A method to stabilize the incident X-ray energy for anomalous diffraction measurements
aSchool of Science, Qilu University of Technology, Changqing District, Jinan, People's Republic of China, bBeijing Synchrotron Radiation Facility, Institute of High Energy Physics, Chinese Academy of Sciences, People's Republic of China, and cSchool of Life Science, University of Science and Technology of China, Hefei 230026, People's Republic of China
*Correspondence e-mail: dongyh@ihep.ac.cn, liup@ihep.ac.cn
A method to calibrate and stabilize the incident X-ray energy for anomalous diffraction data collection is provided and has been successfully used at the single-crystal diffraction beamline 1W2B at the Beijing Synchrotron Radiation Facilities. Employing a feedback loop to control the movement of the double-crystal monochromator, this new method enables the incident X-ray energy to be kept within a 0.2 eV range at the inflection point of the absorption edge.
Keywords: energy stability; monochromator; anomalous diffraction; feedback; absorption edge..
1. Introduction
Anomalous diffraction data, collected using an X-ray energy in the vicinity of an et al., 1990). In order to maximize the anomalous signal, the peak point, which maximizes the imaginary component of the factor f′′, and the inflection point, where the minimum f′ is reached, are often selected for data collection (González, 2003; De la Fortelle & Bricogne, 1997). However, anomalous diffraction data collected at these two points are highly sensitive to energy instability. A small drift of energy can cause a significant change in the of heavy atoms. For example, at the K-edge of selenium a 5.6 eV increase in the incident X-ray energy results in a 5.4× rise in f′′ (varies from 1.2 to 6.3).
are routinely utilized to determine the positions of heavy atoms, and thereafter to solve the in the protein procedure (HendricksonIt is known that after hkl and its Bijvoet mate are no longer equal, and the square of their deviation [] is used as the for heavy-atom position calculations. F(+) and F(−) always appear in different diffraction patterns, so they are affected differently by the energy fluctuation. Thus, energy instability can cause errors in the and may lead to a failure in heavy-atom position calculation and structural determination.
the intensities of a reflectionThere are many methods to reduce the error in . Hendrickson et al. (1990) showed that symmetry-equivalent measurements of F+ and F− can be made simultaneously on the same image if a crystal is aligned in a certain way. It is also feasible to collect data of F+ and F− in adjacent images. That is to say, the orientation of the crystal is rotated in a 0°–180°–1°–181°–2°–182° way. The two methods above allow systematic errors of to be balanced to some extent during data collection.
Another way is to keep the energy of the incident X-rays stabilized during data collection. Evans & Pettifer (1996) used an apparatus to stabilize the incident X-ray energy that employed a back diffraction comparison method of a silicon crystal. In this paper an alternative technique exploiting the steep of the heavy element is provided; it was successfully used at the single-crystal diffraction beamline 1W2B at the Beijing Synchrotron Radiation Facility (BSRF).
2. Description of the method
The basic design concept is based on the feedback amplification principle. As shown in Fig. 1, the absorption factor (μ) is highly sensitive to energy stability at the edge of the Therefore, the can be used to monitor the incident X-ray energy. At a reference point (E0, μ0), an increase or decrease of energy by ΔE can be detected as a rise or fall of the absorption factor by Δμ1 or Δμ2, respectively. Thus, a slight variation in energy can be significantly amplified as a detectable change of the absorption factor in the spectrum.
The arrangement of the optical elements of beamline 1W2B is shown in Fig. 2. Synchrotron beam emitted from the wiggler is reflected by the collimating mirror, the double-crystal monochromator (DCM), the focusing mirror, and incident onto the energy stabilization system. The system is composed of an ion chamber (Oxford; to record the incident I0), a heavy-element sample (abbreviated as `standard sample' hereon) and a fluorescence detector (Lytle, to collect the fluorescence signal If). During anomalous diffraction experiments, a fraction of the incident beam is absorbed by the standard sample, transferred into fluorescence signal, and recorded by the fluorescence detector. Apart from this small amount, most of the energy passes through the standard sample and proceeds to the normal diffraction experiment. If the absorption factor (calculated using the function μ = If/I0) is out of a certain range, the DCM will make a corresponding adjustment to revert to the initial energy.
The standard samples which were used to produce fluorescence signals should be prepared with different valences so as to match the M ml−1, and the thickness of the standard sample was 1 mm. The standard samples were located at 45° from the incident light and the fluorescence detector in order to receive the maximum fluorescence energy. The prepared standard samples covered most of the heavy elements that are frequently used in the method (such as selenium, gold, platinum, mercury, iodine, bromine, ferrum, cuprum, etc.). For each element, most of its common valences were prepared (typical standard samples are listed in Table 1).
of the heavy element inside the crystal to be measured. For example, if we would like to stabilize the X-ray energy at the inflection point of a crystal, it should be located on the steep rising part of the of the standard sample (unnecessarily located exactly at the inflection point of the standard sample). In this case, we can say that there is a match between the crystal and the standard sample. The standard samples can be either in solid or liquid form. The metal samples were covered with Kapton film of thickness 100 nm. This value was optimized by balancing between yielding a sufficient fluorescence signal and maintaining a more direct light (about 3% of the incident light is lost through the film). The liquid and powder samples were encapsulated in cuvettes with mica windows. The concentration of heavy element in the solution or powder sample was 50 m
|
A schematic flowchart illustrating the procedure for stabilizing the incident X-ray energy during anomalous data collection at the inflection point is shown in Fig. 3. Before data collection, the of the standard samples and the crystal were collected and compared. The standard sample whose matches best with that of the crystal was selected for energy monitoring. Once a certain energy E0 at the inflection point of the was chosen for anomalous data collection, the initial absorption factor μ0 and the permitted range of μ (μ0 − Δμ1, μ0 + Δμ2) of the standard sample were calculated based on the The absorption factor μ of the standard sample is monitored continuously during the anomalous data collection procedure. If (μ0 − μ2) < μ < (μ0 + μ1), the data collection procedure would proceed, otherwise a software feedback loop to the DCM would be triggered to recover to the permitted range of μ (μ0 − Δμ1, μ0 + Δμ2). The programs in charge of data acquisition and DCM movements were written using LabVIEW (Johnson, 1997).
Similarly, there are two ways to stabilize energy at the peak point. One way is to use the peak point of the standard sample to monitor the X-ray energy [abbreviated as peak–peak monitoring, Fig. 4(a); a schematic flowchart is shown in Fig. S1A of the supporting information]. In this situation, the of the standard sample should accurately match that of the heavy element inside the crystal at the peak point. The other way is to monitor the incident X-ray energy by the region around the inflection point of the standard sample when the inflection point of the crystal is located on the steep rising part of the [abbreviated as inflection–peak monitoring, Fig. 4(b); a schematic flowchart is shown in Fig. S1B of the supporting information]. The inflection–peak monitoring method can reach a higher energy resolution than the peak–peak monitoring method (0.2 eV versus 0.5 eV). However, it can work only when it is possible to find one standard sample whose inflection region coincides well with the peak region of the crystal. Sometimes this occurs for the same element with different valences. As a result, if possible, it is better to monitor the energy at the peak point with the inflection point of the standard sample.
3. Results
The apparatus described above has been used at the biological macromolecular crystallography beamline 1W2B at BSRF. Typically, the intensities recorded by the fluorescence detector and the ion chamber are both about 106 counts s−1 with a background of about 104 counts s−1. At these count rates, a 5% change of μ can be detected precisely. Corresponding to the resolution of μ, the energy of the incident X-rays can be measured with an accuracy of 0.1 eV and 0.5 eV at the inflection and the peak point of the respectively. When working at an energy of 10 keV, the step size of the Si(111) DCM is about 0.2 eV. Taking together the energy resolution of the equipment and the step size of the DCM, the incident X-ray energy can be kept within the scope of 0.2 eV and 0.5 eV, monitored by the inflection point and the peak point of the standard sample, respectively.
Using this equipment, the energy stability of beamline 1W2B at BSRF was measured by continuous recording of the absorption factor at the inflection point of a standard elemental Se sample over the lifetime of the synchrotron. It is observed that the absorption factor fluctuates in a wave-like manner, with a maximum value of about 1.4 and a minimum value of about 0.8 (Fig. 5a, black triangles). Due to the middle section of the (the part near the inflection point, within the range 0.8 < μ < 1.4) being approximately linear (Fig. 5b), the absorption factors were further transformed into the corresponding energy of the incident beam accordingly. A total energy shift of 1.4 eV was observed over the lifetime of the synchrotron (Fig. 5c, black triangles). Beam attenuation or injection of the synchrotron induces an increase or decrease of heat loading at the monochromator, and leads to a deformation of the monochromator. Thus, the same monochromator position can select different energies at different times. This illustrates that the continuous drift in energy is mainly caused by fluctuations in the Besides the continuous drift, an abrupt change in energy can also be observed (Fig. 5c, black triangles, from 3.7 h to 4.2 h). Considering that the deformation period of the DCM should be much larger, this kind of sudden change of energy was most probably caused by unpredictable beam position movements induced by electron shaking inside the storage ring.
Further, we recorded the absorption factors of the same standard sample during the equivalent synchrotron life-time period while the DCM feedback loop is in operation. The absorption factor was confined within the range 1.1–1.2 (Fig. 5a, red squares) during measurement, corresponding to an incident X-ray energy in the range 12657.7–12657.9 eV (Fig. 5c, red squares). Thus, after using the energy stabilization apparatus, the fluctuation of the X-ray energy was ensured to a limited range of 0.2 eV, significantly superior to the energy shift of 1.4 eV observed earlier.
So far, this method has been successfully used to stabilize the energy of the incident beam when collecting anomalous diffraction data at the crystallography beamline 1W2B. For example, published single-wavelength anomalous diffraction (SAD) data of Se-Met YafQ-DinJ [Protein Data Bank (PDB) id: 4ml0] and Se-Met YafQ mutant H87Q (PDB id: 4mmg) (Liang et al., 2014) was obtained with the aid of this equipment. The of Se within the protein matches well with that of the elemental Se (Fig. 6a). Thus, elemental Se was chosen as a standard sample to monitor the energy when collecting anomalous data at the peak point. The absorption factor of the standard sample was controlled within the range 2.75 < μ < 2.80 by the equipment, and the stability of the incident X-ray energy was kept to 0.5 eV for the duration of data collection. Thereafter, the sample's factors f′ and f′′, which were calculated based on the spectra and by the program CHOOCH (Evans & Pettifer, 2001), were guaranteed within the ranges −6.7 < f′ < −7.3 and 6.35 < f′′ < 6.5, respectively (Fig. 5b). If the energy has not been stabilized by the system, the maximum energy change can reach 1.4 eV, and the factors f′ and f′′ would fluctuate in the ranges −6.0 < f′ < −8.5 and 6.0 < f′′ < 6.5, respectively. This example illustrates the need of the system to keep the energy stable.
4. Discussion
The main properties of the X-ray beam that are essential for a successful SAD/MAD (single-wavelength anomalous diffraction/multi-wavelength anomalous diffraction) data collection include the beam size and energy resolution at the sample position, and stable beam intensity, beam position and beam energy (the beamline specifications of beamline 1W2B are shown in Table 2). The beam size and energy resolution contribute to the reflection width, and are immediately visible in the spot profile of low-resolution reflections. A large beam size and low energy resolution may introduce overlapping between reflections for crystals with very long unit cells. The energy resolution (≤2 × 10−4) and beam spot size (1.0 mm × 0.6 mm) of 1W2B can satisfy the demand of high-precision data acquisition in MAD/SAD. Large fluctuations in the beam intensity produce variable background intensities, while large fluctuations in the beam position shifts the diffraction points positions in different diffraction patterns. These deviations, as long as they are not too large, can be corrected by routine scaling and procedures using auto-indexing programs.
|
Unlike the items above, the energy of the incident beam was deemed to be stable by the auto-indexing programs by default, and there is no way of recognizing energy fluctuations by the scattering patterns directly, even though energy fluctuation is common during synchrotron exposures. It would be desirable, therefore, to design an instrument to stabilize the incident X-ray energy during the exposures. In this investigation, a new method to calibrate and stabilize the incident X-ray energy has been reported and successfully utilized at the 1W2B X-ray crystallography beamline of BSRF. The method employs the inflection or peak point of the
to monitor the X-ray energy. It is of general use to anomalous diffraction or scattering experiments which are highly sensitive to energy stability.In this energy stabilization system, the first priority is to select a suitable standard sample whose
coincides well with that of the crystal. Movement of the depends greatly on the valence of the heavy element. The higher the valence, the further the spectra will move compared with its elemental form. Se-Met protein is mostly used to solve the in structural biology research. An excellent standard sample for Se-Met protein crystals is selenocysteine powder. It ensures that the adjacent anatomic structures of the Se element and the protein crystal are the same as that of the standard sample, and hence ensures the absolute coincidence of the of the crystal and the standard sample. For anomalous diffraction experiments using a crystal containing other heavy elements, the valence of the heavy element inside the crystal is unknown. As a result, a series of standard samples with different valences should be prepared, and comparing the of the crystal and that of the standard samples is essential for selecting the best match.A significant feature of this apparatus is that it is not able to fix energies far away from the absorption edges of standard samples. Fortunately, for most synchrotron techniques including imaging, spectroscopy, normal scattering and diffraction, an energy fluctuation of 1–2 eV (meaning a 0.01–0.02% change in energy when working at a typical synchrotron energy of 10 keV) is negligible. Adequate cooling matching the heat load of the incident X-ray beam (Kudo & Tanida, 2007; Proux et al., 2006) and recalibration of the energy is sufficient to achieve the required stability in this case. On the contrary, anomalous data collected at the inflection and peak points of the are more sensitive to energy instability. The 0.01–0.02% change of energy at the inflection point of the selenium K-edge can lead to an 80% change in the coefficient f′′. As a result, the system is essential for stabilizing the incident X-ray energy and to obtain a more reliable coefficient.
Using the instrument, the energy of the incident beam can be detected at a resolution of about 0.1 eV at the inflection point. If an accumulating counter is used to reduce the noise of the measured absorption factor, the energy resolution can be even higher. However, the smallest rotation step of the DCM at beamline 1W2B is only 0.2 eV. This indicates that the instrument has the potential to achieve even higher energy stability by using an accumulating counter and improving the DCM's rotation platform.
Supporting information
Flowcharts of peak-peak monitoring and inflection-peak monitoring of the incident X-ray energy. DOI: https://doi.org/10.1107/S1600577517006312/rv5065sup1.pdf
Acknowledgements
We thank the staff at 1W2B beamline, BSRF. We also thank Dr Zhi Geng for helpful discussions. This work was supported by grants from the National Basic Research Program of China (2009CB918600) and the National Natural Science Foundation of China.
Funding information
Funding for this research was provided by: National Natural Science Foundation of Chinahttps://dx.doi.org/10.13039/501100001809 (award No. 31600597).
References
Evans, G. & Pettifer, R. F. (1996). Rev. Sci. Instrum. 67, 3428–3433. CrossRef CAS Web of Science Google Scholar
Evans, G. & Pettifer, R. F. (2001). J. Appl. Cryst. 34, 82–86. Web of Science CrossRef CAS IUCr Journals Google Scholar
González, A. (2003). Biol. Crystallogr. 59, 1935–1942. Google Scholar
Hendrickson, W. A., Horton, J. R. & LeMaster, D. M. (1990). EMBO J. 9, 1665–1672. CAS PubMed Web of Science Google Scholar
Johnson, G. W. (1997). LabVIEW Graphical Programming: Practical Applications in Instrumentation and Control. New York: McGraw-Hill School Education Group. Google Scholar
Kudo, T. & Tanida, H. (2007). Rev. Sci. Instrum. 78, 033104. Web of Science CrossRef PubMed Google Scholar
La Fortelle, E. de & Bricogne, G. (1997). Methods Enzymol. 276, 472–494. PubMed Web of Science Google Scholar
Liang, Y. J., Gao, Z. Q., Wang, F., Zhang, Y. L., Dong, Y. H. & Liu, Q. S. (2014). J. Biol. Chem. 289, 21191–21202. Web of Science CrossRef CAS PubMed Google Scholar
Proux, O., Nassif, V., Prat, A., Ulrich, O., Lahera, E., Biquard, X., Menthonnex, J.-J. & Hazemann, J.-L. (2006). J. Synchrotron Rad. 13, 59–68. Web of Science CrossRef CAS IUCr Journals Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.