Effect of X-ray spot size on liquid jet photoelectron spectroscopy

CROSSMARK_Color_square_no_text.svg

aLaboratory for Surface Science and Technology, Department of Materials, ETH Zürich, CH-8093 Zurich, Switzerland, and bSwiss Light Source, Paul Scherrer Institut, CH-5023 Villigen PSI, Switzerland
*Correspondence e-mail: armin.kleibert@psi.ch, matthew.brown@mat.ethz.ch

Edited by M. Yabashi, RIKEN SPring-8 Center, Japan (Received 3 July 2015; accepted 1 September 2015; online 30 September 2015)

A 30 µm pinhole is introduced in the intermediate focus of the SIM beamline at the Swiss Light Source to improve the spot size at the second downstream focus, which is used here for liquid jet X-ray photoelectron spectroscopy experiments. The 30 µm pinhole reduces the beam dimensions from 250 (v) × 100 (h) µm to 75 × 45 µm for a vertical exit slit of 100 µm. The smaller X-ray spot results in a substantial decrease in the gas-phase contribution of the spectra from 40% down to 20% and will help to simplify the interpretation and peak assignments of future experiments.

1. Introduction

Liquid jet (LJ) based X-ray photoelectron spectroscopy (XPS) (Siegbahn & Siegbahn, 1973[Siegbahn, H. & Siegbahn, K. (1973). J. Electron Spectrosc. Relat. Phenom. 2, 319-325.]; Winter & Faubel, 2006[Winter, B. & Faubel, M. (2006). Chem. Rev. 106, 1176-1211.]) is a powerful analytical probe of electronic and geometric structures at liquid interfaces and has found widespread application over the last decade in diverse areas of environmental (Brown et al., 2012[Brown, M. A., Vila, F., Sterrer, M., Thürmer, S., Winter, B., Ammann, M., Rehr, J. J. & van Bokhoven, J. A. (2012). J. Phys. Chem. Lett. 3, 1754-1759.], 2009[Brown, M. A., Winter, B., Faubel, M. & Hemminger, J. C. (2009). J. Am. Chem. Soc. 131, 8354-8355.]), biological (Ottosson et al., 2011[Ottosson, N., Børve, K. J., Spångberg, D., Bergersen, H., Saethre, L. J., Faubel, M., Pokapanich, W., Öhrwall, G., Björneholm, E. & Winter, B. (2011). J. Am. Chem. Soc. 133, 3120-3130.]), soft-matter (Brown et al., 2014[Brown, M. A., Arrigoni, M., Héroguel, F., Beloqui Redondo, A., Giordano, L., van Bokhoven, J. A. & Pacchioni, G. (2014). J. Phys. Chem. C, 118, 29007-29016.]; Brown, Beloqui Redondo, Sterrer et al., 2013[Brown, M. A., Beloqui Redondo, A., Sterrer, M., Winter, B., Pacchioni, G., Abbas, Z. & van Bokhoven, J. A. (2013). Nano Lett. 13, 5403-5407.]) and fundamental (Björneholm et al., 2014[Björneholm, O., Werner, J., Ottosson, N., Öhrwall, G., Ekholm, V., Winter, B., Unger, I. & Söderström, J. (2014). J. Phys. Chem. C, 118, 29333-29339.]; Winter et al., 2004[Winter, B., Weber, R., Widdra, W., Dittmar, M., Faubel, M. & Hertel, I. V. (2004). J. Phys. Chem. A, 108, 2625-2632.]) sciences. One often-limiting factor in the interpretation of the spectra is the overlap of the photoelectron lines originating from liquid with those originating from gas-phase molecules that surround the liquid jet in vacuum or at ambient pressures. This contribution can be exaggerated when the X-ray spot size greatly exceeds that of the liquid jet (typically <30 µm) and nearly suppressed when the two are of the same size. These overlapping peaks may lead to the incorrect assignment of the liquid peaks or prevent peak assignment altogether. The latter is often the case in the valence-band region where multiple peaks are present from both condensed and gas phases.

Modifying the focal X-ray spot profile of a synchrotron beamline can, in principle, be achieved using, for example, Kirkpatrick–Baez mirror pairs or Fresnel zone plates. These approaches require a dedicated beamline and the appropriate endstation design and are, therefore, not easily implemented into existing equipment. Here we demonstrate improved performance for LJ-XPS by introducing a 30 µm circular pinhole in the intermediate focus of the SIM beamline (Flechsig et al., 2010[Flechsig, U., Nolting, F., Fraile Rodríguez, A., Krempaský, J., Quitmann, C., Schmidt, T., Spielmann, S., Zimoch, D., Garrett, R., Gentle, I., Nugent, K. & Wilkins, S. (2010). AIP Conf. Proc. 1234, 319-322.]) at the Swiss Light Source (SLS). In the normal operation conditions of the beamline, the pinhole reduces the spot size at the second refocusing station where the LJ-XPS is operated and results in spectra that have up to 20% lower gas-phase contributions. This approach is straightforward to install and works for all wavelengths. Moreover, it requires virtually no capital investment.

2. Experimental setup

The SIM beamline is equipped with two 3.8 m Apple II undulators in the X11M straight section of the SLS followed by a plane-grating monochromator (PGM) and employs a collimated light scheme (Follath & Senf, 1997[Follath, R. & Senf, F. (1997). Nucl. Instrum. Methods Phys. Res. A, 390, 388-394.]) to provide an intermediate focus for a permanently installed photoemission electron microscope (PEEM). The latter is equipped with a sample manipulator that allows for accurate and stable positioning of the sample in both the focus of the microscope and the intermediate focal plane of the beamline (Le Guyader et al., 2012[Le Guyader, L., Kleibert, A., Fraile Rodríguez, A., El Moussaoui, S., Balan, A., Buzzi, M., Raabe, J. & Nolting, F. (2012). J. Electron Spectrosc. Relat. Phenom. 185, 371-380.]). When retracting the PEEM sample a second re­focusing mirror further downstream provides a nearly 1:1 image of the intermediate focus (Flechsig et al., 2010[Flechsig, U., Nolting, F., Fraile Rodríguez, A., Krempaský, J., Quitmann, C., Schmidt, T., Spielmann, S., Zimoch, D., Garrett, R., Gentle, I., Nugent, K. & Wilkins, S. (2010). AIP Conf. Proc. 1234, 319-322.]) for operation of a variety of exchangeable endstations (Fig. 1[link]a). Here, we mount a precision platinum pinhole (Plano GmbH) to an adapted PEEM sample holder and place the pinhole (instead of a PEEM sample) in the intermediate focus to modify the spot profile at the second focal plane. The effect of the pinhole on the X-ray spot profile in the second focal plane is monitored using a phosphor screen and a CCD camera. XPS measurements are performed using the near ambient pressure photoemission (NAPP) endstation of the SLS (Brown, Beloqui Redondo, Jordan et al., 2013[Brown, M. A., Redondo, A. B., Jordan, I., Duyckaerts, N., Lee, T. M., Ammann, M., Nolting, F., Kleibert, A., Huthwelker, T., Mächler, J. P., Birrer, M., Honegger, J., Wetter, R., Wörner, H. J. & van Bokhoven, J. A. (2013). Rev. Sci. Instrum. 84, 073904.]). A 19 µm liquid jet of 0.05 M NaCl is operated in vacuum (10−4 mbar). O 1s spectra are collected at ∼160 eV kinetic energy (KE) using a photon energy of 700 eV. Gas phase N2 spectra were recorded at 0.3 mbar using the same beamline settings as for the liquid jet experiments. A 200 nm-thick SiNx window separates the ultrahigh vacuum of the beamline from the ambient N2 environment. All spectra were recorded at a pass energy of 50 eV with a 300 µm entrance slit on a Scienta R4000 HiPP-2 analyzer.

[Figure 1]
Figure 1
SIM beamline layout (a) and X-ray spot image after the second toroidal refocusing mirror: (b) without a pinhole and (c) with a 30 µm pinhole in the intermediate focus of the beamline. In both images the exit slit of the beamline is 100 µm and the photon energy is 800 eV.

3. Results and discussion

The X-ray spot profile after the second toroidal refocusing mirror is shown in Fig. 1[link](b) for a photon energy of 800 eV. With a vertical exit slit size (dispersive plane) of 100 µm we measure an X-ray spot size of 250 (h) × 100 (v) µm. Inserting a 30 µm pinhole at the intermediate focus of the beamline reduces the X-ray size to 75 (h) × 45 (v) µm (Fig. 1[link]c). Under these conditions the pinhole decreases the total flux at the second focal plane by ∼85%, as measured by an IRD SXUV 100 photodiode.

O 1s spectra from a 19 µm liquid jet of 0.05 M NaCl operating in vacuum (10−4 mbar) are shown in Fig. 2[link](a) (without pinhole) and Fig. 2[link](b) (with a 30 µm pinhole). The vertical exit slit of the beamline is 10 µm for these experiments and, therefore, any reduction in X-ray spot size is expected primarily in the horizontal (width) direction. Because the liquid jet flows in the vertical direction, the smaller X-ray width will lead to fewer gas-phase molecules being ionized between the liquid jet and the entrance aperture of the hemispherical analyzer. The spectra are deconvoluted into two components, O 1s(gas) at low KE and O 1s(liq) at high KE. Without the pinhole the integrated gas-phase area contributes 40% of the total signal, whereas with the 30 µm pinhole in the intermediate focus of the beamline the gas-phase signal is reduced by a factor of two (Table 1[link]). It becomes immediately clear that introducing this small pinhole in the intermediate focus of the SIM beamline provides significant benefit for LJ-XPS experiments (or any experiment that will require a small X-ray focal size).

Table 1
XPS peak fitting results

  Without a pinhole With a 30 µm pinhole
Peak Contribution (%) FWHM (eV) Contribution (%) FWHM (eV)
19 µm liquid jet of 0.05 M NaCl
 O 1s(liq) 59.9 1.5 79.0 1.5
 O 1s(gas) 40.1 0.5 21.0 0.4
 
0.3 mbar N2(gas)
 N 1s(gas) N/A 0.4 N/A 0.3
[Figure 2]
Figure 2
O 1s spectra from a 19 µm liquid jet of 0.05 M NaCl collected with a photon energy of 700 eV: (a) without a pinhole and (b) with a 30 µm pinhole in the intermediate focus of the beamline. Both spectra are acquired using a 10 µm exit slit on the beamline. The peaks corresponding to the liquid jet, O 1s(liq), and from gas phase, O 1s(gas), are labeled only in (a).

Further, we observe a small decrease in the FWHM of the O 1s(gas) component when using the 30 µm pinhole (0.5 → 0.4 eV, Table 1[link]). For these experiments the hemispherical energy analyzer operates at a theoretical resolution of 0.05 eV, suggesting that the decrease in FWHM originates from an increase in resolution of the beamline when using the 30 µm pinhole. To quantify this observation we performed an additional set of experiments in 0.3 mbar N2 (Fig. 3[link]). The FWHM of the N 1s again shows a decrease when using the 30 µm pinhole (0.5 → 0.4 eV, Table 1[link]) confirming that, in addition to reducing the X-ray spot size, the 30 µm pinhole increases the resolving power of the beamline. The origin of this behavior is a decrease in stray light of the beamline in its typical operating conditions (cutoff by the pinhole). However, because the FWHM of O 1s(liq) is intrinsically broad, due to fluctuations in the local hydrogen-bonding environment of liquid water (Winter & Faubel, 2006[Winter, B. & Faubel, M. (2006). Chem. Rev. 106, 1176-1211.]), this improved resolution is only notable for gas-phase lines.

[Figure 3]
Figure 3
N 1s spectra from 0.3 mbar N2(gas) collected with a photon energy of 700 eV: (red) without a pinhole and (blue) with a 30 µm pinhole in the intermediate focus of the beamline. Both spectra are acquired using a 10 µm exit slit on the beamline. The full widths at half-maximum (FWHM) are shown.

4. Conclusions

A circular 30 µm pinhole introduced at an intermediate focus has been used to decrease the X-ray spot size at the second focal plane of the SIM beamline. This reduced the gas-phase contribution in LJ-XPS spectra by a factor of two and improved the energy resolution by about 25%. These improvements are easy to achieve and will prove useful for future experiments involving high vapor pressure organic solvents such as those employed in Li-ion batteries.

Acknowledgements

The authors acknowledge F. Nolting and U. Flechsig for fruitful discussions and A. Jaggi and P. Schifferle for technical support at the beamline. This work was supported by the Swiss National Science Foundation (No. 153578) and a ETH Research Grant (ETH-20 13-2).

References

First citationBjörneholm, O., Werner, J., Ottosson, N., Öhrwall, G., Ekholm, V., Winter, B., Unger, I. & Söderström, J. (2014). J. Phys. Chem. C, 118, 29333–29339.  Google Scholar
First citationBrown, M. A., Arrigoni, M., Héroguel, F., Beloqui Redondo, A., Giordano, L., van Bokhoven, J. A. & Pacchioni, G. (2014). J. Phys. Chem. C, 118, 29007–29016.  CrossRef CAS Google Scholar
First citationBrown, M. A., Beloqui Redondo, A., Sterrer, M., Winter, B., Pacchioni, G., Abbas, Z. & van Bokhoven, J. A. (2013). Nano Lett. 13, 5403–5407.  CrossRef CAS PubMed Google Scholar
First citationBrown, M. A., Redondo, A. B., Jordan, I., Duyckaerts, N., Lee, T. M., Ammann, M., Nolting, F., Kleibert, A., Huthwelker, T., Mächler, J. P., Birrer, M., Honegger, J., Wetter, R., Wörner, H. J. & van Bokhoven, J. A. (2013). Rev. Sci. Instrum. 84, 073904.  CrossRef PubMed Google Scholar
First citationBrown, M. A., Vila, F., Sterrer, M., Thürmer, S., Winter, B., Ammann, M., Rehr, J. J. & van Bokhoven, J. A. (2012). J. Phys. Chem. Lett. 3, 1754–1759.  Web of Science CrossRef CAS PubMed Google Scholar
First citationBrown, M. A., Winter, B., Faubel, M. & Hemminger, J. C. (2009). J. Am. Chem. Soc. 131, 8354–8355.  CrossRef PubMed CAS Google Scholar
First citationFlechsig, U., Nolting, F., Fraile Rodríguez, A., Krempaský, J., Quitmann, C., Schmidt, T., Spielmann, S., Zimoch, D., Garrett, R., Gentle, I., Nugent, K. & Wilkins, S. (2010). AIP Conf. Proc. 1234, 319–322.  CrossRef Google Scholar
First citationFollath, R. & Senf, F. (1997). Nucl. Instrum. Methods Phys. Res. A, 390, 388–394.  CrossRef CAS Web of Science Google Scholar
First citationLe Guyader, L., Kleibert, A., Fraile Rodríguez, A., El Moussaoui, S., Balan, A., Buzzi, M., Raabe, J. & Nolting, F. (2012). J. Electron Spectrosc. Relat. Phenom. 185, 371–380.  CrossRef CAS Google Scholar
First citationOttosson, N., Børve, K. J., Spångberg, D., Bergersen, H., Saethre, L. J., Faubel, M., Pokapanich, W., Öhrwall, G., Björneholm, E. & Winter, B. (2011). J. Am. Chem. Soc. 133, 3120–3130.  CrossRef CAS PubMed Google Scholar
First citationSiegbahn, H. & Siegbahn, K. (1973). J. Electron Spectrosc. Relat. Phenom. 2, 319–325.  CrossRef CAS Google Scholar
First citationWinter, B. & Faubel, M. (2006). Chem. Rev. 106, 1176–1211.  Web of Science CrossRef PubMed CAS Google Scholar
First citationWinter, B., Weber, R., Widdra, W., Dittmar, M., Faubel, M. & Hertel, I. V. (2004). J. Phys. Chem. A, 108, 2625–2632.  CrossRef CAS Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Volume 22| Part 6| November 2015| Pages 1528-1530
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds