Figure 2
Calibration of the MASSIF-1 Laue [110] diamond monochromator. (a) Scan in transmission mode of the Ir LII absorption edge. The blue line shows the position of the inflection point, the cyan line the angle to which the monochromator is moved to ensure a monochromatic X-ray beam of E = 12.80 keV (λ = 0.969 Å). (b) Scan of the MASSIF-2 monochromator θ angle; the glitch from the MASSIF-1 monochromator can be seen at 12.78 keV before it was moved to the calibrated position. Intensity units are arbitrary.