view article

Figure 20
Closed loop control of the Veritas monochromator mirror with histogram and Fourier analysis of incremental angular encoder signals. The signal is sampled for 10 min with 4.5 ms interval. The Gaussian FWHM is 170 nrad and in the spectrum vibrations peaks at 20 Hz originate from the floor while the peaks above 40 Hz originate from pumps and ventilation.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds