research papers
An endstation for resonant inelastic X-ray scattering studies of solid and liquid samples
aPhoton Science, Deutsches Elektronen-Synchrotron, Notkestrasse 85, 22607 Hamburg, Germany, bStructural Dynamics of (Bio)chemical Systems, Max Planck Institute for Biophysical Chemistry, Am Fassberg 11, 37077 Goettingen, Germany, cInstitute for Material Physics, University of Goettingen, Friedrich-Hund-Platz 1, 37077 Goettingen, Germany, and dInstitute for X-ray Physics, University of Goettingen, Friedrich-Hund Platz 1, Goettingen, Germany
*Correspondence e-mail: simone.techert@desy.de
A novel experimental setup is presented for resonant inelastic X-ray scattering investigations of solid and liquid samples in the soft X-ray region for studying the complex
of (bio)chemical systems. The uniqueness of the apparatus is its high flexibility combined with optimal energy resolution and energy range ratio. The apparatus enables investigation of chemical analyses, which reflects the chemical imprints. The endstation is composed of a main sample chamber, a sample holder for either solid or liquid jet delivery system, and a soft X-ray grating spectrometer for 210–1250 eV with a of ∼1000. It combines for the first time liquid jet technology with a soft X-ray spectrometer based on the variable line spacing principle. This setup was commissioned at the soft X-ray beamline P04 at PETRA III of the Deutsches Elektronen-Synchrotron in Hamburg which is currently the most brilliant storage-ring-based X-ray radiation source in the world. The first results of liquid and solid samples show that this setup allows the detection of photons across an energy range of ∼300 eV. This covers simultaneously the emission lines of life-important elements like carbon, nitrogen and oxygen in a shot-based procedure.Keywords: RIXS; varied line spacing; liquid jet; XAS; XES; FEL.
1. Introduction
Chemical reactions are governed by the properties of the electronic structure. Information about the et al., 2011; Kotani & Shin, 2001). It has become an established method at modern light sources such as synchrotron radiation or free-electron laser facilities due to their continuous improvement and development which has led to a remarkable increase in coherence and short pulse width compared with previous light source facilities. The high enables a broad range of techniques to be utilized which have high photon consumption. With soft the unoccupied (X-ray absorption) and occupied (X-ray emission) atomic and molecular orbitals are probed. A special case is resonant inelastic X-ray scattering (RIXS), where the is set at a resonance.
reveals the chemical pathways of these reactions. Soft is a powerful method of studying the occupied and unoccupied molecular orbitals of systems with low (AmentMajor chemical reactions and biological processes are occurring in liquid environments. Most life-important elements are light elements with an ; Faubel et al., 1988). When using a liquid cell the sample is isolated from the vacuum by thin X-ray transparent windows, e.g. made from Si3N4. However, the windows may interact with the sample. Despite the very small window thickness, absorption of the incoming and outgoing radiation cannot be avoided. This results in photon losses and longer data acquisition times. An advantage of this method is the small sample volume needed to perform the experiment. This can also be a disadvantage though, as radiation damage may occur. The latter can be avoided using the liquid jet. No windows are needed. The sample is refreshed with a repetition rate up to several MHz, depending on the flow rate, nozzle size and X-ray spot size. Yet, there is a high consumption of the compound. This is a major disadvantage when using very precious samples.
in the soft X-ray regime, which requires experimenting under vacuum conditions. Probing direct electronic structures of samples in a liquid phase is an experimental challenge. Two concepts have been introduced to probe liquid samples in a vacuum environment: the liquid cell and liquid jet (Yang & Kirz, 1987For detection of et al., 1986; Nordgren et al., 1989). The Rowland-circle-based X-ray spectrometer offers the possibility to easily increase the energy resolution by detecting higher orders of diffraction at the cost of decreasing In recent years more spectrometers have been used based on the variable line spacing (VLS) design (Ghiringhelli et al., 2006; Harada et al., 2012; Chiuzbăian et al., 2014; Yin et al., 2015). Advantages are the reduced astigmatism and the focal plane geometry, allowing detectors to be utilized in normal incidence. Therefore, this leads to a larger energy range at one detector position. Currently, there exist a variety of RIXS setups for studying liquid samples (Blum et al., 2009; Lange et al., 2010; Tokushima et al., 2008; Kunnus et al., 2012).
an energy-dispersive detector is needed. The first pioneering work used a Rowland-circle-based grating spectrometer to detect soft X-rays (CallcottIn this work we present the novel experimental station, ChemRIXS, for chemical studies of liquid and solid samples utilizing RIXS. This new setup combines for the first time liquid jet technology with a soft X-ray spectrometer based on the VLS design. The advantages of windowless liquid samples in a vacuum coupled with a large energy detection range at sufficient energy resolution enable the investigation of the
of biochemical processes in their natural environment. The setup is also equipped with a recycling device to reduce high sample consumption of the liquid jet technology.In the following the details of the experimental endstation are described. In section 3, results from liquid and solid samples are presented. Very often, chemical analysis requires clear separation of superposing signals of complex systems. In order to emphasize such complexity in real-life samples, results from a mixture of relevant solvents in liquid phase and from samples with complex calgonite structures (solid) are presented.
2. Experimental setup
The endstation consists of a target chamber, the sample delivery system and a soft X-ray spectrometer based on the VLS principle (Yin et al., 2015). The VLS spectrometer is a flexible unit that can also be installed on other target chambers of existing setups. A target holder is available for the liquid jet as well as for solid samples. As solid samples do not present challenges for the vacuum conditions, we concentrate on describing how the vacuum conditions are met when using the liquid jet. The main challenge is to fulfil the vacuum conditions of the beamline and the spectrometer. When we describe the setup we use the z-axis for the height; x is the direction of the interacting synchrotron beam and y is perpendicular to the synchrotron radiation beam. The layout of the setup and its dimensions are illustrated in Fig. 1. The target chamber is made from a 250 mm tube. The central axis of the tube is aligned to y. One end of the tube is terminated by a DN250CF for pumping with a turbo-molecular pump (TMP) at a pumping speed of 1900 l s−1 (N2), the other by a DN100CF flange for attaching the detection device, in our case the VLS spectrometer.
On the top and bottom are DN100CF flanges. These are needed for an x-y-z manipulator and a liquid-nitrogen-cooled trap or the recycling system on the opposite side. Around the sample positions are eight DN40CF flanges for installing devices such as a photodiode to find the superposition of the incoming X-ray beam and liquid jet and a microscope to monitor the state of the sample. One challenge is to fulfil the ultrahigh-vacuum (UHV) conditions of the beamline which usually requires a vacuum of better than 10−8 mbar. The pressure in the main chamber depends on the nozzle size and the sample. Using a running jet with aqueous solutions, the vacuum is typically in the order of 10−4 mbar. The setup uses two differential pumping stages (DPSs) to fulfil the vacuum conditions. The first stage is attached directly to the target chamber via a narrow tube of length 5 cm with an elliptical of 8 mm width and height of 4 mm. Together with a TMP this allows a pressure difference of up to three orders of magnitude to be overcome. The second DPS is a combination of a plate with a 10 mm hole and a TMP which overcomes an additional order of magnitude and fulfils the UHV conditions of the beamline. An HPLC pump delivers the liquid sample through the nozzle into the chamber. The small nozzle aperture and fast flow prevent freezing of the liquid due to evaporative cooling.
Shortly below the nozzle opening is the laminar flow of the sample. Depending on the sample, flow rate and nozzle size, the laminar flow can reach up to several millimetres. After the laminar range the liquid breaks up into droplets (Faubel et al., 1988).
Besides probing the sample in a real liquid environment in a vacuum, the liquid jet offers a continuously refreshed sample and therefore prevents radiation damage, i.e. with a flow rate of 1 ml min−1 for a 20 µm nozzle and a beam spot size of 10 µm, a repetition rate >5 MHz can be reached. A 10 µm X-ray spot size is available at P04 at PETRA III (Viefhaus et al., 2013). A mixing device allows switching between different samples without breaking the vacuum and opening the setup. A recycling system can be installed so that precious samples can be reused. This system consists of a conical formed catcher with a 100 µm hole on top and a membrane pump for pumping the sample out of the catcher of the into a collecting tank. To avoid ice formation while hitting the catcher with aqueous solution, this part can be heated. The whole catcher system is on an x-y-z manipulator to align the recycling system (Charvat et al., 2004). Tests show that the sample consumption can be limited to around 40 ml for a running experiment of 24 h. Therefore, the sample can be reused and precious samples can be studied in small amounts. The time after interaction with the X-ray beam and recycling must be larger than the relaxations time of the system to avoid radiation damage, otherwise artefacts might influence the true signal.
A grazing-incident VLS-design-based soft X-ray spectrometer is used to detect the
after core excitation. To reach the optimum performance of the spectrometer the geometrical conditions must be met as precisely as possible; one requirement is a small source size. The spectrometer is equipped with a slit system with slit sizes ranging from 100 µm down to 1 µm in micrometre steps. Since the liquid jet has a small diameter, in the 10–25 µm range, a small source size can be realised without using the slit system which is also ideal for According to the focal simulation, derivation of the optimal incident angle leads to large discrepancies in the source–detector distance for different energies, while ray-tracing simulations show that small deviations in the source–grating distance are negligible. Ideally, the focal plane is correct at one distance for the whole energy range.The spectrometer is equipped with an ANDOR iKon L CCD detector with 2048 × 2018 pixels with each having a pixel size of 13.5 µm resulting in a total detection area of 27.6 mm × 27.6 mm. Owing to charge spreading, the effective pixel size is around 27 µm (Ghiringhelli et al., 2006). During the measurements the CCD was cooled down to 193 K to reduce thermal noise. Another advantage of this spectrometer is that it can be operated under relatively high vacuum conditions. It is implemented with a DPS between the slit system and the grating which can overcome about two orders of magnitude without losing photon intensity. Fig. 2 shows a schematic layout. A detailed description of the spectrometer can be found elsewhere (Yin et al., 2015).
An additional feature is that the spectrometer can be installed either with the grating in the plane of the synchrotron beam or in the perpendicular direction to improve the ).
depending on the dimensions of the focal spot size on the sample (see Fig. 2The apparatus is also equipped with a photodiode for performing X-ray absorption measurements in total ).
(see Fig. 1The total dimensions of the setup are 2168 mm in length and 690 mm in width. The width from the DPS flange to the middle of the target chamber is 511 mm and can be reduced to 430 mm. This compact design allows the experimental setup to be deployed at different synchrotron radiation and free-electron laser facilities. In the following section, we present experimental results from liquid and solid samples.
3. Results
The results of experiments using liquid and solid samples are presented to illustrate the capability of the setup. As samples, important organic solvents have been chosen. The beam spot was 300 µm × 10 µm for the liquid sample and 1 mm × 1 mm for the solid sample, to reduce radiation damage. For the liquid experiment a ∼25 µm nozzle was used. The beamline resolution was set at ∼2 eV.
3.1. Application of a liquid sample
In order to demonstrate the large energy detection range, Fig. 3 shows the X-ray emission signal for a 1:1 acetonitrile–ethanol mixture. It is probed using an of 540 eV in the liquid phase at an optimal detector position for the nitrogen emission signal. The total exposure time was set to 5 min with a ∼300 µm × 10 µm beam spot. X-ray emission lines of oxygen in first and second order and nitrogen and carbon in first order are observed. In total, with this configuration an energy range of ∼300 eV was covered.
The strongest signal derives from oxygen in first order followed by the second order and carbon and nitrogen signals. The nitrogen and carbon emission lines are the weakest. At 540 eV s electrons compensate the lower signal of this element.
the photons are absorbed mostly by the oxygen in ethanol which explains the strong line of oxygen in first and second order. The number of carbon atoms is higher than the number of nitrogen atoms but the higher cross sections of the nitrogen 1Compared with previous measurements at similar energy resolution (Dierker et al., 2013) the energy range is extended by a factor of about four. This offers the possibility to measure the characteristic emission lines of more than one element at the same time. In this case the three elements carbon, nitrogen and oxygen are shown in Fig. 3.
In order to show the possible energy resolution, the K-edge. Fig. 4 shows the X-ray emission (XE) spectrum of nitrogen for an of 400 eV which is the resonance of the core electron of nitrogen. The emission signal of ethanol vanished as expected due to the low absorption of oxygen.
is tuned to the nitrogenThe spectrum shows the emission signal from the 7a1 and 2e orbitals after core excitation at the nitrogen K-edge. The signal agrees with previously reported results (Dierker et al., 2013). The at 400 eV is determined according to the X-ray detected in total mode using a diode. A practical feature of P04 is its `on the fly' mode which allows X-ray absorption spectra to be recorded relatively quickly (Viefhaus et al., 2013).
3.2. Application to a solid sample
Fig. 5 shows the X-ray of an epitaxial Pr0.7Ca0.3MnO3 (PCMO) thin film excited on the manganese L-edge. To reduce the risk of radiation damage the X-ray beam spot was enlarged to ∼1 mm × 1 mm, which decreased the energy density on the sample at the cost of lower and lower energy resolution. The X-ray shows the oxygen K-emission and the manganese L-emission lines.
A 300 nm-thick film has been deposited by reactive ion beam sputtering on 200 nm of sputtered Pt on MgO single-crystal substrate (Mildner et al., 2015). PCMO shows strong electron–electron and electron–lattice correlations. The band structure near the comprises strongly hybridized bands containing Mn 3d and O 2p states. The shows the wide range of relaxations from Mn 3s to Mn 2p (about 560 eV) as well as from the O 2p–Mn 3d band to O 1s (O K at about 525 eV) and to Mn 2p (L3/2 at about 640–660 eV). of the beam is located at 642 eV.
4. Conclusion
In this work we present a novel endstation for RIXS studies of liquid and solid samples in the soft X-ray regime. ChemRIXS combines liquid jet technology with a soft X-ray VLS spectrometer and offers the advantages of real liquid samples under vacuum and the easy use of a soft X-ray spectrometer based on the VLS principle. ChemRIXS is also equipped with a photodiode to detect
spectra in total mode. The first experimental results of liquid and solid samples showed a working setup with sufficient resolution power. Compared with similar setups the spectrometer can cover an energy range of approximately 300 eV which is almost four times wider compared with the Rowland-circle-based spectrometer with comparable energy resolution. This enables measurement of various element edges using one-shot procedures, which is particularly interesting as a portable analysis endstation for investigation of chemical reactions at free-electron laser facilities.Acknowledgements
This work was supported by SFB755 `Nanoscale Photonic Imaging' and SFB1073 `Atomic Scale Control of Energy Conversion' project C02 of the German Science Foundation (DFG), the Advanced Study Group of the Max Planck Society, Deutsches Elektronen-Synchrotron. ST is grateful to the Funds of the Chemical Industry. We thank the staff of P04 and S. Klumpp for their continuous support.
References
Ament, L., van Veenendaal, M., Devereaux, T., Hill, J. & van den Brink, J. (2011). Rev. Mod. Phys. 83, 705–767. Web of Science CrossRef CAS Google Scholar
Blum, M., Weinhardt, L., Fuchs, O., Bär, M., Zhang, Y., Weigand, M., Krause, S., Pookpanratana, S., Hofmann, T., Yang, W., Denlinger, J. D., Umbach, E. & Heske, C. (2009). Rev. Sci. Instrum. 80, 123102. Web of Science CrossRef PubMed Google Scholar
Callcott, T. A., Tsang, K. L., Zhang, C. H., Ederer, D. L. & Arakawa, E. T. (1986). Rev. Sci. Instrum. 57, 2680. CrossRef Web of Science Google Scholar
Charvat, A., Lugovoj, E., Faubel, M. & Abel, B. (2004). Rev. Sci. Instrum. 75, 1209. Web of Science CrossRef Google Scholar
Chiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J.-M. (2014). Rev. Sci. Instrum. 85, 043108. Web of Science PubMed Google Scholar
Dierker, B., Suljoti, E., Atak, K., Lange, K. M., Engel, N., Golnak, R., Dantz, M., Hodeck, K., Khan, M., Kosugi, N. & Aziz, E. F. (2013). New J. Phys. 15, 093025. Web of Science CrossRef Google Scholar
Faubel, M., Schlemmer, S. & Toennies, J. P. (1988). Z. Phys. D, 10, 269–277. CrossRef CAS Google Scholar
Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006). Rev. Sci. Instrum. 77, 113108. Web of Science CrossRef Google Scholar
Harada, Y., Kobayashi, M., Niwa, H., Senba, Y., Ohashi, H., Tokushima, T., Horikawa, Y., Shin, S. & Oshima, M. (2012). Rev. Sci. Instrum. 83, 013116. Web of Science CrossRef PubMed Google Scholar
Kotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203–246. Web of Science CrossRef CAS Google Scholar
Kunnus, K., Rajkovic, I., Schreck, S., Quevedo, W., Eckert, S., Beye, M., Suljoti, E., Weniger, C., Kalus, C., Grübel, S., Scholz, M., Nordlund, D., Zhang, W., Hartsock, R. W., Gaffney, K. J., Schlotter, W. F., Turner, J. J., Kennedy, B., Hennies, F., Techert, S., Wernet, P. & Föhlisch, A. (2012). Rev. Sci. Instrum. 83, 123109. Web of Science CrossRef Google Scholar
Lange, K. M., Könnecke, R., Ghadimi, S., Golnak, R., Soldatov, M. A., Hodeck, K. F., Soldatov, A. & Aziz, E. F. (2010). Chem. Phys. 377, 1–5. Web of Science CrossRef CAS Google Scholar
Mildner, S., Hoffmann, J., Blöchl, P. E., Techert, S. & Jooss, C. (2015). Phys. Rev. B, 92, 035145. Web of Science CrossRef Google Scholar
Nordgren, J., Bray, G., Cramm, S., Nyholm, R., Rubensson, J.-E. & Wassdahl, N. (1989). Rev. Sci. Instrum. 60, 1690. CrossRef Web of Science Google Scholar
Tokushima, T., Harada, Y., Takahashi, O., Senba, Y., Ohashi, H., Pettersson, L. G. M., Nilsson, A. & Shin, S. (2008). Chem. Phys. Lett. 460, 387–400. Web of Science CrossRef CAS Google Scholar
Viefhaus, J., Scholz, F., Deinert, S., Glaser, L., Ilchen, M., Seltmann, J., Walter, P. & Siewert, F. (2013). Nucl. Instrum. Methods Phys. Res. A, 710, 151–154. Web of Science CrossRef CAS Google Scholar
Yang, B. X. & Kirz, J. (1987). Phys. Rev. B, 36, 1361–1364. CrossRef CAS Web of Science Google Scholar
Yin, Z., Peters, H. B., Hahn, U., Agåker, M., Hage, A., Reininger, R., Siewert, F., Nordgren, J., Viefhaus, J. & Techert, S. (2015). Rev. Sci. Instrum. 86, 093109. Web of Science CrossRef PubMed Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.