beamlines
Design and performance of bending-magnet beamline BL02B at the SSRF
aShanghai Institute of Applied Physics, Chinese Academy of Sciences, Zhangheng Road 239, Pudong District, Shanghai 201800, People's Republic of China, and bShanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Changning Road 865, Shanghai 200050, People's Republic of China
*Correspondence e-mail: guozhi@sinap.ac.cn, wangyong@sinap.ac.cn, tairenzhong@sinap.ac.cn
The BL02B bending-magnet beamline at the Shanghai Synchrotron Radiation Facility (SSRF) has been constructed and is now operational for ambient-pressure e.g. energy-resolving power and focus spot size. experiments show that the energy range extends from 40 to 2000 eV. Argon, nitrogen and neon gas core-shell excitation spectra indicate energy-resolving powers of over 1.4 × 104 @ 244 eV, 1.0 × 104 @ 401 eV and 7.0 × 103 @ 867 eV, respectively. The measured is 1.3 × 1011 photons s−1 @ E/ΔE = 3700 at 244 eV at the expected sample position, for the SSRF electron energy of 3.5 GeV and electron current of 240 mA. The spot sizes are 177 µm × 23 µm and 150 µm × 46 µm at the APPES and PIPOS samples, respectively.
(APPES) and photon-in/photon-out spectroscopy (PIPOS) experimental use. Optical optimization was implemented for realization of high performance,Keywords: bending-magnet source; photon flux; energy-resolving power; spot size.
1. Introduction
High energy-resolving power and high et al., 2017). The application of APPES can provide information about solid or liquid surfaces in the presence of gases under ambient conditions (Salmeron & Schlögl, 2008; Ogletree et al., 2009). When the of the electron is 100 eV, the photoelectron inelastic mean-free path in a gas with a pressure of 1 Torr is approximately only a millimetre. Therefore, a small X-ray spot with high is necessary for obtaining a high signal-to-noise ratio (Liu et al., 2014). High energy resolution is important for the APPES technique (Zhang et al., 2010; Blomberg et al., 2013), which can be used to study the active adsorption site, formation of a new chemical phase, and possible surface reconstruction. PIPOS includes soft soft and resonant inelastic X-ray scattering (RIXS). These techniques also require a high energy resolution and high to provide detailed information for both occupied and unoccupied electron states in the vicinity of the (Himpsel, 2011).
are essential for study of the electronic structures of new energy-related environmentally friendly materials using ambient pressure (APPES) and photon-in/photon-out spectroscopy (PIPOS) (GuoIn most cases, radiation with high et al., 2012). The number of bending magnets used at the Shanghai Synchrotron Radiation Facility (SSRF) is 40, which is twice as many as that of insertion devices. Therefore, it is crucial to utilize bending magnets to tap the potential of synchrotron radiation. A large acceptance for the divergence angle should be implemented to acquire a relatively high However, the large divergence angle leads to great aberration and, thus, a decreased energy resolution as well. The beamline design shall be optimized to make a balance between the and the energy resolution. The pre-focusing mirror is located as close as possible to the source. To fulfil the requirement of a large acceptance angle, the length of the mirrors and gratings must be large. To prevent aberration, an elliptical cylinder mirror (ECM1) and two ellipsoidal mirrors (EM3 and EM4) are used. One can also limit the coma aberration by optimizing the variable-line-spacing gratings (VLSGs) and demagnification of the beamline.
and coherence is generated using undulators, which can be regarded as scarce and expensive sources (FeserOur experiments show that the BL02B beamline can cover the energy range 40–2000 eV with a 11 photons s−1 @ E/ΔE = 3700 and an energy-resolving power of over 104 at 244 eV. The spot sizes on samples are 177 µm × 23 µm and 150 µm × 46 µm for APPES and PIPOS, respectively. These parameters satisfy the demands for high energy resolution and high in APPES and PIPOS experiments.
of 1.3 × 102. Beamline
2.1. Bending-magnet source
The SSRF is a third-generation synchrotron with an electron energy of 3.5 GeV and a low emittance of 3.9 nm rad. The BL02B source is located on a standard bending-magnet section with a magnetic field of 1.27 T and a bending radius of 9.17 m. The electron current is 300 mA. Due to incoherence from the different positions of the electron ring, within the photon energy from 30 eV to 2500 eV, the angular x-axis (horizon) and y-axis (vertical) can be observed to follow a homogeneous and Gaussian distribution in a certain region, respectively. The spot size (FWHM) and angular divergence (FWHM) in the vertical direction for the bending-magnet radiation is shown in Fig. 1. The spot size decreases from 25.9 µm @ 30 eV to 23.1 µm @ 2500 eV, while the angular divergence decreases from 1620 µrad @ 30 eV to 178 µrad @ 2500 eV.
densities along the2.2. Beamline design
The BL02B beamline layout is shown in Fig. 2(a), with the main parameters for the beamline optics listed in Table 1. The lead-out angle for the front-end (the angle between the beam and the linear section of the storage ring) is 1°. A four-blade aperture (Slit1) is located 8.9 m downstream from the bending-magnet source, which can block undesirable photons and define the acceptance angle for the beamline. To obtain higher a large horizontal divergence angle for the bending-magnet beamline is inevitable. Therefore, an elliptical cylinder mirror (ECM1) is located 11.1 m downstream from the source, which is as close as possible; moreover, the mirror length is over 1 m with a grazing-incident angle of 1.5° and a large acceptance angle of 2 mrad. ECM1 can horizontally focus the beam onto the exit slits located at 32.4 m. There are two exit slits in total: Slit2a for the APPES branch and Slit2b for the PIPOS branch, which are alternated by switching the plane mirror (PM2). The monochromator (Mono) is located at 22 m and consists of one internally water-cooled plane mirror (PM1) and three VLSGs. In addition to energy dispersion, the gratings can also vertically focus the beam onto the exit slit. To obtain high energy resolution and grating diffraction efficiency, three gratings are optimized for varying energy ranges. Gratings with line densities of 400 lines mm−1, 800 lines mm−1 and 1100 lines mm−1 cover the energy ranges 40–600 eV, 200–1600 eV and 200–2000 eV, respectively. Due to the large vertical acceptance angle of 0.8 mrad, the −1st diffraction order and a Cff (fixed focus constant) value of 0.6 is chosen for the whole energy range. Since the ECM1 and VLSG focus the beam onto the exit slit plane in the horizontal and vertical directions, respectively, the X-rays on the exit slit plane can be observed as a second source for the downstream ellipsoidal mirrors (EM3 and EM4). The entrance and exit arm lengths are 7 m and 2 m for EM3, and 7.8 m and 3.2 m for EM4, respectively. The two mirrors can be used to focus the beam onto the sample positions for APPES (41.4 m) and PIPOS (43.4 m).
|
3. Beamline design specifications
3.1. Energy resolution
The monochromator is designed to correct optical aberrations using the light-path function method (Harada & Kita, 1980; Kiyokura et al., 1998). Due to the fixed exit slit, optical aberrations degrade the energy resolution of the grating. The main optical aberrations (Itou et al., 1989; Reininger, 2011; Xue et al., 2014) can be minimized using a line density variation on the grating given by
where k0 is the line density at the grating centre, w is the coordinate along the grating length and b2, b3 and b4 are the varied-line-spacing parameters used to defocus the monochromatic beam, correct the coma aberration and spherical aberration, respectively. The defocus term for the light-path function can be expressed as
where α and β are the incidence and diffraction angles, respectively, r1 is the entrance arm length, r2 is the exit arm length, n is the diffraction order and λ is the wavelength. F20 can be zeroed within the whole energy range using a suitable choice of b2 to simultaneously solve the grating equation and equation (2). This yields a unique pair of angles α and β for each photon energy and grating. The coma term is the primary aberration and can be expressed as
At a certain energy, F30 is zero; for the whole energy range covered by each grating, the effect of the coma aberration can be negligible due to its weak value.
Five factors contribute to the energy resolution, including the source size ΔEso, exit slit size ΔEex, tangential slope error for the grating ΔEgr and plane mirror ΔEpl, and the aberration of the beamline ΔEab. The total energy resolution can be expressed as
The slope errors for the plane mirror and gratings are 0.2 µrad and 0.12 µrad in the meridional direction, respectively. When the exit slit size is chosen to be 15 µm, the resolving powers for each grating are calculated as shown in Fig. 3(a). The contributions to the energy broadening ΔEtotal are the source size, exit slit size, slope error of the plane mirror and grating and beamline aberration in sequence. As seen in the figure, the decreases with increasing energy. At energies up to 2500 eV, the remains greater than 3500. The resolution can be verified using a ray-tracing model called the SHADOW code (Tanaka & Kitamura, 2001). As an example, we show the spot patterns obtained at the exit plane for 250 eV and 250.018 eV with an 1100 lines mm−1 grating in Fig. 3(b). The two patterns can be well separated, which indicates a of 14000. Under this condition, the is 13200 in Fig. 3(a), which is slightly worse due to the contribution of the 15 µm exit slit.
3.2. Photon flux
The . The efficiency of the grating depends on the grating duty ratio, line density and groove depth. Neviere's code is used to optimize the grating parameters and obtain the maximum efficiencies for the gratings (Nevière et al., 1974).
at the sample can be calculated by considering the angular divergence limited by Slit1 and the transmission efficiencies for the mirrors and gratings. The vertical divergence for Slit1 is modified according to variation in photon energy. Due to the finite length of ECM1, the divergence in the horizontal direction should be limited to 2 mrad. The surface roughness of all the optical elements is listed in Table 1The . For the 400 lines mm−1, 800 lines mm−1 and 1100 lines mm−1 gratings, the at the PIPOS sample is lower than that obtained at the APPES sample, since an additional switching mirror is used at the PIPOS branch. For the 400 lines mm−1 grating, the maximum @ 150 eV is higher than 1.3 × 1012 photons s−1. For the 800 lines mm−1 and 1100 lines mm−1 gratings, the is approximately flat at low and moderate energy and decreases rapidly at high energy.
for varying vertical divergence is shown in Fig. 43.3. Spot size at the sample
A small spot with high . The spot sizes at the APPES and PIPOS samples are 111 µm × 13.5 µm (FWHM) at 800 eV and 133 µm × 10 µm (FWHM) at 400 eV, respectively.
at the endstation is required for the APPES and PIPOS experiments. The ellipsoidal mirror can reduce the reflection losses compared with the Kirkpatrick–Baez pair. Therefore, two ellipsoidal mirrors are chosen to focus the beam onto the sample positions for APPES and PIPOS. To decrease the spot size, the demagnification ratio for the mirror is up to 3.5:1, with slope errors of 1 µrad × 5 µrad for both mirrors. The spot sizes at the samples with an exit slit of 35 µm are simulated using ray tracing, as shown in Fig. 54. Experimental results
4.1. Energy resolution
An −6 Torr with an exit slit opening of 15 µm. The measured spectrum was a Voigt profile: a convolution of a naturally broadened Lorentzian ΔL with an instrumental Gaussian width ΓG.
was installed downstream of the exit slit to measure the core-shell excitation spectra for several gases, from which we can obtain the energy resolution. A microchannel plate (MCP) was mounted on the chamber to record the total ion yield spectra. The includes a gas inlet system, detection system and data acquisition and processing system. The advantage of using the is that the gas core-shell ionization spectra can be measured under a very low working pressure so that collision broadening becomes negligible. In our experiments, the spectra were recorded for a gas pressure of 10Fig. 6(a) shows the core-shell excitation spectra for Ar obtained using the 1100 lines mm−1 grating. The Ar L2,3 absorption-edge transitions to the Rydberg levels 2p3/2 → 4s, 3d, 4d, 5d, 6d, 7d and 2p1/2 → 4s, 3d, 4d, 5d, 6d were all observed (King et al., 1977; Xue et al., 2010). The 2p3/2 → 4s transition (hν = 244.39 eV) was used to characterize the energy resolution. Assuming a Lorentzian linewidth ΔL = 112 meV (Prince et al., 1999; Strocov et al., 2010), we obtained a Gaussian width of ΓG = 17 meV ± 2 meV and a of 1.4 × 104, which can be compared with a theoretical of 1.3 × 104.
Fig. 6(b) shows the spectra measured for N2 (dotted line) in the range of the N 1s → π* excitation (Sodhi & Brion, 1984). Six transition energies from N 1s (ν = 0) → π* (ν′:ν′ = 0–5) were clearly observed. By fitting the experiment spectra with a Voigt linewidth ΔL = 113 meV, we obtained a Gaussian width of ΓG = 39 meV ± 1 meV and a of 1.0 × 104, which can be compared with the theoretical value of 1.1 × 104.
Fig. 6(c) shows the spectrum for Ne at the 1s Rydberg series measured with the 1100 lines mm−1 grating. The fine structure resonances 1s → 3p (hν = 867.1 eV), 4p, 5p and 6p were all clearly observed. To estimate the resolution, we fitted the Ne spectra with a Lorentzian linewidth of ΔL = 230 meV (Gelius et al., 1974). This yielded a Gaussian width of ΓG = 124 meV ± 3 meV and a of 7.0 × 103, which again conformed well to the theoretical value of 7.2 × 103.
4.2. at the sample
The . For the Ar L2,3 spectra at 244 eV, the experimental obtained for the 400 lines mm−1 grating at the APPES branch is 1.3 × 1011 photons s−1 @ E/ΔE = 3700, while the theoretical is calculated to be 4 × 1011 photons s−1 (0.1% bandwidth)−1 (cf. Fig. 4). Fig. 7 shows the at the APPES and PIPOS expected sample positions as a function of energy. The maximum was 5 × 1011 photons s−1 (0.1% bandwidth)−1 at 125 eV at the APPES expected sample position using the 400 lines mm−1 grating. The experimental coincides with the theoretical results in Fig. 4. The dip in the at approximately 285 eV originated from carbon contamination of the beamline optics (mirrors and gratings). In situ synchrotron-radiation-activated oxygen cleaning is now operational for the removal of such contamination (Warburton & Pianetta, 1992).
at the expected APPES and PIPOS sample positions was measured within the whole energy range using an AXUV100G photodiode. The vertical divergence varies at different energies, with the energy resolution estimated from the gas core-shell excitation spectra measured in Section 4.14.3. Energy range
According to the
principle,where BE is the binding energy of the atomic orbitals from which the electron originates, hν is the photon energy, KE is the of the emitted electron measured by the instrument and Φ is the work function of the spectrometer. A VG Scienta Hipp3 analyzer was used to carry out (XPS) experiments and obtain the The work function for the spectrometer was 4.3 eV and the pass energy was set to be 50 eV. By analyzing the of an electron from a typical we can obtain the photon energy. The (VB)-XPS for an Ag film sample at the APPES endstation was measured by the electron energy analyzer, as shown in Fig. 8(a). A KE(EF) value (dashed line) of 35.6 eV was obtained from VB-XPS for the Ag film. Because the Ag BE is 0 eV, the photon energy was deduced to be 39.9 eV. In addition, Ag 3d XPS was also measured, as shown in Fig. 8(b). The obtained from the XPS profile and binding energy for Ag 3d5/2 obtained from a database (Moulder et al., 1995) was 1639.8 eV and 368.3 eV, respectively. According to equation (5), we calculate the photon energy to be hν = 2014.04 eV. By measuring the Ag photoelectron spectrum, the beamline energy range was found to range from 39.9 eV to 2014.04 eV.
4.4. Spot size at the sample
A high-resolution imaging system was installed at the endstation to measure the spot size. This system consisted of a YAG:Ce scintillator, NAVITAR zoom lens system and CCD camera. The scintillator is 50 µm thick, which generates visible light at X-ray radiation illumination. The zoom lens offers 0.58× to 7× magnification in the detector system. The CCD camera, with a pixel size of 13.5 µm, is used to detect visible light. The spot size at the sample was measured using a vertical exit slit of 35 µm. Fig. 9 shows that the spot sizes are 177 µm × 23 µm and 150 µm × 46 µm at the APPES and PIPOS samples, respectively.
5. APPES and PIPOS endstations
The APPES endstation contains a fast-access load lock, a sample preparation chamber, an analysis chamber and an energy analyzer. The load lock chamber can be evacuated to 10−7 mbar within an hour, which facilitates fast sample transfers. The sample preparation chamber has an ion gun for sputtering and a (LEED) setup for surface structure characterization. The X-rays are let into the analysis chamber through a thin Si3N4 membrane that provides pressure isolation. Through a pre-lens, the analysis chamber is connected to the VG Scienta Hipp3 energy analyzer, which is designed for measurements of under ambient-pressure conditions ranging from ultrahigh-vacuum (UHV) to 20 mbar. The ion gauge and capacitance diaphragm gauge are mounted on the analysis chamber to monitor the gas pressure ranging from 10−10 mbar to 10−4 mbar and from 10−4 mbar to 100 mbar, respectively. Besides, the second differential pumping stage is equipped with a quadrupole that could monitor the reaction gases. We briefly present some results carried out at the APPES endstation. The sample is 100 nm Au film deposited on Si wafer. The pre-lens of the energy analyzer is 0.3 mm distance from the Au film surface. The photon energy and the pass energy for the energy analyzer are chosen to be 500 eV and 50 eV, respectively. Fig. 10(a) shows the Au 4f XPS spectrum under the UHV pressure and with 1.27 mbar N2. Due to the strong scattering of photoelectrons, the spectrum intensity decreases with the N2 pressure increasing.
The PIPOS endstation contains a fast-access load lock, a transfer chamber, a sample preparation chamber and an analysis chamber. The sample surface can be cleaned by argon-ion etching and vacuum annealing on the sample preparation chamber. Furthermore, the film can be grown using the MBE method with different oxygen pressures and its surface orientation can be in situ detected by the LEED setup. The analysis chamber includes low-temperature and a high-temperature sample stages, where the absorption spectra can be measured with the temperature ranging from 70 K to 600 K. Besides, in situ charge–discharge experiments for energy materials can be carried out under the UHV and ambient pressure conditions by using the in situ reaction device. The X-ray absorption spectra can be collected in both total electron yield and total modes simultaneously with a base pressure of approximately 5 × 10−10 Torr. The photon energy can be calibrated with the spectra of the reference samples (CoO, MnO, NiO and SrTiO3), which are measured simultaneously. The X-ray absorption spectra of graphite film are measured using the the total electron yield technique at the PIPOS endstation. Fig. 10(b) shows C K-edge XANES with different X-ray incident angle. Two small peaks are observed near 286.7 eV and 290.3 eV with the angle of 75°. With the angle decreasing, these two peaks weaken gradually, and disappear at the normal incident angle of 0°.
6. Conclusion
The BL02B bending-magnet beamline at the SSRF for high energy resolution and high 11 photons s−1 @ E/ΔE = 3700. The energy-resolving powers are measured to be over 10000 at 244 eV and 401 eV, which coincide with the design values. The spot sizes at the APPES and the PIPOS samples are 177 µm × 23 µm and 150 µm × 46 µm, respectively. Due to the high performance within a wide energy range from 40 eV to 2000 eV, the BL02B beamline can be utilized to carry out novel APPES and PIPOS experiments. The APPES and PIPOS endstations have been built up and can offer a platform for study of electronic structures under UHV and ambient pressure conditions.
has been constructed. To realize high performance, the beamline was optimized by using an elliptical cylinder mirror as the first mirror, three gratings covering different energy ranges, and high demagnification ratio (>3:1) ellipsoidal mirrors. The measured at the APPES sample at 244 eV is 1.3 × 10Funding information
The following funding is acknowledged: National Major Scientific Instruments and Equipment Development of NSFC (grant No. 11227902); National Natural Science Foundation of China (grant No. 11575284; grant No. 11475251; grant No. U1732268; grant No. 11705272).
References
Blomberg, S., Hoffmann, M. J., Gustafson, J., Martin, N. M., Fernandes, V. R., Borg, A., Liu, Z., Chang, R., Matera, S., Reuter, K. & Lundgren, E. (2013). Phys. Rev. Lett. 110, 117601. CrossRef Google Scholar
Feser, M., Howells, M. R., Kirz, J., Rudati, J. & Yun, W. (2012). J. Synchrotron Rad. 19, 751–758. CrossRef CAS IUCr Journals Google Scholar
Gelius, U., Svensson, S., Siegbahn, H., Basilier, E., Faxälv, A. & Siegbahn, K. (1974). Chem. Phys. Lett. 28, 1–7. CrossRef CAS Google Scholar
Guo, Z., Meng, X., Wang, Y., Liu, H., Zhang, X., Li, Z., Xue, L. & Tai, R. (2017). J. Synchrotron Rad. 24, 877–885. CrossRef CAS IUCr Journals Google Scholar
Harada, T. & Kita, T. (1980). Appl. Opt. 19, 3987–3993. CrossRef CAS PubMed Web of Science Google Scholar
Himpsel, F. (2011). Phys. Status Solidi B, 248, 292–298. CrossRef CAS Google Scholar
Itou, M., Harada, T. & Kita, T. (1989). Appl. Opt. 28, 146–153. CrossRef CAS PubMed Google Scholar
King, G. C., Tronc, M., Read, F. H. & Bradford, R. C. (1977). J. Phys. B: At. Mol. Phys. 10, 2479–2495. Google Scholar
Kiyokura, T., Maeda, F. & Watanabe, Y. (1998). J. Synchrotron Rad. 5, 572–574. Web of Science CrossRef CAS IUCr Journals Google Scholar
Liu, X. S., Yang, W. L. & Liu, Z. (2014). Adv. Mater. 26, 7710–7729. CrossRef CAS Google Scholar
Moulder, J., Chastain, J. & King, R. (1995). Handbook of X-ray Photoelectron Spectroscopy: A Reference Book of Standard Spectra for Identification and Interpretation of XPS Data. Eden Prairie: Perkin-Elmer Corporation. Google Scholar
Nevière, M., Vincent, P. & Petit, R. (1974). Nouv. Rev. Opt. 5, 65. Google Scholar
Ogletree, D. F., Bluhm, H., Hebenstreit, E. D. & Salmeron, M. (2009). Nucl. Instrum. Methods Phys. Res. A, 601, 151–160. Google Scholar
Prince, K., Vondráček, M., Karvonen, J., Coreno, M., Camilloni, R., Avaldi, L. & de Simone, M. (1999). J. Electron Spectrosc. Relat. Phenom. 101–103, 141–147. CrossRef CAS Google Scholar
Reininger, R. (2011). Nucl. Instrum. Methods Phys. Res. A, 649, 139–143. Web of Science CrossRef CAS Google Scholar
Salmeron, M. & Schlögl, R. (2008). Surf. Sci. Rep. 63, 169–199. Web of Science CrossRef CAS Google Scholar
Sodhi, R. N. & Brion, C. (1984). J. Electron Spectrosc. Relat. Phenom. 34, 363–372. CrossRef CAS Google Scholar
Strocov, V. N., Schmitt, T., Flechsig, U., Schmidt, T., Imhof, A., Chen, Q., Raabe, J., Betemps, R., Zimoch, D., Krempasky, J., Wang, X., Grioni, M., Piazzalunga, A. & Patthey, L. (2010). J. Synchrotron Rad. 17, 631–643. Web of Science CrossRef CAS IUCr Journals Google Scholar
Tanaka, T. & Kitamura, H. (2001). J. Synchrotron Rad. 8, 1221–1228. Web of Science CrossRef CAS IUCr Journals Google Scholar
Warburton, W. K. & Pianetta, P. (1992). Nucl. Instrum. Methods Phys. Res. A, 319, 240–243. CrossRef Web of Science Google Scholar
Xue, C., Wang, Y., Guo, Z., Wu, Y., Zhen, X., Chen, M., Chen, J., Xue, S., Peng, Z., Lu, Q. & Tai, R. (2010). Rev. Sci. Instrum. 81, 103502. Web of Science CrossRef PubMed Google Scholar
Xue, L., Reininger, R., Wu, Y.-Q., Zou, Y., Xu, Z.-M., Shi, Y.-B., Dong, J., Ding, H., Sun, J.-L., Guo, F.-Z., Wang, Y. & Tai, R.-Z. (2014). J. Synchrotron Rad. 21, 273–279. Web of Science CrossRef CAS IUCr Journals Google Scholar
Zhang, C. J., Grass, M. E., McDaniel, A. H., DeCaluwe, S. C., El Gabaly, F., Liu, Z., McCarty, K. F., Farrow, R. L., Linne, M. A., Hussain, Z., Jackson, G. S., Bluhm, H. & Eichhorn, B. W. (2010). Nat. Mater. 9, 944–949. CrossRef CAS Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.