research papers
Side-bounce beamlines using single-reflection diamond monochromators at Cornell High Energy Synchrotron Source
aCornell High Energy Synchrotron Source, Cornell University, Ithaca, New York, USA, and bMaterials and Manufacturing Directorate, Air Force Research Laboratory, Wright-Patterson Air Force Base, Dayton, Ohio, USA
*Correspondence e-mail: stoupin1@llnl.gov
The design and implementation of new beamlines featuring side-bounce (single-reflection) diamond monochromators at Cornell High Energy Synchrotron Source (CHESS) are described. Undulator radiation is monochromated using an interchangeable set of diamond crystal plates reflecting radiation in the horizontal (synchrotron) plane, where each crystal plate is set to one of the low-index Bragg reflections (111, 220, 311 and 400) in either Bragg or Laue reflection geometries. At the nominal in situ characterization of soft materials during processing, has been outfitted with the described setup.
of 18° these reflections deliver monochromated X-rays with photon energies of 9.7, 15.9, 18.65 and 22.5 keV, respectively. An X-ray mirror downstream of the diamond monochromator is used for rejection of higher radiation harmonics and for initial focusing of the monochromated beam. The characteristics of the X-ray beam entering the experimental station were measured experimentally and compared with the results of simulations. A reasonable agreement is demonstrated. It is shown that the use of selected high-dislocation-density `mosaic' diamond single-crystal plates produced using the chemical vapor deposition method yields a few-fold enhancement in the of the monochromated beam in comparison with that delivered by perfect crystals under the same conditions. At present, the Functional Materials Beamline at CHESS, which is used for time-resolvedKeywords: X-ray monochromators; diamond; beamlines.
1. Introduction
A single-reflection high-heat-load crystal monochromator is a well known concept for monochromatization of synchrotron radiation (Lennie et al., 2007; Jiang et al., 2012; Gomez et al., 2018). The monochromator selects a narrow energy bandwidth from a wide polychromatic (bending magnet, wiggler) or multi-harmonic (undulator) spectrum of incoming synchrotron radiation using a particular Bragg reflection. The monochromated X-ray beam can be used for a variety of experiments as long as the chosen photon energy yields a good signal contrast, and energy tunability in a wide range (ΔE ≳ 10 eV) is not required. Applications are found using X-ray imaging, and many methods of X-ray scattering.
The single-reflection approach has a number of advantages. For one, a convenient space-preserving beamline layout can be implemented, especially if the X-rays are reflected in the horizontal plane (the so-called side-bounce configuration). Additionally, monochromator motion control is relatively straightforward due to relaxed angular tolerances. Losses in diffraction efficiency due to heat-load-induced strain of the et al., 2000). Also, for a single-reflection monochromator radiation heat-load mitigation is simplified (e.g. water cooling versus cryocooling), especially if robust single crystals with low X-ray absorption and high like diamond are employed. These properties of the monochromator's crystal are also essential for beam multiplexing solutions at synchrotrons and XFELs (Als-Nielsen et al., 1994; Grübel et al., 1996; Juanhuix et al., 2014; Dippel et al., 2015; Zhu et al., 2014).
are minimal for a single-reflection monochromator compared with a double-crystal monochromator, where the efficiency may be reduced because the second (thermally undistorted) crystal reflects the divergent beam emanating from the distorted first crystal (BilderbackIn this work, the design, implementation and performance characteristics of two new side-bounce beamlines at the Cornell High Energy Synchrotron Source (CHESS) are reported. These beamlines were built as part of the recent facility upgrade (CHESS-U). Each of the two beamlines, designated beamline 2B and beamline 3B, is equipped with an individual undulator source (Temnykh et al., 2013, 2016) and a diamond multi-crystal side-bounce monochromator operating at a set of four fixed photon energies. One of the main challenges for the beamline implementation was the lack of a stable supply of large diamond plates of high crystal quality [i.e. crystals produced using the high-pressure high-temperature method (HPHT)]. Instead, selected commercially available chemical vapor deposited (CVD) single crystals were used. These crystals featured working regions characterized by a uniformly high dislocation density to increase the reflection intensities, as well as a minimal effective crystal-lattice curvature to prevent unpredictable focusing of the reflected beam and thus any possible substantial increase in the size of the beam profile (Stoupin et al., 2019a).
The key characteristics of the monochromated beam (photon ΔE/E ≃ 10−3, can be tolerated by a large number of experiments. The characteristic sizes of the profiles of CVD diamond-reflected beams in the experimental station were found to be manageable (e.g. 3 mm × 5 mm), and even beneficial in some experiments (e.g. an increased field of view for X-ray radiography). One of the newly constructed beamlines was chosen to host the Functional Materials program at CHESS, where X-ray scattering and X-ray imaging methods are used to perform materials research.
energy bandwidth and dimensions of the beam profile) were measured and compared with the results of ray-tracing simulations for perfect crystals. Our analysis indicates that selected imperfect CVD crystals operating in the Laue (transmission) geometry yield a few-fold enhancement in the of the reflected beam. This increase, leading to a relative energy resolution of about2. Beamline layout and operating principles
The layout of the side-bounce beamlines is shown schematically in Fig. 1. Fig. 1(a) shows the top view illustrating the horizontal (synchrotron) plane, while Figs. 1(b) and 1(c) show the side views.
The X-ray source (S) has finite dimensions and finite angular divergence. The source characteristics, calculated using the SPECTRA program (Tanaka & Kitamura, 2001) taking into account the parameters of the storage ring (Shanks et al., 2019) and the parameters of the undulator (Temnykh et al., 2013, 2016), are summarized in Table 1.
|
The divergent X-ray beam (gray in Fig. 1) emitted by the source is incident on the diamond crystal plate (C) of the monochromator (a symmetric Laue reflection is illustrated). The sizes of individual components, and the divergence of the X-ray beam, are exaggerated for clarity. The reflected beam (aqua) is converging, which illustrates the effect of pseudo-focusing in the Laue geometry (Sanchez del Rio et al., 1995). The beam is either bypassed [Fig. 1(b)] or reflected [Fig. 1(c)] by the harmonic rejection mirror (M) and intercepted by the area detector (AD) placed in the experimental station at a distance corresponding to a nominal sample location. The undulator beam is restricted by a vertical fixed aperture AV = 1 mm placed just upstream of the monochromator. There is no beam-limiting aperture in the horizontal direction. The monochromated beam can be shaped using two sets of variable slits. The first set () is placed just downstream of the monochromator while the second set () is placed near the sample position (upstream of the area detector). The horizontal slit blades are shown schematically in Fig. 1(a). The vertical aperture and vertical slit blades are shown schematically in Fig. 1(b) but omitted from Fig. 1(c) for clarity.
To preserve the original beam direction after reflection from the mirror, the
of the reflection is tilted by adjusting the monochromator crystal's azimuthal angle by an amountwhere θM is the angle of incidence to the X-ray mirror and θC is the of the reflection. A value of 18° was chosen by design for each of the two beamlines. This arrangement leads to a vertical offset of the beam by an amount
where LM is the distance from the crystal to the mirror. Values for LM, along with the source-to-monochromator distance L1 and monochromator-to-sample distance L2, are given in Table 2 for the two beamlines 2B and 3B. Given typical values for θM of a few milliradians (the optimized value depends on the working photon energy), the resulting offset is a few millimetres. While the mirror positions remain fixed, the mirror can be bypassed by an adjustment of Δχ of the crystal. This has been found useful for diagnostic purposes during beamline alignment and commissioning.
|
The side-bounce monochromator of each beamline holds four diamond crystal plates aligned for four different reflections which correspond to four photon energies of the Bragg reflected beam. Switching between the reflections requires translation and angular adjustment of the water-cooled copper block holding the crystal plates. For each reflection, the positions of the monochromator and mirror are stored. The saved positions can be optimized and updated during beamline alignment. After the optimization, switching between different reflections/energies is achieved by a single software command. More details on the monochromator are presented in Section 3.
3. Side-bounce diamond monochromator
3.1. Optomechanical design
The four diamond reflections chosen for the high-heat-load monochromator on the side-bounce beamlines are 111, 220, 131 and 400. The photon energies corresponding to the nominal Bragg angles for these reflections are shown in Table 3. The table also shows the values of the energy bandwidths estimated as
where Δψ ≃ 124 µrad is the divergence of the incident undulator beam and ɛhkl is the intrinsic relative energy width of a given diamond reflection (perfect crystal, π-polarization). Both values are taken as full widths at half-maximum (FWHM). Estimates by equation (3) are valid for the case of nearly perfect crystals and represent the total energy bandwidth of the reflected beam across the FWHM size.
|
Each of the reflections required a separate diamond crystal plate. The arrangement of the plates is shown schematically in Fig. 2(a) for the 3B beamline monochromator. The undulator beam (gray) is incident on a particular reflector. The reflected beam is shown in aqua. In the figure, the 111 reflector is being used. This reflector has a slightly asymmetric Bragg geometry denoted by the asymmetry angle η. A rectangular copper block holding the crystals is tilted at θC with respect to the incident beam. The edges of the block serve as an angular reference for the alignment of the crystal plates on the block. The schematic for the 2B beamline is similar except that all reflectors are aligned in Laue (transmission) geometry.
Fig. 2(b) shows a picture of the block holding the diamond plates. The water-cooled block, which is removable, is made of nickel-plated copper. The reflectors are installed using adjustable clamps, which support the lower portion (approximately half) of each plate and restrain each plate's motion with only minimal applied pressure. The lower portions of the plates are in thermal contact with the block using InGa eutectic (viscous liquid at room temperature), while the upper (working) portions are exposed to the undulator X-ray beam. The basic idea of the described crystal-mounting scheme was based on a preliminary design of diamond monochromators at the Linac Coherent Light Source (LCLS). This preliminary design, however, was not adopted at the LCLS in favor of an alternative solution (Stoupin et al., 2014).
Fig. 2(c) shows the mechanical assembly of the monochromator. The block holding the reflectors is placed, via bellows, in a vacuum chamber connected to the supporting structure. The chamber can be translated in the horizontal (synchrotron) plane using a pair of linear translation stages (x and y directions). These translations enable centering of a particular reflector on the incident beam. A viewport in the chamber enables video monitoring of luminescence in the diamond plates excited by the incident X-ray beam, thus providing visual alignment confirmation. The linear translation stages are installed on a segmented circle stage, which performs azimuthal rotation (χ) of the crystal. The segmented circle stage is placed on an elevation stage (z direction) having a range of motion which permits complete retraction of the plates from the incident beam. Finally, at the base of the assembly is the rotation stage controlling the scattering angle (θ). To accommodate the copper lines that supply the cooling water, the angular ranges of motion for χ and θ are limited to about ±5°. Therefore, accurate placement of the reflectors on the block is essential.
3.2. Choice of the diamond crystals
Based upon the horizontal source size and the divergence parameters of the X-ray source, the nominal horizontal size of the incident X-ray beam at the monochromator's location is calculated to be about bx ≃ 2.5 mm (FWHM). Therefore, at the of θC = 18°, the length of the crystal plate required to intercept the beam size in the Bragg geometry is = 8.1 mm. High-quality HPHT diamond crystal plates of such lengths and a specified crystallographic surface orientation are not widely available. The advantages of using high-quality (nearly perfect) crystals in the Bragg geometry are wavefront preservation and very high reflectivity (nearly 100% for diamond). Fortunately, it was possible to acquire one large plate with a close to 111 surface orientation. The 0.5 mm thick plate was fabricated by New Diamond Technology (Russia). It was characterized with rocking-curve X-ray topography in the double-crystal nondispersive configuration using an Si 220 asymmetric beam conditioner (Stoupin et al., 2016) and a photon energy of ∼8 keV on the 1-BM beamline (Macrander et al., 2016) at the Advanced Photon Source (Argonne National Laboratory). Rocking-curve topographs showing peak position1 δθm and curve width2 Δθ are presented in Fig. 3. The dashed rectangle (∼8.5 × 1.8 mm) shows a region of reasonably good quality. The variation in the curve width in this region was found to be ≲0.5 µrad root-mean-square (r.m.s.) and the variation in the peak position across the region was about 3 µrad (excluding edge effects). The surface orientation was found to be 3.3° off the (111) plane (polishing miscut). The misorientation direction is shown in Fig. 3 using the projection of the 111 reciprocal vector (black arrow) on the crystal surface. The crystal was employed as the 111 reflector (photon energy 9.74 keV) in the monochromator of beamline 3B.
CVD single-crystal plates procured from Applied Diamond were used for all other reflectors. These plates featured lateral sizes of 7 × 7 mm (square), a thickness of 1 mm and a uniform high dislocation density of ρ ≳ 1 × 106 cm−2. The dislocation density was greater than what conventional X-ray topography techniques can resolve (Bowen & Tanner, 1998), as confirmed in our prior study (Stoupin et al., 2019a), which included white-beam Laue X-ray topography. This resulted in an increase in the reflection intensities due to the enhancement of the effective intrinsic energy bandwidth (Stoupin et al., 2018, 2019b). The CVD plates were of two distinct crystallographic orientations: plates of the first type had a nominal (110) edge orientation (i.e. 110 reciprocal vector normal to the edge of the plate), while plates of the second type had a nominal (100) edge orientation. For plates of both types, the nominal orientation of the working 7 × 7 mm surface was (001). The average crystallographic orientation was measured using a Multiwire X-ray back-reflection instrument at Cornell Center for Materials Research. The precision of this instrument was approximately 0.2°. The measured angular deviations from the nominal orientation did not exceed 3°. The measured angular deviations were taken into account during initial alignment of the plates on the copper block. To achieve a convenient arrangement of the plates with respect to the nominal horizontal and vertical directions of the beamline layout, the 111 and 220 Laue reflectors were of the first type [(110)-edge orientation], while the 131 and 400 reflectors were of the second type [(100)-edge orientation]. In this scheme, orientation of the 131 reflector on the block required an azimuthal rotation of ∼18° with respect to the top surface of the block, as illustrated in Fig. 2(b) (the edge of the 131 reflector is tilted with respect to the top surface of the block). Selection of the best crystal plates was performed using X-ray rocking-curve topography (Stoupin et al., 2019a). Maps of the effective radius of curvature in the were calculated using spline interpolation of the rocking-curve peak position across the plates. Plates with working regions of relatively large effective radius of curvature (R0 ≳ 30–70 m) were selected. This criterion was used to prevent defocusing due to an unknown effective lattice curvature and the related substantial increase in the size of the reflected beam.
4. Beamline characterization and simulation
4.1. Experiment and simulation details
In this work we report the results of characterization on beamline 3B due to more detailed measurements conducted on this beamline. The results of studies of the CVD diamond reflectors on beamline 2B were found to be similar, bearing no contradiction to our conclusions. The characterization was performed while the storage ring was operated at a positron current of 50 mA, generating a substantial heat load for the primary optical components (the maximum operational current is set at 200 mA). No indication of a substantial change in the monochromator's performance was found during more recent tests conducted at 125 mA.
Profiles of the monochromated beams were imaged using the area detector (AD, as shown in Fig. 1). The adjustable slits S0 and S1 were fully open. An attenuator (a stack of Si wafers) was placed in front of the area detector to prevent image saturation. With the beam bypassing the X-ray mirror [cf. Fig. 1(a)], it was found that the signal on the detector was affected by higher radiation harmonics. Therefore, interpretation of images collected in this mode requires caution.
X-ray 2-filled of length 6 cm and an online calculator (Revesz, 2007). The mass–energy taken at the corresponding photon energies (Hubbell & Seltzer, 2004) was assumed as the mechanism responsible for the production of in the ionization chamber.
characterization was performed using an NThe slit S0 openings (gaps) were set to = 2 mm and = 6 mm (essentially not limiting the beam in the vertical direction), while the gaps of slit S1 were set to 2 mm × 2 mm.
The undulator K parameter was optimized by maximizing the intensity passing through both slits onto the The radiation bandwidth was measured using a crystal analyzer operating at Si 111 and Si 333 reflections [symmetric Si(111) crystal] using a method developed by Batterman et al. (2021).
Ray tracing was performed using Lux, which is a program based upon the BMAD toolkit (Sagan, 2006) for charged-particle and X-ray simulations. The simulations assumed a Gaussian source with the nominal parameters of Tables 1, 2 and 3. Perfect diamond crystals were modeled with a specified thickness and asymmetry parameters corresponding to the values measured for the actual crystal plates. The aperture AV and the effective crystal size in the horizontal direction were the beam-limiting apertures in the simulations of the full beam profiles. In the simulations of the X-ray slits S0 and S1 (2 mm × 2 mm gaps) were introduced at the corresponding locations. These settings were consistent with the experimental conditions. The measured parameters and those of ray-tracing simulations are summarized in Table 6.
4.2. Diamond 111 Bragg reflector
The data for the 111 reflector represent an important benchmark for evaluation of the beamline performance since, assuming symmetric reflection of a perfect crystal, the radiation wavefront of the reflected beam should be preserved. The beam profile is expected to be that of the corresponding undulator harmonic (Gaussian, to a good approximation), propagated to the observation plate of the detector and clipped by the AV aperture and by the lateral effective size of the crystal in the horizontal direction (∼2 mm). Therefore, the beam profile is expected to be Gaussian but truncated in the vertical direction (the beam size is expected to be comparable with the vertical aperture size of 1 mm) and, less prominently, truncated in the horizontal direction (due to the much larger horizontal divergence and the source size).
The images of the beam profiles collected for the 111 reflector under different conditions are shown in Fig. 4. In these tests, the mirror was set to θM = 3.8 mrad, and the undulator was optimized to achieve maximum at its third harmonic. Fig. 4(a) shows a profile of the beam bypassing the X-ray mirror. Fig. 4(b) shows a profile of the beam reflected from the mirror while the mirror benders' positions are set to minimize the curvature. Fig. 4(c) shows a profile of the beam reflected from the mirror while the mirror benders are set to provide best focusing at the detector position. The contour lines represent the two-dimensional Gaussian fits of the profiles, while the side plots show center slices (in the horizontal and vertical directions) of the profiles (blue lines) and the slices of the corresponding Gaussian fits (black lines).
The measured beam profiles in Figs. 4(a) and 4(b) reveal noticeable deviation from the Gaussian shape in both vertical and horizontal directions. The profile in the vertical direction is affected by the mirror's imperfections. It is difficult to quantify the vertical size exactly (due to the combined effect of the aperture and that of the mirror imperfections), yet it seems to be close to the nominal expected size. The latter can be calculated as
where sx,y is the size of the source and is the source divergence (Table 1). Table 4 summarizes the calculation of the nominal beam sizes according to equation (4), along with the observed values estimated using the Gaussian fits.
|
The size of the focused beam for an ideal focusing lens is , which is approximately 24 µm (FWHM) in the present case. The measured size of the focused beam in Fig. 4(c) is 47 µm. An estimate of the increase in the focused beam size due to the mirror's spherical aberrations was performed using an expression by Susini (1995). The estimated increase was found to be ∼5 µm. Therefore, the discrepancy between the measured size and that of the ideal lens is dominated by other imperfections of the mirror such as slope errors and micro-roughness.
The measured horizontal size of the beam without focusing is about 1 mm less than the nominal horizontal size. Note that it remains approximately the same in the two conditions, which suggests that the presence of higher radiation harmonics [Fig. 4(a)] and imperfections of the X-ray mirror are not the primary reasons for the observed mismatch. The mismatch can be explained by crystal bending due to imperfect mounting. We performed a ray-tracing simulation to quantify the radius of curvature R, which leads to the observed reduction in the horizontal size. The ray tracing was performed for the slightly asymmetric 111 diamond reflection η = 3.3° in the geometry of the experiment (including the beam-limiting aperture, which is a combination of AV and the lateral size of the crystal).3 Fig. 5 shows the simulated beam profile of the 111 reflector assuming R = 70 m and a concave shape with respect to the incident beam.
The case of the 111 reflector includes known objects and quantities (undulator radiation, a nearly perfect crystal affected by mounting strain, an X-ray mirror and a set of apertures/slits) and where X-ray propagation is well understood using the simple principles of geometric optics and available ray-tracing tools. The estimates and simulations show that the performance characteristics of the 111 reflector are not ideal, but the discrepancies can be explained using realistic imperfections.
4.3. CVD diamond Laue reflectors
Scattering from misoriented blocks within an imperfect crystal leads to an increase in the angular divergence, not only in the scattering (horizontal) plane but also in the vertical plane [see e.g. Wuttke (2014)]. As a result, the size of the beam profile inevitably increases in both dimensions. At the same time, the integrated reflectivity of the CVD crystal can be greater by an order of magnitude or more (Stoupin et al., 2018). For those experiments which do not require the high energy resolution provided by perfect crystals (ΔE/E ≃ 10−4), an increase in the total available size of the beam can be considered an advantage provided that the per unit area (flux density) is not reduced. The resulting enlarged beam can then be limited by an aperture and/or focused on a sample using secondary optics (Stoupin et al., 2019b) without a substantial loss in the throughput of the experiment.
The beam profiles of the 220, 131 and 400 CVD diamond Laue reflectors measured in the characterization experiments are shown in Figs. 6(a)–6(c). These profiles were measured for the beams reflected from the X-ray mirror with the mirror's benders set to minimize the mirror's curvature. Adjusting the positions of the benders in the available motion range in an attempt to focus the beam onto the detector did not yield any appreciable reduction in the beam size. This outcome indicates that the distortion of the radiation wavefront by the CVD diamond reflectors does not permit re-imaging of the radiation source.
The measured profiles are compared with the corresponding beam profiles simulated using ray tracing in perfect crystals in Figs. 6(d)–6(f). An increase in the characteristic sizes of the beam profiles is observed for the CVD reflectors compared with the profiles of the corresponding perfect-crystal simulations. The sizes of the increased profiles can be reasonably quantified using 2D Gaussian fits, as shown by the side plots in Fig. 6 showing the center slices of the profiles and the corresponding projections of the 2D Gaussian fits. For all profiles in Fig. 6, the corresponding characteristic sizes (FWHM) are summarized in Table 5. The profiles of the 220 and 400 reflectors permit comfortable beamline operations with nearly flat beam profiles slit-limited to 2 × 2 mm.
|
The profile of the 131 reflector [Fig. 6(b)] has an obvious tilted appearance. The fits include a rotation angle as a fit parameter. The fitted tilt was approximately 24°. The origin of the rotation of the beam profile appears to be related to the direction of a contour of equal lattice orientation in the corresponding CVD diamond plate, as evident from the results of rocking-curve topography (see supporting information for details). This effect could be useful for tailoring the beam profile to the needs of an experiment at the primary stage of beam monochromatization.
5. and energy bandwidth
Table 6 provides a summary of the measured energy bandwidth ΔEe and the Fe along with the values from the ray-tracing simulations (ΔEt, Ft). It also provides the undulator K parameters and the undulator harmonics Un chosen to maximize the undulator output at the corresponding operating photon energies of the reflectors. As mentioned above, due to the high quality of the crystal, the 111 reflector case represents an important benchmark in the analysis of the resulting values. From the comparison of the experimental and simulated bandwidths, it appears that the measurement method somewhat underestimates the ray-tracing result of 2.8 eV (derived for a crystal with a radius of curvature R = 70 m). At the same time, the measured is about 40% less than the simulated value. This mismatch could be related to a number of factors, including non-ideal performance of the undulator and non-ideal alignment of all individual optical components to the common optical axis. Thus, further analysis should take into account possible underestimation of the photon bandwidth and the maximum available Nevertheless, the measured value of the corresponding to the CVD reflectors is greater by a factor of at least 6. A comparative metric that is more relevant for the purpose of performance evaluation is the gain G = Fe/Ft × At/Ae, where At = (from Table 5) and Ae = 2 mm × 2 mm. The gain is reported in the last column of Table 6. The calculation takes into account the characteristic sizes of the simulated beams, which are less than 2 mm × 2 mm (see Table 5).
6. Applications
Recent experiments conducted on the 3B beamline were focused on a direct–ink–write 3D printing process using epoxy-nanoclay fiber-reinforced composites. These studies serve as an example to illustrate the capabilities and range of applicability for the beamline. Among the relevant experiments conducted on the 3B beamline, some were focused on fiber alignment kinetics during 3D printing (Croom et al., 2021), while others were focused on studies of the filler morphology of the printed parts (Trigg et al., 2021).
Fig. 7 (top) shows a schematic of the 3D printing process, involving an ink-extruding nozzle where fiber alignment was investigated using in situ X-ray radiography/phase-contrast imaging (PCI), and the printed roadmap where the morphology of the was investigated using small-angle X-ray scattering (SAXS) in raster-scanning mode.
The in situ X-ray radiography/PCI was performed using the 15.9 keV X-ray beam of the 220 CVD diamond reflector. Sequences of images were acquired using an ANDOR Neo 5.5 CMOS area detector equipped with an LuAG:Ce scintillator crystal using frame rates up to 25 frames per second. The nominal pixel size of the area detector, taking into account the 5× optical magnification, was 1.3 µm. The velocity field was determined using particle-image velocimetry (PIV), which evaluated the motion of fibers through the nozzle (Willert & Gharib, 1991). Accurate PIV measurement with a spatial resolution of ∼100 µm was demonstrated under flow conditions with velocities up to 2.7 mm s−1. Fig. 7(a) shows an example of the calculated velocity field near the tip of the nozzle, superimposed on a selected detector frame (radiograph).
The SAXS mapping of the roadmap's 10 photons s−1. The SAXS data were collected by a Pilatus 200K detector using a `flyscan' mode, in which the sample was translated continuously in the vertical direction while the detector was triggered at each 5 µm (0.1 s) interval. Each vertical sweep of the sample was followed by a horizontal sample translation and vertical carriage return. The mapped area was 1.82 mm × 1.55 mm (vertical × horizontal). Each 2D map took approximately 3.5 h to complete. Fig. 7(b) shows the image of the where the mapped parameter is the extent of nanoplatelet orientation extracted from the SAXS dataset (Trigg et al., 2021).
on the microscale was performed using the 9.7 keV X-ray beam of the 111 diamond reflector. A secondary focusing optics (compound refractive lens) was used to create an ∼2.5 × 10 µm beam with a of approximately 10These recent experiments were performed while the storage ring was operated at a positron current of 50 mA. Further enhancements to the in situ beamline capabilities due to improved image statistics are anticipated upon the planned transition to storage-ring operations at 200 mA.
7. Summary
In summary, we have reported on the design and implementation of beamlines featuring side-bounce (single-reflection) diamond monochromators at the Cornell High Energy Synchrotron Source. Undulator radiation is monochromated using an interchangeable set of diamond crystal plates reflecting radiation in the horizontal (synchrotron) plane, where each crystal plate is set to one of the low-index Bragg reflections 111, 220, 311 or 400. At the nominal
of 18° these reflections deliver monochromated X-rays with photon energies of 9.7, 15.9, 18.65 and 22.5 keV, respectively. The 111 reflector is a nearly perfect diamond crystal plate implemented in the Bragg (reflection) geometry, while the remaining reflectors are selected high-dislocation-density single-crystal CVD diamond plates implemented in the Laue (transmission) geometry.The case of the 111 reflector serves as a benchmark to compare the results of experimental measurements of the reflected beam characteristics (size of the beam profile, e.g. radiography, small- and wide-angle X-ray scattering, or X-ray fluorescence). At the same time, the enlarged sizes of the beam profiles provided by the CVD crystals (resulting from the increased angular divergence upon reflection from an imperfect high-dislocation-density crystal lattice) can be considered an advantage for certain experiments (e.g. an enlarged field of view in radiography).
and energy bandwidth) and those of ray-tracing simulations. A reasonable agreement is demonstrated. In the other cases [crystal plates oriented in the Laue (transmission) geometry] it is shown that the use of the pre-selected CVD diamond plates yields a two- to three-fold enhancement in the of the monochromated beam in comparison with the simulated delivered by a perfect crystal. This enhancement is accompanied by an increase in the energy bandwidth, which can be tolerated by a large number of synchrotron experiments (In this work, beamline characterization was performed while the storage ring was operated at a positron current of 50 mA. No indication of a substantial change in the monochromators' performance was found during more recent tests conducted at 125 mA. A four-fold increase in the reported value of the 13 photons s−1 in certain cases.
is anticipated at the final stage of the storage-ring commissioning where it will be routinely operated at 200 mA. Thus, the of the monochromated beam could reach values as high as 1 × 10During the facility upgrade, two similar beamlines were implemented featuring slightly different monochromator-to-sample distances. At present, one of the implemented beamlines hosts the Functional Materials program, where X-ray scattering and X-ray imaging methods are used to perform materials research (e.g. studies of the direct–ink–write 3D-printing process).
Supporting information
Supporting information file. DOI: https://doi.org/10.1107/S160057752100120X/mo5230sup1.pdf
Footnotes
‡Now at Lawrence Livermore National Laboratory, Livermore, California, USA.
1The peak position was determined as the average of the angular positions (mid-point) of the left slope and right slope of the curve; the slope positions correspond to the half-maximum level.
2The curve width was calculated as the difference between the slope positions, thus representing full width at half-maximum.
3Note that the limited lateral size of the crystal and the asymmetry of the reflection cannot explain the observed discrepancy.
Acknowledgements
The authors are grateful to H. Koerner (AFRL) for encouragement, stimulating interest and discussions. We acknowledge the efforts of CHESS staff on the construction of the beamlines. The effort of A. V. Kolyadin (New Diamond Technology) on the manufacturing and timely delivery of the HPHT diamond crystal plates is greatly appreciated. We thank E. Fontes, J. Brock and E. Arenholz for their support and encouragement.
Funding information
This work is based upon research conducted at the Cornell High Energy Synchrotron Source (CHESS) which is supported by the National Science Foundation under award DMR-1332208, and upon research conducted at the Materials Solutions Network at CHESS (MSN-C) which is supported by the Air Force Research Laboratory under award FA8650-19-2-5220. Support for the Cornell Center for Materials Research facility used in this work was provided through NSF Grant DMR-1719875, part of the NSF MRSEC Program. This research used resources of the Advanced Photon Source, a US Department of Energy (DOE) Office of Science User Facility operated for the DOE Office of Science by Argonne National Laboratory under Contract No. DE-AC02-06CH11357. This research was performed while E. B. Trigg and B. P. Croom held an NRC Research Associateship award at the Air Force Research Laboratory.
References
Als-Nielsen, J., Freund, A., Grübel, G., Linderholm, J., Nielsen, M., del Rio, M. & Sellschop, J. (1994). Nucl. Instrum. Methods Phys. Res. B, 94, 306–318. CAS Google Scholar
Batterman, B., Shastri, S. & Richter, D. (2021). Energy Resolution and Angular Divergence Measurement for Monochromatic Synchrotron Radiation, https://www.chess.cornell.edu/users/calculators/x-ray-calculations-energy-analyzer. Google Scholar
Bilderback, D. H., Freund, A. K., Knapp, G. S. & Mills, D. M. (2000). J. Synchrotron Rad. 7, 53–60. Web of Science CrossRef CAS IUCr Journals Google Scholar
Bowen, D. K. & Tanner, B. K. (1998). Editors. High Resolution X-ray Diffraction and Topography. London: Taylor and Francis. Google Scholar
Croom, B. P., Abbott, A., Kemp, J. W., Rueschhoff, L., Smieska, L., Woll, A., Stoupin, S. & Koerner, H. (2021). Additive Manufacturing, 37, 101701. CrossRef Google Scholar
Dippel, A.-C., Liermann, H.-P., Delitz, J. T., Walter, P., Schulte-Schrepping, H., Seeck, O. H. & Franz, H. (2015). J. Synchrotron Rad. 22, 675–687. Web of Science CrossRef CAS IUCr Journals Google Scholar
Gomez, A., Dina, G. & Kycia, S. (2018). Rev. Sci. Instrum. 89, 063301. Web of Science CrossRef PubMed Google Scholar
Grübel, G., Abernathy, D., Vignaud, G., Sanchez del Rio, M. & Freund, A. (1996). Rev. Sci. Instrum. 67, 3349. Google Scholar
Hubbell, J. & Seltzer, S. (2004). NIST Standard Reference Database 126: X-ray Mass Attenuation Coefficients. https://physics.nist.gov/PhysRefData/XrayMassCoef/tab3.html. Google Scholar
Jiang, Z., Li, X., Strzalka, J., Sprung, M., Sun, T., Sandy, A. R., Narayanan, S., Lee, D. R. & Wang, J. (2012). J. Synchrotron Rad. 19, 627–636. Web of Science CrossRef CAS IUCr Journals Google Scholar
Juanhuix, J., Gil-Ortiz, F., Cuní, G., Colldelram, C., Nicolás, J., Lidón, J., Boter, E., Ruget, C., Ferrer, S. & Benach, J. (2014). J. Synchrotron Rad. 21, 679–689. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lennie, A. R., Laundy, D., Roberts, M. A. & Bushnell-Wye, G. (2007). J. Synchrotron Rad. 14, 433–438. Web of Science CrossRef IUCr Journals Google Scholar
Macrander, A., Erdmann, M., Kujala, N., Stoupin, S., Marathe, S., Shi, X., Wojcik, M., Nocher, D., Conley, R., Sullivan, J., Goetze, K., Maser, J. & Assoufid, L. (2016). AIP Conf. Proc. 1741, 030030. Google Scholar
Revesz, P. (2007). Ion Chamber Flux Calculator. https://www.chess.cornell.edu/users/calculators/ion-chamber-flux-calculator. Google Scholar
Sagan, D. (2006). Nucl. Instrum. Methods Phys. Res. A, 558, 356–359. CrossRef CAS Google Scholar
Sanchez del Rio, M., Grübel, G., Als–Nielsen, J. & Nielsen, M. (1995). Rev. Sci. Instrum. 66, 5148–5152. CAS Google Scholar
Shanks, J., Barley, J., Barrett, S., Billing, M., Codner, G., Li, Y., Liu, X., Lyndaker, A., Rice, D., Rider, N., Rubin, D. L., Temnykh, A. & Wang, S. T. (2019). Phys. Rev. Phys. ST Accel. Beams, 22, 021602. CrossRef Google Scholar
Stoupin, S., Krawczyk, T., Liu, Z. & Franck, C. (2019a). Crystals, 9, 396. Web of Science CrossRef Google Scholar
Stoupin, S., Krawczyk, T., Ruff, J. P. C., Finkelstein, K. D., Lee, H. H. & Huang, R. (2019b). AIP Conf. Proc. 2054, 060019. Google Scholar
Stoupin, S., Ruff, J. P. C., Krawczyk, T. & Finkelstein, K. D. (2018). Acta Cryst. A74, 567–577. CrossRef IUCr Journals Google Scholar
Stoupin, S., Shvyd'ko, Yu., Trakhtenberg, E., Liu, Z., Lang, K., Huang, X., Wieczorek, M., Kasman, E., Hammonds, J., Macrander, A. & Assoufid, L. (2016). AIP Conf. Proc. 1741, 050020. Google Scholar
Stoupin, S., Terentyev, S. A., Blank, V. D., Shvyd'ko, Y. V., Goetze, K., Assoufid, L., Polyakov, S. N., Kuznetsov, M. S., Kornilov, N. V., Katsoudas, J., Alonso-Mori, R., Chollet, M., Feng, Y., Glownia, J. M., Lemke, H., Robert, A., Sikorski, M., Song, S. & Zhu, D. (2014). J. Appl. Cryst. 47, 1329–1336. Web of Science CrossRef CAS IUCr Journals Google Scholar
Susini, J. (1995). Opt. Eng. 34, 361–376. CrossRef Web of Science Google Scholar
Tanaka, T. & Kitamura, H. (2001). J. Synchrotron Rad. 8, 1221–1228. Web of Science CrossRef CAS IUCr Journals Google Scholar
Temnykh, A., Dale, D., Fontes, E., Li, Y., Lyndaker, A., Revesz, P., Rice, D. & Woll, A. (2013). J. Phys. Conf. Ser. 425, 032004. CrossRef Google Scholar
Temnykh, A., Dale, D., Fontes, E., Lyndaker, A., Li, Y., Ruff, J., Revesz, P. & Woll, A. (2016). AIP Conf. Proc. 1741, 020003. Google Scholar
Trigg, E. B., Hmeidat, N. S., Smieska, L. M., Woll, A. R., Compton, B. G. & Koerner, H. (2021). Additive Manufacturing, 37, 101729. CrossRef Google Scholar
Willert, C. & Gharib, M. (1991). Exp. Fluids, 10, 181–193. CrossRef Google Scholar
Wuttke, J. (2014). Acta Cryst. A70, 429–440. Web of Science CrossRef IUCr Journals Google Scholar
Zhu, D., Feng, Y., Stoupin, S., Terentyev, S. A., Lemke, H. T., Fritz, D. M., Chollet, M., Glownia, J. M., Alonso-Mori, R., Sikorski, M., Song, S., van Driel, T. B., Williams, G. J., Messerschmidt, M., Boutet, S., Blank, V. D., Shvyd'ko, Yu. V. & Robert, A. (2014). Rev. Sci. Instrum. 85, 063106. CrossRef PubMed Google Scholar
© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.