Figure 1
(a) A sketch of the IXS geometry. (b) Energy shift and (c) required temperature correction for the analyzer as a function of analyzer position for an Si analyzer near room temperature from equations (8) and (7). The geometry is off-Rowland with L1 = 2005 mm, R = 1900 mm, M = 0.90 and d = 6 mm. The size of the analyzer pixels is neglected. |