Figure 21
Closed loop control of the Veritas monochromator mirror with histogram and Fourier analysis of incremental angular encoder signals when the cooling water and guard vacuum pump is turned off. The signal is sampled for 10 min with 4.5 ms interval. The Gaussian FWHM is 58 nrad and in the spectrum vibrations peaks at 20 Hz originate from the floor while the peaks above 40 Hz originate from pumps and ventilation.