PINK: a tender X-ray beamline for X-ray emission spectroscopy

crossmark logo

aDepartment of Inorganic Spectroscopy, Max Planck Institute for Chemical Energy Conversion, Stiftstrasse 34–36, Mülheim an der Ruhr, Germany, bBerlin Laboratory for Innovative X-ray Technologies (BLiX), Institute of Optics and Atomic Physics, Technical University of Berlin, Hardenbergstrasse 36, Berlin, Germany, and cHelmholtz-Zentrum Berlin für Materialien und Energie, Albert-Einstein-Strasse 15, Berlin, Germany
*Correspondence e-mail: sergey.peredkov@cec.mpg.de

Edited by K. Kvashnina, ESRF – The European Synchrotron, France (Received 12 January 2024; accepted 7 March 2024; online 25 April 2024)

A high-flux beamline optimized for non-resonant X-ray emission spectroscopy (XES) in the tender X-ray energy range has been constructed at the BESSY II synchrotron source. The beamline utilizes a cryogenically cooled undulator that provides X-rays over the energy range 2.1 keV to 9.5 keV. This energy range provides access to XES [and in the future X-ray absorption spectroscopy (XAS)] studies of transition metals ranging from Ti to Cu (Kα, Kβ lines) and Zr to Ag (Lα, Lβ), as well as light elements including P, S, Cl, K and Ca (Kα, Kβ). The beamline can be operated in two modes. In PINK mode, a multilayer monochromator (E/ΔE ≃ 30–80) provides a high photon flux (1014 photons s−1 at 6 keV and 300 mA ring current), allowing non-resonant XES measurements of dilute substances. This mode is currently available for general user operation. X-ray absorption near-edge structure and resonant XAS techniques will be available after the second stage of the PINK commissioning, when a high monochromatic mode (E/ΔE ≃ 10000–40000) will be facilitated by a double-crystal monochromator. At present, the beamline incorporates two von Hamos spectrometers, enabling time-resolved XES experiments with time scales down to 0.1 s and the possibility of two-color XES experiments. This paper describes the optical scheme of the PINK beamline and the endstation. The design of the two von Hamos dispersive spectrometers and sample environment are discussed here in detail. To illustrate, XES spectra of phosphorus complexes, KCl, TiO2 and Co3O4 measured using the PINK setup are presented.

1. Introduction

In studies of complex catalysts and metalloproteins, X-ray emission spectroscopy (XES) and X-ray absorption spectroscopy (XAS) play a major role (Bauer, 2014[Bauer, M. (2014). Phys. Chem. Chem. Phys. 16, 13827-13837.]; Kowalska et al., 2016[Kowalska, J. K., Lima, F. A., Pollock, C. J., Rees, J. A. & DeBeer, S. (2016). Isr. J. Chem. 56, 803-815.]; Castillo et al., 2020[Castillo, R. G., Henthorn, J. T., McGale, J., Maganas, D. & DeBeer, S. (2020). Angew. Chem. Int. Ed. 59, 12965-12975.]; Cutsail III & DeBeer, 2022[Cutsail, G. E. III & DeBeer, S. (2022). ACS Catal. 12, 5864-5886.]; Geoghegan et al., 2022[Geoghegan, B. L., Liu, Y., Peredkov, S., Dechert, S., Meyer, F., DeBeer, S. & Cutsail, G. E. (2022). J. Am. Chem. Soc. 144, 2520-2534.]). XES of transition metals in the Kβ main line and valence-to-core (VtC) spectral regions have attracted particular interest, as these regions provide element-selective information about spin states and coordination environments, respectively.

Recent experiments showed that analysis of weak VtC spectra can help to identify ligands surrounding the probing atom (Lancaster et al., 2011[Lancaster, K. M., Roemelt, M., Ettenhuber, P., Hu, Y., Ribbe, M. W., Neese, F., Bergmann, U. & DeBeer, S. (2011). Science, 334, 974-977.]; Pollock et al., 2013[Pollock, C. J., Grubel, K., Holland, P. L. & DeBeer, S. (2013). J. Am. Chem. Soc. 135, 11803-11808.]; Pollock & DeBeer, 2015[Pollock, C. J. & DeBeer, S. (2015). Acc. Chem. Res. 48, 2967-2975.]; Cutsail et al., 2019[Cutsail, G. E. III, Gagnon, N. L., Spaeth, A. D., Tolman, W. B. & DeBeer, S. (2019). Angew. Chem. Int. Ed. 58, 9114-9119.]; Levin et al., 2020[Levin, N., Peredkov, S., Weyhermüller, T., Rüdiger, O., Pereira, N. B., Grötzsch, D., Kalinko, A. & DeBeer, S. (2020). Inorg. Chem. 59, 8272-8283.]). Another promising approach complements non-resonant XES with advanced resonant-based techniques that help to overcome the limitations of conventional XAS (de Groot, 2001[Groot, F. de (2001). Chem. Rev. 101, 1779-1808.]; Szlachetko et al., 2013[Szlachetko, J., Sá, J., Safonova, O., Smolentsev, G., Szlachetko, M., van Bokhoven, J. & Nachtegaal, M. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 161-165.]; Lima et al., 2013[Lima, F. A., Bjornsson, R., Weyhermüller, T., Chandrasekaran, P., Glatzel, P., Neese, F. & DeBeer, S. (2013). Phys. Chem. Chem. Phys. 15, 20911.]; Glatzel et al., 2013[Glatzel, P., Weng, T.-C., Kvashnina, K., Swarbrick, J., Sikora, M., Gallo, E., Smolentsev, N. & Mori, R. A. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 17-25.]; Castillo et al., 2021[Castillo, R. G., Hahn, A. W., Van Kuiken, B. E., Henthorn, J. T., McGale, J. & DeBeer, S. (2021). Angew. Chem. Int. Ed. 60, 10112-10121.]). These methods are referred to as resonant XES (RXES) or resonant inelastic X-ray spectroscopy (RIXS), and involve recording an emission spectrum for each excitation energy of a regular X-ray absorption scan. Analysis of the resulting two-dimensional RXES (or RIXS plane) may allow, in certain cases, for higher-resolution XAS spectra and diminished background contributions (Hämäläinen et al., 1991[Hämäläinen, K., Siddons, D. P., Hastings, J. B. & Berman, L. E. (1991). Phys. Rev. Lett. 67, 2850-2853.]; Castillo et al., 2017[Castillo, R. G., Banerjee, R., Allpress, C. J., Rohde, G. T., Bill, E., Que, L., Lipscomb, J. D. & DeBeer, S. (2017). J. Am. Chem. Soc. 139, 18024-18033.]; Cutsail et al., 2018[Cutsail, G. E. III, Banerjee, R., Zhou, A., Que, L., Lipscomb, J. D. & DeBeer, S. (2018). J. Am. Chem. Soc. 140, 16807-16820.]). We note, however, that care must be taken in mapping RXES to XAS as there are significant deviations that may be possible depending on the transition metal multiplet structure, as well the exact RIXS plane being analyzed (e.g. 1s2p versus 1s3p).

XAS and XES using hard X-rays above 6 keV have become relatively routine practice today at synchrotron facilities. However, although today there are more than 60 synchrotron facilities worldwide, application for beam time is generally quite competitive and requires a proposal review process that may take up to six months or longer. During the last decade, a lot of effort has been put into the development of setups for performing XAS and XES experiments in laboratories (Malzer et al., 2018[Malzer, W., Grötzsch, D., Gnewkow, R., Schlesiger, C., Kowalewski, F., Van Kuiken, B., DeBeer, S. & Kanngießer, B. (2018). Rev. Sci. Instrum. 89, 113111.]; Seidler et al., 2014[Seidler, G. T., Mortensen, D. R., Remesnik, A. J., Pacold, J. I., Ball, N. A., Barry, N., Styczinski, M. & Hoidn, O. R. (2014). Rev. Sci. Instrum. 85, 113906.]; Błachucki et al., 2019[Błachucki, W., Czapla-Masztafiak, J., Sá, J. & Szlachetko, J. (2019). J. Anal. At. Spectrom. 34, 1409-1415.]). Laboratory-based setups are now more frequently found in research laboratories, providing rapid access to the instrument. Recently, an overview of the abilities of laboratory-based XAS/XES setups was reported (Zimmermann et al., 2020[Zimmermann, P., Peredkov, S., Abdala, P. M., DeBeer, S., Tromp, M., Müller, C. & van Bokhoven, J. A. (2020). Coord. Chem. Rev. 423, 213466.]).

While the hard X-ray regime has become increasingly accessible at both synchrotrons and in-house laboratories, access to setups in the tender X-ray regime is limited. However, many elements with 1s core XES in the 2–5 keV energy region, including P, S, Cl and Ca, play important roles in both biochemistry and catalysis (Mori et al., 2010[Mori, R. A., Paris, E., Giuli, G., Eeckhout, S. G., Kavčič, M., Žitnik, M., Bučar, K., Pettersson, L. G. M. & Glatzel, P. (2010). Inorg. Chem. 49, 6468-6473.]; Petric et al., 2015[Petric, M., Bohinc, R., Bučar, K., Žitnik, M., Szlachetko, J. & Kavčič, M. (2015). Anal. Chem. 87, 5632-5639.]; Qureshi et al., 2021[Qureshi, M., Nowak, S. H., Vogt, L. I., Cotelesage, J. J. H., Dolgova, N. V., Sharifi, S., Kroll, T., Nordlund, D., Alonso-Mori, R., Weng, T.-C., Pickering, I. J., George, G. N. & Sokaras, D. (2021). Phys. Chem. Chem. Phys. 23, 4500-4508.]). In addition, second row transition metal complexes (Mo, Ru, Rh etc.) actively used in photochemistry, catalysis and/or bioinorganic chemistry have also very informative Lβ, Lγ emission lines in the 2–5 keV energy region (Levin et al., 2020[Levin, N., Peredkov, S., Weyhermüller, T., Rüdiger, O., Pereira, N. B., Grötzsch, D., Kalinko, A. & DeBeer, S. (2020). Inorg. Chem. 59, 8272-8283.]). Experiments conducted in the tender X-ray region can present several challenges due to higher photon absorption in media, increased radiation damage and a more pronounced scattering background. All of this imposes additional requirements on the optical design of the beamline, the vacuum system and the sample environment. Access to XES at synchrotron facilities in the tender X-ray region can be very limited and highly competitive (Mori et al., 2010[Mori, R. A., Paris, E., Giuli, G., Eeckhout, S. G., Kavčič, M., Žitnik, M., Bučar, K., Pettersson, L. G. M. & Glatzel, P. (2010). Inorg. Chem. 49, 6468-6473.]; Rehanek et al., 2018[Rehanek, J., Milne, C. J., Szlachetko, J., Czapla-Masztafiak, J., Schneider, J., Huthwelker, T., Borca, C. N., Wetter, R., Patthey, L. & Juranić, P. (2018). J. Synchrotron Rad. 25, 16-19.]; Butorin et al., 2018[Butorin, S. M., Kvashnina, K. O., Klintenberg, M., Kavčič, M., Žitnik, M., Bučar, K., Gougeon, P., Gall, P., Candolfi, C. & Lenoir, B. (2018). ACS Appl. Energy Mater. 1, 4032-4039.]; Abraham et al., 2019[Abraham, B., Nowak, S., Weninger, C., Armenta, R., Defever, J., Day, D., Carini, G., Nakahara, K., Gallo, A., Nelson, S., Nordlund, D., Kroll, T., Hunter, M. S., van Driel, T., Zhu, D., Weng, T.-C., Alonso-Mori, R. & Sokaras, D. (2019). J. Synchrotron Rad. 26, 629-634.]; Rovezzi et al., 2020[Rovezzi, M., Harris, A., Detlefs, B., Bohdan, T., Svyazhin, A., Santambrogio, A., Degler, D., Baran, R., Reynier, B., Noguera Crespo, P., Heyman, C., Van Der Kleij, H.-P., Van Vaerenbergh, P., Marion, P., Vitoux, H., Lapras, C., Verbeni, R., Kocsis, M. M., Manceau, A. & Glatzel, P. (2020). J. Synchrotron Rad. 27, 813-826.]; Qureshi et al., 2021[Qureshi, M., Nowak, S. H., Vogt, L. I., Cotelesage, J. J. H., Dolgova, N. V., Sharifi, S., Kroll, T., Nordlund, D., Alonso-Mori, R., Weng, T.-C., Pickering, I. J., George, G. N. & Sokaras, D. (2021). Phys. Chem. Chem. Phys. 23, 4500-4508.]). Moreover, the availability of non-synchrotron based instruments for such experiments is also quite limited (Petric & Kavčič, 2016[Petric, M. & Kavčič, M. (2016). J. Anal. At. Spectrom. 31, 450-457.]; Malzer et al., 2018[Malzer, W., Grötzsch, D., Gnewkow, R., Schlesiger, C., Kowalewski, F., Van Kuiken, B., DeBeer, S. & Kanngießer, B. (2018). Rev. Sci. Instrum. 89, 113111.]; Holden et al., 2017[Holden, W. M., Hoidn, O. R., Ditter, A. S., Seidler, G. T., Kas, J., Stein, J. L., Cossairt, B. M., Kozimor, S. A., Guo, J., Ye, Y., Marcus, M. A. & Fakra, S. (2017). Rev. Sci. Instrum. 88, 073904.], 2018[Holden, W. M., Seidler, G. T. & Cheah, S. (2018). J. Phys. Chem. A, 122, 5153-5161.]; Abramson et al., 2023[Abramson, J. E., Holden, W. M., Rivera-Maldonado, R. A., Velian, A., Cossairt, B. M. & Seidler, G. T. (2023). J. Anal. At. Spectrom. 38, 1125-1134.]).

The goal of the current project is to offer world-class capabilities for studying complex catalysts with a strong emphasis on non-resonant XES in the tender X-ray energy region in combination with high sample throughput and time-resolved (∼50–100 ms) measurements that enable operando investigations. Another promising technique – resonant XES, which includes VtC-enhanced XAS offering the possibility to obtain ligand-selective XAS data (Hall et al., 2014[Hall, E. R., Pollock, C. J., Bendix, J., Collins, T. J., Glatzel, P. & DeBeer, S. (2014). J. Am. Chem. Soc. 136, 10076-10084.]; Maganas et al., 2017[Maganas, D., DeBeer, S. & Neese, F. (2017). Inorg. Chem. 56, 11819-11836.]) – will be implemented at the PINK beamline in the near future.

2. Beamline overview

The PINK branch was constructed in collaboration with the Helmholtz-Zentrum Berlin (HZB) on the EMIL beamline (Follath et al., 2013[Follath, R., Hävecker, M., Reichardt, G., Lips, K., Bahrdt, J., Schäfers, F. & Schmid, P. (2013). J. Phys. Conf. Ser. 425, 212003.]; Hendel et al., 2016[Hendel, S., Schäfers, F., Hävecker, M., Reichardt, G., Scheer, M., Bahrdt, J. & Lips, K. (2016). AIP Conf. Proc. 1741, 030038]) at the BESSY II synchrotron source and is depicted in Fig. 1[link]. X-rays are provided by one of two canted undulators installed at the EMIL beamline. A planar cryogenic in-vacuum U17 undulator provides photons from 700 eV to 10 keV and the PINK branch accepts the 2.1 keV to 9.5 keV energy range (3rd to 13th harmonics). In order to optimize the branch for photon-hungry VtC XES experiments, the number of optical elements was minimized to reduce flux losses. There are adjustable slits for the beam shaping at the front of each of the optical elements.

[Figure 1]
Figure 1
Layout of the PINK instrument showing the position (in m) of the beamline components with respect to the center of the U17 undulator. Beamline optics: toroidal (R = 2614923 mm, r = 197 mm) mirror M1, optional DCM, cylindrical (r = 907.5 mm) multilayer monochromator M2; motorized apertures: AU1, AU2, AU3, AU4; filters: F1 – set of water-cooled diamond attenuators (5 µm, 10 µm and 20 µm), F2 – three sets of six attenuators each (C, Al and Ta); optical beam-position monitors: BPM1, BPM2 and BPM3; Quadrant I0 monitor: Q; photon shutter: PS; vacuum windows: W1 – 12 µm Be, W2 – 10 µm diamond, W3 – 8 µm, 13 µm or 25 µm Kapton; sample: S. Atmospheric VH#1 and vacuum VH#2 von Hamos spectrometers; energy-dispersion axis: E.

The optical system has two operation modes: a primary high flux or `PINK' mode with photon flux up to 1014 photons s−1 at 300 mA of the ring current serves for non-resonant XES. In the high-flux mode, the X-ray beam passes only two optical elements. A bilayer-coated (8 nm Rh top layer and 35 nm Pt bottom layer) toroidal mirror (M1) is set at 0.4° grazing incidence angle. The water-cooled M1 mirror (which is utilized for all branches that use radiation from the U17 undulator) collimates the X-ray beam vertically and focuses it horizontally at a distance of 23 m. Unfortunately, horizontal focusing capabilities of the toroidal mirror M1 do not fully correspond to the designed parameters. The theoretical horizontal focus size is ∼500 µm full width at half-maximum (FWHM), but in actual use it has a correlation with AU1 and AU3 apertures opening and varies between 500 µm and 2 mm. A set of exchangeable diamond filters (5 µm, 10 µm and 20 µm thick) is installed after M1. These filters cut low-energy radiation photons from the first undulator harmonic. This filtering action serves to reduce the power load on the downstream optical elements. The second, particularly important optical element, is a water-cooled multilayer (ML) monochromator (M2). It consists of nine vertically stacked horizontal stripes of multilayer material deposited on the three cylindrical mirrors (three stripes on each mirror). There are two kinds of multilayers: Cr/B4C and W/B4C. In the high-flux mode, M2 is operated at a fixed grazing incident angle of 2° and performs three tasks. The first is switching of the X-ray beam into the PINK branch. The second is focusing the beam vertically at the sample position. Due to the slightly different quality of the substrates measured, the vertical focus size varies from the theoretical to the measured values of 25 µm to 35 µm (FWHM). The third function of the M2 is photon-energy selection. Reflectivity of individual stripes as a function of the photon energy for a 2° angle of incidence is plotted in Fig. 2[link]. At this angle, the ML monochromator can pass photons with nine fixed energies. Each stripe corresponds to one photon energy (see Table 1[link]). The photon bandwidth of the multilayer stripes varies from 65 eV to 155 eV (FWHM) and almost the full photon flux of a single undulator harmonics can pass the monochromator generating the so-called `PINK beam'.

Table 1
Energy and bandwidth of the multilayer monochromator

Mirror 1 2 3
Stripe 1 2 3 1 2 3 1 2 3
Energy (keV) 2.3 3.0 4.0 5.2 5.8 6.8 7.3 8.0 9.5
FWHM (eV) 155 117 80 65 115 90 123 110 100
[Figure 2]
Figure 2
Reflectivity of the multilayer monochromator at a 2° incident angle. K- (black) and L-edges (blue) of chemical elements are depicted at the top.

The PINK photon flux at the sample position is reported in Fig. 3[link]. For the flux measurements, a Hamamatsu Si diode was inserted into a direct beam attenuated by F2 filters. At higher energies the measured flux agrees with the values calculated using the xrt code (Klementiev & Chernikov, 2014[Klementiev, K. & Chernikov, R. (2014). Proc. SPIE, 9209, 92090A.]). The discrepancy increases at lower photon energies. This can be associated with the fact that flux measurements at energies below 5–6 keV become more sensitive to the thicknesses of vacuum windows and absorbing foils. Additionally, at lower energies, we have to reduce the AU3 horizontal aperture size in order to decrease the power on the ML to avoid overheating of the mirror and shape the beam into the designed horizontal size of 500 µm (FWHM).

[Figure 3]
Figure 3
Calculated (red solid circles) and measured (red open circles) photon flux at the sample position for the high-flux (PINK) mode using a multilayer monochromator. Flux values are calculated and measured with the front aperture AU1 opened to 0.8 mm × 1.5 mm. The symbols Hn indicate the undulator harmonics used, where n is a harmonic number. The estimated flux in high-monochromatization mode with a DCM inserted is also shown.

Experiments involving RXES/XAS techniques can be done in a high monochromatic mode with reduced photon flux in which the DCM monochromator is inserted into the beam. This second operation mode is under commissioning at present. The outgoing monochromatic beam is 20 mm higher than a PINK beam and hence we have to raise the ML monochromator as well as the endstation. Since the monochromatic beam also passes the ML monochromators, its energy has to be set to the region needed for RXES/XANES measurement. This can be done by changing the angle of incidence of the ML monochromator. By varying the M2 angle in the range 2° ± 0.15°, we can shift the energy of the ML monochromator by approximately ±500 eV, thus nine multilayer stripes cover almost the full 2.1 keV to 9.5 keV energy range. The beamline vacuum tubing/apertures accept ±0.17° horizontal deflection of the beam. The maximum horizontal beam displacement at the sample position is ±80 mm. The endstation is situated on a movable motorized platform and connected to the beamline via flexible bellows that allow it to be adjusted to the beam height. Any measurements involving the DCM are done at a fixed M2 angle; no ML monochromator scanning is allowed, and the photon flux impinging on the sample is a product of convolution of the monochromatic beam with a multilayer mirror reflectivity curve.

3. Experimental station

3.1. Vacuum system

Using tender X-rays places additional constraints on the vacuum system. The PINK setup has three isolated areas. A 12 µm-thick and 7 mm-diameter Be window braised into a CF40 flange manufactured by Materion protects the ultra-high-vacuum part of the beamline and the storage ring. An in-house-made 10 µm CVD diamond vacuum window (6 mm diameter) separates 10−6 mbar vacuum of the diagnostic chamber incorporated into the endstation and a sample environment area operated between 10−6 mbar and 10 mbar pressure. When experiments are run under He atmosphere at ambient pressure, an additional self-made 25 µm Kapton window is installed downstream of the diamond window.

3.2. Beam diagnostic

Three beam-position monitors (BPMs) are installed for X-ray beam alignment: 3.5 m downstream of M2 (BPM1), 2 m before (BPM2) and 0.7 m after (BPM3) the sample position (as shown in Fig. 1[link]). The BPMs are equipped with motorized 100 µm-thick YAG and 20 µm-thick CVD diamond screens rotated through 45° about a vertical axis to the beam propagation direction and monitored by high-resolution video cameras. A 260 mm-long diagnostic chamber comprises a set of attenuators, I0 monitor and a photon shutter (F2, Q and PS in Fig. 1[link]). Three linear piezo-positioners carry a diode, two apertures (500 µm and 30 µm in diameter) and three sets of CVD diamond, Al and Ta foils of different thicknesses that can be combined for attenuation optimization. The transmission coefficient can be set from 0.9 to 0.0001 over the whole energy range. The X-ray exposure of the sample is controlled by a vacuum fast beam shutter (PS) Uniblitz XRSR6 that opens only at the acquisition time to prevent continuous exposure of the sample. The shutter has a large 6 mm aperture and an opening/closing time of ∼20 ms. The maximum operating frequency is 2 Hz.

The intensity and beam-position monitor I0 uses an array of four 10 mm × 10 mm PIN photodiodes (S3590-09 Hamamatsu) that detect radiation scattered from a foil in backscattering radiation geometry (Tono et al., 2011[Tono, K., Kudo, T., Yabashi, M., Tachibana, T., Feng, Y., Fritz, D., Hastings, J. & Ishikawa, T. (2011). Rev. Sci. Instrum. 82, 023108.]). There are three low-Z-material foils fixed on a linear piezo stage: 10 µm CVD diamond, 2 µm Si3N4 and 25 µm Kapton.

3.3. Dispersive spectrometer

High incoming photon flux and excessive radiation sensitivity of biomolecules determined the choice of the analyzer geometry. Dispersive von Hamos spectrometers detect X-rays of a range of energies simultaneously (van Hámos, 1933[Hámos, L. von (1933). Ann. Phys. 409, 716-724.]; Vane et al., 1988[Vane, C. R., Smith, M. S. & Raman, S. (1988). A Versatile, High-Efficient, High-Resolution von Hamos Bragg Crystal X-ray Spectrometer. Oak Rodge National Laboratory, Oak Ridge, Tennessee, USA.]; Hoszowska et al., 1996[Hoszowska, J., Dousse, J.-C., Kern, J. & Rhême, C. (1996). Nucl. Instrum. Methods Phys. Res. A, 376, 129-138.]). They are suitable for time-resolved measurements and are actively used at free-electron laser facilities (Alonso-Mori et al., 2012[Alonso-Mori, R., Kern, J., Sokaras, D., Weng, T.-C., Nordlund, D., Tran, R., Montanez, P., Delor, J., Yachandra, V. K., Yano, J. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 073114.]; Szlachetko et al., 2017[Szlachetko, J., Nachtegaal, M., Grolimund, D., Knopp, G., Peredkov, S., Czapla-Masztafiak, J. & Milne, C. (2017). Appl. Sci. 7, 899.]; Canton et al., 2023[Canton, S. E., Biednov, M., Pápai, M., Lima, F. A., Choi, T., Otte, F., Jiang, Y., Frankenberger, P., Knoll, M., Zalden, P., Gawelda, W., Rahaman, A., Møller, K. B., Milne, C., Gosztola, D. J., Zheng, K., Retegan, M. & Khakhulin, D. (2023). Adv. Sci. 10, 2206880.]). Nowadays, von Hamos spectrometers are becoming increasingly popular at synchrotrons and laboratories (Malzer et al., 2018[Malzer, W., Grötzsch, D., Gnewkow, R., Schlesiger, C., Kowalewski, F., Van Kuiken, B., DeBeer, S. & Kanngießer, B. (2018). Rev. Sci. Instrum. 89, 113111.]; Kalinko et al., 2020[Kalinko, A., Caliebe, W. A., Schoch, R. & Bauer, M. (2020). J. Synchrotron Rad. 27, 31-36.]; Zimmermann et al., 2020[Zimmermann, P., Peredkov, S., Abdala, P. M., DeBeer, S., Tromp, M., Müller, C. & van Bokhoven, J. A. (2020). Coord. Chem. Rev. 423, 213466.]). The PINK setup is equipped with two dispersive von Hamos analyzers that can be operated simultaneously. The current design achieves unique two-color XES collection capabilities that offer scope for more detailed investigation of catalysts and metalloenzymes with more than one metal at the active sites. Designing the spectrometers, we aimed to strike a balance between resolution, efficiency, the space available for the sample environment and cost. Our main concern at first was to obtain the highest efficiency while maintiaining a moderate energy resolution because the natural width of Kβ, Lβ and VtC lines is usually larger than 1–1.5 eV.

A von Hamos spectrometer resolution is mainly affected by (1) the X-ray beam spot size at the sample position, (2) the analyzer crystal radius of curvature and diffraction angle, (3) the sample surface orientation relative to the dispersive crystal and the X-ray penetration depth into the sample, and (4) the detector pixel size. The small vertical size of the PINK beam (30 µm FWHM) offers an advantage to an analyzer with energy dispersion in the vertical direction. Higher resolution can be achieved by increasing the bending radius, but a price must be paid in the loss of efficiency. Building a large crystal array can increase the solid angle of collection (Alonso-Mori et al., 2012[Alonso-Mori, R., Kern, J., Sokaras, D., Weng, T.-C., Nordlund, D., Tran, R., Montanez, P., Delor, J., Yachandra, V. K., Yano, J. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 073114.]; Kalinko et al., 2020[Kalinko, A., Caliebe, W. A., Schoch, R. & Bauer, M. (2020). J. Synchrotron Rad. 27, 31-36.]). The downside of this approach is a dramatic increase of costs and the extra space required. Thus, we have prioritized a single-crystal solution.

We did not investigate cylindrically bent crystals with very short radii (R ≤ 250 mm) because using them significantly limits room for a sample environment. That aside, manufacturing of bent short-radius crystals without elastic deformations – which can significantly reduce their intrinsic resolution – is extremely challenging. Striping of the crystal is an alternative way of manufacturing short-working-distance crystals (Szlachetko et al., 2012[Szlachetko, J., Nachtegaal, M., de Boni, E., Willimann, M., Safonova, O., Sa, J., Smolentsev, G., Szlachetko, M., van Bokhoven, J. A., Dousse, J.-C., Hoszowska, J., Kayser, Y., Jagodzinski, P., Bergamaschi, A., Schmitt, B., David, C. & Lücke, A. (2012). Rev. Sci. Instrum. 83, 103105.], 2017[Szlachetko, J., Nachtegaal, M., Grolimund, D., Knopp, G., Peredkov, S., Czapla-Masztafiak, J. & Milne, C. (2017). Appl. Sci. 7, 899.]). Here the cylindrical surface is formed from an array of small flat stripes, but manufacturing capabilities and the quality of such crystals are also very limited. A strip bent crystal also has a larger focus size. A bent crystal focuses the source width to detector down to 1:1. A striped crystal forms an image on the detector with a width in the focusing direction that is twice the stripe width.

Optimization of the spectrometer geometry was done using xrt ray-tracing software (https://xrt.readthedocs.io). Ray tracing was performed for a collection of nine crystals that efficiently covered the 2 keV to 9.5 keV energy range (see Fig. 4[link]). Curvature radii of 500 mm, 350 mm and 250 mm were chosen for the simulation. We also considered both bent and striped crystals. The ray-tracing simulations for ideal bent crystals of R = 250 mm showed that, for a beam size of 30 µm × 500 µm (FWHM, V × H) and Bragg angles in the range 80° to 50°, resolution varies between 0.2 eV and 0.4 eV, and for a 1 mm strip bent crystal between 0.2 eV and 0.8 eV. For larger radii, the resolution varies between 0.1 eV and 0.3 eV. Both PINK spectrometers are designed to accept R = 250 mm and R = 350 mm analyzer crystals. Currently only shorter radius R = 250 mm crystals are available. The bent crystals were produced by Bourevestnik (St Petersburg, Russia). The strip bent crystals with a segment width of 1 mm were manufactured in collaboration with the Paul Scherrer Institute (PSI). The analyzer crystal dimension in the focusing (horizontal) direction is 100 mm and in the dispersive direction varies from 25 mm to 50 mm depending on the crystal. The specifications of both spectrometers are presented in Table 2[link].

Table 2
Working specification of the atmospheric and vacuum von Hamos spectrometers installed at the PINK beamline

All measured samples were in powder form unless indicated by an asterisk.

  Atmospheric VH#1 Vacuum VH#2
Crystal radius R = 250 (mm)
Available crystals Bent: Si(100), Si(110), Si(111), Si(311), quartz(1012)
Strip bent: Si(100), Si(1110), Si(111), Ge(100), Ge(110), Ge(111), Si(310)
Crystal dimensions Quartz(1012): 25 mm × 100 mm; Si(311): 30 mm × 100 mm; other: 50 mm × 100 mm
Bragg angle 45–82° (possible) Chamber 1: 59–81°
60–82° (in use) Chamber 2: 49–61° (S, Mo)
Energy range 6–10 eV 2–6 eV
Energy window 300–500 eV 20–80 eV
Energy resolution 0.4–1 eV
Solid angle (0.9–1.5) × 10−4 sr eV−1
Detector Mythen2 1K X (8 mm × 64 mm) CCD GreatEyes (6 mm × 26 mm)
Eiger2 R 500K (40 mm × 80 mm)
Typical acquisition time Co VtC (Co2O3): 20 min P Kβ (KH2PO4): 3 min
Fe VtC (FeCl3 10 mM): 12 h* P Kβ (Na3ADP, 30 mM): 40 min*
Cu VtC (CuCl): 40 min K VtC (KCl): 5 min
  Ru Lβ2 {[Ru(tacn)2](PF6)2}: 1 h
  Ru Lγ1 [Ru(NH3)6Cl3]: 15 min
[Figure 4]
Figure 4
Energy ranges available for the nine-crystal library at PINK: Si(111), Ge(111), Quartz (1012), Si(110), Ge(110), Si(311), Si(100), Ge(100) and Si(310). The majority of the 2 keV to 9.5 keV energy range can be covered within 80° to 50° Bragg angles.
3.3.1. Atmospheric von Hamos spectrometer

The atmospheric von Hamos spectrometer is designed for measurements above 6 keV (VH#1 on Fig. 1[link]). The analyzer crystal is oriented at 90° with respect to the X-ray propagation direction in order to reduce the elastic scattering background and at 45° to the sample surface. The analyzer focuses X-rays in the horizontal direction and spatially separates different energies in the vertical direction. The spectrometer is built on the constant exit-direction scheme where the Bragg angle of a dispersive crystal is changed by the crystal rotation. In this context, `constant exit direction' denotes the consistent position of the crystal relative to the sample. This is in contrast to a classical von Hamos scheme where linear translation is used. In the geometry used at the PINK endstation, the center of the dispersive crystal constantly stays in the horizontal plane passing through the incoming X-ray beam, and the source of X-ray emission and the analyzer crystal always stay on the same axis. This has two advantages. Independent of the crystal Bragg angle, the maximum vertical opening needed to illuminate the 50 mm-high crystal is only 11.4°, which allows the vertical height of the exit window on the sample chamber to be minimized. Additionally, in this configuration, the spectrometer is effectively insensitive to sample thickness or propagation depth of the X-ray beam into the sample (which increases with higher incoming X-ray energies).

The mechanical implementation of the spectrometer is shown in Fig. 5[link] (left). The crystal (C) and the detector (D) are mechanically parallel at the distance r = 250 mm. The sample(S)-to-crystal distance a = [r/\sin(\Theta_{\rm Bragg})] is adjusted by the linear stage m1, moving the whole spectrometer as a unit. The angle of incidence of the X-rays onto the crystal is adjusted by means of a curved motorized rail segment (m3) where (R) is the axis of rotation. The same curved segment also rotates the linear stage m2 carrying the detector. A crystal-to-detector distance is set by the linear stage m2. Final adjustment of the dispersive image on the detector is done with a pair of piezo-stages rotating the crystal (yaw and roll).

[Figure 5]
Figure 5
View of the PINK atmospheric von Hamos spectrometer with a fixed-exit direction VH#1 (left) and vacuum von Hamos spectrometer VH#2 (right), where E is the energy dispersion axis. S – sample; C – analyzer crystal; D – detector; m1, m2 and m3 – motorized axes; F1 – exchangable flange; F2 – flange with hinges; R – rotation axis of the crystal.

Two detectors are currently available. The first is a 2D Eiger2 R 500K (75 µm × 7 5 µm pixel size, 512 × 1030 pixels). The second detector is a 1D Mythen2 1K X (50 µm × 8 mm stripe size, 1280 stripes). The advantage of using a 2D detector is easier alignment of the crystal by controlling the focus size. The design of the spectrometer allows Bragg angles between 82° and 45° to be reached. The upper Bragg angle is limited by the sample environment chamber dimensions. The typical energy window visible by the 80 mm-long Eiger detector is 300 eV to 500 eV. The average solid angle collected by the crystal and imaged by the detector is ∼0.05 sr and (0.9–1.5) × 10−4 sr eV−1 depending on the span of Bragg angles. For crystals with R = 250 mm, the X-ray path in air between the sample and the detector is approximately 500–540 mm. At 7 keV the air transmission remains above 40%. Therefore, the use of an He bag, which is commonly employed to minimize X-ray absorption in the air for spectrometers with bending radii of 0.5 m to 1 m, is not strictly necessary. This makes operating the spectrometer more convenient but at the cost of reduced signal.

3.3.2. Vacuum von Hamos spectrometer

The vacuum von Hamos spectrometer is designed for measurements between 2 keV and 6 keV and has a working pressure of ∼10−5 mbar. The vacuum spectrometer is fixed at 45° to the X-ray propagation direction and perpendicular to a sample surface (VH#2 in Fig. 1[link]). The detector is located above the sample.

A fixed-exit direction scheme has a number of advantages over a conventional von Hamos design (see Section 3.3.1[link]) but its engineering implementation in the case of a vacuum spectrometer is challenging. It requires a large vacuum chamber with an in-vacuum detector, water cooling for the detector and a number of motion stages. We implemented a classical von Hamos scheme but tilted 20° about the vertical axis (see Fig. 5[link], right). When the center of the dispersive crystal (C) is in the incident beam, the horizontal plane of emitted X-rays from a sample (S) illuminates the center of the crystal at a 70° Bragg angle. An analyzer crystal is mounted on an in-vacuum linear stage m1 inside the vacuum chamber. By moving the crystal along the m1 axis the Bragg angle can be changed by 70 ± 11°. In this case, the dispersive crystal is only slightly out of the horizontal plane and mimics the fixed-exit direction scheme. There is a pair of motorized blades in front of the crystal. These blades help to mask unused crystal surface in order to reduce diffuse scattering background. The upper blade carries a diode measuring the total fluorescence yield (TFY). This signal is also used to monitor synchronization of the fast shutter (Fig. 1[link], PS) and the CCD.

A windowless GreatEyes CCD detector (256 × 1024 pixels, 26 µm × 26 µm pixel size) is mounted on an exchangeable vacuum flange (F1) to mechanically guarantee the detector position and parallelism of the crystal and the detector (D). The field of view of the 26.6 mm-long CCD camera varies from 20 eV to 80 eV depending on the energy and Bragg angle.

Switching between XES measurements for different emission lines takes about 1.5–2 h. It requires venting of the spectrometer chamber, changing a dispersive crystal and mounting the detector to a new position. To make this procedure easier, the crystal mount and all electrical feedthroughs are fixed to a door with hinges and quick clamps (F2). We have a collection of seven exchangeable flanges (F1) and each flange has 2–3 pre-cut windows for the CCD mounting.

For the 2300–2500 eV and 5400–5500 eV energy ranges there is no available Si, Ge or quartz crystal with working Bragg angles larger than 60°. XES measurements of S Kβ, Mo Lβ and V Kβ lines demand operation of the vacuum spectrometer at 50° to 57° Bragg angles. Especially for these measurements, we manufactured a second vacuum chamber for a von Hamos spectrometer tilted by 40°. It covers Bragg angles of 50° ± 11°. The chambers use the same base frame and can be easily exchanged by a small crane.

3.3.3. Energy calibration

The most commonly used energy calibration procedure of an emission spectrometer applied at a synchrotron facility employs monochromatic light. Monochromatic X-rays provided by a monochromator were elastically scattered from a light-Z target placed at the sample position and recorded by an analyzer which gives a precise energy calibration. At the PINK beamline in high-flux mode there is no access to monochromatic light and, as an alternative, emission lines from reference samples are used for the calibration procedure. An example of an 4d-to-2p XES spectrum of the Ru complex is shown in Fig. 6[link]. This measurement was done with the vacuum von Hamos spectrometer using an SiO2(1012) crystal (d = 2.282 Å) with R = 250 mm (Levin et al., 2020[Levin, N., Peredkov, S., Weyhermüller, T., Rüdiger, O., Pereira, N. B., Grötzsch, D., Kalinko, A. & DeBeer, S. (2020). Inorg. Chem. 59, 8272-8283.]). In the present geometry, a 1 inch-long CCD detector can capture an approximately 50 eV-wide energy window. For energy calibration, KCl powder and Pd foil samples were used. To establish the energy scale, the positions of the Cl Kβ1, Pd Lα2 and Lα1 emission lines were picked by fitting both reference spectra with two Voigt profiles. Corresponding energies of 2815.6 eV, 2833.29 eV and 2838.61 eV for these three peaks were taken from the work of Thompson & Vaughan (2009[Thompson, A. & Vaughan, D. (2009). X-ray Data Booklet. Center for X-ray Optics and Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, USA.]) and Liu et al. (2004[Liu, Z., Yuge, K. & Kawai, J. (2004). At. Spectrosc. 59, 93-99.]) and converted to Bragg angles,

[\Theta_{\rm Bragg}(n) = \arctan\left(\,{{12398.419} \over {2d\,[{\rm \AA}]\,E\,[{\rm eV}]}}\,\right).]

For a von Hamos spectrometer, geometry X-rays diffracted at the Bragg angle ΘBragg arrive at the detector at a position described by

[\Theta_{\rm Bragg}(n) = \arctan\left({{2R} \over {a_{0}+ndx}}\right),]

where n is the detector pixel number, dx is the pixel size, a0 is the distance from the sample to the first pixel on the detector and R is the curvature radius of the crystal. By fitting this function with two free parameters R and a0, a one-to-one correspondence between the Bragg angle and the detector pixel number is obtained. By converting the Bragg angles to corresponding energies, we obtain the energy calibration E(n) of the spectrometer. This function is quasilinear for a 1 inch CCD detector, but the 80 mm-long Eiger chip already shows substantial E(n) nonlinearity. This energy calibration procedure has to be repeated after any realignment of the beamline or analyzer optics, or after sample exchange.

[Figure 6]
Figure 6
Ru 4d-to-2p XES spectrum (top) and reference spectrum (bottom) used for the energy calibration procedure. The [Ru(tacn)2](PF6)2 spectrum was adapted from the work by Levin et al. (2020[Levin, N., Peredkov, S., Weyhermüller, T., Rüdiger, O., Pereira, N. B., Grötzsch, D., Kalinko, A. & DeBeer, S. (2020). Inorg. Chem. 59, 8272-8283.]),

Reproducibility of the energy calibration process relies mainly on the accuracy of the peak-fitting procedure applied to reference spectra. In day to day measurements, the reproducibility is generally within a single detector pixel and is typically less than ±30 meV. We repeated K Kβ main line XES measurements performed on KCl, as outlined in Section 4[link] (see Fig. 13), with a two-year interval after the beamline upgrade and vacuum spectrometer rearrangement. Energy calibration was achieved using Sb Lα1 (3604.72 eV) and Lα2 (3595.32 eV) lines measured at the Sb foil. The energy of the Kβ13 peak exhibited variations within the range ±50 meV.

3.4. Sample environment

Radiation damage is an important consideration for XES/XAS studies of many molecular complexes and metalloproteins (Van Schooneveld & DeBeer, 2015[Schooneveld, M. M. van & DeBeer, S. (2015). J. Electron Spectrosc. Relat. Phenom. 198, 31-56.]). The PINK beamline delivers a high photon flux, particularly in the tender X-ray energy range. This elevated photon flux amplifies the acquisition of valid signals but concurrently escalates the rate of radiation-induced damage. Consequently, the damage rate becomes notably pronounced at lower photon energies within the range 2 keV to 5 keV, primarily due to the reduction in the attenuation length of the material. There are three primary ways to measure radiation-sensitive specimens: scanning of the sample, using high flow rate liquid cells and subjecting samples to cryogenic temperatures (Meents et al., 2010[Meents, A., Gutmann, S., Wagner, A. & Schulze-Briese, C. (2010). Proc. Natl Acad. Sci. USA, 107, 1094-1099.]). We have incorporated all three of these approaches into our experimental protocol. There are two sample chambers that can be incorporated into the PINK vacuum system. Both chambers as well as the samples inside are oriented at 45° to the incident beam.

3.4.1. LiquidSEC

A liquid sample environment chamber (LiquidSEC) is available for room-temperature (RT) experiments, which operates under vacuum conditions of 2 mbar to 10 mbar or an He atmosphere up to 1 bar. The vertical dimension of the chamber is deliberately limited to 10 cm to create additional room for both detectors, thereby facilitating measurements at Bragg angles of up to ∼82°. Two linear piezo-stages provide fast (up to 2 mm s−1) vertical and horizontal positioning of a sample carrier that can hold four solid samples and a liquid-flow cell or electrochemical cell. A miniature camera installed inside the chamber is used for alignment and control of operating the flow/electrochemical cells. The chamber has two exit windows for fluorescent radiation: a 13 µm-thick Kapton window (31 mm × 18 mm) oriented towards the atmospheric von Hamos spectrometer (VH#1 on Fig. 1[link]) and an 8 µm-thick Kapton window (18 mm × 18 mm) that separates the sample environment chamber and the second vacuum von Hamos spectrometer (VH#2 on Fig. 1[link]). The later window enables sample exchange without venting the vacuum spectrometer. The chamber has two feedthroughs for liquid supplies and a DSUB-9 socket for connecting a potentiostat and an LED light source for studying photoinduced reactions. KF40 flanges are used to connect the sample chamber to the diagnostic chamber from the upstream side and to the BPM3 chamber on the downstream side. Continuous scanning of a sample irradiation area at high speeds (500–1000 µm s−1) helps to reduce the radiation dose at a given a spot. By varying the scanning speed, we can reduce the instantaneous irradiation time down to 100 ms to 50 ms during one scan over the sample area. An example of radiation damage of the [Ru(NH3)6]Cl3 complex under an incoming photon flux of 1013 photons s−1 is shown in Fig. 7[link]. A 64 mm2 powder sample was scanned with 500 µm s−1 velocity that is equivalent to 100 ms radiation at the spot. The measurement was repeated ten times with increasing radiation dose. The Lγ1 XES spectra presented show no changes between 0.1 s and 0.2 s radiation time. The first changes appeared around 2963 eV after ∼0.4 s and became more pronounced after 1 s. The conventional scanning spectrometers also allow fast sample motion, but analysis of the spectra requires additional normalization by the total fluorescence signal, which can be challenging.

[Figure 7]
Figure 7
Radiation damage of [Ru(NH3)6]Cl3 under 1013 photons s−1 photon flux. The instantaneous irradiation time during one scan over the sample area is 0.1 s. The measurement time for one pass is 4 min.
3.4.2. CryoSEC

Cryogenic conditions for XES/XAS measurements are highly desirable at synchrotron facilities, as the low temperatures help to reduce the radiation damage of the samples, stabilize reactive complexes, prevent decomposition of intermediates and enable preparation of frozen solutions. The highly intense PINK beam bears up to several watts of energy onto the 30 µm × 500 µm spot. For example, at 3 keV in water, 85% of the X-ray photons are absorbed within 100 µm. Powder and biological samples typically have poor thermal conductivity. To efficiently cool down samples and avoid hot spots, an He exchange gas must be used.

A cryogenic sample environment chamber (CryoSEC) was designed entirely by the sample environment group of HZB (see Figs. 8[link] and 9[link]). The CryoSEC consists of an internal cold sample chamber (T2) that can be operated under 10 mbar to 1000 mbar He pressure and outer insulation vacuum chambers (T1). The insulation chamber is connected to the diagnostic chamber on the front side and to the BPM3 chamber on the back side, and maintained at 10−5 mbar pressure. To transfer heat from the sample to a cold Cu exchange tube, He gas at a pressure of 5–20 mbar is used. This tube is connected to the cold finger of the cryocooler via a Cu braid. The Sumitomo RDK-500B one-stage closed-cycle He cryocooler (Cryo) is situated at the bottom of the chamber and has a cooling capacity of 40 W. It can cool the sample down to temperatures as low as 30 K. Heating elements inside and around the cold chamber are used to heat the setup. A Lakeshore Model 336 temperature controller regulates the displex performance and temperature of the exchange gas. The typical time needed to cool down the setup from RT or warm up to RT is 3 h. The whole CryoSEC is mounted on moveable rails (R). It can be detached from the beamline and removed for maintenance while the LiquidSEC sample environment is in use.

[Figure 8]
Figure 8
View of the cryogenic sample environment chamber CryoSEC. T1 – vacuum chamber; Kapton vacuum windows: W2 – 8 µm, W3 – 25 µm; m1 – motor for horizontal motion of the sample carrier; R – THK rails; Cryo – closed-cycle cryocooler; B – bellow; D – vibration dumper.
[Figure 9]
Figure 9
Scheme of a part of the CryoSEC connected to the beamline and the spectrometers. T1 – insulation tube; T2 – cold copper tube; P – pin centering the inner copper tube; C – sample carrier; S – irradiated sample; m1 – linear stage providing horizontal motion of a sample carrier (90 mm); m2 – vertical linear piezo stage (±8 mm); H – heater for the piezo-stage; G1 and G2 – graphenic carbon 1 µm-thick windows; Kapton vacuum windows: W1 – 13 µm, W2 – 8 µm, W3 – 25 µm; F – fluorescent light emitted from the sample in the direction of the spectrometers.

Studies of dilute samples in the 2 keV to 6 keV photon energy range require proper attention to vacuum windows manufacturing. Attenuation of the incoming and fluorescent photons should be minimized. However at the same time the windows in the cold sample chamber should withstand a 1 atm differential pressure at 30 K. In addition, the incoming window should not degrade under a high-flux beam. The insulation chamber has exit windows of a similar geometry to the LiquidSEC (see Section 3.4.1[link]). The cold chamber has an exchangeable Al flange with two windows glued with cryogenic epoxy (Stycast 2850FT, Catalyst23LV) to the flange. A 8 mm × 34 mm (H × W) 8 µm-thick Kapton window (W1) serves as an exit window for the fluorescent radiation. The choice of material for the entrance window is more challenging. Even thin 8 µm Kapton will be quickly burned by intense photon beams at energies of 2 keV to 4 keV. Be is used in cryogenic applications but is toxic and becomes fragile at low temperatures. A good alternative is a 1 µm-thick graphenic carbon window (G1, G2) produced by KETEK for RT applications (Huebner et al., 2015[Huebner, S., Miyakawa, N., Kapser, S., Pahlke, A. & Kreupl, F. (2015). IEEE Trans. Nucl. Sci. 62, 588-593.]). Our test showed that this window can be also used under cryo-temperatures. This graphenic carbon window has an opening of 7 mm and 90% transmission at 2 keV. We have not observed any degradation after one year of use at the PINK beamline.

During the design phase, emphasis was placed on efficient sample-exchange capabilities because sample changes occur when the cold chamber is opened to atmospheric air. A 1 m-long sample stick with a motorized head is connected to a vacuum flange firmly attached to horizontal rail system (not shown in Fig. 8[link]), allowing easy sliding of the sample stick from and into the cold chamber. The head of the sample stick is driven by a stepper motor (m1) and can travel 90 mm horizontally. An Attocube cryogenic linear piezo-stage ANPx321 (m2) is attached to the head and provides vertical motion of the sample carrier by ±8 mm. There is an additional heater (H) attached to the back side the piezo-stage. It is used to locally heat the piezo-stage if the cryostat becomes blocked due to ice accumulation. The sample carrier (C) made from PEEK can hold up to seven samples. Samples are loaded into slots via openings on the top of the carrier and forced against the front part of carrier by a Cu–Be spring. Sample loading can be comfortably done in liquid nitrogen or in a glovebox. Afterwards the loaded sample carrier can be contained in a vessel with liquid nitrogen. The sample carrier has a trapezoidal cross-section and can be slid into a machined response groove at the piezo-stage with a single, effortless motion. After venting of the sample chamber with He, the sample-exchange procedure usually takes 15–20 s and can be accomplished by one person. Furthermore, this procedure can be conducted while the cold chamber remains under cryogenic temperature.

A number of XES measurements on S, Fe and Cu complexes using the CryoSEC have already been performed at the PINK beamline and published in the literature (Levin et al., 2020[Levin, N., Peredkov, S., Weyhermüller, T., Rüdiger, O., Pereira, N. B., Grötzsch, D., Kalinko, A. & DeBeer, S. (2020). Inorg. Chem. 59, 8272-8283.]; Mathe et al., 2021[Mathe, Z., McCubbin Stepanic, O., Peredkov, S. & DeBeer, S. (2021). Chem. Sci. 12, 7888-7901.]; Geoghegan et al., 2022[Geoghegan, B. L., Liu, Y., Peredkov, S., Dechert, S., Meyer, F., DeBeer, S. & Cutsail, G. E. (2022). J. Am. Chem. Soc. 144, 2520-2534.]; Gerz et al., 2021[Gerz, I., Jannuzzi, S. A. V., Hylland, K. T., Negri, C., Wragg, D. S., Øien-Ødegaard, S., Tilset, M., Olsbye, U., DeBeer, S. & Amedjkouh, M. (2021). Eur. J. Inorg. Chem. 2021, 4762-4775.]; Hou et al., 2023[Hou, K., Börgel, J., Jiang, H. Z. H., SantaLucia, D. J., Kwon, H., Zhuang, H., Chakarawet, K., Rohde, R. C., Taylor, J. W., Dun, C., Paley, M. V., Turkiewicz, A. B., Park, J. G., Mao, H., Zhu, Z., Alp, E. E., Zhao, J., Hu, M. Y., Lavina, B., Peredkov, S., Lv, X., Oktawiec, J., Meihaus, K. R., Pantazis, D. A., Vandone, M., Colombo, V., Bill, E., Urban, J. J., Britt, R. D., Grandjean, F., Long, G. J., DeBeer, S., Neese, F., Reimer, J. A. & Long, J. R. (2023). Science, 382, 547-553.]; Liu et al., 2023[Liu, Y., Chatterjee, S., Cutsail, G. E. III, Peredkov, S., Gupta, S. K., Dechert, S., DeBeer, S. & Meyer, F. (2023). J. Am. Chem. Soc. 145, 18477-18486.]). Tests conducted with frozen solutions of Fe and Cu containing samples demonstrated that it is feasible to obtain Kβ VtC emission spectra even at concentrations as low as 10 mM to 20 mM. However, these measurements are time-consuming, typically taking 8 h to 12 h to complete, and are not suitable for samples that are sensitive to radiation. Considering the elevated flux offered by the PINK beamline at lower energies and the augmented fluorescence cross-section of both light elements (Kβ) and transition elements (Lβ), we anticipate achieving lower concentrations ranging from 3 mM to 5 mM for these elements. At present, this is hindered by the significant contribution of scattering background below 4 keV, which begins to impact the experimental spectrum. Any minor imperfection on the sample surface, presence of gas bubbles or similar factors can introduce individual background patterns and artificial features on the detector. Usually 10% to 30% of the sample surface produces such `bad' spectra. Correction of such irregular background requires additional analysis of the detector images and can be challenging.

3.5. Control system

The efficiency and success of an experiment depend on the stability and ease of use of a control system. In addition, the endstation control requires high flexibility and maintainability. Development of the control system was based on three principles. First, use of free and open source software, and avoiding any proprietary software or drivers. Second, development of a friendly framework that enables integration with Python language and HDF5 data format. Third, users and staff should be able to write their own control and acquisition components on demand. After an extended series of tests, a combination of Experimental Physics and Industrial Control System (EPICS) as a base (https://epics-controls.org), CS-Studio and Phoebus as a GUI interface (https://www.controlsystemstudio.org), and PShell (https://github.com/paulscherrerinstitute/pshell) as an acquisition framework were chosen. The beamline IT infrastructure is based on a Cisco network switch and two Dell PowerEdge servers with multiple network cards. One server uses a series of virtual machines running the different control system services for hardware communication and control. The second server is used for data storage and data analysis. To better organize data exchange, the network is divided into several sub-networks for cameras and detectors, vacuum interlock, hardware communication and guest internet access. Vacuum interlock is managed by a programmable logic controller (PLC). The PLC reads analog signals from multiple vacuum sensors across the beamline and with the combination of vacuum valves it provides control and safe operation of the multiple vacuum components of the beamline. All computers used at the beamline run a long term service (LTS) version of Ubuntu Linux. An electronic logbook based on ELOG is available for users of the beamline to log their experimental information. During operation of the beamline, specific details such as datafile names, sample names, sample motion parameters and detector settings are automatically recorded using Python scripts, which store this information in Google Docs. Each project has its dedicated logbook for organization and data management.

4. Results

In this section, the capabilities of the PINK beamline and a comparison with other setups are presented.

A test of the quality of the striped crystal manufacturing was carried out by analysis of the Fe Kα1, 2 spectrum recorded with the atmospheric spectrometer using an Si(333) striped crystal set at a 68° Bragg angle (see Fig. 10[link]). A ray-tracing simulation gives the resolution for a 1 mm striped crystal with R = 250 mm of ∼0.3 eV. The observed emission line yields both the Lorentzian natural linewidth and the Gaussian instrument broadening. We utilized the lmfit Python library for fitting purposes. We employed an asymmetric Lorentzian function where the distribution width differs between the left and right slopes. Each emission line was then represented as a convolution of this asymmetric Lorentzian profile with a Gaussian profile. The experimental spectrum was fitted using this newly constructed function. Taking a Lorentzian width of Fe Kα1 and Fe Kα2 lines to 2.55 eV and 3.14 eV, respectively (Hölzer et al., 1997[Hölzer, G., Fritsch, M., Deutsch, M., Härtwig, J. & Förster, E. (1997). Phys. Rev. A, 56, 4554-4568.]), the resulting experimental Gaussian broadening, the FWHM of which reflects the spectrometer resolution, was approximately 0.4 eV. This value closely matches that predicted by the ray-tracing value.

[Figure 10]
Figure 10
Normalized Kα1, 2 XES spectrum of Fe foil (red open circles) recorded on the atmospheric spectrometer with a fixed-exit direction using a strip bent Si(111) crystal and a best fit to the data (blue line).

Resolution capabilities of the vacuum von Hamos spectrometer are demonstrated in Fig. 11[link]. In the work by Swarbrick et al. (2010[Swarbrick, J. C., Kvashnin, Y., Schulte, K., Seenivasan, K., Lamberti, C. & Glatzel, P. (2010). Inorg. Chem. 49, 8323-8332.]), it was shown how VtC XES can be applied in the identification of different ligands in Ti complexes. Recent calculations carried out by Miaja-Avila et al. (2021[Miaja-Avila, L., O'Neil, G. C., Joe, Y. I., Morgan, K. M., Fowler, J. W., Doriese, W. B., Ganly, B., Lu, D., Ravel, B., Swetz, D. S. & Ullom, J. N. (2021). X-ray Spectrom. 50, 9-20.]) predicted splitting of the Kβ2, 5 emission line on two peaks having different relative intensities. Follow-up measurements done with a microcalorimeter spectrometer could resolve Kβ′ and Kβ2, 5 lines but could not resolve the predicted fine structure of the Kβ2, 5 line due to the low resolution of the microcalorimeter (∼4 eV).

[Figure 11]
Figure 11
XES VtC spectrum of TiO2 in the form of rutile and anatase.

XES VtC measurements of rutile and anatase powders were performed at the PINK beamline using the vacuum von Hamos spectrometer with an Si(400) striped crystal at Bragg angles of 66.4–69°. In this geometry, the estimated spectrometer resolution was expected to be 0.8 eV or better. The excitation energy was set to 5800 eV. Samples were measured using commercially available powders (Sigma–Aldrich, >99.9% purity). Powders were finely ground, packed into a 1 mm Al cell and sealed from both sides with 30 µm-thick Kapton tape. For the energy calibration procedure, Kα1, 2 emission lines [4952.2 eV and 4944.64 eV (Thompson & Vaughan, 2009[Thompson, A. & Vaughan, D. (2009). X-ray Data Booklet. Center for X-ray Optics and Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, USA.])] of a V powder sample were recorded. The predicted splitting of the Kβ2, 5 line is clearly visible in the experimental spectra shown in Fig. 11[link].

The lightest element that is reachable at the PINK beamline is P. Solid spectra of orthophosphates NaH2PO4 and Na2HPO4 were measured in earlier work by Petric et al. (2015[Petric, M., Bohinc, R., Bučar, K., Žitnik, M., Szlachetko, J. & Kavčič, M. (2015). Anal. Chem. 87, 5632-5639.]). In Fig. 12[link], P Kβ solution spectra of NaH2PO4 and Na2HPO4 are shown, collected at RT using a liquid cell installed in the LiquidSEC sample chamber (Mathe et al., 2021[Mathe, Z., McCubbin Stepanic, O., Peredkov, S. & DeBeer, S. (2021). Chem. Sci. 12, 7888-7901.]). The data were collected at an excitation energy of 4 keV (incoming flux 2 × 1013 photons s−1) using the vacuum spectrometer. An Si(111) crystal accepted Bragg angles of 69.3° to 66.7° with the corresponding energy range 2110 eV to 2150 eV. The analyzer resolution was estimated to be ∼0.3 eV. For the energy calibration procedure, a powder NaH2PO4 standard was chosen. The Kβ XES NaH2PO4 spectrum was fitted with four Voigt profiles that correspond to 2139.5 eV, 2137.9 eV, 2135.3 eV and 2123.4 eV (Petric et al., 2015[Petric, M., Bohinc, R., Bučar, K., Žitnik, M., Szlachetko, J. & Kavčič, M. (2015). Anal. Chem. 87, 5632-5639.]).

[Figure 12]
Figure 12
Solution P Kβ spectra of HPO42− and H2PO4.

An example Kβ spectrum of a KCl powder sample measured at the PINK beamline with a vacuum spectrometer using an Si(220) strip bent crystal set at 64° Bragg angle is shown in Fig. 13[link]. The spectrum has three well resolved peaks. The main line Kβ1, 3 at 3589.9 eV is attributed to the 3p–1s transition. The approximately 5× lower intensity Kβx satellite peak at ∼3593.5 eV can be explained by double ionization of the K 3p level. A low-intensity signal at 3602 eV is assigned to the Kβ5 VtC transition (Deslattes, 1964[Deslattes, R. D. (1964). Phys. Rev. 133, A390-A398.]; Best, 1971[Best, P. E. (1971). Phys. Rev. B, 3, 4377-4382.]). A more detailed analysis of VtC spectra of K salts will be published soon. For comparison, a KCl spectrum measured at the LabXES laboratory setup using a metal jet X-ray source and a von Hamos spectrometer with a ring HAPG crystal is presented.

[Figure 13]
Figure 13
Comparison of K Kβ spectra of a KCl powder sample measured at the PINK beamline and LabXES laboratory setup at CEC MPI; the measurement times were 200 s and 1.5 h, respectively.

Scanning spectrometers based on Johann geometry are traditionally characterized with better signal-to-noise ratio than von Hamos spectrometers. Indeed, the detected photon flux per energy unit is usually much lower for long-radius von Hamos spectrometers, but shortening the distance between the sample and the dispersive crystal should improve the von Hamos spectrometer performance. Fig. 14[link] presents the Co Kβ VtC spectrum of Co3O4 powder recorded at the I20 beamline (Diamond Light Source) with a Johann emission spectrometer based on a 1 m Rowland circle and the PINK short-radius atmospheric spectrometer [striped Si(620) crystal, R = 250 mm]. The spectra demonstrate comparable signal-to-noise ratios and resolutions well suited for VtC XES experiments. Both spectra have the same incoming photon flux of ∼1013 photons s−1 and total measurement time t = 15 min; but in the case of the Johann spectrometer, the detector collection time is smaller due to the necessity to scan the analyzer crystal. An additional bonus of the PINK von Hamos spectrometer is a detected energy range of ∼350 eV, thus both the Kβ2,5 and the Kβ1,3 main lines were recorded in the same time. During a VtC spectra analysis, the main line can be useful for normalization procedures.

[Figure 14]
Figure 14
Comparison of VtC XES of Co3O4 powder samples measured at the Diamond I20 beamline (blue) and PINK setup (red); the measurement time was 15 min.

5. Conclusions

We have described the PINK beamline and its specialized endstation optimized for non-resonant XES studies within the tender X-ray energy range. Our primary focus is investigating VtC photon-hungry emission lines. Photon flux ranges from 1013 photons s−1 to 1014 photons s−1 covering the energy range 2.1 keV to 9.5 keV. This range encompasses the Kα and Kβ lines of elements from P to Cu, as well as the Lα and Lβ lines of second row transition metals including Mo, Ru, Rh, Pd etc.

The endstation is equipped with two von Hamos dispersive spectrometers, finely tuned for the energy ranges 2 keV to 6 keV and 6 keV to 9.5 keV. Users have access to two sample environment chambers: the LiquidSEC chamber for experiments with solid and liquid samples at RT, including the option for electrochemical experiments using an electrochemical cell under vacuum or an He atmosphere; the CryoSEC chamber is designed for frozen samples and operates at 30 K, accommodating up to seven samples with a fast sample-exchange system.

By combining high flux, quick sample scanning, cryogenic temperatures and the capabilities of a dispersive von Hamos spectrometer to record the full spectrum in a single shot, we can obtain radiation-damage-free VtC spectra for sensitive and dilute complexes. We have provided case studies for P, Ru, Ti and Co complexes to showcase the beamline capabilities. The PINK beamline is fully operational and open to external users. In future work, the current setup will be expanded to include resonant XES and VtC-enhanced XAS techniques.

6. Data availability

The data used for the figure production are available at https://doi.org/10.17617/3.4ZB7ZR.

Footnotes

Presently at Experiments Division, ALBA Synchrotron Light Source, Carrer de la Llum 2–26, Cerdanyola del Vallès, 08290 Barcelona, Spain.

Acknowledgements

The authors thank Konstantin Klementiev (MAXIV) and Alexandr Kalinko (DESY) for insightful discussions. We are indebted to the BESSY II staff and especially the Sample Environment Group for their help in installation and commissioning of the PINK beamline. Natalia Levin Rojas took part in preparation, measurements and data analysis of the Ru complexes. The authors thank Olaf Rüdiger and Olivia McCubbin Stepanic for providing samples and expertise. Some of the KCl XES spectra were measured using LabXES at the MPI CEC by F. Strunk. We thank Diamond Light Source for beam time (proposal No. SP21526). Open access funding enabled and organized by Projekt DEAL.

Funding information

The Max Planck Society is acknowledged for funding.

References

First citationAbraham, B., Nowak, S., Weninger, C., Armenta, R., Defever, J., Day, D., Carini, G., Nakahara, K., Gallo, A., Nelson, S., Nordlund, D., Kroll, T., Hunter, M. S., van Driel, T., Zhu, D., Weng, T.-C., Alonso-Mori, R. & Sokaras, D. (2019). J. Synchrotron Rad. 26, 629–634.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationAbramson, J. E., Holden, W. M., Rivera-Maldonado, R. A., Velian, A., Cossairt, B. M. & Seidler, G. T. (2023). J. Anal. At. Spectrom. 38, 1125–1134.  CrossRef CAS Google Scholar
First citationAlonso-Mori, R., Kern, J., Sokaras, D., Weng, T.-C., Nordlund, D., Tran, R., Montanez, P., Delor, J., Yachandra, V. K., Yano, J. & Bergmann, U. (2012). Rev. Sci. Instrum. 83, 073114.  Web of Science PubMed Google Scholar
First citationBauer, M. (2014). Phys. Chem. Chem. Phys. 16, 13827–13837.  Web of Science CrossRef CAS PubMed Google Scholar
First citationBest, P. E. (1971). Phys. Rev. B, 3, 4377–4382.  CrossRef Google Scholar
First citationBłachucki, W., Czapla-Masztafiak, J., Sá, J. & Szlachetko, J. (2019). J. Anal. At. Spectrom. 34, 1409–1415.  Google Scholar
First citationButorin, S. M., Kvashnina, K. O., Klintenberg, M., Kavčič, M., Žitnik, M., Bučar, K., Gougeon, P., Gall, P., Candolfi, C. & Lenoir, B. (2018). ACS Appl. Energy Mater. 1, 4032–4039.  Web of Science CrossRef CAS Google Scholar
First citationCanton, S. E., Biednov, M., Pápai, M., Lima, F. A., Choi, T., Otte, F., Jiang, Y., Frankenberger, P., Knoll, M., Zalden, P., Gawelda, W., Rahaman, A., Møller, K. B., Milne, C., Gosztola, D. J., Zheng, K., Retegan, M. & Khakhulin, D. (2023). Adv. Sci. 10, 2206880.  CrossRef Google Scholar
First citationCastillo, R. G., Banerjee, R., Allpress, C. J., Rohde, G. T., Bill, E., Que, L., Lipscomb, J. D. & DeBeer, S. (2017). J. Am. Chem. Soc. 139, 18024–18033.  CrossRef CAS PubMed Google Scholar
First citationCastillo, R. G., Hahn, A. W., Van Kuiken, B. E., Henthorn, J. T., McGale, J. & DeBeer, S. (2021). Angew. Chem. Int. Ed. 60, 10112–10121.  Web of Science CSD CrossRef CAS Google Scholar
First citationCastillo, R. G., Henthorn, J. T., McGale, J., Maganas, D. & DeBeer, S. (2020). Angew. Chem. Int. Ed. 59, 12965–12975.  CrossRef CAS Google Scholar
First citationCutsail, G. E. III, Banerjee, R., Zhou, A., Que, L., Lipscomb, J. D. & DeBeer, S. (2018). J. Am. Chem. Soc. 140, 16807–16820.  CrossRef CAS PubMed Google Scholar
First citationCutsail, G. E. III & DeBeer, S. (2022). ACS Catal. 12, 5864–5886.  Google Scholar
First citationCutsail, G. E. III, Gagnon, N. L., Spaeth, A. D., Tolman, W. B. & DeBeer, S. (2019). Angew. Chem. Int. Ed. 58, 9114–9119.  CrossRef CAS Google Scholar
First citationDeslattes, R. D. (1964). Phys. Rev. 133, A390–A398.  CrossRef Google Scholar
First citationFollath, R., Hävecker, M., Reichardt, G., Lips, K., Bahrdt, J., Schäfers, F. & Schmid, P. (2013). J. Phys. Conf. Ser. 425, 212003.  CrossRef Google Scholar
First citationGeoghegan, B. L., Liu, Y., Peredkov, S., Dechert, S., Meyer, F., DeBeer, S. & Cutsail, G. E. (2022). J. Am. Chem. Soc. 144, 2520–2534.  CSD CrossRef CAS PubMed Google Scholar
First citationGerz, I., Jannuzzi, S. A. V., Hylland, K. T., Negri, C., Wragg, D. S., Øien–Ødegaard, S., Tilset, M., Olsbye, U., DeBeer, S. & Amedjkouh, M. (2021). Eur. J. Inorg. Chem. 2021, 4762–4775.  CSD CrossRef CAS PubMed Google Scholar
First citationGlatzel, P., Weng, T.-C., Kvashnina, K., Swarbrick, J., Sikora, M., Gallo, E., Smolentsev, N. & Mori, R. A. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 17–25.  Web of Science CrossRef CAS Google Scholar
First citationGroot, F. de (2001). Chem. Rev. 101, 1779–1808.  Web of Science CrossRef PubMed Google Scholar
First citationHall, E. R., Pollock, C. J., Bendix, J., Collins, T. J., Glatzel, P. & DeBeer, S. (2014). J. Am. Chem. Soc. 136, 10076–10084.  Web of Science CrossRef CAS PubMed Google Scholar
First citationHämäläinen, K., Siddons, D. P., Hastings, J. B. & Berman, L. E. (1991). Phys. Rev. Lett. 67, 2850–2853.  CrossRef PubMed CAS Web of Science Google Scholar
First citationHámos, L. von (1933). Ann. Phys. 409, 716–724.  Google Scholar
First citationHendel, S., Schäfers, F., Hävecker, M., Reichardt, G., Scheer, M., Bahrdt, J. & Lips, K. (2016). AIP Conf. Proc. 1741, 030038  Google Scholar
First citationHolden, W. M., Hoidn, O. R., Ditter, A. S., Seidler, G. T., Kas, J., Stein, J. L., Cossairt, B. M., Kozimor, S. A., Guo, J., Ye, Y., Marcus, M. A. & Fakra, S. (2017). Rev. Sci. Instrum. 88, 073904.  Web of Science CrossRef PubMed Google Scholar
First citationHolden, W. M., Seidler, G. T. & Cheah, S. (2018). J. Phys. Chem. A, 122, 5153–5161.  Web of Science CrossRef CAS PubMed Google Scholar
First citationHölzer, G., Fritsch, M., Deutsch, M., Härtwig, J. & Förster, E. (1997). Phys. Rev. A, 56, 4554–4568.  CrossRef CAS Web of Science Google Scholar
First citationHoszowska, J., Dousse, J.-C., Kern, J. & Rhême, C. (1996). Nucl. Instrum. Methods Phys. Res. A, 376, 129–138.  CrossRef CAS Web of Science Google Scholar
First citationHou, K., Börgel, J., Jiang, H. Z. H., SantaLucia, D. J., Kwon, H., Zhuang, H., Chakarawet, K., Rohde, R. C., Taylor, J. W., Dun, C., Paley, M. V., Turkiewicz, A. B., Park, J. G., Mao, H., Zhu, Z., Alp, E. E., Zhao, J., Hu, M. Y., Lavina, B., Peredkov, S., Lv, X., Oktawiec, J., Meihaus, K. R., Pantazis, D. A., Vandone, M., Colombo, V., Bill, E., Urban, J. J., Britt, R. D., Grandjean, F., Long, G. J., DeBeer, S., Neese, F., Reimer, J. A. & Long, J. R. (2023). Science, 382, 547–553.  CSD CrossRef CAS PubMed Google Scholar
First citationHuebner, S., Miyakawa, N., Kapser, S., Pahlke, A. & Kreupl, F. (2015). IEEE Trans. Nucl. Sci. 62, 588–593.  CrossRef CAS Google Scholar
First citationKalinko, A., Caliebe, W. A., Schoch, R. & Bauer, M. (2020). J. Synchrotron Rad. 27, 31–36.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKlementiev, K. & Chernikov, R. (2014). Proc. SPIE, 9209, 92090A.  Google Scholar
First citationKowalska, J. K., Lima, F. A., Pollock, C. J., Rees, J. A. & DeBeer, S. (2016). Isr. J. Chem. 56, 803–815.  Web of Science CrossRef CAS Google Scholar
First citationLancaster, K. M., Roemelt, M., Ettenhuber, P., Hu, Y., Ribbe, M. W., Neese, F., Bergmann, U. & DeBeer, S. (2011). Science, 334, 974–977.  Web of Science CrossRef CAS PubMed Google Scholar
First citationLevin, N., Peredkov, S., Weyhermüller, T., Rüdiger, O., Pereira, N. B., Grötzsch, D., Kalinko, A. & DeBeer, S. (2020). Inorg. Chem. 59, 8272–8283.  CrossRef CAS PubMed Google Scholar
First citationLima, F. A., Bjornsson, R., Weyhermüller, T., Chandrasekaran, P., Glatzel, P., Neese, F. & DeBeer, S. (2013). Phys. Chem. Chem. Phys. 15, 20911.  CrossRef PubMed Google Scholar
First citationLiu, Y., Chatterjee, S., Cutsail, G. E. III, Peredkov, S., Gupta, S. K., Dechert, S., DeBeer, S. & Meyer, F. (2023). J. Am. Chem. Soc. 145, 18477–18486.  CSD CrossRef CAS PubMed Google Scholar
First citationLiu, Z., Yuge, K. & Kawai, J. (2004). At. Spectrosc. 59, 93–99.  CrossRef Google Scholar
First citationMaganas, D., DeBeer, S. & Neese, F. (2017). Inorg. Chem. 56, 11819–11836.  CrossRef CAS PubMed Google Scholar
First citationMalzer, W., Grötzsch, D., Gnewkow, R., Schlesiger, C., Kowalewski, F., Van Kuiken, B., DeBeer, S. & Kanngießer, B. (2018). Rev. Sci. Instrum. 89, 113111.  Web of Science CrossRef PubMed Google Scholar
First citationMathe, Z., McCubbin Stepanic, O., Peredkov, S. & DeBeer, S. (2021). Chem. Sci. 12, 7888–7901.  CrossRef CAS PubMed Google Scholar
First citationMeents, A., Gutmann, S., Wagner, A. & Schulze-Briese, C. (2010). Proc. Natl Acad. Sci. USA, 107, 1094–1099.  Web of Science CrossRef CAS PubMed Google Scholar
First citationMiaja–Avila, L., O'Neil, G. C., Joe, Y. I., Morgan, K. M., Fowler, J. W., Doriese, W. B., Ganly, B., Lu, D., Ravel, B., Swetz, D. S. & Ullom, J. N. (2021). X-ray Spectrom. 50, 9–20.  CAS Google Scholar
First citationMori, R. A., Paris, E., Giuli, G., Eeckhout, S. G., Kavčič, M., Žitnik, M., Bučar, K., Pettersson, L. G. M. & Glatzel, P. (2010). Inorg. Chem. 49, 6468–6473.  Web of Science CrossRef CAS PubMed Google Scholar
First citationPetric, M., Bohinc, R., Bučar, K., Žitnik, M., Szlachetko, J. & Kavčič, M. (2015). Anal. Chem. 87, 5632–5639.  CrossRef CAS PubMed Google Scholar
First citationPetric, M. & Kavčič, M. (2016). J. Anal. At. Spectrom. 31, 450–457.  CrossRef CAS Google Scholar
First citationPollock, C. J. & DeBeer, S. (2015). Acc. Chem. Res. 48, 2967–2975.  Web of Science CrossRef CAS PubMed Google Scholar
First citationPollock, C. J., Grubel, K., Holland, P. L. & DeBeer, S. (2013). J. Am. Chem. Soc. 135, 11803–11808.  CrossRef CAS PubMed Google Scholar
First citationQureshi, M., Nowak, S. H., Vogt, L. I., Cotelesage, J. J. H., Dolgova, N. V., Sharifi, S., Kroll, T., Nordlund, D., Alonso-Mori, R., Weng, T.-C., Pickering, I. J., George, G. N. & Sokaras, D. (2021). Phys. Chem. Chem. Phys. 23, 4500–4508.  CrossRef CAS PubMed Google Scholar
First citationRehanek, J., Milne, C. J., Szlachetko, J., Czapla-Masztafiak, J., Schneider, J., Huthwelker, T., Borca, C. N., Wetter, R., Patthey, L. & Juranić, P. (2018). J. Synchrotron Rad. 25, 16–19.  Web of Science CrossRef IUCr Journals Google Scholar
First citationRovezzi, M., Harris, A., Detlefs, B., Bohdan, T., Svyazhin, A., Santambrogio, A., Degler, D., Baran, R., Reynier, B., Noguera Crespo, P., Heyman, C., Van Der Kleij, H.-P., Van Vaerenbergh, P., Marion, P., Vitoux, H., Lapras, C., Verbeni, R., Kocsis, M. M., Manceau, A. & Glatzel, P. (2020). J. Synchrotron Rad. 27, 813–826.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSchooneveld, M. M. van & DeBeer, S. (2015). J. Electron Spectrosc. Relat. Phenom. 198, 31–56.  Google Scholar
First citationSeidler, G. T., Mortensen, D. R., Remesnik, A. J., Pacold, J. I., Ball, N. A., Barry, N., Styczinski, M. & Hoidn, O. R. (2014). Rev. Sci. Instrum. 85, 113906.  Web of Science CrossRef PubMed Google Scholar
First citationSwarbrick, J. C., Kvashnin, Y., Schulte, K., Seenivasan, K., Lamberti, C. & Glatzel, P. (2010). Inorg. Chem. 49, 8323–8332.  Web of Science CrossRef CAS PubMed Google Scholar
First citationSzlachetko, J., Nachtegaal, M., de Boni, E., Willimann, M., Safonova, O., Sa, J., Smolentsev, G., Szlachetko, M., van Bokhoven, J. A., Dousse, J.-C., Hoszowska, J., Kayser, Y., Jagodzinski, P., Bergamaschi, A., Schmitt, B., David, C. & Lücke, A. (2012). Rev. Sci. Instrum. 83, 103105.  Web of Science CrossRef PubMed Google Scholar
First citationSzlachetko, J., Nachtegaal, M., Grolimund, D., Knopp, G., Peredkov, S., Czapla–Masztafiak, J. & Milne, C. (2017). Appl. Sci. 7, 899.  Web of Science CrossRef Google Scholar
First citationSzlachetko, J., Sá, J., Safonova, O., Smolentsev, G., Szlachetko, M., van Bokhoven, J. & Nachtegaal, M. (2013). J. Electron Spectrosc. Relat. Phenom. 188, 161–165.  CrossRef CAS Google Scholar
First citationThompson, A. & Vaughan, D. (2009). X-ray Data Booklet. Center for X-ray Optics and Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, USA.  Google Scholar
First citationTono, K., Kudo, T., Yabashi, M., Tachibana, T., Feng, Y., Fritz, D., Hastings, J. & Ishikawa, T. (2011). Rev. Sci. Instrum. 82, 023108.  Web of Science CrossRef PubMed Google Scholar
First citationVane, C. R., Smith, M. S. & Raman, S. (1988). A Versatile, High-Efficient, High-Resolution von Hamos Bragg Crystal X-ray Spectrometer. Oak Rodge National Laboratory, Oak Ridge, Tennessee, USA.  Google Scholar
First citationZimmermann, P., Peredkov, S., Abdala, P. M., DeBeer, S., Tromp, M., Müller, C. & van Bokhoven, J. A. (2020). Coord. Chem. Rev. 423, 213466.  Web of Science CrossRef Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.