Figure 8
CirPAD image obtained from the scattering of a glass sample, with an energy of 18 keV and for an integration time of 60 s. The left column shows the first 10 modules and the right column shows the other 10 modules. The mean value of the signal per pixel is around 2000 counts. The bright vertical bars correspond to the `double' pixel in the chip edge. The intense horizontal lines originate from the pixels close to the guard ring (Fig. 2[link]). The edge of the last module (lower left corner) was damaged during installation. Note that the arrangement of the modules in the figure does not represent the physical detector geometry.