Simulating and optimizing compound refractive lens-based X-ray microscopes

CROSSMARK_Color_square_no_text.svg

aDepartment of Physics, Technical University of Denmark, 2800 Kgs Lyngby, Denmark, and bEuropean Synchrotron Radiation Facility, 71 Avenue des Martyrs, 38000 Grenoble, France
*Correspondence e-mail: husimo@fysik.dtu.dk

Edited by G. E. Ice, Oak Ridge National Laboratory, USA (Received 2 March 2016; accepted 24 December 2016; online 20 February 2017)

A comprehensive optical description of compound refractive lenses (CRLs) in condensing and full-field X-ray microscopy applications is presented. The formalism extends ray-transfer matrix analysis by accounting for X-ray attenuation by the lens material. Closed analytical expressions for critical imaging parameters such as numerical aperture, spatial acceptance (vignetting), chromatic aberration and focal length are provided for both thin- and thick-lens imaging geometries. These expressions show that the numerical aperture will be maximized and chromatic aberration will be minimized at the thick-lens limit. This limit may be satisfied by a range of CRL geometries, suggesting alternative approaches to improving the resolution and efficiency of CRLs and X-ray microscopes.

1. Introduction

Many recent advances in synchrotron X-ray imaging can be attributed to X-ray focusing optics (Ice et al., 2011[Ice, G. E., Budai, J. D. & Pang, J. W. L. (2011). Science, 334, 1234-1239.]). These optics may operate via three possible principles: (i) diffraction, such as in Fresnel zone plates (Kirz, 1974[Kirz, J. (1974). J. Opt. Soc. Am. 64, 301-309.]) and multilayer Laue lenses (Kang et al., 2006[Kang, H. C., Maser, J., Stephenson, G. B., Liu, C., Conley, R., Macrander, A. T. & Vogt, S. (2006). Phys. Rev. Lett. 96, 127401.]); (ii) total reflection, such as in Kirkpatrick–Baez (Kirkpatrick & Baez, 1948[Kirkpatrick, P. & Baez, A. V. (1948). J. Opt. Soc. Am. 38, 766-774.]) and Wolter (Wolter, 1952[Wolter, H. (1952). Ann. Phys. 445, 94-114.]) mirrors, and lobster-eye (Inneman et al., 1999[Inneman, A. V., Hudec, R., Pina, L. & Gorenstein, P. (1999). Proc. SPIE, 3766, 72-79.]) and Kumakhov (Kumakhov & Komarov, 1990[Kumakhov, M. A. (1990). Phys. Rep. 191, 289-350.]) lenses; and (iii) refraction, such as in prisms (Cederstrom et al., 2000[Cederström, B., Cahn, R. N., Danielsson, M., Lundqvist, M. & Nygren, D. R. (2000). Nature (London), 404, 951.]) and compound refractive lenses (CRLs) (Snigirev et al., 1996[Snigirev, A., Kohn, V. G., Snigireva, I. I. & Lengeler, B. (1996). Nature (London), 384, 49-51.]). In the hard X-ray regime (E > 15 keV), CRLs (linear arrays of refractive lenslets) are widely used due to their relatively low cost, ease-of-use and efficiency. Furthermore, their focal length can be actively varied by adjusting the number of lenslets (Vaughan et al., 2011[Vaughan, G. B. M., Wright, J. P., Bytchkov, A., Rossat, M., Gleyzolle, H., Snigireva, I. & Snigirev, A. (2011). J. Synchrotron Rad. 18, 125-133.]). However, the spatial resolution of CRL-based imaging systems is typically 50–100 nm (Schroer et al., 2005[Schroer, C. G., Kurapova, O., Patommel, J., Boye, P., Feldkamp, J., Lengeler, B., Burghammer, M., Riekel, C., Vincze, L., van der Hart, A. & Küchler, M. (2005). Appl. Phys. Lett. 87, 124103.]), worse than that of other optics at comparable energies: 7 nm (Yamauchi et al., 2011[Yamauchi, K., Mimura, H., Kimura, T., Yumoto, H., Handa, S., Matsuyama, S., Arima, K., Sano, Y., Yamamura, K., Inagaki, K., Nakamori, H., Kim, J., Tamasaku, K., Nishino, Y., Yabashi, M. & Ishikawa, T. (2011). J. Phys. Condens. Matter, 23, 394206.]), 8 nm (Morgan et al., 2015[Morgan, A. J., Prasciolu, M., Andrejczuk, A., Krzywinski, J., Meents, A., Pennicard, D., Graafsma, H., Barty, A., Bean, R. J., Barthelmess, M., Oberthuer, D., Yefanov, O., Aquila, A., Chapman, H. N. & Bajt, S. (2015). Sci. Rep. 5, 9892.]) and 20 nm (Vila-Comamala et al., 2012[Vila-Comamala, J., Pan, Y., Lombardo, J., Harris, W. M., Chiu, W. K., David, C. & Wang, Y. (2012). J. Synchrotron Rad. 19, 705-709.]) have been reported from microscopes based on mirrors, multilayer Laue lenses and Fresnel zone plates, respectively. Nonetheless, the advantages of CRLs make them uniquely suitable for in situ experiments where efficiency, large working distances and high X-ray energies are required. Under such circumstances, improving the spatial resolution of CRLs could facilitate new opportunities for multi-scale characterization.

One route to improving spatial resolution is by optimizing the CRL geometry (Chen et al., 2014[Chen, Z., Xie, H., Deng, B., Du, G., Jiang, H. & Xiao, T. (2014). Chin. Opt. Lett. 12, 123401.]). Numerical optimization requires concise analytical expressions for parameters such as focal length, transmission and aberration. Furthermore, these expressions are essential for imaging techniques that involve sampling data in grids such as ptychography (Schroer et al., 2008[Schroer, C. G., Boye, P., Feldkamp, J. M., Patommel, J., Schropp, A., Schwab, A., Stephan, S., Burghammer, M., Schöder, S. & Riekel, C. (2008). Phys. Rev. Lett. 101, 090801.]), scanning X-ray microscopy (Schroer et al., 2005[Schroer, C. G., Kurapova, O., Patommel, J., Boye, P., Feldkamp, J., Lengeler, B., Burghammer, M., Riekel, C., Vincze, L., van der Hart, A. & Küchler, M. (2005). Appl. Phys. Lett. 87, 124103.]) or dark-field X-ray microscopy (Simons et al., 2015[Simons, H., King, A., Ludwig, W., Detlefs, C., Pantleon, W., Schmidt, S., Snigireva, I., Snigirev, A. & Poulsen, H. F. (2015). Nat. Commun. 6, 6098.]). The optical theory of CRLs and CRL-based imaging systems has been addressed by various approaches such as ray-transfer matrices (RTMs) (Protopopov & Valiev, 1998[Protopopov, V. V. & Valiev, K. A. (1998). Opt. Commun. 151, 297-312.]; Pantell et al., 2003[Pantell, R. H., Feinstein, J., Beguiristain, H. R., Piestrup, M. A., Gary, C. K. & Cremer, J. T. (2003). Appl. Opt. 42, 719-723.]) [including Gaussian beam variants (Poulsen & Poulsen, 2014[Poulsen, S. O. & Poulsen, H. F. (2014). Met. Mater. Trans. A, 45, 4772-4779.])], Monte Carlo ray tracing (Sanchez~del Rio & Alianelli, 2012[Sanchez del Rio, M. & Alianelli, L. (2012). J. Synchrotron Rad. 19, 366-374.]), wavefront propagation methods (Kohn, 2003[Kohn, V. G. (2003). J. Exp. Theor. Phys. 97, 204-215.]) and others (Lengeler et al., 1999[Lengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153-1167.]). While these have greatly furthered the design and implementation of CRLs, no single formalism fulfills the core requirements for optimization: (i) simple, closed expressions, (ii) broad applicability to both condensing and full-field imaging systems, and (iii) consideration of both the thin-lens (where the focal length of the CRL far exceeds its length) and thick-lens conditions (where this approximation is no longer valid).

We present a formalism for CRL-based imaging systems utilizing an RTM approach to model archetypal X-ray imaging systems in a lens-by-lens manner, thus accounting for both thin- and thick-lens conditions. Attenuation by the lens material is calculated using RTMs to trace the ray position through the CRL. We provide exact analytical expressions for focal length, numerical aperture, spatial resolution, vignetting and chromatic aberration among other key optical parameters. These expressions form the basis of an efficient parametric optimization of the CRL and imaging geometry, which ultimately provides suggestions for future lens development routes.

2. RTM formalism for CRLs

2.1. Assumed CRL and lenslet geometry

This formalism assumes a one-dimensional (1D) focusing geometry valid for both axisymmetric two-dimensional and planar 1D CRLs. The CRL is comprised of N identical parabolic and non-kinofirm lenslets (Fig. 1[link]), each with radius of curvature R, aperture 2Y and center-to-center distance between successive lenslets T such that Y = (RT)1/2. For manufacturing reasons, lenslets have a small distance between the parabolic apices (i.e. a web thickness) Tweb that affects attenuation. There may also be a gap between adjacent lenslets, implying that the physical lenslet thickness Tphys is less than T. Such geometries limit Y and necessitate defining the physical aperture 2Yphys such that Yphys = [R(Tphys-Tweb)]1/2.

[Figure 1]
Figure 1
CRL and lenslet geometry assumed in this formalism. A single refracting lenslet element is shown in blue, annotated with symbolic dimensions.

2.2. Background to the RTM approach and focusing behavior

RTM analysis is a paraxial ray-tracing approach that assumes all rays propagate nearly parallel to the optical axis. It does not intrinsically consider diffraction and total reflection; however, these may be introduced ad hoc. The approach treats each photon as a ray with transverse position y and angle to the longitudinal optical axis α, within an optical system defined by a matrix [{\bf M}] (i.e. an RTM) that transforms an incident ray [(\,y_0,\alpha_0)] into an exit ray [(\,y_1,\alpha_1)],

[\left[\matrix{ y_1 \cr \alpha_1 }\right] = {\bf M} \left[\matrix{ y_0 \cr \alpha_0}\right]. \eqno(1)]

RTMs of compound systems may then be determined by multiplying the RTMs for the individual optical components. Previous RTM analyses of CRLs (Protopopov & Valiev, 1998[Protopopov, V. V. & Valiev, K. A. (1998). Opt. Commun. 151, 297-312.]; Pantell et al., 2003[Pantell, R. H., Feinstein, J., Beguiristain, H. R., Piestrup, M. A., Gary, C. K. & Cremer, J. T. (2003). Appl. Opt. 42, 719-723.]; Poulsen & Poulsen, 2014[Poulsen, S. O. & Poulsen, H. F. (2014). Met. Mater. Trans. A, 45, 4772-4779.]) show that a single refractive X-ray lenslet can be described by three such components: a free-space propagation by T/2, a refracting thin lens with focal length f = [R/(2\delta)] (where δ is the refractive decrement) and a final free-space propagation by T/2. Because f is many times larger than T, each lenslet behaves like an ideal thin lens with the following transfer matrix,

[{\bf M} = \left[\matrix{ 1 & T/2 \cr 0 & 1 }\right] \left[\matrix{ 1 & 0 \cr -1/f & 1 }\right] \left[\matrix{ 1 & T/2 \cr 0 & 1 }\right]. \eqno(2)]

As the CRL is a linear array (i.e. a stack) of identical lenslets, its transfer matrix is [{\bf M}^{\,N}] = [({\bf M})^N]. This can be calculated through the matrix eigendecomposition theorem (Poulsen & Poulsen, 2014[Poulsen, S. O. & Poulsen, H. F. (2014). Met. Mater. Trans. A, 45, 4772-4779.]) (see §S1 of the supporting information for derivation),

[{\bf M}^{\,N} = \left[\matrix{ \cos(N\varphi) & f\varphi\sin(N\varphi) \cr -\sin(N\varphi)/(\,f\varphi) & \cos(N\varphi) }\right]. \eqno(3)]

Within this paraxial treatment, the parameter φ can be expressed as [\varphi] = (T/f )1/2 = [Y/[\,f(2\delta)^{1/2}]]. Thus, [1/(\,f\varphi)] is the refractive power of the CRL per unit length (Lengeler et al., 1999[Lengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153-1167.]) while [(2\delta)^{1/2}] is the critical angle for total external reflection (Schroer & Lengeler, 2005[Schroer, C. G. & Lengeler, B. (2005). Phys. Rev. Lett. 95, 054802.]).

The trigonometric terms in equation (3)[link] imply periodicity with respect to [N\varphi]. While attenuation by the lens means that CRLs practically operate within the first half-period (i.e. [0 \leq N\varphi \leq \pi]), optical behavior differs markedly between the thin-lens limit (i.e. [NT \ll f_N] and correspondingly [N\varphi \ll 1]) and the general thick lens case (i.e. all values of [N\varphi]). This formalism provides both cases in order to give straightforward access to the most common and important optical parameters for the vast majority of CRL geometries.

From equation (3)[link], the focal length of the CRL as measured from its exit surface is given by the following two expressions (derived in §S2 of the supporting information), which are identical to those given in the literature (Poulsen & Poulsen, 2014[Poulsen, S. O. & Poulsen, H. F. (2014). Met. Mater. Trans. A, 45, 4772-4779.]; Lengeler et al., 1999[Lengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153-1167.]),

[f_N = \left\{ \matrix{ \, -M^{\,N}_{11}/M^{\,N}_{21} = f\varphi\cot(N\varphi) \hfill & \,{\rm{general\,\,case}},_{\vphantom{\Big|}} \hfill \cr \eqalign{ & f/N-NT/3 \cr& } \hfill & \eqalign{ & {\rm{for}}\,\,N\varphi\ll1\cr&\,\,{\rm{(thin\,\,lens).}} } \hfill } \right. \eqno(4)]

2.3. Ray transfer path

In order to predict the attenuation of the rays as they traverse the CRL, the RTM approach must be extended. Specifically, we require an expression for the position and angle of a given ray at the center of the nth lenslet [(\,y_n,\alpha_n)] as a function of its incident state [(\,y_0,\alpha_0)]. To this end, we compute the RTM of the CRL after the nth lenslet and back-propagate by T/2,

[\eqalignno{ \left[\matrix{ y_n \cr \alpha_n }\right] &= \left[\matrix{ 1 & -T/2 \cr 0 & 1 }\right] \,{\bf M}^{\,n} \left[\matrix{ y_0 \cr \alpha_0 }\right] \cr& = \left[\matrix{ M_{11}^{\,n} -{{T}\over{2}} M_{21}^{\,n} & M_{12}^{\,n} -{{T}\over{2}}_{\vphantom{\big|}} M_{22}^{\,n} \cr M_{21}^{\,n} & M_{22}^{\,n} }\right] \left[\matrix{ y_0 \cr \alpha_0 }\right]. &(5)}]

Inserting from equation (3)[link] and simplifying (see §S3 of the supporting information), [(\,y_n,\alpha_n)] are then

[\eqalign{ y_n & = \left[\,y_0^2+\left(\alpha_0\,f\varphi\right)^2\right]^{1/2}\cos\left[\!\left(n-{{1}\over{2}}\right)\varphi + \tan^{-1}\left( {{ \alpha_0\,f\varphi }\over{ y_0 }} \right)\!\right], \cr \alpha_n & = \left[\alpha_0^2+\left({{y_0}\over{f\varphi}}\right)^2\right]^{1/2} \sin\left[n\varphi+\tan^{-1} \left( {{ \alpha_0\,f\varphi }\over{ y_0 }} \right)\right]. } \eqno(6)]

Within the CRL, all rays have a sinusoidal trajectory that varies with distance nT with a period of [2\pi Y/(2\delta)^{1/2}]. The physical aperture of the lenslets Yphys bounds this trajectory however, imposing the following criteria for participation in the focusing process,

[\left|\,y_n\right| \le Y_{\rm{phys}} \,\,\,\forall\,\,n. \eqno(7)]

Rays may be excluded due to total reflection by the parabolic lenslet surfaces. In this case, the criteria for participation is [R/(2y_n)-\alpha_n] > [(2\delta)^{1/2}]. However, we note that, at typical X-ray energies, such reflection effects are only significant for lenslet geometries with impractically large values of T/R. As such, we do not consider them further.

2.4. Attenuation in CRLs

The attenuation U of a ray passing through a single lenslet at a distance y from the optical axis depends on the absorption coefficient μ of the lens material and the local material thickness t(y) = Tweb + y2/R. Since the paraxial approximation implies that the variation of y and t(y) within the lenslet is negligible, the attenuation of the X-rays by the absorbing lenslet can be simply expressed using the Beer–Lambert law,

[U(\,y) = \exp{(-\mu T_{\rm{web}})} \exp{\left({{-\mu y^2}\over{R}}\right)} \, H{\left({{y} \over {Y_{\rm{phys}}}}\right)}. \eqno(8)]

Here, H is a box function of width 2Yphys that enforces the criteria in equation (7)[link]. The cumulative transmission UN(yn) of a ray as it travels through N lenslets is then the product of the individual attenuation contributions from each lenslet,

[\eqalignno{ U_N(\,y_n) & = \prod_{n\,=\,1}^N\,U(\,y_n) &(9) \cr& = \exp{(-N\mu T_{\rm{web}})} \exp{\left({{-\mu}\over{R}} \sum_{n\,=\,1}^N y_{n}^2\right)} \, \prod_{n\,=\,1}^N H{\left(\,{{y_n}\over{Y_{\rm{phys}}}}\right)}.}]

The central expression [\textstyle\sum_{n\,=\,1}^Ny_n^{\,2}] is a geometric sum that can be solved analytically (see §S4 of the supporting information). As all yn are a linear function of [(\,y_0,\alpha_0)], UN has a Gaussian dependence on both parameters. Combined with the conditions imposed by H, this results in a bounded two-dimensional Gaussian transmission distribution, as shown in Fig. 2[link].

[Figure 2]
Figure 2
Transmission function [U_N(\,y_0,\alpha_0)] for a typical beryllium CRL at 17 keV with parameters N = 50, R = 50 µm, T = 2 mm, Yphys = 0.5 mm. The white region represents [(\,y_0,\alpha_0)] values excluded due to the constraints in equation (7)[link], which can be approximated by two pairs of dashed lines corresponding to the entrance (horizontal pair) and exit (slanted pair) of the CRL.

2.5. Effective aperture and transmission efficiency of a CRL

The spatial acceptance function for a homogeneous and parallel incident beam can then be calculated from [U_N(\,y_0,\alpha_0)] by inserting [\alpha_0] = 0 into equation (6)[link] (see §S5 of the supporting information). This results in a 1D Gaussian transmission profile in y0 with a root mean square (RMS) value [\sigma_D] of

[\sigma_D = \left\{ \matrix{ \,(R/\mu{N})^{1/2}\left[1+{\rm{sinc}}{(2N\varphi)}\right]^{-1/2} \hfill & {\rm{general\,\,case}},_{\vphantom{\bigg|}} \hfill \cr \eqalign{ & (R/2\mu{N})^{1/2}\left[1+(N\varphi)^2/6\right] \cr& } \hfill & \eqalign{ & \!\!\1{\rm{for\,\,}} N\varphi\ll1 \cr& \,\,{\rm{(thin\,\,lens).}} } }\right. \eqno(10)]

From this, we can calculate the effective aperture Deff: the diameter of a circular pinhole with the same total transmitted intensity as a (two-dimensional) CRL made from rotationally symmetric paraboloids. For all values of [N\varphi], we find

[D_{\rm{eff}} = 2\sqrt{2}\sigma_D \left[ 1-\exp\left(- {{ Y_{\rm{phys}}^{\,2} }\over{ 2\sigma_D^2 }} \right) \right]. \eqno(11)]

The transmission efficiency is then given by

[t = {{D_{\rm{eff}}} \over {2Y_{\rm{phys}}}}. \eqno(12)]

These expressions for [\sigma_D], Deff and t provide a convenient means to compare the attenuation-limited properties of CRLs independent of the optical system they are operating in. Coupled with the expression for fN [equation (4)[link]], they constitute the simplest way to characterize CRL performance.

3. CRL-based imaging systems

3.1. The imaging condition

The general imaging case describes both condensing (Schroer et al., 2005[Schroer, C. G., Kurapova, O., Patommel, J., Boye, P., Feldkamp, J., Lengeler, B., Burghammer, M., Riekel, C., Vincze, L., van der Hart, A. & Küchler, M. (2005). Appl. Phys. Lett. 87, 124103.]) and full-field (Lengeler et al., 1999[Lengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153-1167.]) geometries comprising a source, a lens (either as an objective or condenser) and an image/detection plane. Note that these geometries are mathematically identical; condensing can be seen as full-field imaging of the source with a magnification ratio of less than one (shown schematically in Fig. 3[link]).

[Figure 3]
Figure 3
Comparison of CRL-based imaging systems. Full-field transmission X-ray microscopy (rear) versus a condensing system that demagnifies a (typically Gaussian) X-ray source (front).

In an imaging configuration, a ray originating from the source plane at [(\,y_{\rm{s}},\alpha_{\rm{s}})] travels a distance d1 to the objective, where it is transformed by the CRL RTM [{\bf M}^{\,N}] before travelling a distance d2 to a point [(\,y_{\rm{d}},\alpha_{\rm{d}})] on the detector plane. This transformation between [(\,y_{\rm{s}},\alpha_{\rm{s}})] and [(\,y_{\rm{d}},\alpha_{\rm{d}})] can be expressed as

[\left[ \matrix{ y_{\rm{d}} \cr \alpha_{\rm{d}} }\right] = \left[\matrix{ 1 & d_2 \cr 0 & 1 }\right] \,{\bf M}^{\,N} \left[\matrix{ 1 & d_1 \cr 0 & 1 }\right] \left[\matrix{ y_{\rm{s}} \cr \alpha_{\rm{s}} }\right] \equiv {\bf K} \left[\matrix{ y_{\rm{s}} \cr \alpha_{\rm{s}} }\right], \eqno(13)]

where [{\bf K}] is the matrix

[{\bf K} = \left[\matrix{ K_{11} & K_{12} \cr K_{21} & K_{22} }\right] \eqno(14)]

whose components can be expressed in terms of [{\bf M}^{\,N}] as

[\eqalign{ K_{11} & = M_{11}^{\,N} + d_2 M_{21}^{\,N}, \cr K_{12} & = M_{12}^{\,N} + d_1 \left(M_{11}^{\,N}+d_2 M_{21}^{\,N}\right) + d_2 M_{22}^{\,N}, \cr K_{21} & = M_{21}^{\,N}, \cr K_{22} & = d_1 M_{21}^{\,N} + M_{22}^{\,N}.} \eqno(15)]

The imaging condition implies that K12 = 0, which leads to the general imaging formula (derivation in §S6 of the supporting information)

[{{1} \over {d_1}} + {{1} \over {d_2}} - {{1} \over {f_N}} + {{f\varphi\tan(N\varphi)} \over {d_1 d_2}} = 0. \eqno(16)]

The magnification of the imaging system [{\cal M}] (a positive number) is defined by [{\cal M}] = -yd/ys = -K11. Combined with equation (16)[link], this gives a set of two equations, which can be solved to give exact expressions for magnification [{\cal M}] and the imaging distances d1 and d2,

[{\cal M} = \left\{ \matrix{ \,(d_2/f\varphi)\sin(N\varphi)-\cos(N\varphi) \hfill & \,{\rm{general\,\,case,_{\vphantom{\Big|}}}} \hfill \cr \eqalign{ & (N\varphi)^2/2+d_2N/f-1 \cr& }\hfill & \eqalign{ & {\rm{for}}\,\,N\varphi\ll1\cr&\,{\rm{(thin\,\,lens),}}} \hfill } \right. \eqno(17)]

[d_1 = \left\{ \matrix{ \, f_N \left[1 + {{1}\over{{\cal M}\cos(N\varphi)}}\right]_{\vphantom{\bigg|}} \hfill & \,{\rm{general\,\,case}}, \hfill \cr \eqalign{ & (\,f/N)\left[1+{{1}\over{{\cal M}}}\right]-{{N T}/{2}} \cr& } \hfill & \eqalign{ & {\rm{if\,\,}}N\varphi\ll1\cr&\,\,{\rm{(thin\,\,lens)}},} \hfill } \right. \eqno(18)]

[d_2 = \left\{ \matrix{ f_N \left[1 + {{ {\cal M}}\over{\cos(N\varphi)}}\right]_{\vphantom{\bigg|}} \hfill & {\rm{general\,\,case}}, \hfill \cr f/N\left[1+{\cal M}\right]-{NT}/{2} \hfill & {\rm{if}}\,\, N\varphi\ll1\,\,{\rm{(thin\,\,lens).}} \hfill } \right. \eqno(19)]

Note that d1 and d2 can never be negative in equations (18)[link] and (19)[link], resulting in the following conditions on [N\varphi],

[\!\!\!0 \le N\varphi \le \left\{ \matrix{ \eqalign{ & \cos^{-1}\left({{-1}/{{\cal M}}}\right) \cr&}\hfill & \eqalign{ & {\rm{when\,\,}}{\cal M}\,\gt\,1 \cr& \,(i.e.\,\,{\rm{full}}\hbox{-}{\rm{field\,\,imaging}}),_{\vphantom{\big|}}} \hfill \cr \eqalign{ & \cos^{-1}\left(-{\cal M}\right) \cr&} \hfill & \eqalign{ & {\rm{when\,\,}} {\cal M}\,\lt\,1\cr&\,(i.e.\,\, {\rm{condensing}}). } \hfill } \right. \eqno(20)]

The number of lenses N necessary to achieve a given magnification [{\cal M}] and source-to-detector distance L = d1+d2+NT can be calculated by rewriting the imaging formula [equation (16)[link]] as follows,

[N = {{1}\over{\varphi}} \left[\!\sin^{-1}\left( {{ {\cal M} + ({1}/{{\cal M}}) }\over{ \left\{[(L-NT)/f\varphi]^2 + 4\right\}^{1/2} }} \right) + \tan^{-1}\left({{2f\varphi} \over {L-NT}}\right)\!\! \right]. \eqno(21)]

Despite a factor in the denominator, N can nonetheless be evaluated iteratively since [L \gg NT] (solution by fixed-point method). Furthermore, as N must be an integer number, d1 and d2 must be adjusted to fulfill the imaging condition. Consequently, a small deviation of the magnification [{\cal M}] from the target value must be accepted.

We observe that equation (13)[link] can also be used to describe the geometry of the back focal plane, which for these imaging systems is located at d2 = fN. The intensity distribution at the back focal plane is closely related to the Fourier transform of the object in the sample plane, and as such can be used as a means for quantifying micro- and nano-scale periodicity (Ershov et al., 2013[Ershov, P., Kuznetsov, S., Snigireva, I., Yunkin, V., Goikhman, A. & Snigirev, A. (2013). J. Appl. Cryst. 46, 1475-1480.]). Inserting d2 = fN into equations (13)[link]–(15)[link] gives K11 = 0 and

[y_{\rm{d}} = {{f\varphi}\over{\sin{(N\varphi)}}} \, \alpha_{\rm{s}}. \eqno(22)]

Hence, X-rays emerging from the sample plane at the angle αs will converge to position yd in the back focal plane.

3.2. Attenuation in imaging systems

The spatial and angular acceptance of the CRL are defining characteristics of imaging systems, jointly defined by the attenuation properties of the lens and the specific geometry of the system. To find expressions for them, we first use equation (6)[link] to provide yn as a function of a ray's position and angle in the source plane, [(\,y_{\rm{s}},\alpha_{\rm{s}})],

[\eqalignno{y_n = {}&\left(d_1\alpha_{\rm{s}}+y_{\rm{s}}\right) \cos\left[\left(n-{{1}\over{2}}\right)\varphi\right] \cr& +f\varphi\,\alpha_{\rm{s}}\sin\left[\left(n-{{1}\over{2}}\right)\varphi\right]. &(23)}]

Inserting this into UN [equation (9)[link]] gives the complete acceptance function of the imaging system, which can be rewritten as follows (derivation in §S7 of the supporting information),

[\eqalignno{ U_N\left(\,y_{\rm{s}},\alpha_{\rm{s}}\right) = {}& \exp\left(-N\mu T_{\rm{web}}\right) \exp\left[- {{ \left(\alpha_{\rm{s}}+\gamma y_{\rm{s}}\right)^2 }\over{ 2\sigma_{\rm{a}}^2 }} \right] \cr& \times\exp\left(-{{y_{\rm{s}}^2}\over{2\sigma_{\rm{v}}^2}}\right) H\left(\,y_{\rm{s}},\alpha_{\rm{s}}\right). &(24)}]

Notably, this is the product of a prefactor, a Gaussian with RMS [\sigma_{\rm{a}}] and offset coefficient γ describing the system's angular acceptance, another Gaussian with RMS [\sigma_{\rm{v}}] decribing the system's spatial acceptance (i.e. vignetting) and a box function [H(\,y_{\rm{s}},\alpha_{\rm{s}})] representing the system's physical aperture. This physical aperture Yphys imposes a sharp cut-off to the acceptances [see equation (7)[link]]. In the same manner as equation (9)[link], this is represented by [\prod_{n\,=\,1}^N H[\,y_n(\,y_{\rm{s}},\alpha_{\rm{s}})/Y]] which, from equation (23)[link], is well approximated by

[\eqalignno{ & H\left(\,y_{\rm{s}},\alpha_{\rm{s}}\right) = \cr& \left\{ \matrix{ H\left({{d_1\alpha_{\rm{s}}+y_{\rm{s}}}\over{Y_{\rm{phys}}}}\right) H\left\{ {{ \left[\left(d_1\alpha_{\rm{s}}+y_{\rm{s}}\right)^2+\left(\,f\alpha_{\rm{s}}\varphi\right)^2\right]^{1/2}}\over{Y_{\rm{phys}}}}\right\} \hfill & {\rm{general\,\,case}}, \hfill \cr H\left({{d_1\alpha_{\rm{s}}+y_{\rm{s}}}\over{Y_{\rm{phys}}}}\right) \hfill & {\rm{for\,\,}}N\varphi\ll1\,\,{\rm{(thin\,\,lens}}). \hfill } \right. &(25)}]

The angular acceptance, defined by its RMS [\sigma_{\rm{a}}] and offset [\gamma y_{\rm{s}}], describes the range of angles over which the lens collects radiation emitted from a point ys on the source plane and ultimately defines the theoretical image resolution achievable by the system. At any point in the field of view (i.e. at any ys), the system will have a Gaussian angular acceptance with RMS [\sigma_{\rm{a}}] given by

[\sigma_{\rm{a}} = \left\{ \matrix{ \eqalign{ & \left\{{{R}\over{\mu N[d_1^{\,2}+(\,f\varphi)^2]}}\right\}^{1/2}\cr&\times \left\{1+{{1}\over{N}}-{{1}\over{N\varphi}}\sin[(N+1)\varphi] \right. \cr&\left.\times\cos\left[(N-1)\varphi+2\tan^{-1}\left(\,{{d_1}\over{f\varphi}}\right)\right]\right\}_{\vphantom{\Bigg|}}^{-1/2} } \hfill & \!\!{\rm{general\,\,case}}, \hfill \cr \eqalign{ & \left\{ {{R}\over{\mu N\left[d_1^{\,2}\,+\,(\,f\varphi)^2\right]}}\right\}^{1/2} \left[{{6\,+\,(N\varphi)^2}\over{6\sqrt{2}}}\right] \cr& = \sigma_D\gamma } \hfill & \eqalign{ & \!\!\!{\rm{for\,\,}}N\varphi\ll1\cr&\,\,({\rm{thin\,\,lens}}).} \hfill } \right. \eqno(26)]

The offset coefficient γ describes the median angle accepted by the lens at a particular point on the sample plane ys. Notably, in the thin-lens case the characteristic distance [1/\gamma] is the weighted average of the imaging distance d1 and the period of the sinusoidal divided by [2\pi]. The full derivation for γ is provided in §S7 of the supporting information.

The spatial acceptance has RMS [\sigma_{\rm{v}}] and describes the reduction in brightness from the center of the optical axis towards its periphery, i.e. the maximum achievable field-of-view of the system. This is defined as the total acceptance of the system integrated across all incident angles [\alpha_{\rm{s}}]. The general expression for the RMS of the vignetting function is given by

[\sigma_{\rm{v}} = {{\delta}\over{\mu\sigma_{\rm{a}}}} \left[N^{\,2}\varphi^2-\sin^2(N\varphi)\right]^{-1/2}. \eqno(27)]

The leading term in the thin-lens limit is of the order [(N\varphi)^{-3}], meaning that the vignetting does not originate from the material attenuation at the thin-lens limit. Instead, the thin-lens vignetting function, Ivys), is defined by the physical aperture of the small stack of lenses,

[\eqalignno{ & I_{\rm{v}}(\,y_{\rm{s}}) = \cr& \left\{ \matrix{ \, 1 \hfill & \, {\rm{for}}\,\left|\,y_{\rm{s}}\right|\,\lt\,Y_{\rm{phys}},_{\vphantom{\Big|}} \hfill \cr \eqalign{ & 1-\textstyle{{Y_{\rm{phys}}-\left|\,y_{\rm{s}}\right|}\over{Y_{\rm{phys}}-\tan\left[\left(d_1\,+\,{{NT}\over{2}}\right)\tan^{-1}\left({{2Y}\over{NT}}\right)\right] }} \cr&}\hfill & \eqalign{ & {\rm{for}}\,\,Y_{\rm{phys}}\,\lt\,\left|\,y_{\rm{s}}\right|\,\lt\, \cr& \!\tan\left[\textstyle\left(d_1+{{NT}\over{2}}\right)\tan^{-1}\left({{2Y}\over{NT}}\right)\right],_{\vphantom{\Big|}} } \hfill \cr \, 0 \hfill & \, {\rm{otherwise}}. \hfill } \right. &(28)}]

The numerical aperture (NA) is a dimensionless number characterizing the range of angles accepted by the imaging system, and is therefore naturally related to [\sigma_{\rm{a}}]. The parameter is regularly used in the context of visible-light systems where this range is sharply defined by the physical aperture of the system. However, the Gaussian nature of the X-ray acceptance function of typical thick CRLs makes such approaches inappropriate. Instead, the definition used for Gaussian laser systems is used here (Born & Wolf, 1999[Born, M. & Wolf, E. (1999). Principles of Optics, 7th ed. Cambridge University Press.]), in which the NA at a given position ys is given in terms of the ray angle where the transmission drops to e-2,

[{\rm{NA}} = y\gamma + {{ \sigma_{\rm{a}} \left(4\sigma_{\rm{v}}^2-y_{\rm{s}}^2\right)^{1/2} }\over{ \sigma_{\rm{v}} }}. \eqno(29)]

Note that this reduces to NA = [2\sigma_{\rm{a}}] at the center of the field-of-view.

3.3. Spatial resolution

Analytical expressions for the spatial resolution of the magnified image can now be derived from the magnification [{\cal M}] and the angular acceptance [\sigma_{\rm{a}}] of the imaging system. Spatial resolution may be defined as the minimum distinguishable distance [\Delta y_{\rm{s}}] between two points at the source/sample plane (Born & Wolf, 1999[Born, M. & Wolf, E. (1999). Principles of Optics, 7th ed. Cambridge University Press.]). The degree of blurring of these points due to diffraction and the aberration inherent in lens-based optical systems is described by the point spread function (PSF), which can be calculated from the Fourier transform of the CRL pupil function (Born & Wolf, 1999[Born, M. & Wolf, E. (1999). Principles of Optics, 7th ed. Cambridge University Press.]). Neglecting aberration, the effective pupil function Py0) for the center of the image can be derived from equation (24)[link] by substituting ys = 0 and [\alpha_{\rm{s}}] = y0/d1, where y0 is the ray position at the CRL entrance (i.e. n = 0) (see §S8 of the supporting information),

[P(\,y_0) = \exp\left( {{-y_0^2 }\over{ 2\sigma_{\rm{a}}^2d_1^{\,2} }}\right) H\left({{y_0}\over{Y_{\rm{pup}}}}\right), \eqno(30)]

where Ypup [\equiv] [Y_{\rm{phys}}/\left[1+\left(\,f\varphi/d_1\right)^2\right]^{1/2}]. For a wavenumber k = [2\pi/\lambda] at X-ray wavelength λ, and ignoring constant prefactors, the point-spread intensity function at the source plane, PSF( ys), is then given by

[\eqalignno{ {\rm{PSF}}(\,y_{\rm{s}}) & = \left| \,\int\limits_{-\infty}^{\infty} P(\,y_0)\exp\left(-i\,{{k}\over{d_1}}\,y_0y_{\rm{s}}\right)\,{\rm{d}}y_0 \,\,\right|^2 \cr & = \exp\left(-k^2\sigma_{\rm{a}}^2y_{\rm{s}}^2\right) \left[{\rm{erf}}\left( {{ Y_{\rm{pup}}+ik\sigma_{\rm{a}}^2y_{\rm{s}} }\over{ \sqrt{2}\sigma_{\rm{a}} }} \right) \right. \cr & \quad \left.+ \,\,{\rm{erf}}\left( {{ Y_{\rm{pup}}-ik\sigma_{\rm{a}}^2y_{\rm{s}} }\over{ \sqrt{2}\sigma_{\rm{a}} }} \right)\right]^2. &(31)}]

The PSF has two components: a Gaussian with RMS σPSF = [1/(\sqrt{2}k\sigma_{\rm{a}})] and a complex term comprising two error functions representing the effect of the physical aperture Yphys. The relative contributions of these components are shown in Fig. 4[link]. Notably, in the absence of a physical aperture (i.e. if the CRL is limited by the attenuation of the lens material), only the Gaussian component of the PSF remains.

[Figure 4]
Figure 4
Transmission function (a) and PSF (b) for a typical beryllium CRL at 17 keV with parameters [{\cal M}] = 20, N = 50, R = 50 µm, T = 2 mm, Yphys = 0.15 mm. The solid black line corresponds to the cumulative response of the CRL, while the dotted red and blue lines correspond to the contributions from the material absorption and physical aperture, respectively.

In classical optical systems, the resolution is often defined by the Rayleigh criterion, where two PSFs are regarded distinguishable when the maximum of one PSF coincides with the first minimum of the other (Born & Wolf, 1999[Born, M. & Wolf, E. (1999). Principles of Optics, 7th ed. Cambridge University Press.]). However, this is inappropriate in the case of a Gaussian or near-Gaussian PSF such as for CRLs, where such a minimum may not be present. Instead, we propose that the resolution be defined by the separation distance [\Delta y_{\rm{s}}] between two PSFs corresponding to a contrast ratio of C (where C is small when the contrast is poor). Using equation (32)[link], this can be determined by numerically solving

[{{ {\rm{PSF}}(0)+{\rm{PSF}}(\Delta y_{\rm{s}}) }\over{ {\rm{PSF}}({\Delta y_{\rm{s}}}/{2})+{\rm{PSF}}(-{\Delta y_{\rm{s}}}/{2})}} = 1-C. \eqno(32)]

In the case of absorption-limited (i.e. Gaussian) CRLs, this gives a function in terms of λ, [\sigma_{\rm{a}}] and C,

[\Delta y_{\rm{d}} = \big[0.06905-0.1019\log(1-C)\big]^{1/2}({{\lambda}/{\sigma_{\rm{a}}}}). \eqno(33)]

The value of C necessary to distinguish details depends on the sampling statistics. In the case of low-intensity (e.g. dynamic) measurements, C should be greater than the equivalent for the Rayleigh criteria (approximately 0.26).

3.4. Chromatic aberration

Imaging with a wide energy bandwidth can be advantageous due to the significant increase in photon flux. The bandwidth is ultimately defined by the type of X-ray source and conditioning optics: the raw spectrum from an undulator (i.e. the pink beam) is typically of the order of [10^{-2}\Delta E/E], while monochromators can provide bandwidths from 10-2 to 10-4. As most conditioning schemes use a diffraction-based monochromator, the energy spectrum is typically Gaussian. As such, the spectrum can be defined around a nominal energy E0 and intensity I0, in terms of the energy perturbation , defined by E/E0 = [1+\varepsilon] and the RMS bandwidth [\sigma_{\rm{e}}],

[I_\varepsilon = I_0 {{1} \over {\sqrt{2\pi}\sigma_{\rm{e}}}} \exp{\left({{-\varepsilon^2} \over {2\sigma_{\rm{e}}^2}}\right)}. \eqno(34)]

Small energy perturbations alter δ from its nominal value [\delta_0] according to [\delta] [\approx] [\delta_0/(1+\varepsilon)^2]. Since the focal length of the lenslets depend on δ, CRLs are chromatic by nature. Under an ideal imaging condition at [\varepsilon] = 0, a ray departing from the center of the sample plane will strike the center of the detector plane, i.e. yd = ys = 0. At a slightly different photon energy, [\varepsilon \neq 0]; however, the same ray will be displaced in the detector plane, [y_{\rm{d}} \neq y_{\rm{s}}] = 0. The position, yd, at which an incident ray from ys = 0 strikes the detector, can be approximated by inserting the chromatic expression for δ above into equation (13)[link], and Taylor expanding to first order in ,

[\eqalignno{ y_{\rm{d}} &= \alpha_{\rm{s}} \left[\left(d_1+d_2\right)\cos(N\varphi) + \left(f\varphi - {{ d_1 d_2 }\over{ f\varphi }} \right) \sin(N\varphi)\right] \cr& = \varepsilon\alpha_{\rm{s}}d_{\rm{ch}}, &(35)}]

where dch is a distance term for the imaging system given in terms of the nominal values for φ and f by

[\eqalignno{ d_{\rm{ch}} = {}& N\varphi_0 \left(\, {{ d_1 d_2 }\over{ f_0\varphi_0 }} - f_0\varphi_0\right) \cos\left(N\varphi_0\right) \cr& + \left[ \,{{ d_1 d_2 }\over{ f_0\varphi_0 }} +f_0\varphi_0+N\varphi_0\left(d_1+d_2\right)\right] \sin\left(N\varphi_0\right). &(36)}]

The ray is attenuated depending on its incident angle [\alpha_{\rm{s}}]. Noting from equation (35)[link] that this angle may be written as [\alpha_{\rm{s}}] = [y_{\rm{d}}/(d_{\rm{ch}}\varepsilon)], the spatio-chromatic intensity distribution of the rays on the detector plane, [I_{\rm{d}}(\varepsilon,y_{\rm{d}})], is then determined from equations (31)[link] and (34)[link] as (see §S9 of the supporting information)

[I_{\rm{d}}(\varepsilon,y_{\rm{d}}) = I_0 {{ 1 }\over{ \sqrt{2\pi}\sigma_{\rm{e}} }} \exp\left( {{ -\varepsilon^2 }\over{ 2\sigma_{\rm{e}}^2 }} \right) \exp\left[ {{ -y_{\rm{d}}^2 }\over{ 2(\sigma_{\rm{a}}d_{\rm{ch}}\varepsilon)^2 }} \right]. \eqno(37)]

This distribution (Fig. 5[link]) is consistent with experimental results in the literature (Falch et al., 2016[Falch, K. V., Detlefs, C., Di Michiel, M., Snigireva, I., Snigirev, A. & Mathiesen, R. H. (2016). Appl. Phys. Lett. 109, 054103.]) and illustrates that the chromatic spread of intensity becomes broader as deviates further from zero.

[Figure 5]
Figure 5
Spatio-chromatic intensity distribution for a typical beryllium CRL at 17 keV with parameters [{\cal M}] = 20, N = 50, R = 50 µm, T = 2 mm, Yphys = 0.5 mm at 17 keV (a). Point-spread function of the same Be CRL and configuration, assuming a bandwidth of [\sigma_{\rm{e}}] = 10-3 (b).

The point-spread function as a result of this chromatic behavior, PSFch, is determined by integrating Id across , and normalizing yd by the nominal magnification [{\cal M}],

[\eqalignno{ {\rm{PSF}}_{\rm{ch}} & = I_0 {{ 1 }\over{ \sqrt{2\pi}\sigma_{\rm{e}} }} \int\limits_{-\infty}^{\infty} \exp\left( {{ -\varepsilon^2 }\over{ 2\sigma_{\rm{e}}^2 }} \right) \exp\left( {{ -y_{\rm{s}}^{2} }\over{ 2\left(\sigma_{\rm{a}}d_{\rm{ch}}{\cal M}\varepsilon\right)^2 }} \right)\,{\rm{d}}\varepsilon \cr & = I_0\exp\left( {{ -\left| y_{\rm{s}} \right| }\over{ \sigma_{\rm{e}}\sigma_{\rm{a}}{\cal M}d_{\rm{ch}} }} \right). &(38)}]

Notably, this is a Laplace distribution described by the characteristic width [\sigma_{\rm{ch}}],

[\sigma_{\rm{ch}} = \left\{ \matrix{ \sigma_{\rm{a}}\sigma_{\rm{e}}{\cal{M}}d_{\rm{ch}}\hfill & {\rm{general\,\,case,_{\vphantom{\big|}}}} \hfill \cr \sigma_{\rm{a}}\sigma_{\rm{e}}{\cal{M}}(2d_1 d_2N/f\,) \hfill & {\rm{for\,\,}} N\varphi\ll1\,\,({\rm{thin\,\,lens}}). \hfill } \right. \eqno(39)]

The combined PSF from a chromatic and diffraction-limited system will be a convolution of the two PSFs in equations (31)[link] and (38)[link], shown in Fig. 5[link]. While this does not have a convenient analytical expression, it can be readily solved by numerical methods.

4. Optimization and the thick-lens limit

The analytical expressions for key optical parameters such as spatial resolution, vignetting and aberration provide a foundation for numerical optimization. In the most general case, the aim is to optimize some figure-of-merit function with respect to X-ray wavelenth (λ), CRL geometry (R, T, N) and imaging geometry (d1, [{\cal M}] or L, [{\cal M}]) for a given material (δ, μ).

In imaging systems, optimizing NA and [\sigma_{\rm{a}}] can improve the diffraction- and chromatic-limited resolution [equations (31)[link] and (38)[link]] while increasing transmission efficiency. Because [\sigma_{\rm{a}}] is directly coupled to [\sigma_{\rm{v}}] [equation (27)[link]], it also determines the image vignetting profile. The multidimensional problem of optimizing [\sigma_{\rm{a}}] for a given wavelength can, to a good approximation, be greatly simplified by assuming that N is large, i.e. N+1 [\approx] N-1 [\approx] N. Then equation (26)[link] becomes

[\sigma_{\rm{a}} = \left( {{ \delta }\over{ \mu NT}} \right)^{1/2} g(N\varphi,{\cal M}), \eqno(40)]

where [g(N\varphi,{\cal M})] is the expression

[\eqalignno{ g(N\varphi,{\cal M}) = {}& \left[ {{ 1 }\over{ 2 }} \left\{ 1+\left[ {{ \cos(N\varphi)+1/{\cal M} }\over{ \sin(N\varphi) }} \right]^2 \right\} \right. \cr& \left. \times \bigg( 1-{\rm{sinc}} (N\varphi) \right. \cr& \left. \times \cos \left\{ N\varphi+2\tan^{-1} \left[ {{ \cos(N\varphi)+1/{\cal M} }\over{ \sin(N\varphi) }} \right] \right\} \bigg) \right]^{-1/2}. \cr&&(41)}]

Equation (40)[link] confirms the well known belief that optical performance (i.e. [\sigma_{\rm{a}}]) is maximized when [\delta/\mu] is large (Snigirev et al., 1996[Snigirev, A., Kohn, V. G., Snigireva, I. I. & Lengeler, B. (1996). Nature (London), 384, 49-51.]). However, the expression also dictates that [g(N\varphi,{\cal M})] be maximized too. Plotting [g(N\varphi,{\cal M})] for the allowable range of [0\,\lt\,N\varphi\,\lt\,\cos^{-1}{(-1/{\cal M})}] and some typical [{\cal M}] values (Fig. 6[link]) shows that g reaches a maximum value at large [N\varphi].

[Figure 6]
Figure 6
Value of [g(N\varphi,{\cal M})] for different values of [{\cal M}].

Correspondingly, there is a global optimum [\sigma_{\rm{a}}] at [N\varphi] = [\cos^{-1}{(-1/{\cal M})}]. For typical values of [{\cal M}], this thick-lens limit is approximately at [N\varphi] = [\pi/2], and is associated with some unusual imaging geometries,

[\eqalign{ &f_N = 0, \qquad d_1 = {{Y}\over{{\cal M}\sqrt{2\delta}}}, \qquad d_2 = {{Y{\cal M}}\over{\sqrt{2\delta}_{\vphantom{\big|}}}}, \cr& {{d_2} \over {d_1}} = {\cal M}^{\,2}, \qquad L = {{Y({\cal M}+1/{\cal M})} \over {\sqrt{2\delta}}}.} \eqno(42)]

As discussed, the angular acceptance [\sigma_{\rm{a}}] reaches its maximum at the thick-lens limit close to [N\varphi] = [\pi/2]. As the magnification approaches infinity, [\sigma_{\rm{a}}], [\sigma_{\rm{v}}] and [\sigma_c] are

[\eqalign{ & \sigma_{\rm{a}} = \delta \left({{\pi} \over {\mu Y}}\right)^{1/2}, \qquad \sigma_{\rm{v}} = {{8} \over {\left[Y \mu \pi(\pi^2-4)\right]^{1/2} }}, \cr& \sigma_c = \left({{Y\delta\pi}\over{2\mu}}\right)^{1/2} \,\,\left[{{\pi}\over{2}} \left({\cal M}^2+1\right)+2{\cal M}\right].} \eqno(43)]

The first of these expressions show that the lenslet aperture Y is the critical geometrical term that fundamentally limits the optimal angular acceptance of CRL-based imaging systems. Similarly, Y is also a fundamental term for the vignetting profile and the chromatic spread. These expressions clearly demonstrate the need to miniaturize the lenslet geometry, as reducing Y is the only path to increasing the acceptance for a given set of [\delta,\mu] parameters. Most importantly, the fact that Y = [\sqrt{RT}] implies that there are a range of lenslet geometries that will satisfy the following conditions for optimization,

[N\left({{2 \delta T} \over {R}}\right)^{1/2} = {{\pi} \over {2}}, \qquad Y = \sqrt{RT} \le {{\pi \ln(2) \delta} \over {\mu}}. \eqno(44)]

5. Discussion

Following previous RTM formalisms (Protopopov & Valiev, 1998[Protopopov, V. V. & Valiev, K. A. (1998). Opt. Commun. 151, 297-312.]; Pantell et al., 2003[Pantell, R. H., Feinstein, J., Beguiristain, H. R., Piestrup, M. A., Gary, C. K. & Cremer, J. T. (2003). Appl. Opt. 42, 719-723.]; Poulsen & Poulsen, 2014[Poulsen, S. O. & Poulsen, H. F. (2014). Met. Mater. Trans. A, 45, 4772-4779.]), this implementation is shown to be a versatile and effective tool for many aspects of the optimization of CRL-based X-ray imaging systems. It is particularly suitable for calculating optical parameters for systems operating in the thick-lens condition. Being a linear formalism, the approach does not account for spherical or higher-order aberrations; however, some effects of diffraction and refraction can be included. Similarly, CRL geometries where the lenslet profile varies along the CRL thickness direction [i.e. adiabatic lenses (Schroer & Lengeler, 2005[Schroer, C. G. & Lengeler, B. (2005). Phys. Rev. Lett. 95, 054802.]; Chen et al., 2014[Chen, Z., Xie, H., Deng, B., Du, G., Jiang, H. & Xiao, T. (2014). Chin. Opt. Lett. 12, 123401.])] are not intrinsically accounted for, but can be calculated numerically by substituting a function of, for example, r(n) or T(n) for R or T, respectively. In the case of astigmatic lenses such as two-dimensional CRLs produced by interdigitating 1D chips (Simons et al., 2016[Simons, H., Stöhr, F., Michael-Lindhard, J., Jensen, F., Hansen, O., Detlefs, C. & Poulsen, H. F. (2016). Opt. Commun. 359, 460-464.]), the RTM formalism can be readily extended to either two 2×2 systems or a single 4×4 system that can calculate the astigmatism analytically.

The optimization demonstrated in this work was largely unconstrained and thus represents a simplified imaging configuration. Applying the practical constraints of real beamline hutch geometries therefore requires additional constraints to the optimization. However, a significant advantage of the RTM formalism is its versatility and mathematical simplicity compared with, for example, Monte Carlo and wave propagation methods. As such, the number of free parameters could be increased without necessarily resulting in impractical computation times.

A major result of this work is that the spatial resolution is globally optimized at the thick-lens limit: [N\varphi] = [\cos^{-1}{(-1/{\cal M})}] [\approx] [\pi/2]. This can be observed directly in Fig. 7[link], where [\sigma_{\rm{a}}] is maximized and [\sigma_{\rm{ch}}] is minimized at this limit. Furthermore, the field-of-view and imaging geometries also remain practical at this thick-lens limit: [\sigma_{\rm{v}}] is above 200 µm, L is 4 m and d1 is 10 mm at the optimum [\sigma_{\rm{a}}] in the example presented in Fig. 7[link].

[Figure 7]
Figure 7
Critical optical parameters as a function of [N\varphi] for a typical imaging system with [{\cal M}] = 20, R = 50 µm, T = 2 mm, Yphys = 0.5 mm at 17 keV with [\sigma_{\rm{e}}] = 10-3. Shown are the sample–objective and total imaging distances (a), the CRL focal length and transmission efficiency (b), the angular acceptance and vignetting (c) and chromatic aberration width (d).

The majority of CRL microscopes described in the literature (Simons et al., 2015[Simons, H., King, A., Ludwig, W., Detlefs, C., Pantleon, W., Schmidt, S., Snigireva, I., Snigirev, A. & Poulsen, H. F. (2015). Nat. Commun. 6, 6098.]) operate quite far from this optimimum configuration, implying that there may be significant resolution gains from increasing the focal power of CRLs beyond their current state. This is particularly true in the case of pink-beam X-ray imaging, where the chromatic blurring is minimized at the thick-lens limit.

The formalism indicates that smaller R and T will offer superior performance, making a strong case for miniaturizing the lenslet geometry. However, the optimization also shows that the global maximum for [\sigma_{\rm{a}}] can be reached by any lens geometry that satisfies equation (44)[link]. Thus, significant gains can still be had by reducing R or T alone. This is particularly important when manufacturing CRLs, as certain processes may prevent the realisation of specific lenslet geometries. For example, the performance of a CRL comprising indented two-dimensional metal lenslets may be improved by reducing the thickness of the lenslet rather than reducing the parabola radius, thus avoiding the shape-error and aberration that usually accompanies the production of small radii.

This formalism and optimization approach are key for the design of X-ray imaging experiments and instruments. The global optimum at small imaging distances implies that imaging systems with a small footprint will be optically superior. Such systems can offer improved mechanical stability and ultimately higher spatial resolution, and are therefore recommended on the basis of this work. The calculations performed here also shed light on the expected gains from upcoming fourth-generation synchrotron sources, which are characterized by unprecedented brilliance and a smaller energy bandwidth. As demonstrated in Fig. 7[link], the contribution of chromatic aberration to the PSF is expected to significantly decrease for lenses whose geometry approaches the thick-lens limit. We therefore predict that high-speed three-dimensional X-ray microscopy may be possible without the use of a monochromator.

6. Conclusions

This work describes a means of calculating and optimizing the optical properties of CRLs and some of the most common CRL-based imaging systems. Specifically:

(i) We developed a formalism with closed analytical expressions for key optical parameters, such as focal length, imaging distances, vignetting, spatial resolution and chromatic aberration. The expressions are relevant to the vast majority of X-ray microscopes, and pertain to both full-field imaging and condensing systems as well as thin- and thick-lens conditions.

(ii) We carried out an efficient global optimization on the archetypal X-ray imaging system. While this example was geometrically unconstrained, we note that practical limits to, for example, imaging distances can be easily incorporated.

(iii) The optimization identified that the optimum spatial resolution for any material and energy will occur at the thick-lens limit of [N\varphi] = [\cos^{-1}{(-1/{\cal M})}]. Hence, manufacturing CRLs at this thick-lens limit is an opportunity for resolution enhancement.

(iv) This implies that the optimum resolution may be reached with many different CRL geometries. This creates a significant opportunity for tailoring CRL geometries for specific manufacturing processes, e.g. lithography or indentation, etc.

(v) Chromatic aberration is reduced near this thick-lens limit. This means that larger energy bandwidths can be used with thick lenses, potentially increasing imaging flux and providing a new opportunity for high-speed dynamic imaging experiments.

Ultimately, we hope that the expressions and optimization approach described here can be applied to improve the performance and design of full-field X-ray microscopes. For existing lens materials and manufacturing technology such as the example given in Fig. 7[link], one can expect such optimizations to yield improvements in numerical aperture and spatial resolution by a factor of two or more, while simultaneously allowing greater X-ray energy bandwidth and flux. With the advent of many new instruments, both for the laboratory and in synchrotrons, we believe this capability is of significant contemporary relevance.

Supporting information


Acknowledgements

We thank Frederik Stöhr and Ray Barrett for useful discussions. In addition, we are grateful to the ESRF for providing beam time on ID06 and Danscatt for travel funding. HFP and SRA acknowledge support from the ERC grant `d-TXM'. HS acknowledges support from a DFF-FTP individual postdoc grant.

References

First citationBorn, M. & Wolf, E. (1999). Principles of Optics, 7th ed. Cambridge University Press.  Google Scholar
First citationCederström, B., Cahn, R. N., Danielsson, M., Lundqvist, M. & Nygren, D. R. (2000). Nature (London), 404, 951.  Google Scholar
First citationChen, Z., Xie, H., Deng, B., Du, G., Jiang, H. & Xiao, T. (2014). Chin. Opt. Lett. 12, 123401.  Google Scholar
First citationErshov, P., Kuznetsov, S., Snigireva, I., Yunkin, V., Goikhman, A. & Snigirev, A. (2013). J. Appl. Cryst. 46, 1475–1480.  CrossRef CAS IUCr Journals Google Scholar
First citationFalch, K. V., Detlefs, C., Di Michiel, M., Snigireva, I., Snigirev, A. & Mathiesen, R. H. (2016). Appl. Phys. Lett. 109, 054103.  CrossRef Google Scholar
First citationIce, G. E., Budai, J. D. & Pang, J. W. L. (2011). Science, 334, 1234–1239.  Web of Science CrossRef CAS PubMed Google Scholar
First citationInneman, A. V., Hudec, R., Pina, L. & Gorenstein, P. (1999). Proc. SPIE, 3766, 72–79.  CrossRef Google Scholar
First citationKang, H. C., Maser, J., Stephenson, G. B., Liu, C., Conley, R., Macrander, A. T. & Vogt, S. (2006). Phys. Rev. Lett. 96, 127401.  Web of Science CrossRef PubMed Google Scholar
First citationKirkpatrick, P. & Baez, A. V. (1948). J. Opt. Soc. Am. 38, 766–774.  CrossRef PubMed CAS Web of Science Google Scholar
First citationKirz, J. (1974). J. Opt. Soc. Am. 64, 301–309.  CrossRef CAS Web of Science Google Scholar
First citationKohn, V. G. (2003). J. Exp. Theor. Phys. 97, 204–215.  Web of Science CrossRef CAS Google Scholar
First citationKumakhov, M. A. (1990). Phys. Rep. 191, 289–350.  CrossRef CAS Web of Science Google Scholar
First citationLengeler, B., Schroer, C., Tümmler, J., Benner, B., Richwin, M., Snigirev, A., Snigireva, I. & Drakopoulos, M. (1999). J. Synchrotron Rad. 6, 1153–1167.  Web of Science CrossRef IUCr Journals Google Scholar
First citationMorgan, A. J., Prasciolu, M., Andrejczuk, A., Krzywinski, J., Meents, A., Pennicard, D., Graafsma, H., Barty, A., Bean, R. J., Barthelmess, M., Oberthuer, D., Yefanov, O., Aquila, A., Chapman, H. N. & Bajt, S. (2015). Sci. Rep. 5, 9892.  CrossRef PubMed Google Scholar
First citationPantell, R. H., Feinstein, J., Beguiristain, H. R., Piestrup, M. A., Gary, C. K. & Cremer, J. T. (2003). Appl. Opt. 42, 719–723.  CrossRef PubMed Google Scholar
First citationPoulsen, S. O. & Poulsen, H. F. (2014). Met. Mater. Trans. A, 45, 4772–4779.  CrossRef CAS Google Scholar
First citationProtopopov, V. V. & Valiev, K. A. (1998). Opt. Commun. 151, 297–312.  CrossRef CAS Google Scholar
First citationSanchez del Rio, M. & Alianelli, L. (2012). J. Synchrotron Rad. 19, 366–374.  Web of Science CrossRef IUCr Journals Google Scholar
First citationSchroer, C. G., Boye, P., Feldkamp, J. M., Patommel, J., Schropp, A., Schwab, A., Stephan, S., Burghammer, M., Schöder, S. & Riekel, C. (2008). Phys. Rev. Lett. 101, 090801.  Web of Science CrossRef PubMed Google Scholar
First citationSchroer, C. G., Kurapova, O., Patommel, J., Boye, P., Feldkamp, J., Lengeler, B., Burghammer, M., Riekel, C., Vincze, L., van der Hart, A. & Küchler, M. (2005). Appl. Phys. Lett. 87, 124103.  Web of Science CrossRef Google Scholar
First citationSchroer, C. G. & Lengeler, B. (2005). Phys. Rev. Lett. 95, 054802.  CrossRef PubMed Google Scholar
First citationSimons, H., King, A., Ludwig, W., Detlefs, C., Pantleon, W., Schmidt, S., Snigireva, I., Snigirev, A. & Poulsen, H. F. (2015). Nat. Commun. 6, 6098.  Web of Science CrossRef PubMed Google Scholar
First citationSimons, H., Stöhr, F., Michael-Lindhard, J., Jensen, F., Hansen, O., Detlefs, C. & Poulsen, H. F. (2016). Opt. Commun. 359, 460–464.  CrossRef CAS Google Scholar
First citationSnigirev, A., Kohn, V. G., Snigireva, I. I. & Lengeler, B. (1996). Nature (London), 384, 49–51.  CrossRef CAS Google Scholar
First citationVaughan, G. B. M., Wright, J. P., Bytchkov, A., Rossat, M., Gleyzolle, H., Snigireva, I. & Snigirev, A. (2011). J. Synchrotron Rad. 18, 125–133.  Web of Science CrossRef IUCr Journals Google Scholar
First citationVila-Comamala, J., Pan, Y., Lombardo, J., Harris, W. M., Chiu, W. K., David, C. & Wang, Y. (2012). J. Synchrotron Rad. 19, 705–709.  Web of Science CrossRef IUCr Journals Google Scholar
First citationWolter, H. (1952). Ann. Phys. 445, 94–114.  CrossRef Google Scholar
First citationYamauchi, K., Mimura, H., Kimura, T., Yumoto, H., Handa, S., Matsuyama, S., Arima, K., Sano, Y., Yamamura, K., Inagaki, K., Nakamori, H., Kim, J., Tamasaku, K., Nishino, Y., Yabashi, M. & Ishikawa, T. (2011). J. Phys. Condens. Matter, 23, 394206.  Web of Science CrossRef PubMed Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.