view article

Figure 2
(a) Measured vertical and horizontal profiles of the micro-focused beam obtained using a knife-edge-scan setup giving a focus beam 4.1 µm (V) × 8.5 µm (H) in size. The red rhombuses are the measured data whereas the solid lines are the fitted Gaussian profiles. The data were collected with ∼1012 photons s−1 at a photon energy of 12.659 keV. (b) The 11C Micro-MX experimental hutch showing a six-axis robot arm, an MD2S diffractometer and a Pilatus3 6M detector.