view article

Figure 4
Optimization of ICCD gate timing parameters. (a) The ICCD collection efficiency, η(t′, 140 ns), of a 140 ns duration gate with varying gate delay from first light arrival at the ICCD photocathode, t′, as predicted by equation (3)[link]. (b) The relative collection efficiencies, F1(t′) and F−1(t′), of the previous X-ray pulse and next X-ray pulse, respectively, as a function of gate delay. F1(t′) represents the relative weight of the after-image due to the previous pulse in comparison to the weight of the pulse of interest.