AP-XPS beamline, a platform for operando science at Pohang Accelerator Laboratory

CROSSMARK_Color_square_no_text.svg

aDepartment of Physics and Photon Science, Gwangju Institute of Science and Technology, Gwangju 61005, Republic of Korea, bAdvanced Nano-Surface Research Group, Korea Basic Science Institute, Daejeon 34133, Republic of Korea, and cBeamline Research Division, Pohang Accelerator Laboratory, Pohang 37674, Republic of Korea
*Correspondence e-mail: cla100@postech.ac.kr, bsmun@gist.ac.kr, kjkim@postech.ac.kr

Edited by Y. Amemiya, University of Tokyo, Japan (Received 11 September 2019; accepted 15 December 2019; online 28 January 2020)

Beamline 8A (BL 8A) is an undulator-based soft X-ray beamline at Pohang Accelerator Laboratory. This beamline is aimed at high-resolution ambient-pressure X-ray photoelectron spectroscopy (AP-XPS), soft X-ray absorption spectroscopy (soft-XAS) and scanning photoemission microscopy (SPEM) experiments. BL 8A has two branches, 8A1 SPEM and 8A2 AP-XPS, that share a plane undulator, the first mirror (M1) and the monochromator. The photon beam is switched between the two branches by changing the refocusing mirrors after the monochromator. The acceptance angle of M1 is kept glancing at 1.2°, and Pt is coated onto the mirrors to achieve high reflectance, which ensures a wide photon energy range (100–2000 eV) with high resolution at a photon flux of ∼1013 photons s−1. In this article, the main properties and performance of the beamline are reported, together with selected experiments performed on the new beamline and experimental system.

1. Introduction

Currently, scientists and engineers are searching for possible solutions to mitigate serious imminent global issues such as climate change, deficit energy resource and air pollution. These issues seem to be independent, yet they are closely interconnected. Solutions can be found from various fields of science and engineering, for example environmental science, catalysis, nanotechnology and energy technology, and, in particular, material science can provide a starting platform for any possible solution.

In material science, identifying or modifying the `surface/interface' of the materials is the most important process as most physical or chemical reactions take places at the surface or interfaces (Ertl, 2015[Ertl, G. (2015). Nobel Lecture: Reactions at Surfaces: From Atoms to Complexity, doi: 10.1002/anie.200800480.]). While many analytical techniques are developed for studying the surface/interface, electron- or ion-based techniques provide highly surface sensitive information due to their short inelastic mean free path, e.g. ion scattering, secondary ion mass spectroscopy and X-ray photoemission spectroscopy. These tools have proven to be highly effective in studying the gas/solid surface (or interface) under ultrahigh vacuum (UHV) conditions. However, in the case of the liquid/solid surface (or interface) or the gas/solid interface under elevated pressure conditions, electron- or ion-based tools cannot be used.

In fact, since the early days of surface science, there has been a huge amount of effort in adapting UHV-based surface-science techniques to ambient-pressure (AP) systems (Siegbahn, 1969[Siegbahn, K. (1969). ESCA Applied to Free Molecules. Amsterdam: North Holland.]; Somorjai, 1978[Somorjai, G. A. (1978). Science, 201, 489-497.]; Joyner et al., 1979[Joyner, R. W., Roberts, M. W. & Yates, K. (1979). Surf. Sci. 87, 501-509.]). Any surface chemical reaction mechanism occurring under UHV can differ significantly from that taking place under ambient-pressure conditions. As in situ surface studies have become increasingly important over the years, a number of innovative operando spectroscopy and microscopy techniques have also been developed (Shen, 1989[Shen, Y. R. (1989). Nature, 337, 519-525.]; Hu et al., 1995[Hu, J., Xiao, X. D., Ogletree, D. F. & Salmeron, M. (1995). Science, 268, 267-269.]; Hansen et al., 2001[Hansen, T. W., Wagner, J. B., Hansen, P. L., Dahl, S., Topsøe, H. & Jacobsen, C. J. H. (2001). Science, 294, 1508-1510.]; Hendriksen & Frenken, 2002[Hendriksen, B. L. M. & Frenken, J. W. M. (2002). Phys. Rev. Lett. 89, 046101.]; Donald, 2003[Donald, A. M. (2003). Nat. Mater. 2, 511-516.]; Hüfner, 2003[Hüfner, S. (2003). Photoelectron Spectroscopy: Principles and Applications, 3rd rev. ed. Berlin, Heidelberg: Springer.]; Ferrer et al., 2007[Ferrer, S., Ackermann, M. D. & Lundgren, E. (2007). MRS Bull. 32, 1010-1014.]; Forsberg et al., 2007[Forsberg, J., Duda, L. C., Olsson, A., Schmitt, T., Andersson, J., Nordgren, J., Hedberg, J., Leygraf, C., Aastrup, T., Wallinder, D. & Guo, J. H. (2007). Rev. Sci. Instrum. 78, 083110.]). Ambient-pressure X-ray photoemission spectroscopy (AP-XPS) is one of the prominent techniques that can provide surface/interface information under near ambient pressure conditions, i.e. in the ∼mbar range (Siegbahn, 1969[Siegbahn, K. (1969). ESCA Applied to Free Molecules. Amsterdam: North Holland.]; Joyner et al., 1979[Joyner, R. W., Roberts, M. W. & Yates, K. (1979). Surf. Sci. 87, 501-509.]; Ruppender et al., 1990[Ruppender, H. J., Grunze, M., Kong, C. W. & Wilmers, M. (1990). Surf. Interface Anal. 15, 245-253.]; Ogletree et al., 2002[Ogletree, D. F., Bluhm, H., Lebedev, G., Fadley, C. S., Hussain, Z. & Salmeron, M. (2002). Rev. Sci. Instrum. 73, 3872-3877.]; Salmeron & Schlögl, 2008[Salmeron, M. & Schlögl, R. (2008). Surf. Sci. Rep. 63, 169-199.]). AP-XPS, built with differential pumping electrostatic lens system schemes, has clearly made a big contribution to the operando science community (Bluhm, 2010[Bluhm, H. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 71-84.]).

X-ray photoemission spectroscopy (XPS), the predecessor of AP-XPS, has been demonstrated as an invaluable technique in the study of filled electronic states of solids, as well as helping to determine the nature of interactions between solid surfaces and molecular species. But there is one main barrier of the technique – that XPS measurements on clean surfaces for surface science are required to be performed under UHV conditions (∼10−10 mbar). This limitation was partially overcome with the development of differentially pumped AP-XPS systems, which began in the early 1970s (Siegbahn & Siegbahn, 1973[Siegbahn, H. & Siegbahn, K. (1973). J. Electron Spectrosc. Relat. Phenom. 2, 319-325.]). Despite much effort, they suffered from poor electron yields due to the scattering of electrons in the gas phase. Then, with advances in electrostatic lens systems coupled with differential pumping of electron analyzer and high-flux synchrotron radiation sources, AP-XPS measurements became possible in the 100 mbar pressure range. AP-XPS has been recognized by scientific communities as an important in situ tool to study water, environmental science, catalysis and many other important fields (Salmeron & Schlögl, 2008[Salmeron, M. & Schlögl, R. (2008). Surf. Sci. Rep. 63, 169-199.]; Starr et al., 2013[Starr, D. E., Liu, Z., Hävecker, M., Knop-Gericke, A. & Bluhm, H. (2013). Chem. Soc. Rev. 42, 5833-5857.]; Shavorskiy et al., 2014[Shavorskiy, A., Karslioglu, O., Zegkinoglou, I. & Bluhm, H. (2014). Synchrotron Radiat. News, 27(2), 14-23.]). Since 2000, AP-XPS systems have been continuously installed at most major synchrotron radiation facilities around the world: ALS, BESSY, SSRL, MAX IV, SLS, ALBA, SOLEIL, SPring-8, Diamond, SSRF, and so on. Using these systems, many researchers have already published meaningful results in various fields (Artiglia et al., 2017[Artiglia, L., Orlando, F., Roy, K., Kopelent, R., Safonova, O., Nachtegaal, M., Huthwelker, T. & van Bokhoven, J. A. (2017). J. Phys. Chem. Lett. 8, 102-108.]; Timm et al., 2018[Timm, R., Head, A. R., Yngman, S., Knutsson, J. V., Hjort, M., McKibbin, S. R., Troian, A., Persson, O., Urpelainen, S., Knudsen, J., Schnadt, J. & Mikkelsen, A. (2018). Nat. Commun. 9, 1412.]; Kim et al., 2018[Kim, J., Park, W. H., Doh, W. H., Lee, S. W., Noh, M. C., Gallet, J. J., Bournel, F., Kondoh, H., Mase, K., Jung, Y., Mun, B. S. & Park, J. Y. (2018). Sci. Adv. 4, eaat3151.]; Yu et al., 2019[Yu, L., Takagi, Y., Nakamura, T., Sakata, T., Uruga, T., Tada, M., Iwasawa, Y., Masaoka, S. & Yokoyama, T. (2019). J. Phys. Chem. C, 123, 603-611.]; Soler et al., 2019[Soler, L., Casanovas, A., Ryan, J., Angurell, I., Escudero, C., Pérez-Dieste, V. & Llorca, J. (2019). ACS Catal. 9, 3641-3647.]; Diulus et al., 2019[Diulus, J. T., Frederick, R. T., Li, M., Hutchison, D. C., Olsen, M. R., Lyubinetsky, I., Árnadóttir, L., Garfunkel, E. L., Nyman, M., Ogasawara, H. & Herman, G. S. (2019). ACS Appl. Mater. Interfaces, 11, 2526-2534.]; Tesch et al., 2019[Tesch, M. F., Bonke, S. A., Jones, T. E., Shaker, M. N., Xiao, J., Skorupska, K., Mom, R., Melder, J., Kurz, P., Knop-Gericke, A., Schlögl, R., Hocking, R. K. & Simonov, A. N. (2019). Angew. Chem. Int. Ed. 58, 3426-3432.]; Cai et al., 2019[Cai, J., Dong, Q., Han, Y., Mao, B. H., Zhang, H., Karlsson, P. G., Åhlund, J., Tai, R. Z., Yu, Y. & Liu, Z. (2019). Nucl. Sci. Tech. 30, 81.]).

Recently, beamline 8A (BL 8A) at Pohang Accelerator Laboratory (PAL) has been successfully reconstructed and an AP-XPS end-station is installed. Since 1999, BL 8A, the first undulator beamline at PAL, has been dedicated for surface/interface and material science by providing scanning photoemission microscopy (SPEM) (8A1) and high-resolution photoemission spectroscopy (HR-PES) (8A2) (Shin et al., 1999[Shin, H.-J., Chung, Y. & Kim, B. (1999). J. Electron Spectrosc. Relat. Phenom. 101-103, 985-989.]). The undulator at BL 8A was designed to produce photon energies in the range 50–2000 eV, and BL 8A has been one of the most popular beamlines at PAL as a user-friendly end-station that can provide both microscopic and spectroscopic techniques for surface/interface science. However, the practically available photon energy range at the end-station was gradually reduced to 100–800 eV as beamline components including the optical system went out of exhaustive operation. User groups in surface/interface and material science have made continuous requests to recover the higher energy above 2000 eV and introduce an additional advanced photoemission method, e.g. AP-XPS, for operando science. Finally, in December 2014, PAL and KBSI signed a Memorandum of Understanding for the construction of a new beamline and AP-XPS experimental system to initiate a joint research program in emerging fields of science.

In this paper, we outline the main properties and performance of the beamline as characterized during commissioning at the PAL storage ring (PLS II). For fertilizing new operando science, the beamline was designed with a wide photon energy range (100–2000 eV) with high photon resolution. A high photon flux, ∼1013 photon s−1, was achieved by keeping the M1 mirror acceptance angle glancing at 1.2° as well as minimizing the number of total reflecting mirrors, four for the 8A1 SPEM branch and six for the 8A2 AP-XPS branch. We will also discuss the experimental capabilities and present a few showcase experiments which have been performed on the new beamline.

2. Beamline overview

2.1. Photon source for BL 8A

The U6.8 undulator for BL 8A receives electron bunches of 3 GeV and 400 mA from the storage ring of the PAL synchrotron (Shin et al., 2013[Shin, S., Kwon, S., Kim, D. T., Kim, D. E., Kim, M., Kim, S. H., Kim, S. C., Kim, J., Kim, C., Park, B., Park, S. S., Park, S. J., Park, E., Son, Y., Yoon, J., Lee, B., Lee, E., Lee, J. W., Lee, H. S., Joo, Y., Choi, J., Ha, T., Hwang, W., Hwang, I., Lee, J. Y., Oh, B., Lee, C. H., Lee, H. S., Kim, J. Y., Hwang, J. Y., Nam, S. H. & Cho, M. (2013). J. Instrum. 8, P01019.]; Hwang et al., 2014[Hwang, I., Huang, J. Y., Kim, M., Lee, B. J., Kim, C., Choi, J. Y., Kim, M. H., Lee, H. S., Moon, D., Lee, E. H., Kim, D. E., Nam, S. H., Shin, S. & Cho, M. (2014). Rev. Sci. Instrum. 85, 055113.]). The electron velocity is 0.9999999963733c and the Lorentz factor γ, given as γ = 1/(1 − v2/c2)1/2, is 11742. A total of 48 permanent magnets are displaced with a periodicity of 6.8 cm and the total length of the undulator is 3.3 m. The mechanically allowed gap between the upper and lower magnet arrays is from 16 to 90 mm, for which the deflection parameter K of the undulator is from 5.769 to 0.105, given by the following equation,

[K = {{eB_0\,\lambda_{\rm{u}}}\over{2\pi m_{\rm{e}}c}},\eqno(1)]

where B0 is the magnetic field at the centre of the undulator, me is the electron mass and λu is the period of the permanent magnets in the undulator. B0 is a function of the gap and the geometry of the undulator.

With the given value of the parameter K, the photon energy E of the nth harmonics, which is radiated with emission angle θ referred to the beam propagating direction, is determined by the following equation,

[E = {{ 2n\gamma^{\,2}hc }\over{ \lambda_{\rm{u}}\left[1+(K^{\,2}/2)+\theta^{\,2}\gamma^{\,2}\right] }}. \eqno(2)]

As a result of equation (2)[link], the U6.8 undulator covers the energy range from 71.236 eV to 1248.94 eV for the first-harmonic radiation. Including the third and fifth harmonics, the undulator can cover the entire energy range that the beamline was designed for, i.e. 100–2000 eV, with a flux density of 1017 photons s−1 mrad−2 (0.1% bandwidth)−1 theoretically as shown in Fig. 1[link]. The radiated beam is linearly polarized.

[Figure 1]
Figure 1
Calculated photon flux density of the U6.8 undulator radiation. The lines coloured black, red and blue represent the first, third and fifth harmonics, respectively. The inset shows a schematic diagram of the undulator.

2.2. Beamline specifications

The beamline optics are designed to accommodate the needs of the SPEM and AP-XPS end-stations, providing X-rays over a wide energy range from 100 to 2000 eV. The undulator can operate in the energy range from 70 eV up to 3000 eV, which corresponds to operation up to the third harmonic. The photon beam is collimated vertically by the first M1 mirror and delivered to the monochromator. The incident angle on the M1 mirror is 1.2° to ensure high reflectance at higher energy (∼1700 eV). The plane-grating monochromator (PGM) system is designed with SX-700-based optics to provide X-rays over a wide energy range (Petersen & Baumgärtel, 1980[Petersen, H. & Baumgärtel, H. (1980). Nucl. Instrum. Methods, 172, 191-193.]). Fig. 2[link] shows the conceptual design of the PGM. The operation of the monochromator follows the grating equation described by [m\lambda/d] = [2\cos\theta\sin(\theta+\beta)], where m is the diffraction order, λ is the photon wavelength and d is the line spacing corresponding to the ruling density of the grating. D, the vertical distance offset between the entering and exiting X-ray paths, is 15 mm. By having a pre-collimated beam on the monochromator from the M1 mirror along the vertical direction, the fix constant, Cff, is allowed to vary freely. Cff is given by Cff = cosβ/cosα, and 2θ, the variable inclusive angle, is given by 2θ = αβ, where α > 0, β < 0 and 2θ are shown in Fig. 2[link]. This additional flexibility of Cff enables the monochromator to be optimized for enhanced flux and improved energy resolution.

[Figure 2]
Figure 2
Schematic drawing of the operation mechanism of the PGM. The grating module rotates around the green pivotal point and the centre of the pre-mirror rotates around the blue pivotal point close to the grating, which always transfers white light to the centre of the grating.

Two different gratings are installed in a single monochromator unit, where the low-energy grating (LEG) is for the photon energy range 100–1100 eV, and the high-energy grating (HEG) is for the photon energy range 900–2000 eV. The LEG is ruled with a 400 lines mm−1 line density and 16 nm groove depth, and HEG is ruled with a 500 lines mm−1 line density and 8 nm groove depth.

The diffracted beam from the monochromator is focused onto the exit slit and switched by inserting the M3-1 (SPEM, 8A1) or M3-2 (AP-XPS, 8A2) toroidal mirror. As a result, the monochromatic beam after the M3 mirrors focuses onto the exit slit with a beam size of 210 µm (horizontal, H) × 5.9 µm (vertical, V) for 8A1 and 280 µm (H) × 7.1 µm (V) for 8A2. The aperture size of the exit slit can be determined by controlling four-way blades (xy). Finally it is designed to achieve an energy resolving power of more than 5000 in the energy range that each grating can provide, when the vertical aperture size is kept at 40 µm, and LEG and HEG are operated with Cff = 2.5 and Cff = 1.7, respectively, with negative diffraction order.

In the SPEM (8A1) beamline, this monochromatic beam passes through a Fresnel zone plate and reaches the sample with a beam size of 100 nm. In the AP-XPS (8A2) beamline, Kirkpatrick–Baez (KB) mirrors, located at the end of the photon transfer line, tightly refocus the monochromatic beam onto the AP-XPS sample position. The FWHM of the beam size at the focal spot of the AP-XPS sample is less than 50 µm (V) × 50 µm (H). The beamline's optical layout and design are schematically shown in Fig. 3[link]. The optical and geometrical specifications of each optical element are listed in Table 1[link].

Table 1
Beamline optics specifications

Name of optics M1 mirror M2 pre-mirror Gratings M3-1 (8A1) / M3-2 (8A2) HKB (8A2) VKB (8A2)
Position from source 19 m 22 m 23.5 m 22 m / 25 m 19 m 19 m
Shape Cylindrical Plane Plane Toroidal Cylindrical Cylindrical
    LEG: 400 l mm−1      
    HEG: 500 l mm−1      
Function Vertical collimation Just reflection Diffraction Focusing to exit slit Horizontal refocusing Vertical refocusing
Tangential radius Infinite Infinite Infinite 32000 cm / 40000 cm 10000 cm 7550 cm
Sagittal radius 8012.3 mm Infinite Infinite 4400 mm / 3138 mm Infinite Infinite
Substrate material Glidcop (internal cooling) Glidcop (internal cooling) Single-crystal silicon Single-crystal silicon Single-crystal silicon Single-crystal silicon
Incident angle 1.2° (in and out) Variable (0.9°–5.0°) Variable (0.9°–5.0°) 2.0° / 1.5° 1.5° (in and out) 2.0° (in and out)
Coating thickness (Å) Au (600 Å) Au (600 Å) Au (600 Å) Au (600 Å) Au (600 Å) Au (600 Å)
Beam size @ 830 eV (µm) 1240 × 610 1330 × 610 1360 × 610 1370 × 4060 / 1370 × 4060 760 × 2370 570 × 2900
Footprint @ 830 eV (µm) 58400 × 606 1330 × 17600 1360 × 119900 49500 × 4060 / 54600 × 4060 29100 × 2370 570 × 78700
[Figure 3]
Figure 3
Optical layout and design of BL 8A. From the undulator, M1 horizontal focusing mirror, monochromator system with M2 and gratings, M3 and M3′ branching mirror for SPEM and AP-XPS, respectively, and exit slits and KB mirror system for AP-XPS are located.

3. Beamline performance: photon resolution and photon flux

Fig. 4[link] shows the nitro­gen K-edge absorption spectrum of gas-phase N2, which indicates the resolving power of the grating. The measurement was taken using a gas cell located between the exit slit and the KB mirrors of beamline 8A2 with an exit slit opening of 40 µm (V) × 200 µm (H). The gas cell was composed of two parallel plates and filled with N2 gas at a pressure of 10−5 mbar. One of the plates is biased by +100 V and pushes photo-ionized N2 molecules, and the other detects the collision of ions to itself by measuring current outputs.

[Figure 4]
Figure 4
Photo-absorption spectrum of the nitro­gen K-edge taken under 10−5 mbar N2 gas pressure. The best deconvolution using a Voight function allows an 80 meV Lorentzian width and 40 meV Gaussian width.

The spectrum was deconvoluted with a Voigt function for which the Lorentzian and Gaussian widths were 80 meV and 40 meV, respectively. Since the Lorentzian width originates from lifetime broadening, the smaller value for this beamline compared with those of other soft X-ray beamlines (Lee & Shin, 2001[Lee, M. K. & Shin, H. J. (2001). Nucl. Instrum. Methods Phys. Res. A, 467-468, 508-511.]; Watanabe et al., 2004[Watanabe, M., Toyoshima, A., Azuma, Y., Hayaishi, T., Yan, Y. & Yagishita, A. (2004). Proc. SPIE, 3150, 58-66.]) can be due to using lower N2 gas pressure and providing fewer collision opportunities between ions and electrons. Additionally, the low signal-to-noise ratio of these spectra may come from the low N2 gas pressure. At 400 eV photon energy, the resolving power E/ΔE is estimated to be 400 eV/40 meV = 10000, which matches precisely the designed value of the LEG.

Also, the photon flux at the end-station, which will be used as an important value in actual experiments, was measured using a silicon photodiode (IRD 100G) located right after the exit slit for the 8A1 SPEM end-station (Gullikson et al., 1996[Gullikson, E. M., Korde, R., Canfield, L. R. & Vest, R. E. (1996). J. Electron Spectrosc. Relat. Phenom. 80, 313-316.]). In the most frequently used energy range of 100–2000 eV in both beamlines, the photon flux could be estimated as ∼1013 photons s−1.

4. Experimental station

4.1. Windowless aperture system with differential pumping stages: buffering the pressure gap between the beamline and the experimental chamber

The main obstacle in performing AP-XPS experiments is due to the elevated pressure in the main chamber in which experiments are conducted, which conflicts with the UHV requirement of the beamline. In many cases, a photon-transparent window, e.g. silicon nitride (Si3N4) or aluminium membrane (Al, thickness of ∼100 nm), has been installed between the beamline and AP-XPS experiment chamber to maintain the UHV condition of the beamline. BL 8A2, however, adopts a windowless differential pumping station (SPECS, Germany), installed between the experimental system and beamline with an aperture system with 200 µm-diameter capillary. Using four pumping stages and four apertures (with diameters of 6 mm, 5 mm, 4 mm and 3 mm in order from upstream), the pressure at the KB-mirror chamber can be maintained at ∼10−9 mbar.

4.2. Experimental components in AP-XPS

The 8A2 AP-XPS end-station is aimed at surface studies under near ambient pressure conditions (up to 25 mbar, a pressure where liquid-phase water can exist at room temperature) as well as UHV conditions. The end-station consists of preparation and analysis chambers with two manipulators: one for preparing a well defined sample surface in UHV conditions and the other for AP-XPS measurements, as can be seen in Fig. 5(a)[link]. The preparation chamber includes the usual equipment for cleaning and annealing surfaces (Ar+ ion sputter gun and e-beam heater) and characterization of surface structure (LEED optics). There is a load-lock chamber for transferring the sample into the UHV chamber without venting the whole system. The analysis chamber, which has a backfilling configuration, is equipped with a SPECS Phoibos NAP 150 hemispherical analyzer for XPS. A Cr Kα X-ray source (hν = 5.417 keV) is also installed in the main chamber as an auxiliary photon source, and allows the bulk property to be analyzed with a higher probing depth by detecting photoelectrons with high kinetic energy. On the sample manipulator in the analysis chamber there are three sample stages for cooling down the sample using liquid nitro­gen, annealing the sample using an IR laser (Ostec, up to 1500 K) and direct-current heating of semiconductor samples, respectively. Fig. 5(b)[link] shows the sample configuration inside the sample analysis chamber. Brief information on the AP-XPS experimental system is summarized in Table 2[link].

Table 2
Brief information for the AP-XPS experimental system

Electron analyzer SPECS Phoibos NAP 150 hemispherical analyser with 2D delay line detector
Photon energy 100–2000 eV, 5417 eV (Cr X-ray source)
Pressure UHV ∼25 mbar
Temperature control Liquid nitro­gen (150 K)
IR laser heater (up to 1500 K)
Direct-current heating (up to 1500 K)
Gases N2, O2, CO2, CO, Ar, Ne…
Equipment Low-energy electron diffraction (LEED)
Ion gun
Electron beam heater
Residual gas analyser (RGA)
Applicable research area Physics at surface and interface
Gas-phase heterogeneous catalysis
Electrochemistry (battery, fuel cell, corrosion, electrocatalysis)
Photovoltaic
Environmental science
Biological systems…
[Figure 5]
Figure 5
(a) Main experimental components of the AP-XPS end-station. (b) Beam spot site inside the analysis chamber. The angle between the analyser cone and beamline is 55° and the angle between the analyser cone and the Cr X-ray source is 79°.

4.3. Probing depth change of SiO2/Si on photon energy

To test the beamline photon energy and the analyzer, XPS measurements were carried out on a SiO2/Si (100) sample for various photon energies. Fig. 6[link] shows the Si 2p XPS spectral changes for various photon energies under UHV conditions. For ease of comparison and avoidance of the difference by photoionization cross section, the spectra are normalized by the maximum intensity of the bulk Si 2p3/2 peak. Since the natural oxide was already formed on the surface, the oxide peak of Si 2p, whose intensity varies with photon energy, follows the typical escape depth curve of an electron. Using a photon energy of 150 eV to 1100 eV for the LEG, all the data points were collected only once with a 0.1 s dwell time and 20 meV step size at the most optimized position. The pass energy was 2 eV from 150 eV to 700 eV and 10 eV from 800 eV to 1100 eV. When the Si 2p core-level was measured using photon energies of 1586 eV and 1950 eV, for the HEG, the accumulation number is increased to overcome the reduced photon flux by five times and ten times, respectively, with 0.1 s dwell time and 50 eV pass energy. In summary, we would like to demonstrate the practical photon energy range by measuring the Si 2p core-level on SiO2/Si (100) and prove that the available photon energy range is 100–2000 eV at BL 8A.

[Figure 6]
Figure 6
Si 2p spectra corresponding to photon energies from 150 eV to 1950 eV. The spectra are normalized with respect to the intensity of the Si 2p3/2 peak. For the LEG, the energy region from 150 eV to 1100 eV is covered, while the HEG covered 1586 eV to 1950 eV.

4.4. Intensity change of Au 4f7/2 under N2 environments

A simple performance test of AP-XPS was carried out with an Au reference. Fig. 7(a)[link] shows the intensity change of Au 4f7/2 for a gradual change of the N2 pressure at a photon energy of 520 eV. All of the spectra in Fig. 7[link] are acquired with a 0.1 s dwell time, 50 eV pass energy, 50 meV step energy, 3 mm × 25 mm analyzer slit and 40 µm (V) × 200 µm (H) 8A2 exit slit. All signals are normalized using the beam current at the KB mirror. As the pressure of N2 is varied from 10−9 to 0.1 mbar, Au 4f7/2 has almost the same intensity. At 1 mbar, the signal intensity starts to decrease and reduces rapidly in the 1–10 mbar range. At 10 mbar, the signal is very low due to the short inelastic mean free path of the photo-electrons which cannot overcome the distance between the sample and the nozzle of the spectrometer. In practice, signal is collected sufficiently to distinguish components in the spectrum up to 7–8 mbar in the case of N2 gas for photon energies of the LEG. Fig. 7(b)[link] shows the Au 4f7/2 spectra at photon energies of 750 eV and 950 eV at 10 mbar. It shows an increased intensity because of the prolonged inelastic mean free path of the electrons at higher photon energy.

[Figure 7]
Figure 7
(a) Au 4f intensity change at elevated N2 pressure. The black dots show the maximum intensities of Au 4f7/2 peaks acquired from the spectra in the inset of (a). (b) Raw spectra of Au 4f taken under a 10 mbar N2 gas environment. All the spectra are collected in one scan.

4.5. Photo-induced nitro­gen doping on graphene/Ge at near ambient pressure

An investigation of the photo-induced effect during the N2 gas pressurization onto graphene/Ge is a good example of AP-XPS. The graphene on Ge (110) substrate system has been reported as a good stage for dry-transferring of graphene (Kim et al., 2013[Kim, K. J., Yang, S., Park, Y., Lee, M., Kim, B. & Lee, H. (2013). J. Phys. Chem. C, 117, 2129-2134.]; Yang et al., 2019[Yang, S. J., Choi, S., Odongo Ngome, F. O., Kim, K. J., Choi, S. Y. & Kim, C. J. (2019). Nano Lett. 19, 3590-3596.]), but many of its properties are not yet characterized. For the investigation, two identical graphene/Ge samples are prepared as previously reported (Yang et al., 2019[Yang, S. J., Choi, S., Odongo Ngome, F. O., Kim, K. J., Choi, S. Y. & Kim, C. J. (2019). Nano Lett. 19, 3590-3596.]), and each is located at different stages in the main chamber manipulator to compare the effects of photon irradiation by exposing only one of them to a synchrotron beam (hν = 700 eV) during the pressurization. N2 gas pressure was controlled from UHV to 10 mbar to trace the generation of nitro­gen species on the surface of the sample, and then pumped out to UHV again. By that process, the same atmospheric conditions are applied to both samples. During the whole experiment, no heat or other gases are applied on the samples. Fig. 8[link] shows the fitting results of C 1s, Ge 3d and N 1s spectra of the exposed sample measured at UHV before and after pressurization, and the LEED patterns of the exposed and unexposed samples.

[Figure 8]
Figure 8
UHV states of exposed graphene/Ge (110) (a) before and (b) after pressurization, and (c) LEED patterns of final N 1s states of exposed and unexposed graphene/Ge samples to the synchrotron beam. N-related species are produced immediately after arrival at the 10 mbar condition under the irradiation. All the spectra are taken at 700 eV photon energy and 50 eV pass energy.

As shown in Fig. 8(b)[link], several components of N 1s at binding energy 399.3 eV (NA), 400.7 eV (NB) and 402 eV (NC) were found after the pressurization only for the photon-irradiated sample and only for the irradiated spot. These components are not reduced or removed even by annealing or ageing in UHV. Simultaneously, C 1s of final states has an enhanced shoulder in the higher binding energy region compared with its initial state, shown in Fig. 8(a)[link]. On the other hand, Ge 3d shows no change during the process. Additionally, beam exposure did not affect the LEED pattern of graphene as shown in Fig. 8(c)[link]. Therefore, it can be inferred that nitro­gen atoms have some interaction with carbon atoms by irradiation but do not destroy the overall 2D structure of graphene. In previous investigations (Shao et al., 2010[Shao, Y., Zhang, S., Engelhard, M. H., Li, G., Shao, G., Wang, Y., Liu, J., Aksay, I. A. & Lin, Y. (2010). J. Mater. Chem. 20, 7491-7496.]; Usachov et al., 2011[Usachov, D., Vilkov, O., Grüneis, A., Haberer, D., Fedorov, A., Adamchuk, V. K., Preobrajenski, A. B., Dudin, P., Barinov, A., Oehzelt, M., Laubschat, C. & Vyalikh, D. V. (2011). Nano Lett. 11, 5401-5407.]; Mokhtar Mohamed et al., 2018[Mokhtar Mohamed, M., Mousa, M. A., Khairy, M. & Am, A. A. (2018). ACS Omega, 3, 1801-1814.]; Xu et al., 2018[Xu, H., Ma, L. & Jin, Z. (2018). J. Energy Chem. 27, 146-160.]), mainly three different types of nitro­gen bonding to carbon atoms are reported in the N-doped graphene: pyridinic, pyrrolic and graphitic bonding. Based on this, the enhanced shoulder of the C 1s peak could be deconvoluted into three peaks at binding energy 286.7 eV (CB), 288.0 eV (CC) and 289.2 eV (CD), and they might be coupled with NA, NB and NC, respectively.

Since only the exposed sample has nitro­gen-related species during the pressurization, those components can be thought of as being produced by the photo-induced effect. This result is quite encouraging because it can be a new method to synthesize N-doped graphene. Further research will be conducted for more detailed information.

5. Summary

The AP-XPS beamline offers a platform for operando experiments at UHV and ambient-pressure conditions. With the combination of a soft X-ray (100–2000 eV) and Cr X-ray source, two photon sources allows the electronic structure of matter to be studied from UHV to elevated pressures up to 25 mbar both at the surface and in the bulk. Beamline performance tests demonstrated that the beamline meets the design parameters very well. The estimated photon resolving power through the N K-edge is more than 10000, and the photon flux measured using an IRD photodiode is ∼1013 photon s−1. AP-XPS measurements of the Si and Au sample under elevated pressure also meet the expected performance while AP-XPS spectra of the graphene/Ge system under nitro­gen pressure demonstrated the presence of the photo-induced effect. The beamline and AP-XPS end-station are currently open to public users at the 3.0 GeV ring at PAL.

Funding information

Financial support was provided by the National Research Foundation of Korea (2015R1A5A1009962, NRF-2019R1A2C2008052) and the GIST Research Institute Grant funded by the Gwangju Institute of Science and Technology 2019. Experiments at PLS-II were supported in part by MSICT and POSTECH, and NRF-2018R1D1A1B07048177, NRF-2016R1D1A1B03936348, NRF-2019M3D1A1079309. Building of the APXPS end-station was funded by MSICT-KBSI-C39119.

References

First citationArtiglia, L., Orlando, F., Roy, K., Kopelent, R., Safonova, O., Nachtegaal, M., Huthwelker, T. & van Bokhoven, J. A. (2017). J. Phys. Chem. Lett. 8, 102–108.  Web of Science CrossRef CAS PubMed Google Scholar
First citationBluhm, H. (2010). J. Electron Spectrosc. Relat. Phenom. 177, 71–84.  Web of Science CrossRef CAS Google Scholar
First citationCai, J., Dong, Q., Han, Y., Mao, B. H., Zhang, H., Karlsson, P. G., Åhlund, J., Tai, R. Z., Yu, Y. & Liu, Z. (2019). Nucl. Sci. Tech. 30, 81.  Web of Science CrossRef Google Scholar
First citationDiulus, J. T., Frederick, R. T., Li, M., Hutchison, D. C., Olsen, M. R., Lyubinetsky, I., Árnadóttir, L., Garfunkel, E. L., Nyman, M., Ogasawara, H. & Herman, G. S. (2019). ACS Appl. Mater. Interfaces, 11, 2526–2534.  Web of Science CrossRef CAS PubMed Google Scholar
First citationDonald, A. M. (2003). Nat. Mater. 2, 511–516.  Web of Science CrossRef PubMed CAS Google Scholar
First citationErtl, G. (2015). Nobel Lecture: Reactions at Surfaces: From Atoms to Complexity, doi: 10.1002/anie.200800480.  Google Scholar
First citationFerrer, S., Ackermann, M. D. & Lundgren, E. (2007). MRS Bull. 32, 1010–1014.  Web of Science CrossRef CAS Google Scholar
First citationForsberg, J., Duda, L. C., Olsson, A., Schmitt, T., Andersson, J., Nordgren, J., Hedberg, J., Leygraf, C., Aastrup, T., Wallinder, D. & Guo, J. H. (2007). Rev. Sci. Instrum. 78, 083110.  Web of Science CrossRef PubMed Google Scholar
First citationGullikson, E. M., Korde, R., Canfield, L. R. & Vest, R. E. (1996). J. Electron Spectrosc. Relat. Phenom. 80, 313–316.  CrossRef CAS Web of Science Google Scholar
First citationHansen, T. W., Wagner, J. B., Hansen, P. L., Dahl, S., Topsøe, H. & Jacobsen, C. J. H. (2001). Science, 294, 1508–1510.  Web of Science CrossRef PubMed CAS Google Scholar
First citationHendriksen, B. L. M. & Frenken, J. W. M. (2002). Phys. Rev. Lett. 89, 046101.  Web of Science CrossRef PubMed Google Scholar
First citationHu, J., Xiao, X. D., Ogletree, D. F. & Salmeron, M. (1995). Science, 268, 267–269.  CrossRef PubMed CAS Web of Science Google Scholar
First citationHüfner, S. (2003). Photoelectron Spectroscopy: Principles and Applications, 3rd rev. ed. Berlin, Heidelberg: Springer.  Google Scholar
First citationHwang, I., Huang, J. Y., Kim, M., Lee, B. J., Kim, C., Choi, J. Y., Kim, M. H., Lee, H. S., Moon, D., Lee, E. H., Kim, D. E., Nam, S. H., Shin, S. & Cho, M. (2014). Rev. Sci. Instrum. 85, 055113.  Web of Science CrossRef PubMed Google Scholar
First citationJoyner, R. W., Roberts, M. W. & Yates, K. (1979). Surf. Sci. 87, 501–509.  CrossRef CAS Web of Science Google Scholar
First citationKim, J., Park, W. H., Doh, W. H., Lee, S. W., Noh, M. C., Gallet, J. J., Bournel, F., Kondoh, H., Mase, K., Jung, Y., Mun, B. S. & Park, J. Y. (2018). Sci. Adv. 4, eaat3151.  Web of Science CrossRef PubMed Google Scholar
First citationKim, K. J., Yang, S., Park, Y., Lee, M., Kim, B. & Lee, H. (2013). J. Phys. Chem. C, 117, 2129–2134.  Web of Science CrossRef CAS Google Scholar
First citationLee, M. K. & Shin, H. J. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 508–511.  Web of Science CrossRef CAS Google Scholar
First citationMokhtar Mohamed, M., Mousa, M. A., Khairy, M. & Am, A. A. (2018). ACS Omega, 3, 1801–1814.  Web of Science CrossRef CAS PubMed Google Scholar
First citationOgletree, D. F., Bluhm, H., Lebedev, G., Fadley, C. S., Hussain, Z. & Salmeron, M. (2002). Rev. Sci. Instrum. 73, 3872–3877.  Web of Science CrossRef CAS Google Scholar
First citationPetersen, H. & Baumgärtel, H. (1980). Nucl. Instrum. Methods, 172, 191–193.  CrossRef Web of Science Google Scholar
First citationRuppender, H. J., Grunze, M., Kong, C. W. & Wilmers, M. (1990). Surf. Interface Anal. 15, 245–253.  CrossRef CAS Web of Science Google Scholar
First citationSalmeron, M. & Schlögl, R. (2008). Surf. Sci. Rep. 63, 169–199.  Web of Science CrossRef CAS Google Scholar
First citationShao, Y., Zhang, S., Engelhard, M. H., Li, G., Shao, G., Wang, Y., Liu, J., Aksay, I. A. & Lin, Y. (2010). J. Mater. Chem. 20, 7491–7496.  Web of Science CrossRef CAS Google Scholar
First citationShavorskiy, A., Karslioglu, O., Zegkinoglou, I. & Bluhm, H. (2014). Synchrotron Radiat. News, 27(2), 14–23.  CrossRef Google Scholar
First citationShen, Y. R. (1989). Nature, 337, 519–525.  CrossRef CAS Web of Science Google Scholar
First citationShin, H.-J., Chung, Y. & Kim, B. (1999). J. Electron Spectrosc. Relat. Phenom. 101–103, 985–989.  Web of Science CrossRef CAS Google Scholar
First citationShin, S., Kwon, S., Kim, D. T., Kim, D. E., Kim, M., Kim, S. H., Kim, S. C., Kim, J., Kim, C., Park, B., Park, S. S., Park, S. J., Park, E., Son, Y., Yoon, J., Lee, B., Lee, E., Lee, J. W., Lee, H. S., Joo, Y., Choi, J., Ha, T., Hwang, W., Hwang, I., Lee, J. Y., Oh, B., Lee, C. H., Lee, H. S., Kim, J. Y., Hwang, J. Y., Nam, S. H. & Cho, M. (2013). J. Instrum. 8, P01019.  Google Scholar
First citationSiegbahn, H. & Siegbahn, K. (1973). J. Electron Spectrosc. Relat. Phenom. 2, 319–325.  CrossRef CAS Google Scholar
First citationSiegbahn, K. (1969). ESCA Applied to Free Molecules. Amsterdam: North Holland.  Google Scholar
First citationSoler, L., Casanovas, A., Ryan, J., Angurell, I., Escudero, C., Pérez-Dieste, V. & Llorca, J. (2019). ACS Catal. 9, 3641–3647.  Web of Science CrossRef CAS Google Scholar
First citationSomorjai, G. A. (1978). Science, 201, 489–497.  CrossRef PubMed CAS Web of Science Google Scholar
First citationStarr, D. E., Liu, Z., Hävecker, M., Knop-Gericke, A. & Bluhm, H. (2013). Chem. Soc. Rev. 42, 5833–5857.  Web of Science CrossRef CAS PubMed Google Scholar
First citationTesch, M. F., Bonke, S. A., Jones, T. E., Shaker, M. N., Xiao, J., Skorupska, K., Mom, R., Melder, J., Kurz, P., Knop-Gericke, A., Schlögl, R., Hocking, R. K. & Simonov, A. N. (2019). Angew. Chem. Int. Ed. 58, 3426–3432.  Web of Science CrossRef CAS Google Scholar
First citationTimm, R., Head, A. R., Yngman, S., Knutsson, J. V., Hjort, M., McKibbin, S. R., Troian, A., Persson, O., Urpelainen, S., Knudsen, J., Schnadt, J. & Mikkelsen, A. (2018). Nat. Commun. 9, 1412.  Web of Science CrossRef PubMed Google Scholar
First citationUsachov, D., Vilkov, O., Grüneis, A., Haberer, D., Fedorov, A., Adamchuk, V. K., Preobrajenski, A. B., Dudin, P., Barinov, A., Oehzelt, M., Laubschat, C. & Vyalikh, D. V. (2011). Nano Lett. 11, 5401–5407.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWatanabe, M., Toyoshima, A., Azuma, Y., Hayaishi, T., Yan, Y. & Yagishita, A. (2004). Proc. SPIE, 3150, 58–66.  CrossRef Google Scholar
First citationXu, H., Ma, L. & Jin, Z. (2018). J. Energy Chem. 27, 146–160.  Web of Science CrossRef Google Scholar
First citationYang, S. J., Choi, S., Odongo Ngome, F. O., Kim, K. J., Choi, S. Y. & Kim, C. J. (2019). Nano Lett. 19, 3590–3596.  Web of Science CrossRef CAS PubMed Google Scholar
First citationYu, L., Takagi, Y., Nakamura, T., Sakata, T., Uruga, T., Tada, M., Iwasawa, Y., Masaoka, S. & Yokoyama, T. (2019). J. Phys. Chem. C, 123, 603–611.  Web of Science CrossRef CAS Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds