view article

Figure 5
(a, b) The homogeniety of the Ge(111) C- and D-crystals was tested by conducting rocking scans of the (555) reflection in 5 mm steps along a ∼100 mm region. For clarity the maximum of each curve was normalized to unity. (c) Peak center and (d) FWHM as a function of crystal position were extracted by fitting the curves in (a, b). The dashed horizontal line in (d) is the expected theoretical value.