Figure 6
Effect of the non-linear intensity behaviour of the ePix10k detector in low gain before and after applying the corrections described here and in more detail by van Driel et al. (2015b ) on an ePix10k dataset containing diffuse liquid scattering from a water sample as seen in th einsert of Fig. 5 . The images at three different intensities were evaluated in relation to the image around reference intensity ic. The resulting deviation [(Ix/I0x)/(IIc/I0Ic)-1] is shown for the three reference intensities x = a,b,c for the uncorrected data as well as data corrected with a polynomial of order G = 1,2,3 showing the need for at least a second-order polynomial to successfully correct most of the non-linear detector dependency of the ePix10k. |