Trace-element
sensitivity: a stress test for a new XRF multi-detectoraElettra Sincrotrone Trieste, Basovizza, Trieste, Italy, bINFN – Trieste, Padriciano, Trieste, Italy, cFondazione Bruno Kessler – FBK, Trento, Italy, dTIFPA – INFN, Trento, Italy, ePolitecnico di Milano, Como, Italy, fINFN – Milano, Milano, Italy, gDepartment of Mathematics, Computer Science, and Physics, University of Udine, Udine, Italy, hDepartment of Chemical and Geological Science, University of Cagliari, Cagliari, Italy, iDepartment of Medical Imaging, University of Saskatchewan, Saskatoon, Canada SK S7N 5A2, and jDepartment of Science, University of Roma Tre, Roma, Italy
*Correspondence e-mail: ilaria.carlomagno@elettra.eu
X-ray absorption fine-structure (XAFS) spectroscopy can assess the chemical speciation of the elements providing their coordination and d elements down to the p.p.m. range. The potentialities of the detector are presented through a case study: the speciation of ultra-diluted elements (Fe, Mn and Cr) in geological rocks from a calcareous formation related to the dispersal processes from Ontong (Java) volcanism (mid-Cretaceous). Trace-elements speciation is crucial in evaluating the impact of geogenic and anthropogenic harmful metals on the environment, and to evaluate the risks to human health and ecosystems. These results show that the new detector is suitable for collecting spectra of 3d elements in trace amounts in a calcareous matrix. The data quality is high enough that quantitative data analysis could be performed to determine their chemical speciation.
information generally hidden to other techniques. In the case of trace elements, achieving a good quality signal poses several challenges, as it requires high counting statistics and detector linearity. Here, a new multi-element detector is presented, specifically designed to probe the chemical speciation of trace 31. Introduction
; Parsons et al., 2002; Mastelaro & Zanotto, 2018; Kuzmin & Chaboy, 2014). can be applied to studies over a wide range of temperature (from a few K to some thousands of K) and pressure (Koningsberger & Prins, 1988; Bunker, 2010). Owing to these characteristics, is widely exploited in geological and environmental sciences where the materials of interest are often highly heterogeneous both in chemical composition and physical status, presenting mixtures ranging from crystalline to amorphous phases (see Quartieri, 2015; De Giudici et al., 2015; Brown Jr & Sturchio, 2002; Jones, 1999, and references therein). The possibility to identify the chemical species to which the elements belong is one of the major strengths of the technique when dealing with environmental science: in fact, the impact of an element in terms of hazardousness, for example, is determined not only by its concentration but also, crucially, by its chemical nature (Tack & Verloo, 1995). The latter also determines the element bioavailability, mobility, pathways and toxicity (Tack & Verloo, 1995; Hettiarachchi et al., 2017), thus affecting specific ecosystem functions and components. Therefore, chemical speciation is key information to understand environmental risks to human health and ecosystems, and can help in the design of remedial actions to prevent the dispersion of pollutants (Scheckel & Ford, 2010). In paleoenvironmental science, the variability of trace elements in rocks is a useful proxy for insights into geological events such as regional dispersion of 3d metals over paleoceans and disruptive volcanic events (Kravtsova, 2020; Terzano et al., 2019). For this reason, the simultaneous assessment of chemical concentration and mineralogical speciation of trace elements, possibly carried out in unique analytical routines, is crucial to understand environmental processes.
is an element-selective technique, sensitive to the local coordination chemistry and to the valence state of the absorber, independently from its physical state, crystallographic phase (crystalline, amorphous) or morphology (Koningsberger & Prins, 1988Synchrotron radiation ). The XANES features generally represent roughly 10% of the total signal of the absorbing element while, at higher energies, the features may decrease even below the 1% of the absorption signal. Therefore, measuring the signal requires much stricter specifications to the detection system than those needed for XRF. In particular, when the element of interest is highly diluted, as in the case of trace elements, the Poisson noise associated with fluorescence photons emitted from the atoms of the matrices represents a major obstacle to the measurement.
(SR-XRF) is widely used to individuate and quantify the elements in samples. Compared with SR-XRF, in fluorescence mode poses additional challenges due to the nature of the signal itself, represented by the fine-structure oscillations (XAFS) of the appearing in the near-edge (XANES) and extended post-edge region (EXAFS) (Bunker, 2010A possible approach to cope with these constraints is to increase the incident X-ray et al., 2018). However, the high of the X-ray beams may induce radiation damage in photosensitive systems (e.g. oxides of Fe and Mn) and organic materials, while the high energy resolution provided by the analyzers (ΔE/E < 10−3) is often not required for the purpose of experiments, making them more suitable for specific techniques such as resonant inelastic X-ray scattering (RIXS) (Brookes et al., 2018). Solid state detectors based on Si and Ge semiconductors possess a more suitable energy resolution (ΔE/E ≃ 10−2) (Cramer et al., 1988; Fabjan & Schopper, 2020; Chatterji et al., 2016) and represent a valuable alternative. In particular, Si detectors using the sideward depletion method such as silicon drift detectors (SDDs) (Gatti & Rehak, 1984; Gatti et al., 1984) enable spectroscopic measurements with an energy resolution close to the Fano limit (Fano, 1947) when coupled to custom ultra-low-noise front-end electronics (Mele et al., 2021). However, especially for diluted samples, the yield of specific fluorescence lines of the trace elements might be low, while the overall impinging on the detector, mainly originating from the elastic and and the from the matrix, is usually extremely high. In such situations, the use of multi-element detectors is indispensable and of paramount importance since single, albeit large, area spectroscopic detectors would suffer from pile-up and saturation effects. In addition, microcrystalline phases in specimens can give rise to Bragg peaks that spoil the spectra with spikes, glitches or even paralyze the detector at given energies. Dividing the acquisition area into independent elements is a suitable way to cope with these problems. If Bragg peaks affect few detector elements at selected energies, the signal of the saturated elements can be removed during post-processing analysis (Meneghini et al., 2009). However, when intense Bragg peaks excite the total fluorescence of the sample so affecting the signal of several detector elements, then the multidetector geometry does not help with respect to other solutions. A number of custom-made and commercial off-the-shelf detector systems are already available and, for most of them, the manufacturing companies (such as, but not limited to, Mirion, Hitachi and Ketek) offer multi-element solutions (4–32 elements) based on individual SDDs, while PNsensor provides also monolithic multi-channel silicon sensors (Strüder, 2000). Regarding Ge detectors, Canberra (now Mirion Technology) has provided for instance 32 and 64 channel detectors (Chatterji et al., 2016), and is currently offering high-purity monolithic Ge detectors with up to 64 elements. Although the application of off-the-shelf detectors in experiments in fluorescence mode is a convenient approach, only custom tailored detector systems are able to thoroughly exploit the potential of state-of-the-art beamlines (Thompson et al., 2000). The MAIA detector for instance (Ryan et al., 2014; Chen et al., 2017) is an excellent example of a custom-made solution featuring an array of 20 × 20 individual silicon diodes or small-sized silicon drift detectors with 1 mm2 active area. Another system developed by academic and industrial partners is the ARDESIA detector, based on a monolithic array of four circular or squarish SDDs (collimated active area of 12.56 mm2 or 15 mm2), which has been successfully installed and tested at different synchrotron radiation beamlines (Bellotti et al., 2016; Hafizh et al., 2019).
(high counts) and to use monochromators on the (crystal analyzers) to sharply select the absorption signal of the element of interest, as implemented on insertion device beamlines (Diaz-MorenoIn this framework, several Italian institutions have started collaborating on the design of a multi-element SDD detector tailored for the specific needs of the et al., 2009). The research has focused on the study of ultra-diluted systems through in fluorescence mode and aimed at achieving: (i) higher than 80%, in particular in the low-energy region of the beamline (2.4–10 keV), (ii) both room-temperature operation and cooling of the sensors, suitable for automated beam time use, (iii) coverage of a wide solid angle of ≥30% at 10 cm from the sample, (iv) energy resolution ΔE/E ≤ 180 eV with peaking time of 1 µs. The sensors are produced by the Fondazione Bruno Kessler (FBK), a silicon foundry in Trento, Italy, that has developed a special production process yielding a substantial reduction of leakage currents in the sensors allowing the operation of large area detectors with excellent energy resolution even at room temperature (Bertuccio et al., 2015, 2016).
beamline at Elettra Sincrotrone Trieste (Di CiccoHere we discuss the application of this detector in a d metal traces (Fe, Mn and Cr) in a calcareous formation. The very high dilution of the elements of interest, in particular for Cr reaching the p.p.m. concentration, represents a valuable stress test for the device.
study of 32. Experimental section
2.1. The multi-element XRF detector
At present, the multi-element detector (Rachevski et al., 2019; Cirrincione, 2019; Bufon et al., 2019) is assembled tiling up eight independent monolithic modules, each of which comprises eight linearly aligned SDDs (Gatti et al., 1984) [Fig. 1(a)].
Each module is equipped with a Peltier cell and its own water pipeline to purge the residual heat of the former when the detector is operated in cooling mode. A new design technique (INFN, 2018) has been employed to maximize the active area of the sensors keeping the instrument as compact as possible. The sensors have been fabricated on 450 µm-thick, n-type high-purity silicon wafers with a resistivity of 9 kΩ cm. Each sensor module has typical dimensions of 26.65 mm × 5.65 mm, and it features eight SDD cells of 3 mm × 3 mm active area (0.478 area ratio). The active over total area ratio for the whole detector is 0.339. The size of the single SDD cell was decided on the basis of simulations that took into account not only the total active area but also the acquisition solid angle and the dead-time following each event. It has been noted that the increase in the number of cells, which divide the same total area, although initially dramatically increasing the efficiency, becomes less and less important when the dimensions of the cells are very small (also due to the dead zones introduced by the collimators and separations between cells). Therefore, a solution of 3 mm × 3 mm turned out to be the best compromise between efficiency optimization and constructive complexity of the system.
On the front side of each monolithic module a shallow and uniformly implanted p+ entrance window covers all eight SDD cells. This guarantees high efficiency even at very low photon energies down to and below the carbon Kα emission line at 277 eV (Bufon et al., 2018, 2019). The back side of each SDD cell is segmented into concentric drift cathodes (p+ rings) and a small collecting anode (n+ pad) in the centre (Rachevski et al., 2019; Cirrincione, 2019), which is virtually grounded through the attached front-end electronics. By using an integrated voltage divider, every drift cathode is biased with decreasing potential generating a gradient drift field, which is pointing away from the collecting anode. The interaction of X-ray photons with the bulk silicon creates electron–hole pairs. These charge carriers drift along the electric field lines, with the electrons being eventually collected at the anode. The number of generated electrons per photon, i.e. the released charge, is proportional to the characteristic energy of the X-ray photon. The photo-generated electrons are converted into a voltage proportional to the photon energy by the front-end electronics, connected to the readout anode via wire bonds. The front-end electronics is made up of custom-made, ultra-low noise, monolithic charge-sensitive pre-amplifiers operating in pulse reset mode (Bertuccio et al., 2014). This system will be referred to as `SIRIO' in the following. SIRIO enables low dead-time and high-count-rate measurements (some 106 photons s−1) with energy resolution better than 150 eV FWHM at the Mn Kα emission line (5.9 keV) at room temperature. Each of the SIRIO output signals from a monolithic module is filtered and further amplified by CR-RC2 analog pre-shapers with selectable peaking time of 0.3 µs or 1.2 µs. The peak-shaped signals are then sampled by multichannel analog-to-digital converters (ADCs) with a resolution of 12 bits at 40 MHz sampling rate, and further filtered with trapezoidal finite impulse response (FIR) filters of variable length (peaking time ranges of 0.4–1.6 µs and 1.3–3.2 µs, respectively) in order to guarantee the maximum energy resolution even at a count rate of ∼2.5 × 105 counts s−1 cell−1. Digital filtering, settings and complete monitoring of the instrument are realized within high-performance FPGAs, which send the acquired data to a host computer via a TCP/IP protocol. The detector discussed here employs a total of eight FPGA boards and eight TCP/IP transceivers. This modular structure allows an up- or down-scaling towards larger or smaller detection systems. In both cases, each individual cell is operated independently and includes different preset digital filters. The detector acquisition software ensures the proper alignment and calibration of the energy spectra for all the cells. Moreover, the software offers the possibility to separately save the information related to each cell (raw and pre-processed energy spectra) while its multichannel analyzers (MCAs) allow to select specific energy regions from the total XRF spectrum of each cell.
The system provides an efficient pile-up rejection and can identify the count rate excess on each cell accounting for the events that can be correctly analysed and ignoring the ones that should be rejected. The software also performs a preliminary data treatment on-the-fly, during the data acquisition, which includes the count rate dead-time correction, resulting in a clean, ready-to-use energy spectrum for each cell (Ciatto et al., 2004). The detector read-out electronics coupled with the specifically developed software provides an automatic (rate-induced) or manually selected cell disabling feature allowing to prevent system paralysis in case of extremely high count rates (e.g. Bragg peaks).
As depicted in Fig. 2, the high rate behaviour follows that of a paralyzed detector, which is higher than that reported in data sheets of the majority of commercial devices or reported for custom-made solutions considering the maximum output count rate ( Ryan et al., 2014) but somewhat lower than what is reported by Hafizh et al. (2019), which operates at shorter peaking times (≤128 ns). In Fig. 3 the energy resolution (FWHM) for the Mn Kα emission line at 5.9 keV is shown as a function of the peaking time, revealing little difference between the cooled detector (red line) and room-temperature operation (black line), demonstrating the very low noise level associated with the SDD leakage current. While the room-temperature behaviour is unique in comparison with other multi-element detectors, the energy resolution for the cooled detector is still somewhat inferior to that reported by Strüder (2000) and Bellotti et al. (2016). Furthermore, the fact that the distal cells are more prone to Compton contributions compared with the centre cell can be exploited for morphological studies of the sample (Billè et al., 2016).
2.2. Samples and chemical analysis
The samples for our case study were collected on a calcareous formation in Cava Colonella (so-called chalkstone formation di Bari) in Murge, Apulia, Italy (Istituto Superiore per la Protezione e la Ricerca Ambientale, 2008). This geological formation is related to the Ontong Java volcanism which is associated with the global environmental perturbations in the mid-Cretaceous. In samples like these, the possibility to understand the speciation of some 3d metals like Fe, Mn and Cr is specifically useful to mark dispersal processes from these geological events.
For chemical analysis, samples were ground in an agate mortar. Acid digestion was carried out on 0.2 g of each sample by the microwave ETHOS One, Advanced Microwave Digestion System, Milestone. A high-purity mixture of 4 ml of Milli-Q water (<0.1 µS cm−1) and 6 ml of HCl (37% Suprapur) was added to the solids into microwave vessels. Samples were processed together with a blank and a reference material prepared with the same mixture. After cooling, the mixtures were transferred into Teflon beakers rinsing the vessels with a few ml of Milli-Q water. These mixtures were heated in a hot-plate (110°C); following evaporation, 3 ml of concentrated HNO3 (65% Suprapur) were added three times. Finally, the mixture was filtered (0.4 µm), and the solution was made up to 50 ml final volume using Milli-Q water. Trace elements were quantified by inductively coupled plasma optical emission spectrometry (ICP-OES; ARL Fisons ICP Analyzer 3520 B), and inductively coupled plasma (ICP-MS; PerkinElmer, Elan/DRC-e, USA). To evaluate the accuracy (<5%) and precision (<5%) of the chemical analysis, EnviroMAT Drinking Water (SCP Science), EP-H-3 and EP-L-3 reference solutions were used. Concentrations of Fe, Mn and Cr are reported in Table 1.
|
2.3. measurements and data treatment
The et al., 2009) at Elettra synchrotron radiation facility. The multi-element detector was installed in the experimental hutch [Fig. 1(c)] oriented at 90° with respect to the incoming beam in the horizontal plane, in order to minimize the signal from the sample. An Al filter was placed in front of the detector in order to attenuate the intense background fluorescence signal coming from the lighter element (mainly Ca) matrix. The Al filter allowed the detector count rate to remain above the 90% linearity response region.
measurements for the presented case study were collected at the beamline (Di CiccoThe samples were ground in fine powders, mixed with light-absorbing matrix (cellulose) and pressed into thin solid pellets. The pellets were then mounted in a vacuum cell, with their surface at 45° with respect to the incident X-ray beam and the detector. The sample-to-detector distance was 2–4 cm covering a solid angle of 11–3%. A raw XRF spectrum from sample C20 is shown in Fig. 4 as an example: the weakness of the fluorescence signal from elements of interest (Fe, Mn and Cr) with respect to the total fluorescence is evident. For the four samples, the fraction of the counts related to the three inspected elements (Ctsel/Ctstot) was approximately: Cr, 1 × 10−3; Mn, 2 × 10−3; Fe, 8 × 10−3.
I0 was measured by a gas-filled and a full spectrum was recorder by each of the 64 detector elements as described above. The raw fluorescence signals were individually treated to identify and remove spoiled signals (saturation, spikes) while the dead-time correction has been implemented applying a not-paralyzable detection model (Ciatto et al., 2004). The fluorescence signals corresponding to the Kα emission of the absorbing elements were obtained from the regions of interest (ROIs), and the absorption signal α was calculated as
data were acquired in energy step scanning mode using a Si (111) double-crystal monochromator. For each energy point the incident beamwhere the sum runs over the n active detector cells at that energy point. For each sample several spectra αj were collected and summed up.
Additional Fe, Mn and Cr K-edge spectra were measured in transmission geometry on a set of reference compounds to be used for quantitative XANES analysis through linear combination analysis (LCA) (see below). The raw data were processed using the well known procedures (Bunker, 2010) for data treatment including pre-edge background subtraction, post-edge bare ion model definition, and edge jump normalization.
The Fe K-edge normalized spectra of the three investigated samples are shown in the left-hand panel of Fig. 5. Fe is the most concentrated element among the ones investigated. The Fe Kα fluorescence signal corresponds roughly to 10−3 of the total XRF signal in each detector element; a single scan in the XANES region required approximately 40 min. The normalized XANES spectra shown in Fig. 5 required roughly 2 h per sample (3–4 scans). Such an Fe concentration is quite routinely investigated at facilities and barely accessible with single-element detectors. The Fe data allowed to test the procedures for the preliminary on-the fly data treatment.
The Mn concentration in the investigated samples was approximately one order of magnitude lower than that of Fe, the Mn Kα fluorescence signal corresponding roughly to some 10−4 of the total XRF signal in each detector element. A larger number of energy scans (8–12 per sample) was then required to obtain suitable statistics for a quantitative XANES analysis. Each of the normalized Mn K-edge spectra presented in the central panel of Fig. 5 required 4–6 h of total measurement time. The statistical noise is slightly larger than for Fe K-edge spectra due to the lower Mn concentration and represents a suitable compromise for an appropriate quantitative XANES analysis.
The few-p.p.m. Cr concentration in the samples represents a challenge to test the limits of the multi-detector for XANES acquisition. Due to the long acquisition times and the limited beam time allocated for the experiment, the Cr K-edge was investigated just for one sample: the one with the lowest Cr concentration of 2.9 p.p.m. The Cr Kα fluorescence signal, being less than 10−4 of the total XRF signal, corresponds to a few hundred counts per second in each detector element. The spectrum presented in the right-hand panel of Fig. 5 required about 12 h, averaging 18 energy scans. Despite the higher statistical noise in the data, the main spectral features are suitable for reliable quantitative analysis to assess the Cr local coordination chemistry in the sample.
3. Data analysis and results
Figure 6 shows the Fe K-edge spectra of our samples together with those from relevant reference samples. In all our samples the Fe spectra are consistent with mainly Fe3+ valence state. Noticeably, the XANES spectra depict changes in the edge shoulder around 7125 eV that rises with lowering Fe concentration (i.e. following the order: C9, C20, C11, C13; details in Table 1). Such a shoulder seems to be related to the presence of Fe-carbonate (FeCO3) for which XANES depicts an evident peak at such energy. The observed trend suggests the raising of the FeCO3 phase fraction for the lower-concentrated Fe-samples.
The Mn K-edge experimental spectra depict a double-peak structure at 6550 eV and 6560 eV. Looking at the XANES spectra of reference compounds, the double peak is likely associated with an Mn-carbonate (MnCO3) phase. Changes among the samples mainly occur to the intensity of the main peak at 6550 eV.
The Cr K-edge XANES spectrum measured from sample C9 is presented in Fig. 6 compared with the Cr K-edge spectra from the relevant reference compounds. The double-peak shape of the white line clearly stands out around 6010 eV followed by a secondary peak around 6025 eV. These features are typical of the Cr2O3 phase, while they are absent in the other references.
To quantitatively assess the electronic state and coordination chemistry of Fe, Mn and Cr in these samples, the XANES spectra were analysed using XANES spectra from reference compounds measured in transmission at the ).
beamline. The LCA describes the local structure of the sample as a combination of phases whose local coordination chemistry and valence state are similar to those of the references (Benfatto & Meneghini, 2015After preliminary attempts, we could adequately reproduce the Fe K-edge XANES using three phases. The results of the analysis are presented in Table 2 for all the samples, while Fig. 7 reports the best fit (red line) to the experimental data (black dots) for sample C13, i.e. the one having the lowest Fe concentration. Table 2 points out that the Fe in our samples is present mostly as a carbonate phase similar to the siderite (FeCO3) and an Fe-oxide phase locally similar to ferrihydrite [FeO(OH)] structure. Traces of hematite are revealed in sample C13.
|
The LCA of the data collected at the Mn edge revealed the presence of two phases: an Mn carbonate MnCO3 and an Mn3+ oxide similar to Mn2O3. The LCA results, reported in Table 3, show that the Mn carbonate-like phase fraction is larger for the samples with lower Mn content, while the Mn3+ oxide phase fraction is the lowest for the C9 sample, the one having the highest Mn concentration. The LCA best fit is presented in Fig. 8 for sample C11, having the lowest Mn concentration.
|
The LCA was applied to the Cr K-edge XANES on the C9 sample, having less than 3 p.p.m. Cr concentration. The LCA best fit, reported in Fig. 9, required three reference spectra: a major contribution, around 77%, from the Cr3+ oxide phase (like Cr2O3); a Cr-carbonate phase (like CrCH3COO), around 16%; and a weak but statistically significant contribution from Cr6+ of about 7%, as shown in Table 4.
|
4. Discussion and conclusions
A general realization has emerged that there is a clear and widening gulf between the increasing et al., 2014; Hafizh et al., 2019). Owing to their large solid angle coverage they help to reduce acquisition times thus enabling higher user throughput on the beamlines. The system described here, designed expressly for has the ability of measuring large rates without saturating the detector, which is of pivotal importance for diluted samples. For this purpose, a highly pixelated detector divided into 64 small elements has been designed, fabricated and tested. At the moment the detector is designed to operate with peaking times ≥0.4 µs, resulting in a maximum count rate of 2.5 × 105 counts s−1 cell−1, comparable with the performance of other state-of-the-art detectors with the same peaking time. In these conditions, however, the detector is capable of handling a total count rate ≥10 × 106 counts s−1, which to our knowledge is higher than any other detector system currently available. If required, with a modification of the acquisition scheme (analog high-pass filter instead of a CR-RC2 scheme) and sacrificing marginally energy resolution it is possible to reduce the minimum peaking time to the 100 ns level or below, reaching count rates of the order of 1 × 106 counts s−1 cell−1, thus 64 MHz for the entire detector.
delivered at the sample stage by the new-generation synchrotron radiation sources, and the capacity of the detectors in fully exploiting such abundance of photons. Off-the-shelf detectors are certainly attractive solutions for many beamlines in general and for XRF beamlines in particular. However, especially in experiments, for the thorough exploitation of their potential, a tailored detection system serving specific purposes is required. The use of multi-element detection systems opens new perspectives for X-ray research as for instance demonstrated by the MAIA detector or the Ardesia system (RyanFor the cooled multi-element detector described here the energy resolution, especially for low input count rates, is marginally inferior to the single-element SDD system relying on Cube technology (Cube, 2012). In the present system, it is a matter of some pick-up on the bias lines, which will be remedied in the next version of the detector. As an additional and valuable feature it is noteworthy that the spatial resolution provided by the system enables morphological studies of the sample (Billè et al., 2016).
The potentialities of our detector system were presented through a case study about trace elements in rocks across stratigraphic sequences and geochemical association, a valuable tool to gain insight into geological events. Understanding the 3d metals mobility across oceans and their incorporation in sediments requires not only the knowledge of geochemical concentrations but also their mineralogical speciation. This work presented the use of a prototype high-performance detector to collect XANES spectra at few p.p.m. concentration for 3d metals Fe, Mn and Cr and providing a good quality signal, suitable for quantitative chemical speciation.
Our data show that Fe2+ and Mn2+ can form their calcite series bearing carbonate, but they also form Fe3+ and Mn3+ oxides. This suggests that Fe and Mn carbonates can get through partial oxidative dissolution with precipitation of secondary oxides. However, biological interfaces often control oxidation reactions and then two different oxidative states of trace elements can appear due to the primary carbonate formation process. Cr is mainly present as Cr3+ oxide with weak traces of Cr6+. Interestingly, the double peak in the XANES white line is consistent with Cr K-edge spectra of Ca3Cr2(SiO4)3 (Liu et al., 2020) suggesting a major fraction of Cr in the calcium silicate environment. The well resolved XANES features suggest a relatively ordered and highly crystalline environment. Some amount of Cr3+ is found in a disordered atomic environment; owing to the weakness of this contribution (around 16%) it is difficult to assess the chemical speciation: Cr acetate is a candidate but the similarity of Cr K-edge XANES in other amorphous environments like Cr3+-doped glasses or oxy-hydroxides (Ohta, 2015; Liu et al., 2020) prevents achieving a definitive answer. These results prove that the new multidetector system provides good quality XANES spectra from trace elements down to the p.p.m. concentration, suitable for quantitative analysis.
Acknowledgements
The authors thank R. Graziano and R. Graspini for collecting the samples. Open access funding provided by Elettra Sincrotrone Trieste SCpA within the CRUI-CARE Agreement.
Funding information
This work has been carried out within the ReDSoX-2 INFN research project, supported with the contribution of the Italian Ministry of University and Research (MIUR) within the EUROFEL Project and the contribution of Fondazione Bruno Kessler (FBK).
References
Bellotti, G., Butt, A., Carminati, M., Fiorini, C., Balerna, A., Piemonte, C., Zorzi, N. & Bombelli, L. (2016). 2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD), 29 October–6 November 2016, Strasbourg, France. IEEE. Google Scholar
Benfatto, M. & Meneghini, C. (2015). In Synchrotron Radiation, edited by S. Mobilio, F. Boscherini and C. Meneghini. Heidelberg, Berlin: Springer. Google Scholar
Bertuccio, G., Ahangarianabhari, M., Graziani, C., Macera, D., Shi, Y., Gandola, M., Rachevski, A., Rashevskaya, I., Vacchi, A., Zampa, G., Zampa, N., Bellutti, P., Giacomini, G., Picciotto, A., Piemonte, C. & Zorzi, N. (2016). IEEE Trans. Nucl. Sci. 63, 400–406. Web of Science CrossRef CAS Google Scholar
Bertuccio, G., Ahangarianabhari, M., Graziani, C., Macera, D., Shi, Y., Rachevski, A., Rashevskaya, I., Vacchi, A., Zampa, G., Zampa, N., Bellutti, P., Giacomini, G., Picciotto, A. & Piemonte, C. (2015). J. Instrum. 10, P01002. Google Scholar
Bertuccio, G., Macera, D., Graziani, C. & Ahangarianabhari, M. (2014). 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 8–15 November 2014, Seattle, WA, USA. IEEE. Google Scholar
Billè, F., Kourousias, G., Luchinat, E., Kiskinova, M. & Gianoncelli, A. (2016). At. Spectrosc. 122, 23–30. Google Scholar
Brookes, N., Yakhou-Harris, F., Kummer, K., Fondacaro, A., Cezar, J., Betto, D., Velez-Fort, E., Amorese, A., Ghiringhelli, G., Braicovich, L., Barrett, R., Berruyer, G., Cianciosi, F., Eybert, L., Marion, P., van der Linden, P. & Zhang, L. (2018). Nucl. Instrum. Methods Phys. Res. A, 903, 175–192. Web of Science CrossRef CAS Google Scholar
Brown, G. E. Jr & Sturchio, N. C. (2002). Rev. Mineral. Geochem. 49, 1–115. Web of Science CrossRef CAS Google Scholar
Bufon, J., Altissimo, M., Aquilanti, G., Bellutti, P., Bertuccio, G., Billè, F., Borghes, R., Borghi, G., Cautero, G., Ciano, S., Cicuttin, A., Cirrincione, D., Crespo, M. L., Fabiani, S., Ficorella, F., Gandola, M., Gianoncelli, A., Giuressi, D., Grisonich, R., Kourousias, G., Mannatunga, K. S., Mele, F., Menk, R. H., Olivi, L., Orzan, G., Picciotto, A., Rachevski, A., Rashevskaya, I., Sammartini, M., Schillani, S., Stolfa, A., Zampa, G., Zampa, N., Zorzi, N. & Vacchi, A. (2019). AIP Conf. Proc. 2054, 060061. Google Scholar
Bufon, J., Schillani, S., Altissimo, M., Bellutti, P., Bertuccio, G., Billè, F., Borghes, R., Borghi, G., Cautero, G., Cirrincione, D., Fabiani, S., Ficorella, F., Gandola, M., Gianoncelli, A., Giuressi, D., Kourousias, G., Mele, F., Menk, R. H., Picciotto, A., Rachevski, A., Rashevskaya, I., Sammartini, M., Stolfa, A., Zampa, G., Zampa, N., Zorzi, N. & Vacchi, A. (2018). J. Instrum. 13, C03032. Web of Science CrossRef Google Scholar
Bunker, G. (2010). Introduction to XAFS. Cambridge University Press. Google Scholar
Chatterji, S., Dennis, G. J., Helsby, W. I., Dent, A., Diaz-Moreno, S., Cibin, G. & Tartoni, N. (2016). AIP Conf. Proc. 1741, 040030. Google Scholar
Chen, W., Elliott, D., Giacomini, G., Kuczewski, A., Pinelli, D., Rumaiz, A. K., Siddons, D. P. & Smith, G. (2017). 2017 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 21–28 October 2017, Atlanta, GA, USA. Google Scholar
Ciatto, G., d'Acapito, F., Boscherini, F. & Mobilio, S. (2004). J. Synchrotron Rad. 11, 278–283. Web of Science CrossRef CAS IUCr Journals Google Scholar
Cicco, A. D., Aquilanti, G., Minicucci, M., Principi, E., Novello, N., Cognigni, A. & Olivi, L. (2009). J. Phys. Conf. Ser. 190, 012043. CrossRef Google Scholar
Cirrincione, D. (2019). Nuov. Cim. C, 42, Article 235. Google Scholar
Cramer, S., Tench, O., Yocum, M. & George, G. (1988). Nucl. Instrum. Methods Phys. Res. A, 266, 586–591. CrossRef Web of Science Google Scholar
Cube, X. (2012). CUBE – The CMOS Preamplifier for Radiation Detectors, https://www.xglab.it/products/cube/. Sold under licenses from Politecnico di Milano (patents jp2012-172331, jp2012-172332) and Oxford Instruments (patents ep2710733b, gb2491111b, us9397626 and jp6441675). Google Scholar
De Giudici, G., Lattanzi, P. & Medas, D. (2015). In Synchrotron Radiation, edited by S. Mobilio, F. Boscherini and C. Meneghini. Heidelberg, Berlin: Springer. Google Scholar
Diaz-Moreno, S., Amboage, M., Basham, M., Boada, R., Bricknell, N. E., Cibin, G., Cobb, T. M., Filik, J., Freeman, A., Geraki, K., Gianolio, D., Hayama, S., Ignatyev, K., Keenan, L., Mikulska, I., Mosselmans, J. F. W., Mudd, J. J. & Parry, S. A. (2018). J. Synchrotron Rad. 25, 998–1009. Web of Science CrossRef IUCr Journals Google Scholar
Fabjan, C. W. & Schopper, H. (2020). Particle Physics Reference Library Volume 2: Detectors for Particles and Radiation. Springer. Google Scholar
Fano, U. (1947). Phys. Rev. 72, 26–29. CrossRef CAS Web of Science Google Scholar
Gatti, E. & Rehak, P. (1984). Nucl. Instrum. Methods Phys. Res. 225, 608–614. CrossRef CAS Web of Science Google Scholar
Gatti, E., Rehak, P. & Walton, J. T. (1984). Nucl. Instrum. Methods Phys. Res. A, 226, 129–141. CrossRef CAS Web of Science Google Scholar
Hafizh, I., Bellotti, G., Carminati, M., Utica, G., Gugiatti, M., Balerna, A., Tullio, V., Lepore, G., Borghi, G., Ficorella, F., Picciotto, A., Zorzi, N., Capsoni, A., Coelli, S., Bombelli, L. & Fiorini, C. (2019). J. Instrum. 14, P06027. Google Scholar
Hettiarachchi, G. M., Donner, E. & Doelsch, E. (2017). J. Environ. Qual. 46, 1139–1145. Web of Science CrossRef CAS PubMed Google Scholar
INFN, (2018). Infn-technology transfer – detector with optimized effective area, https://web.infn.it/TechTransfer/index.php/en/schede-brevetti-en/769-zampa-en. Google Scholar
Istituto Superiore per la Protezione e la Ricerca Ambientale (2008). Carta Geologica d'Italia. Google Scholar
Jones, K. W. (1999). In Microscopic X-rav Fluorescence Analysis. John Wiley & Sons. Google Scholar
Koningsberger, D. & Prins, R. (1988). X-ray Absorption: Principles, Applications, Techniques of EXAFS, SEXAFS, and XANES. New York: Wiley-Interscience. Google Scholar
Kravtsova, A. (2020). J. Surface Investig. 14, 135–149. Web of Science CrossRef CAS Google Scholar
Kuzmin, A. & Chaboy, J. (2014). IUCrJ, 1, 571–589. Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
Liu, P., Wang, Q., Jung, H. & Tang, Y. (2020). Energy Fuels, 34, 14333–14343. Web of Science CrossRef CAS Google Scholar
Mastelaro, V. R. & Zanotto, E. D. (2018). Materials, 11, 204. Web of Science CrossRef Google Scholar
Mele, F., Gandola, M. & Bertuccio, G. (2021). IEEE Trans. Nucl. Sci. 68, 379–383. Web of Science CrossRef CAS Google Scholar
Meneghini, C., Di Matteo, S., Monesi, C., Neisius, T., Paolasini, L., Mobilio, S., Natoli, C., Metcalf, P. A. & Honig, J. M. (2009). J. Phys. Condens. Matter, 21, 355401. Web of Science CrossRef PubMed Google Scholar
Ohta, A. (2015). Geostand. Geoanalyt. Res. 39, 87–103. Web of Science CrossRef CAS Google Scholar
Parsons, J. G., Aldrich, M. V. & Gardea-Torresdey, J. L. (2002). Appl. Spectrosc. Rev. 37, 187–222. Web of Science CrossRef CAS Google Scholar
Quartieri, S. (2015). Synchrotron Radiation, pp. 641–660. Springer. Google Scholar
Rachevski, A., Ahangarianabhari, M., Aquilanti, G., Bellutti, P., Bertuccio, G., Borghi, G., Bufon, J., Cautero, G., Ciano, S., Cicuttin, A., Cirrincione, D., Crespo, M. L., Fabiani, S., Ficorella, F., Gandola, M., Giuressi, D., Mannatunga, K. S., Mele, F., Menk, R. H., Olivi, L., Orzan, G., Picciotto, A., Rashevskaya, I., Sammartini, M., Schillani, S., Zampa, G., Zampa, N., Zorzi, N. & Vacchi, A. (2019). Nucl. Instrum. Methods Phys. Res. A, 936, 719–721. Web of Science CrossRef CAS Google Scholar
Ryan, C. G., Siddons, D. P., Kirkham, R., Li, Z. Y., de Jonge, M. D., Paterson, D. J., Kuczewski, A., Howard, D. L., Dunn, P. A., Falkenberg, G., Boesenberg, U., De Geronimo, G., Fisher, L. A., Halfpenny, A., Lintern, M. J., Lombi, E., Dyl, K. A., Jensen, M., Moorhead, G. F., Cleverley, J. S., Hough, R. M., Godel, B., Barnes, S. J., James, S. A., Spiers, K. M., Alfeld, M., Wellenreuther, G., Vukmanovic, Z. & Borg, S. (2014). J. Phys. Conf. Ser. 499, 012002. CrossRef Google Scholar
Scheckel, K. G. & Ford, R. G. (2010). Develop. Soil Sci. 34, 147–169. CrossRef CAS Google Scholar
Strüder, L. (2000). Nucl. Instrum. Methods Phys. Res. A, 454, 73–113. CAS Google Scholar
Tack, F. & Verloo, M. G. (1995). Int. J. Environ. Anal. Chem. 59, 225–238. CrossRef CAS Web of Science Google Scholar
Terzano, R., Denecke, M. A., Falkenberg, G., Miller, B., Paterson, D. & Janssens, K. (2019). Pure Appl. Chem. 91, 1029–1063. Web of Science CrossRef CAS PubMed Google Scholar
Thompson, A., Mills, D., Naday, S., Hormes, J., Gruner, S., Siddons, P., Arthur, J., Wehlitz, R. & Padmore, H. (2000). White paper based on a workshop held in Washington DC, 30–31 October 2000. Google Scholar
This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.