research papers
Smaller capillaries improve the small-angle X-ray scattering signal and sample consumption for biomacromolecular solutions
aEuropean Molecular Biology Laboratory (EMBL), Hamburg Outstation c/o DESY, Notkestrasse 85, 22607 Hamburg, Germany, and bEuropean Molecular Biology Laboratory (EMBL), Grenoble Outstation, 71 Avenue des Martyrs, 38042 Grenoble, France
*Correspondence e-mail: martin.schroer@embl-hamburg.de
Radiation damage by intense X-ray beams at modern synchrotron facilities is one of the major complications for biological small-angle X-ray scattering (SAXS) investigations of macromolecules in solution. To limit the damage, samples are typically measured under a laminar flow through a cell (typically a capillary) such that fresh solution is continuously exposed to the beam during measurement. The diameter of the capillary that optimizes the scattering-to-absorption ratio at a given X-ray wavelength can be calculated a priori based on fundamental physical properties. However, these well established scattering and absorption principles do not take into account the radiation susceptibility of the sample or the often very limited amounts of precious biological material available for an experiment. Here it is shown that, for biological solution SAXS, capillaries with smaller diameters than those calculated from simple scattering/absorption criteria allow for a better utilization of the available volumes of radiation-sensitive samples. This is demonstrated by comparing two capillary diameters di (di = 1.7 mm, close to optimal for 10 keV; and di = 0.9 mm, which is nominally sub-optimal) applied to study different protein solutions at various flow rates. The use of the smaller capillaries ultimately allows one to collect higher-quality SAXS data from the limited amounts of purified biological macromolecules.
Keywords: small-angle X-ray scattering; protein solution; radiation damage; flow through sample environment.
1. Introduction
Small-angle X-ray scattering (SAXS) is used to probe the structure of biological macromolecules in solution (Graewert & Svergun, 2013; Skou et al., 2014a; Svergun & Koch, 2003; Svergun et al., 2013; Schroer & Svergun, 2018). The most advanced studies are conducted at synchrotron SAXS beamlines, partially or fully dedicated to structural biology (Blanchet et al., 2015; Pernot et al., 2013; David & Pérez, 2009; Inoue et al., 2013; Acerbo et al., 2015; Classen et al., 2013; Kirby et al., 2013; Li et al., 2016). The SAXS data collection on biomacromolecular solutions presents specific experimental challenges. The solutes are often very dilute (over 99% water by volume) and have low X-ray contrast resulting in very weak scattering intensities upon subtraction of the solvent scattering. Synchrotron SAXS capitalizes on the of the X-ray source to improve the signal intensity; however, the high also has a downside. As the samples are in aqueous solutions, the photolytic production of free radicals and solvated electrons from the solvent by the X-ray beam may cause radiation damage and the formation of macromolecular aggregates that may render SAXS profiles uninterpretable (Garrison, 1987; Maleknia et al., 2001; Kuwamoto et al., 2004). Since the observable effect of radiation damage depends not simply on the dose deposited on the sample but largely on the tendency of the solute to aggregate, it is not possible to adequately predict the magnitude of the expected effect prior to the measurement. Complicating matters, the sample itself may only be available in limited quantities. Therefore, there is a need to balance SAXS data collection between the effects of potential radiation damage and sample consumption to maximize the data quality, i.e. optimize the signal-to-noise ratios in the recorded weak scattering intensities from undamaged samples. Mitigating the effects of radiation damage is particularly important for biological SAXS (bioSAXS) measurements (Brooks-Bartlett et al., 2017; Hopkins & Thorne, 2016; Jeffries et al., 2015) and, in order to reduce this effect, several schemes can be employed. These include continuous sample flow, beam attenuation, the addition of small `scavenger' molecules or cryo-cooling (Fischetti et al., 2003; Jeffries et al., 2015, 2016; Kirby et al., 2016; Kuwamoto et al., 2004; Meisburger et al., 2013, Sahle et al., 2015). In the following, we show that the geometry of the sample exposure cell can also be optimized to limit the impact of radiation damage and improve the measured SAXS profiles from radiation-sensitive samples.
The magnitude of the SAXS intensity, I, measured from a dilute protein solution can be described as
where s is the wavevector momentum transfer that relates to the X-ray wavelength λ and the scattering angle 2θ as s = 4πsinθ/λ. Here, c denotes the solute concentration, I(0) the P(s) the normalized form factor [i.e. P(0) = 1], t the exposure time, AXray the X-ray beam-area at the sample position, d the sample path length and μ(E) the total Intuitively, reducing the time in the X-ray beam decreases I(s); however, measuring samples as a set of sequential short exposures, as opposed to one long exposure, is useful to detect the onset of radiation damage and ensure that only the data collected from the sample unaffected by beam-induced changes are averaged, thereby improving the data accuracy. The total is dependent on the X-ray absorption and scattering cross sections at a given X-ray photon energy E. Equation (1) indicates that the strongest scattering signal at a fixed concentration is given for the sample path length dopt = 1/μ (see §S1, Fig. S1, of the supporting information), which is typically met in solution SAXS measurements. In practice, cylindrical glass capillaries with diameters between 1 and 2 mm are used, as these match the optimum sample path length of dopt = 1.0 mm (at E = 8 keV) and dopt = 1.9 mm (at E = 10 keV) for water, which is a reasonable approximation for diluted aqueous macromolecular samples.
This calculation of the optimal path length that maximizes the scattering signal only takes into account the stochastic X-ray interaction with the sample as described by μ(E) without accounting for the effect of radiation damage. The latter may lead to the formation of diverse solute states, mostly aggregates in this study, which all yield different form factors P(s). It is from P(s) that structural information can be obtained and, ultimately, a successful bioSAXS experiment needs to maximize I(s) without incurring X-ray induced changes in P(s) during the X-ray exposure.
Since macromolecular samples have limited in-beam lifetimes and are often available in limited quantities, it is necessary to optimize the SAXS signal that can be collected as a function of the sample volume while minimizing radiation damage. The diameter of the cylindrical capillary influences the intensity I(s), the absorption and therefore defines how efficiently the sample volume is used. The beam area at many SAXS beamlines is in the range AXray = 100–500 µm × 100–500 µm and consequently a large quantity of the sample, if housed within a 1–2 mm capillary, is not exposed at all and exits the capillary unmeasured. Capillaries with smaller diameters increase the ratio of illuminated to non-illuminated volume, thus improving the sample utilization. Moreover, for continuous-flow experiments, smaller capillaries enable higher flow speeds refreshing the sample through the beam faster and thus reducing the cumulative effects of radiation damage. Therefore, although smaller-diameter capillaries based on first principles described in equation (1) lower the scattering intensity, the decrease in path length may be outweighed by a more efficient data collection scheme to ultimately yield higher-quality scattering data from undamaged samples. In this work we present comparative studies demonstrating that, indeed, capillaries with diameters smaller than the calculated optimum thickness do provide overall improvements to the SAXS profiles measured from radiation-sensitive biological samples.
2. Experimental details
2.1. Sample preparation
Powdered ultrapure chicken egg-white lysozyme was purchased from Affymetrix USB. Bovine serum albumin (BSA; lyophilized powder, crystallized, >98%) was purchased from Sigma. In all instances the protein samples were freshly prepared at room temperature (T ≃ 20°C), just prior to the SAXS measurements. Individual aliquots of the respective protein powders were separately dissolved in 50 ml of 40 mM KH2PO4, 60 mM KCl, pH 6.8 [this phosphate buffer was chosen to increase the susceptibility of both proteins to radiation damage (Jeffries et al., 2016)]. The solutes were 0.22 µm filtered to produce lysozyme and BSA samples at a final protein concentration of c = 4.4 mg ml−1 and c = 2.9 mg ml−1, respectively. The final concentration was evaluated by UV absorption using a NanoDrop spectrophotometer. The A280nm absorption coefficients expressed as E0.1% (= 1 mg ml−1) were calculated from the primary amino acid sequence of each protein using ProtParam (Gasteiger et al., 2005: BSA E0.1% = 0.614; lysozyme E0.1% = 2.653). The same phosphate buffer used to prepare the protein samples was also employed as the matched solvent blank for the subtraction of the background scattering.
2.2. SAXS data collection
SAXS measurements were performed at the EMBL P12 bioSAXS beamline, EMBL/PETRA III, Hamburg, Germany (Blanchet et al., 2015; Hajizadeh et al., 2018), using an X-ray photon energy of E = 10 keV (λ = 0.124 nm) with a beam size of 200 µm × 350 µm (V × H, full width at half-maximum, FWHM) and a of F = 5 × 1012 photons s−1 at the sample position. Standard batch mode measurements were performed using the robotic P12 sample changer (Blanchet et al., 2015; Round et al., 2015) both with and without continuous in-capillary sample flow through the beamline (horizontally oriented under vacuum). Before and after the protein solutions the corresponding buffer solutions were measured. Between each measurement the capillaries were extensively cleaned to remove any deposits due to radiation-induced aggregates.
Two types of quartz glass capillaries with different diameter were used: (1) a capillary with an outer diameter of do = 1.8 mm and inner diameter of di = 1.7 mm, close to the optimum path length (dopt = 1.9 mm) for 10 keV X-rays, and (2) a capillary with do = 1.00 mm and di = 0.91 mm. Both capillaries were purchased from Hilgenberg, Germany. In the following, they will be referred to as the di = 1.7 mm and di = 0.9 mm capillaries, respectively. For all measurements, a constant sample volume of Vtot = 20 µl was loaded into the capillaries that were maintained at a constant temperature of T = 10°C.
SAXS data were recorded using a PILATUS 2M photon-counting detector (DECTRIS, Switzerland) from samples and buffers using exactly the same conditions at several different flow rates and total exposure times of the solutions as described in the Results. For all experiments, two-dimensional (2D) SAXS data were recorded as a sequential set of images (frames) for every 50 ms of exposure time (45 ms collection time + 5 ms detector readout time) during the course of X-ray exposure.
2.3. SAXS data analysis
The 2D images on the detector were used to estimate the number of photons scattered by the proteins in each sample. First, the scattering patterns for the samples and corresponding buffers were scaled to the storage ring current to take into account small variations in the intensity of the incident X-rays. Then the data collected from the buffer were subtracted from that of the respective sample to obtain subtracted 2D images of the proteins in solution. For each subtracted 2D image, the number of photons counted by the pixels in the range 0.12 nm−1 < s < 2.9 nm−1 were summed to obtain the integrated scattered photon count. The summation was conducted until the frames showed radiation damage yielding the integrated photon count, Iint, from the undamaged macromolecules in the sample.
To assess the time point of the onset of radiation damage and to obtain 1D-reduced scattering profiles for subsequent analysis, the unsubtracted 2D images underwent data reduction using the P12 beamline SASFLOW pipeline (Franke et al., 2012). Briefly, the 2D SAXS patterns were normalized to the transmitted beam and azimuthally averaged. The resulting SAXS curves were then analysed using the correlation map method (CorMap; Franke et al., 2015) whereby the one-dimensional (1D) profiles from the sequentially acquired images were pairwise compared for statistical similarity across the entire s-range (relative to the first data frame). Only the frames above a similarity significance threshold of α ≥ 0.01 were used to generate the (frame-)averaged SAXS profile for the given sample and buffer. The averaged protein SAXS data were corrected for the averaged buffer/background scattering, measured under the exact same conditions, to yield the SAXS signal from the protein, 〈I(s)〉, without radiation damage. In the pure buffer solutions studied no radiation damage was observed across the individual data frames as assessed using CorMap.
In addition to the CorMap pairwise evaluation of each data frame, all individual images measured from a sample underwent data reduction and buffer subtraction using the averaged buffer data to produce a set of single SAXS curves I(s) from which the consistency of the Rg across frames was evaluated using the Guinier approximation [I(s) I(0)exp(-Rg 2 s 2/3), where I(0) denotes the extrapolated at zero angle (Guinier, 1939)].
While CorMap assesses systematic (correlated) +1 or −1 variances (co-variances) in the intensities between data frames across the entire data range, the Rg assessment from the Guinier approximation uses only intensities at the very low angles calculated from a linear fit to the data [when expressed as lnI(s) versus s2, to an Rgsmax = 1.3]. The linear fit from which Rg is calculated from the (proportional to the negative slope) is affected by both signal-to-noise and the associated errors on the data. As a result, the Rg error estimates may be relatively large, especially when using a single data frame. On comparing the data frames, the Rg approach does not take into account systematic discrepancies that may occur during the linear fitting (linear correlation coefficients are insufficient to detect such discrepancies). Therefore, the joint use of CorMap and the Guinier Rg provides a reliable estimate of the onset of damage. CorMap also has the advantage of assessing whether systematic discrepancies between the compared intensities occur outside of the Guinier region (covering the cases when the effects of damage manifest themselves at higher angles, i.e. at shorter length scales).
As the main aim of averaging the single SAXS curves unaffected by radiation damage is to reduce the noise in the data, it is essential to assess the effect of changing the capillary diameter on the resulting noise level. Therefore, the 〈I(s)〉 curves were further processed with the program AUTOGNOM from the ATSAS program package (Petoukhov et al., 2012; Franke et al., 2017) to obtain the smoothed reciprocal-space fit to the experimental, undamaged data, IAG(s) (from the calculated probable real-space distance distribution). In all instances, reciprocal-space fit CorMap p-values were greater than 0.01, indicating that no systematic deviations are present between the calculated fit and the experimental data (Franke et al., 2015). The normalized deviation Δ〈I(s)〉 of the experimental SAXS curves, 〈I(s)〉, from the smoothed AUTOGNOM IAG(s) function was computed for each sample such that
From this normalized difference curve, a scalar quantity DEV was determined as
i.e. the total sum of the modulus of the normalized deviations divided by the number of experimental data points ns. For noisy data, it is expected that the fluctuations with respect to the smooth curve IAG(s) will be high, yielding larger DEV, while lower DEV values point to lower relative noise levels. For each protein species the same respective s-range and a similar number of data points ns were used to compute the DEV value.
In order to determine the quality and information content in the SAXS curves from which structural information for modelling can be determined, the radially averaged data were further analysed with the ATSAS program SHANUM (Konarev & Svergun, 2015). With this approach, the numbers of reliably detectable Shannon channels MS and therefore the maximum momentum transfer in the useful data range sopt were determined for both capillaries.
3. Results
3.1. Sample geometry and effective sample exposure volumes
Fig. 1(a) shows the images taken from a video camera for online visualization of the two cylindrical sample capillaries with the internal diameters di = 1.7 mm and di = 0.9 mm placed in the evacuated sample exposure unit during the automated sample loading. The video microscope detects the meniscus of the sample and places it such that the X-ray beam fully hits the solution that is subsequently measured either at a fixed position or with the sample flow enabled. In a continuous-flow operation, the flow speed is automatically adjusted such that the full volume of the sample is passed through the incident X-ray beam during the total exposure time. The data are acquired as a set of individual images, typically 20 × 50 ms during the exposure time. In Fig. 1(a) the sample flows horizontally from left to right, with the X-ray beam oriented at 90° to the capillary, progressing through the sample from the bottom of the image to the top, with the horizontal beam dimension parallel to the flow direction and the vertical beam dimension out of the plane of the image. The beam is aligned to hit the centre of the capillary, maximizing the path length through the sample.
We shall distinguish the following volumes inside the capillary: the sample volume needed to fill the capillary, Vsample, which is approximated by the size of the beam in the horizontal direction multiplied by the cross-sectional area of the capillary; the volume of Vsample actually exposed in the X-ray beam, Vbeam, that corresponds to the beam size in both horizontal and vertical directions multiplied by the capillary diameter; and the sample volume exposed when flowing the sample Vexp. Fig. 1(b) shows the geometry schematically. As can be seen, the volume exposed at a given time is noticeably smaller than Vsample. The corresponding volumes consumed but not exposed by the beam are, for fixed sample measurements, ΔVloss,fix = Vsample − Vbeam, and for the flowing samples ΔVloss,flow = Vtot − Vexp. The ratio between Vbeam and Vsample (and also of Vexp and the total sample volume Vtot) reflects the sample consumption efficiency. Table 1 shows the corresponding values for di = 1.7 mm and di = 0.9 mm capillaries where the effect of reducing the internal diameter is immediately obvious under the flow conditions. Indeed, a significantly higher proportion of the total sample is exposed to the X-ray beam upon reducing the capillary diameter.
|
For the measurements in the fixed position, the larger capillary needs about 3.5 times more volume for filling compared with the smaller one but only a 1.9 times larger volume is actually exposed to the beam and thus contributes to the SAXS signal. When flowing a total solution volume of Vtot = 20 µl, for the di = 0.9 mm capillary the exposed volume is even about 1.9 times larger than for di = 1.7 mm. This results in a consumption efficiency of 42% (di = 0.9 mm) in comparison with 22% (di = 1.7 mm).
3.2. Data collection without sample flowing
Besides the overall sample consumption, the integrated scattering intensity Iint from the protein solutions in both capillaries is of major relevance for successful SAXS measurements. For the background-corrected 2D data, the averaged integrated scattering intensity per unit time t and per 1 mg ml−1 of sample, Iint/(tc), has been determined for lysozyme and BSA protein samples (Fig. 2a). These are corrected for the absorption of the quartz glass capillaries by dividing through the respective transmissions T.
For di = 1.7 mm, the scattering intensity in the fixed position is higher than that for di = 0.9 mm. Thus, as expected from theory, decreasing the capillary diameter leads to a decrease of the integrated scattering per unit time. On the other hand, the scattering intensity per time and per volume sample loaded, Vsample, is better for the smaller capillary (Fig. 2b). Thus, a higher sample consumption efficiency accompanied by a lower scattering intensity still may yield overall improved scattering intensity statistics if sufficient number of SAXS curves without radiation damage can be measured.
The onset of the radiation damage can be seen by observing the evolution of the Rg with time. Figs. 2(c) and 2(d) show the averaged over several repetitions as a function of exposure time t for fixed samples in both capillaries for lysozyme and BSA, respectively. For the lysozyme sample at fixed position (Fig. 2c), Rg increases already after t = 45 ms, independently of the capillary diameter. For longer exposure times, the CorMap analysis does not find any statistical similarity (black cross). In this situation, no reliable information can be obtained from the data, as the radiation damage is already present in the first SAXS curve. In the case of BSA, which is slightly less radiation sensitive, a similar result is found (Fig. 2d).
Summarizing, for the fixed sample, the onset of radiation damage does not depend on the capillary diameter, whereas the total scattering intensity per time and per volume sample, Vsample, is higher for the smaller capillary. Thus, the lower sample consumption for the di = 0.9 mm capillary allows for a larger number of repeated measurements which, in turn, would allow one to obtain a larger summed scattering intensity. A natural way of repeating the experiments uses the sample flow within the capillary, and this is discussed in the next section.
3.3. Data collection with flowing sample
3.3.1. Flow rates
The influence of the capillary size on the measured SAXS intensities from samples under continuous flow through the X-ray beam was studied using a fixed total sample volume Vtot for various sample delivery volume flow rates Q (from Q = 10 µl s−1 to Q = 40 µl s−1). The corresponding linear sample flow speed v = , is obviously higher for the di = 0.9 mm capillary compared with the di = 1.7 mm capillary if the samples are delivered at the same flow rate, resulting in shorter average dwell times tdwell = w/v of the protein solution in the X-ray beam of width w (in this case, 350 µm) and thus a lower average radiation dose (see §S2 of the supporting information).
The flow properties of liquids in capillaries are usually characterized by the Reynolds number Re = ρvdi/η (Rott, 1990), where ρ denotes the density of the fluid and η the For the solution samples described here (consisting of 99.5% by mass of water) and for the given capillary diameters, the highest Reynolds number is Re = 43 (T = 10°C). This value is much lower than the critical value of Re ≃ 2000 (Rott, 1990) required for the breakdown of laminar flow, and it is thus assumed that, for the dilute protein samples, the laminar flow regime is maintained.
The flow measurements were performed for a total sample volume of Vtot = 20 µl. Different flow rates through the X-ray beam were obtained by changing the total number of short (texp = 50 ms) sequentially measured SAXS exposures, NSAXS, such that
(Note: texp = 50 ms is the sum of collection time tcoll = 45 ms per curve and a read-out time of tread = 5 ms of the Pilatus 2M detector.) The total exposure time, ttot, of a sample, i.e. the maximum illumination of each sample in the X-ray beam (irrespective of radiation damage), is given by ttot = texpNSAXS. The experimental parameters summarized in Table 2 correspond to the SAXS data collected for the two protein samples in both capillaries at different flow rates using the same total volume.
|
The laminar flow of protein solutions can be described by a simple parabolic radial velocity profile (Rogers, 1992). At the capillary walls, the flow velocity approaches zero, and radiation damage is expected to be stronger (Nielsen et al., 2012). For the given linear flow speeds v, the velocity profiles for both capillaries at different flow rates indicate a much thinner boundary layer of slowly moving sample for the di = 0.9 mm capillary, and hence a lower probability of the deposition of protein and aggregates (see §S3, Fig. S2).
3.3.2. Influence of the capillary size on the onset of the radiation damage
Fig. 3 shows a series of single 1D-SAXS curves of lysozyme at the flow rate Q = 20 µl s−1 for di = 1.7 mm and di = 0.9 mm capillaries. In both cases, 20 single SAXS profiles were collected using continuous flow. For di = 1.7 mm, the radiation damage is apparent after an exposure of t = 245 ms, as is observed by a systematic increase in the scattering intensities near zero angle and an increase in the negative slope at very low values of s (s < 0.5 nm−1, Fig. 3a). In comparison, for the di = 0.9 mm capillary, the onset of radiation damage is significantly extended whereby the formation of aggregates is not apparent in the data until t = 495 ms (Fig. 3b).
The onset of the radiation damage is reflected yet more clearly in the Rg, and Fig. 4 shows Rg as a function of exposure time t for different volume flow rates in both capillaries. Flowing the sample at Q = 10 µl s−1 somewhat reduces the apparent Rg compared with the fixed sample but only slightly diminishes the radiation damage (Fig. 4a). For di = 1.7 mm, after t = 90 ms, radiation damage is already detectable, whereas, for di = 0.9 mm, data up to t = 200 ms can be averaged. It appears that the size and/or amount of aggregates in the smaller capillary is lower than for the larger capillary, possibly thanks to higher linear flow speed and thus lower radiation dose in the former case. The increase in Rg at t = 800 ms for di = 0.9 mm may be attributed to the formation of larger aggregates similar to those present for di = 1.7 mm, which are, however, quickly removed after t = 1000 ms due to the volume flow. This effect could be similar to the vortex-ring flow which circulates the sample solution between the centre and the wall of the capillary as reported previously for capillary flow (Nielsen et al., 2012).
As expected, less radiation damage is observed when the flow rate increases. Notably, however, the number of radiation-damage-free SAXS curves for Q = 20 µl s−1 (Fig. 4b) and Q = 40 µl s−1 (Fig. 4c) is higher for the di = 0.9 mm capillary than for the di = 1.7 mm. Additionally, for flowing samples, Rg of the aggregates formed due to radiation damage reaches a plateau value, which decreases continuously with increasing flow rate. Similar results are obtained for BSA (see §S4, Fig. S3), for which the total effect of radiation damage is slightly lower and the solutions are even partially under-exposed, i.e. actually more SAXS profiles could be taken before the radiation damage sets in.
3.3.3. Influence of the capillary size and flow rates on the integrated scattering
When the samples are measured at a continuous flow the smaller-diameter capillary is clearly more effective than the larger capillary in limiting the rate of radiation damage of the radiation-sensitive protein samples. This stems from the higher linear flow speed, v, and the associated increase in replenishing the sample through the capillary. As a result, larger numbers of individual SAXS data frames can be collected in succession that, after the frame averaging, improve the signal-to-noise ratios in the resultant SAXS curves. Indeed, the final integrated scattering Iint collected from the undamaged samples is higher for the di = 0.9 mm capillary compared with the di = 1.7 mm capillary, i.e. the reduction in scattering intensity for the 0.9 mm system as predicted by equation (1) (and as discussed for the fixed position results, §3.2) is offset by faster flows reducing the radiation damage combined with a higher proportion of sample volume being illuminated during the total exposure time.
It is instructive to analyse how long the samples can be exposed to the X-ray beam as a function of flow rate before radiation damage sets in, and this can be also easily tested during a SAXS measurement. Fig. 5(a) shows the exposure time, tav, for which the lysozyme sample can be measured before the onset of radiation damage plotted against the flow rate Q. An increase of tav with increasing flow rate and, more importantly, a larger exposure time for di = 0.9 mm compared with di = 1.7 mm is observed before radiation damage sets in. The figure also shows the total exposure time ttot for the entire set of the SAXS curves. In the case of di = 1.7 mm, radiation damage is present for all flow rates studied; for di = 0.9 mm at Q = 40 µl s−1, nearly all SAXS curves are similar and can be averaged.
Fig. 5(b) shows the integrated scattering intensity from the SAXS patterns before showing radiation damage, Iint, determined for lysozyme as a function of for both capillaries. Increasing the rate from Q = 10 µl s−1 to 20 µl s−1, the integrated intensity of the intact sample increases as fewer SAXS curves exhibit radiation damage. In the case of di = 1.7 mm, Iint is the same within the error for Q = 20 µl s−1 and Q = 40 µl s−1. A similar result is observed for the smaller capillary.
In the case of BSA, the situation is somewhat different, as the protein is less radiation sensitive (Figs. 5c, 5d) (Jeffries et al., 2015). Fig. 5(d) shows Iint as a function of flow rate. For the larger capillary, the optimum flow condition with the largest Iint is at Q = 20 µl s−1 whereas for lower or higher flow rates less scattering intensity from native BSA is recordable. For di = 0.9 mm, Iint is larger for Q = 10 µl s−1 and 20 µl s−1 than for di = 1.7 mm. Only at Q = 40 µl s−1 is the signal from the larger sample path length higher. More interesting to note is that the optimum conditions are present at Q = 10 µl s−1 where the highest signal is detectable.
These findings and the difference to the data for lysozyme can be directly understood by looking at the Q dependence of tav and ttot (Fig. 5c). In the case of the highest flow rate, no radiation damage is present for both capillaries, all curves can be averaged, and thus the total SAXS intensity is higher for di = 1.7 mm because more proteins give rise to the signal (however, at the cost of a less efficient sample consumption, see §3.1). With decreasing flow rate, radiation damage sets in for the large capillary, leading to the best measured condition for Q = 20 µl s−1 (longest exposure time tav). In contrast, for di = 0.9 mm, nearly no radiation damage is present for all conditions studied where the sample is moved. Thus, the sample is in this case `under-exposed' as more frames could have been collected before the onset of the radiation damage.
3.3.4. Influence of the capillary size on the 1D-data noise level
Ultimately, the goal for the structural bioSAXS experiments is to collect the optimal SAXS pattern (with the lowest noise possible) from a given sample volume, and the noise level for the averaged SAXS curves is discussed in detail below.
Fig. 6(a) shows the frame-averaged SAXS curves 〈I(s)〉 for the lysozyme sample obtained in both capillaries together with smoothed AUTOGNOM curves for Q = 20 µl s−1. The scattering intensity per frame for the di = 1.7 mm capillary is higher than for the di = 0.9 mm capillary as discussed before (§3.2) but the noise level is changed due to the larger number of frames used for averaging. The noise in the averaged curves (Figs. 6a and 6b) is more clear in the normalized difference between the experimental data and the smoothed curve, Δ〈I(s)〉 [see equation (2), Fig. 6c]. As can be seen, Δ〈I(s)〉 is fluctuating around zero, whereas the amplitude of these fluctuations is slightly larger for di = 1.7 mm.
To quantify the differences, the parameter DEV has been computed for all 〈I(s)〉 following equation (3), and the results are shown in Fig. 6(d). For the fixed sample (Q = 0 µl s−1), DEV is higher in di = 0.9 mm because the radiation damage is already visible after 45 ms in both capillaries and the highest signal is obtained for the optimal path length. As mentioned before, radiation damage present already for the first frames cannot be ruled out and therefore this case is not considered further. For the flowing samples, DEV is smaller for di = 0.9 mm than for di = 1.7 mm, indicating a lower relative noise. The lowest values are reached for Q = 20 µl s−1 and 40 µl s−1, the rates that also yield both similar strong total scattering intensities Iint and the times tav without radiation damage.
The useful data range as well as the optimum number of Shannon channels obtained for the two capillaries at the different flow rates do not show an obvious dependence on the capillary diameter. For fixed samples, the useful s-range is limited to sopt < 4.0 nm−1 and MS = 8. When flowing the protein solutions, the scattering intensity at larger angles can be used to determine structural information (sopt < 4.5 nm−1 and MS = 10).
For the more stable BSA samples, which are under-exposed for the di = 0.9 mm capillary, the basic findings are similar (§S4, Fig. S4). Here, the difference between both capillaries is not as pronounced as for the lysozyme samples because of similar strong integrated scattering intensity.
4. Conclusion
We have shown that for highly radiation-sensitive samples, such as biological macromolecules in solution, the use of cylindrical capillaries with path length smaller than the theoretically calculated optimum can yield a more efficient sample consumption and better quality SAXS data under continuous flow of the samples through the X-ray beam. This effect has been demonstrated by comparing two capillary diameters (di = 1.7 mm, close to optimal for 10 keV; di = 0.9 mm, sub-optimal) at various volume flow rates.
The determined SAXS curves for the different settings have been analysed in terms of sample consumption, onset of radiation damage, integrated scattering intensity, and noise level. Based on these parameters, it was found that capillary diameters of di = 0.9 mm yield better SAXS curves for highly radiation-sensitive samples compared with those of di = 1.7 mm. For the smaller diameter the sample consumption is more efficient allowing for more repetitions using a given sample volume. For the same volume flow rates, higher flow speeds are achieved in smaller capillaries, which leads to a faster replenishing of the sample thus diminishing the radiation damage. The larger number of damage-free single SAXS curves and, thus, the overall higher integrated scattering intensity outweighs the effect of the sub-optimal path length. Based on these results, the use of smaller capillaries can be advantageous for standard batch mode measurements, as performed at several bioSAXS beamlines. These findings are confirmed by additional measurements performed at a different SAXS beamline (BM29, ESRF) with different experimental parameters (see §S5).
In contrast to macromolecular crystallography where the radiation damage at cryo-temperatures depends directly on the dose deposited on the sample and can thus be taken into account before the measurement (Garman, 2010), in bioSAXS the damage strongly depends on the protein's tendency to form aggregates, which is not necessarily known a priori and may also be buffer-dependent. Optimizing the data collection, as shown here, will allow SAXS data to be safely obtained with reduced radiation damage.
Based on the comparison of the data from lysozyme and BSA under different flow rates, two scenarios were found: (i) the sample exhibits significant radiation damage and is thus heavily over-exposed; (ii) the sample does not show large damage and is under-exposed. The optimum condition would be the situation in which as many SAXS curves as possible are collected that are free from radiation damage. A criterion for this could be to compare the exposure time without radiation damage, tav, and the total exposure time, ttot, for different volume flow rates. Whereas in the case of di = 0.9 mm for lysozyme an optimum condition might be present for slightly lower flow rates than Q = 40 µl s−1 and increased numbers of SAXS curves, in the case of BSA even slightly lower flow rates than Q = 10 µl s−1 can yield longer life times. The quantities tav and ttot as well as Q can be easily determined and changed during the SAXS measurement on a given sample to yield the optimum conditions providing the longest in-beam integrity. Alternatively, the sample could be flown at high speed; interactive data analysis would indicate if the sample is damaged. If this is not the case, the measurement could be repeated using the same sample until radiation damage is noticed. This approach is basically not limited to linear flows but might also be used with oscillating flows.
The presented results are of special relevance for the beamlines conducting bioSAXS measurements on solutions, both with and without automatic sample changers. The continuous decrease of X-ray beam sizes into the micrometre regime and the development of more brilliant X-ray sources has spurred the use of microfluidic devices, in particular to study rapid mixing and shearing of protein solutions with limited sample consumption (Brennich et al., 2011; Graceffa et al., 2013; Pham et al., 2017; Schwemmer et al., 2016; Skou et al., 2014b; Wieland et al., 2016). Due to an even smaller path length down to several tens of micrometres and the strong background contribution from scattering of the wall material, optimized schemes for data collection, like the one proposed here, would be of importance.
Optimal designs for flow-through environments would be highly asymmetric with a thickness of dopt and a vertical dimension in the range of the X-ray beam, i.e. several micrometres. However, the reproducible fabrication of such a device is a non-trivial task. Moreover, for the selection of the proper material the mechanical stability, vacuum capability, radiation robustness, as well as a low background signal over a large s-range have to be taken into account. Using standard cylindrical glass capillaries with a reduced diameter is an easy alternative to other miniaturization approaches. Overall, the use of smaller capillary appears as the most straightforward way to conduct measurements and obtain the high-quality data free from radiation damage.
5. Related literature
The following reference, not citied in the main body of the paper, has been cited in the supporting information: Henke et al. (1993).
Supporting information
Sections S1 to S5; Tables S1 to S2; Figures S1 to S5. DOI: https://doi.org/10.1107/S1600577518007907/co5096sup1.pdf
Acknowledgements
The authors thank Tomer Cohen for support during the BM29 measurements.
Funding information
The following funding is acknowledged: Bundesministerium für Bildung und Forschung/Röntgen-Ångström cluster project `TT-SAS' (grant No. 05K16YEA to MAS); Deutsche Forschungsgemeinschaft (grant No. GZ: SV 9/5-1 to AYG); EMBL Interdisciplinary Postdoc Programme (EIPOD) und Marie Curie COFUND (award to MAG); Horizon 2020: iNEXT (grant No. 653706).
References
Acerbo, A. S., Cook, M. J. & Gillilan, R. E. (2015). J. Synchrotron Rad. 22, 180–186. Web of Science CrossRef CAS IUCr Journals Google Scholar
Blanchet, C. E., Spilotros, A., Schwemmer, F., Graewert, M. A., Kikhney, A., Jeffries, C. M., Franke, D., Mark, D., Zengerle, R., Cipriani, F., Fiedler, S., Roessle, M. & Svergun, D. I. (2015). J. Appl. Cryst. 48, 431–443. Web of Science CrossRef CAS IUCr Journals Google Scholar
Brennich, M. E., Nolting, J. F., Dammann, C., Nöding, B., Bauch, S., Herrmann, H., Pfohl, T. & Köster, S. (2011). Lab Chip, 11, 708–716. Web of Science CrossRef CAS PubMed Google Scholar
Brooks-Bartlett, J. C., Batters, R. A., Bury, C. S., Lowe, E. D., Ginn, H. M., Round, A. & Garman, E. F. (2017). J. Synchrotron Rad. 24, 63–72. Web of Science CrossRef IUCr Journals Google Scholar
Classen, S., Hura, G. L., Holton, J. M., Rambo, R. P., Rodic, I., McGuire, P. J., Dyer, K., Hammel, M., Meigs, G., Frankel, K. A. & Tainer, J. A. (2013). J. Appl. Cryst. 46, 1–13. Web of Science CrossRef CAS IUCr Journals Google Scholar
David, G. & Pérez, J. (2009). J. Appl. Cryst. 42, 892–900. Web of Science CrossRef CAS IUCr Journals Google Scholar
Fischetti, R. F., Rodi, D. J., Mirza, A., Irving, T. C., Kondrashkina, E. & Makowski, L. (2003). J. Synchrotron Rad. 10, 398–404. Web of Science CrossRef CAS IUCr Journals Google Scholar
Franke, D., Jeffries, C. M. & Svergun, D. I. (2015). Nat. Methods, 12, 419–422. Web of Science CrossRef CAS PubMed Google Scholar
Franke, D., Kikhney, A. G. & Svergun, D. I. (2012). Nucl. Instrum. Methods Phys. Res. A, 689, 52–59. Web of Science CrossRef CAS Google Scholar
Franke, D., Petoukhov, M. V., Konarev, P. V., Panjkovich, A., Tuukkanen, A., Mertens, H. D. T., Kikhney, A. G., Hajizadeh, N. R., Franklin, J. M., Jeffries, C. M. & Svergun, D. I. (2017). J. Appl. Cryst. 50, 1212–1225. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garman, E. F. (2010). Acta Cryst. D66, 339–351. Web of Science CrossRef CAS IUCr Journals Google Scholar
Garrison, W. M. (1987). Chem. Rev. 87, 381–398. CrossRef CAS Web of Science Google Scholar
Gasteiger, E., Hoogland, C., Gattiker, A., Duvaud, S., Wilkins, M. R., Appel, R. D. & Bairoch, A. (2005). Protein Identification and Analysis Tools on the ExPASy Server, in The Proteomics Protocols Handbook, edited by J. M. Walker, pp. 571–607. Totowa: Humana Press. Google Scholar
Graceffa, R., Nobrega, R. P., Barrea, R. A., Kathuria, S. V., Chakravarthy, S., Bilsel, O. & Irving, T. C. (2013). J. Synchrotron Rad. 20, 820–825. Web of Science CrossRef CAS IUCr Journals Google Scholar
Graewert, M. A. & Svergun, D. I. (2013). Curr. Opin. Struct. Biol. 23, 748–754. Web of Science CrossRef CAS PubMed Google Scholar
Guinier, A. (1939). Ann. Phys. 11, 161–237. CrossRef Google Scholar
Hajizadeh, N. R., Franke, D. & Svergun, D. I. (2018). J. Synchrotron Rad. 25, 906–914. Web of Science CrossRef IUCr Journals Google Scholar
Henke, B. L., Gullikson, E. M. & Davis, J. C. (1993). At. Data Nucl. Data Tables, 54, 181–342. CrossRef CAS Web of Science Google Scholar
Hopkins, J. B. & Thorne, R. E. (2016). J. Appl. Cryst. 49, 880–890. Web of Science CrossRef CAS IUCr Journals Google Scholar
Inoue, K., Doutch, J. & Terrill, N. (2013). Bunseki Kagaku, 62, 565–570. Web of Science CrossRef Google Scholar
Jeffries, C. M., Graewert, M. A., Blanchet, C. E., Langley, D. B., Whitten, A. E. & Svergun, D. I. (2016). Nat. Protoc. 11, 2122–2153. CrossRef CAS PubMed Google Scholar
Jeffries, C. M., Graewert, M. A., Svergun, D. I. & Blanchet, C. E. (2015). J. Synchrotron Rad. 22, 273–279. Web of Science CrossRef CAS IUCr Journals Google Scholar
Kirby, N., Cowieson, N., Hawley, A. M., Mudie, S. T., McGillivray, D. J., Kusel, M., Samardzic-Boban, V. & Ryan, T. M. (2016). Acta Cryst. D72, 1254–1266. Web of Science CrossRef IUCr Journals Google Scholar
Kirby, N. M., Mudie, S. T., Hawley, A. M., Cookson, D. J., Mertens, H. D. T., Cowieson, N. & Samardzic-Boban, V. (2013). J. Appl. Cryst. 46, 1670–1680. Web of Science CrossRef CAS IUCr Journals Google Scholar
Konarev, P. V. & Svergun, D. I. (2015). IUCrJ, 2, 352–360. Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
Kuwamoto, S., Akiyama, S. & Fujisawa, T. (2004). J. Synchrotron Rad. 11, 462–468. Web of Science CrossRef CAS IUCr Journals Google Scholar
Li, N., Li, X., Wang, Y., Liu, G., Zhou, P., Wu, H., Hong, C., Bian, F. & Zhang, R. (2016). J. Appl. Cryst. 49, 1428–1432. Web of Science CrossRef CAS IUCr Journals Google Scholar
Maleknia, S. D., Ralston, C. Y., Brenowitz, M. D., Downard, K. M. & Chance, M. R. (2001). Anal. Biochem. 289, 103–115. Web of Science CrossRef PubMed CAS Google Scholar
Meisburger, S. P., Warkentin, M., Chen, H., Hopkins, J. B., Gillilan, R. E., Pollack, L. & Thorne, R. E. (2013). Biophys. J. 104, 227–236. Web of Science CrossRef CAS PubMed Google Scholar
Nielsen, S. S., Møller, M. & Gillilan, R. E. (2012). J. Appl. Cryst. 45, 213–223. Web of Science CrossRef CAS IUCr Journals Google Scholar
Pernot, P., Round, A., Barrett, R., De Maria Antolinos, A., Gobbo, A., Gordon, E., Huet, J., Kieffer, J., Lentini, M., Mattenet, M., Morawe, C., Mueller-Dieckmann, C., Ohlsson, S., Schmid, W., Surr, J., Theveneau, P., Zerrad, L. & McSweeney, S. (2013). J. Synchrotron Rad. 20, 660–664. Web of Science CrossRef CAS IUCr Journals Google Scholar
Petoukhov, M. V., Franke, D., Shkumatov, A. V., Tria, G., Kikhney, A. G., Gajda, M., Gorba, C., Mertens, H. D. T., Konarev, P. V. & Svergun, D. I. (2012). J. Appl. Cryst. 45, 342–350. Web of Science CrossRef CAS IUCr Journals Google Scholar
Pham, N., Radajewski, D., Round, A., Brennich, M., Pernot, P., Biscans, B., Bonneté, F. & Teychené, S. (2017). Anal. Chem. 89, 2282–2287. Web of Science CrossRef Google Scholar
Rogers, D. F. (1992). Laminar Flow Analysis. Cambridge University Press. Google Scholar
Rott, N. (1990). Annu. Rev. Fluid Mech. 22, 1–12. CrossRef Web of Science Google Scholar
Round, A., Felisaz, F., Fodinger, L., Gobbo, A., Huet, J., Villard, C., Blanchet, C. E., Pernot, P., McSweeney, S., Roessle, M., Svergun, D. I. & Cipriani, F. (2015). Acta Cryst. D71, 67–75. Web of Science CrossRef IUCr Journals Google Scholar
Sahle, C. J., Henriquet, C., Schroer, M. A., Juurinen, I., Niskanen, J. & Krisch, M. (2015). J. Synchrotron Rad. 22, 1555–1558. Web of Science CrossRef IUCr Journals Google Scholar
Schroer, M. A. & Svergun, D. I. (2018). Emerg. Top. Life Sci. 2, 69–79. CrossRef Google Scholar
Schwemmer, F., Blanchet, C. E., Spilotros, A., Kosse, D., Zehnle, S., Mertens, H. D. T., Graewert, M. A., Rössle, M., Paust, N., Svergun, D. I., von Stetten, F., Zengerle, R. & Mark, D. (2016). Lab Chip, 16, 1161–1170. Web of Science CrossRef CAS Google Scholar
Skou, M., Skou, S., Jensen, T. G., Vestergaard, B. & Gillilan, R. E. (2014b). J. Appl. Cryst. 47, 1355–1366. Web of Science CrossRef IUCr Journals Google Scholar
Skou, S., Gillilan, R. E. & Ando, N. (2014a). Nat. Protoc. 9, 1727–1739. Web of Science CrossRef Google Scholar
Svergun, D. I. & Koch, M. H. J. (2003). Rep. Prog. Phys. 66, 1735–1782. Web of Science CrossRef CAS Google Scholar
Svergun, D. I., Koch, M. H. J., Timmins, P. A. & May, R. P. (2013). Small Angle X-ray and Neutron Scattering from Solutions of Biological Macromolecules. Oxford University Press. Google Scholar
Wieland, D. C. F., Garamus, V. M., Zander, T., Krywka, C., Wang, M., Dedinaite, A., Claesson, P. M. & Willumeit-Römer, R. (2016). J. Synchrotron Rad. 23, 480–486. Web of Science CrossRef CAS IUCr Journals Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.