research papers
Characterization of a submicro-X-ray fluorescence setup on the B16 beamline at Diamond Light Source
aAtominstitut, TU Wien, Vienna, Austria, bLaboratory of Environmental Chemistry and Bioanalytics, Department of Analytical Chemistry, Institute of Chemistry, Eötvös Loránd University, Budapest, Hungary, cDiamond Light Source Ltd, Harwell Science and Innovation Campus, Didcot, UK, and dKarlsruhe Institute of Technology (KIT), Karlsruhe, Germany
*Correspondence e-mail: mirjam.rauwolf@tuwien.ac.at, anna.turyanskaya@tuwien.ac.at
An −1 fg−1 and 4 ag in 1000 s, respectively (for As Kα excited with 17 keV), are presented in order to demonstrate the capabilities of this setup. Sample measurements of a human bone and a single cell performed at B16 are presented in order to illustrate the suitability of the setup in biological applications.
setup has been tested on the B16 beamline at the Diamond Light Source synchrotron with two different excitation energies (12.7 and 17 keV). This setup allows the scanning of thin samples (thicknesses up to several micrometers) with a sub-micrometer resolution (beam size of 500 nm × 600 nm determined with a 50 µm Au wire). Sensitivities and detection limits reaching values of 249 counts sKeywords: SR submicro-XRF; beamline B16; elemental imaging; scanning; human bone; cells.
1. Introduction
Synchrotron-radiation-induced et al., 2010; Holt et al., 2013; West et al., 2017). SR-XRF imaging can be performed in scanning or in full-field mode. While the ideal analysis mode depends on the elements of interest and their concentration, in our experience imaging of trace elements works best in scanning mode. In our previous study we demonstrated that, in bone, Ca (the major element of bone) was nicely imaged with both a full-field color X-ray camera setup at the BAMline at BESSY II and at the scanning confocal micro-XRF setup at the FLUO beamline at ANKA. However, whereas Zn (a in bone) was quite easy to image in scanning mode, it was barely possible to detect in full-field mode. Therefore, the scanning mode is our preferred choice for imaging bone samples (Rauwolf et al., 2017). Nowadays, the smallest beam sizes (and therefore highest resolution) available for scanning SR-XRF imaging reach a few tens of nanometers (Martínez-Criado et al., 2016; Lemelle et al., 2017). However, for many biological samples such high-resolution beams are already too small to scan meaningful areas of interest on the samples on a reasonable time scale. For instance, the samples of our interest are bone tissue and cells. Within the bone we are mostly interested in osteons (diameter of 200 µm). Diameters of the cells can vary between 10 and 120 µm. As one can imagine, scanning such an object with such a small beam would require considerable time. Therefore, submicro-beam would be a perfect choice, allowing measurements of such structures and, moreover, allowing resolving of substructures, such as cement lines in bone (several micrometers thick). To find the ideal SR-XRF imaging method for the investigation of distributions and possible substructures (in the micrometer range) in biological samples, such as bone, single cells and other tissues (Ugarte et al., 2016; Mihucz et al., 2016), SR-XRF at the B16 beamline at Diamond Light Source was tested with different samples and standards. Test measurements were performed with two different excitation energies, 12.7 and 17 keV.
(SR-XRF) analysis with micro- or nanometer resolution is a powerful method for non-destructively investigating element distribution in a wide variety of samples (Janssens2. Materials and methods
2.1. SR-XRF setup
The SR-XRF setup was installed for multiple experimental sessions at the B16 bending magnet test beamline of the Diamond Light Source synchrotron (see Fig. 1), a beamline committed to be easily adaptable and adjustable to the particular requirements of various experiments (Sawhney et al., 2010, 2011). The Diamond Light Source storage ring operates in top-up mode (3 GeV electron energy, 300 mA ring current). For the SR submicro-XRF experiment, the X-ray energy was set with an RuB4C double-multilayer monochromator (FMB Oxford Ltd, Oxford, UK), which can be used to produce a high-flux monochromatic beam in the range 8–20 keV (Diamond, 2017a). The beam was focused with a Kirkpatrick–Baez (KB) mirror pair (Diamond Light Source Ltd, in-house design and construction). The sample stage consisted of piezo crystal-based motors (each of the xyz axes is an ECS5050 piezo stage; Attocube Systems AG, Munich, Germany) and a high-purity methyl acrylate sample holder, which was 3D-printed from a clear photoreactive resin (Form 2, Formlabs Inc., Somerville, USA) at Diamond Light Source. The angle between the primary beam and sample was 45°. Additionally, the setup was equipped with an optical microscope (Optem Fusion, Qioptiq), which was focused onto the back side of the samples (see Fig. 1). This allowed identification of the regions of interest quickly. The XRF radiation was detected at an angle of 45° to the sample or 90° to the primary beam. Note that, due to the setup's geometry, in particular the angle of 45° between primary beam and sample, the horizontal resolution worsens with increasing sample thickness. Additionally, as the microscope sees the sample from the back side, the sample should be transparent in order to find the areas of interest. Therefore, the recommended sample thickness should be in a range up to several micrometers.
The size of the focused beam was determined using a knife-edge scan of a 50 µm Au wire. Measurements of standards were performed at 17 keV and 12.7 keV to optimize the excitation for different elements (once for elements up to Sr and once for those up to Zn).
2.1.1. Setup for 17 keV
The 17 keV measurements were recorded with a Vortex four-element detector (Vortex-ME4, Hitachi Ltd, Tokyo, Japan) with a total active area of 200 µm2 and SII electronics. The beam resolution (full width at half-maximum, FWHM) determined with a 50 µm Au cross was about 500 nm (vertical) × 600 nm (horizontal).
To characterize the setup, especially for thin samples, a NIST standard (Standard Reference Material 2783 – Air Particulate on Filter Media; NIST, Gaithersburg, USA) was measured. This standard is quite heterogeneous as one can see from the distribution maps of Ca, Fe, Cr and Zn in Fig. 2. The certified values can only be considered to be valid within the stated uncertainty if an area of at least 1 cm2 is measured (NIST, 2011). As one can easily see, this is too large an area to analyze with a sub-micrometer beam. Additionally, a measurement time of 120 s per pixel was required for this standard to produce an evaluable spectrum. To obtain sensitivity and lower limits of detection in reasonable time, the area scanned was limited to 6 × 6 pixels (considering our beam size this results in an area of about 10.8 µm2). Between each point measured the sample was moved by 10 µm (step size). Additionally to the loaded filter, a blank filter (also part of the NIST SRM 2783) was scanned in the same way. This allowed determination of the components of the signal that originate from the blank filter alone. Sum spectra for the loaded and the blank filters (see Fig. 2, top left) were fitted using AXIL-QXAS, version 3.6 (IAEA, 2009). Sensitivities and lower limits of detection (LLD) were calculated using the fitted result of the loaded filter corrected with the net counts of the blank filter. The uncertainties for the sensitivities were calculated by using the uncertainties (about 95%) given in the NIST certificate (NIST, 2011) as well as two times the fit error (2σ) of the fitted net counts of the loaded filter. Note that, due to the considerably smaller measured area (compared with the given recommendation), the error bars shown in Fig. 2 do not correspond to a confidence interval of 95% and are only given as a rough indication. As one can see from Fig. 2, the sensitivities obtained ranged from about 8 counts s−1 fg−1 for Ca Kα to 249 counts s−1 fg−1 for As Kα, while lower limits of detection ranged from 169 to 4 ag in 1000 s. Sensitivity values shown in Fig. 2 can also be found in Table S1 of the supporting information.
To put these values into perspective, a comparison with the LLDs for Mn and Cu reached at the ID16B-NA beamline at the ESRF is given in Table 1.
|
2.1.2. Setup for 12.7 keV
The measurements with an 2 active area; Canberra counting electronics). The beam resolution (FWHM) was also about 500 nm (vertical) × 600 nm (horizontal).
of 12.7 keV were performed with a different detector, a single silicon drift detector (SDD) (Vortex 90EX, Hitachi Ltd, Tokyo, Japan; 50 mmWith the same step size and measurement time as described in §2.1.1, the NIST standard SRM 2783 (both loaded and blank filters) (NIST, 2011) was also measured with 12.7 keV In total, 41 spectra (41 × 1 pixels) were recorded and the results are shown in Fig. 3. The sensitivities obtained ranged from about 23 counts s−1 fg−1 for Ca Kα to 396 counts s−1 fg−1 for Zn Kα while lower limits of detection ranged from 92 to 5 ag in 1000 s. For the 12.7 keV excitation it was also possible to obtain sensitivity and LLD values for Ni Kα. This was not possible for the 17 keV excitation due to the contributions to the Ni signal from the blank filter and the lower sensitivities. The high sensitivity and LLD values for Cr Kα are caused by a single Cr hotspot in the scan. In this hotspot Cr reaches 16408 counts in 60 s while all other pixels in the scan have values between 22 and 1046 counts in 60 s. The sensitivity values presented in Fig. 3 can also be found in Table S2 of the supporting information.
To further characterize the B16 submicro-XRF setup, sensitivities and LLD (in µg g−1) were also determined for a thick homogeneous standard reference material (NIST SRM 1412 – Multicomponent Glass; NIST, Gaithersburg, USA) (National Bureau of Standards, 1985). Again, a scan area of 6 × 6 pixels was measured (with a step size of 10 µm). Measurement time was 20 s. Sensitivities for K- and L-lines [in counts s−1 (µg g−1)−1] as well as the sum spectrum and the LLD (µg g−1 in 1000 s) are shown in Figs. 4 and 5, respectively.
A comparison of the LLDs for Ca, Fe and Cu reached at the P06 beamline at PETRA III can be seen in Table 2.
|
3. Scanned samples – results
To show the potential of the XRF setup on B16, scans from three projects are presented in this section. The analysis of three different sample types was carried out at 17 keV. An Au test structure was measured to show the resolution capabilities of the setup. The results of an area measured in bone and the elemental maps gained by measuring a single cell demonstrate the potential of the XRF setup tested at B16 for biological samples.
3.1. Gold test structure
This sample was produced for the Atominstitut (ATI) by the Karlsruhe Nano Micro Facility (KNMF) by means of electron beam lithography. A four-inch silicon wafer was coated with a start layer (10 nm Ti, 100 nm Au) for electroplating by vapor deposition. The structures were written as trenches in a 1.5 µm-thick PMMA layer with an electron beam writer (EBPG5200Z, Raith). The dose was 800 µC cm−2. Development took place in a mixture of MIBK and IPA (1:1). The trenches had been filled with gold of thickness 1 mm by electroplating. Residual resist was removed by oxygen plasma. The sample is a Ti-coated (over the whole surface) Si wafer. On top of the Ti layer there are Au structures in the shape of 100 µm × 100 µm squares and consecutive stripes of different widths between 1 and 10 µm. A microscope image of the sample is shown in Fig. 6.
Two areas of this sample were scanned with the XRF setup on B16: the square shown at the bottom right (of the microscope image, Fig. 6) as well as parts of the line structure on the bottom left. From top to bottom, the first stripe of the line structure has a width of 10 µm and is followed by a gap of 10 µm; the second stripe has a width of 9 µm, followed by a 9 µm gap. The stripe widths and stripe gaps decrease in 1 µm steps down to the smallest stripe which has a width of 1 µm. Both structures were scanned with a step size of 1 µm and a measurement time of 1 s. Measurements for the two Au maps took about 50 min for the stripes and 8 h for the square. The time given above includes the time needed for the communication of the measurement software with the detector and motor movements time (about 1 s per pixel). The results from the scanned Au structure confirm that 1 µm structures are easily resolved with the XRF setup on B16.
3.2. Human bone biopsy sample
The sample is a vertebral body biopsy and was prepared as a 4 µm (thickness) cut. This cut was sandwiched between two 8 µm Kapton foils and fastened between Plexiglas frames. The bone sample was thus fixed on the sample carrier, which can be seen in Fig. 1.
Fig. 7 shows a microscope image of the area of the sample measured (top left), the Ca distribution of an overview scan (bottom left) and the Ca, Zn, Sr and Co distribution of the final high-resolution scan. The overview scan was performed with a step size of 20 µm and a measurement time per pixel of 2 s. The overview area was 400 µm × 360 µm and the scan took about 20 min.
The high-resolution scan (step size 1 µm) was performed in the area marked with a blue frame in the microscope image and was optimized for measuring Zn and therefore a considerably longer measurement time (20 s per pixel) was needed. The whole scan took about 15 h. The scan area was 11 µm × 237 µm. Fig. 7 demonstrates that the setup is not only capable of imaging thin structures in bone samples (especially visible in the Zn map) but also that the resulting elemental maps can easily be correlated to microscope images using rough scans.
3.3. Cancer cell
These measurements were performed as part of a project investigating the localization of metals within cells. Human A-375 malignant melanoma cells (ECACC, Salisbury, UK) were grown on 7.5 mm × 7.5 mm low-stress silicon nitride windows with a thickness of 500 nm (Norcada, Edmonton, AB, Canada). Cells were treated with copper sulfate and chelators. A more detailed description of the sample treatment is given by Gaál et al. (2018a,b).
Generally, a maximum of 100 cells per total area is sufficient. Usually a cell suspension of half a million cells in 2 ml of medium is prepared and applied to the plate. The cells will adhere to the membrane. If too many cells have been placed on the membrane, the process should be started from the beginning; if too little, then more cells should be pipetted. If this method is used, we can scan individual cells and small groups of cells on the same membrane.
The cell sample area was measured for 5 s per pixel with a step size of 0.5 µm. The area size was 20 µm × 20 µm (41 × 41 pixels). The complete scan took about 2.8 h. The elemental maps in Fig. 8 show the suitability of the setup for investigating elemental distributions within a single cell.
4. Conclusion
We characterized an XRF setup on beamline B16 at Diamond Light Source using two different excitation energies (12.7 and 17 keV) as well as two different standard samples. The beam size was determined as 500 nm × 600 nm and thus the 1 µm structures of the Au test sample were easily resolved. For the −1 fg−1 for Ca K to 249 counts s−1 fg−1 for As K. The lower limits of detection ranged from 169 ag (Ca) to 4 ag (As) in 1000 s. For 12 keV, sensitivities obtained ranged from about 23 counts s−1 fg−1 for Ca K to 396 counts s−1 fg−1 for Zn K. For the lower limits of detection, values from 92 ag (Ca) to 5 ag (Zn) in 1000 s were achieved. Although the detection limits presented do not reach the impressive values of the P06 beamline at PETRA III (due to the considerably lower flux) and of ID16B-NA at ESRF (due to the larger beam size), the setup is very recommendable for imaging biological samples, which we have illustrated by means of scans on a bone sample and a single cancer cell. Additionally, it should be mentioned that at Diamond Light Source there exists the possibility to first scan a larger sample area at I18 with micrometer resolution (for example 2 × 2 µm) followed by further investigation of smaller areas of interest with sub-micrometer resolution at B16 (Ugarte et al., 2016). In conclusion, the XRF setup on the B16 beamline is very well suited for studying thin samples and especially for biological applications.
of 17 keV the following sensitivities were obtained: 8 counts sSupporting information
Tables S1 and S2: Calculation of the sensitivities for 17 keV and 12.7 keV. DOI: https://doi.org/10.1107/S1600577518006203/pp5117sup1.pdf
Acknowledgements
We thank Diamond Light Source for access to beamline B16 (MT12705-1, MT12869-1, MT15267-1, MT16260-1) that contributed to the results presented here. We thank Nikul Khunti at the B21 SAXS beamline at Diamond Light Source for 3D-printing the sample carrier. We acknowledge the Karlsruhe Nano Micro Facility (KNMF, https://www.kit.edu/knmf) of the Karlsruhe Institute of Technology (KIT) for the production of the structured gold sample described in §3.1. We would like to thank Paul Roschger for the provision of the bone sample and Jan Matthew Welch for the constructive criticism of the manuscript.
Funding information
Funding for this research was provided by: Austrian Science Fund (grant No. P27715 to CS).
References
Boesenberg, U., Ryan, C. G., Kirkham, R., Siddons, D. P., Alfeld, M., Garrevoet, J., Núñez, T., Claussen, T., Kracht, T. & Falkenberg, G. (2016). J. Synchrotron Rad. 23, 1550–1560. Web of Science CrossRef CAS IUCr Journals Google Scholar
Diamond (2017a). B16 – Facility & Equipment Availability, https://www.diamond.ac.uk/Beamlines/Materials/B16/status.html. Google Scholar
Diamond (2017b). B16 – beamline schematic, https://www.diamond.ac.uk/Beamlines/Materials/B16/schematic.html. Google Scholar
Gaál, A., Mihucz, V. G., Bosze, S., Szabó, I., Baranyi, M., Horváth, P., Streli, C. & Szoboszlai, N. (2018a). Microchem. J. 136(Suppl. C), 227–235. Google Scholar
Gaál, A., Orgován, G., Mihucz, V. G., Pape, I., Ingerle, D., Streli, C. & Szoboszlai, N. (2018b). J. Trace Elem. Med. Biol. 47, 79–88. Google Scholar
Holt, M., Harder, R., Winarski, R. & Rose, V. (2013). Annu. Rev. Mater. Sci. 43, 183–211. Web of Science CrossRef CAS Google Scholar
IAEA (2009). Quantitative X-ray Analysis System, No. 21 in Computer Manual Series. Vienna: IAEA. (https://www-pub.iaea.org/books/IAEABooks/7884/Quantitative-X-ray-Analysis-System.) Google Scholar
Janssens, K., De Nolf, W., Van Der Snickt, G., Vincze, L., Vekemans, B., Terzano, R. & Brenker, F. E. (2010). Trends Anal. Chem. 29, 464–478. Web of Science CrossRef Google Scholar
Laforce, B., Schmitz, S., Vekemans, B., Rudloff, J., Garrevoet, J., Tucoulou, R., Brenker, F. E., Martinez-Criado, G. & Vincze, L. (2014). Anal. Chem. 86, 12369–12374. Web of Science CrossRef CAS PubMed Google Scholar
Lemelle, L., Simionovici, A., Schoonjans, T., Tucoulou, R., Enrico, E., Salomé, M., Hofmann, A. & Cavalazzi, B. (2017). Trends Anal. Chem. 91(Suppl. C), 104–111. Google Scholar
Martínez-Criado, G., Villanova, J., Tucoulou, R., Salomon, D., Suuronen, J.-P., Labouré, S., Guilloud, C., Valls, V., Barrett, R., Gagliardini, E., Dabin, Y., Baker, R., Bohic, S., Cohen, C. & Morse, J. (2016). J. Synchrotron Rad. 23, 344–352. Web of Science CrossRef IUCr Journals Google Scholar
Mihucz, V. G., Meirer, F., Polgári, Z., Réti, A., Pepponi, G., Ingerle, D., Szoboszlai, N. & Streli, C. (2016). J. Biol. Inorg. Chem. 21, 241–249. Web of Science CrossRef Google Scholar
National Bureau of Standards (1985). Standard Reference Material 1412 – Multicomponent Glass, https://www-s.nist.gov/srmors/certificates/1412.pdf. Google Scholar
NIST (2011). Standard Reference Material 2783 – Air Particulate on Filter Media, https://www-s.nist.gov/srmors/certificates/2783.pdf. Google Scholar
Rauwolf, M., Turyanskaya, A., Roschger, A., Prost, J., Simon, R., Scharf, O., Radtke, M., Schoonjans, T., Guilherme Buzanich, A., Klaushofer, K., Wobrauschek, P., Hofstaetter, J. G., Roschger, P. & Streli, C. (2017). J. Synchrotron Rad. 24, 307–311. Web of Science CrossRef IUCr Journals Google Scholar
Sawhney, K. J. S., Dolbnya, I. P., Scott, S. M., Tiwari, M. K., Preece, G. M., Alcock, S. G. & Malandain, A. W. (2011). Proc. SPIE, 8139, 813908. CrossRef Google Scholar
Sawhney, K. J. S., Dolbnya, I. P., Tiwari, M. K., Alianelli, L., Scott, S. M., Preece, G. M., Pedersen, U. K. & Walton, R. D. (2010). AIP Conf. Ser. 1234, 387–390. Google Scholar
Ugarte, M., Geraki, K. & Jeffery, G. (2016). Metallomics, 8, 1071–1080. Web of Science CrossRef Google Scholar
West, M., Ellis, A. T., Streli, C., Vanhoof, C. & Wobrauschek, P. (2017). J. Anal. At. Spectrom. 32, 1629–1649. Web of Science CrossRef Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.