Development of the Soft X-ray AGM–AGS RIXS beamline at the Taiwan Photon Source

crossmark logo

aNational Synchrotron Radiation Research Center, Hsinchu 30076, Taiwan, and bDepartment of Physics, National Tsing Hua University, Hsinchu 30013, Taiwan
*Correspondence e-mail: djhuang@nsrrc.org.tw

Edited by M. Yabashi, RIKEN SPring-8 Center, Japan (Received 4 June 2020; accepted 18 March 2021; online 30 March 2021)

We report on the development of a high-resolution and highly efficient beamline for soft X-ray resonant inelastic X-ray scattering (RIXS) located at the Taiwan Photon Source. This beamline adopts an optical design that uses an active grating monochromator (AGM) and an active grating spectrometer (AGS) to implement the energy com­pensation principle of grating dispersion. Active gratings are utilized to diminish defocus, coma and higher-order aberrations, as well as to decrease the slope errors caused by thermal deformation and optical polishing. The AGS is mounted on a rotatable granite platform to enable momentum-resolved RIXS measurements with scattering angles over a wide range. Several high-precision instruments developed in-house for this beamline are described briefly. The best energy resolution obtained from this AGM–AGS beamline was 12.4 meV at 530 eV, achieving a resolving power of 4.2 × 104, while the bandwidth of the incident soft X-rays was kept at 0.5 eV. To demonstrate the scientific impact of high-resolution RIXS, we present an example of momentum-resolved RIXS measurements on a high-temperature superconducting cuprate, i.e. La2–xSrxCuO4. The measurements reveal the A1g buckling phonons in superconducting cuprates, opening a new opportunity to investigate the coupling between these phonons and charge-density waves.

1. Introduction

The energy dispersion of low-energy elementary excitations in momentum space reflects the fundamental physical properties of materials. Resonant inelastic X-ray scattering (RIXS) is a powerful technique to probe these excitations with momentum resolution, and provides direct information about the dynamics arising from fluctuations of spin, charge and orbital degrees of freedom (Kotani & Shin, 2001[Kotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203-246.]; Ament et al., 2011a[Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011a). Rev. Mod. Phys. 83, 705-767.]).

The process of X-ray absorption in a material and the subsequent re-emission of an X-ray of different energy is known as inelastic X-ray scattering (IXS) (Schülke, 1989[Schülke, W. (1989). Nucl. Instrum. Methods Phys. Res. A, 280, 338-348.]). If the energy of the incident photons is tuned to an absorption resonance in which a core-level electron is excited to an unoccupied state, the subsequent X-ray emission spectrum depends strongly on the incident photon energy; this process is called resonant IXS, i.e. RIXS (Kotani & Shin, 2001[Kotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203-246.]; Ament et al., 2011a[Ament, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011a). Rev. Mod. Phys. 83, 705-767.]). It is also a scattering process in which the energy and momentum of the scattered X-rays conform to conservation rules, thus providing information about the energy and momentum of elementary excitations, such as dd, charge-transfer, plasmon, magnon and phonon excitations, etc., of quantum materials (Lee et al., 2014[Lee, W., Lee, J., Nowadnick, E., Gerber, S., Tabis, W., Huang, S., Strocov, V., Motoyama, E., Yu, G., Moritz, B., Huang, H. Y., Wang, R. P., Huang, Y. B., Wu, W. B., Chen, C. T., Huang, D. J., Greven, M., Schmitt, T., Shen, Z. X. & Devereaux, T. P. (2014). Nat. Phys. 10, 883-889.]; Minola et al., 2015[Minola, M., Dellea, G., Gretarsson, H., Peng, Y., Lu, Y., Porras, J., Loew, T., Yakhou, F., Brookes, N., Huang, Y., Pelliciari, J., Schmitt, T., Ghiringhelli, G., Keimer, B., Braicovich, L. & Le Tacon, M. (2015). Phys. Rev. Lett. 114, 217003.]; Huang et al., 2016[Huang, H., Jia, C., Chen, Z., Wohlfeld, K., Moritz, B., Devereaux, T., Wu, W., Okamoto, J., Lee, W., Hashimoto, M., He, Y., Shen, Z. X., Yoshida, Y., Eisaki, H., Mou, C. Y., Chen, C. T. & Huang, D. J. (2016). Sci. Rep. 6, 19657.], 2017[Huang, H.-Y., Chen, Z.-Y., Wang, R.-P., de Groot, F. M., Wu, W.-B., Okamoto, J., Chainani, A., Singh, A., Li, Z.-Y., Zhou, J.-S., Jeng, H. T., Guo, G. Y., Park, J. G., Tjeng, L. H., Chen, C. T. & Huang, D. J. (2017). Nat. Commun. 8, 15929.]; Fabbris et al., 2016[Fabbris, G., Meyers, D., Okamoto, J., Pelliciari, J., Disa, A., Huang, Y., Chen, Z.-Y., Wu, W., Chen, C., Ismail-Beigi, S., Ahn, C. H., Walker, F. J., Huang, D. J., Schmitt, T. & Dean, M. P. (2016). Phys. Rev. Lett. 117, 147401.]; Tomiyasu et al., 2017[Tomiyasu, K., Okamoto, J., Huang, H.-Y., Chen, Z.-Y., Sinaga, E. P., Wu, W.-B., Chu, Y.-Y., Singh, A., Wang, R.-P., de Groot, F., Chainani, A., Ishihara, S., Chen, C. T. & Huang, D. J. (2017). Phys. Rev. Lett. 119, 196402.]; Hepting et al., 2018[Hepting, M., Chaix, L., Huang, E. W., Fumagalli, R., Peng, Y. Y., Moritz, B., Kummer, K., Brookes, N. B., Lee, W. C., Hashimoto, M., Sarkar, T., He, J. F., Rotundu, C. R., Lee, Y. S., Greene, R. L., Braicovich, L., Ghiringhelli, G., Shen, Z. X., Devereaux, T. P. & Lee, W. S. (2018). Nature, 563, 374-378.]; Suzuki et al., 2018[Suzuki, H., Minola, M., Lu, Y., Peng, Y., Fumagalli, R., Lefrançois, E., Loew, T., Porras, J., Kummer, K., Betto, D., Ishida, S., Eisaki, H., Hu, C., Zhou, X., Haverkort, M. W., Brookes, N. B., Braicovich, L., Ghiringhelli, G., Le Tacon, M. & Keimer, B. (2018). NPJ Quantum Mater. 3, 65.]; Miao et al., 2019[Miao, H., Fumagalli, R., Rossi, M., Lorenzana, J., Seibold, G., Yakhou-Harris, F., Kummer, K., Brookes, N., Gu, G., Braicovich, L., Ghiringhelli, G. & Dean, M. (2019). Phys. Rev. X, 9, 031042.]). The resonance effect significantly enhances the scattering cross section and offers a probe of elementary excitations with elemental and chemical selectivity. In addition, RIXS is a photon-in and photon-out technique that has been applied to explore matter in various phases.

Despite its unique advantages, RIXS was unpopular among available spectrometric techniques because of a lack of good energy resolution and a very weak signal intensity. In the past decade, RIXS has, however, become widely accepted as one of the most powerful tools to investigate the properties of materials in terms of elementary excitations. After extensive developments of instrumentation, significant improvements in energy resolution and measurement efficiency have been achieved in the regime of soft X-ray energy. For example, the energy resolution of the AXES monochromator and spectrometer (Dallera et al., 1996[Dallera, C., Puppin, E., Trezzi, G., Incorvaia, N., Fasana, A., Braicovich, L., Brookes, N. B. & Goedkoop, J. B. (1996). J. Synchrotron Rad. 3, 231-238.]; Ghiringhelli et al., 1998[Ghiringhelli, G., Tagliaferri, A., Braicovich, L. & Brookes, N. B. (1998). Rev. Sci. Instrum. 69, 1610-1615.]) improved from 500 meV at 530 eV in 1996 to 50 meV in 2013, which was achieved at the ADRESS beamline of the Swiss Light Source (Ghiringhelli et al., 2006b[Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006b). Rev. Sci. Instrum. 77, 113108.], 2012[Ghiringhelli, G., Braicovich, L., Schmitt, T. & Strocov, V. (2012). Synchrotron Radiat. News, 25(4), 16-22.]; Schmitt et al., 2013[Schmitt, T., Strocov, V. N., Zhou, K. J., Schlappa, J., Monney, C., Flechsig, U. & Patthey, L. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 38-46.]; Dinardo et al., 2007[Dinardo, M. E., Piazzalunga, A., Braicovich, L., Bisogni, V., Dallera, C., Giarda, K., Marcon, M., Tagliaferri, A. & Ghiringhelli, G. (2007). Nucl. Instrum. Methods Phys. Res. A, 570, 176-181.]) on replacing the microchannel plate detector with a charge-coupled device (CCD) detector (Dinardo et al., 2007[Dinardo, M. E., Piazzalunga, A., Braicovich, L., Bisogni, V., Dallera, C., Giarda, K., Marcon, M., Tagliaferri, A. & Ghiringhelli, G. (2007). Nucl. Instrum. Methods Phys. Res. A, 570, 176-181.]) and switching to a Dragon-type monochromator (Chen, 1987[Chen, C. T. (1987). Nucl. Instrum. Methods Phys. Res. A, 256, 595-604.]; Chen & Sette, 1989[Chen, C. T. & Sette, F. (1989). Rev. Sci. Instrum. 60, 1616-1621.]). These improvements allowed researchers to study the dd excitations of 3d transition-metal oxides with an effective energy resolution (Ghiringhelli et al., 2004[Ghiringhelli, G., Brookes, N. B., Annese, E., Berger, H., Dallera, C., Grioni, M., Perfetti, L., Tagliaferri, A. & Braicovich, L. (2004). Phys. Rev. Lett. 92, 117406.], 2006a[Ghiringhelli, G., Matsubara, M., Dallera, C., Fracassi, F., Tagliaferri, A., Brookes, N. B., Kotani, A. & Braicovich, L. (2006a). Phys. Rev. B, 73, 035111.]). The SAXES spectrometer at the Swiss Light Source (Ghiringhelli et al., 2006b[Ghiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006b). Rev. Sci. Instrum. 77, 113108.]) enabled measurements of magnetic excitations in cuprate superconductors (Braicovich et al., 2010[Braicovich, L., van den Brink, J., Bisogni, V., Sala, M. M., Ament, L. J. P., Brookes, N. B., De Luca, G. M., Salluzzo, M., Schmitt, T., Strocov, V. N. & Ghiringhelli, G. (2010). Phys. Rev. Lett. 104, 077002.]; Le Tacon et al., 2011[Le Tacon, M., Ghiringhelli, G., Chaloupka, J., Sala, M. M., Hinkov, V., Haverkort, M. W., Minola, M., Bakr, M., Zhou, K. J., Blanco-Canosa, S., Monney, C., Song, Y. T., Sun, G. L., Lin, C. T., De Luca, G. M., Salluzzo, M., Khaliullin, G., Schmitt, T., Braicovich, L. & Keimer, B. (2011). Nat. Phys. 7, 725-730.]). Several soft X-ray RIXS spectrometers have also been built in other synchrotron facilities, e.g. SEXTANTS at SOLEIL (Chiuzbăian et al., 2014[Chiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J. M. (2014). Rev. Sci. Instrum. 85, 043108.]), BL07LSU at SPring-8 (Yamamoto et al., 2014[Yamamoto, S., Senba, Y., Tanaka, T., Ohashi, H., Hirono, T., Kimura, H., Fujisawa, M., Miyawaki, J., Harasawa, A., Seike, T., Takahashi, S., Nariyama, N., Matsushita, T., Takeuchi, M., Ohata, T., Furukawa, Y., Takeshita, K., Goto, S., Harada, Y., Shin, S., Kitamura, H., Kakizaki, A., Oshima, M. & Matsuda, I. (2014). J. Synchrotron Rad. 21, 352-365.]) and iRIXS at ALS (Chuang et al., 2017[Chuang, Y.-D., Shao, Y.-C., Cruz, A., Hanzel, K., Brown, A., Frano, A., Qiao, R., Smith, B., Domning, E., Huang, S.-W., Wray, L. A., Lee, W. S., Shen, Z. X., Devereaux, T. P., Chiou, J. W., Pong, W. F., Yashchuk, V. V., Gullikson, E., Reininger, R., Yang, W., Guo, J., Duarte, R. & Hussain, Z. (2017). Rev. Sci. Instrum. 88, 013110.]). Furthermore, beamline ID32 at the European Synchrotron Radiation Facility achieved a resolution of 30 meV at the Cu L3-edge (Brookes et al., 2018[Brookes, N. B., Yakhou-Harris, F., Kummer, K., Fondacaro, A., Cezar, J. C., Betto, D., Velez-Fort, E., Amorese, A., Ghiringhelli, G., Braicovich, L., Barrett, R., Berruyer, G., Cianciosi, F., Eybert, L., Marion, P., van der Linden, P. & Zhang, L. (2018). Nucl. Instrum. Methods Phys. Res. A, 903, 175-192.]). This beamline enables RIXS intensity mapping as a function of momentum transfer through rotation of the spectrometer in ultra-high vacuum (UHV). A polarimeter has also been installed, allowing the polarization analysis of scattered photons. Recent years have seen an enhanced development of new high-resolution soft X-ray RIXS instruments, including beamline I21 at the Diamond Light Source, beamline SIX at the National Synchrotron Light Source II (Jarrige et al., 2018[Jarrige, I., Bisogni, V., Zhu, Y., Leonhardt, W. & Dvorak, J. (2018). Synchrotron Radiat. News, 31(2), 7-13.]), the VERITAS soft X-ray RIXS beamline at MAX IV, the soft X-ray spectrometer PEAXIS at BESSY II (Schulz et al., 2020[Schulz, C., Lieutenant, K., Xiao, J., Hofmann, T., Wong, D. & Habicht, K. (2020). J. Synchrotron Rad. 27, 238-249.]) and beamline 41A at the Taiwan Photon Source (TPS) (Huang & Chen, 2018[Huang, D. J. & Chen, C. T. (2018). Synchrotron Radiat. News, 31(2), 3-6.]).

To meet the two stringent requirements, i.e. high resolution and high efficiency, in soft X-ray RIXS experiments, a design concept of an active grating monochromator (AGM) and an active grating spectrometer (AGS) based on the principle of energy com­pensation of grating dispersion was conceived in 2002 and published in 2004 (Fung et al., 2004[Fung, H. S., Chen, C. T., Huang, L. J., Chang, C. H., Chung, S. C., Wang, D. J., Tseng, T. C. & Tsang, K. L. (2004). AIP Conf. Proc. 705, 655-658.]). In the AGM–AGS design, the efficiency of the RIXS measurements becomes greatly enhanced on increasing the bandwidth of the incident photons, while maintaining the energy resolution. The energy-com­pensation principle for RIXS has been tested successfully at the Taiwan Light Source (TLS) beamline 05A (Lai et al., 2014[Lai, C. H., Fung, H. S., Wu, W. B., Huang, H. Y., Fu, H. W., Lin, S. W., Huang, S. W., Chiu, C. C., Wang, D. J., Huang, L. J., Tseng, T. C., Chung, S. C., Chen, C. T. & Huang, D. J. (2014). J. Synchrotron Rad. 21, 325-332.]). Our theoretical simulations indicate that a resolving power better than 105 for photon energies from 400 to 1000 eV is achievable with an AGM–AGS beamline, motivating us to build a new soft X-ray RIXS beamline at TPS (Huang & Chen, 2018[Huang, D. J. & Chen, C. T. (2018). Synchrotron Radiat. News, 31(2), 3-6.]).

In this article, we report on the development of the soft X-ray AGM–AGS RIXS beamline located at TPS port 41A. This article is organized as follows. In §2[link], we introduce the RIXS branch of TPS 41A, including the design concept, a summary of the precision instruments developed in-house for this beamline and their performance. In §3[link], we discuss the commissioning results, including the adjustment of the grating surface profile, resolution optimization and RIXS measurements on the phonon excitations of high-temperature superconducting cuprates (HTSC). A summary and future plan follow in §4[link].

2. RIXS of TPS beamline 41A

2.1. Photon source

Beamline 41A at TPS com­poses two branches, i.e. high-resolution RIXS and coherent soft X-ray scattering. Both branches share the same monochromator, slits and front-end focusing optics. The photon source originates from two elliptically polarized undulators (EPU) in tandem in a 12 m straight section with a double-minimum β function to enhance the brilliance. Each EPU magnet has a length of 3.2 m and a period of 48 mm. The brilliance of the EPU tandem is designed to be greater than 1 × 1020 photons s−1 mrad−2 mm−2 (0.1% band­width)−1 in the energy range from 400 to 1200 eV; the photon flux of the central cone exceeds 1 × 1015 photons s−1. In this energy range, the calculated beam sizes are about 386 and 28–35 µm at the full width at half-maximum (FWHM) in the horizontal and vertical directions, respectively; the beam divergences are, respectively, 42–61 and 33–52 µrad at FWHM, in the horizontal and vertical directions, depending on photon energy. Detailed parameters of EPU 48 are reported elsewhere (Chang et al., 2010[Chang, C.-H., Hwang, C. S., Chang, C. H., Chen, H. H., Huang, M. H., Lin, F.-Y., Huang, J.-C. & Wu, C. M. (2010). Proceedings of the First International Particle Accelerator Conference (IPAC2010), 23-28 May 2010, Kyoto, Kyoto, Japan, pp. 3183-3185. WEPD042.]; Chung et al., 2016[Chung, T. Y., Chang, C. H., Jan, J. C., Hwang, C. S., Chiu, M. S., Kuo, C. C., Chang, C. H., Tseng, F. H., Huang, J. C., Liu, Y. C., et al. (2016). Proceedings of the 7th International Particle Accelerator Conference (IPAC2016), 8-13 May 2016, Busan, Korea, pp. 4054-4056. THPOW049]).

2.2. Focusing optics and slits

The RIXS branch includes the front-end, focusing mirrors, slits, AGM grating, sample station, AGS grating and detector. Fig. 1[link] illustrates its optical layout and Fig. 2[link] shows the beamline photographs. The first optical element is a horizontal focusing mirror (HFM) located at 25.805 m from the centre of the EPU tandem. After that, a vertical plane mirror (VPM) and a vertical focusing mirror (VFM) are located at 1.2 and 2.4 m from the HFM, respectively. The VFM focuses the photon beam onto a vertical slit, i.e. the entrance slit of the AGM, with a demagnification of 28.2. The entrance-slit assembly com­prises two water-cooled diamond blades. Through a flexure-based high-precision positioning mechanism, the translation and the tilt of the two blades are controlled independently with four actuators with resolutions of 0.002 µm and 0.02 µrad, respectively. The opening of the entrance slit and its centre can be varied continuously from 0 to 4000 µm. The monochromator uses an active grating located 4 m after the vertical entrance slit to focus the incident soft X-rays vertically onto the sample located 2.5 m after the AGM grating.

[Figure 1]
Figure 1
Optical layout of the AGM–AGS RIXS. Abbreviations for the optical elements are defined in the text. Distances between optical elements are summarized in Tables 1[link] and 2[link].
[Figure 2]
Figure 2
Photographs of the RIXS set-up installed at TPS beamline 41A. The AGS grating and the EMCCD detector are placed on two separate granite platforms. Through an air-cushion mechanism, the spectrometer can swing over a wide scattering angle from 17 to 163° about a vertical axis at the sample position. Inset photographs show the water-cooled active grating and slit assemblies.

The bandwidth of the soft X-rays dispersed from the AGM grating is selected with an exit slit made of tungsten carbide. Through a translational stage driven with a UHV-com­patible piezo motor, six discrete openings, i.e. 5, 10, 20, 50, 100 and 200 µm, can be selected. The exit slit placed 27.5 mm before the sample also sets the vertical beam size of the incident soft X-rays on the sample. In the horizontal direction, soft X-rays are focused in two steps, first with the HFM which has a demagnification of 8.3, followed by a horizontal refocusing mirror mirror M4 placed 0.6 m before the sample with a demagnification of 10.3. The designed horizontal beam size on the sample is 4.5 µm at FWHM. The soft X-rays scattered from the sample are focused with another horizontal focusing mirror M5, which has a collection angle of 18 mrad to enhance the efficiency of the AGS. The spectrometer uses an active grating located 2.5 m after the sample to focus the scattered soft X-rays vertically onto a two-dimensional (2D) detector located 5.5 m after the AGS grating. The major parameters of the focusing mirrors are summarized in Table 1[link].

Table 1
Major parameters of the mirrors used at the RIXS branch of TPS beamline 41A

All mirrors have an Au-coated surface. HFM is with a Glidcop substrate, while others are with an Si substrate. All values of r1, r2 and radius are given in units of m.

Optic Type r1 r2 Radius Deviation angle (°)
HFM cylindrical 25.805 3.100 211 3
VPM plane 3
VFM active 28.205 1.000 74 3
M4 plane–elliptical 6.200 0.600 42 3
M5 plane–elliptical 0.750 9.500 36 4
†Radius at the mirror centre.
‡HFM, VPM and VFM are also known as M1, M2 and M3, respectively.

2.3. AGM–AGS scheme

The designs of the monochromator and spectrometer are based on the energy-com­pensation principle of grating dispersion (Fung et al., 2004[Fung, H. S., Chen, C. T., Huang, L. J., Chang, C. H., Chung, S. C., Wang, D. J., Tseng, T. C. & Tsang, K. L. (2004). AIP Conf. Proc. 705, 655-658.]; Lai et al., 2014[Lai, C. H., Fung, H. S., Wu, W. B., Huang, H. Y., Fu, H. W., Lin, S. W., Huang, S. W., Chiu, C. C., Wang, D. J., Huang, L. J., Tseng, T. C., Chung, S. C., Chen, C. T. & Huang, D. J. (2014). J. Synchrotron Rad. 21, 325-332.]). Incident X-rays from the entrance slit are diffracted, dispersed and focused onto the sample with the AGM grating; similarly, scattered X-rays are diffracted, dispersed and focused onto the 2D detector with the AGS grating. The entrance and exit arms of the AGM are, respectively, r1 and r2; those of the AGS are r1′ and r2′, respectively. The AGM–AGS scheme requires that the two gratings have an identical groove density n0 at the grating centre and that r1′ = r2. As shown in Fig. 1[link], the inelastically scattered X-rays with the same energy loss but different incident energies have the same dispersion from the AGS, so implementing the energy-com­pensation principle of grating dispersion. The AGM–AGS scheme has two important features: (i) the resolution and spectral-weight distribution of RIXS are insensitive to the incident bandwidth as long as the selected bandwidth is smaller than the core-hole lifetime width; (ii) the measurement efficiency is proportional to the selected bandwidth, because the energy-loss spectrum is the summation of the inelastic scattering excited by the incident photons within the bandwidth. Therefore, the AGM–AGS scheme can greatly enhance the efficiency of RIXS measurement while maintaining the energy resolution.

2.4. Active gratings

To implement the AGM–AGS scheme for high-resolution RIXS, two high-precision active gratings are required. Using a multi-actuator bender, one can obtain a surface profile of the grating with a high-degree polynomial to diminish defocus, coma and higher-order aberrations, as well as to decrease the slope errors caused by thermal deformation and optical polishing. In this RIXS beamline, each active grating is made of a varied-line-spacing plane grating mounted on a high-precision 25-actuator bender (Kao et al., 2019[Kao, K. Y., Hua, C. Y., Fung, H. S., Lin, S. W., Chao, H. Y., Yeh, S. C., Chung, S. C., Huang, D. J. & Chen, C. T. (2019). AIP Conf. Proc. 2054, 060005.]); its surface slope is monitored with an in-position long-trace profiler (LTP) (Lin et al., 2019[Lin, S. W., Wang, D. J., Fu, H. W., Tsai, H. M., Hua, C. Y., Kuo, C. Y., Hsu, M. Y., Kao, K. Y., Yin, G. C., Fung, H. S., Perng, S. Y. & Chang, C. F. (2019). Rev. Sci. Instrum. 90, 021716.]). Table 2[link] lists the optical parameters of the AGM and AGS gratings.

Table 2
Optical parameters of the AGM and AGS gratings

Both gratings have a laminar groove profile with a duty ratio of 2:3 and a groove depth between 8 and 6 nm, with a position-dependent density n(x) = n0 + n1x, where x is defined in §3.2[link].

  AGM AGS
Size (L × W × T) (mm3) 186 × 45 × 10 186 × 45 × 10
Substrate Si Si
Coating Au Au
Radius (m) 85–155 50–60
Entrance arm (m) r1 = 4.0 r1′ = 2.5
Exit arm (m) r2 = 2.5 r2′ = 5.5
n0 (grooves mm−1) 1200 1200
n1 (grooves mm−2) 0.80 −0.32 (low energy)
    −0.13 (high energy)

2.5. Optical tables and AGS rotation platform

The implementation of high-resolution RIXS requires mechanical adjustments and supports with great precision and stability for all mirrors and gratings. To fulfill these requirements, a high-precision, highly rigid and high-load optical table was designed and constructed for the HFM, VPM, VFM, AGM and AGS. This optical table can support a UHV chamber for an optical element up to 1000 kg in weight, with adjustments in all six degrees of freedom, i.e. three translations and three rotations. The resolution and repeatability of translation adjustments are 0.01 and 0.05 µm, respectively; those of rotation adjustments are 0.02 and 0.1 µrad, respectively. With com­parable resolution and repeatability, an all-flexure-made high-precision optical table with adjustments in five degrees of freedom was designed and constructed for M4, M5 and the 2D detector.

To facilitate the variation of the scattering angle, horizontal refocusing mirror M5, the AGS grating and the 2D detector are mounted on a rotational platform com­posed of two movable air-cushioned granite blocks with a high-precision quickly detachable connecting bridge in between, as shown in Fig. 2[link]. This design can minimize the ground micro-vibrational effects during data acquisition. M5 and the AGS grating are placed on one block, with the 2D detector on the other. During a rotation of the platform, the two blocks are connected with the bridge; the air gap between the granite block and the granite floor is kept at 30–50 µm. Coupled with a unique UHV-com­patible scattering chamber without a differential pumping, a wide scattering angle of 146°, i.e. from 17 to 163°, can be achieved.

2.6. Sample manipulation

The sample manipulator enables linear motions along three orthogonal directions with a resolution of 1 µm. These motions are driven with a translational stage using UHV stepper motors and optical encoders. Fig. 3[link] shows a photograph of this manipulator, which is mounted on top of an in-vacuum single-axis goniometer. Thus, our sample stage has four degrees of freedom, i.e. x, y, z and θ. The azimuthal rotation can be pre-adjusted in the air with a specially designed sample holder. The sample holder made of oxygen-free high-conductivity copper is connected to a liquid helium cryostat through copper braids. The sample holder is thermally isolated from the linear stage through a Vespel rod. The sample can be transferred in vacuum through a load-lock chamber and can be cooled to 20 K. In addition, the sample holder is electrically insulated from the cryostat, which enables us to perform measurements of X-ray absorption spectra (XAS) in the total-electron-yield mode. To minimize the radiation-induced damage and saturation of the detector, two shutters have been installed, one before mirror M4 and the other before the 2D detector.

[Figure 3]
Figure 3
Photograph of the sample manipulator inside the RIXS chamber. The sample xyz stage was developed using UHV-com­patible stepper motors and optical encoders. The sample position can be aligned with the incident X-ray beam with a resolution of 1 µm. Two photodiodes are installed to measure the incident photon flux and fluorescence.

2.7. Soft X-ray 2D detector

The scattering cross section of RIXS is typically small and its signal is weak. To achieve a detection scheme near photon counting, a detector of high sensitivity and low noise is required because only a few hundred electron–hole pairs can be generated with a silicon-based detector for soft X-rays. Fig. 4[link] shows a photograph of a custom-made electron-multiplication CCD (EMCCD) detector without an antireflection coating at a grazing angle of 10°. The pixel size of the EMCCD is 13.5 µm and the grazing incidence gives an effective pixel size of 2.3 µm. The EMCCD is back-illuminated to avoid absorption of photons by the integrated circuits on the front side of the wafer. The detector head is cooled to −85°C. Through the electron-multiplication mechanism, the signal-to-noise ratio is enhanced. In addition, two four-jaw apertures are installed before both gratings to block stray light.

[Figure 4]
Figure 4
The RIXS 2D image detector. A photograph of the detector mounted on a granite floor is shown on the left. Details of a customized EMCCD sensor located inside the chamber marked with a red circle are revealed on the right.

2.8. Instrumentation control and user interface

The beamline instrumentation control for data acquisition is implemented on three levels: (1) the intrinsic commands of commercial devices, (2) the algorithm and software for the group control of commercial devices, and (3) the user-level commands and interface. Levels 2 and 3 are developed in-house using the Fortran and Python languages, respectively; the communication between them is based on EPICS tools. Fig. 5[link] shows the screen of the user interface, which is com­posed of four panels and a user-oriented command-line input (CLI). The first panel displays the RIXS 2D image recorded by the EMCCD. The second panel displays the real-time status of measurement parameters, including the positions and orientations of optical elements, monochromator and spectrometer energies, sample rotation angle (θ), scattering angle (2θ), photon flux, sample temperature, shutter status, etc. The third panel shows the command history. The fourth panel presents a graph of RIXS spectra, XAS, θ–2θ angular scan, etc.

[Figure 5]
Figure 5
Screenshot of the user interface, which is com­posed of four panels and one CLI. Panel 1 shows the measured RIXS 2D image. Panel 2 displays the experimental parameters in real time. Panels 3 and 4 show the command history and the one-dimensional (1D) spectra for RIXS or XAS, respectively.

3. Commissioning results

3.1. Beam size

To achieve high-resolution RIXS measurements, a small soft X-ray source is required. The knife-edge scan method was used to measure the beam profile and to obtain the beam size at the slit positions. Figs. 6[link](a) and 6[link](b) show the measured beam-intensity profiles and their derivatives as a function of the knife-edge position of the vertical and horizontal entrance slits, respectively. The derivatives of the measured profile reveal that the beam sizes at the vertical and horizontal entrance slits are 1.97 and 44.38 µm at FWHM, respectively, in satisfactory agreement with the designed values. Fig. 6[link](c) plots the horizontal beam profile at the sample position; the focused beam size is 3.23 µm at FWHM in the horizontal direction.

[Figure 6]
Figure 6
Measurements of the soft X-ray beam profile. (a, b) Measurements of the beam profile at the entrance slits for the vertical and horizontal directions, respectively. (c) The horizontal beam profile of the soft X-rays at the sample position. The beam profile was obtained on detecting the beam intensity with a photodiode after a blade moving along the transverse direction. The beam intensity from the photodiode as a function of the blade position is plotted with black circles, together with a green curve showing its Fourier filtered function; the beam profile after the differentiation of the photodiode signal is plotted with a blue curve for its Fourier filtered function. The measured beam sizes of (a), (b) and (c) are 1.97, 44.38 and 3.23 µm at FWHM, respectively.

3.2. Adjustment of the grating surface profile

To operate the AGM–AGS optical system, the surface profile of each of the two active gratings must be adjusted to meet its target profile for a given incident photon energy. The target profile yt(x) is expressed as a polynomial function,

[y_{t}(x) = \sum_{k\,=\,0}c_{k}\,x^{\,k}, \eqno(1)]

in which x is the position along the longitudinal direction of the grating and ck is the kth coefficient of the polynomial. Our theoretical simulations indicate that a third-degree polynomial profile is sufficient to achieve an energy-resolving power of 2 × 104, whereas a fourth-degree polynomial is necessary for an energy-resolving power of 105. To facilitate the adjustment of the profile of the grating surface, an in-position LTP system was used to measure the slope of this profile. With the measured slope function, the target slope function and the set of actuator-response functions as inputs, one can use an iterative algorithm to deduce a set of actuator incremental values to adjust the surface profile to match the target profile. The details of the adjustment are reported elsewhere (Kao et al., 2019[Kao, K. Y., Hua, C. Y., Fung, H. S., Lin, S. W., Chao, H. Y., Yeh, S. C., Chung, S. C., Huang, D. J. & Chen, C. T. (2019). AIP Conf. Proc. 2054, 060005.]). Fig. 7[link] plots typical LTP measurement results, including the measured slope function and its target slope polynomial function, the difference between them and the measured surface profile of the grating. The slope difference between the measured and target functions was minimized to a root-mean-square (r.m.s.) value less than 0.25 µrad for both the AGM and the AGS grating. Because of the limited precision and accuracy of the in-position LTP instrument, the obtained slope difference is larger than the slope error of 0.1 µrad r.m.s. of the polished surface provided by the manufacturer.

[Figure 7]
Figure 7
Slope measurements of an active grating with an in-position LTP. (a) The measured grating slope and its target polynomial slope function yt′(x) = c1x + 2c2x + 3c3x2 + 4c4x3, in which c1 = 1.95 × 10−5, c2 = 9.08 × 10−6, c3 = −3.85 × 10−9 and c4 = 5.94 × 10−14. The function yt′(x) is vertically offset for clarity. (b) The difference between the measured and target slope functions. The r.m.s. of this difference is 0.23 µrad. (c) The profile function of the grating surface obtained from the integration of the measured slope, with the height at the grating centre defined as zero.

3.3. Resolution optimization

We measured the linewidth of the elastic scattering from a W/B4C multilayer (ML) to tune the energy resolution. At the beginning, the grating surface profile was adjusted using the 25 actuators to match the initial target profile from the theoretical simulations of the designed beamline. Because of the alignment imperfections of the beamline, various sets of polynomial coefficients, i.e. [c2, c3, c4], were tested as the target profile to optimize the energy resolution. Fig. 8[link] shows the elastic scattering of 530 eV soft X-rays from the ML near a reflection condition with an entrance slit of 4 µm and an exit slit of 100 µm, which corresponds to a bandwidth of 0.5 eV of incident soft X-rays. The best achieved spectral resolution was 12.4 meV, which is defined as the minimum separation between two spectral lines when they can be resolved through a criterion that the core-to-wing ratio is smaller than the ratio of 0.9272 of two identical Gaussian functions separated by their FWHM.

[Figure 8]
Figure 8
Elastic scattering spectra of a W/B4C multilayer measured in a near reflection condition at 530 eV. The measured scattering spectrum, which has an FWHM of 14.8 meV, is plotted with black circles, together with a green curve showing its Fourier filtered function. The spectrum plotted along with a blue curve for its Fourier filtered function is a duplicate of this measured spectrum. When the separation between these two spectra is 12.4 meV, the core-to-wing ratio in their summation plotted together with a red curve for its Fourier filtered function is 0.926. The measured spectrum was obtained without applying any pixel elimination with an intensity threshold on the 2D image recorded in 1 s.

3.4. Photon flux on the sample

As RIXS is a photon-demanding technique, a large photon flux on the sample is essential to achieve high-resolution measurements. The AGM–AGS scheme allows us to increase the bandwidth of the incident photons on the sample while maintaining the energy resolution. To verify this condition, we measured the energy resolution of the elastic scattering of 530 eV soft X-rays with the exit-slit opening set at 50, 100 and 200 µm. The measured photon flux increased linearly from 3.6 × 1012 to 1.3 × 1013 photons s−1 with increased exit-slit opening. The measured energy resolution remained nearly unchanged, while the exit-slit opening increased. These data are consistent with our previous results (Lai et al., 2014[Lai, C. H., Fung, H. S., Wu, W. B., Huang, H. Y., Fu, H. W., Lin, S. W., Huang, S. W., Chiu, C. C., Wang, D. J., Huang, L. J., Tseng, T. C., Chung, S. C., Chen, C. T. & Huang, D. J. (2014). J. Synchrotron Rad. 21, 325-332.]) and demonstrate that the AGM–AGS scheme is also applicable to high-resolution soft X-ray RIXS.

3.5. Phonon excitations of superconducting cuprates

The dressing of electrons by phonons in a material plays an important role in its novel electronic properties; in particular, the coupling between electrons and phonons in HTSC is pivotal yet remains under debate. RIXS has the potential to measure the strength of the coupling of electrons to lattice excitations. It has been shown theoretically that high-resolution RIXS provides direct, element-specific and momentum-resolved information about the electron–phonon coupling (Ament et al., 2011b[Ament, L. J. P., van Veenendaal, M. & van den Brink, J. (2011b). EPL (Europhysics Lett), 95, 27008.]; Devereaux et al., 2016[Devereaux, T. P., Shvaika, A. M., Wu, K., Wohlfeld, K., Jia, C. J., Wang, Y., Moritz, B., Chaix, L., Lee, W.-S., Shen, Z.-X., Ghiringhelli, G. & Braicovich, L. (2016). Phys. Rev. X, 6, 041019.]). Here, we present O K-edge RIXS measurements on La1.88Sr0.12CuO4 (LSCO) to reveal its momentum-resolved phonon excitations.

La2–xSrxCuO4 is a doped Mott insulator (Imada et al., 1998[Imada, M., Fujimori, A. & Tokura, Y. (1998). Rev. Mod. Phys. 70, 1039-1263.]) com­posed of CuO2 conducting layers with relatively weak coupling between the layers. Its mother com­pound La2CuO4 is an insulating antiferromagnet, in which the Cu ion has a 3d9 electronic configuration, with a hole of symmetry x2 − y2. The system is a charge-transfer insulator because strong correlation effects split the conduction band into the upper and lower Hubbard bands. These correlation effects manifest themselves in O K-edge X-ray absorption in which an electron is excited from the 1s core level to the 2p band. The polarization-dependent XAS measurements revealed that the doped holes are distributed mainly throughout the O 2px,y orbital in the CuO2 plaquette and are hybridized with the Cu [3d_{x^{2} - y^{2}}] orbital to form a spin singlet [3d^{\,9}\underline{L}], termed a Zhang–Rice singlet (ZRS), in which [\underline{L}] denotes a ligand hole (Zhang & Rice, 1988[Zhang, F. C. & Rice, T. M. (1988). Phys. Rev. B, 37, 3759-3761.]; Chen et al., 1992[Chen, C. T., Tjeng, L. H., Kwo, J., Kao, H. L., Rudolf, P., Sette, F. & Fleming, R. M. (1992). Phys. Rev. Lett. 68, 2543-2546.]). In the XAS of La2CuO4, there exists a prepeak at the absorption edge arising from the O 2p band hybridized with the upper Hubbard band (UHB) of Cu 3d. In a hole-doped cuprate, a lower-energy XAS feature resulting from ZRS emerges and grows linearly with hole concentration in the under-doped regime, as plotted in Fig. 9[link](a) (Chen et al., 1991[Chen, C. T., Sette, F., Ma, Y., Hybertsen, M. S., Stechel, E. B., Foulkes, W. M. C., Schulter, M., Cheong, S.-W., Cooper, A. S., Rupp, L. W. Jr, Batlogg, B., Soo, Y. L., Ming, Z. H., Krol, A. & Kao, Y. H. (1991). Phys. Rev. Lett. 66, 104-107.]). That is, hole doping manifests itself in the spectral-weight transfer from UHB to ZRS as a consequence of electron correlations. The existence of ZRS enables measurements of phonon excitations using O K-edge RIXS. Fig. 9[link](b) shows the RIXS spectrum of LSCO measured with incident X-rays set at 528.4 eV, i.e. the ZRS feature. The inset of Fig. 9[link](b) illustrates the scattering geometry. The in-plane momentum transfer q was along the direction (π, 0), with q = 0.12[{{2\pi} \over {a}}]; the out-of-plane momentum transfer was q = 1.0[{{2\pi} \over {c}}], in which a and c denote the lattice parameters. The RIXS spectrum in Fig. 9[link](b) shows an intense elastic peak and pro­noun­ced phonon excitations. We fitted the measured RIXS spectrum with four peaks, i.e. an elastic scattering peak and three phonon peaks, using Voigt functions. The observed three phonon excitations at 75, 45 and 20 meV originate from the half-breathing mode (Cohen et al., 1989[Cohen, R. E., Pickett, W. E. & Krakauer, H. (1989). Phys. Rev. Lett. 62, 831-834.]; Chaplot et al., 1995[Chaplot, S., Reichardt, W., Pintschovius, L. & Pyka, N. (1995). Phys. Rev. B, 52, 7230-7242.]; McQueeney et al., 1999[McQueeney, R. J., Petrov, Y., Egami, T., Yethiraj, M., Shirane, G. & Endoh, Y. (1999). Phys. Rev. Lett. 82, 628-631.]; Fukuda et al., 2005[Fukuda, T., Mizuki, J., Ikeuchi, K., Yamada, K., Baron, A. & Tsutsui, S. (2005). Phys. Rev. B, 71, 060501.]; Giustino et al., 2008[Giustino, F., Cohen, M. L. & Louie, S. G. (2008). Nature, 452, 975-978.]), A1g/B1g buckling modes (Cohen et al., 1989[Cohen, R. E., Pickett, W. E. & Krakauer, H. (1989). Phys. Rev. Lett. 62, 831-834.]; McQueeney et al., 1999[McQueeney, R. J., Petrov, Y., Egami, T., Yethiraj, M., Shirane, G. & Endoh, Y. (1999). Phys. Rev. Lett. 82, 628-631.]; Burns & Dacol, 1990[Burns, G. & Dacol, F. (1990). Phys. Rev. B, 41, 4747-4748.]; Giustino et al., 2008[Giustino, F., Cohen, M. L. & Louie, S. G. (2008). Nature, 452, 975-978.]; Sugai et al., 2013[Sugai, S., Takayanagi, Y., Hayamizu, N., Muroi, T., Shiozaki, R., Nohara, J., Takenaka, K. & Okazaki, K. (2013). J. Phys. Condens. Matter, 25, 475701.]; Johnston et al., 2010[Johnston, S., Vernay, F., Moritz, B., Shen, Z.-X., Nagaosa, N., Zaanen, J. & Devereaux, T. (2010). Phys. Rev. B, 82, 064513.]) and the vibrational mode involved with La/Sr (Sugai, 1989[Sugai, S. (1989). Phys. Rev. B, 39, 4306-4315.]; Kim et al., 2017[Kim, G., Christiani, G., Logvenov, G., Choi, S., Kim, H.-H., Minola, M. & Keimer, B. (2017). Phys. Rev. Mater. 1, 054801.]), respectively. These high-resolution RIXS measurements open a new opportunity to investigate the coupling between phonons and charge-density waves in HTSC.

[Figure 9]
Figure 9
(a) The O K-edge XAS of LSCO at 23 K measured with the fluorescence yield mode with X-rays of σ polarization. (b) RIXS spectrum of LSCO with the incident energy tuned to ZRS with q = 0.12[{{2\pi} \over {a}}] and q = 1.0[{{2\pi} \over {c}}]. The duration of exposure to record the RIXS data shown in the black circles was 2 h. The red curve is the summation of the fitted com­ponents and a linear background function. Other colour curves are fitted Voigt functions for elastic scattering and phonon excitations. The inset illustrates the scattering geometry with q = kinkout, in which kin and kout are incident and scattered wavevectors, respectively. a and c are crystallographic axes of LSCO. The scattering angle 2θ was 150°.

4. Summary and future plan

We have designed, constructed and commissioned a high-resolution and highly efficient RIXS beamline based on the energy-com­pensation principle of grating dispersion. The achieved resolving power was 4.2 × 104 at a photon energy of 530 eV, with the bandwidth of the incident soft X-rays set at 0.5 eV. With this high resolution, we observed three phonon excitations of LSCO. A new in-vacuum LTP instrument with a high precision of 0.005 µrad r.m.s. has been developed recently. It will be installed in the beamline to greatly improve the accuracy of the measured slope function of the grating surface, aiming to reach an ultra-high resolving power of 105. For our AGM–AGS design, the resolution is limited to the contribution of the grating slope error. Fig. 10[link] presents the estimated resolution contributed from the slope error of the grating, showing that such a goal is achievable if we can minimize the grating slope error to be smaller than 0.05 µrad r.m.s. We also plan to install a polarimeter for the polarization analysis of scattered soft X-rays and to develop a high-spatial-resolution and highly efficient soft X-ray 2D detector.

[Figure 10]
Figure 10
Estimated resolution contributed from the slope error of the grating as a function of photon energy. Red squares and blue circles represent the combined resolution of the AGM and AGS gratings contributed from slope errors of 0.05 and 0.25 µrad r.m.s., respectively. The slope error contribution [\Delta\lambda_{\rm SE}] is calculated through the grating equation [m\lambda = d(\sin\alpha+\sin\beta)], i.e. [\Delta\lambda_{\rm SE} = d(\cos\alpha+\cos\alpha)\Delta\phi/m], where λ and m are the wavelength and grating diffraction order, respectively, α and β are the grating incidence and diffraction angles from the surface normal, respectively, Δϕ is the grating slope error and d is the grating period. Black solid lines serve as a guide to the eye.

Acknowledgements

We acknowledge the NSRRC staff for technical support; in particular, we thank D. J. Wang for useful discussions. We also acknowledge L. Y. Chang for his help during the commissioning stage. Thanks are due to The Ministry of Science and Technology of Taiwan and MOST for providing partial support to this work.

Funding information

Funding for this research was provided by: The Ministry of Science and Technology of Taiwan (grant No. 103-2112-M-213-008-MY3); MOST (grant No. 108-2923-M-213-001).

References

First citationAment, L. J. P., van Veenendaal, M., Devereaux, T. P., Hill, J. P. & van den Brink, J. (2011a). Rev. Mod. Phys. 83, 705–767.  Web of Science CrossRef CAS Google Scholar
First citationAment, L. J. P., van Veenendaal, M. & van den Brink, J. (2011b). EPL (Europhysics Lett), 95, 27008.  Google Scholar
First citationBraicovich, L., van den Brink, J., Bisogni, V., Sala, M. M., Ament, L. J. P., Brookes, N. B., De Luca, G. M., Salluzzo, M., Schmitt, T., Strocov, V. N. & Ghiringhelli, G. (2010). Phys. Rev. Lett. 104, 077002.  Web of Science CrossRef PubMed Google Scholar
First citationBrookes, N. B., Yakhou-Harris, F., Kummer, K., Fondacaro, A., Cezar, J. C., Betto, D., Velez-Fort, E., Amorese, A., Ghiringhelli, G., Braicovich, L., Barrett, R., Berruyer, G., Cianciosi, F., Eybert, L., Marion, P., van der Linden, P. & Zhang, L. (2018). Nucl. Instrum. Methods Phys. Res. A, 903, 175–192.  Web of Science CrossRef CAS Google Scholar
First citationBurns, G. & Dacol, F. (1990). Phys. Rev. B, 41, 4747–4748.  CrossRef CAS Web of Science Google Scholar
First citationChang, C.-H., Hwang, C. S., Chang, C. H., Chen, H. H., Huang, M. H., Lin, F.-Y., Huang, J.-C. & Wu, C. M. (2010). Proceedings of the First International Particle Accelerator Conference (IPAC2010), 23–28 May 2010, Kyoto, Kyoto, Japan, pp. 3183–3185. WEPD042.  Google Scholar
First citationChaplot, S., Reichardt, W., Pintschovius, L. & Pyka, N. (1995). Phys. Rev. B, 52, 7230–7242.  CrossRef CAS Web of Science Google Scholar
First citationChen, C. T. (1987). Nucl. Instrum. Methods Phys. Res. A, 256, 595–604.  CrossRef Web of Science Google Scholar
First citationChen, C. T. & Sette, F. (1989). Rev. Sci. Instrum. 60, 1616–1621.  CrossRef CAS Web of Science Google Scholar
First citationChen, C. T., Sette, F., Ma, Y., Hybertsen, M. S., Stechel, E. B., Foulkes, W. M. C., Schulter, M., Cheong, S.-W., Cooper, A. S., Rupp, L. W. Jr, Batlogg, B., Soo, Y. L., Ming, Z. H., Krol, A. & Kao, Y. H. (1991). Phys. Rev. Lett. 66, 104–107.  CrossRef PubMed CAS Web of Science Google Scholar
First citationChen, C. T., Tjeng, L. H., Kwo, J., Kao, H. L., Rudolf, P., Sette, F. & Fleming, R. M. (1992). Phys. Rev. Lett. 68, 2543–2546.  CrossRef PubMed CAS Web of Science Google Scholar
First citationChiuzbăian, S. G., Hague, C. F., Avila, A., Delaunay, R., Jaouen, N., Sacchi, M., Polack, F., Thomasset, M., Lagarde, B., Nicolaou, A., Brignolo, S., Baumier, C., Lüning, J. & Mariot, J. M. (2014). Rev. Sci. Instrum. 85, 043108.  Web of Science PubMed Google Scholar
First citationChuang, Y.-D., Shao, Y.-C., Cruz, A., Hanzel, K., Brown, A., Frano, A., Qiao, R., Smith, B., Domning, E., Huang, S.-W., Wray, L. A., Lee, W. S., Shen, Z. X., Devereaux, T. P., Chiou, J. W., Pong, W. F., Yashchuk, V. V., Gullikson, E., Reininger, R., Yang, W., Guo, J., Duarte, R. & Hussain, Z. (2017). Rev. Sci. Instrum. 88, 013110.  Web of Science CrossRef PubMed Google Scholar
First citationChung, T. Y., Chang, C. H., Jan, J. C., Hwang, C. S., Chiu, M. S., Kuo, C. C., Chang, C. H., Tseng, F. H., Huang, J. C., Liu, Y. C., et al. (2016). Proceedings of the 7th International Particle Accelerator Conference (IPAC2016), 8–13 May 2016, Busan, Korea, pp. 4054–4056. THPOW049  Google Scholar
First citationCohen, R. E., Pickett, W. E. & Krakauer, H. (1989). Phys. Rev. Lett. 62, 831–834.  CrossRef PubMed CAS Web of Science Google Scholar
First citationDallera, C., Puppin, E., Trezzi, G., Incorvaia, N., Fasana, A., Braicovich, L., Brookes, N. B. & Goedkoop, J. B. (1996). J. Synchrotron Rad. 3, 231–238.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationDevereaux, T. P., Shvaika, A. M., Wu, K., Wohlfeld, K., Jia, C. J., Wang, Y., Moritz, B., Chaix, L., Lee, W.-S., Shen, Z.-X., Ghiringhelli, G. & Braicovich, L. (2016). Phys. Rev. X, 6, 041019.  Google Scholar
First citationDinardo, M. E., Piazzalunga, A., Braicovich, L., Bisogni, V., Dallera, C., Giarda, K., Marcon, M., Tagliaferri, A. & Ghiringhelli, G. (2007). Nucl. Instrum. Methods Phys. Res. A, 570, 176–181.  Web of Science CrossRef CAS Google Scholar
First citationFabbris, G., Meyers, D., Okamoto, J., Pelliciari, J., Disa, A., Huang, Y., Chen, Z.-Y., Wu, W., Chen, C., Ismail-Beigi, S., Ahn, C. H., Walker, F. J., Huang, D. J., Schmitt, T. & Dean, M. P. (2016). Phys. Rev. Lett. 117, 147401.  Web of Science CrossRef PubMed Google Scholar
First citationFukuda, T., Mizuki, J., Ikeuchi, K., Yamada, K., Baron, A. & Tsutsui, S. (2005). Phys. Rev. B, 71, 060501.  Web of Science CrossRef Google Scholar
First citationFung, H. S., Chen, C. T., Huang, L. J., Chang, C. H., Chung, S. C., Wang, D. J., Tseng, T. C. & Tsang, K. L. (2004). AIP Conf. Proc. 705, 655–658.  CrossRef Google Scholar
First citationGhiringhelli, G., Braicovich, L., Schmitt, T. & Strocov, V. (2012). Synchrotron Radiat. News, 25(4), 16–22.  CrossRef Google Scholar
First citationGhiringhelli, G., Brookes, N. B., Annese, E., Berger, H., Dallera, C., Grioni, M., Perfetti, L., Tagliaferri, A. & Braicovich, L. (2004). Phys. Rev. Lett. 92, 117406.  Web of Science CrossRef PubMed Google Scholar
First citationGhiringhelli, G., Matsubara, M., Dallera, C., Fracassi, F., Tagliaferri, A., Brookes, N. B., Kotani, A. & Braicovich, L. (2006a). Phys. Rev. B, 73, 035111.  Web of Science CrossRef Google Scholar
First citationGhiringhelli, G., Piazzalunga, A., Dallera, C., Trezzi, G., Braicovich, L., Schmitt, T., Strocov, V. N., Betemps, R., Patthey, L., Wang, X. & Grioni, M. (2006b). Rev. Sci. Instrum. 77, 113108.  Web of Science CrossRef Google Scholar
First citationGhiringhelli, G., Tagliaferri, A., Braicovich, L. & Brookes, N. B. (1998). Rev. Sci. Instrum. 69, 1610–1615.  Web of Science CrossRef CAS Google Scholar
First citationGiustino, F., Cohen, M. L. & Louie, S. G. (2008). Nature, 452, 975–978.  Web of Science CrossRef PubMed CAS Google Scholar
First citationHepting, M., Chaix, L., Huang, E. W., Fumagalli, R., Peng, Y. Y., Moritz, B., Kummer, K., Brookes, N. B., Lee, W. C., Hashimoto, M., Sarkar, T., He, J. F., Rotundu, C. R., Lee, Y. S., Greene, R. L., Braicovich, L., Ghiringhelli, G., Shen, Z. X., Devereaux, T. P. & Lee, W. S. (2018). Nature, 563, 374–378.  Web of Science CrossRef CAS PubMed Google Scholar
First citationHuang, D. J. & Chen, C. T. (2018). Synchrotron Radiat. News, 31(2), 3–6.  CrossRef Google Scholar
First citationHuang, H., Jia, C., Chen, Z., Wohlfeld, K., Moritz, B., Devereaux, T., Wu, W., Okamoto, J., Lee, W., Hashimoto, M., He, Y., Shen, Z. X., Yoshida, Y., Eisaki, H., Mou, C. Y., Chen, C. T. & Huang, D. J. (2016). Sci. Rep. 6, 19657.  Web of Science CrossRef PubMed Google Scholar
First citationHuang, H.-Y., Chen, Z.-Y., Wang, R.-P., de Groot, F. M., Wu, W.-B., Okamoto, J., Chainani, A., Singh, A., Li, Z.-Y., Zhou, J.-S., Jeng, H. T., Guo, G. Y., Park, J. G., Tjeng, L. H., Chen, C. T. & Huang, D. J. (2017). Nat. Commun. 8, 15929.  Web of Science CrossRef PubMed Google Scholar
First citationImada, M., Fujimori, A. & Tokura, Y. (1998). Rev. Mod. Phys. 70, 1039–1263.  Web of Science CrossRef CAS Google Scholar
First citationJarrige, I., Bisogni, V., Zhu, Y., Leonhardt, W. & Dvorak, J. (2018). Synchrotron Radiat. News, 31(2), 7–13.  CrossRef Google Scholar
First citationJohnston, S., Vernay, F., Moritz, B., Shen, Z.-X., Nagaosa, N., Zaanen, J. & Devereaux, T. (2010). Phys. Rev. B, 82, 064513.  Web of Science CrossRef Google Scholar
First citationKao, K. Y., Hua, C. Y., Fung, H. S., Lin, S. W., Chao, H. Y., Yeh, S. C., Chung, S. C., Huang, D. J. & Chen, C. T. (2019). AIP Conf. Proc. 2054, 060005.  Google Scholar
First citationKim, G., Christiani, G., Logvenov, G., Choi, S., Kim, H.-H., Minola, M. & Keimer, B. (2017). Phys. Rev. Mater. 1, 054801.  Web of Science CrossRef Google Scholar
First citationKotani, A. & Shin, S. (2001). Rev. Mod. Phys. 73, 203–246.  Web of Science CrossRef CAS Google Scholar
First citationLai, C. H., Fung, H. S., Wu, W. B., Huang, H. Y., Fu, H. W., Lin, S. W., Huang, S. W., Chiu, C. C., Wang, D. J., Huang, L. J., Tseng, T. C., Chung, S. C., Chen, C. T. & Huang, D. J. (2014). J. Synchrotron Rad. 21, 325–332.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationLee, W., Lee, J., Nowadnick, E., Gerber, S., Tabis, W., Huang, S., Strocov, V., Motoyama, E., Yu, G., Moritz, B., Huang, H. Y., Wang, R. P., Huang, Y. B., Wu, W. B., Chen, C. T., Huang, D. J., Greven, M., Schmitt, T., Shen, Z. X. & Devereaux, T. P. (2014). Nat. Phys. 10, 883–889.  Web of Science CrossRef CAS Google Scholar
First citationLe Tacon, M., Ghiringhelli, G., Chaloupka, J., Sala, M. M., Hinkov, V., Haverkort, M. W., Minola, M., Bakr, M., Zhou, K. J., Blanco-Canosa, S., Monney, C., Song, Y. T., Sun, G. L., Lin, C. T., De Luca, G. M., Salluzzo, M., Khaliullin, G., Schmitt, T., Braicovich, L. & Keimer, B. (2011). Nat. Phys. 7, 725–730.  Web of Science CrossRef CAS Google Scholar
First citationLin, S. W., Wang, D. J., Fu, H. W., Tsai, H. M., Hua, C. Y., Kuo, C. Y., Hsu, M. Y., Kao, K. Y., Yin, G. C., Fung, H. S., Perng, S. Y. & Chang, C. F. (2019). Rev. Sci. Instrum. 90, 021716.  Web of Science CrossRef PubMed Google Scholar
First citationMcQueeney, R. J., Petrov, Y., Egami, T., Yethiraj, M., Shirane, G. & Endoh, Y. (1999). Phys. Rev. Lett. 82, 628–631.  Web of Science CrossRef CAS Google Scholar
First citationMiao, H., Fumagalli, R., Rossi, M., Lorenzana, J., Seibold, G., Yakhou-Harris, F., Kummer, K., Brookes, N., Gu, G., Braicovich, L., Ghiringhelli, G. & Dean, M. (2019). Phys. Rev. X, 9, 031042.  Google Scholar
First citationMinola, M., Dellea, G., Gretarsson, H., Peng, Y., Lu, Y., Porras, J., Loew, T., Yakhou, F., Brookes, N., Huang, Y., Pelliciari, J., Schmitt, T., Ghiringhelli, G., Keimer, B., Braicovich, L. & Le Tacon, M. (2015). Phys. Rev. Lett. 114, 217003.  Web of Science CrossRef PubMed Google Scholar
First citationSchmitt, T., Strocov, V. N., Zhou, K. J., Schlappa, J., Monney, C., Flechsig, U. & Patthey, L. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 38–46.  Web of Science CrossRef CAS Google Scholar
First citationSchülke, W. (1989). Nucl. Instrum. Methods Phys. Res. A, 280, 338–348.  Google Scholar
First citationSchulz, C., Lieutenant, K., Xiao, J., Hofmann, T., Wong, D. & Habicht, K. (2020). J. Synchrotron Rad. 27, 238–249.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSugai, S. (1989). Phys. Rev. B, 39, 4306–4315.  CrossRef CAS Web of Science Google Scholar
First citationSugai, S., Takayanagi, Y., Hayamizu, N., Muroi, T., Shiozaki, R., Nohara, J., Takenaka, K. & Okazaki, K. (2013). J. Phys. Condens. Matter, 25, 475701.  Web of Science CrossRef PubMed Google Scholar
First citationSuzuki, H., Minola, M., Lu, Y., Peng, Y., Fumagalli, R., Lefrançois, E., Loew, T., Porras, J., Kummer, K., Betto, D., Ishida, S., Eisaki, H., Hu, C., Zhou, X., Haverkort, M. W., Brookes, N. B., Braicovich, L., Ghiringhelli, G., Le Tacon, M. & Keimer, B. (2018). NPJ Quantum Mater. 3, 65.  Google Scholar
First citationTomiyasu, K., Okamoto, J., Huang, H.-Y., Chen, Z.-Y., Sinaga, E. P., Wu, W.-B., Chu, Y.-Y., Singh, A., Wang, R.-P., de Groot, F., Chainani, A., Ishihara, S., Chen, C. T. & Huang, D. J. (2017). Phys. Rev. Lett. 119, 196402.  Web of Science CrossRef PubMed Google Scholar
First citationYamamoto, S., Senba, Y., Tanaka, T., Ohashi, H., Hirono, T., Kimura, H., Fujisawa, M., Miyawaki, J., Harasawa, A., Seike, T., Takahashi, S., Nariyama, N., Matsushita, T., Takeuchi, M., Ohata, T., Furukawa, Y., Takeshita, K., Goto, S., Harada, Y., Shin, S., Kitamura, H., Kakizaki, A., Oshima, M. & Matsuda, I. (2014). J. Synchrotron Rad. 21, 352–365.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationZhang, F. C. & Rice, T. M. (1988). Phys. Rev. B, 37, 3759–3761.  CrossRef CAS Web of Science Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.