Figure 9
Results of the performance evaluation of the spectrometer rotation. Elastic scattering from a copper plate was measured at the O K-edge at each 2θ angle, and (a) the FWHM and (b) the peak position were evaluated. In the entire 2θ range, the γ angle was set −0.15° away from the calculated value to align the Trident paths. |