A statistical approach to correct X-ray response non-uniformity in microstrip detectors for high-accuracy and high-resolution total-scattering measurements

CROSSMARK_Color_square_no_text.svg

aRIKEN SPring-8 Center, 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5148, Japan, bJST, PRESTO, 4-1-8 Honcho, Kawaguchi, Saitama 332-0012, Japan, cGraduate School of Material Science, University of Hyogo, 3-2-1 Kouto, Kamigori-cho, Ako-gun, Hyogo 678-1297, Japan, dInternational Institute for Carbon-Neutral Energy Research (WPI-I2CNER), Kyushu University, 744 Moto-oka, Nishi-ku, Fukuoka 819-0395, Japan, eDepartment of Chemistry, Faculty of Science, Kyushu University, 744 Moto-oka, Nishi-ku, Fukuoka 819-0395, Japan, and fResearch and Utilization Division, Japan Synchrotron Radiation Research Institute (JASRI, SPring-8), 1-1-1 Kouto, Sayo-cho, Sayo-gun, Hyogo 679-5198, Japan
*Correspondence e-mail: katok@spring8.or.jp

Edited by Y. Amemiya, University of Tokyo, Japan (Received 9 July 2018; accepted 9 February 2019; online 5 April 2019)

An unbiased approach to correct X-ray response non-uniformity in microstrip detectors has been developed based on the statistical estimation that the scattering intensity at a fixed angle from an object is expected to be constant within the Poisson noise. Raw scattering data of SiO2 glass measured by a microstrip detector module was found to show an accuracy of 12σPN at an intensity of 106 photons, where σPN is the standard deviation according to the Poisson noise. The conventional flat-field calibration has failed in correcting the data, whereas the alternative approach used in this article successfully improved the accuracy from 12σPN to 2σPN. This approach was applied to total-scattering data measured by a gapless 15-modular detector system. The quality of the data is evaluated in terms of the Bragg reflections of Si powder, the diffuse scattering of SiO2 glass, and the atomic pair distribution function of TiO2 nanoparticles and Ni powder.

1. Introduction

The reliability of results obtained by X-ray structural analysis depends on accuracy and precision in diffraction and scattering intensity from a sample. In particular, this applies to atomic pair distribution function (PDF) analysis of total-scattering data since real-space structure information can be obtained via no structural model. The total-scattering PDF method has been employed to characterize the local structure of crystalline materials as well as amorphous materials (Egami & Billinge, 2012[Egami, T. & Billinge, S. J. L. (2012). Underneath the Bragg Peaks: Structural Analysis of Complex Materials, Pergamon Materials Series, Vol. 16, 2nd ed. Amsterdam: Elsevier.]). In contrast, it is widely recognized that Rietveld analysis of Bragg diffraction data provides fundamental understanding on the average structure of crystalline materials. In brief, the two methods have been considered to be complementary to each other in crystal structural analysis.

The term `total scattering' is derived from treating Bragg and diffuse scattering on an equal basis. In other words, total-scattering experiments are considered to be on the extension of conventional powder diffraction experiments. Nevertheless, data measurements for PDF analysis have been conducted independently of measurements for Rietveld analysis. In general, PDF analysis requires total-scattering data in a wide range of scattering vector, Q, which has a magnitude |Q| = Q = 4π sin(2θ/2)/λ, where λ is the wavelength of scattered X-rays and 2θ is the scattering angle. In contrast, Rietveld analysis needs high-Q-resolution data. The problem is how to obtain total-scattering data with both high Q resolution and wide Q space, which are considered to be in a trade-off relationship. Such ideal data would facilitate multi-scale PDF analysis. Among other local structure probes such as extended X-ray absorption fine structure, the total-scattering PDF method has a potential for investigating heterogeneous phenomena in functional nanomaterials (Skrobas et al., 2017[Skrobas, K., Stelmakh, S., Gierlotka, S. & Palosz, B. F. (2017). J. Appl. Cryst. 50, 1821-1829.]).

The rapid-acquisition PDF (RA-PDF) method (Chupas et al., 2003[Chupas, P. J., Qiu, X., Hanson, J. C., Lee, P. L., Grey, C. P. & Billinge, S. J. L. (2003). J. Appl. Cryst. 36, 1342-1347.], 2007[Chupas, P. J., Chapman, K. W. & Lee, P. L. (2007). J. Appl. Cryst. 40, 463-470.]), which is typical of total-scattering beamlines at synchrotron facilities, can collect total-scattering data with the maximum Q value (Qmax) above 30 Å−1 by a combination of high-energy X-rays and a large area detector. The Q resolution (∼10−2 Å−1) is moderate for diffuse scattering but is insufficient for Bragg reflections because of the use of energies of around 100 keV, a short sample-to-detector distance and a low-resolution detector. As a consequence, the PDF obtained by RA-PDF undergoes significant damping at longer inter­atomic distances (Saleta et al., 2017[Saleta, M. E., Eleotério, M., Mesquita, A., Mastelaro, V. R. & Granado, E. (2017). J. Synchrotron Rad. 24, 1098-1104.]). On the other hand, point detectors such as scintillation counters allow high-resolution measurements by setting analyzer crystals (Saleta et al., 2017[Saleta, M. E., Eleotério, M., Mesquita, A., Mastelaro, V. R. & Granado, E. (2017). J. Synchrotron Rad. 24, 1098-1104.]). It is not, however, technically feasible to cover a wide range of Q at the same time even by using multiple point detectors. For that reason, such a scanning method cannot be applied to in situ experiments under non-ambient conditions, where a scattering pattern varies every moment.

A number of powder diffraction beamlines (Haverkamp & Wallwork, 2009[Haverkamp, R. G. & Wallwork, K. S. (2009). J. Synchrotron Rad. 16, 849-856.]; Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]; Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]; Saleta et al., 2017[Saleta, M. E., Eleotério, M., Mesquita, A., Mastelaro, V. R. & Granado, E. (2017). J. Synchrotron Rad. 24, 1098-1104.]) at synchrotron facilities have employed an array of microstrip detectors such as MYTHEN (DECTRIS Ltd, Baden-Daettwil, Switzerland) (Schmitt et al., 2003[Schmitt, B., Brönnimann, C., Eikenberry, E. F., Gozzo, F., Hörmann, C., Horisberger, R. & Patterson, B. (2003). Nucl. Instrum. Methods Phys. Res. A, 501, 267-272.]). In addition to powder diffraction data, such a MYTHEN detector array can measure high-Q-resolution (∼10−4 Å−1) total-scattering data in a wide range of Q (20–30 Å−1) even by using moderate energies of around 20–30 keV. Indeed, the detector array has been tested for the PDF method (Haverkamp & Wallwork, 2009[Haverkamp, R. G. & Wallwork, K. S. (2009). J. Synchrotron Rad. 16, 849-856.]; Saleta et al., 2017[Saleta, M. E., Eleotério, M., Mesquita, A., Mastelaro, V. R. & Granado, E. (2017). J. Synchrotron Rad. 24, 1098-1104.]). However, multi-scale PDF analysis at interatomic distances above 100 Å has never been conducted, which is likely to be because of X-ray response non-uniformity (XRNU) in MYTHEN. In X-ray detectors, XRNU has been widely recognized as one of the most critical problems for structural analysis. To correct XRNU, the conventional flat-field calibration has been applied to MYTHEN detector arrays (Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]; Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]). In this calibration, each module is illuminated with a roughly flat X-ray pattern with a specific energy, which is obtained by scanning a stable beam. The XRNU on each channel is calibrated based on the difference between raw data and the completely flat pattern. The calibration can correct the relatively large XRNU as observed at both ends of channels (Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]) but has failed in correcting the small XRNU found at intensities above 104 photons because of an approximation to the completely flat pattern, which is impossible to obtain.

Here we propose an approach to correct XRNU in microstrip detectors without using any approximation, which could be a replacement for the conventional flat-field calibration. The alternative approach is based on the statistical estimation that the scattering intensity at a fixed angle from an object is expected to be constant within the Poisson noise. The improvements in total-scattering data measured by a MYTHEN-modular detector system are discussed from the point of view of two figures of merit for X-ray detectors and PDFs.

2. Instruments and methods

2.1. X-ray total-scattering detector

2.1.1. Detector module

Among other X-ray detectors, a photon-counting microstrip detector, MYTHEN, was selected as a promising candidate for high-accuracy and high-resolution total-scattering measurements. The detector module can meet the demands for spatial resolution and dynamic range in such measurements. An interval between two adjacent strips is 50 µm, which corresponds to the spatial resolution of the module. The counter has a dynamic range of the order of 107. In addition, such a photon-counting system can separate signal from noise by setting the optimal threshold energy for each incident energy to exclude readout and dark noise. The module is composed of an electronic board and a Si sensor, the dimensions of which are 64 mm by 8 mm. The sensor with a thickness of 1 mm was employed to improve quantum efficiency at high energies.

2.1.2. Detector system

To measure total-scattering data in a wide range of Q in parallel, 15 modules without housing were assembled to build a detector system, which was installed on the Debye–Scherrer camera of the RIKEN Materials Science beamline BL44B2 (Kato et al., 2010[Kato, K., Hirose, R., Takemoto, M., Ha, S., Kim, J., Higuchi, M., Matsuda, R., Kitagawa, S. & Takata, M. (2010). AIP Conf. Proc., 1234, 875-878.]; Kato & Tanaka, 2016[Kato, K. & Tanaka, H. (2016). Adv. Phys. X, 1, 55-80.]) of SPring-8 (Fig. 1[link]). These modules were arranged in a pseudo curve on the 2θ stage of the camera. The distance between the sample position and the sensor surface was fixed at 286.48 (1) mm so that the angular resolution is ∼0.01° in 2θ. The detector system was characterized by its gapless arrangement in the vertical direction as shown in Fig. 1[link](a). Adjacent modules overlapped each other by 140 channels, which correspond to 1.4° in 2θ. The overlap can detect the identical Bragg reflection by two adjacent modules. Comparing these two intensities leads to the evaluation of granularity and preferred orientation of powder samples. Thanks to the arrangement, total-scattering data in the vertical direction can be collected up to 153° in 2θ simultaneously, which corresponds to 30 Å−1 in Qmax in the case of 30 keV in incident energy. In contrast, there is a 0.5 mm gap in the horizontal direction between adjacent modules. Including the insensitive edge (1 mm) of the sensor, the width of ±1.25 mm from the horizontal center is lost in data. The full width of the sensor is 8 mm. The aperture for scattered X-rays can be adjusted using a brazen mask of 2 mm thickness in front of the sensor [Fig. 1[link](b)]. The mask edge was aligned along the sensor edge within an accuracy of 0.3% in aperture. The XRNU observed in the detector system is affected by the misalignment of the mask to the sensor and the roughness of the mask edge. Taking into account the umbrella effect at lower angles and relatively low statistics at higher angles, the apertures of the 15 modules were fixed for the present experiments as follows: 0.50 mm for module 1 (M1), 0.75 mm for M2, 1.00 mm for M3, 1.25 mm for M4, 1.50 mm for M5, 1.75 mm for M6, 2.00 mm for M7, 2.25 mm for M8, 2.50 mm for M9, 2.75 mm for M10, 3.00 mm for M11, 3.25 mm for M12, 3.50 mm for M13, 3.75 mm for M14 and 4.00 mm for M15 [Fig. 1[link](c)]. The detector system is controlled using a measurement program written by the graphical programming software LabVIEW (National Instruments Corporation, Austin, TX, USA) via the detector control system DCS24 (DECTRIS Ltd).

[Figure 1]
Figure 1
The gapless MYTHEN-modular detector system. (a) A schematic view of the gapless arrangement with five modules. (b) A corresponding photograph of (a), where masks can be seen in front of the sensors. (c) A photograph of the 15-modular system, which has been installed on the Debye–Scherrer camera at BL44B2 of SPring-8. Note that the photographs in (b) and (c) were taken without housing.

2.2. Approach to correct XRNU

2.2.1. Concept

Our approach to correct XRNU is based on the statistical estimation that the scattering intensity at a fixed angle from an object is expected to be constant within the Poisson noise. The conventional flat-field calibration employs an approximation to the perfectly uniform illumination of X-rays, which is impossible to obtain. In contrast, the alternative approach is characterized by no need for such an approximation. For example, the scattering intensity distribution at a fixed angle range Δ2θ measured by module position 1 [Figs. 2[link](a) and 2[link](b)] should be in agreement with that at Δ2θ by position 2 [Figs. 2[link](c) and 2[link](d)] within the Poisson noise. In practice, however, these distributions are in disagreement because of various systematic errors including XRNU. The approach is expected to correct XRNU based on the difference between the two distributions at Δ2θ measured by different channels. Therefore, there should be no systematic difference in XRNU between compared channels. In other words, the key to success in correction is randomness in XRNU between channels. From this fact, the module should not be partitioned every 128 channels, which corresponds to an architectural unit of the readout chip.

[Figure 2]
Figure 2
Schematic drawing of the statistical approach at two module positions using one module. (a) The setup and (b) the scattering data at module position 1. Likewise, (c) the setup and (d) the scattering data at module position 2. Without any systematic error, the intensity distribution at Δ2θ in (b) is expected to be consistent with that at Δ2θ in (d) within the Poisson noise.
2.2.2. Divided-accumulation technique

The statistical approach requires that the scattering intensity at a fixed angle from an object should not exceed 0.1% in relative fluctuation, which corresponds to the Poisson noise at 106 photons. The fluctuation of the ring current at SPring-8 is controlled below 0.05% by the top-up injection in any electron bunch mode. We, nevertheless, found it difficult to maintain the 0.1% fluctuation because of the thermal instability of optical and measuring systems. To average out fluctuations in scattering intensity, we have employed a divided-accumulation technique, where two measurements at module positions 1 and 2 [Figs. 2[link](a) and 2[link](c)] are repeated until the total intensity at each position reaches the order of 106 photons. The important point is that measurement time at each position is fixed so that the cycle of measurements is sufficiently shorter than that of fluctuations in scattering intensity. After a series of measurements, multiple data at the same position are summed up to yield a two-data set. The technique can be applied to detectors without readout noise such as photon-counting-type detectors since a large amount of data is accumulated.

2.2.3. Scattering object and conditions

A SiO2 glass rod with a diameter of 3.5 mm was used as the scattering object for the approach. Among others, the object can give rise to high and reproducible scattering intensity in a wide range of Q. The wavelength of incident X-rays was set at 0.45 Å, taking into account the quantum efficiency of the 1 mm-Si sensor. The threshold energy in the module was set at half of the incident energy to avoid double counts. The wavelength was calibrated by the use of the K absorption edge of In with a precision of 10−4 Å. The horizontal and vertical sizes of the incident slit were 3 mm and 1 mm, respectively, which is smaller than the object size to reduce the influence of fluctuations in beam position.

2.2.4. Single-step process

In Fig. 2[link], only two module positions are illustrated for simplicity. Based on the concept, in practice, measurements were conducted at 160 positions, which is referred to as the single-step process. One module has 1280 channels, which are indexed from 1 to 1280 in order. The 1280 channels were partitioned into every eight channels, which corresponds to about 0.08° in 2θ in the sample-to-detector distance. The partition can be expressed in terms of a 8 × 160 matrix,

[{\rm{{\bf A} = }}\,\,{\left({{{{a}}_{{{ij}}}}} \right)_{\matrix{\scriptstyle {{\rm{1}} \le {{i}} \le {{l,\, l = 8}}}\hfill \cr \scriptstyle {{\rm{1}} \le {{j}} \le {{m,\, m = 160}}}\hfill \cr } }}{\rm{ = }}\,\left({\matrix{ {\rm{1}}\! & \!{\rm{9}}\! & \!\cdots\! &\! {{\rm{1265}}}\! & \!{{\rm{1273}}} \cr {\rm{2}}\! & \!{{\rm{10}}}\! &\! \cdots\! &\! {{\rm{1266}}}\! &\! {{\rm{1274}}} \cr \!\vdots\! &\! \vdots\! &\! \cdots\! &\! \vdots\! &\! \vdots\! \cr {\rm{7}}\! &\! {{\rm{15}}}\! &\! \cdots\! &\! {{\rm{1271}}} \!&\! {{\rm{1279}}} \cr {\rm{8}} \!&\! {{\rm{16}}} \!&\! \cdots \!&\! {{\rm{1272}}} \!&\! {{\rm{1280}}} \cr } } \right). \eqno(1)]

The process was combined with the divided-accumulation technique described in Section 2.2.2[link], in which case the following sequence of measurements was repeated until the total intensity at a fixed module angle 2θM reached the order of 106 photons:

(1) One-minute measurement at 2θM,

(2) One-minute measurement at 2θM + 0.08°,

(3) One-minute measurement at 2θM + 0.08° × 2,

  :

(160) One-minute measurement at 2θM + 0.08° × 159.

Multiple one-minute data measured at the same module angle were summed up to yield a 160-data set. Correction factors css for individual channels based on the single-step were calculated by taking an average of the intensity measured by 160 different channels at a fixed scattering angle 2θ according to the following expression,

[{c_{\rm ss}}\left({{{{a}}_{{{ij}}}}} \right){{ \,=\, }}{{\mathop \sum \nolimits_{{{j = 1}}}^{{m}} {{{y}}_{{\rm{2}}\theta }}\left({{{{a}}_{{{ij}}}}} \right)} \over m}{\rm{}}{{\rm{1}} \over {{{{y}}_{{\rm{2\theta }}}}\left({{{{a}}_{{{ij}}}}} \right)}}, \eqno(2)]

where [{{{y}}_{{\rm{2}}\theta }}] is the scattering intensity at 2θ for each channel. Note that 32 channels at both ends (channels 1–16 and 1265–1280) were eliminated from the calculation since their intensities were unreliable because of the unpredictable response.

2.2.5. Multi-step process

The number of measurements in the single-step process is 160, which takes several-ten hours to obtain intensity of the order of 106 photons. To reduce the total time maintaining the correction effects, a multi-step process based on the single-step was developed. The process consists of five steps. The number of measurements in the multi-step is 24, which is 15% of that in the single-step.

For the first-step calculation, 1280 channels were partitioned into every 80 channels, which corresponds to 0.8° in 2θ, with a 80 × 16 matrix,

[\eqalign {{{{\rm {\bf A}1}}}={\left({{{a}}{{\rm{1}}_{{{ij}}}}} \right)_{\matrix{\scriptstyle {{\rm{1}}\le{{i}}\le{{{l}}_{\rm{1}}}{\rm{,}}\,{{{l}}_{\rm{1}}}{\rm{ = 80}}} \hfill \cr \scriptstyle{{\rm{1}}\le{{j}}\le{{{m}}_{\rm{1}}}{\rm{,}}\,{{{m}}_{\rm{1}}}{\rm{ = 16}}} \hfill \cr } }}=\left({\matrix{ {\rm{1}} & {{\rm{81}}} & \cdots & {{\rm{1121}}} & {{\rm{1201}}} \cr {\rm{2}} & {{\rm{82}}} & \cdots & {{\rm{1122}}} & {{\rm{1202}}} \cr \vdots & \vdots & \cdots & \vdots & \vdots \cr {{\rm{79}}} & {{\rm{159}}} & \cdots & {{\rm{1119}}} & {{\rm{1279}}} \cr {{\rm{80}}} & {{\rm{160}}} & \cdots & {{\rm{1200}}} & {{\rm{1280}}} \cr } } \right).}\eqno(3)]

In the same way as the single-step, the following sequence was repeated:

(1) Three-minute measurement at 2θM1,

(2) Three-minute measurement at 2θM1 + 0.8°,

(3) Three-minute measurement at 2θM1 + 0.8° × 2,

  :

(16) Three-minute measurement at 2θM1 + 0.8° × 15.

Using the 16-data set, correction factors c1 were calculated by the following expression,

[{{c1}}\left({{{a}}{{\rm{1}}_{{{ij}}}}} \right){{ \,=\, }}{{\mathop \sum \nolimits_{{{j = 1}}}^{{{{m}}_{\rm{1}}}} {{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{1}}}}}\left({{{a}}{{\rm{1}}_{{{ij}}}}} \right)} \over {{{{m}}_{\rm{1}}}}}{\rm{}}{{\rm{1}} \over {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{1}}}}}\left({{{a}}{{\rm{1}}_{{{ij}}}}} \right)}}, \eqno(4)]

where [{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{1}}}}}] is the scattering intensity at a fixed scattering angle 2θ1 for each channel. Note that 32 channels at both ends (channels 1–16 and 1265–1280) were eliminated from the calculation as in the single-step process.

For the second-step calculation, 1280 channels were repartitioned into every 40 channels, which corresponds to 0.4° in 2θ, with a three-dimensional 40 × 2 × 16 matrix,

[\eqalignno{{\rm{{\bf A}2}}=\,\,\,&{\left({{{a}}{{\rm{2}}_{{{ijk}}}}} \right)_{\matrix{\scriptstyle {{\rm{1}} \le {{i}} \le {{{l}}_{\rm{2}}}{\rm{,}}\,{{{l}}_{\rm{2}}}{\rm{\, = 40}}}\hfill \cr \scriptstyle{{\rm{1}} \le {{j}} \le {{{m}}_{\rm{2}}}{\rm{,}}\,{{{m}}_{\rm{2}}}{\rm{\, = 2}}}\hfill \cr \scriptstyle{{\rm{1}} \le {{k}} \le {{{n}}_{\rm{2}}}{\rm{,}}\,{{{n}}_{\rm{2}}}{\rm{\, = 16}}}\hfill \cr } }}=\left({\matrix{ {\rm{1}} & {{\rm{41}}} \cr {\rm{2}} & {{\rm{42}}} \cr \vdots & \vdots \cr {{\rm{39}}} & {{\rm{79}}} \cr {{\rm{40}}} & {{\rm{80}}} \cr } } \right){\rm{,}}\cr &\left({\matrix{ {{\rm{81}}} & {{\rm{121}}} \cr {{\rm{82}}} & {{\rm{122}}} \cr \vdots & \vdots \cr {{\rm{119}}} & {{\rm{159}}} \cr {{\rm{120}}} & {{\rm{160}}} \cr } } \right) \cdots \left({\matrix{ {{\rm{1201}}} & {{\rm{1241}}} \cr {{\rm{1202}}} & {{\rm{1242}}} \cr \vdots & \vdots \cr {{\rm{1239}}} & {{\rm{1279}}} \cr {{\rm{1240}}} & {{\rm{1280}}} \cr } } \right). &(5)}]

Similarly, the following sequence was repeated:

(1) Three-minute measurement at 2θM2,

(2) Three-minute measurement at 2θM2 + 0.4°.

Using the two-data set, correction factors c2 were calculated by the following expression,

[{{c2}}\left({{{a}}{{\rm{2}}_{{{ijk}}}}} \right){{ = }}{{\mathop \sum \nolimits_{{{j = 1}}}^{{{{m}}_{\rm{2}}}} \left\{ {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{2}}}}}\left({{{a}}{{\rm{2}}_{{{ijk}}}}} \right)\,{{c1}}\left({{{a}}{{\rm{2}}_{{{ijk}}}}} \right)} \right\}} \over {{{{m}}_{\rm{2}}}}}{\rm{}}{{\rm{1}} \over {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{2}}}}}\left({{{a}}{{\rm{2}}_{{{ijk}}}}} \right)\,{{c1}}\left({{{a}}{{\rm{2}}_{{{ijk}}}}} \right)}}. \eqno(6)]

For the third-step calculation, 1280 channels were repartitioned into every 20 channels, which corresponds to 0.2° in 2θ, with a three-dimensional 20 × 2 × 32 matrix,

[\eqalignno{{\rm{{\bf A}3}}=\,\,&{\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)_{\matrix{\scriptstyle {{\rm{1}} \le {{i}} \le {{{l}}_{\rm{3}}}{\rm{,}}\,{{{l}}_{\rm{3}}}{\rm{\, = 20}}}\hfill \cr \scriptstyle{{\rm{1}} \le {{j}} \le {{{m}}_{\rm{3}}}{\rm{,}}\,{{{m}}_{\rm{3}}}{\rm{\, = 2}}}\hfill \cr \scriptstyle {{\rm{1}} \le {{k}} \le {{{n}}_{\rm{3}}}{\rm{,}}\,{{{n}}_{\rm{3}}}{\rm{\, = 32}}}\hfill \cr } }}=\left({\matrix{ {\rm{1}} & {{\rm{21}}} \cr {\rm{2}} & {{\rm{22}}} \cr \vdots & \vdots \cr {{\rm{19}}} & {{\rm{39}}} \cr {{\rm{20}}} & {{\rm{40}}} \cr } } \right){\rm{,}}\cr &\left({\matrix{ {{\rm{41}}} & {{\rm{61}}} \cr {{\rm{42}}} & {{\rm{62}}} \cr \vdots & \vdots \cr {{\rm{59}}} & {{\rm{79}}} \cr {{\rm{60}}} & {{\rm{80}}} \cr } } \right) \cdots \left({\matrix{ {{\rm{1241}}} & {{\rm{1261}}} \cr {{\rm{1242}}} & {{\rm{1262}}} \cr \vdots & \vdots \cr {{\rm{1259}}} & {{\rm{1279}}} \cr {{\rm{1260}}} & {{\rm{1280}}} \cr } } \right). &(7)}]

Similarly, the following sequence was repeated:

(1) Three-minute measurement at 2θM3,

(2) Three-minute measurement at 2θM3 + 0.2°.

Using the two-data set, correction factors c3 were calculated by the following expression,

[\eqalignno{{{c3}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)=&{{\mathop \sum \nolimits_{{{j = 1}}}^{{{{m}}_{\rm{3}}}} \left\{ {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{3}}}}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)\,{{c1}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)\,{{c2}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)} \right\}} \over {{{{m}}_{\rm{3}}}}}\cr &\times{\rm{}}{{\rm{1}} \over {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{3}}}}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)\,{{c1}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)\,{{c2}}\left({{{a}}{{\rm{3}}_{{{ijk}}}}} \right)}}. &(8)}]

For the fourth-step calculation, 1280 channels were repartitioned into every 10 channels, which corresponds to 0.1° in 2θ, with a three-dimensional 10 × 2 × 64 matrix,

[\eqalignno{{\rm{{\bf A}4}}=\,\,&{\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)_{\matrix{\scriptstyle {{\rm{1}} \le {{i}} \le {{{l}}_{\rm{4}}}{\rm{,}}\,{{{l}}_{\rm{4}}}{\rm{\, = 10}}}\hfill \cr \scriptstyle{{\rm{1}} \le {{j}} \le {{{m}}_{\rm{4}}}{\rm{,}}\,{{{m}}_{\rm{4}}}{\rm{\, = 2}}}\hfill \cr \scriptstyle {{\rm{1}} \le {{k}} \le {{{n}}_{\rm{4}}}{\rm{,}}\,{{{n}}_{\rm{4}}}{\rm{\, = 64}}}\hfill \cr } }}=\left({\matrix{ {\rm{1}} & {{\rm{11}}} \cr {\rm{2}} & {{\rm{12}}} \cr \vdots & \vdots \cr {\rm{9}} & {{\rm{19}}} \cr {{\rm{10}}} & {{\rm{20}}} \cr } } \right){\rm{,}}\cr &\left({\matrix{ {{\rm{21}}} & {{\rm{31}}} \cr {{\rm{22}}} & {{\rm{32}}} \cr \vdots & \vdots \cr {{\rm{29}}} & {{\rm{39}}} \cr {{\rm{30}}} & {{\rm{40}}} \cr } } \right) \cdots \left({\matrix{ {{\rm{1261}}} & {{\rm{1271}}} \cr {{\rm{1262}}} & {{\rm{1272}}} \cr \vdots & \vdots \cr {{\rm{1269}}} & {{\rm{1279}}} \cr {{\rm{1270}}} & {{\rm{1280}}} \cr } } \right). &(9)}]

Similarly, the following sequence was repeated:

(1) Three-minute measurement at 2θM4,

(2) Three-minute measurement at 2θM4 + 0.1°.

Using the two-data set, correction factors c4 were calculated by the following expression,

[\eqalignno{{{c4}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)=&{{\mathop \sum \nolimits_{{{j = 1}}}^{{{{m}}_{\rm{4}}}} \left\{ {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{4}}}}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)\,{{c1}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)\,{{c2}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)\,{{c3}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)} \right\}} \over {{{{m}}_{\rm{4}}}}} \cr &\times{{\rm{1}} \over {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{4}}}}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)\,{{c1}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)\,{{c2}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)\,{{c3}}\left({{{a}}{{\rm{4}}_{{{ijk}}}}} \right)}}. &(10)}]

For the fifth-step calculation, 1280 channels were repartitioned into every five channels, which corresponds to 0.05° in 2θ, with a three-dimensional 5 × 2 × 128 matrix,

[\eqalignno{{\rm{{\bf A}5}}=\,\,&{\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right)_{\matrix{\scriptstyle {{\rm{1}} \le {{i}} \le {{{l}}_{\rm{5}}}{\rm{,}}\,{{{l}}_{\rm{5}}}{\rm{\, = 5}}}\hfill \cr \scriptstyle{{\rm{1}} \le {{j}} \le {{{m}}_{\rm{5}}}{\rm{,}}\,{{{m}}_{\rm{5}}}{\rm{\, = 2}}}\hfill \cr \scriptstyle{{\rm{1}} \le {{k}} \le {{{n}}_{\rm{5}}}{\rm{,}}\,{{{n}}_{\rm{5}}}{\rm{\, = 128}}}\hfill \cr } }}=\left({\matrix{ {\rm{1}} & {\rm{6}} \cr {\rm{2}} & {\rm{7}} \cr {\rm{3}} & {\rm{8}} \cr {\rm{4}} & {\rm{9}} \cr {\rm{5}} & {{\rm{10}}} \cr } } \right){\rm{,}}\cr &\left({\matrix{ {{\rm{11}}} & {{\rm{16}}} \cr {{\rm{12}}} & {{\rm{17}}} \cr {{\rm{13}}} & {{\rm{18}}} \cr {{\rm{14}}} & {{\rm{19}}} \cr {{\rm{15}}} & {{\rm{20}}} \cr } } \right) \cdots \left({\matrix{ {{\rm{1271}}} & {{\rm{1276}}} \cr {{\rm{1272}}} & {{\rm{1277}}} \cr {{\rm{1273}}} & {{\rm{1278}}} \cr {{\rm{1274}}} & {{\rm{1279}}} \cr {{\rm{1275}}} & {{\rm{1280}}} \cr } } \right). &(11)}]

Similarly, the following sequence was repeated:

(1) Three-minute measurement at 2θM5,

(2) Three-minute measurement at 2θM5 + 0.05°.

Using the two-data set, correction factors c5 were calculated by the following expression,

[\eqalignno{{{c5}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right)=&{{\mathop \sum \nolimits_{{{j = 1}}}^{{{{m}}_{\rm{5}}}} \left\{ {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{5}}}}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c1}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c2}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c3}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c4}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right)} \right\}} \over {{{{m}}_{\rm{5}}}}}\cr &\times{{\rm{1}} \over {{{{y}}_{{\rm{2}}{{\rm{\theta }}_{\rm{5}}}}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c1}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c2}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c3}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right){{c4}}\left({{{a}}{{\rm{5}}_{{{ijk}}}}} \right)}}. \,(12)}]

The final correction factors cms for individual channels based on the multi-step process is expressed by

[{c_{\rm ms}} = c1 \times c2 \times c3 \times c4 \times c5. \eqno(13)]

2.2.6. Application to the detector system

In Sections 2.2.4[link] and 2.2.5[link], one module is considered for simplicity. In practice, however, the multi-step process combined with the divided-accumulation technique was applied to the 15-modular system with 1280 × 15 = 19200 channels. Our approach to correct XRNU has another advantage in that the number of measurements for correction factors is not influenced by the number of modules since the SiO2 glass scattering is continuously distributed in a wide range of Q. In brief, measurements for correction factors of 15 modules can be carried out in parallel.

2.3. Figure of merit for X-ray detectors

To examine the quality of scattering data measured by X-ray detectors, we have defined total fractional uncertainty (TFU) as a figure of merit,

[{\rm TFU} = {{{\sigma _I}} \over {\bar I}} 100 = {{{\Bigl\{{\bigl[{1/(N - 1)}\bigr]\mathop \sum \nolimits_i {{\left({{I_i} - \bar I} \right)}^2}}\Bigr\}}^{1/2} } \over {\bar I}}\times 100, \eqno(14)]

where Ii is the scattering intensity of channel i, [\bar I] is the average of Ii, [{\sigma _I}] is the standard deviation of Ii and N is the number of channels.

In addition, another figure of merit has been introduced as the ratio of [{\sigma _I}] to the standard deviation based on the Poisson noise [{\sigma _{\rm PN}}] (Poisson noise ratio or PNR),

[{\rm{PNR = }}{{{\sigma _I}} \over {{\sigma _{\rm PN}}}} = {\Biggl\{{{{\bigl[{1/(N - 1)}\bigr]\mathop \sum \nolimits_i {{\left({{I_i} - \bar I} \right)}^2}} } \over {{\bar I} }}\Biggr\}}^{1/2}. \eqno(15)]

2.4. Sample measurements and PDF analysis

To consider the quality of corrected data from various aspects, four different kinds of samples were measured by the detector system using the incident X-ray wavelength of 0.45 Å, which is the same as that in measurements for correction factors. The threshold energy in the detector system was set at half of the incident energy, which is the same as that in measurements for correction factors. The four samples were: a SiO2 glass rod, 3.5 mm in diameter; Si powder (Standard Reference Material 640d; NIST, Gaithersburg, MD, USA), filled into a Lindemann-glass capillary (No. 14; Hilgenberg GmbH, Malsfeld, Germany), 0.3 mm in diameter; anatase-type TiO2 powder, ∼10 nm in diameter, filled into a Lindemann-glass capillary, 0.3 mm in diameter; and Ni powder (NI-314010; The Nilaco Corporation, Tokyo, Japan), filled into a Lindemann-glass capillary, 0.1 mm in diameter. In addition, empty capillaries were measured under the same experimental conditions for PDF analysis. Intensity data indexed from channels 1 to 19200 were converted into 2θ and Q by the calibration using a line-position and -shape standard for powder diffraction, LaB6 powder (Standard Reference Material 660b; NIST). PDF analysis was performed using the xPDFsuite software (Yang et al., 2014[Yang, X., Juhas, P., Farrow, C. L. & Billinge, S. J. L. (2014). arXiv:1402.3163.]). For the analysis, Qmax and the minimum Q value were 27 Å−1 and 0.9 Å−1, respectively.

It is known that counting efficiency is influenced by counting rate in photon-counting-type detectors, yielding XRNU between channels in microstrip detectors. In the present study, the photon arrival rate from samples was set at between 102 and 103 photons s−1, which is much lower than the maximum counting rate (105 photons s−1). This means that counting loss becomes insignificant.

3. Results

3.1. Correction factors and their application

The application of the multi-step process to the 15-modular system yielded correction factors from channels 1 to 19200. Correction factors from channels 14500 to 15000 at each step are shown in Fig. 3[link] among these factors. The first-step factors c1 fluctuate between 0.98 and 1.03 [Fig. 3[link](a)]. The fluctuations of the second-step factors c2 are much smaller than those of the first-step [Fig. 3[link](b)]. Step by step, the fluctuations of the correction factors decrease, indicating that the multi-step functions effectively [Figs. 3[link](c), 3[link](d) and 3[link](e)].

[Figure 3]
Figure 3
Correction factors from channels 14500 to 15000 at each step in the multi-step process: (a) the first-step factors c1 in equation (4[link]), (b) the second-step factors c2 in equation (6[link]), (c) the third-step factors c3 in equation (8[link]), (d) the fourth-step factors c4 in equation (10[link]) and (e) the fifth-step factors c5 in equation (12[link]).

The scattering data of SiO2 glass were measured by the detector system with 19200 channels and then corrected by correction factors at each step in the multi-step process. Fig. 4[link] only shows the data measured from channels 14500 and 15000, which corresponds to ∼123.6°–128.6° in 2θ. The average intensity of the range is ∼5.7 × 106 photons. According to the Poisson statistics, the standard deviation of the range is expected to be the square root of 5.7 × 106 (= 2387). However, the uncorrected data deviate sharply from the expected values [Fig. 4[link](a)]. The data corrected by the first-step factors c1 indicate that the deviations decrease [Fig. 4[link](b)]. Step by step, the deviations show a gradual decrease [Figs. 4[link](c), 4[link](d), 4[link](e) and 4[link](f)].

[Figure 4]
Figure 4
The scattering data of SiO2 glass from channels 14500 to 15000 at each step in the multi-step process: (a) uncorrected; (b) multiplied by c1 in equation (4[link]); (c) multiplied by c1 in equation (4[link]) and c2 in equation (6[link]); (d) multiplied by c1 in equation (4[link]), c2 in equation (6[link]) and c3 in equation (8[link]); (e) multiplied by c1 in equation (4[link]), c2 in equation (6[link]), c3 in equation (8[link]) and c4 in equation (10[link]); and (f) multiplied by cms in equation (13[link]).

Fig. 5[link] shows histograms of the scattering data of SiO2 glass from channels 14500 to 15000 at each step, which have the horizontal axis of intensity and the vertical axis of number of channels. The uncorrected data have a wide distribution with a standard deviation of 55010 photons [Fig. 5[link](a)]. The application of the first-step factors c1 reduces the deviation to 16462 photons [Fig. 5[link](b)]. Step by step, the distributions become sharp, indicating an approach to the Poisson distribution [Figs. 5[link](c), 5[link](d) and 5[link](e)]. The final standard deviation is 9082 photons at 5.7 × 106 photons, which is reduced by a factor of six relative to that of the uncorrected data [Fig. 5[link](f)].

[Figure 5]
Figure 5
Intensity distribution of SiO2 glass from channels 14500 and 15000 at each step. (a) The uncorrected, (b) first-step, (c) second-step, (d) third-step, (e) fourth-step and (f) fifth-step distributions. The Poisson distribution is also plotted for comparison.

3.2. TFU and PNR

Fig. 6[link](a) shows the TFU (i: 14500–15000, N = 501) of the corrected and uncorrected scattering data of SiO2 glass as a function of average intensity, which is defined as a figure of merit for X-ray detectors in equation (14[link]). To compare the systematic error such as XRNU with the statistical error, the fractional uncertainty curve based on the Poisson statistics was also plotted. In the uncorrected data, the deviations from the Poisson noise are significant at intensities above 104 photons, reaching a plateau above 105 photons. In contrast, the corrected data are found to be close to the Poisson noise even at intensities above 106 photons.

[Figure 6]
Figure 6
(a) TFU (i: 14500–15000, N = 501) and (b) PNR (i: 14500–15000, N = 501) of the uncorrected and corrected scattering data of SiO2 glass, as a function of average intensity. The Poisson curve is also plotted for comparison.

Fig. 6[link](b) shows the PNR (i: 14500–15000, N = 501) of the corrected and uncorrected scattering data of SiO2 glass as a function of average intensity, which is defined as another figure of merit for X-ray detectors in equation (15[link]). The PNR of the uncorrected data increases exponentially with increasing average intensity, resulting in 23.0 at 5.7 × 106 photons. In contrast, the corrected data maintain the PNR of 2.3 even at 1.4 × 106 photons, leading to 3.8 at 5.7 × 106 photons.

3.3. Effects of the multi-step process and the mask

To examine improvements in data quality by the multi-step process and the mask in front of the sensor, the TFU (i: 14500–15000, N = 501) in the SiO2 glass scattering data with the order of 106 photons in intensity was plotted as a function of measurement time for correction factors in Fig. 7[link]. The single-step without the mask shows a gradual decrease in TFU with increasing time, resulting in 0.25% at 80 h. On the other hand, the TFU in the multi-step without the mask reaches 0.25% within 8 h, which is an order of magnitude shorter than the single-step. The significant reduction in measurement time indicates that some multiplier effects work on data correction in the multi-step. Furthermore, the TFU in the multi-step is further improved by the use of the mask. The observations suggest that the mask is useful in reducing some noise, which might be caused by an exposure of scattered X-rays to the wire bonding. Notice that the TFU before the XRNU correction with the mask was comparable with that without the mask. The results indicate that the accuracy of the alignment is sufficient to discuss XRNU.

[Figure 7]
Figure 7
TFU (i: 14500–15000, N = 501) of the corrected scattering data of SiO2 glass with intensity of the order of 106 photons, as a function of measurement time for correction factors. The single-step process without the mask and the multi-step with and without the mask are indicated.

3.4. Improvements in total-scattering data

Total-scattering measurements take into account both Bragg and diffuse scattering. This section describes the improvements in Bragg reflections of Si powder and diffuse scattering of SiO2 glass by our approach.

3.4.1. Bragg reflections

Si powder was measured to examine the improvements in Bragg reflections. Fig. 8[link](a) shows the forbidden reflection 222 in the uncorrected and corrected data, the intensity of which is three orders of magnitude lower than that of the primary reflection 111. The forbidden reflection is clearly observed in the corrected data as compared with the uncorrected data. Fig. 8[link](b) shows higher-order reflections around 25 Å−1 in Q, the intensity of which is four orders of magnitude lower than that of 111. These reflections in the uncorrected data are hidden under noise, whereas those in the corrected data appear at the expected peak positions.

[Figure 8]
Figure 8
Part of Bragg reflections in the uncorrected and corrected data of Si powder. (a) The forbidden reflection 222 and (b) higher-order reflections. For clarity, the corrected data are shifted slightly lower. Note that each higher-order reflection is indicated by the primary index.
3.4.2. Diffuse scattering

The scattering data of SiO2 glass were converted into the reduced structure function F(Q) = Q[S(Q) − 1] to examine the improvements in diffuse scattering. Fig. 9[link] shows the F(Q) obtained based on the uncorrected and corrected data as a function of scattering intensity (103 to 106 photons). The F(Q) based on the corrected and uncorrected data with 103 photons is too noisy to identify diffuse scattering especially at Q above 20 Å−1, which is caused by insufficient statistics [Fig. 9[link](a)]. With increasing intensity, the uncorrected data undergo almost no change above 104 photons, whereas the corrected data are found to be improved at high Q according to photon statistics [Figs. 9[link](b), 9[link](c) and 9[link](d)].

[Figure 9]
Figure 9
Reduced structure function F(Q) based on the uncorrected and corrected SiO2 glass data with intensities of the order of (a) 103, (b) 104, (c) 105 and (d) 106 photons. Note that the moving average (±10 channels) was conducted in the uncorrected and corrected data used for the F(Q) to improve statistics at high Q.

3.5. Improvements in PDFs

The total-scattering PDF, which can be obtained without any structural models, is considered to be one of the figures of merit for X-ray detectors. This section describes the improvements by our approach in the PDF of two crystalline powders with different crystallite sizes.

3.5.1. Nanometre crystal

The PDF analysis of anatase-type TiO2 nanoparticles was performed every 20 Å in interatomic distance. Fig. 10[link] shows the fitting results in the range 1.5–20 Å and 80–100 Å, the latter of which is close to the crystallite size. The reliability factors Rw based on the uncorrected data indicate 23% and 65% at the short and long ranges, respectively [Figs. 10[link](a) and 10[link](b)]. On the other hand, the Rw based on the corrected data are 16% and 38% at the short and long ranges, respectively [Figs. 10[link](c) and 10[link](d)]. The improvements in Rw indicate that the PDF from the corrected data is more reliable than that from the uncorrected data.

[Figure 10]
Figure 10
PDF G(r) of TiO2 nanoparticles fitted by the anatase-type structural model. Firstly, (a) and (b) show the results based on the uncorrected data in the range 1.5–20 Å and 80–100 Å in interatomic distance r, respectively. Secondly, (c) and (d) show the results based on the corrected data in the range 1.5–20 Å and 80–100 Å in r, respectively. Reliability factors Rw and the difference between observed and calculated PDFs are shown at the upper right and below in each figure, respectively.
3.5.2. Micrometre crystal

Fig. 11[link] shows the PDF G(r) of Ni powder up to 1000 Å in interatomic distance r. The G(r) based on the uncorrected data appears to be similar to that based on the corrected data [Fig. 11[link](a)]. The amplitude of the G(r) is two orders of magnitude larger than that of the difference between the uncorrected and corrected PDFs [Fig. 11[link](b)]. On the other hand, the amplitude of the G(r) is damped towards the r corresponding to the crystallite size, whereas that of the difference is almost constant at r above 300 Å. Fig. 11[link](c) shows the ratio of the difference to G(r). The ratio is found to increase with increasing r. The results indicate that the systematic error caused by XRNU affects the PDF at higher r rather than at lower r since the amplitude of the PDF at higher r is much smaller than that at lower r.

[Figure 11]
Figure 11
PDF G(r) of Ni powder up to 1000 Å in interatomic distances. Firstly, (a) is the G(r) based on the uncorrected and corrected data, which are shifted up and down, respectively. Secondly, (b) is the difference in G(r) between the uncorrected and corrected data and (c) is the ratio of the difference to the G(r).

4. Discussion

The present study was intended to develop a statistical approach to correct XRNU in microstrip detectors. Indeed, the multi-step process combined with the divided-accumulation technique reduced the PNR from 11.6 to 2.3 at the intensity of 1.4 × 106 photons, indicating an approach to the Poisson noise. As a result, a number of hidden Bragg reflections of Si powder, the intensity of which is four orders of magnitude lower than that of 111, were clearly observed. The diffuse scattering of SiO2 glass was obtained up to 27 Å−1 in Q. The resultant PDF of TiO2 nanoparticles was improved in a wide range of interatomic distances.

4.1. Origin of XRNU

In contrast to the Poisson noise, the TFU of the uncorrected data of SiO2 glass reaches a plateau at intensities above 104 photons, as shown in Fig. 6[link](a). The difference in TFU between the Poisson noise and the uncorrected data is estimated to be ∼0.8% at 106 photons. In photon-counting detectors, the dispersion of a threshold energy is recognized as one of the most important factors for XRNU. The threshold dispersion in MYTHEN can be reduced by the threshold-equalization technique (trimming) using the internal 6-bit digital-to-analog converter by a factor of 15 (Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]). Nevertheless, it has been reported that there still remains a dispersion of nearly 1% (Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]), which is larger than the Poisson noise above 104 photons. The remaining dispersion after the trimming is found to be close to the TFU in the uncorrected data. It follows from these arguments that most of the XRNU is caused by the threshold dispersion. Moreover, it is known that the distribution of the threshold energy is a function of ambient temperature as well as threshold energy. For this reason, the trimming for MYTHEN has been conducted at a fixed temperature for specific energies. In the present study, the same experimental booth temperature was maintained during measurements of correction factors and samples, minimizing the XRNU. In other words, correction factors must be collected every time the threshold energy and the ambient temperature are changed.

4.2. Alternative to the flat-field calibration

The flat-field calibration, which has been conventionally used to correct XRNU in X-ray detectors, has been applied to MYTHEN (Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]; Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]). The conventional calibration requires that a whole detector should be illuminated by uniform X-rays with a specific energy. We, however, have found it difficult to generate such X-rays to a precision of 0.1%, which corresponds to the Poisson noise at 106 photons. Another problem is that the conventional calibration does not take into account the Poisson noise. Therefore, the large XRNU as observed at both ends of channels can be corrected (Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]), whereas the conventional calibration has failed in correcting the XRNU at intensities above 104 photons. In contrast, the alternative approach is based on the statistical estimation that the scattering intensity at a fixed angle from an object is expected to be constant within the Poisson noise. The approach is also characterized by the use of any scattering object to correct XRNU in principle. We emphasize that our approach has succeeded in exploiting the dynamic range and the spatial resolution in MYTHEN for the first time.

4.3. XRNU correction for short-scale PDF

Let us compare our experimental method with the RA-PDF method described in Section 1[link], where a flat-panel detector was employed, in terms of the reduced structure function F(Q) of SiO2 glass. In RA-PDF, the F(Q) at high Q has been reported to be improved by increasing the measurement time, leading to a Qmax of 35 Å−1 (Chupas et al., 2007[Chupas, P. J., Chapman, K. W. & Lee, P. L. (2007). J. Appl. Cryst. 40, 463-470.]). Similarly, the F(Q) based on our corrected data is found to be improved according to photon statistics (Fig. 9[link]). Area detectors such as the flat panel and the imaging plate can average out XRNU between pixels when two-dimensional data are converted into one-dimensional data for data analysis. Such an averaging effect cannot be expected in microstrip detectors such as MYTHEN. From these arguments, we emphasize that XRNU in microstrip detectors should be corrected in terms of short-scale PDF analysis as well as multi-scale analysis.

4.4. Need for the XRNU correction in MYTHEN

MYTHEN has been employed as a detector for high-resolution and high-throughput powder diffraction experiments at synchrotron facilities (Bergamaschi et al., 2010[Bergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653-668.]; Thompson et al., 2011[Thompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637-648.]). Moreover, total-scattering measurements have been attempted taking advantage of its resolution (Haverkamp & Wallwork, 2009[Haverkamp, R. G. & Wallwork, K. S. (2009). J. Synchrotron Rad. 16, 849-856.]; Saleta et al., 2017[Saleta, M. E., Eleotério, M., Mesquita, A., Mastelaro, V. R. & Granado, E. (2017). J. Synchrotron Rad. 24, 1098-1104.]). Such attempts, however, have not been successful in obtaining the reliable PDF over a wide range of interatomic distances. Here let us compare the present study with the previous study in terms of the PDF analysis of anatase-TiO2 nanoparticles. In the previous study, the Rw for the distance range 1.5–20 Å was 26% (Haverkamp & Wallwork, 2009[Haverkamp, R. G. & Wallwork, K. S. (2009). J. Synchrotron Rad. 16, 849-856.]), which is larger than that of the present study by 10% [Fig. 10[link](c)]. The discrepancy in Rw between the two results may be attributed to not only the difference in Qmax but also the XRNU at high Q. In the present study, the Rw for the range 80–100 Å is improved by the XRNU correction by 27%, which is, however, still higher than that for the range 1.5–20 Å. This could be ascribed to non-uniformity in crystallite size and shape since the range is close to the crystallite size. In addition, the PDF analysis of micrometre crystal Ni clearly indicates that the impact of XRNU on PDF is significant with increasing interatomic distance [Fig. 11[link](c)]. Based on these findings, we emphasize that the XRNU in MYTHEN needs to be corrected to obtain the reliable PDF over a wide range of distances.

4.5. Stepwise updating of correction factors in the multi-step process

As shown in Fig. 7[link], the single-step process takes about 80 h to reach 0.25% in TFU, whereas the multi-step process takes about 8 h to reach the same TFU. For the single-step, the correction factors css in equation (2[link]) were calculated based on 160 measurements, in which case 1280 channels were partitioned every eight channels. For the multi-step, although the partitioning in the fifth step was close to that of the single step, the correction factors cms in equation (13[link]) were calculated based on only 24 measurements. The substantial reduction in hours and measurements can be interpreted by introducing global and local correction factors. In the first step of the multi-step, 16 measurements were performed by shifting the detector system by 80 channels to obtain the correction factors c1 for each channel, which can be defined as global correction factors for the second step. By using the global factors, only two measurements should be performed by shifting the detector system by half of 80 channels to obtain c2, which can be regarded as local correction factors for the second step. Note that there is no need to carry out 32 (= 1280/40) measurements by using global factors before calculating local factors. Next, c1 × c2 can be used as the global factors for the third step. Indeed, local correction factors c2 in equation (6[link]), c3 in equation (8[link]), c4 in equation (10[link]) and c5 in equation (12[link]) were obtained based on global factors c1, c1 × c2, c1 × c2 × c3 and c1 × c2 × c3 × c4, respectively. As a result, correction factors at each step are stepwise updated as shown in Fig. 3[link]. That is the reason why the multi-step process can significantly save hours and measurements to correct XRNU.

4.6. Limitations of our approach

We acknowledge that there are several limitations of our approach to correct XRNU. As shown in Fig. 6[link](a), there remains a discrepancy of ∼0.1% between the corrected data and the Poisson noise at 106 photons. The discrepancy could be ascribed to some random errors, except for the statistical error which is impossible for the approach to correct. Fig. 6[link](b) shows the systematic discrepancy in PNR as a function of average intensity. These results imply that the random error such as the thermal noise, which is negligible at intensities below 105 photons, becomes significant at intensities above 105 photons. From this point of view, the mask in front of the sensor plays an important role in reducing part of the random error (Fig. 7[link]). Based on these arguments, we conclude that our approach is powerful in correcting systematic errors such as XRNU.

In the present study, measurements for correction factors took as long as ∼80 h to evaluate the time evolution of the TFU. Compared with the typical measurement time for a sample, this time is too long. As shown in Fig. 7[link], however, the TFU reachs a plateau in about 30 h, indicating that one day would be sufficient thanks to the stepwise updating described in Section 4.5[link]. Furthermore, correction factors need to be collected for each threshold energy since most of the XRNU in MYTHEN are considered to be caused by the dispersion of the threshold energy. In other words, once correction factors for a fixed threshold energy are obtained, these factors can always be applied to scattering data measured at the same energy.

5. Conclusion

We have succeeded in developing an unbiased approach to correct the XRNU in microstrip detectors such as MYTHEN and evaluated the quality of total-scattering data in terms of the TFU and PNR, which are proposed as figures of merit for X-ray detectors in the present article, as well as F(Q) and PDFs. The principle of our approach could be applied to not only microstrip detectors but also pixel detectors such as PILATUS (DECTRIS Ltd). Unlike the conventional flat-field calibration, our alternative approach can minimize XRNU over a wide range of intensities, making best use of the dynamic range of the detectors. Consequently, the approach facilitates a discrimination between the systematic error, such as the XRNU, and the random error, such as the thermal and Poisson noise, leading to the development of new X-ray detectors with a PNR of 1. We believe that the present study provides a new benchmark for the evaluation of X-ray detectors and their data. In the near future, we will develop a data-driven approach to the PDF based on high-accuracy (PNR ≃ 2 at 106 photons) and high-resolution (ΔQ ≃ 10−4 Å−1) total-scattering measurements over a wide range of Q (∼30 Å−1) to obtain 0.1 Å-resolution PDFs at interatomic distances above 100 Å, leading to the visualization of heterogeneous phenomena in functional nanomaterials.

Acknowledgements

The authors thank Mr Kazuya Shigeta and Mr Kenta Ishino of RIKEN SPring-8 Center for their skilled technical contributions and fruitful discussions; and Professor Masaki Takata of Tohoku University and Dr Tetsuya Ishikawa of RIKEN SPring-8 Center for their support and encouragement. The synchrotron radiation experiments were performed at the BL44B2 of SPring-8 with the approval of RIKEN (Proposal Nos. 20130087, 20140055, 20150056 and 20160056).

Funding information

This work was supported by JSPS KAKENHI Grant Number JP16K04870 (to KK) and JST, PRESTO Grant Number JPMJPR1872, Japan (to KK).

References

First citationBergamaschi, A., Cervellino, A., Dinapoli, R., Gozzo, F., Henrich, B., Johnson, I., Kraft, P., Mozzanica, A., Schmitt, B. & Shi, X. (2010). J. Synchrotron Rad. 17, 653–668.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationChupas, P. J., Chapman, K. W. & Lee, P. L. (2007). J. Appl. Cryst. 40, 463–470.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationChupas, P. J., Qiu, X., Hanson, J. C., Lee, P. L., Grey, C. P. & Billinge, S. J. L. (2003). J. Appl. Cryst. 36, 1342–1347.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationEgami, T. & Billinge, S. J. L. (2012). Underneath the Bragg Peaks: Structural Analysis of Complex Materials, Pergamon Materials Series, Vol. 16, 2nd ed. Amsterdam: Elsevier.  Google Scholar
First citationHaverkamp, R. G. & Wallwork, K. S. (2009). J. Synchrotron Rad. 16, 849–856.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKato, K., Hirose, R., Takemoto, M., Ha, S., Kim, J., Higuchi, M., Matsuda, R., Kitagawa, S. & Takata, M. (2010). AIP Conf. Proc., 1234, 875–878.  CrossRef CAS Google Scholar
First citationKato, K. & Tanaka, H. (2016). Adv. Phys. X, 1, 55–80.  Google Scholar
First citationSaleta, M. E., Eleotério, M., Mesquita, A., Mastelaro, V. R. & Granado, E. (2017). J. Synchrotron Rad. 24, 1098–1104.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSchmitt, B., Brönnimann, C., Eikenberry, E. F., Gozzo, F., Hörmann, C., Horisberger, R. & Patterson, B. (2003). Nucl. Instrum. Methods Phys. Res. A, 501, 267–272.  Web of Science CrossRef CAS Google Scholar
First citationSkrobas, K., Stelmakh, S., Gierlotka, S. & Palosz, B. F. (2017). J. Appl. Cryst. 50, 1821–1829.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationThompson, S. P., Parker, J. E., Marchal, J., Potter, J., Birt, A., Yuan, F., Fearn, R. D., Lennie, A. R., Street, S. R. & Tang, C. C. (2011). J. Synchrotron Rad. 18, 637–648.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationYang, X., Juhas, P., Farrow, C. L. & Billinge, S. J. L. (2014). arXiv:1402.3163.  Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.