Calculated energy-dependent transmission of the (n = 3) harmonic X-rays using a modified scheme with four diffractions by the Si(111) monochromator crystals, as well as four diffractions by the Si(220) harmonic-rejection crystal pair. The harmonic transmission is defined as the flux ratio of the outgoing to the incoming harmonic radiation incident at the harmonic-rejection crystal pair. Compared with Fig. 3[link](b), it is apparent that the transmission of the (n = 3) harmonic X-rays is reduced by another five orders of magnitude by introducing two additional reflections in the Si(111) monochromator.