Development of the Soft X-ray AGM–AGS RIXS beamline at the Taiwan Photon Source
aNational Synchrotron Radiation Research Center, Hsinchu 30076, Taiwan, and bDepartment of Physics, National Tsing Hua University, Hsinchu 30013, Taiwan
*Correspondence e-mail: djhuang@nsrrc.org.tw
We report on the development of a high-resolution and highly efficient beamline for soft X-ray resonant inelastic X-ray scattering (RIXS) located at the Taiwan Photon Source. This beamline adopts an optical design that uses an active grating monochromator (AGM) and an active grating spectrometer (AGS) to implement the energy compensation principle of grating dispersion. Active gratings are utilized to diminish defocus, coma and higher-order aberrations, as well as to decrease the slope errors caused by thermal deformation and optical polishing. The AGS is mounted on a rotatable granite platform to enable momentum-resolved RIXS measurements with scattering angles over a wide range. Several high-precision instruments developed in-house for this beamline are described briefly. The best energy resolution obtained from this AGM–AGS beamline was 12.4 meV at 530 eV, achieving a 4, while the bandwidth of the incident soft X-rays was kept at 0.5 eV. To demonstrate the scientific impact of high-resolution RIXS, we present an example of momentum-resolved RIXS measurements on a high-temperature superconducting cuprate, i.e. La2–xSrxCuO4. The measurements reveal the A1g buckling phonons in superconducting cuprates, opening a new opportunity to investigate the coupling between these phonons and charge-density waves.
of 4.2 × 10Keywords: soft X-ray beamline; soft X-ray optics; grating monochromator; active grating; resonant inelastic X-ray scattering; soft X-ray spectra.
1. Introduction
The energy dispersion of low-energy elementary excitations in momentum space reflects the fundamental physical properties of materials. Resonant inelastic X-ray scattering (RIXS) is a powerful technique to probe these excitations with momentum resolution, and provides direct information about the dynamics arising from fluctuations of spin, charge and orbital ; Ament et al., 2011a).
(Kotani & Shin, 2001The process of X-ray absorption in a material and the subsequent re-emission of an X-ray of different energy is known as ). If the energy of the incident photons is tuned to an absorption resonance in which a core-level electron is excited to an unoccupied state, the subsequent X-ray depends strongly on the incident photon energy; this process is called resonant i.e. RIXS (Kotani & Shin, 2001; Ament et al., 2011a). It is also a scattering process in which the energy and momentum of the scattered X-rays conform to conservation rules, thus providing information about the energy and momentum of elementary excitations, such as d–d, charge-transfer, plasmon, magnon and phonon excitations, etc., of quantum materials (Lee et al., 2014; Minola et al., 2015; Huang et al., 2016, 2017; Fabbris et al., 2016; Tomiyasu et al., 2017; Hepting et al., 2018; Suzuki et al., 2018; Miao et al., 2019). The significantly enhances the scattering and offers a probe of elementary excitations with elemental and chemical selectivity. In addition, RIXS is a photon-in and photon-out technique that has been applied to explore matter in various phases.
(Schülke, 1989Despite its unique advantages, RIXS was unpopular among available spectrometric techniques because of a lack of good energy resolution and a very weak signal intensity. In the past decade, RIXS has, however, become widely accepted as one of the most powerful tools to investigate the properties of materials in terms of elementary excitations. After extensive developments of instrumentation, significant improvements in energy resolution and measurement efficiency have been achieved in the regime of soft X-ray energy. For example, the energy resolution of the AXES monochromator and spectrometer (Dallera et al., 1996; Ghiringhelli et al., 1998) improved from 500 meV at 530 eV in 1996 to 50 meV in 2013, which was achieved at the ADRESS beamline of the Swiss Light Source (Ghiringhelli et al., 2006b, 2012; Schmitt et al., 2013; Dinardo et al., 2007) on replacing the microchannel plate detector with a charge-coupled device (CCD) detector (Dinardo et al., 2007) and switching to a Dragon-type monochromator (Chen, 1987; Chen & Sette, 1989). These improvements allowed researchers to study the d–d excitations of 3d transition-metal oxides with an effective energy resolution (Ghiringhelli et al., 2004, 2006a). The SAXES spectrometer at the Swiss Light Source (Ghiringhelli et al., 2006b) enabled measurements of magnetic excitations in cuprate superconductors (Braicovich et al., 2010; Le Tacon et al., 2011). Several soft X-ray RIXS spectrometers have also been built in other synchrotron facilities, e.g. SEXTANTS at SOLEIL (Chiuzbăian et al., 2014), BL07LSU at SPring-8 (Yamamoto et al., 2014) and iRIXS at ALS (Chuang et al., 2017). Furthermore, beamline ID32 at the European Synchrotron Radiation Facility achieved a resolution of 30 meV at the Cu L3-edge (Brookes et al., 2018). This beamline enables RIXS intensity mapping as a function of momentum transfer through rotation of the spectrometer in ultra-high vacuum (UHV). A polarimeter has also been installed, allowing the polarization analysis of scattered photons. Recent years have seen an enhanced development of new high-resolution soft X-ray RIXS instruments, including beamline I21 at the Diamond Light Source, beamline SIX at the National Synchrotron Light Source II (Jarrige et al., 2018), the VERITAS soft X-ray RIXS beamline at MAX IV, the soft X-ray spectrometer PEAXIS at BESSY II (Schulz et al., 2020) and beamline 41A at the Taiwan Photon Source (TPS) (Huang & Chen, 2018).
To meet the two stringent requirements, i.e. high resolution and high efficiency, in soft X-ray RIXS experiments, a design concept of an active grating monochromator (AGM) and an active grating spectrometer (AGS) based on the principle of energy compensation of grating dispersion was conceived in 2002 and published in 2004 (Fung et al., 2004). In the AGM–AGS design, the efficiency of the RIXS measurements becomes greatly enhanced on increasing the bandwidth of the incident photons, while maintaining the energy resolution. The energy-compensation principle for RIXS has been tested successfully at the Taiwan Light Source (TLS) beamline 05A (Lai et al., 2014). Our theoretical simulations indicate that a resolving power better than 105 for photon energies from 400 to 1000 eV is achievable with an AGM–AGS beamline, motivating us to build a new soft X-ray RIXS beamline at TPS (Huang & Chen, 2018).
In this article, we report on the development of the soft X-ray AGM–AGS RIXS beamline located at TPS port 41A. This article is organized as follows. In §2, we introduce the RIXS branch of TPS 41A, including the design concept, a summary of the precision instruments developed in-house for this beamline and their performance. In §3, we discuss the commissioning results, including the adjustment of the grating surface profile, resolution optimization and RIXS measurements on the phonon excitations of high-temperature superconducting cuprates (HTSC). A summary and future plan follow in §4.
2. RIXS of TPS beamline 41A
2.1. Photon source
Beamline 41A at TPS composes two branches, i.e. high-resolution RIXS and coherent soft X-ray scattering. Both branches share the same monochromator, slits and front-end focusing optics. The photon source originates from two elliptically polarized undulators (EPU) in tandem in a 12 m straight section with a double-minimum β function to enhance the Each EPU magnet has a length of 3.2 m and a period of 48 mm. The of the EPU tandem is designed to be greater than 1 × 1020 photons s−1 mrad−2 mm−2 (0.1% bandwidth)−1 in the energy range from 400 to 1200 eV; the of the central cone exceeds 1 × 1015 photons s−1. In this energy range, the calculated beam sizes are about 386 and 28–35 µm at the full width at half-maximum (FWHM) in the horizontal and vertical directions, respectively; the beam divergences are, respectively, 42–61 and 33–52 µrad at FWHM, in the horizontal and vertical directions, depending on photon energy. Detailed parameters of EPU 48 are reported elsewhere (Chang et al., 2010; Chung et al., 2016).
2.2. Focusing optics and slits
The RIXS branch includes the front-end, focusing mirrors, slits, AGM grating, sample station, AGS grating and detector. Fig. 1 illustrates its optical layout and Fig. 2 shows the beamline photographs. The first optical element is a horizontal focusing mirror (HFM) located at 25.805 m from the centre of the EPU tandem. After that, a vertical plane mirror (VPM) and a vertical focusing mirror (VFM) are located at 1.2 and 2.4 m from the HFM, respectively. The VFM focuses the photon beam onto a vertical slit, i.e. the entrance slit of the AGM, with a demagnification of 28.2. The entrance-slit assembly comprises two water-cooled diamond blades. Through a flexure-based high-precision positioning mechanism, the translation and the tilt of the two blades are controlled independently with four actuators with resolutions of 0.002 µm and 0.02 µrad, respectively. The opening of the entrance slit and its centre can be varied continuously from 0 to 4000 µm. The monochromator uses an active grating located 4 m after the vertical entrance slit to focus the incident soft X-rays vertically onto the sample located 2.5 m after the AGM grating.
The bandwidth of the soft X-rays dispersed from the AGM grating is selected with an exit slit made of tungsten carbide. Through a translational stage driven with a UHV-compatible piezo motor, six discrete openings, i.e. 5, 10, 20, 50, 100 and 200 µm, can be selected. The exit slit placed 27.5 mm before the sample also sets the vertical beam size of the incident soft X-rays on the sample. In the horizontal direction, soft X-rays are focused in two steps, first with the HFM which has a demagnification of 8.3, followed by a horizontal refocusing mirror mirror M4 placed 0.6 m before the sample with a demagnification of 10.3. The designed horizontal beam size on the sample is 4.5 µm at FWHM. The soft X-rays scattered from the sample are focused with another horizontal focusing mirror M5, which has a collection angle of 18 mrad to enhance the efficiency of the AGS. The spectrometer uses an active grating located 2.5 m after the sample to focus the scattered soft X-rays vertically onto a two-dimensional (2D) detector located 5.5 m after the AGS grating. The major parameters of the focusing mirrors are summarized in Table 1.
‡HFM, VPM and VFM are also known as M1, M2 and M3, respectively. |
2.3. AGM–AGS scheme
The designs of the monochromator and spectrometer are based on the energy-compensation principle of grating dispersion (Fung et al., 2004; Lai et al., 2014). Incident X-rays from the entrance slit are diffracted, dispersed and focused onto the sample with the AGM grating; similarly, scattered X-rays are diffracted, dispersed and focused onto the 2D detector with the AGS grating. The entrance and exit arms of the AGM are, respectively, r1 and r2; those of the AGS are r1′ and r2′, respectively. The AGM–AGS scheme requires that the two gratings have an identical groove density n0 at the grating centre and that r1′ = r2. As shown in Fig. 1, the inelastically scattered X-rays with the same energy loss but different incident energies have the same dispersion from the AGS, so implementing the energy-compensation principle of grating dispersion. The AGM–AGS scheme has two important features: (i) the resolution and spectral-weight distribution of RIXS are insensitive to the incident bandwidth as long as the selected bandwidth is smaller than the core-hole lifetime width; (ii) the measurement efficiency is proportional to the selected bandwidth, because the energy-loss spectrum is the summation of the excited by the incident photons within the bandwidth. Therefore, the AGM–AGS scheme can greatly enhance the efficiency of RIXS measurement while maintaining the energy resolution.
2.4. Active gratings
To implement the AGM–AGS scheme for high-resolution RIXS, two high-precision active gratings are required. Using a multi-actuator bender, one can obtain a surface profile of the grating with a high-degree polynomial to diminish defocus, coma and higher-order aberrations, as well as to decrease the slope errors caused by thermal deformation and optical polishing. In this RIXS beamline, each active grating is made of a varied-line-spacing plane grating mounted on a high-precision 25-actuator bender (Kao et al., 2019); its surface slope is monitored with an in-position long-trace profiler (LTP) (Lin et al., 2019). Table 2 lists the optical parameters of the AGM and AGS gratings.
|
2.5. Optical tables and AGS rotation platform
The implementation of high-resolution RIXS requires mechanical adjustments and supports with great precision and stability for all mirrors and gratings. To fulfill these requirements, a high-precision, highly rigid and high-load optical table was designed and constructed for the HFM, VPM, VFM, AGM and AGS. This optical table can support a UHV chamber for an optical element up to 1000 kg in weight, with adjustments in all six i.e. three translations and three rotations. The resolution and repeatability of translation adjustments are 0.01 and 0.05 µm, respectively; those of rotation adjustments are 0.02 and 0.1 µrad, respectively. With comparable resolution and repeatability, an all-flexure-made high-precision optical table with adjustments in five was designed and constructed for M4, M5 and the 2D detector.
To facilitate the variation of the scattering angle, horizontal refocusing mirror M5, the AGS grating and the 2D detector are mounted on a rotational platform composed of two movable air-cushioned granite blocks with a high-precision quickly detachable connecting bridge in between, as shown in Fig. 2. This design can minimize the ground micro-vibrational effects during data acquisition. M5 and the AGS grating are placed on one block, with the 2D detector on the other. During a rotation of the platform, the two blocks are connected with the bridge; the air gap between the granite block and the granite floor is kept at 30–50 µm. Coupled with a unique UHV-compatible scattering chamber without a differential pumping, a wide scattering angle of 146°, i.e. from 17 to 163°, can be achieved.
2.6. Sample manipulation
The sample manipulator enables linear motions along three orthogonal directions with a resolution of 1 µm. These motions are driven with a translational stage using UHV stepper motors and optical encoders. Fig. 3 shows a photograph of this manipulator, which is mounted on top of an in-vacuum single-axis goniometer. Thus, our sample stage has four i.e. x, y, z and θ. The azimuthal rotation can be pre-adjusted in the air with a specially designed sample holder. The sample holder made of oxygen-free high-conductivity copper is connected to a liquid helium cryostat through copper braids. The sample holder is thermally isolated from the linear stage through a Vespel rod. The sample can be transferred in vacuum through a load-lock chamber and can be cooled to 20 K. In addition, the sample holder is electrically insulated from the cryostat, which enables us to perform measurements of X-ray absorption spectra (XAS) in the total-electron-yield mode. To minimize the radiation-induced damage and saturation of the detector, two shutters have been installed, one before mirror M4 and the other before the 2D detector.
2.7. Soft X-ray 2D detector
The scattering shows a photograph of a custom-made electron-multiplication CCD (EMCCD) detector without an antireflection coating at a grazing angle of 10°. The pixel size of the EMCCD is 13.5 µm and the grazing incidence gives an effective pixel size of 2.3 µm. The EMCCD is back-illuminated to avoid absorption of photons by the integrated circuits on the front side of the wafer. The detector head is cooled to −85°C. Through the electron-multiplication mechanism, the signal-to-noise ratio is enhanced. In addition, two four-jaw apertures are installed before both gratings to block stray light.
of RIXS is typically small and its signal is weak. To achieve a detection scheme near a detector of high sensitivity and low noise is required because only a few hundred electron–hole pairs can be generated with a silicon-based detector for soft X-rays. Fig. 42.8. Instrumentation control and user interface
The beamline instrumentation control for data acquisition is implemented on three levels: (1) the intrinsic commands of commercial devices, (2) the algorithm and software for the group control of commercial devices, and (3) the user-level commands and interface. Levels 2 and 3 are developed in-house using the Fortran and Python languages, respectively; the communication between them is based on EPICS tools. Fig. 5 shows the screen of the user interface, which is composed of four panels and a user-oriented command-line input (CLI). The first panel displays the RIXS 2D image recorded by the EMCCD. The second panel displays the real-time status of measurement parameters, including the positions and orientations of optical elements, monochromator and spectrometer energies, sample rotation angle (θ), scattering angle (2θ), sample temperature, shutter status, etc. The third panel shows the command history. The fourth panel presents a graph of RIXS spectra, θ–2θ angular scan, etc.
3. Commissioning results
3.1. Beam size
To achieve high-resolution RIXS measurements, a small soft X-ray source is required. The knife-edge scan method was used to measure the beam profile and to obtain the beam size at the slit positions. Figs. 6(a) and 6(b) show the measured beam-intensity profiles and their derivatives as a function of the knife-edge position of the vertical and horizontal entrance slits, respectively. The derivatives of the measured profile reveal that the beam sizes at the vertical and horizontal entrance slits are 1.97 and 44.38 µm at FWHM, respectively, in satisfactory agreement with the designed values. Fig. 6(c) plots the horizontal beam profile at the sample position; the focused beam size is 3.23 µm at FWHM in the horizontal direction.
3.2. Adjustment of the grating surface profile
To operate the AGM–AGS optical system, the surface profile of each of the two active gratings must be adjusted to meet its target profile for a given incident photon energy. The target profile yt(x) is expressed as a polynomial function,
in which x is the position along the longitudinal direction of the grating and ck is the kth coefficient of the polynomial. Our theoretical simulations indicate that a third-degree polynomial profile is sufficient to achieve an energy-resolving power of 2 × 104, whereas a fourth-degree polynomial is necessary for an energy-resolving power of 105. To facilitate the adjustment of the profile of the grating surface, an in-position LTP system was used to measure the slope of this profile. With the measured slope function, the target slope function and the set of actuator-response functions as inputs, one can use an iterative algorithm to deduce a set of actuator incremental values to adjust the surface profile to match the target profile. The details of the adjustment are reported elsewhere (Kao et al., 2019). Fig. 7 plots typical LTP measurement results, including the measured slope function and its target slope polynomial function, the difference between them and the measured surface profile of the grating. The slope difference between the measured and target functions was minimized to a root-mean-square (r.m.s.) value less than 0.25 µrad for both the AGM and the AGS grating. Because of the limited precision and accuracy of the in-position LTP instrument, the obtained slope difference is larger than the slope error of 0.1 µrad r.m.s. of the polished surface provided by the manufacturer.
3.3. Resolution optimization
We measured the linewidth of the 4C multilayer (ML) to tune the energy resolution. At the beginning, the grating surface profile was adjusted using the 25 actuators to match the initial target profile from the theoretical simulations of the designed beamline. Because of the alignment imperfections of the beamline, various sets of polynomial coefficients, i.e. [c2, c3, c4], were tested as the target profile to optimize the energy resolution. Fig. 8 shows the of 530 eV soft X-rays from the ML near a reflection condition with an entrance slit of 4 µm and an exit slit of 100 µm, which corresponds to a bandwidth of 0.5 eV of incident soft X-rays. The best achieved spectral resolution was 12.4 meV, which is defined as the minimum separation between two spectral lines when they can be resolved through a criterion that the core-to-wing ratio is smaller than the ratio of 0.9272 of two identical Gaussian functions separated by their FWHM.
from a W/B3.4. on the sample
As RIXS is a photon-demanding technique, a large 12 to 1.3 × 1013 photons s−1 with increased exit-slit opening. The measured energy resolution remained nearly unchanged, while the exit-slit opening increased. These data are consistent with our previous results (Lai et al., 2014) and demonstrate that the AGM–AGS scheme is also applicable to high-resolution soft X-ray RIXS.
on the sample is essential to achieve high-resolution measurements. The AGM–AGS scheme allows us to increase the bandwidth of the incident photons on the sample while maintaining the energy resolution. To verify this condition, we measured the energy resolution of the of 530 eV soft X-rays with the exit-slit opening set at 50, 100 and 200 µm. The measured increased linearly from 3.6 × 103.5. Phonon excitations of superconducting cuprates
The dressing of electrons by phonons in a material plays an important role in its novel electronic properties; in particular, the coupling between electrons and phonons in HTSC is pivotal yet remains under debate. RIXS has the potential to measure the strength of the coupling of electrons to lattice excitations. It has been shown theoretically that high-resolution RIXS provides direct, element-specific and momentum-resolved information about the electron–phonon coupling (Ament et al., 2011b; Devereaux et al., 2016). Here, we present O K-edge RIXS measurements on La1.88Sr0.12CuO4 (LSCO) to reveal its momentum-resolved phonon excitations.
La2–xSrxCuO4 is a doped Mott insulator (Imada et al., 1998) composed of CuO2 conducting layers with relatively weak coupling between the layers. Its mother compound La2CuO4 is an insulating antiferromagnet, in which the Cu ion has a 3d9 with a hole of symmetry x2 − y2. The system is a charge-transfer insulator because strong correlation effects split the into the upper and lower Hubbard bands. These correlation effects manifest themselves in O K-edge X-ray absorption in which an electron is excited from the 1s core level to the 2p band. The polarization-dependent measurements revealed that the doped holes are distributed mainly throughout the O 2px,y orbital in the CuO2 plaquette and are hybridized with the Cu orbital to form a spin singlet , termed a Zhang–Rice singlet (ZRS), in which denotes a ligand hole (Zhang & Rice, 1988; Chen et al., 1992). In the of La2CuO4, there exists a prepeak at the arising from the O 2p band hybridized with the upper Hubbard band (UHB) of Cu 3d. In a hole-doped cuprate, a lower-energy feature resulting from ZRS emerges and grows linearly with hole concentration in the under-doped regime, as plotted in Fig. 9(a) (Chen et al., 1991). That is, hole doping manifests itself in the spectral-weight transfer from UHB to ZRS as a consequence of electron correlations. The existence of ZRS enables measurements of phonon excitations using O K-edge RIXS. Fig. 9(b) shows the RIXS spectrum of LSCO measured with incident X-rays set at 528.4 eV, i.e. the ZRS feature. The inset of Fig. 9(b) illustrates the scattering geometry. The in-plane momentum transfer q∥ was along the direction (π, 0), with q∥ = 0.12; the out-of-plane momentum transfer was q⊥ = 1.0, in which a and c denote the lattice parameters. The RIXS spectrum in Fig. 9(b) shows an intense elastic peak and pronounced phonon excitations. We fitted the measured RIXS spectrum with four peaks, i.e. an peak and three phonon peaks, using Voigt functions. The observed three phonon excitations at 75, 45 and 20 meV originate from the half-breathing mode (Cohen et al., 1989; Chaplot et al., 1995; McQueeney et al., 1999; Fukuda et al., 2005; Giustino et al., 2008), A1g/B1g buckling modes (Cohen et al., 1989; McQueeney et al., 1999; Burns & Dacol, 1990; Giustino et al., 2008; Sugai et al., 2013; Johnston et al., 2010) and the vibrational mode involved with La/Sr (Sugai, 1989; Kim et al., 2017), respectively. These high-resolution RIXS measurements open a new opportunity to investigate the coupling between phonons and charge-density waves in HTSC.
4. Summary and future plan
We have designed, constructed and commissioned a high-resolution and highly efficient RIXS beamline based on the energy-compensation principle of grating dispersion. The achieved 4 at a photon energy of 530 eV, with the bandwidth of the incident soft X-rays set at 0.5 eV. With this high resolution, we observed three phonon excitations of LSCO. A new in-vacuum LTP instrument with a high precision of 0.005 µrad r.m.s. has been developed recently. It will be installed in the beamline to greatly improve the accuracy of the measured slope function of the grating surface, aiming to reach an ultra-high of 105. For our AGM–AGS design, the resolution is limited to the contribution of the grating slope error. Fig. 10 presents the estimated resolution contributed from the slope error of the grating, showing that such a goal is achievable if we can minimize the grating slope error to be smaller than 0.05 µrad r.m.s. We also plan to install a polarimeter for the polarization analysis of scattered soft X-rays and to develop a high-spatial-resolution and highly efficient soft X-ray 2D detector.
was 4.2 × 10Acknowledgements
We acknowledge the NSRRC staff for technical support; in particular, we thank D. J. Wang for useful discussions. We also acknowledge L. Y. Chang for his help during the commissioning stage. Thanks are due to The Ministry of Science and Technology of Taiwan and MOST for providing partial support to this work.
Funding information
Funding for this research was provided by: The Ministry of Science and Technology of Taiwan (grant No. 103-2112-M-213-008-MY3); MOST (grant No. 108-2923-M-213-001).
References
Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011a). Rev. Mod. Phys. 83, 705–767. Web of Science CrossRef CAS Google Scholar
Ament, L. J. P., van Veenendaal, M. & van den Brink, J. (2011b). EPL (Europhysics Lett), 95, 27008. Google Scholar
Braicovich, L., van den Brink, J., Bisogni, V., Sala, M. M., Ament, L. J. P., Brookes, N. B., De Luca, G. M., Salluzzo, M., Schmitt, T., Strocov, V. N. & Ghiringhelli, G. (2010). Phys. Rev. Lett. 104, 077002. Web of Science CrossRef PubMed Google Scholar
Brookes, N. B., Yakhou-Harris, F., Kummer, K., Fondacaro, A., Cezar, J. C., Betto, D., Velez-Fort, E., Amorese, A., Ghiringhelli, G., Braicovich, L., Barrett, R., Berruyer, G., Cianciosi, F., Eybert, L., Marion, P., van der Linden, P. & Zhang, L. (2018). Nucl. Instrum. Methods Phys. Res. A, 903, 175–192. Web of Science CrossRef CAS Google Scholar
Burns, G. & Dacol, F. (1990). Phys. Rev. B, 41, 4747–4748. CrossRef CAS Web of Science Google Scholar
Chang, C.-H., Hwang, C. S., Chang, C. H., Chen, H. H., Huang, M. H., Lin, F.-Y., Huang, J.-C. & Wu, C. M. (2010). Proceedings of the First International Particle Accelerator Conference (IPAC2010), 23–28 May 2010, Kyoto, Kyoto, Japan, pp. 3183–3185. WEPD042. Google Scholar
Chaplot, S., Reichardt, W., Pintschovius, L. & Pyka, N. (1995). Phys. Rev. B, 52, 7230–7242. CrossRef CAS Web of Science Google Scholar
Chen, C. T. (1987). Nucl. Instrum. Methods Phys. Res. A, 256, 595–604. CrossRef Web of Science Google Scholar
Chen, C. T. & Sette, F. (1989). Rev. Sci. Instrum. 60, 1616–1621. CrossRef CAS Web of Science Google Scholar
Chen, C. T., Sette, F., Ma, Y., Hybertsen, M. S., Stechel, E. B., Foulkes, W. M. C., Schulter, M., Cheong, S.-W., Cooper, A. S., Rupp, L. W. Jr, Batlogg, B., Soo, Y. L., Ming, Z. H., Krol, A. & Kao, Y. H. (1991). Phys. Rev. Lett. 66, 104–107. CrossRef PubMed CAS Web of Science Google Scholar
Chen, C. T., Tjeng, L. H., Kwo, J., Kao, H. L., Rudolf, P., Sette, F. & Fleming, R. M. (1992). Phys. Rev. Lett. 68, 2543–2546. CrossRef PubMed CAS Web of Science Google Scholar
Chiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J. M. (2014). Rev. Sci. Instrum. 85, 043108. Web of Science PubMed Google Scholar
Chuang, Y.-D., Shao, Y.-C., Cruz, A., Hanzel, K., Brown, A., Frano, A., Qiao, R., Smith, B., Domning, E., Huang, S.-W., Wray, L. A., Lee, W. S., Shen, Z. X., Devereaux, T. P., Chiou, J. W., Pong, W. F., Yashchuk, V. V., Gullikson, E., Reininger, R., Yang, W., Guo, J., Duarte, R. & Hussain, Z. (2017). Rev. Sci. Instrum. 88, 013110. Web of Science CrossRef PubMed Google Scholar
Chung, T. Y., Chang, C. H., Jan, J. C., Hwang, C. S., Chiu, M. S., Kuo, C. C., Chang, C. H., Tseng, F. H., Huang, J. C., Liu, Y. C., et al. (2016). Proceedings of the 7th International Particle Accelerator Conference (IPAC2016), 8–13 May 2016, Busan, Korea, pp. 4054–4056. THPOW049 Google Scholar
Cohen, R. E., Pickett, W. E. & Krakauer, H. (1989). Phys. Rev. Lett. 62, 831–834. CrossRef PubMed CAS Web of Science Google Scholar
Dallera, C., Puppin, E., Trezzi, G., Incorvaia, N., Fasana, A., Braicovich, L., Brookes, N. B. & Goedkoop, J. B. (1996). J. Synchrotron Rad. 3, 231–238. CrossRef CAS Web of Science IUCr Journals Google Scholar
Devereaux, T. P., Shvaika, A. M., Wu, K., Wohlfeld, K., Jia, C. J., Wang, Y., Moritz, B., Chaix, L., Lee, W.-S., Shen, Z.-X., Ghiringhelli, G. & Braicovich, L. (2016). Phys. Rev. X, 6, 041019. Google Scholar
Dinardo, M. E., Piazzalunga, A., Braicovich, L., Bisogni, V., Dallera, C., Giarda, K., Marcon, M., Tagliaferri, A. & Ghiringhelli, G. (2007). Nucl. Instrum. Methods Phys. Res. A, 570, 176–181. Web of Science CrossRef CAS Google Scholar
Fabbris, G., Meyers, D., Okamoto, J., Pelliciari, J., Disa, A., Huang, Y., Chen, Z.-Y., Wu, W., Chen, C., Ismail-Beigi, S., Ahn, C. H., Walker, F. J., Huang, D. J., Schmitt, T. & Dean, M. P. (2016). Phys. Rev. Lett. 117, 147401. Web of Science CrossRef PubMed Google Scholar
Fukuda, T., Mizuki, J., Ikeuchi, K., Yamada, K., Baron, A. & Tsutsui, S. (2005). Phys. Rev. B, 71, 060501. Web of Science CrossRef Google Scholar
Fung, H. S., Chen, C. T., Huang, L. J., Chang, C. H., Chung, S. C., Wang, D. J., Tseng, T. C. & Tsang, K. L. (2004). AIP Conf. Proc. 705, 655–658. CrossRef Google Scholar
Ghiringhelli, G., Braicovich, L., Schmitt, T. & Strocov, V. (2012). Synchrotron Radiat. News, 25(4), 16–22. CrossRef Google Scholar
Ghiringhelli, G., Brookes, N. B., Annese, E., Berger, H., Dallera, C., Grioni, M., Perfetti, L., Tagliaferri, A. & Braicovich, L. (2004). Phys. Rev. Lett. 92, 117406. Web of Science CrossRef PubMed Google Scholar
Ghiringhelli, G., Matsubara, M., Dallera, C., Fracassi, F., Tagliaferri, A., Brookes, N. B., Kotani, A. & Braicovich, L. (2006a). Phys. Rev. B, 73, 035111. Web of Science CrossRef Google Scholar
Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006b). Rev. Sci. Instrum. 77, 113108. Web of Science CrossRef Google Scholar
Ghiringhelli, G., Tagliaferri, A., Braicovich, L. & Brookes, N. B. (1998). Rev. Sci. Instrum. 69, 1610–1615. Web of Science CrossRef CAS Google Scholar
Giustino, F., Cohen, M. L. & Louie, S. G. (2008). Nature, 452, 975–978. Web of Science CrossRef PubMed CAS Google Scholar
Hepting, M., Chaix, L., Huang, E. W., Fumagalli, R., Peng, Y. Y., Moritz, B., Kummer, K., Brookes, N. B., Lee, W. C., Hashimoto, M., Sarkar, T., He, J. F., Rotundu, C. R., Lee, Y. S., Greene, R. L., Braicovich, L., Ghiringhelli, G., Shen, Z. X., Devereaux, T. P. & Lee, W. S. (2018). Nature, 563, 374–378. Web of Science CrossRef CAS PubMed Google Scholar
Huang, D. J. & Chen, C. T. (2018). Synchrotron Radiat. News, 31(2), 3–6. CrossRef Google Scholar
Huang, H., Jia, C., Chen, Z., Wohlfeld, K., Moritz, B., Devereaux, T., Wu, W., Okamoto, J., Lee, W., Hashimoto, M., He, Y., Shen, Z. X., Yoshida, Y., Eisaki, H., Mou, C. Y., Chen, C. T. & Huang, D. J. (2016). Sci. Rep. 6, 19657. Web of Science CrossRef PubMed Google Scholar
Huang, H.-Y., Chen, Z.-Y., Wang, R.-P., de Groot, F. M., Wu, W.-B., Okamoto, J., Chainani, A., Singh, A., Li, Z.-Y., Zhou, J.-S., Jeng, H. T., Guo, G. Y., Park, J. G., Tjeng, L. H., Chen, C. T. & Huang, D. J. (2017). Nat. Commun. 8, 15929. Web of Science CrossRef PubMed Google Scholar
Imada, M., Fujimori, A. & Tokura, Y. (1998). Rev. Mod. Phys. 70, 1039–1263. Web of Science CrossRef CAS Google Scholar
Jarrige, I., Bisogni, V., Zhu, Y., Leonhardt, W. & Dvorak, J. (2018). Synchrotron Radiat. News, 31(2), 7–13. CrossRef Google Scholar
Johnston, S., Vernay, F., Moritz, B., Shen, Z.-X., Nagaosa, N., Zaanen, J. & Devereaux, T. (2010). Phys. Rev. B, 82, 064513. Web of Science CrossRef Google Scholar
Kao, K. Y., Hua, C. Y., Fung, H. S., Lin, S. W., Chao, H. Y., Yeh, S. C., Chung, S. C., Huang, D. J. & Chen, C. T. (2019). AIP Conf. Proc. 2054, 060005. Google Scholar
Kim, G., Christiani, G., Logvenov, G., Choi, S., Kim, H.-H., Minola, M. & Keimer, B. (2017). Phys. Rev. Mater. 1, 054801. Web of Science CrossRef Google Scholar
Kotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203–246. Web of Science CrossRef CAS Google Scholar
Lai, C. H., Fung, H. S., Wu, W. B., Huang, H. Y., Fu, H. W., Lin, S. W., Huang, S. W., Chiu, C. C., Wang, D. J., Huang, L. J., Tseng, T. C., Chung, S. C., Chen, C. T. & Huang, D. J. (2014). J. Synchrotron Rad. 21, 325–332. Web of Science CrossRef CAS IUCr Journals Google Scholar
Lee, W., Lee, J., Nowadnick, E., Gerber, S., Tabis, W., Huang, S., Strocov, V., Motoyama, E., Yu, G., Moritz, B., Huang, H. Y., Wang, R. P., Huang, Y. B., Wu, W. B., Chen, C. T., Huang, D. J., Greven, M., Schmitt, T., Shen, Z. X. & Devereaux, T. P. (2014). Nat. Phys. 10, 883–889. Web of Science CrossRef CAS Google Scholar
Le Tacon, M., Ghiringhelli, G., Chaloupka, J., Sala, M. M., Hinkov, V., Haverkort, M. W., Minola, M., Bakr, M., Zhou, K. J., Blanco-Canosa, S., Monney, C., Song, Y. T., Sun, G. L., Lin, C. T., De Luca, G. M., Salluzzo, M., Khaliullin, G., Schmitt, T., Braicovich, L. & Keimer, B. (2011). Nat. Phys. 7, 725–730. Web of Science CrossRef CAS Google Scholar
Lin, S. W., Wang, D. J., Fu, H. W., Tsai, H. M., Hua, C. Y., Kuo, C. Y., Hsu, M. Y., Kao, K. Y., Yin, G. C., Fung, H. S., Perng, S. Y. & Chang, C. F. (2019). Rev. Sci. Instrum. 90, 021716. Web of Science CrossRef PubMed Google Scholar
McQueeney, R. J., Petrov, Y., Egami, T., Yethiraj, M., Shirane, G. & Endoh, Y. (1999). Phys. Rev. Lett. 82, 628–631. Web of Science CrossRef CAS Google Scholar
Miao, H., Fumagalli, R., Rossi, M., Lorenzana, J., Seibold, G., Yakhou-Harris, F., Kummer, K., Brookes, N., Gu, G., Braicovich, L., Ghiringhelli, G. & Dean, M. (2019). Phys. Rev. X, 9, 031042. Google Scholar
Minola, M., Dellea, G., Gretarsson, H., Peng, Y., Lu, Y., Porras, J., Loew, T., Yakhou, F., Brookes, N., Huang, Y., Pelliciari, J., Schmitt, T., Ghiringhelli, G., Keimer, B., Braicovich, L. & Le Tacon, M. (2015). Phys. Rev. Lett. 114, 217003. Web of Science CrossRef PubMed Google Scholar
Schmitt, T., Strocov, V. N., Zhou, K. J., Schlappa, J., Monney, C., Flechsig, U. & Patthey, L. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 38–46. Web of Science CrossRef CAS Google Scholar
Schülke, W. (1989). Nucl. Instrum. Methods Phys. Res. A, 280, 338–348. Google Scholar
Schulz, C., Lieutenant, K., Xiao, J., Hofmann, T., Wong, D. & Habicht, K. (2020). J. Synchrotron Rad. 27, 238–249. Web of Science CrossRef CAS IUCr Journals Google Scholar
Sugai, S. (1989). Phys. Rev. B, 39, 4306–4315. CrossRef CAS Web of Science Google Scholar
Sugai, S., Takayanagi, Y., Hayamizu, N., Muroi, T., Shiozaki, R., Nohara, J., Takenaka, K. & Okazaki, K. (2013). J. Phys. Condens. Matter, 25, 475701. Web of Science CrossRef PubMed Google Scholar
Suzuki, H., Minola, M., Lu, Y., Peng, Y., Fumagalli, R., Lefrançois, E., Loew, T., Porras, J., Kummer, K., Betto, D., Ishida, S., Eisaki, H., Hu, C., Zhou, X., Haverkort, M. W., Brookes, N. B., Braicovich, L., Ghiringhelli, G., Le Tacon, M. & Keimer, B. (2018). NPJ Quantum Mater. 3, 65. Google Scholar
Tomiyasu, K., Okamoto, J., Huang, H.-Y., Chen, Z.-Y., Sinaga, E. P., Wu, W.-B., Chu, Y.-Y., Singh, A., Wang, R.-P., de Groot, F., Chainani, A., Ishihara, S., Chen, C. T. & Huang, D. J. (2017). Phys. Rev. Lett. 119, 196402. Web of Science CrossRef PubMed Google Scholar
Yamamoto, S., Senba, Y., Tanaka, T., Ohashi, H., Hirono, T., Kimura, H., Fujisawa, M., Miyawaki, J., Harasawa, A., Seike, T., Takahashi, S., Nariyama, N., Matsushita, T., Takeuchi, M., Ohata, T., Furukawa, Y., Takeshita, K., Goto, S., Harada, Y., Shin, S., Kitamura, H., Kakizaki, A., Oshima, M. & Matsuda, I. (2014). J. Synchrotron Rad. 21, 352–365. Web of Science CrossRef CAS IUCr Journals Google Scholar
Zhang, F. C. & Rice, T. M. (1988). Phys. Rev. B, 37, 3759–3761. CrossRef CAS Web of Science Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.