apparent for low-Z elements, which have relatively small mass absorption coefficients , and for which at the relatively low energy of their fluorescence is relatively large.
Two, under the given approximations, rotations of the surface normal out of the xy plane (for the polar coordinate φ within ±90°) do not change θ and the detected fluorescence intensity. Thus, whenever possible, objects should be mounted so that their dominant surface curvature runs perpendicular to the detector–incident-beam (xy) plane. For instance, mounting a cylindrical sample with its axis in the xy plane avoids position-dependent modulation of the fluorescence intensities.
Finally, the angle effect vanishes as the detector position approaches the incident beam, and is maximal when the detector is perpendicular to the beam. For polarized incident radiation (e.g. synchrotron radiation), Compton scattering is greatly reduced when the detector is perpendicular to the beam, in the plane of polarization (Dzubay et al., 1974). To minimize Compton background, typical XRF configurations place the detector at α = 90° and the sample normal at θ = 45°. However, for samples with appreciable surface relief, a better approach is to orient the sample's `mean' normal in the −y direction, and to set the detector at as small an angle α as practicable (typically about 45°). The very high counting rates possible using modern detectors make the Compton background less problematic.
Orienting the sample's mean normal and the translation stage's normal in the −y direction (so that translations occur in the xz plane) solves another problem too: if the stage is at any other angle, sample topography will vary the interaction point of the beam in stage xz coordinates, and therefore in the image coordinates. This distorts the images, sometimes discontinuously, and can lead to difficulty in registering the XRFI scan maps with actual features on the sample.
The surface and detector angles also affect the , we can find the mean depth at which the recorded fluorescent photons are generated,
of the X-rays, and therefore the surface sensitivity of the measurement. From equation (3)Fig. 4 shows the variation in mean depth with angle in a calcite sample. This effect further complicates analysis of XRFI data from samples with relief.
3. Correction of XRFI maps for the effects of surface relief
If the approximate matrix of a sample is known and is spatially uniform, and the non-matrix elements are present in trace concentrations, then good approximations to can be calculated. In that case the only undetermined variable in equation (5) is θ. Therefore, if (where x,z are coordinates on the sample surface) can be determined, the XRFI data can be corrected for surface relief, at least to the extent that the approximations leading to equation (5) hold.
The most straightforward way to find is to spatially profile the surface of the object, e.g. with a laser displacement sensor or via stereo photogrammetry. The three-dimensional profile must then be registered with the XRFI data. Profiling may not always be feasible due to equipment and time constraints.
This raises the question: is it possible to correct for angle based on the fluorescence data alone? Often, the matrix of a sample contains one XRF-visible element that is present in roughly uniform and high concentration over the entire surface; for instance, calcium in marble, limestone or gypsum. In this case we can assume that the reference element is, in fact, uniformly distributed, and use the intensity of one of its fluorescence lines to invert equation (5) and find . Equation (5) can then be used to correct the XRFI maps for each trace element.
We have found the following inversion and correction procedure to be effective:
(i) Given the matrix composition, compute the et al. (2004) and Schoonjans et al. (2011).
for the energy range of interest. We use the data of Brunetti(ii) Choose a single fluorescence peak from the fitted spectra, for a major matrix element that is expected to be uniformly distributed (e.g. Ca-Kα for limestone, marble and gypsum).
(iii) Let I(x,z) be the intensity (i.e. number of counts) of this peak at points x,z in the scan; I0 the median of I in a relatively flat region of the sample; and k0 the geometric constant [according to equation (4)] for this region.
(iv) Compute
(v) Invert equation (4) to find θ.
Next, with k(x,z) determined by (7), calculate a correction factor as a function of the energy E of the fluorescence,
Finally, apply this correction to the XRF spectrum at each pixel of the scan, re-fit the peaks in the corrected spectra, and then generate corrected XRFI maps for each element of interest.
4. Application to an incised tablet
As an example of the application of the above inversion and correction procedure, we examined a gypsum (CaSO4·2H2O) tablet from what is now Iraq, dated ca 1000 BCE, shown in Fig. 5. The tablet is deeply incised with cuneiform figures. XRFI measurements were performed at beamline F3 of the Cornell High Energy Synchrotron Source (CHESS), using an incident X-ray beam energy of 17.0 keV, an X-ray beam size of 200 µm square, a stage angle of 90° and a detector angle of 45°. The detector was a four-element Vortex ME4 silicon drift detector from SII Nanotechnology. The sample stage was continuously scanned at 2 mm s−1, and spectra determined for steps (pixels) of size 200 µm. Fluorescence spectra were fit using the PyMCA software library (Solé et al., 2007). Fig. 6 shows the uncorrected Ca-Kα and Pb-Lα fluorescence maps for the outlined scan region in Fig. 5.
Calcium should be roughly uniformly distributed throughout the tablet, but the Ca map shows pronounced intensity variations that correlate with variations in surface angle. The detector was located to the right of the image, leading to apparently enhanced calcium fluorescence from the left-hand sides of the incisions.
The Pb map appears to indicate larger Pb concentrations in the incisions. However, it is not clear whether this is a geometric artifact since, like Ca, Pb fluorescence is stronger on the left-hand sides of the incisions. This tablet illustrates typical complications in interpreting fluorescence maps from non-planar objects.
Applying the inversion procedure outlined above and using the Ca-Kα peak at 3.69 keV as the reference peak yields the map of surface orientation angle θ shown at the top of Fig. 7. Fig. 7, bottom left, shows a rendering of the surface with angles as shown at the top, shaded as if illuminated by a light from the approximate position of the detector; the luminance L was computed using Lambertian shading, with L . Interestingly, this shows the pattern of relief more clearly than the optical photograph shown at the bottom right; it is surprisingly accurate, considering the restrictive approximations under which it was computed. Note that, while the fluorescence intensity gives no direct information about φ rotations of the local surface normal (out of the xy plane), only is needed to reconstruct the surface topography.
Using the surface orientation map in Fig. 7, the XRFI maps can now be corrected. We first check the corrected Ca-Kβ map (4.01 keV) map. The Ca-Kβ peak was measured and fitted independently of the Ca-Kα peak, so, if our model is valid, the corrected Ca-Kβ map shown at the upper left in Fig. 8 should be almost flat. Indeed, it is, aside from small regions where the surface was at an angle θ < −α = −45° and so was blocked from the detector's view.
Fig. 8 also shows corrected Pb and Ti fluorescence maps. In contrast to Fig. 6, Pb now appears evenly distributed between the left and right sides of the incisions. We are confident, therefore, that there actually is an increased concentration of Pb (and also Ti) within the incised features. These trace elements could have been introduced from the tools used to carve the figures [titanium has been associated with bronze-age tin production (Adriaens, 1996)], paint used to decorate the tablet, or contaminants which, for example, have been scrubbed from the field of the tablet but not from the protected valleys.
We emphasize that in calculating the corrections we have assumed that the sample composition normal to the average surface plane is uniform within the X-ray attenuation length (roughly 600 µm). In the limiting case where all trace elements were concentrated within a very thin i.e. much thinner than the attenuation length), there would be a smaller correction to the concentrations, proportional to (where γ is the angle between the surface normal and −b), given by = . Applying this latter correction produced maps that were qualitatively similar to the uncorrected ones. In either case, there is evidently an increased concentration of Pb and Ti in the incisions.
(Also note that our correction procedure assumes that elements of interest do not affect the detected fluorescence from the matrix reference element. For example, a thick Pb e.g. from a pigment) would absorb and therefore mask the Ca fluorescence that we use as a reference, and equation (7) would then underestimate the surface angle. For the specimen considered here, there is no visual or XRF evidence of such layers; the elements of interest are present in trace concentrations.
(5. Conclusions
We have derived simple expressions for the modulation of
intensity by surface orientation relative to the incident beam, and we have demonstrated a correction procedure applicable to maps of such samples. The analysis assumes that the major elemental constituents of the matrix are homogeneous throughout the sample, that the surface is locally flat on the scale of the X-ray beam, and that absorption of matrix element fluorescence by layers can be neglected. These assumptions should be approximately valid for a wide range of samples.A key advantage of this method is its simplicity, both in the requisite experimental apparatus (a single detector, and no separate apparatus for determining surface topography) and in its analysis. However, the estimates of surface angles might be improved by using multiple detectors, e.g. positioned in pairs on opposite sides of the beam. One approach to dual-detector topography correction was reported by Smilgies et al. (2012). However, that work simply used the sum and difference of the maps from the two detectors, and, as equation (5) shows, the variation of fluorescence intensity with angle is not linear, even in the case of a flat homogeneous surface. Fitting equation (7) to fluorescence maps from multiple detectors would likely yield better corrections, and would reduce the occurrence of `blind spots' where the detector's view is blocked by another part of the sample. Finally, the recently developed highly parallel annular detectors (Ryan et al., 2010) raise an intriguing possibility: analyzed using methods similar to those outlined here, a single scan might provide extremely detailed reconstructions of both the composition and geometry of a sample.
Acknowledgements
Experimental work was conducted at the Cornell High Energy Synchrotron Source, which is supported by the NSF and NIH/NIGMS via NSF award DMR-1332208. We thank Joseph Greene and Adam Aja of the Harvard Semitic Museum for the loan of the sample, Darren Dale and Arthur Woll of CHESS for technical assistance, and Kevin Clinton for collaboration on imaging Greek and Latin inscriptions on stone that led to the present work. This work was supported in part by grants from the Samuel H. Kress Foundation and the Mellon Foundation. ECG also acknowledges support from the National Science Foundation under grant DMR-0805240.
References
Adriaens, A. (1996). Microchim. Acta, 124, 89–98. CrossRef CAS Web of Science Google Scholar
Bergmann, U. & Knox, K. T. (2009). Proc. SPIE, 7247, 724702. CrossRef Google Scholar
Bonizzoni, L., Maloni, A. & Milazzo, M. (2006). X-ray Spectrom. 35, 390–399. Web of Science CrossRef CAS Google Scholar
Brunetti, A., Sanchez del Rio, M., Golosio, B., Simionovici, A. & Somogyi, A. (2004). Spectrochim. Acta B, 59, 1725–1731. Web of Science CrossRef Google Scholar
Dik, J., Janssens, K., Snickt, G. V. D., van der Loeff, L., Rickers, K. & Cotte, M. (2008). Anal. Chem. 80, 6436–6442. Web of Science CrossRef PubMed CAS Google Scholar
Dzubay, T. G., Jarrett, B. V. & Jaklevic, J. M. (1974). Nucl. Instrum. Methods, 115, 297–299. CrossRef CAS Web of Science Google Scholar
Limburg, K. E., Huang, R. & Bilderback, D. H. (2007). X-ray Spectrom. 36, 336–342. Web of Science CrossRef CAS Google Scholar
Lintern, M., Anand, R., Ryan, C. & Paterson, D. (2013). Nat. Commun. 4, 2274. Web of Science CrossRef PubMed Google Scholar
Mantler, M. & Schreiner, M. (2000). X-ray Spectrom. 29, 3–17. CrossRef CAS Google Scholar
Powers, J., Dimitrova, N., Huang, R., Smilgies, D.-M., Bilderback, D. H. & Thorne, R. E. (2005). Z. Papyrol. Epigr. 152, 221–227. Google Scholar
Powers, J., Smilgies, D.-M., Geil, E. C., Clinton, K., Peachin, M. & Thorne, R. E. (2009). J. Archaeol. Sci. 36, 343–350. Web of Science CrossRef Google Scholar
Rivers, M. L., Sutton, S. R. & Jones, K. W. (1992). X-ray Microscopy III, edited by A. G. Michette, G. R. Morrison and C. J. Buckley, Vol. 67 of Springer Series in Optical Sciences, pp. 212–216. Berlin: Springer. Google Scholar
Ryan, C. G., Siddons, D. P., Kirkham, R., Dunn, P. A., Kuczewski, A., Moorhead, G., De Geronimo, G., Paterson, D. J., de Jonge, M. D., Hough, R. M., Lintern, M. J., Howard, D. L., Kappen, P. & Cleverley, J. (2010). AIP Conf. Proc. 1221, 9–17. CrossRef CAS Google Scholar
Schoonjans, T., Brunetti, A., Golosio, B., del Rio, M. S., Solé, V., Ferrero, C. & Laszlo, V. (2011). Spectrochim. Acta B, 66, 776–784. Web of Science CrossRef CAS Google Scholar
Scot, V., Fermandez, J. E., Vincze, L. & Janssens, K. (2007). Nucl. Instrum. Methods Phys. Res. B, 263, 204–208. Web of Science CrossRef CAS Google Scholar
Smilgies, D.-M., Powers, J. A., Bilderback, D. H. & Thorne, R. E. (2012). J. Synchrotron Rad. 19, 547–550. Web of Science CrossRef IUCr Journals Google Scholar
Solé, V., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). Spectrochim. Acta B, 62, 63–68. Google Scholar
Trojek, T. (2011). J. Anal. At. Spectrom. 26, 1253–1257. Web of Science CrossRef CAS Google Scholar
Trojek, T. (2012). Appl. Radiat. Isotopes, 70, 1206–1209. Web of Science CrossRef CAS Google Scholar
Trojek, T., Čecàk, T. & Musílek, L. (2010). Nucl. Instrum. Methods Phys. Res. A, 619, 266–269. Web of Science CrossRef CAS Google Scholar
Trojek, T. & Wegrzynek, D. (2010). Nucl. Instrum. Methods Phys. Res. A, 619, 311–315. Web of Science CrossRef CAS Google Scholar
Urbach, H. P. & de Bokx, P. K. (2001). Phys. Rev. B, 63, 085408. CrossRef Google Scholar