view article

Figure 11
Temperature measured in the diffraction endstation hutch under static conditions for 6 h. Temperature of the sample (blue), the KB-vacuum chamber (green) and the granite (red). The temperature of the sample fluctuates more than for heavier components. 2σ deviations for respective temperature are 83 mK (sample), 9 mK (chamber) and 8 mK (granite).