Performance and characterization of the FinEstBeAMS beamline at the MAX IV Laboratory
aMAX IV Laboratory, Lund University, PO Box 118, SE-22100 Lund, Sweden, bNano and Molecular Systems Research Unit, University of Oulu, PO Box 3000, FI-90014 Oulu, Finland, cInstitute of Physics, University of Tartu, W. Ostwald Street 1, EE-51014 Tartu, Estonia, dComputational Physics Laboratory, Tampere University, PO Box 692, FI-33014 Tampere, Finland, eInstitute of Solid State Physics, University of Latvia, 8 Kengaraga iela, LV-1063 Riga, Latvia, fDepartment of Physics and Astronomy, University of Turku, FI-20014 Turku, Finland, gBESSY-II, Helmholtz-Zentrum Berlin für Materialien und Energie, Albert-Einstein-Straße 15, 12489 Berlin, Germany, and hSurface Science Group, Laboratory of Photonics, Physics Unit, Tampere University, PO Box 692, FI-33014 Tampere, Finland
*Correspondence e-mail: kirill.chernenko@maxiv.lu.se
FinEstBeAMS (Finnish–Estonian Beamline for Atmospheric and Materials Sciences) is a multidisciplinary beamline constructed at the 1.5 GeV storage ring of the MAX IV synchrotron facility in Lund, Sweden. The beamline covers an extremely wide photon energy range, 4.5–1300 eV, by utilizing a single elliptically polarizing undulator as a radiation source and a single grazing-incidence plane grating monochromator to disperse the radiation. At photon energies below 70 eV the beamline operation relies on the use of optical and thin-film filters to remove higher-order components from the monochromated radiation. This paper discusses the performance of the beamline, examining such characteristics as the quality of the gratings, photon energy calibration, photon energy resolution, available
polarization quality and focal spot size.Keywords: MAX IV; photon energy resolution; photon flux; plane grating monochromator; beam polarization.
1. Introduction
The FinEstBeAMS beamline at the MAX IV Laboratory was designed to fulfil the various needs of the Estonian, Finnish and other Nordic user communities in gas-phase electron and ion spectroscopies, surface science, and et al., 2017).
research, while also providing a modern platform to any users of synchrotron radiation with their own experimental end stations. Therefore, the new beamline had to afford an extensive photon energy range from ultraviolet to soft X-rays, good performance in terms of photon energy resolution and (often needed in gas-phase electron spectroscopy), reasonably small focal spot sizes to allow moderate spatial resolution in surface studies, and a possibility of probing samples with light of variable polarization. A further restriction in the design was to have only one undulator and one monochromator despite the extended operation range. A solution for how these demanding goals could be achieved was presented in the design paper of the FinEstBeAMS beamline (PärnaThe photon source of FinEstBeAMS is an elliptically polarizing undulator (EPU). It has a very long magnetic period, 95.2 mm, for the effective production of photons in the ultraviolet spectral region at the 1.5 GeV storage ring. The long magnetic period inevitably compromises the performance of the beamline in the soft X-ray region but usable 10 photons per second) is available up to 1000 eV with a good (ca 5000). The plane grating monochromator of the FinEstBeAMS beamline is of the SX700 type (Petersen & Baumgärtel, 1980). At present, two plane Au-coated gratings are available. The first one, with 600 lines mm−1 line density, can be used at photon energies from 15 to 1300 eV. The second plane grating operates in the photon energy range from 4.5 to 50 eV and has an exceptionally sparse line density of 92 lines mm−1. The combination of the low-line-density grating and grazing-incidence optics on the beamline results in poor high-order suppression in monochromated light. Therefore, each of the two beamline branches contains specific filter units, where six different filters can be mounted, purifying the radiation further for low-photon-energy experiments.
(>10At present, FinEstBeAMS has three experimental end stations: the gas-phase end station (GPES) for electron and ion spectroscopy and for photoelectron–photoion coincidence spectroscopy of low-density matter (Kooser et al., 2020), the solid-state end station (SSES) for photoelectron and X-ray absorption spectroscopy of surfaces and interfaces, and the endstation (PLES) for luminescence spectroscopy of solids (Pankratov et al., 2019). As the vacuum requirements are significantly different in surface and gas-phase studies, the SSES and GPES have been placed on separate branch lines, which are accordingly called the solid-state and gas-phase branches. The PLES has been designed to be a movable end station and, when in use, it is installed on the gas-phase branch (in which case the GPES is moved upstream and the photon beam passes through it). The GPES and PLES have been in regular user operation since April 2019. The SSES accepts its first regular users in 2021.
In this paper, we give a short overview of the beamline design, report the results of the optical characterization of the plane gratings, and present the results of the beamline performance characterization, including photon energy resolution and
in the accessible energy range, accuracy of the photon energy calibration, quality of the polarization properties of the photon beam, and focal beam spot size. We also discuss how the achieved performance meets the estimations made on the basis of optical element characterization.2. Overview of the beamline
The optical layout of FinEstBeAMS is shown in Fig. 1. The photon source, an elliptically polarized undulator (EPU) of the APPLE II type (Sasaki et al., 1993), was built in-house (Thiel et al., 2018; Tarawneh et al., 2019) based on the design used at Helmholtz-Zentrum Berlin (HZB) (Bahrdt et al., 2001). It contains four arrays of magnets that can be shifted independently in the longitudinal direction in order to control the radiation polarization. The beamline design follows the concept of the collimated plane grating monochromator (Follath, 2001). A side-cooled toroidal mirror (M1) is the first optical element in the beamline, mounted 12.00 m from the centre of the undulator. It collimates the beam in both directions, horizontally and vertically. After M1, two pairs (vertical and horizontal) of precise baffles (BAFF) are used to limit the monochromator aperture (acceptance angle). The monochromator was manufactured by FMB Feinwerk und Messtechnik GmbH, Berlin. It contains an internally cooled plane mirror (M2) and three slots for side-cooled plane gratings. Currently, two Au-coated plane gratings are installed (PG1, 600 lines mm−1 and PG2, 92 lines mm−1). The selected grating disperses incoming radiation in the vertical plane and – with the adjusted position and incidence angle of the M2 mirror – guides the central energy of the outgoing radiation in a direction parallel to the incoming beam. The branch line used in the experiments is selected by inserting either of the two toroidal focusing mirrors, M3GP or M3SS, into the beam path. That mirror focuses the dispersed radiation at the exit slit, which is located 6.00 m downstream from the focusing mirror. After the exit slit, each branch line contains a higher-order suppression unit (HOS), where two optical filters (fused silica, MgF2) and four thin-film metal filters (In, Sn, Mg, Al) are mounted on two linear manipulators; i.e. each manipulator has a holder for three filters. These sets of filters allow the suppression of higher-order radiation at photon energies below 72 eV. An ellipsoidal mirror (M4GP or M4SS) refocuses the monochromated radiation to the end station in each branch. It deflects the beam sideways, keeping it in the same horizontal plane. The entrance and exit arms of the refocusing mirror have been chosen to be 15% longer in the gas-phase branch than in the solid-state branch, so that the end stations are physically not mounted exactly side by side.
The beamline utilizes a standard MAX IV control system based on IcePAP motion controllers (Sjöblom et al., 2016) and Tango software packages with the Sardana framework (Coutinho et al., 2011).
3. Optics characterization
The mirrors and grating blanks were provided by various commercial vendors, whereas the gratings were manufactured at the grating laboratory of HZB (Siewert et al., 2018) using a mechanical diamond ruling method. First, the grating blanks were covered with a ruling gold layer and the line pattern was applied by a mechanical ruling process. The obtained pattern was then etched by an ion beam, forming a blazed surface. In the last step, the surface was coated with a reflective gold layer, which provides optimal efficiency in the required photon energy range. The design goal of covering the whole photon energy range with a single monochromator demands an extremely sparsely ruled grating for low photon energies. The optimal blaze angle for low-energy operation can be several degrees, which, combined with a low line density, results in a very large material deformation applied by the grating ruling machine. The ruling gold layer for the 92 lines mm−1 grating had to be about three times thicker than for the 600 lines mm−1 grating in order to allow ruling of a blaze angle of approximately 8°. The profile height of the ruled lines was increased from 225 nm for the 600 lines mm−1 grating to >1500 nm for the 92 lines mm−1 grating, resulting in more trenching during pattern transfer into silicon. This leads to rounding of a large part of the blaze facet, and identifying the final blaze angle by linear assumption does not result in the real grating profile.
Optical components installed in the beamline have to withstand the powerful beam generated by the EPU at FinEstBeAMS. Their quality is essential to ensure all the required beam characteristics, like a small focal spot size, polarization, spectral resolution and a sufficiently high ex situ at the at-wavelength metrology facility at BESSY-II (Schäfers et al., 2016) of HZB. The slope errors, and shape parameters like the radii of curvature, were studied by means of the BESSY-NOM (Siewert et al., 2004). The NOM allows measurement of the slope error of reflective surfaces for the spatial frequency range from about 1 mm up to a full aperture length (Siewert et al., 2016), which corresponds to a range of spatial frequencies sufficient to cause a deformation of the wavefront and thus have a direct impact on the coherent part of the photons. The mid- and high-spatial-frequency shape error was measured by means of a white-light interferometer (Wyant, 2002) using magnifications 10×, 20× and 50×. The micro-roughness on the grooves of the gratings was inspected by means of an atomic force microscope (Nanosurf Nanite). The parameters of the mirrors were presented in the design paper by Pärna et al. (2017). Table 1 shows the same parameters of the gratings as studied at the BESSY-II Optics Lab. One can see that the micro-roughness on the blaze facet is much higher for the 92 lines mm−1 grating. This is caused by the high ruling-layer thickness needed for the low-line-density grating: the gold grain roughness is transferred into the silicon. Usually, the gold of the blaze facets is compressed during the ruling process and a facet roughness of 0.8–1 nm r.m.s. for blaze angles ≥1.5° can be achieved. The gold grains in the 92 lines mm−1 profiles are only compressed in the surface region, leading to a much larger facet roughness.
of photons in a wide energy range. Therefore, it was decided to inspect the beam- and wave-front shaping parameters of all the mirrors and gratings to verify their performance. The micro-roughness on the surface of optical elements is of additional interest because it causes scattering and consequently a loss of photons. These studies were performed
|
The performance of both gratings was also characterized at the at-wavelength metrology facility at BESSY-II (Schäfers et al., 2016) and using a reflectometer (Sokolov et al., 2016). Fig. 2 shows the measured efficiencies of the two gratings in the energy range 20–1000 eV. The efficiencies for the second and third orders are given in `transmission' mode, where a photon energy value is divided by the corresponding diffraction order m.
4. Beamline performance
4.1. Photon energy repeatability and calibration
The advantage of illuminating the grating with collimated light is that the grating angles α and β can be chosen freely. The ratio cosα/cosβ is called a fixed-focus constant (cff) and it actually defines many of the properties of the dispersed radiation (Petersen, 1982; Petersen et al., 1995). The primary operation mode of the monochromator uses cff = 2.25, which provides a good compromise between high-order suppression, photon energy resolution and The monochromator performance presented below was obtained with cff = 2.25 unless otherwise mentioned. The control system, however, allows the operation of the monochromator with different cff values, thus giving the possibility of increasing the photon resolution or if required by the user.
The angular range of the grating cradle limits the low-energy side of the 92 lines mm−1 grating to 4.4 eV and that of the 600 lines mm−1 grating to about 15 eV. The highest achievable energies are effectively determined by the acceptance of the M2 mirror: the practical limits in the primary operation mode are about 50 and 1300 eV for the 92 and 600 lines mm−1 gratings, respectively.
According to the manufacturer's data, the angular repeatability of the mechanical system is 0.24 µrad for the M2 mirror and 0.036 µrad for the gratings, which determines a limit for the photon energy repeatability. Fig. 3 shows the limits of the photon energy repeatability calculated for both gratings using the above-mentioned values of angular repeatability (dashed lines). During a measurement, the photon energy repeatability is also determined by the settings of the motion controllers, which are programmed to have a positioning tolerance (a dead band) to avoid high-frequency noise from the encoders and to have a reasonably short settling time. The dead band is 0.625 µrad for the motion controllers of both M2 and PG. The solid lines in Fig. 3 show the calculated photon energy repeatability, as determined by the settings of the motion controllers. As can be seen, errors in photon energy arising from the mechanical repeatability of the system, shown by dashed curves in Fig. 3, are practically negligible compared with errors arising from positioning repeatability defined by the motion controllers. Therefore, the solid lines in Fig. 3 provide information on a minimum step size to be used in a scan measurement. To obtain a reasonably even-spaced photon energy scale, the step size should not be less than the value of the solid line at the corresponding photon energy in the figure.
The photon energy scale has been calibrated by the method described by Weiss et al. (2001). The general idea is to measure the position of a gas absorption feature with a known energy using different cff values. For a perfectly aligned monochromator, the photon energy E diffracted to the exit slit is described by the grating equation,
where h is Planck's constant, c is the speed of light, k is the diffraction order, N is the line density of the grating, θ is the plane mirror angle, α is the incident angle and β is the diffraction angle. If the directional angles of M2 and PG differ from their ideal values, the photon energy observed at the exit slit, Eobs, would not correspond to the expected value:
where Δθ and Δβ are the errors in the plane mirror and diffraction angles, respectively. Using equation (1), equation (2) can be expressed as a function of the photon energy, angular errors and cff:
Therefore, if the energy E of the absorption feature is known, the experimentally obtained dependence of the observed peak position on the cff value can be approximated with equation (3), thus calculating the values of Δθ and Δβ. Then the errors in the directional angles of M2 and PG can easily be found.
The calibration method was implemented by measuring total ion yield (TIY) spectra with an ion time-of-flight (TOF) spectrometer installed at the GPES. The measurements were made using the N 1s → π*, v = 1 transition of N2 at 401.10 eV (Sodhi & Brion, 1984) for the 600 lines mm−1 grating and the 2p1/2 → 13d transition of Ne at 21.5806 eV (Ito et al., 1988) for the 92 lines mm−1 grating. Several iterations of the calibration procedure were made until no systematic shift in the peak position was observed with the change in cff. A residual shift that had no dependence on cff still remained between the observed and known photon energies. It arose from the misalignment of the roll angle of the M3 mirror and was corrected by changing that angle. After the calibration procedure, the following transitions were investigated with cff = 2.25: Ne 1s → 3p at 867.12 eV (Sodhi & Brion, 1984), Ne 2p1/2 → 13d at 21.5806 eV (Ito et al., 1988), Ne 2s → 3p at 45.5442 eV (Schulz et al., 1996), N2 N 1s → π*, v = 1 at 401.10 eV (Sodhi & Brion, 1984), Ar 2p3/2 → 4s at 244.39 eV (King et al., 1977) and Xe 5p1/2 → 9s at 12.8887 eV (Berkowitz, 1979; Johnson et al., 1980). The obtained residual shifts between the known and observed photon energies are represented in Fig. 3 by red dots. Some of these transitions were also re-measured with different cff values. Slight shifts in the observed peak positions were observed; however, the behaviour of the shifts was not systematic, but rather noise-like. As these energy shifts were reproducible, a conclusion was made that they arise because of the limit of absolute accuracy of the mechanical system, which is typically worse than its mechanical repeatability, thus representing an inherent characteristic of the monochromator. The error bars in black at chosen points in Fig. 3 represent one standard deviation of the peak positions obtained using different cff values.
Calibration can be considered excellent if error bars cross the zero level, which would mean that the photon energy scale is calibrated on the level of the mechanical accuracy of the monochromator. As can be seen from Fig. 3, this condition is not achieved in the low photon energy range of the 600 lines mm−1 grating. We believe that this residual shift arises from a slight vertical misalignment of the exit slit of the monochromator. Even in that case, features in both absorption spectroscopy and photoemission spectroscopy will show up very close to their expected energy positions. In high-accuracy experiments, a separate calibration of the photon energy scale is needed at the time of the measurement, no matter how well the general calibration of the monochromator is performed.
4.2. Photon energy resolution
The photon energy resolution of the beamline was investigated by measuring the TIY spectra of several gases at the GPES. For that purpose, the Ne 1s → 3p, Ne 2p → 13d, N2 N 1s → π*, Ar 2p3/2 → 4s and Xe 5p1/2 → 9s resonances were used. Some selected spectra are displayed in Fig. 4.
The experimental core-excitation spectra were fitted with Voigt profiles, where the Lorentzian widths in the final fits were fixed to the following values based on the data in the literature: 111 meV for the Ar 2p3/2 → 4s excitation (Kato et al., 2007; Nicolas & Miron, 2012), 112 meV for the N 1s → π* excitations in N2 (Kato et al., 2007) and 256 meV for the Ne 1s → 3p excitation (Coreno et al., 1999; Kato et al., 2006). These values do not represent our suggestions for the lifetime broadenings of the core-excited states concerned, but the sole purpose of this analysis was to estimate the photon energy resolution. The Ne 2p and Xe 5p excitation spectra were approximated by Gaussian functions, as the contribution of lifetime widths in these spectra was assumed to be negligible. The experimental photon energy resolutions, obtained with a 10 µm slit, and the corresponding resolving powers are presented in Table 2. The dependences of the photon energy resolution as a function of the monochromator exit slit width were measured as well. They are shown with dots for the corresponding spectra in the insets of Fig. 4.
|
The photon energy resolution of the beamline was estimated by taking into account four main factors: source size, exit slit width, slope errors of the optical elements, and the diffraction limit. For the sake of simplicity, it was assumed that all of these factors follow a Gaussian distribution, and thus quadratic summing can be used to estimate the overall resolution. It should be noted, however, that the spatial distribution of the source is not Gaussian (Onuki & Elleaume, 2002) and the contribution of the exit slit would be described more adequately by a rectangular function. The calculated resolution agrees well with the experimental values, except for the cases where the diffraction limit of the grating (at low energy range) or the exit slit width gives a dominant term to the photon energy resolution. In these cases, the calculations tend to overestimate the resolution, i.e. underestimate the In the case of the exit slit contribution, this overestimation was taken into account by introducing a multiplier of 0.8 into the exit slit term. Examples of comparisons between the calculated (solid line) and experimentally obtained resolution (dots) are presented in the insets of Fig. 4.
In this way, we could estimate the photon energy resolution over a wide energy range and with different values of slit widths (Fig. 5). The calculated values are expected to reproduce the experimental resolution well, apart from a slight overestimation in the low energy range. In Fig. 5, dashed–dotted lines are shown as eye guides for some values of constant resolving power.
4.3. Beam polarization
The undulator contains four arrays of magnets that can be shifted independently in the longitudinal direction, allowing full control of the radiation polarization. So far, the undulator control system has been commissioned to operate in two modes, helical and inclined. In the inclined mode, two diagonal arrays of magnets move antiparallel, producing radiation with a zero phase shift between the horizontal and vertical components of the electric field vector, whereas in the helical mode the arrays of magnets move parallel, resulting in a phase shift of π/2. The magnitude of the magnets' shift changes the ratio between the vertical and horizontal components of the electric field vector. In the inclined mode, it allows us to obtain linear horizontal and vertical polarizations, as well as to have linearly polarized light with a chosen angle of inclination. In the helical mode, it is possible to obtain circularly polarized light by equalizing both components of the electric field vector. Changing the direction of the magnets' shift allows the generation of left or right circular polarization.
As the radiation propagates through the beamline, its polarization can change as the reflections at optical elements change the ratio between the vertical and horizontal components of the electric field vector and introduce an additional phase shift between them. Therefore, the polarization state of the radiation at the end of the beamline needs to be characterized and the undulator parameters adjusted, namely the magnetic gap and the longitudinal shift of the magnets, in order to counteract the changes introduced by the beamline optics. The exceptions are horizontal and vertical linear polarizations, as in these cases one of the components of the electric field vector equals zero, so the polarization state is not affected if all beamline components are properly aligned.
Polarization measurements in the UV range are often performed with a two-stage device, composed of a retarder stage and an analyzer stage with a coupled detector (Schledermann & Skibowski, 1971). The full characterization of polarization requires the acquisition of a two-dimensional intensity map with retarder and analyzer angles as independent variables, resulting in a significant number of individual measurements. We adopted a less time-consuming method where only a single analyzer mirror was used, similar to the method described by Cubric et al. (1999): if the radiation is fully polarized, this provides information about linear and circular polarization.
It can be shown that the dependence of the intensity of the polarized radiation I on the analyser mirror angle η is described by the equation
where the visibility V and phase shift Ψ are parameters defined by the radiation polarization and the properties of the mirror. Note that visibility is a function of photon energy.
If one assumes that light is fully polarized (S0 = 1), the can be found as:
where Vref is the reference visibility value, which is the visibility for fully linearly polarized radiation. The reference visibility value can be obtained from separate measurements using horizontal or vertical polarization as they are not affected by the beamline. The parameter PL in these equations is the linear polarization degree, PL = (S1 2 + S2 2)1/2, which characterizes the amount of linear polarization. The sign of S3 defines right or left polarization helicity. It cannot be found from the present measurement but can be easily predicted by calculations.
As mentioned before, the described method is only valid when the radiation is fully polarized, S0 = 1. Calculations made using measured magnetic fields in the undulator gap predict a high degree of polarization: the amount of unpolarized radiation is in the range of 0.1–0.7% for photon energies from 4.5 to 120 eV. To test the degree of polarization experimentally, the He 1s photoelectron spectrum was measured with horizontal and vertical linear polarizations at photon energies of 40 eV (with the 92 lines mm−1 grating) and 100 eV (with the 600 lines mm−1 grating). The Scienta R4000 electron analyzer of the GPES was positioned in the horizontal direction and it was operated with a pass energy of 10 eV. Analysis of the measured spectra intensities gave values of 99.46% (at 40 eV) and 99.17% (at 100 eV) for the degree of linear polarization. These values are underestimates of the real degree of polarization because the electron lens of the electron spectrometer has a large acceptance angle of 16°.
Utilizing the method described above, the undulator parameters were optimized in the range from 20 to 200 eV for two cases: circular polarization (both left and right), and inclined polarization with a `magic angle' (54.7°) between the horizontal direction and electric field vector.
As the phase shift induced by the beamline cannot be reversed in the inclined and helical modes of the undulator, the resulting polarization state will, strictly speaking, always be elliptical. We can characterize the quality of the polarization state by its linear polarization degree PL. Fig. 6 shows the energy dependence of the degree of linear polarization measured (dots) and calculated (lines) for circular (black) and inclined (blue) polarizations. Circles and solid lines represent measurements and calculations for the 600 lines mm−1 grating, whereas diamonds and dashed lines represent measurements and calculations for the 92 lines mm−1 grating. Qualitatively, the experimental behaviour of PL is in good agreement with the calculations, even though the calculations seem to underestimate the phase delay introduced by the beamline optics. Using the 600 lines mm−1 grating, the polarization quality remains quite good down to 50 eV. At lower photon energies, the incident angles on M2 and the grating become steep, which introduces a large phase delay. The 92 lines mm−1 grating is operated at much shallower incident angles. Its use preserves an acceptable quality of the polarization at low photon energies, but at the cost of lower photon energy resolution.
The EPU of FinEstBeAMS will be upgraded to operate in a universal mode (Bahrdt & Wüstefeld, 2011), which will allow the phase shift of the undulator to be adjusted as well. After its implementation we expect to provide pure circular and inclined polarization of radiation over the full energy range.
The possibility of using different polarizations greatly expands the experimental potential of the beamline. For example, the phenomenon of a `dark corridor' in graphene is believed to be the result of a photoemission interference effect, verified in experiments using linearly polarized radiation (Bostwick et al., 2007; Mucha-Kruczyński et al., 2008); the electronic near the von Hove singularity in graphene has been studied using circular polarized radiation (Hwang & Hwang, 2015; Liu et al., 2011). Fig. 7 presents angle-resolved (ARPES) data of a mixed monolayer–bilayer graphene on a 6H-SiC(0001) sample measured at the SSES. The obtained result agrees very well with the data found in the literature (Marchenko et al., 2018), indicating a high degree of circular polarization at the end of the beamline.
| Figure 7 of graphene using a photon energy of 22 eV. The point is set as a reference 0 in momentum space ( |
4.4. Photon flux
The undulator produces its maximum power (2.4 kW at the minimum gap of 14.0 mm) and provides the widest operational energy range in the horizontal polarization mode. Horizontal polarization is also considered to be the default operation mode for experiments.
Apart from the exit slit width, the
that can be exploited in an experiment depends on a combination of the monochromator input acceptance and selected photon energy (also relative to the central cone energy of the used undulator harmonic). This combination also determines other important properties of the photon beam. For instance, the use of the central cone energy and a relatively small acceptance allows one to reach a small focal spot size in the experiments and reduces the contribution of even undulator harmonics. On the other hand, selecting a photon energy below the central cone energy and a larger acceptance allows one to achieve a higher photon flux.In general, it is a complicated task to optimize the beamline parameters to obtain desirable beam properties among photon energy resolution, et al., 2009) and the limited radiation hardness of the samples.
size and shape of the beam spot, and presence of even harmonics, especially taking into account that different applications have rather diverse requirements for the photon beam. For example, and resolution are often the most crucial parameters in whereas the spot size and presence of higher-order radiation may not play important roles. In contrast, excitation measurements are significantly disturbed by the presence of higher-order radiation in the photon beam, whereas they do not really need the highest photon energy resolution provided by the monochromator because the studied luminescence spectra are influenced by solid-state broadening effects. Also, our experience indicates that high photon fluxes provided by an undulator often cannot be used due to excitation density effects (see KirmAs a reference, we present ) when the storage ring was operating with a 300 mA electron beam current.
curves measured with an acceptance of 215 µrad × 260 µrad (horizontal × vertical) in the photon energy range 4.5–500 eV and with an acceptance of 90 µrad × 135 µrad in the photon energy range 500–1300 eV (Fig. 8The current response from an Opto Diode AXUV-100 photodiode installed in the PLES was used to obtain the value of the ). For the same reason, the photon energy set by the monochromator was chosen to be higher than the peak energy value of the corresponding undulator harmonic, allowing us to decrease the intensity ratio between the second- and first-order radiation at low photon energies. For photon energies higher than 70 eV, the monochromator was tuned to the photon energy corresponding to the maximum of the undulator harmonic. In order to obtain the highest possible successive undulator harmonics were used, namely the first, third, seventh and eleventh in the energy range 4.5–600 eV, and the 21st, 23rd, 25th, 27th, 31st, 35th, 39th and 43rd in the energy range 600–1300 eV.
This response was corrected for the diode's taken from the manufacturer's data sheet. In the photon energy range 4.5–70 eV, different filters were used in order to reduce the impact of higher diffraction orders (curves obtained with the respective filters are represented with dashed–dotted lines in Fig. 8In Fig. 8, black and blue curves represent the measured photon fluxes obtained with the 600 and 92 lines mm−1 gratings, respectively, using a fixed monochromator exit slit width of 15 µm. These data are compared against the simulated results (dots and diamonds). The undulator was estimated using the SPECTRA software (Version 10.2; Tanaka & Kitamura, 2001) for the harmonics described earlier. The ray-tracing model in RAY-UI (Baumgärtel et al., 2016, 2019) was utilized to estimate the transmission of a photon entering the acceptance aperture to the sample plane. This transmission, multiplied with the undulator obtained in SPECTRA, presents the simulated expected on the sample. Note that all the figure errors of the delivered optical components were included in the ray-tracing simulations, and the actual acceptance used in the measurements was implemented in the calculations of the source characteristics as well as in the ray tracing.
As can be seen in Fig. 8, the correspondence between the experimental and simulated results is better for the measurement taken with the 600 lines mm−1 grating. However, there is a discrepancy on the higher energy side, the difference being about an order of magnitude at 1000 eV. The differences between the experiment and the simulations can be attributed to contamination of the mirrors and gratings. A clear indication of carbon contamination is seen in the measured at around 280 eV. The calculated at low energies also does not take into account the transmission of the filters that were used for the measurements at energies below 50 eV. The transmission of the filters varies in a range approximately from 10 to 80% depending on the type of filter. To reduce the effect of carbon contamination, a steady oxygen flow will be applied to the mirrors receiving a significant load of the synchrotron light. Such a method, called oxygen cleaning, has been shown to be effective on other beamlines at MAX IV.
The and 5, one can estimate the and resolution for a chosen value of the exit slit width and find the best compromise between them suitable for the required experiment.
increases proportionally with increasing exit slit width. Using Figs. 84.5. Focal beam spot size
Gas-phase electron spectroscopy and
techniques do not have strict requirements for the beam spot size; they can be satisfied with a beam spot of a couple of hundred micrometres without significant losses in performance. In contrast, electron spectroscopy of surfaces and interfaces, conducted at the SSES, could benefit from a smaller beam spot. However, a high beam divergence in the low energy range and space constraints in the experimental hall resulted in the choice of a relatively small demagnification factor of 2.6 for the M4 mirrors on both branches and modest spot sizes at the end stations. With an expected horizontal spot size of 220–320 µm at the monochromator exit slit, the focal beam spot size should be of the order of 100 × 100 µm in the major part of the operation range for a beam unrestricted horizontally and monochromator exit slit sizes of several hundred micrometres. A pair of horizontal baffles has been installed before the monochromator exit slit, allowing us to decrease the horizontal size of the beam spot if needed in some applications.For the beam spot visualization, an yttrium aluminium oxide garnet crystal doped with cerium (YAG:Ce) was installed on a sample holder in the analysis chamber of the SSES. The crystal had a metal grid on the surface with 1 mm step and 200 µm ticks for size reference. A long-focus optical microscope with a CCD camera was mounted on a viewport of the analysis chamber to record images. Photographs were taken through the viewport at an angle relative to the measurement plane, which resulted in quadrilateral scale units [Fig. 9(a)]. White lines were added to the photographs in order to scale separately the line profiles of the horizontal and vertical spot sizes. The beam spot size was then determined as the full width at half-maximum (FWHM) of the beam spot profile.
Fig. 9(b) presents the dependence of the vertical beam spot size on the monochromator exit slit width at photon energies of 25 and 100 eV. As can be seen, the vertical beam size deviates from the ideal dependence for slit widths below 80 µm and approaches a constant value of 20–25 µm. The horizontal beam spot size of the unrestricted beam practically shows no dependence on photon energy and remains in the range of 80–100 µm for photon energies above 10 eV. With the use of the horizontal baffles at the exit slit, it was possible to limit the horizontal beam spot size down to about 30 µm.
These values are in good agreement with the ray-tracing calculations presented by Pärna et al. (2017), where a beam spot of 100 µm × 20 µm was predicted using a 50 µm monochromator exit slit width and a horizontally unrestricted beam at a photon energy of 17 eV.
5. Conclusion
The baseline commissioning of the FinEstBeAMS beamline has been completed. The monochromator of the beamline is equipped with a 92 lines mm−1 grating for the photon energy range 4.5–50 eV and a 600 lines mm−1 grating for the photon energy range 15–1300 eV. A of ∼5000 has been achieved at 13 eV photon energy with the 92 lines mm−1 grating and of ∼11000 at 400 eV photon energy with the 600 lines mm−1 grating. Overall, the photon energy resolution agrees well with calculations based on the parameters of the optical elements. For the convenience of users, we have presented in graphical form the estimated photon energy resolution for different slit widths over the whole operation range of the FinEstBeAMS beamline.
Four polarization modes are available for users: horizontal linear and vertical linear polarizations, circular polarization (left and right), and `magic angle' inclined linear polarization, mostly intended for gas-phase users. Using horizontal polarization, the beamline can deliver above 2 × 1011 photons s−1 in the photon energy range below 10 eV at a of 3000, and more than 1013 photons s−1 at a of 5000 in the photon energy range 50–150 eV. At higher photon energies, the decreases, but it is still ∼109 photons s−1 at 1253.6 eV. A spot size of ∼90 µm (horizontal) × 25 µm (vertical) has been measured at the solid-state end station, in good agreement with design values.
FinEstBeAMS has accepted regular users at the gas-phase and
end stations since April 2019. Regular user operation at the SSES begins in 2021. Future commissioning work of the beamline will focus on circular and inclined polarization modes, optimization of the beamline acceptance, oxygen cleaning of the optics and characterization of beamline performance using different fixed-focus constants.Acknowledgements
The authors thank the staff of the MAX IV Laboratory for assistance during design, construction, commissioning and operation of the FinEstBeAMS beamline.
Funding information
The following funding is acknowledged: European Regional Development Fund (grant No. TK-141 HiTechDevices 2014-2020.4.01.15-0011 to University of Tartu; grant No. MAX-TEENUS 2014-2020.4.01.20-0278 to University of Tartu; grant No. Eesti Kiirekanal SLOFY11156T/1 to University of Tartu); Estonian Research Council (grant No. PRG-629 to University of Tartu); Jane & Aatos Erkko Foundation (grant No. SOFUS); Business Finland (grant No. 1464/31/2019); Academy of Finland (grant No. 319042; grant No. 326461; grant No. 326406; grant No. 320165); University of Oulu; University of Turku; Tampere University; University of Tartu.
References
Bahrdt, J., Frentrup, W., Gaupp, A., Scheer, M., Gudat, W., Ingold, G. & Sasaki, S. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 21–29. Web of Science CrossRef CAS Google Scholar
Bahrdt, J. & Wüstefeld, G. (2011). Phys. Rev. ST Accel. Beams, 14, 040703. Web of Science CrossRef Google Scholar
Baumgärtel, P., Grundmann, P., Zeschke, T., Erko, A., Viefhaus, J., Schäfers, F. & Schirmacher, H. (2019). AIP Conf. Proc. 2054, 060034. Google Scholar
Baumgärtel, P., Witt, M., Baensch, J., Fabarius, M., Erko, A., Schäfers, F. & Schirmacher, H. (2016). AIP Conf. Proc. 1741, 040016. Google Scholar
Berkowitz, J. (1979). Photoabsorption, Photoionization, and Photoelectron Spectroscopy, p. 181. New York: Academic Press. Google Scholar
Bostwick, A., Ohta, T., Seyller, T., Horn, K. & Rotenberg, E. (2007). Nat. Phys. 3, 36–40. Web of Science CrossRef CAS Google Scholar
Coreno, M., Avaldi, L., Camilloni, R., Prince, K. C., de Simone, M., Karvonen, J., Colle, R. & Simonucci, S. (1999). Phys. Rev. A, 59, 2494–2497. Web of Science CrossRef CAS Google Scholar
Coutinho, T., Cuní, G., Fernández-Carreiras, D., Klora, J., Pascual-Izarra, C., Reszela, Z., Suñé, R., Homs, A., Taurel, E. & Rey, V. (2011). Proceedings of the 13th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2011), 10–14 October 2014, Grenoble, France, pp. 607–609. Google Scholar
Cubric, D., Cooper, D. R., Lopes, M. C. A., Bolognesi, P. & King, G. C. (1999). Meas. Sci. Technol. 10, 554–558. Web of Science CrossRef CAS Google Scholar
Follath, R. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 418–425. Web of Science CrossRef CAS Google Scholar
Hwang, H. & Hwang, C. (2015). J. Electron Spectrosc. Relat. Phenom. 198, 1–5. Web of Science CrossRef CAS Google Scholar
Ito, K., Ueda, K., Yoshino, K., Morioka, Y. & Namioka, T. (1988). J. Opt. Soc. Am. B, 5, 2006–2014. CrossRef CAS Web of Science Google Scholar
Johnson, W. R., Cheng, K. T., Huang, K.-N. & Le Dourneuf, M. (1980). Phys. Rev. A, 22, 989–997. CrossRef CAS Web of Science Google Scholar
Kato, M., Morishita, Y., Koike, F., Fritzsche, S., Yamaoka, H., Tamenori, Y., Okada, K., Matsudo, T., Gejo, T., Suzuki, I. H. & Saito, N. (2006). J. Phys. B At. Mol. Opt. Phys. 39, 2059–2069. Web of Science CrossRef CAS Google Scholar
Kato, M., Morishita, Y., Oura, M., Yamaoka, H., Tamenori, Y., Okada, K., Matsudo, T., Gejo, T., Suzuki, I. H. & Saito, N. (2007). J. Electron Spectrosc. Relat. Phenom. 160, 39–48. Web of Science CrossRef CAS Google Scholar
King, G. C., Tronc, M., Read, F. H. & Bradford, R. C. (1977). J. Phys. B: At. Mol. Phys. 10, 2479–2495. Google Scholar
Kirm, M., Nagirnyi, V., Feldbach, E., De Grazia, M., Carré, B., Merdji, H., Guizard, S., Geoffroy, G., Gaudin, J., Fedorov, N., Martin, P., Vasil'ev, A. & Belsky, A. (2009). Phys. Rev. B, 79, 233103. Web of Science CrossRef Google Scholar
Kooser, K., Kivimäki, A., Turunen, P., Pärna, R., Reisberg, L., Kirm, M., Valden, M., Huttula, M. & Kukk, E. (2020). J. Synchrotron Rad. 27, 1080–1091. Web of Science CrossRef CAS IUCr Journals Google Scholar
Liu, Y., Bian, G., Miller, T. & Chiang, T.-C. (2011). Phys. Rev. Lett. 107, 166803. Web of Science CrossRef PubMed Google Scholar
Marchenko, D., Evtushinsky, D. V., Golias, E., Varykhalov, A., Seyller, T. & Rader, O. (2018). Sci. Adv. 4, eaau0059. Web of Science CrossRef PubMed Google Scholar
Mucha-Kruczyński, M., Tsyplyatyev, O., Grishin, A., McCann, E., Fal'ko, V. I., Bostwick, A. & Rotenberg, E. (2008). Phys. Rev. B, 77, 195403. Google Scholar
Nicolas, C. & Miron, C. (2012). J. Electron Spectrosc. Relat. Phenom. 185, 267–272. Web of Science CrossRef CAS Google Scholar
Onuki, H. & Elleaume, P. (2002). Undulators, Wigglers and Their Applications. Boca Raton: CRC Press. Google Scholar
Pankratov, V., Pärna, R., Kirm, M., Nagirnyi, V., Nõmmiste, E., Omelkov, S., Vielhauer, S., Chernenko, K., Reisberg, L., Turunen, P., Kivimäki, A., Kukk, E., Valden, M. & Huttula, M. (2019). Radiat. Meas. 121, 91–98. Web of Science CrossRef CAS Google Scholar
Pärna, R., Sankari, R., Kukk, E., Nõmmiste, E., Valden, M., Lastusaari, M., Kooser, K., Kokko, K., Hirsimäki, M., Urpelainen, S., Turunen, P., Kivimäki, A., Pankratov, V., Reisberg, L., Hennies, F., Tarawneh, H., Nyholm, R. & Huttula, M. (2017). Nucl. Instrum. Methods Phys. Res. A, 859, 83–89. Google Scholar
Petersen, H. (1982). Opt. Commun. 40, 402–406. CrossRef Web of Science Google Scholar
Petersen, H. & Baumgärtel, H. (1980). Nucl. Instrum. Methods, 172, 191–193. CrossRef Web of Science Google Scholar
Petersen, H., Jung, C., Hellwig, C., Peatman, W. B. & Gudat, W. (1995). Rev. Sci. Instrum. 66, 1–14. CrossRef CAS Web of Science Google Scholar
Sasaki, S., Kakuno, K., Takada, T., Shimada, T., Yanagida, K. & Miyahara, Y. (1993). Nucl. Instrum. Methods Phys. Res. A, 331, 763–767. CrossRef Web of Science Google Scholar
Schäfers, F., Bischoff, P., Eggenstein, F., Erko, A., Gaupp, A., Künstner, S., Mast, M., Schmidt, J.-S., Senf, F., Siewert, F., Sokolov, A. & Zeschke, T. (2016). J. Synchrotron Rad. 23, 67–77. Web of Science CrossRef IUCr Journals Google Scholar
Schledermann, M. & Skibowski, M. (1971). Appl. Opt. 10, 321–326. CrossRef CAS PubMed Web of Science Google Scholar
Schulz, K., Domke, M., Püttner, R., Gutiérrez, A., Kaindl, G., Miecznik, G. & Greene, C. (1996). Phys. Rev. A, 54, 3095–3112. CrossRef CAS PubMed Web of Science Google Scholar
Siewert, F., Noll, T., Schlegel, T., Zeschke, T. & Lammert, H. (2004). AIP Conf. Proc. 705, 847–850. CrossRef Google Scholar
Siewert, F., Löchel, B., Buchheim, J., Eggenstein, F., Firsov, A., Gwalt, G., Kutz, O., Lemke, S., Nelles, B., Rudolph, I., Schäfers, F., Seliger, T., Senf, F., Sokolov, A., Waberski, C., Wolf, J., Zeschke, T., Zizak, I., Follath, R., Arnold, T., Frost, F., Pietag, F. & Erko, A. (2018). J. Synchrotron Rad. 25, 91–99. Web of Science CrossRef CAS IUCr Journals Google Scholar
Siewert, F., Zeschke, T., Arnold, T., Paetzelt, H. & Yashchuk, V. V. (2016). Rev. Sci. Instrum. 87, 051907. Web of Science CrossRef PubMed Google Scholar
Sjöblom, P., Lindberg, M., Forsberg, J., Persson, A. G., Urpelainen, S. & Såthe, C. (2016). AIP Conf. Proc. 1741, 030045. Google Scholar
Sodhi, R. N. S. & Brion, C. E. (1984). J. Electron Spectrosc. Relat. Phenom. 34, 363–372. CrossRef CAS Web of Science Google Scholar
Sokolov, A., Bischoff, P., Eggenstein, F., Erko, A., Gaupp, A., Künstner, S., Mast, M., Schmidt, J.-S., Senf, F., Siewert, F., Zeschke, T. & Schäfers, F. (2016). Rev. Sci. Instrum. 87, 052005. Web of Science CrossRef PubMed Google Scholar
Tanaka, T. & Kitamura, H. (2001). J. Synchrotron Rad. 8, 1221–1228. Web of Science CrossRef CAS IUCr Journals Google Scholar
Tarawneh, H., Thiel, A. & Ebbeni, M. (2019). Proceedings of the 13th International Conference on Synchrotron Radiation Instrumentation (SRI2018), 11–15 June 2018, Taipei, Taiwan, p. 030023. American Institute of Physics. Google Scholar
Thiel, A., Ebbeni, M. & Tarawneh, H. (2018). Proceedings of the 10th Mechanical Engineering Design of Synchrotron Radiation Equipment and Instrumentation (MEDSI2018), 25–29 June 2018, Paris, France, pp. 6–10. Geneva: JACOW Publishing. Google Scholar
Weiss, M. R., Follath, R., Sawhney, K. J. S. & Zeschke, T. (2001). Nucl. Instrum. Methods Phys. Res. A, 467–468, 482–484. Web of Science CrossRef CAS Google Scholar
Wyant, J. C. (2002). Proc. SPIE, 4737, 98–107. CrossRef Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.