Typical vertical intensity histograms on the detector for different values of the vertical position of the knife. Histograms are plotted with a vertical shift, and the red vertical arrows show the knife-edge position from before intersecting the beam (the lowest histogram) to fully blocking the beam (the highest histogram). The knife-edge relative position to the focus along the beam and mirror pitch offset are, respectively, +1.0 mm (downstream) and 0 µrad (a) and −3.6 mm (upstream) and −50 µrad (b). The intensity drops remain in a vertical line for ideal focusing.