beamlines\(\def\hfill{\hskip 5em}\def\hfil{\hskip 3em}\def\eqno#1{\hfil {#1}}\)

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775

ID30A-3 (MASSIF-3) – a beamline for macromolecular crystallography at the ESRF with a small intense beam

CROSSMARK_Color_square_no_text.svg

aEuropean Synchrotron Radiation Facility, 71 Avenue des Martyrs, 38000 Grenoble, France, bInstitut de Biosciences et Biotechnologies de Grenoble (BIG) – Laboratoire Chimie et Biologie des Métaux (LCBM), Université Grenoble Alpes, CNRS, CEA, 38000 Grenoble, France, cInstitut de Biologie Structurale (IBS), Université Grenoble Alpes, CEA, CNRS, Grenoble, France, and dEMBL Outstation Grenoble, 71 Avenue des Martyrs, 38000 Grenoble, France
*Correspondence e-mail: vonstett@esrf.fr, muellerd@esrf.fr

Edited by I. Schlichting, Max Planck Institute for Medical Research, Germany (Received 22 November 2019; accepted 20 March 2020; online 29 April 2020)

ID30A-3 (or MASSIF-3) is a mini-focus (beam size 18 µm × 14 µm) highly intense (2.0 × 1013 photons s−1), fixed-energy (12.81 keV) beamline for macromolecular crystallography (MX) experiments at the European Synchrotron Radiation Facility (ESRF). MASSIF-3 is one of two fixed-energy beamlines sited on the first branch of the canted undulator setup on the ESRF ID30 port and is equipped with a MD2 micro-diffractometer, a Flex HCD sample changer, and an Eiger X 4M fast hybrid photon-counting detector. MASSIF-3 is recommended for collecting diffraction data from single small crystals (≤15 µm in one dimension) or for experiments using serial methods. The end-station has been in full user operation since December 2014, and here its current characteristics and capabilities are described.

1. Introduction

In 2008, the ESRF launched Phase I of its upgrade program, an integral part of which was the replacement and modernization, as part of the `Upgrade Program Beamline 10' (UPBL10; Theveneau et al., 2013[Theveneau, P., Baker, R., Barrett, R., Beteva, A., Bowler, M. W., Carpentier, P., Caserotto, H., Sanctis, D., Dobias, F., Flot, D., Guijarro, M., Giraud, T., Lentini, M., Leonard, G. A., Mattenet, M., McCarthy, A. A., McSweeney, S. M., Morawe, C., Nanao, M., Nurizzo, D., Ohlsson, S., Pernot, P., Popov, A. N., Round, A., Royant, A., Schmid, W., Snigirev, A., Surr, J. & Mueller-Dieckmann, C. (2013). J. Phys. Conf. Ser. 425, 012001.]) project, of the ageing ID14 complex with a new suite of end-stations for structural biology. The experience gained from the operation of ID14 (e.g. diamond beam splitters; Wakatsuki et al., 1998[Wakatsuki, S., Belrhali, H., Mitchell, E. P., Burmeister, W. P., McSweeney, S. M., Kahn, R., Bourgeois, D., Yao, M., Tomizaki, T. & Theveneau, P. (1998). J. Synchrotron Rad. 5, 215-221.]) and of ID23 (canted straight section with two independent X-ray sources; Nurizzo et al., 2006[Nurizzo, D., Mairs, T., Guijarro, M., Rey, V., Meyer, J., Fajardo, P., Chavanne, J., Biasci, J.-C., McSweeney, S. & Mitchell, E. (2006). J. Synchrotron Rad. 13, 227-238.]; Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]) helped to design state-of-the-art end-stations. The UPBL10 project produced three beamlines for macromolecular crystallography (MX) in a canted configuration [ID30A-1 (MASSIF-1; Bowler et al., 2015[Bowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540-1547.]) and ID30A-3 (MASSIF-3) on the ID30A branch, and ID30B (McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.]) on the other branch] as well as a beamline, BM29 (Pernot et al., 2013[Pernot, P., Round, A., Barrett, R., De Maria Antolinos, A., Gobbo, A., Gordon, E., Huet, J., Kieffer, J., Lentini, M., Mattenet, M., Morawe, C., Mueller-Dieckmann, C., Ohlsson, S., Schmid, W., Surr, J., Theveneau, P., Zerrad, L. & McSweeney, S. (2013). J. Synchrotron Rad. 20, 660-664.]), dedicated to BioSAXS. Locating UPBL10 close to ID29 (de Sanctis et al., 2012[Sanctis, D. de, Beteva, A., Caserotto, H., Dobias, F., Gabadinho, J., Giraud, T., Gobbo, A., Guijarro, M., Lentini, M., Lavault, B., Mairs, T., McSweeney, S., Petitdemange, S., Rey-Bakaikoa, V., Surr, J., Theveneau, P., Leonard, G. A. & Mueller-Dieckmann, C. (2012). J. Synchrotron Rad. 19, 455-461.]) created an `MX Village', facilitating the common usage of ancillary technique laboratories and sharing services including the icOS in crystallo optical spectroscopy facility (von Stetten et al. 2015[Stetten, D. von, Giraud, T., Carpentier, P., Sever, F., Terrien, M., Dobias, F., Juers, D. H., Flot, D., Mueller-Dieckmann, C., Leonard, G. A., de Sanctis, D. & Royant, A. (2015). Acta Cryst. D71, 15-26.]), a laboratory for the high-pressure cryo-cooling of biological macromolecules (HPMX; Lafumat et al., 2016[Lafumat, B., Mueller-Dieckmann, C., Leonard, G., Colloc'h, N., Prangé, T., Giraud, T., Dobias, F., Royant, A., van der Linden, P. & Carpentier, P. (2016). J. Appl. Cryst. 49, 1478-1487.]; van der Linden et al., 2014[Linden, P. van der, Dobias, F., Vitoux, H., Kapp, U., Jacobs, J., Mc Sweeney, S., Mueller-Dieckmann, C. & Carpentier, P. (2014). J. Appl. Cryst. 47, 584-592.]), sample storage facilities both at 20°C and at cryogenic temperature, a sample preparation laboratory, and a data processing and backup computing area.

Conceived when CCD-based detectors were the staple device employed on synchrotron-based MX facilities, the original idea for the ID30A (MASSIF) beamlines was to create a hub of highly automated fixed-energy beamlines for the rapid characterization and sorting of samples. The best samples would then be redirected to the most appropriate ESRF beamlines [i.e. ID23-1 (Nurizzo et al., 2006[Nurizzo, D., Mairs, T., Guijarro, M., Rey, V., Meyer, J., Fajardo, P., Chavanne, J., Biasci, J.-C., McSweeney, S. & Mitchell, E. (2006). J. Synchrotron Rad. 13, 227-238.]), ID23-2 (Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]), ID29 (de Sanctis et al., 2012[Sanctis, D. de, Beteva, A., Caserotto, H., Dobias, F., Gabadinho, J., Giraud, T., Gobbo, A., Guijarro, M., Lentini, M., Lavault, B., Mairs, T., McSweeney, S., Petitdemange, S., Rey-Bakaikoa, V., Surr, J., Theveneau, P., Leonard, G. A. & Mueller-Dieckmann, C. (2012). J. Synchrotron Rad. 19, 455-461.]), ID30B (McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.])] for the much more time-consuming step of the optimized collection of full diffraction data sets. However, the advent of fast hybrid photon-counting detectors and shutterless data collection (Förster et al., 2019[Förster, A., Brandstetter, S. & Schulze-Briese, C. (2019). Philos. Trans. R. Soc. A, 377, 20180241.]) greatly reduced the time needed for the collection of high-quality diffraction data sets and therefore this concept was modified. MASSIF-1 is now a beamline dedicated to completely automatic MX experiments (Bowler et al., 2015[Bowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540-1547.]). MASSIF-3, described here, evolved into a mini-beam end-station equipped with an MD2 micro-diffractometer (Perrakis et al., 1999[Perrakis, A., Cipriani, F., Castagna, J.-C., Claustre, L., Burghammer, M., Riekel, C. & Cusack, S. (1999). Acta Cryst. D55, 1765-1770.]; Cipriani et al., 2007[Cipriani, F., Felisaz, F., Lavault, B., Brockhauser, S., Ravelli, R., Launer, L., Leonard, G. & Renier, M. (2007). AIP Conf. Proc. 879, 1928-1931.]), a Flex-HCD sample changer (McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.]) and an Eiger X 4M hybrid photon-counting detector (Casanas et al., 2016[Casanas, A., Warshamanage, R., Finke, A. D., Panepucci, E., Olieric, V., Nöll, A., Tampé, R., Brandstetter, S., Förster, A., Mueller, M., Schulze-Briese, C., Bunk, O. & Wang, M. (2016). Acta Cryst. D72, 1036-1048.]) which can operate with frame rates of up to 750 Hz. In addition to being an excellent vehicle for standard MX experiments based around ESRF-developed automation, this setup has already proven very useful in the application of new diffraction data collection methods [i.e. MeshAndCollect (Zander et al., 2015[Zander, U., Bourenkov, G., Popov, A. N., de Sanctis, D., Svensson, O., McCarthy, A. A., Round, E., Gordeliy, V., Mueller-Dieckmann, C. & Leonard, G. A. (2015). Acta Cryst. D71, 2328-2343.])] and room-temperature synchrotron serial crystallography (SSX) data collection using sample delivery systems such as the high-viscosity extruder (HVE) system developed by Botha et al. (2015[Botha, S., Nass, K., Barends, T. R. M., Kabsch, W., Latz, B., Dworkowski, F., Foucar, L., Panepucci, E., Wang, M., Shoeman, R. L., Schlichting, I. & Doak, R. B. (2015). Acta Cryst. D71, 387-397.]).

2. Beamline overview

2.1. X-ray source and optics hutches

Before the extremely brilliant source upgrade (EBS; https://www.esrf.eu/about/upgrade) of the ESRF storage ring (see below), the X-ray source for the ID30A branch comprised two 1.4 m undulators with a periodicity of 21.2 mm (U21.2) operated in air with a minimum gap of 11 mm and optimized to emit their highest brilliance at an energy of 12.9 keV. A schematic overview of MASSIF-3 is shown in Fig. 1[link]. There are two independent optical hutches (OH), OH1 that is shared with ID30B, and OH2 that contains most of the optical elements for the MASSIF beamlines. The canting angle between the sources for ID30A and ID30B is ±2.2 mrad, providing sufficient space between the two beams to install all optical equipment. Table 1[link] summarizes the current (and future, see below) beamline characteristics.

Table 1
Beamline parameters of ID30A-3 (MASSIF-3)

  2014–2019 From 2020
Source 2 × 1.4 m U21.2 in-air undulators (gap setting: 11 mm) 1 × 2.3 m U21.2 in-air undulator
Source size (FWHM, H × V) 412 µm × 6.2 µm 28 µm × 6.1 µm
Divergence (H × V) 11.5 µrad × 5.1 µrad 7.2 µrad × 5.1 µrad
Vertical focusing 11 white-beam Be CRL 11 white-beam Be CRL
Horizontal focusing Multi-layer mirror (B4C/[Ru/B4C]70/Cr) (ΔE/E = 2.3%) Be CRL
Monochromator Si(111) Si(111)
Photon energy 12.812 keV 12.812 keV
Beam size at the sample (FWHM, H × V) 18 µm × 14 µm 15 µm diameter
Beam flux 2.0 × 1013 photons s−1 4.0 × 1013 photons s−1§
Diffractometer MD2S MD2S
Sample changer FlexHCD FlexHCD
Detector Dectris Eiger X 4M Dectris Eiger X 4M
†Measured by scanning a tungsten wire through the beam.
‡Measured with 100% transmission and 200 mA filling mode in the storage ring.
§Estimated based on the new source characteristics.
¶Refurbished.
[Figure 1]
Figure 1
Schematic layout of the MASSIF-3 beamline. Major components are shown with their distance relative to the source. The monochromator/beamsplitter for MASSIF-2 is shown only for consistency but is currently not in place.
2.1.1. OH1 configuration

OH1 contains a set of ESRF standard high-power primary slits (Marion & Zhang, 2003[Marion, P. & Zhang, L. (2004). AIP Conf. Proc. 705, 320-323.]), an ESRF high-vacuum white-beam viewer (wbv-a) and a photon absorber. The white-beam viewer is equipped with a Basler GigE camera that displays the fluorescence of a water-cooled CVD (chemical vapour deposition) diamond that can be inserted into the X-ray beam using a pneumatic actuator when this is required. The CVD diamond also acts as the scattering element for an integrated diode such that the intensity of the beam can be measured, for example, during alignment of the primary slits.

2.1.2. OH2 configuration

The MASSIF-specific OH2 houses most optical elements for the MASSIF beamlines as well as a long, lead-shielded vacuum tube which allows the white beam of the ID30B branch pass through the hutch. This setup ensures that ID30B is not affected in case of an intervention in OH2.

The first element in OH2 is a set of water-cooled cylindrical white-beam compound refractive lenses (CRLs, Snigirev et al., 1996[Snigirev, A., Kohn, V., Snigireva, I. & Lengeler, B. (1996). Nature, 384, 49-51.]) made from beryllium. This device focuses the white beam vertically, producing a vertical beam size of 15 µm at the sample position (EH2, see below) of MASSIF-3. A synthetic asymmetric Laue [110]-cut diamond, 250 µm thick and mounted on a water-cooled copper support, is then used as a monochromator for MASSIF-1, deviating X-rays with an energy of 12.834 keV into its experimental hutch (EH1; Bowler et al., 2015[Bowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540-1547.]). The advantage of the Laue [110] geometry is that the Bragg planes diffracting X-rays are almost 90° relative to the diamond surface, which maximizes the transmission of the white beam (deprived of its 12.834 keV component). The transmitted white beam then impinges on a liquid-nitro­gen-cooled single Si(111) crystal with a bandwidth of 1.5 × 10−4 used as a monochromator for MASSIF-3, which delivers a beam at a photon energy of 12.812 keV into its experimental hutch (EH2). This energy on the high-energy remote side of the selenium edge was chosen in order to be able to use the anomalous scattering contribution from selenium for de novo phasing experiments. Immediately downstream of the monochromator, an energy calibration device has been installed. This contains a wheel with windows for up to six different pure metal foils and, coupled with a retractable direct diode, read by a Keithley picoammeter, allows for easy, automated energy calibration by recording of X-ray absorption spectra. Currently, the scanning of the LII absorption edge of an iridium foil (12.824 keV, 100 µm foil) is used for the (re)calibration of the MASSIF-3 monochromator to its nominal energy (12.812 keV). A second white-beam viewer (wbv-f), identical to that located in OH1, is installed before the Si(111) monochromator to facilitate the alignment of the beamline and for X-ray beam diagnostic purposes. A monochromatic beam viewer (mbv-31), consisting of a cerium-doped YAG screen that can be pneumatically inserted into the X-ray beam, has been installed after the monochromator. Here also, a Basler GigE camera allows to visualize the beam footprint on the YAG, which additionally serves as a scattering element for a diode (read by a Keithley picoammeter) to measure relative beam intensities.

2.2. EH2 configuration

The experimental hutch of MASSIF-3 contains both the beamline's horizontal focusing element and its sample position and sample environment. Horizontal focusing of the X-ray beam is achieved using a bent multilayer mirror (B4C/[Ru/B4C]70/Cr, manufactured in-house at ESRF) located 1.0 m upstream of the sample position. Combined with the vertical focusing provided by the white-beam CRLs (see above), this results in a focal spot size of 18 µm × 14 µm [FWHM as measured by scanning a tungsten wire across the beam; Fig. 2(c)][link] at the sample position. Potential contamination of diffraction intensities by X-ray photons of higher harmonic reflections from the Si(111) monochromator (mostly the third-harmonic reflections with photon energy ∼38.4 keV) are intrinsically suppressed or rejected due to the combination of three factors: (i) the X-ray detector's quantum efficiency drastically decreases above 20 keV, (ii) the focusing distance of the CRLs depends on the photon energy such that higher energy photons are focused far behind the sample position, and (iii) the reflectivity of the multi-layer mirror drastically decreases at higher energies (Bigault et al., 2003[Bigault, T., Ziegler, E., Morawe, C., Hustache, R., Massonnat, J.-Y. & Rostaing, G. (2003). Proc. SPIE, 5195, 12-20.]).

[Figure 2]
Figure 2
Experimental setup of MASSIF-3. (a) Photograph of the sample environment and (b) setup of the micro-spectrophotometer with the X-ray beam indicated as a red arrow and the optical pathway in transparent orange. (c) 3D plot of the X-ray beam profile measured with a 5 µm pinhole (abscissae in micrometres and ordinate in arbitrary units). (d) Image of the X-ray beam at the sample position as viewed through the on-axis viewer using the built-in scintillator (at 0.06% transmission and a storage ring current of 180 mA).

Upstream of the horizontally focusing mirror, a high-vacuum attenuator box containing a total of eight absorbing foils [two pyrocarbon (1 and 2 mm in thickness) and six Al (0.1, 0.2, 0.35, 0.5, 1, and 1.5 mm thickness)] is installed. This allows to accurately adjust the photon flux at the sample in small steps. Downstream of the mirror, a pair of beam-defining secondary slits (ss2) (JJ-XRAY, Lyngby, Denmark) has been installed to allow for the optimal alignment of the mirror and also to remove parasitic X-rays scattered from the mirror. This pair of slits is mounted in a `slitbox' which also contains a scattering foil diode (I0), read by a Keithley picoammeter which allows constant monitoring of the X-ray beam intensity, a rotating millisecond fast shutter controlled by ESRF-developed IcePAP electronics (Janvier et al., 2013[Janvier, N., Clement, J. M. & Fajardo, P. (2013). Proceedings of the 14th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2013), 6-11 October 2013, San Francisco, CA, USA, pp. 766-769. TUPPC08.]), and a second scattering foil diode (I1) also read through a Keithley picoammeter which allows monitoring of the X-ray beam intensity after the fast shutter. As at the other ESRF Structural Biology Group MX beamlines, the I0 reading is correlated with photon flux using a calibrated diode periodically placed at the sample position, thus allowing the constant measurement of photon flux impinging of the sample.

2.3. Sample environment

The MASSIF-3 sample environment [Fig. 2(a)[link]] comprises a MD2 micro-diffractometer (Arinax, Voreppe, France; permanent loan from EMBL Grenoble Outstation), an Eiger X 4M detector (Dectris, Baden-Daettwil, Switzerland; Casanas et al., 2016[Casanas, A., Warshamanage, R., Finke, A. D., Panepucci, E., Olieric, V., Nöll, A., Tampé, R., Brandstetter, S., Förster, A., Mueller, M., Schulze-Briese, C., Bunk, O. & Wang, M. (2016). Acta Cryst. D72, 1036-1048.]; Förster et al., 2019[Förster, A., Brandstetter, S. & Schulze-Briese, C. (2019). Philos. Trans. R. Soc. A, 377, 20180241.]) mounted on a fast translation table, and a FlexHCD robotic sample changer (McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.]). Samples on the rotation axis of the MD2 are kept at cryogenic temperature using a 700 series cryo-stream (Oxford Cryostreams, Oxford, UK) equipped with an annealing system (Giraud et al., 2009[Giraud, T., Dobias, F., Gabadinho, J., Rey-Bakaikoa, V., Nurizzo, D., Leonard, G. A. & Mueller-Dieckmann, C. (2009). J. Appl. Cryst. 42, 125-128.]). The entire experimental setup [Fig. 2(a)[link]], including the slitbox described above, is installed on a large granite table to allow for the alignment of all elements relative to the X-ray beam using three vertical (tz1, tz2 and tz3) and two horizontal (tyf and tyb) motors. These motors are operated in a closed-loop feedback mode and can be moved independently, or conjointly as pseudo motors for controlling the table height (thgt), translation (ttrans), roll (txrot), pitch (tyrot), and yaw (tzrot). The MD2 micro-diffractometer is, in contrast to all other MX beamlines, not permanently equipped with a Mini-Kappa goniometer head (MK3, Brockhauser et al., 2013[Brockhauser, S., Ravelli, R. B. G. & McCarthy, A. A. (2013). Acta Cryst. D69, 1241-1251.]) to keep the sphere of confusion to less than 1 µm. However, the mounting of an MK3 device is available upon request.

The maximum speed of the MD2 rotation axis is currently 720° s−1, which allows, in combination with the fast Eiger X 4M detector, the recording of complete datasets in less than a second (Table 2[link]). Additional built-in accessories of the MD2 include beam-cleaning apertures and a YAG screen for automatic alignment of the rotation axis and the X-ray beam [Fig. 2(d)[link]]. The MD2 is equipped with a standard G-shaped beamstop/capillary setup, but optionally a motorized beamstop with a maximum distance of 120 mm from the sample can be mounted, leaving more space for experiments that require installation of specific setups that are significantly bulkier than standard SPINE pins. The FlexHCD robotic sample changer allows to use SPINE sample holders stored either in SC3 pucks (Cipriani et al., 2006[Cipriani, F., Felisaz, F., Launer, L., Aksoy, J.-S., Caserotto, H., Cusack, S., Dallery, M., di-Chiaro, F., Guijarro, M., Huet, J., Larsen, S., Lentini, M., McCarthy, J., McSweeney, S., Ravelli, R., Renier, M., Taffut, C., Thompson, A., Leonard, G. A. & Walsh, M. A. (2006). Acta Cryst. D62, 1251-1259.]) or in UniPuck (https://smb.slac.stanford.edu/robosync/Universal_Puck) containers. Each system needs its own specific FlexHCD gripper to transfer sample holders, and the exchange of grippers can be performed automatically within less than a minute via the beamline control GUI MXCuBE (https://MXCuBE.github.io/MXCuBE; Gabadinho et al., 2010[Gabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700-707.]; Oscarsson et al., 2019[Oscarsson, M., Beteva, A., Flot, D., Gordon, E., Guijarro, M., Leonard, G., McSweeney, S., Monaco, S., Mueller-Dieckmann, C., Nanao, M., Nurizzo, D., Popov, A., von Stetten, D., Svensson, O., Rey-Bakaikoa, V., Chado, I., Chavas, L., Gadea, L., Gourhant, P., Isabet, T., Legrand, P., Savko, M., Sirigu, S., Shepard, W., Thompson, A., Mueller, U., Nan, J., Eguiraun, M., Bolmsten, F., Nardella, A., Milàn-Otero, A., Thunnissen, M., Hellmig, M., Kastner, A., Schmuckermaier, L., Gerlach, M., Feiler, C., Weiss, M. S., Bowler, M. W., Gobbo, A., Papp, G., Sinoir, J., McCarthy, A., Karpics, I., Nikolova, M., Bourenkov, G., Schneider, T., Andreu, J., Cuní, G., Juanhuix, J., Boer, R., Fogh, R., Keller, P., Flensburg, C., Paciorek, W., Vonrhein, C., Bricogne, G. & de Sanctis, D. (2019). J. Synchrotron Rad. 26, 393-405.]).

Table 2
Diffraction data collection and refinement statistics for FAE and trypsin

Numbers in parentheses are for the innermost and outermost shell, respectively.

Protein FAE Trypsin-EDNA Trypsin-fast
Data collection
Beamline MASSIF-3 MASSIF-3 MASSIF-3
Energy (keV) 12.812 12.812 12.812
Oscillation range (°) 135 140 140
Increment (°) 0.05 0.1 0.2
Images per second 500 500 746
Exposure time per image (ms) 2.0 2.0 1.34
Total collection time (s) 5.4 2.8 0.93
Angular speed (° s−1) 25 50 149
Flux transmission (%) 1 1 3
Resolution range (Å) 48.9–1.80 (48.9–9.00, 1.84–1.80) 43.8–1.43 (43.8–5.54, 1.48–1.43) 43.8–1.43 (43.8–5.55, 1.48–1.43)
Space group P212121 P212121 P212121
Unit-cell dimensions (Å) 65.5, 108.4, 112.9 54.3, 58.0, 66.9 54.3, 57.9, 66.9
Unique reflections 73998 (678, 4353) 39219 (781, 3669) 39193 (778, 3704)
Rmeas (%) 11.7 (5.6, 97.9) 9.7 (4.3, 156.4) 10.0 (4.4, 171.9)
Rp.i.m. (%) 5.0 (2.4, 41.6) 4.2 (1.9, 68.4) 4.3 (2.0, 75.1)
I/σ(I)〉 11.0 (32.2, 1.7) 11.4 (31.6, 0.9) 11.6 (32.3, 0.9)
σano 1.113 (2.402, 0.732)    
CC1/2 (%) 99.6 (99.8, 75.0) 99.8 (99.8, 47.0) 99.8 (99.8, 43.0)
Completeness (%) 98.5 (97.0, 98.4) 99.0 (99.5, 95.9) 99.1 (99.1, 97.2)
Multiplicity 5.2 (4.3, 5.3) 5.3 (4.8, 4.8) 5.3 (4.7, 4.9)
Wilson B-factor (Å2) 13.7 15.7 15.7
 
Phasing and refinement
Resolution range (Å) 48.9–1.80 43.8–1.43 43.8–1.43
No. of reflections 70226 37260 37232
CCweak / CCall 22.9 / 37.3    
Contrast original / inverted 0.242 / 0.864    
No. of automatically built residues
SHELXE 529    
ARP/WARP 561    
Rwork / Rfree (%) 17.5 / 20.0 17.7 / 19.8 18.5 / 20.6
No. of protein atoms 4588 1629 1629
No. of water molecules 438 232 232
No. of ions and buffer molecules 8 Cd, 1 glycerol 1 Ca, 6 SO4, 1 benzamidine 1 Ca, 6 SO4, 1 benzamidine
Mean B 19.8 10.9 11.1
R.m.s. deviations      
 Bond length (Å) 0.012 0.013 0.013
 Bond angle (°) 1.77 1.78 1.78
PDB code 6y8g   6swv

The Eiger X 4M detector provides diffraction images in HDF5 format (https://portal.hdfgroup.org/display/support), which allows the storage of many (typically 100) images in a single file together with extensive metadata, thus reducing file system clutter. Most crystallographic software programs [e.g. XDS (Kabsch, 2010[Kabsch, W. (2010). Acta Cryst. D66, 125-132.]), DIALS (Winter et al., 2018[Winter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85-97.]); CrystFEL (White et al., 2012[White, T. A., Kirian, R. A., Martin, A. V., Aquila, A., Nass, K., Barty, A. & Chapman, H. N. (2012). J. Appl. Cryst. 45, 335-341.])] are able to handle this data format, but for certain steps of automatic data processing [namely strategy calculations using EDNA/BEST (Bourenkov & Popov, 2006[Bourenkov, G. P. & Popov, A. N. (2006). Acta Cryst. D62, 58-64.]; Incardona et al., 2009[Incardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872-879.])] diffraction images are automatically converted to miniCBF format (Bernstein & Hammersley, 2006[Bernstein, H. J. & Hammersley, A. P. (2006). International Tables for Crystallography, Vol. G, pp. 37-43. International Union of Crystallography.]). The maximum resolution that can be achieved with the Eiger X 4M at the shortest sample-detector distance available on MASSIF-3 is 1.52 Å at the edge of the detector, 1.19 Å at its corners.

MASSIF-3 is currently recommended for diffraction data collection from small crystals (<15 µm) or from larger crystals where use of a relatively small beam will allow the use of helical data collections (Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]) or the identification of the best diffracting segment of the crystal for a diffraction experiment (i.e. diffraction cartography; Bowler et al., 2010[Bowler, M. W., Guijarro, M., Petitdemange, S., Baker, I., Svensson, O., Burghammer, M., Mueller-Dieckmann, C., Gordon, E. J., Flot, D., McSweeney, S. M. & Leonard, G. A. (2010). Acta Cryst. D66, 855-864.]]). As on all other ESRF MX beamlines, a HC1 humidity device (Arinax, Moirans, France) that allows the recording of diffraction data under controlled relative humidity at room temperature is available upon user request (Sanchez-Weatherby et al., 2009[Sanchez-Weatherby, J., Bowler, M. W., Huet, J., Gobbo, A., Felisaz, F., Lavault, B., Moya, R., Kadlec, J., Ravelli, R. B. G. & Cipriani, F. (2009). Acta Cryst. D65, 1237-1246.]; Russi et al., 2011[Russi, S., Juers, D. H., Sanchez-Weatherby, J., Pellegrini, E., Mossou, E., Forsyth, V. T., Huet, J., Gobbo, A., Felisaz, F., Moya, R., McSweeney, S. M., Cusack, S., Cipriani, F. & Bowler, M. W. (2011). J. Struct. Biol. 175, 236-243.]). The beamline is also well adapted for experiments requiring the on-line collection of in crystallo fluorescence and/or UV/vis absorption spectra exploiting the use of a microspectrophotometer [McGeehan et al., 2009[McGeehan, J., Ravelli, R. B. G., Murray, J. W., Owen, R. L., Cipriani, F., McSweeney, S., Weik, M. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 163-172.]; von Stetten et al., 2015[Stetten, D. von, Giraud, T., Carpentier, P., Sever, F., Terrien, M., Dobias, F., Juers, D. H., Flot, D., Mueller-Dieckmann, C., Leonard, G. A., de Sanctis, D. & Royant, A. (2015). Acta Cryst. D71, 15-26.]; Fig. 2[link](b)]. Additionally, MASSIF-3 is an excellent vehicle for serial synchrotron crystallography (SSX) experiments using a high-viscosity extruder (HVE; Botha et al., 2015[Botha, S., Nass, K., Barends, T. R. M., Kabsch, W., Latz, B., Dworkowski, F., Foucar, L., Panepucci, E., Wang, M., Shoeman, R. L., Schlichting, I. & Doak, R. B. (2015). Acta Cryst. D71, 387-397.], results to be published elsewhere) in close collaboration with the group of Ilme Schlichting from the MPI for Medical Research in Heidelberg, Germany.

2.4. Beamline control and data collection

Low-level control of all beamline components and communication with the PMAC controller of the MD2 is provided by the newly developed ESRF standard software BLISS (Beamline Instrumentation Support Software; https://gitlab.esrf.fr/bliss/bliss; Guijarro et al., 2017[Guijarro, M., Beteva, A., Coutinho, T. M., Dominguez, M. C., Homs, A., Meyer, J. M., Michel, V., Papillon, E., Perez, M. & Petitdemanage, S. (2017). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017), 8-13 October 2017, Barcelona, Spain, pp. 1060-1066. WEBPL05.]). BLISS commands are mostly accessed transparently from the beamline control GUI MXCuBE2 (Gabadinho et al., 2010[Gabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700-707.]; Oscarsson et al., 2019[Oscarsson, M., Beteva, A., Flot, D., Gordon, E., Guijarro, M., Leonard, G., McSweeney, S., Monaco, S., Mueller-Dieckmann, C., Nanao, M., Nurizzo, D., Popov, A., von Stetten, D., Svensson, O., Rey-Bakaikoa, V., Chado, I., Chavas, L., Gadea, L., Gourhant, P., Isabet, T., Legrand, P., Savko, M., Sirigu, S., Shepard, W., Thompson, A., Mueller, U., Nan, J., Eguiraun, M., Bolmsten, F., Nardella, A., Milàn-Otero, A., Thunnissen, M., Hellmig, M., Kastner, A., Schmuckermaier, L., Gerlach, M., Feiler, C., Weiss, M. S., Bowler, M. W., Gobbo, A., Papp, G., Sinoir, J., McCarthy, A., Karpics, I., Nikolova, M., Bourenkov, G., Schneider, T., Andreu, J., Cuní, G., Juanhuix, J., Boer, R., Fogh, R., Keller, P., Flensburg, C., Paciorek, W., Vonrhein, C., Bricogne, G. & de Sanctis, D. (2019). J. Synchrotron Rad. 26, 393-405.]), but, for troubleshooting, a terminal interface is also available. All motors except those of the MD2 and that of the millisecond fast shutter are controlled by ESRF-developed IcePAP electronics (Janvier et al., 2013[Janvier, N., Clement, J. M. & Fajardo, P. (2013). Proceedings of the 14th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2013), 6-11 October 2013, San Francisco, CA, USA, pp. 766-769. TUPPC08.]) and pneumatically driven devices are controlled by WAGO control modules (https://www.wago.com).

Users collect diffraction data via the MXCuBE2 GUI (Gabadinho et al., 2010[Gabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700-707.]; Oscarsson et al., 2019[Oscarsson, M., Beteva, A., Flot, D., Gordon, E., Guijarro, M., Leonard, G., McSweeney, S., Monaco, S., Mueller-Dieckmann, C., Nanao, M., Nurizzo, D., Popov, A., von Stetten, D., Svensson, O., Rey-Bakaikoa, V., Chado, I., Chavas, L., Gadea, L., Gourhant, P., Isabet, T., Legrand, P., Savko, M., Sirigu, S., Shepard, W., Thompson, A., Mueller, U., Nan, J., Eguiraun, M., Bolmsten, F., Nardella, A., Milàn-Otero, A., Thunnissen, M., Hellmig, M., Kastner, A., Schmuckermaier, L., Gerlach, M., Feiler, C., Weiss, M. S., Bowler, M. W., Gobbo, A., Papp, G., Sinoir, J., McCarthy, A., Karpics, I., Nikolova, M., Bourenkov, G., Schneider, T., Andreu, J., Cuní, G., Juanhuix, J., Boer, R., Fogh, R., Keller, P., Flensburg, C., Paciorek, W., Vonrhein, C., Bricogne, G. & de Sanctis, D. (2019). J. Synchrotron Rad. 26, 393-405.]) which is the standard interface on all MX beamlines at the ESRF. The metadata of each data collection are stored in the ISPyB database (Delagenière et al., 2011[Delagenière, S., Brenchereau, P., Launer, L., Ashton, A. W., Leal, R., Veyrier, S., Gabadinho, J., Gordon, E. J., Jones, S. D., Levik, K. E., McSweeney, S. M., Monaco, S., Nanao, M., Spruce, D., Svensson, O., Walsh, M. A. & Leonard, G. A. (2011). Bioinformatics, 27, 3186-3192.]) and can be recalled through the web-service EXI (https://exi.esrf.fr). Specific automated workflows are available and executed from within MXCuBE for complex tasks such as fast mesh scans, X-ray centring, helical data collections (Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]), MeshAndCollect (Zander et al., 2015[Zander, U., Bourenkov, G., Popov, A. N., de Sanctis, D., Svensson, O., McCarthy, A. A., Round, E., Gordeliy, V., Mueller-Dieckmann, C. & Leonard, G. A. (2015). Acta Cryst. D71, 2328-2343.]), as well as MXpress workflows for fully automatic data collection (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]; McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.]). MASSIF-3 also features a more advanced workflow for helical data collection (Fig. 3[link]). This workflow is especially useful for cases where crystals are thin elongated needles or plates and where start and end points for the helical trajectory are difficult to assign due to the opaqueness and/or refraction (i.e. lensing effects) of frozen mother liquor. In this workflow, initial start/stop points for the trajectory are refined and precisely determined by X-ray diffraction using two orthogonal line scans at each point. A data collection strategy taking into account that the radiation damage will be distributed over a larger crystal volume (Zeldin et al., 2013[Zeldin, O. B., Gerstel, M. & Garman, E. F. (2013). J. Synchrotron Rad. 20, 49-57.]) is then calculated using characterization images collected at the midpoint of the trajectory and is subsequently executed (Fig. 3[link]).

[Figure 3]
Figure 3
Schematic representation of the procedure for optimized helical data collection. Two orthogonal line scans (green arrows) at each end of the needle-shaped crystal (blue) precisely locate the start and end positions. Then a characterization is performed in the middle of the crystal (red dot), and the data set is collected along the crystal (orange line).

As on the other MX beamlines at the ESRF, all collected datasets are automatically processed using a series of software pipelines [EDNAproc, GrenADeS (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]), XDSAPP (Sparta et al., 2016[Sparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085-1092.]), DIALS (Winter et al., 2018[Winter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85-97.]), autoPROC (Vonrhein et al., 2011[Vonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293-302.])], the results of which are stored in the ISPyB database and are accessible, along with files containing the reduced data, through the web-service EXI (https://exi.esrf.fr; Delagenière et al., 2011[Delagenière, S., Brenchereau, P., Launer, L., Ashton, A. W., Leal, R., Veyrier, S., Gabadinho, J., Gordon, E. J., Jones, S. D., Levik, K. E., McSweeney, S. M., Monaco, S., Nanao, M., Spruce, D., Svensson, O., Walsh, M. A. & Leonard, G. A. (2011). Bioinformatics, 27, 3186-3192.]).

Data backup is easily handled, using an in-house-developed GUI that allows the users to select a specific beamline, experiment number, date, and USB drive, and to automatically copy their data in parallel to the experiment.

3. Results

Opened to external users in December 2014, diffraction data collected on MASSIF-3 have, at the time of writing, contributed to 150 publications (source: ESRF-ILL Library; https://www.epn-campus.eu/1/library/the-joint-ill-esrf-library/) and have resulted in more than 150 depositions in the Protein Data Bank (PDB; wwPDB consortium, 2019[wwPDB consortium (2019). Nucl. Acid Res. 47, D520-D528.]; https://biosync.sbkb.org/stats.do?stats_sec=RGNL&stats_focus_lvl=SITE&stats_site=ESRF). Table 2[link] and Fig. 4[link] illustrate the high quality of data that can be routinely collected at the beamline.

[Figure 4]
Figure 4
Indicators for substructure solution as output from SHELXD and a section of the resulting electron density map for the FAE diffraction data set. The top panel shows CCall versus CCweak, the middle panel d′′/sig versus resolution, and the bottom panel the electron density [prepared with PyMOL v1.8 (Schrödinger, LLC) of residue Gln127 which exists in a double conformation (2FoFc) map at 1.5σ is shown in blue chicken wire, positive electron density for the minor conformation of residue 127 at 2.5σ in green, and the model in stick representation].

In a first experiment, we collected diffraction data from a Se-Met derivatized crystal of ferulic acid esterase (FAE; Prates et al., 2001[Prates, J. A., Tarbouriech, N., Charnock, S. J., Fontes, C. M., Ferreira, L. M. & Davies, G. J. (2001). Structure, 9, 1183-1190.]) with the aim of de novo structure determination using the SAD technique (f′′ = 3.7 e at 12.834 keV). Data were measured according to a strategy proposed by an EDNA characterization (Incardona et al., 2009[Incardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872-879.]) at 100 K from a crystal cryo-protected using Paratone N and were automatically processed (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]). Subsequent de novo phasing performed with hkl2map (Pape & Schneider, 2004[Pape, T. & Schneider, T. R. (2004). J. Appl. Cryst. 37, 843-844.]) and SHELXC/D/E (Usón & Sheldrick, 2018[Usón, I. & Sheldrick, G. M. (2018). Acta Cryst. D74, 106-116.]) found all 16 Se sites (8 per chain), and automatic building with ARP/WARP (Langer et al., 2008[Langer, G., Cohen, S. X., Lamzin, V. S. & Perrakis, A. (2008). Nat. Protoc. 3, 1171-1179.]) resulted in a model containing 280 and 281 residues (out of 297 residues) for the two FAE monomers in the asymmetric unit which was then further refined using Coot (Emsley et al., 2010[Emsley, P., Lohkamp, B., Scott, W. G. & Cowtan, K. (2010). Acta Cryst. D66, 486-501.]) and refmac5 (Kovalevskiy et al., 2018[Kovalevskiy, O., Nicholls, R. A., Long, F., Carlon, A. & Murshudov, G. N. (2018). Acta Cryst. D74, 215-227.]). A representative section of electron density is shown in Fig. 4[link] which illustrates OMIT difference density for a double conformation for residue Gln127 (corresponding to Gln915 in PDB access code 1gkk).

In a second experiment, bovine trypsin purchased as lyophilized powder from Sigma (reference: T9201) was crystallized as follows: A solution of 40 mg ml−1 trypsin, 10 mg ml−1 benzamidine (Sigma B-6506) and 3 mM CaCl2 was mixed in a 1:1 ratio with reservoir solution [2 M (NH4)2SO4 in 100 mM Tris-HCl pH 8.5] at room temperature. Crystals appeared after a few days and were cryo-protected using 20% (w/v) glycerol. Diffraction data were then collected at 100 K using two different strategies: the first based on a strategy provided by EDNA/BEST (500 imgages s−1, 50° s−1 omega rotation, total data collection time 2.8 s, trypsin-EDNA in Table 2[link]), the second using a manually determined strategy optimized for data collection speed (746 images s−1, 149° s−1 omega rotation, total data collection time 0.93 s, trypsin-fast in Table 2[link]). While the `slow' diffraction data set has been collected to show a standard data collection, the `fast' diffraction data set has been collected to show the potential of the beamline. As can be seen in Table 2[link], the two data sets are of comparable quality (automatic processing using EDNAproc) with both giving clear structure solutions, requiring only little additional refinement (adding waters, benzamidine and other buffer molecules, adjusting side chains, etc.), using the ESRF automatic molecular replacement pipeline (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]). The structure refinement based on the trypsin-fast dataset (PDB access code 6swv) demonstrates that on MASSIF-3 complete, high-quality diffraction datasets can be collected in less than a second (Table 2[link]). The combination of the fast read-out detector and a fast rotation goniometer (up to 720° s−1) allows collection of diffraction data at room temperature from crystals (Gotthard et al., 2019[Gotthard, G., Aumonier, S., De Sanctis, D., Leonard, G., von Stetten, D. & Royant, A. (2019). IUCrJ, 6, 665-680.]). The combination of a few dozen data wedges derived from different diffraction data sets allows the collection of time-resolved data in the millisecond range (Aumonier et al., submitted).

4. Post ESRF-EBS configuration

The ESRF-EBS upgrade, which will be completed in the autumn of 2020, has a number of implications for MASSIF-3. Notable amongst these (Table 1[link]) will be a reduction in the horizontal source size (FWHM) from 412 µm to 28 µm while the vertical source size will stay the same. The horizontal source divergence will also be reduced from 11.5 µrad to 7.2 µrad. As described above, horizontal focusing on MASSIF-3 is currently achieved using a bent multilayer mirror, a choice that was made in order to accept the full horizontal footprint (approximately 76 mm) of the X-ray beam 56 m from the source. Post-ESRF-EBS the reduced horizontal footprint will make it possible to replace the multilayer mirror by a set of horizontally focusing CRLs, further simplifying the beamline setup and operation.

The upgrade of the ESRF storage ring also requires a reduction in the length of the ID30 straight section. As a result, the current ID30A X-ray source (2 × 1.4 m U21.2 undulators) will be replaced by a single U21.2 device 2.3 m in length. This, coupled with the acceptance of the entire X-ray beam by the vertically focusing white-beam CRLs aperture situated in OH1, should result in a more than twofold increase in the available flux at the sample position (i.e. ∼4 × 1013 photons s−1). Thus, while our current aim is to maintain a MASSIF-3 focal spot size of ∼15 µm in diameter, post-EBS the end-station should become an even more attractive vehicle for MX.

5. Conclusion

ID30A-3 (MASSIF-3) is a fixed-energy (12.81 keV), highly intense (>1013 photons s−1) and highly automated mini-beam (18 µm × 14 µm) end-station dedicated to protein crystallography at the ESRF. The beamline has been in full user operation since December 2014 with, at the time of writing, data collected at the beamline contributing to more than 150 PDB depositions (https://www.rcsb.org; Berman et al., 2003[Berman, H., Henrick, K. & Nakamura, H. (2003). Nat. Struct. Mol. Biol. 10, 980.]) and more than 150 publications (https://epn-library.esrf.fr). MASSIF-3 is currently recommended for diffraction data collection from small crystals or from larger crystals where use of a relatively small beam will allow the use of helical data collections or the identification of the best diffracting segment of a crystal for a diffraction experiment. Moreover, the relatively small focal spot size is conducive to efficient, high-quality experiments exploiting multi-sample or serial methods. Post-ESRF-EBS, simplification of the beamline's optical setup coupled with an increase in the available flux at the sample position will ensure that the end station remains an extremely attractive vehicle for MX in the coming decade.

Footnotes

Present address: EMBL Hamburg Outstation, c/o DESY, Notkestrasse 85, 22607 Hamburg, Germany.

§Present address: NSLS II, Brookhaven National Laboratory, PO Box 5000, Upton, NY 11973-5000, USA.

Acknowledgements

The authors gratefully acknowledge the ESRF past and present management for their support. We also would like to thank the Joint Structural Biology Group (JSBG) members and the ESRF support groups (TID, BIG, safety group, alignment group, computing and network) for invaluable help in this project. We thank Matthew Bowler (EMBL Grenoble) for helping to define the scientific case for the beamline and for his involvement in the initial stages of the project. We thank Ulrike Kapp (ESRF) for preparing FAE and trypsin crystals. ESRF beam time, granted through the in-house research programme, is acknowledged.et al., 2014[Casanas, A., Warshamanage, R., Finke, A. D., Panepucci, E., Olieric, V., Nöll, A., Tampé, R., Brandstetter, S., Förster, A., Mueller, M., Schulze-Briese, C., Bunk, O. & Wang, M. (2016). Acta Cryst. D72, 1036-1048.]

References

First citationBerman, H., Henrick, K. & Nakamura, H. (2003). Nat. Struct. Mol. Biol. 10, 980.  Web of Science CrossRef Google Scholar
First citationBernstein, H. J. & Hammersley, A. P. (2006). International Tables for Crystallography, Vol. G, pp. 37–43. International Union of Crystallography.  Google Scholar
First citationBigault, T., Ziegler, E., Morawe, C., Hustache, R., Massonnat, J.-Y. & Rostaing, G. (2003). Proc. SPIE, 5195, 12–20.  CrossRef Google Scholar
First citationBotha, S., Nass, K., Barends, T. R. M., Kabsch, W., Latz, B., Dworkowski, F., Foucar, L., Panepucci, E., Wang, M., Shoeman, R. L., Schlichting, I. & Doak, R. B. (2015). Acta Cryst. D71, 387–397.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBourenkov, G. P. & Popov, A. N. (2006). Acta Cryst. D62, 58–64.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBowler, M. W., Guijarro, M., Petitdemange, S., Baker, I., Svensson, O., Burghammer, M., Mueller-Dieckmann, C., Gordon, E. J., Flot, D., McSweeney, S. M. & Leonard, G. A. (2010). Acta Cryst. D66, 855–864.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540–1547.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBrockhauser, S., Ravelli, R. B. G. & McCarthy, A. A. (2013). Acta Cryst. D69, 1241–1251.  Web of Science CrossRef IUCr Journals Google Scholar
First citationCasanas, A., Warshamanage, R., Finke, A. D., Panepucci, E., Olieric, V., Nöll, A., Tampé, R., Brandstetter, S., Förster, A., Mueller, M., Schulze-Briese, C., Bunk, O. & Wang, M. (2016). Acta Cryst. D72, 1036–1048.  Web of Science CrossRef IUCr Journals Google Scholar
First citationCipriani, F., Felisaz, F., Launer, L., Aksoy, J.-S., Caserotto, H., Cusack, S., Dallery, M., di-Chiaro, F., Guijarro, M., Huet, J., Larsen, S., Lentini, M., McCarthy, J., McSweeney, S., Ravelli, R., Renier, M., Taffut, C., Thompson, A., Leonard, G. A. & Walsh, M. A. (2006). Acta Cryst. D62, 1251–1259.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationCipriani, F., Felisaz, F., Lavault, B., Brockhauser, S., Ravelli, R., Launer, L., Leonard, G. & Renier, M. (2007). AIP Conf. Proc. 879, 1928–1931.  CrossRef CAS Google Scholar
First citationDelagenière, S., Brenchereau, P., Launer, L., Ashton, A. W., Leal, R., Veyrier, S., Gabadinho, J., Gordon, E. J., Jones, S. D., Levik, K. E., McSweeney, S. M., Monaco, S., Nanao, M., Spruce, D., Svensson, O., Walsh, M. A. & Leonard, G. A. (2011). Bioinformatics, 27, 3186–3192.  Web of Science PubMed Google Scholar
First citationEmsley, P., Lohkamp, B., Scott, W. G. & Cowtan, K. (2010). Acta Cryst. D66, 486–501.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationFlot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107–118.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationFörster, A., Brandstetter, S. & Schulze-Briese, C. (2019). Philos. Trans. R. Soc. A, 377, 20180241.  Google Scholar
First citationGabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700–707.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGiraud, T., Dobias, F., Gabadinho, J., Rey-Bakaikoa, V., Nurizzo, D., Leonard, G. A. & Mueller-Dieckmann, C. (2009). J. Appl. Cryst. 42, 125–128.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGotthard, G., Aumonier, S., De Sanctis, D., Leonard, G., von Stetten, D. & Royant, A. (2019). IUCrJ, 6, 665–680.  Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
First citationGuijarro, M., Beteva, A., Coutinho, T. M., Dominguez, M. C., Homs, A., Meyer, J. M., Michel, V., Papillon, E., Perez, M. & Petitdemanage, S. (2017). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017), 8–13 October 2017, Barcelona, Spain, pp. 1060–1066. WEBPL05.  Google Scholar
First citationIncardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872–879.  Web of Science CrossRef IUCr Journals Google Scholar
First citationJanvier, N., Clement, J. M. & Fajardo, P. (2013). Proceedings of the 14th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2013), 6–11 October 2013, San Francisco, CA, USA, pp. 766–769. TUPPC08.  Google Scholar
First citationKabsch, W. (2010). Acta Cryst. D66, 125–132.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKovalevskiy, O., Nicholls, R. A., Long, F., Carlon, A. & Murshudov, G. N. (2018). Acta Cryst. D74, 215–227.  Web of Science CrossRef IUCr Journals Google Scholar
First citationLafumat, B., Mueller-Dieckmann, C., Leonard, G., Colloc'h, N., Prangé, T., Giraud, T., Dobias, F., Royant, A., van der Linden, P. & Carpentier, P. (2016). J. Appl. Cryst. 49, 1478–1487.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationLanger, G., Cohen, S. X., Lamzin, V. S. & Perrakis, A. (2008). Nat. Protoc. 3, 1171–1179.  Web of Science CrossRef PubMed CAS Google Scholar
First citationLinden, P. van der, Dobias, F., Vitoux, H., Kapp, U., Jacobs, J., Mc Sweeney, S., Mueller-Dieckmann, C. & Carpentier, P. (2014). J. Appl. Cryst. 47, 584–592.  Web of Science CrossRef IUCr Journals Google Scholar
First citationMarion, P. & Zhang, L. (2004). AIP Conf. Proc. 705, 320–323.  CrossRef CAS Google Scholar
First citationMcCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249–1260.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMcGeehan, J., Ravelli, R. B. G., Murray, J. W., Owen, R. L., Cipriani, F., McSweeney, S., Weik, M. & Garman, E. F. (2009). J. Synchrotron Rad. 16, 163–172.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMonaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804–810.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationNurizzo, D., Mairs, T., Guijarro, M., Rey, V., Meyer, J., Fajardo, P., Chavanne, J., Biasci, J.-C., McSweeney, S. & Mitchell, E. (2006). J. Synchrotron Rad. 13, 227–238.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationOscarsson, M., Beteva, A., Flot, D., Gordon, E., Guijarro, M., Leonard, G., McSweeney, S., Monaco, S., Mueller-Dieckmann, C., Nanao, M., Nurizzo, D., Popov, A., von Stetten, D., Svensson, O., Rey-Bakaikoa, V., Chado, I., Chavas, L., Gadea, L., Gourhant, P., Isabet, T., Legrand, P., Savko, M., Sirigu, S., Shepard, W., Thompson, A., Mueller, U., Nan, J., Eguiraun, M., Bolmsten, F., Nardella, A., Milàn-Otero, A., Thunnissen, M., Hellmig, M., Kastner, A., Schmuckermaier, L., Gerlach, M., Feiler, C., Weiss, M. S., Bowler, M. W., Gobbo, A., Papp, G., Sinoir, J., McCarthy, A., Karpics, I., Nikolova, M., Bourenkov, G., Schneider, T., Andreu, J., Cuní, G., Juanhuix, J., Boer, R., Fogh, R., Keller, P., Flensburg, C., Paciorek, W., Vonrhein, C., Bricogne, G. & de Sanctis, D. (2019). J. Synchrotron Rad. 26, 393–405.  Web of Science CrossRef IUCr Journals Google Scholar
First citationPape, T. & Schneider, T. R. (2004). J. Appl. Cryst. 37, 843–844.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPernot, P., Round, A., Barrett, R., De Maria Antolinos, A., Gobbo, A., Gordon, E., Huet, J., Kieffer, J., Lentini, M., Mattenet, M., Morawe, C., Mueller-Dieckmann, C., Ohlsson, S., Schmid, W., Surr, J., Theveneau, P., Zerrad, L. & McSweeney, S. (2013). J. Synchrotron Rad. 20, 660–664.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPerrakis, A., Cipriani, F., Castagna, J.-C., Claustre, L., Burghammer, M., Riekel, C. & Cusack, S. (1999). Acta Cryst. D55, 1765–1770.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPrates, J. A., Tarbouriech, N., Charnock, S. J., Fontes, C. M., Ferreira, L. M. & Davies, G. J. (2001). Structure, 9, 1183–1190.  Web of Science CrossRef PubMed CAS Google Scholar
First citationRussi, S., Juers, D. H., Sanchez-Weatherby, J., Pellegrini, E., Mossou, E., Forsyth, V. T., Huet, J., Gobbo, A., Felisaz, F., Moya, R., McSweeney, S. M., Cusack, S., Cipriani, F. & Bowler, M. W. (2011). J. Struct. Biol. 175, 236–243.  Web of Science CrossRef CAS PubMed Google Scholar
First citationSanchez-Weatherby, J., Bowler, M. W., Huet, J., Gobbo, A., Felisaz, F., Lavault, B., Moya, R., Kadlec, J., Ravelli, R. B. G. & Cipriani, F. (2009). Acta Cryst. D65, 1237–1246.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSanctis, D. de, Beteva, A., Caserotto, H., Dobias, F., Gabadinho, J., Giraud, T., Gobbo, A., Guijarro, M., Lentini, M., Lavault, B., Mairs, T., McSweeney, S., Petitdemange, S., Rey-Bakaikoa, V., Surr, J., Theveneau, P., Leonard, G. A. & Mueller-Dieckmann, C. (2012). J. Synchrotron Rad. 19, 455–461.  Web of Science CrossRef IUCr Journals Google Scholar
First citationSnigirev, A., Kohn, V., Snigireva, I. & Lengeler, B. (1996). Nature, 384, 49–51.  CrossRef CAS Web of Science Google Scholar
First citationSparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085–1092.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationStetten, D. von, Giraud, T., Carpentier, P., Sever, F., Terrien, M., Dobias, F., Juers, D. H., Flot, D., Mueller-Dieckmann, C., Leonard, G. A., de Sanctis, D. & Royant, A. (2015). Acta Cryst. D71, 15–26.  Web of Science CrossRef IUCr Journals Google Scholar
First citationSvensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757–1767.  Web of Science CrossRef IUCr Journals Google Scholar
First citationTheveneau, P., Baker, R., Barrett, R., Beteva, A., Bowler, M. W., Carpentier, P., Caserotto, H., Sanctis, D., Dobias, F., Flot, D., Guijarro, M., Giraud, T., Lentini, M., Leonard, G. A., Mattenet, M., McCarthy, A. A., McSweeney, S. M., Morawe, C., Nanao, M., Nurizzo, D., Ohlsson, S., Pernot, P., Popov, A. N., Round, A., Royant, A., Schmid, W., Snigirev, A., Surr, J. & Mueller-Dieckmann, C. (2013). J. Phys. Conf. Ser. 425, 012001.  CrossRef Google Scholar
First citationUsón, I. & Sheldrick, G. M. (2018). Acta Cryst. D74, 106–116.  Web of Science CrossRef IUCr Journals Google Scholar
First citationVonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293–302.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWakatsuki, S., Belrhali, H., Mitchell, E. P., Burmeister, W. P., McSweeney, S. M., Kahn, R., Bourgeois, D., Yao, M., Tomizaki, T. & Theveneau, P. (1998). J. Synchrotron Rad. 5, 215–221.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWhite, T. A., Kirian, R. A., Martin, A. V., Aquila, A., Nass, K., Barty, A. & Chapman, H. N. (2012). J. Appl. Cryst. 45, 335–341.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWinter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85–97.  Web of Science CrossRef IUCr Journals Google Scholar
First citationwwPDB consortium (2019). Nucl. Acid Res. 47, D520–D528.  Google Scholar
First citationZander, U., Bourenkov, G., Popov, A. N., de Sanctis, D., Svensson, O., McCarthy, A. A., Round, E., Gordeliy, V., Mueller-Dieckmann, C. & Leonard, G. A. (2015). Acta Cryst. D71, 2328–2343.  Web of Science CrossRef IUCr Journals Google Scholar
First citationZeldin, O. B., Gerstel, M. & Garman, E. F. (2013). J. Synchrotron Rad. 20, 49–57.  Web of Science CrossRef CAS IUCr Journals Google Scholar